Brainlab Elements (7.0); Brainlab Elements Image Fusion (5.0); Brainlab Elements Image Fusion Angio (1.0); Brainlab Elements Contouring (5.0); Brainlab Elements Fibertracking (3.0); Brainlab Elements BOLD MRI Mapping (1.0)
K243633 · Brainlab AG · QIH · Jun 13, 2025 · Radiology
Device Facts
Record ID
K243633
Device Name
Brainlab Elements (7.0); Brainlab Elements Image Fusion (5.0); Brainlab Elements Image Fusion Angio (1.0); Brainlab Elements Contouring (5.0); Brainlab Elements Fibertracking (3.0); Brainlab Elements BOLD MRI Mapping (1.0)
Applicant
Brainlab AG
Product Code
QIH · Radiology
Decision Date
Jun 13, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K243633 · Jun 13, 2025
Brainlab Elements (7.0); Brainlab Elements Image Fusion (5.0); Brainlab Elements Image Fusion Angio (1.0); Brainlab Elements Contouring (5.0); Brainlab Elements Fibertracking (3.0); Brainlab Elements BOLD MRI Mapping (1.0)
Brainlab AG
Retrospective de-identified 3D CE-T1 MR images from 412 patients
Retrospective clinical image data from multiple sites in the US and Europe were used to validate the performance (Dice, Precision, Recall) of the AI Tumor Segmentation algorithm for cranial tumors.
Elements AI Tumor Segmentation Validation; Retrospective performance validation
Adult patients with one or multiple contrast-enhancing cranial tumors (metastases, meningiomas, gliomas, etc.) and control patients without tumors.; Sample Size: 412 patients (595 scans, 1878 annotations); Number of Sites: Multiple clinical sites in the US and Europe
Manual ground-truth segmentations (annotations)
Dice coefficient, Precision, Recall
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Cranial tumor segmentation
Machine learning algorithm powered by the Anatomical Patient Model
Dice >= 0.7, Recall >= 0.8, Precision >= 0.8
Mean Dice: 0.75, Mean Precision: 0.86, Mean Recall: 0.85
—
—
Validation study: 595 scans from 412 patients (1878 annotations) from multiple clinical sites in the US and Europe.
>1 (US radiologists and non US radiologists)
Indications for Use
Brainlab Elements Image Fusion is an application for the co-registration of image data within medical procedures by using rigid and deformable registration methods. It is intended to align anatomical structures between data sets. It is not intended for diagnostic purposes. Brainlab Elements Image Fusion is indicated for planning of cranial and extracranial surgical treatments and preplanning of cranial and extracranial radiotherapy treatments. Brainlab Elements Image Fusion Angio is a software application that is intended to be used for the co-registration of cerebrovascular image data. It is not intended for diagnostic purposes. Brainlab Elements Image Fusion Angio is indicated for planning of cranial surgical treatments and preplanning of cranial radiotherapy treatments. Brainlab Elements Fibertracking is an application for the processing and visualization of cranial white matter tracts based on Diffusion Weighted Imaging (DWI) data for use in treatment planning procedures. It is not intended for diagnostic purposes. Brainlab Elements Fibertracking is indicated for planning of cranial surgical treatments and preplanning of cranial radiotherapy treatments. Brainlab Elements Contouring provides an interface with tools and views to outline, refine, combine and manipulate structures in patient image data. It is not intended for diagnostic purposes. Brainlab Elements Contouring is indicated for planning of cranial and extracranial surgical treatments and preplanning of cranial and extracranial radiotherapy treatments. Brainlab Elements BOLD MRI Mapping provides tools to analyze blood oxygen level dependent data (BOLD MRI Data) to visualize the activation signal. It is not intended for diagnostic purposes. Brainlab Elements BOLD MRI Mapping is indicated for planning of cranial surgical treatments.
