K243642 · Bard Peripheral Vascular, Inc. · NEU · Mar 24, 2025 · General, Plastic Surgery
Device Facts
Record ID
K243642
Device Name
UltraCor™ Twirl™ Breast Tissue Marker
Applicant
Bard Peripheral Vascular, Inc.
Product Code
NEU · General, Plastic Surgery
Decision Date
Mar 24, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4300
Device Class
Class 2
Indications for Use
The UltraCor™ Twirl™ Breast Tissue Marker is intended for use to attach to soft breast tissue. including axillary lymph nodes, to radiographically mark the location of the biopsy procedure.
Device Story
Device is an implantable radiographic marker for long-term localization of breast biopsy sites. Consists of a disposable 17g x 10cm beveled needle applicator with depth markings and a locking plunger. Contains a Nitinol wireform marker in 'Curls' or 'Clover' shapes. Operated by clinicians to deploy the marker into the biopsy site. Marker remains in tissue to provide radiographic visibility for follow-up imaging (X-ray, mammography, ultrasound, MRI). Facilitates accurate identification of biopsy locations for subsequent clinical management.
Clinical Evidence
Bench testing only. Performance evaluated via marker differentiation, visibility (ultrasound, stereotactic, X-ray/mammography, MRI), retention, deployment accuracy, deployment force, corrosion, nickel ion release, and transformation temperature. Comprehensive biocompatibility testing (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, genotoxicity, and implantation studies) performed on both wireform and applicator.
Technological Characteristics
Implantable radiographic marker made of Nitinol. Applicator is a 17g x 10cm beveled needle with locking plunger. Marker shapes: Curls and Clover. Non-active, mechanical deployment. Biocompatibility verified per ISO/FDA standards. Sterilization method and shelf-life consistent with predicate.
Indications for Use
Indicated for use in patients requiring radiographic marking of biopsy sites in soft breast tissue, including axillary lymph nodes.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
UltraCor™ Twirl™ Breast Tissue Marker (Ring shape; K180061)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 24, 2025
Bard Peripheral Vascular, Inc. Meghan Mckelvey Regulatory Affairs Manager 1625 West 3rd Street Tempe, Arizona 85281
Re: K243642
Trade/Device Name: UltraCor™ Twirl™ Breast Tissue Marker Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: NEU Dated: November 25, 2024 Received: November 26, 2024
Dear Meghan Mckelvey:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# TEK N. LAMICHHANE -S
Tek N. Lamichhane, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K243642
Device Name
The UltraCor™ Twirl™ Breast Tissue Marker (Curls and Clover)
Indications for Use (Describe)
The UltraCor™ Twirl™ Breast Tissue Marker is intended for use to attach to soft breast tissue. including axillary lymph nodes, to radiographically mark the location of the biopsy procedure.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary 21 CFR 878.4300
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a summary of the information upon which substantial equivalence determination is based is as follows:
#### 1. Submitter Information:
| Applicant: | Bard Peripheral Vascular, Inc.<br>1625 West 3rd Street<br>Tempe, Arizona 85281 |
|------------|--------------------------------------------------------------------------------|
| Phone: | 602-830-5063 |
| Contact: | Meghan McKelvey, Regulatory Affairs Manager |
| Date: | November 26, 2024 |
#### 2. Subject Device-K243642:
| Device Trade Name: | UltraCor™ Twirl™ Breast Tissue Marker (Curls and Clover) |
|----------------------|----------------------------------------------------------|
| Classification Name: | Marker, Radiographic, Implantable (Product Code: NEU) |
| Review Panel: | General & Plastic Surgery |
| Regulation Number: | 21 CFR 878.4300 |
#### 3. Predicate Device:
UltraCor™ Twirl™ Breast Tissue Marker (Ring shape; K180061)
#### 4. Device Description:
The UltraCor™ Twirl™ Breast Tissue Marker (Curls and Clover) consists of a disposable beveled needle applicator containing a Nitinol radiographic marker is intended for long-term radiographic marking of the tissue site. The applicator has a beveled 17g x 10cm needle with 1 cm depth marks and a locking plunger. Each marker shape is deployed from the beveled needle tip into the tissue site.
{5}------------------------------------------------
## 5. Indications for Use of Device:
The UltraCor™ Twirl™ Breast Tissue Marker is intended for use to attach to soft breast tissue, including axillary lymph nodes, to radiographically mark the location of the biopsy procedure.