Device Story
Brainlab Elements (7.0) is a suite of software applications for medical image processing, co-registration, segmentation, and visualization. Inputs include 3D DICOM imaging (CT, MRI, PET, US, DWI, BOLD, DSA). Applications perform rigid/deformable image fusion, white matter tract visualization, and anatomical contouring. Contouring 5.0 utilizes a machine learning algorithm for semi-automatic tumor segmentation. Fibertracking 3.0 uses deterministic and probabilistic (CSD) algorithms for tractography. Used in hospital/clinical settings by neurosurgeons, ENT/CMF surgeons, and radiation oncologists. Outputs are 3D DICOM objects, activation maps, and fiber bundles used for surgical/radiotherapy planning. The device assists clinicians in visualizing anatomy and planning interventions; it does not provide diagnostic results.
Clinical Evidence
No clinical prospective/retrospective trials. Performance validation for AI Tumor Segmentation used 595 scans from 412 patients (adults, diverse demographics). Primary endpoints: Dice coefficient, precision, and recall. Results: Mean Dice 0.75, Precision 0.86, Recall 0.85 across all tumor types (metastases, meningiomas, gliomas, nerve tumors). Accuracy tests for Virtual iMRI and Fibertracking were performed using retrospective image datasets to validate functionality against state-of-the-art approaches.
Technological Characteristics
Software-based medical image management and processing system. Operates on Windows 10/Server 2016/2019/2022. Connectivity: Networked/Workstation-based. Algorithms: Atlas-based segmentation, rigid/deformable registration, deterministic/probabilistic (CSD) fibertracking, and ML-based tumor segmentation (CNN-based). Hardware requirements: Minimum 4-12 physical cores, 8-64GB RAM, DirectX 12 compatible graphics. No physical materials; purely software.
Indications for Use
Indicated for planning of cranial and extracranial surgical and radiotherapy treatments. Patient population includes adults requiring treatment planning for cranial or extracranial conditions; no specific demographic, regional, or cultural limitations.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 13, 2025
Brainlab AG
Sadwini Suresh
QM Consultant
Olof-Palme-Str.9
Munich, 81829
Germany
Re: K243633
Trade/Device Name: Brainlab Elements (7.0);
Brainlab Elements Image Fusion (5.0);
Brainlab Elements Image Fusion Angio (1.0);
Brainlab Elements Contouring (5.0);
Brainlab Elements Fibertracking (3.0);
Brainlab Elements BOLD MRI Mapping (1.0)
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical image management and processing system
Regulatory Class: Class II
Product Code: QIH, JAK, LLZ
Dated: May 27, 2025
Received: May 27, 2025
Dear Sadwini Suresh:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243633 - Sadwini Suresh
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K243633 - Sadwini Suresh
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Daniel M. Krainak, Ph.D.
Assistant Director
DHT8C: Division of Radiological Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
## Submission Number (if known)
K243633
## Device Name
Brainlab Elements (7.0);
Brainlab Elements Image Fusion (5.0);
Brainlab Elements Image Fusion Angio (1.0);
Brainlab Elements Contouring (5.0);
Brainlab Elements Fibertracking (3.0);
Brainlab Elements BOLD MRI Mapping (1.0)
## Indications for Use (Describe)
Brainlab Elements Image Fusion is an application for the co-registration of image data within medical procedures by using rigid and deformable registration methods. It is intended to align anatomical structures between data sets. It is not intended for diagnostic purposes.
Brainlab Elements Image Fusion is indicated for planning of cranial and extracranial surgical treatments and preplanning of cranial and extracranial radiotherapy treatments.
Brainlab Elements Image Fusion Angio is a software application that is intended to be used for the co-registration of cerebrovascular image data. It is not intended for diagnostic purposes.
Brainlab Elements Image Fusion Angio is indicated for planning of cranial surgical treatments and preplanning of cranial radiotherapy treatments.
Brainlab Elements Fibertracking is an application for the processing and visualization of cranial white matter tracts based on Diffusion Weighted Imaging (DWI) data for use in treatment planning procedures. It is not intended for diagnostic purposes.
Brainlab Elements Fibertracking is indicated for planning of cranial surgical treatments and preplanning of cranial radiotherapy treatments.
Brainlab Elements Contouring provides an interface with tools and views to outline, refine, combine and manipulate structures in patient image data. It is not intended for diagnostic purposes.
Brainlab Elements Contouring is indicated for planning of cranial and extracranial surgical treatments and preplanning of cranial and extracranial radiotherapy treatments.
Brainlab Elements BOLD MRI Mapping provides tools to analyze blood oxygen level dependent data (BOLD MRI Data) to visualize the activation signal. It is not intended for diagnostic purposes.
Brainlab Elements BOLD MRI Mapping is indicated for planning of cranial surgical treatments.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}
BRAINLAB
K243633
# 510(k) Summary
June 11, 2025
| General Information | |
| --- | --- |
| Manufacturer | Brainlab AG; Olof-Palme Str.9; 81829, Munich, Germany |
| Establishment Registration | 8043933 |
| Trade Name | Brainlab Elements (7.0);
Brainlab Elements Image Fusion (5.0);
Brainlab Elements Image Fusion Angio (1.0);
Brainlab Elements Contouring (5.0);
Brainlab Elements Fibertracking (3.0);
Brainlab Elements BOLD MRI Mapping (1.0) |
| Classification Name | Automated Radiological Image Processing Software |
| Product Code | QIH; JAK; LLZ |
| Regulation Number | 892.2050 |
| Regulatory Class | II |
| Panel | Radiology |
| Predicate Device | K223106; Brainlab Elements 6.0 |
| Reference Device | K212397; StealthStation S8 Cranial v2.0 |
| Contact Information | |
| Primary Contact | Alternate Contact |
| Sadwini Suresh
QM Consultant
Phone: +49 89 99 15 68 0
Email: regulatory.affairs@brainlab.com | Chiara Cunico
Senior Manager Regulatory Affairs
Phone: +49 89 99 15 68 0
Email: chiara.cunico@brainlab.com |
# 1. Indications for Use
Brainlab Elements Image Fusion is an application for the co-registration of image data within medical procedures by using rigid and deformable registration methods. It is intended to align anatomical structures between data sets. It is not intended for diagnostic purposes.
Brainlab Elements Image Fusion is indicated for planning of cranial and extracranial surgical treatments and preplanning of cranial and extracranial radiotherapy treatments.
Brainlab Elements Image Fusion Angio is a software application that is intended to be used for the co-registration of cerebrovascular image data. It is not intended for diagnostic purposes.
PAGE 1 OF 19
{5}
BRAINLAB
Brainlab Elements Image Fusion Angio is indicated for planning of cranial surgical treatments and preplanning of cranial radiotherapy treatments.
Brainlab Elements Fibertracking is an application for the processing and visualization of cranial white matter tracts based on Diffusion Weighted Imaging (DWI) data for use in treatment planning procedures. It is not intended for diagnostic purposes.
Brainlab Elements Fibertracking is indicated for planning of cranial surgical treatments and preplanning of cranial radiotherapy treatments.
Brainlab Elements Contouring provides an interface with tools and views to outline, refine, combine and manipulate structures in patient image data. It is not intended for diagnostic purposes.
Brainlab Elements Contouring is indicated for planning of cranial and extracranial surgical treatments and preplanning of cranial and extracranial radiotherapy treatments.
Brainlab Elements BOLD MRI Mapping provides tools to analyze blood oxygen level dependent data (BOLD MRI Data) to visualize the activation signal. It is not intended for diagnostic purposes.
Brainlab Elements BOLD MRI Mapping is indicated for planning of cranial surgical treatments.
## 2. Device Description
The Brainlab Elements are applications and background services for processing of medical images including functionalities such as data transfer, image co-registration, image segmentation, contouring and other image processing.
They consist of the following software applications:
1. Image Fusion 5.0
2. Image Fusion Angio 1.0
3. Contouring 5.0
4. BOLD MRI Mapping 1.0
5. Fibertracking 3.0
This device is a successor of the Predicate Device Brainlab Elements 6.0 (K223106).
Brainlab Elements Image Fusion is an application for the co-registration of image data within medical procedures by using rigid and deformable registration methods.
Brainlab Elements Image Fusion Angio is a software application that is intended to be used for the co-registration of cerebrovascular image data. It allows co-registration of 2D digital subtraction angiography images to 3D vascular images in order to combine flow and location information. In
PAGE 2 OF 19
{6}
BRAINLAB
particular, 2D DSA (digital subtraction angiography) sequences can be fused to MRA, CTA and 3D DSA sequences.