## 6. Technological Comparison to Predicate Devices:
The technological characteristics of the subject device are substantially equivalent to those of the predicate device, in terms of following:
- . Intended Use
- . Indications for Use
- . Performance Characteristics
- . Target Population
- . Fundamental Scientific Technology
- . Operating Principle (Mechanism of Action)
- Wireform (Marker) Material ●
- . Sterility Assurance Level and Method of Sterilization
- . Shelf-Life
- . Packaging Configuration
The subject device and the predicate device are different in the following manner:
- . Marker Shapes
- . Applicator Handle Overmold Colorants
# 7. Performance Testing Summary:
To demonstrate substantial equivalence of the subject device to the predicate device, the technological characteristics and performance criteria were evaluated. Using the FDA Guidance document, "Design Control Guidance for Medical Device Manufacturers," dated March 11, 1997, and internal risk assessments procedures, the following non-clinical tests were performed:
- . Marker Differentiation (Stereotactic or X-Ray/Mammography)
- . Marker Visibility (Ultrasound, Stereotactic, X-Ray/Mammography, MRI)
- . Marker Retention Testing
- . Marker Deployment Accuracy
- Marker Deployment Force ●
- Marker Deployment .
- . Corrosion Testing of Wireform (Marker)
{6}------------------------------------------------
- Nickel Ion Release Testing .
- Transformation Temperature Testing ●
- MRI Testing, and ●
- . Biocompatibility Testing was conducted on both the wireform (marker) and the applicator as presented in Table 1.
## Table 1: Biocompatibility Testing of UltraCor Twirl Breast Tissue Marker (Curls and Clover shapes)
| Implantable Wireform (Marker) | |
|----------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| Test Method | Results |
| Chemical characterization | Pass |
| Exhaustive Extraction at 50°C for 72 hours<br>(in triplicate) | SVOC by GC/MS<br>VOC by GC/MS by Headspace<br>ICP/MS<br>NVOC by UPLC/MS |
| Cytotoxicity | |
| MEM Cell Cytotoxicity | Pass |
| Sensitization (Kligman Maximization) | Pass |
| Irritation / Intracutaneous Reactivity<br>(Intracutaneous Injection) | Pass |
| Acute Systemic Toxicity | Pass |
| Material Mediated Pyrogenicity | Pass |
| Subchronic Toxicity Study in Rats<br>(13 weeks) | Pass |
| Genotoxicity | |
| • AMES Assay | • Pass |
| • Mouse Lymphoma Assay | • Pass |
| Implantation | |
| • 1 week | • Pass |
| • 4 weeks | • Pass |
| • 12 weeks | • Pass |
| Toxicology<br>(Toxicological Risk Assessment) | Pass |
| Applicator | |
| Aqueous Physicochemical Testing | Pass<br>Extract - Purifed Water |
| Non-Aqueous Physicochemical Testing | Pass<br>Extract - Isopropyl Alcohol |
| Exaggerated Extraction | Pass<br>Extract - Purified Water<br>Extract - Isopropyl Alcohol<br>Extract - Cyclohexane |
| | |
| | Extract - MEM Elution |
| | Pass |
| Sensitization-Kligman Guinea Pig<br>Maximization | Extract - 0.9% Sodium Chloride |
| | Extract - Cottonseed oil |
| | Pass |
| Irritation or Intracutaneous Reactivity | Extract - 0.9% Sodium Chloride |
| | Extract - Cottonseed oil |
| | Pass |
| Acute Systemic Toxicity | Extract - 0.9% Sodium Chloride |
| | Extract - Sesame oil |
| | Pass |
| Material-Mediated Pyrogenicity | Extract - 0.9% Sodium Chloride |
| | Pass |
| Chemical Characterization | SVOC by GC/MS |
| | VOC by GC/MS by Headspace |
| | ICP/MS |
| | NVOC by UPLC/MS |
{7}------------------------------------------------
The biologic and chemical characterization test results demonstrate that the subject device is biocompatible for its intended use. The technological characteristics and performance criteria of the UltraCor™ Twirl™ Breast Tissue Marker (Curls and Clover) are comparable to the predicate device and that they perform as safely and as effectively as the legally marketed predicate device.
# 8. Conclusion:
The UltraCor™ Twirl™ Breast Tissue Marker (Curls and Clover) are substantially equivalent to the legally marketed predicate device, the UltraCor™ Twirl™ Breast Tissue Marker (Ring shape; K180061).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.