Brainlab Elements Contouring provides an interface with tools and views to outline, refine, combine and manipulate structures in patient image data. The output is saved as 3D DICOM segmentation object and can be used for further processing and treatment planning.
BOLD MRI Mapping provides methods to analyze task-based (block-design) functional magnet resonance images (fMRI). It provides a user interface with tools and views in order to visualize activation maps and generate 3D objects that can be used for further treatment planning.
Brainlab Elements Fibertracking is an application for the processing and visualization of information based upon Diffusion Weighted Imaging (DWI) data, i.e. to calculate and visualize cranial white matter tracts in selected regions of interest, which can be used for treatment planning procedures.
# 3. Substantial Equivalence
| Predicate Number | Name |
| --- | --- |
| Subject Device | Brainlab Elements, Brainlab Elements Contouring (5.0), Brainlab Elements Fibertracking (3.0), Brainlab Elements Image Fusion (5.0), Brainlab Elements Image Fusion Angio (1.0), Brainlab Elements BOLD MRI Mapping (1.0) |
| Predicate device (K223106) | Brainlab Elements, Brainlab Elements Contouring (4.5), Brainlab Elements Fibertracking (2.0), Brainlab Elements Image Fusion (4.5), Brainlab Elements Image Fusion Angio (1.0), Brainlab Elements BOLD MRI Mapping (1.0) |
| Reference device (K212397) | StealthStation S8 Cranial v2.0 |
# Image Fusion 5.0
| Topic/Feature | Primary Predicate Device (Brainlab Elements Image Fusion 4.5 (K223106)) | Subject Device (Brainlab Elements Image Fusion 5.0) |
| --- | --- | --- |
| Indications for use | Brainlab Elements Image Fusion is an application for the co-registration of image data within medical procedures by using rigid and deformable registration methods. It is | Brainlab Elements Image Fusion is an application for the co-registration of image data within medical procedures by using rigid and deformable registration methods. It is intended to align |
| | with the image information. It is used to identify and evaluate the image information. It is used to identify and evaluate the image information. It is used to identify and evaluate the image information. It is used to identify and evaluate the image information. It is used to identify and evaluate the image information. It is used to identify and evaluate the image information. It is used to identify and evaluate the image information. It is used to identify and evaluate the image information. It is used to identify and evaluate the image information. It is used to identify and evaluate the image information. It is used to identify and evaluate the image information. It is used to identify and evaluate the image information. It is used to identify and evaluate the image information. | |
{7}
BRAINLAB
| Topic/Feature | Primary Predicate Device (Brainlab Elements Image Fusion 4.5 (K223106)) | Subject Device (Brainlab Elements Image Fusion 5.0) |
| --- | --- | --- |
| | intended to align anatomical structures between data sets.
The device itself does not have clinical indications. | anatomical structures between data sets. It is not intended for diagnostic purposes.
Brainlab Elements Image Fusion is indicated for planning of cranial and extracranial surgical treatments and preplanning of cranial and extracranial radiotherapy treatments. |
| Operator profile | The intended users are medical professionals. Typical users are:
- Image Guided Surgery (IGS):
Neurosurgeons, Ear-Nose-Throat (ENT)
surgeons and Cranio-Maxillofacial (CMF)
surgeons including their assistants
Radiotherapy (RT): medical professionals who perform radiation treatment planning (medical physicists, radiation oncologists, dosimetrists, physicians, etc.) | The intended users are medical professionals. Typical users are:
- Image Guided Surgery (IGS): Surgeons trained in the areas of neurosurgery, spine and trauma surgery, ear-nose-throat (ENT) surgery and craniomaxillofacial (CMF) surgery including their assistants.
Radiotherapy (RT): medical professionals who perform radiation treatment planning (medical physicists, radiation oncologists, dosimetrists, physicians, etc.). |
| Patient population | In general, there are no demographic, regional or cultural limitations for patients. It is up to the user to decide if the system shall be used to assist a certain procedure. | There are no demographic, regional or cultural limitations for patients. |
| Conditions of use | The system shall be used in a hospital office environment or rooms appropriate for surgical interventions or radiotherapy planning. | The system shall be used in a hospital office environment or rooms appropriate for surgical interventions or radiotherapy planning. |
| Computer hardware requirements | Brainlab Elements can be used on hardware that fulfills the defined minimum requirements:
- Operating System: Windows 8.1 64bit
- Minimum 4 logical cores
- Minimum RAM: 6 GB
- Graphics: Direct X compatible
- Display Resolution: 1920 x 1080 (Full HD) | All platforms which fulfill the minimum requirements:
IGS Workstation:
- Operating system: Windows 10
- minimum 4 physical cores
- RAM: 8 GB
- Graphics: DirectX 12 compatible
- Display resolution: 1920 x 1080 (Full HD)
IGS Server / Virtual Machine
- Operating System: Windows Server 2016
- RAM 16GB
RT Workstation:
- Operating System: Windows 10
- minimum 6 physical cores
- RAM: 48GB
- Graphics card: DirectX 12
- Display resolution: 1920 x 1080 (Full HD)
RT Server / Virtual Machine
- Operating System: Windows Server 2016
- minimum 12 physical cores |
PAGE 4 OF 19
{8}
BRAINLAB
| Topic/Feature | Primary Predicate Device (Brainlab Elements Image Fusion 4.5 (K223106)) | Subject Device (Brainlab Elements Image Fusion 5.0) |
| --- | --- | --- |
| | | - RAM: 64GB (+16GB per additional user) |
| Operating system | - Windows Server 2012 R2 / 2016 / 2019
- Windows 8.1
Windows 10 | - Windows Server 2016
- Windows Server 2019
- Windows Server 2022
- Windows 10 |
| Input Devices | Touch and mouse/keyboard control Touch and mouse/keyboard control | Touch and mouse/keyboard control Touch and mouse/keyboard control |
| Image Data Sets Format | DICOM 3D imaging Modalities, e.g. CT/XT, MRI, NM/PET, OT, are supported for automatic fusion.
Other modalities are supported for manual fusion (alignment) | DICOM 3D imaging Modalities, e.g. CT/XT, MRI, NM/PET, OT, are supported for automatic fusion.
Other modalities are supported for manual fusion (alignment) |
| Rigid co-registration, rigid image fusion | Automatic rigid co-registration (rigid fusion) of 3D DICOM image data: CT (incl. low-dose CT), MR, OT (e.g. Perfusion), MR-DTI, MR-BOLD, NM/PET, US (ultrasound).
Manual rigid co-registration (rigid fusion) of other (unknown) 3D DICOM image data. | Automatic rigid co-registration (rigid fusion) of 3D DICOM image data: CT (incl. low-dose CT), MR, OT (e.g. Perfusion), MR-DWI, MR-BOLD, NM/PET, US (ultrasound).
Manual rigid co-registration (rigid fusion) of other (unknown) 3D DICOM image data. |
| Application View Layout and Views | GUI Technology: HTML
Single view layout:
- Axial view is displayed by default
- The user is able to switch to coronal and sagittal views
- A toolbar is available on the right side to access the data selection, fusion tools and verification tools
- In the header, the user is able to approve or reject the fusion result | GUI Technology: HTML
Single view layout:
- Axial view is displayed by default
- The user is able to switch to coronal and sagittal views
- A toolbar is available on the right side to access the data selection, fusion tools and verification tools
- In the header, the user is able to approve or reject the fusion result |
| Region of Interest | The co-registration can be restricted to a specific region of interest of the datasets by means of a box. | The co-registration can be restricted to a specific region of interest of the datasets by means of a box and an ellipsoid. |
| Rigid co-registration, rigid image fusion | Automatic rigid co-registration (rigid fusion) of 3D DICOM image data: CT (incl. low-dose CT), MR, OT (e.g. Perfusion), MR-DTI, MR-BOLD, NM/PET, US (ultrasound).
Manual rigid co-registration (rigid fusion) of other (unknown) 3D DICOM image data. | Automatic rigid co-registration (rigid fusion) of 3D DICOM image data: CT (incl. low-dose CT), MR, OT (e.g. Perfusion), MR-DWI, MR-BOLD, NM/PET, US (ultrasound).
Manual rigid co-registration (rigid fusion) of other (unknown) 3D DICOM image data. |
| Deformable Registration / Elastic Image Fusion: Distortion | Distortion Correction for cranial image data: CT-MR, MR-MR (incl. DTI) | Distortion Correction for cranial image data: CT-MR, MR-MR (incl. DTI) |
PAGE 5 OF 19
{9}
BRAINLAB
| Topic/Feature | Primary Predicate Device (Brainlab Elements Image Fusion 4.5 (K223106)) | Subject Device (Brainlab Elements Image Fusion 5.0) |
| --- | --- | --- |
| Correction
Cranial | | |
| Deformable Registration / Elastic Image Fusion: Curvature Correction Spine | Curvature Correction for spinal image data: CT-CT, CT-MR (incl. low-dose CT scanner) | Curvature Correction for spinal image data: CT-CT, CT-MR (incl. low-dose CT scanner) |
| Deformable Registration / Elastic Image Fusion: Virtual iMRI Cranial | Calculation of a brain shift (“virtual iMRI”) simulation, based on intra-operative CT or MR image data, offering a simulation of pre-operative planning data within an intra-operative situation.
The intra-operative resection cavity is grayed out in the simulation result.
With the acceptance of a Virtual iMRI simulation, the original variant of an image set and existing co-registrations are removed from the current data selection | Calculation of a brain shift (“virtual iMRI”) simulation, based on intra-operative CT, MR, or US image data, offering a simulation of pre-operative planning data within an intra-operative situation.
The intra-operative resection cavity is blacked out in the simulation result.
Automatic scrolling is activated as soon as a simulation result is available.
With the acceptance of a Virtual iMRI simulation, the original variant of an image set and existing co-registrations are kept in the current data selection |
| Verification Tools: Spy Glass | A spy glass superimposes the first image set in a frame over the second image set. | A spy glass superimposes the first image set in a frame over the second image set. |
| Verification Tools: Blending | Blending allows to view structures in two image sets at the same time by using complimentary colors to distinguish both image sets. | Blending allows to view structures in two image sets at the same time by using complimentary colors to distinguish both image sets. |
| Verification Tools: Spy Glass for Deformable Co-Registrations | For deformable registrations the spy glass feature is combined with a heatmap and a deformation grid, highlighting areas where the deformed image set has been (locally) modified and giving a global impression about the deformation | For deformable registrations the spy glass feature is combined with a heatmap and a deformation grid, highlighting areas where the deformed image set has been (locally) modified and giving a global impression about the deformation |
| Verification Tools: Blending for deformable co-registrations | For deformable registrations the blending feature is combined with a deformation grid, which gives a global impression about the deformation while its colored grid lines highlight areas where the deformed image set has been (locally) modified. | For deformable registrations the blending feature is combined with a deformation grid, which gives a global impression about the deformation while its colored grid lines highlight areas where the deformed image set has been (locally) modified.
For Virtual iMRI with Ultrasound, the weather radar, a specific type of blending with distinct colors, has been implemented. |
PAGE 6 OF 19
{10}
BRAINLAB
| Topic/Feature | Primary Predicate Device (Brainlab Elements Image Fusion 4.5 (K223106)) | Subject Device (Brainlab Elements Image Fusion 5.0) |
| --- | --- | --- |
| Contrast Clearance Analysis | Calculation of treatment response assessment maps for the analysis of contrast agent clearance. | Calculation of treatment response assessment maps for the analysis of contrast agent clearance |
Image Fusion Angio 1.0.2
| Topic/Feature | Primary Predicate Device (Brainlab Elements Image Angio 1.0.2 (K223106)) | Subject Device (Brainlab Elements Image Fusion Angio 1.0.2) |
| --- | --- | --- |
| Indications for use | Brainlab Elements Image Fusion Angio is a software application that is intended to be used for the co-registration of cerebrovascular image data.
The device itself does not have clinical indications. | Brainlab Elements Image Fusion Angio is a software application that is intended to be used for the co-registration of cerebrovascular image data. It is not intended for diagnostic purposes. Brainlab Elements Image Fusion Angio is indicated for planning of cranial surgical treatments and preplanning of cranial radiotherapy treatments. |
| Operator profile | The intended users are medical professionals. Typical users are:
- Image Guided Surgery (IGS):
Neurosurgeons, Ear-Nose-Throat (ENT)
surgeons and Cranio-Maxillofacial (CMF)
surgeons including their assistants
Radiotherapy (RT): medical professionals who perform radiation treatment planning (medical physicists, radiation oncologists, dosimetrists, physicians, etc.) | The intended users are medical professionals. Typical users are:
- Image Guided Surgery (IGS): Surgeons trained in the areas of neurosurgery including their assistants.
Radiotherapy (RT): medical professionals who perform radiation treatment planning (medical physicists, radiation oncologists, dosimetrists, physicians, etc.). |
| Computer hardware requirements | Brainlab Elements can be used on hardware that fulfills the defined minimum requirements:
- Operating System: Windows 8.1 64bit
- Minimum 4 logical cores
- Minimum RAM: 6 GB
- Graphics: Direct X compatible
Display Resolution: 1920 x 1080 (Full HD) | All platforms which fulfill the minimum requirements:
IGS Workstation:
- Operating system: Windows 10
- minimum 4 physical cores
- RAM: 8 GB
- Graphics: DirectX 12 compatible
- Display resolution: 1920 x 1080 (Full HD)
IGS Server / Virtual Machine
- Operating System: Windows Server 2016
- RAM 16GB
RT Workstation:
- Operating System: Windows 10
- minimum 6 physical cores |
PAGE 7 OF 19
{11}
BRAINLAB
| Topic/Feature | Primary Predicate Device (Brainlab Elements Image Angio 1.0.2 (K223106)) | Subject Device (Brainlab Elements Image Fusion Angio 1.0.2) |
| --- | --- | --- |
| | | - RAM: 48GB
- Graphics card: DirectX 12
- Display resolution: 1920 x 1080 (Full HD)
RT Server / Virtual Machine
- Operating System: Windows Server 2016
- minimum 12 physical cores
- RAM: 64GB (+16GB per additional user) |
Contouring 5.0
| Topic/Feature | Primary Predicate Device (Brainlab Elements Contouring 4.5 (K223106)) | Subject Device (Brainlab Elements Contouring 5.0) |
| --- | --- | --- |
| Indications for use | Brainlab Elements Contouring provides an interface with tools and views to outline, refine, combine and manipulate structures in patient image data. The generated 3D structures are not intended to create physical replicas used for diagnostic purposes.
The device itself does not have clinical indications | Brainlab Elements Contouring provides an interface with tools and views to outline, refine, combine and manipulate structures in patient image data. It is not intended for diagnostic purposes. Brainlab Elements Contouring is indicated for planning of cranial and extracranial surgical treatments and preplanning of cranial and extracranial radiotherapy treatments. |
| Patient population | In general, there are no demographic, regional or cultural limitations for patients. It is up to the user to decide if the system shall be used to assist a certain procedure. | There are no demographic, regional or cultural limitations for patients. |
| Conditions of use | The system shall be used in a hospital office environment or rooms appropriate for surgical interventions or radiotherapy planning. | The system shall be used in a hospital office environment or rooms appropriate for surgical interventions or radiotherapy planning. |
| Operator profile | The intended users are medical professionals. Typical users are:
- Image Guided Surgery (IGS): Neurosurgeons, Ear-Nose-Throat (ENT) surgeons and Cranio-Maxillofacial (CMF) surgeons including their assistants
Radiotherapy (RT): medical professionals who perform radiation treatment planning | The intended users are medical professionals. Typical users are:
- Image Guided Surgery (IGS): Surgeons trained in the areas of neurosurgery, spine and trauma surgery, ear-nose-throat (ENT) surgery and craniomaxillofacial (CMF) surgery including their assistants.
Radiotherapy (RT): medical professionals who perform radiation treatment planning (medical |
PAGE 8 OF 19
{12}
BRAINLAB
| Topic/ Feature | Primary Predicate Device (Brainlab Elements Contouring 4.5 (K223106)) | Subject Device (Brainlab Elements Contouring 5.0) |
| --- | --- | --- |
| | (medical physicists, radiation oncologists, dosimetrists, physicians, etc.) | physicists, radiation oncologists, dosimetrists, physicians, etc.). |
| Computer hardware requirements | Brainlab Elements can be used on hardware that fulfills the defined minimum requirements:
- Operating System: Windows 8.1 64bit
- Minimum 4 logical cores
- Minimum RAM: 6 GB
- Graphics: Direct X compatible
- Display Resolution: 1920 x 1080 (Full HD) | IGS Workstation:
- Operating system: Windows 10
- minimum 4 physical cores
- RAM: 8 GB
- Graphics: DirectX 12 compatible
- Display resolution: 1920 x 1080 (Full HD) IGS Server / Virtual Machine:
- Operating System: Windows Server 2016
- RAM 16GB
RT Workstation:
- Operating System: Windows 10
- minimum 6 physical cores
- RAM: 48GB
- Graphics card: DirectX 12
- Display resolution: 1920 x 1080 (Full HD) RT Server / Virtual Machine:
- Operating System: Windows Server 2016
- minimum 12 physical cores
- RAM: 64GB (+16GB per additional user)
If Anatomical Patient Model 1.1 (APM 1.1) shall be used, the following minimum requirements have to be fulfilled:
- CPU: 12 (virtual) cores
- RAM: 24 GB
If the Cranial Tumor Segmentation feature of Contouring shall be supported, the following GPU requirements have to be fulfilled:
- GPU vRAM: 8GB
- GPU driver: ≥ 472.5* (*cuda 11.8 tool kit compatibility)
- CUDA compute capability: > 5.2 |
| Operating system | - Windows Server 2012 R2 / 2016 / 2019
- Windows 8.1
- Windows 10 | - Windows Server 2016
- Windows Server 2019
- Windows Server 2022
- Windows 10 |
| Input Devices | Touch and mouse/keyboard control | Touch and mouse/keyboard control |
| Automatic Segmentation: Cranial Tumors | Not available | Cranial tumors are auto-segmented as 3D objects in image sets with supported modality |
PAGE 9 OF 19
{13}
BRAINLAB
| Topic/ Feature | Primary Predicate Device (Brainlab Elements Contouring 4.5 (K223106)) | Subject Device (Brainlab Elements Contouring 5.0) |
| --- | --- | --- |
| | | (MR-t1 contrast enhanced) by means of a machine learning algorithm.
Similar to other anatomical structures the user can review, adjust or reject the tumor object. |
| Automatic Segmentation: Anatomical objects | The application automatically creates segmentation objects by using an atlas-based automatic segmentation of the brain.
Workflow-relevant segmentation objects are created automatically or, if required by a workflow, on a single button click. | The application automatically creates segmentation objects by using an atlas-based automatic segmentation of the brain.
Workflow-relevant segmentation objects are created automatically or, if required by a workflow, on a single button click.
In addition auto-segmented objects can optionally also be retrieved via the Anatomical Patient Model 1.1 by means of a setting. |
| Landmark Support | The application can display annotations. Annotations cannot be changed or removed. | The application can be used to display, define and edit annotations and registration markers. |
| Visualization: Blending Mode | Not available | The application can be used to display two fused images within one view. A slider is available to blend from one image to another. |
| Abnormity Detection | If an anomaly is detected, the application automatically focuses the views on this anomaly.
Anomaly detection is either the result of the atlas-based automatic segmentation of the brain or, on platforms with appropriate GPUs, the result of a machine learning algorithm. The GPU requirements are listed in the UniversalAtlasPerformer Test Plan ML Abnormity Detection. | If an anomaly is detected, the application automatically focuses the views on this anomaly.
Anomaly detection is either the result of the atlas-based automatic segmentation of the brain or, on platforms with appropriate GPUs, the result of a machine learning algorithm. The GPU requirements are listed in the UniversalAtlasPerformer Test Plan ML Abnormity Detection. |
| Automatic Segmentation: Anatomical objects – Segmentation Templates | The application also offers the possibility to define customized segmentation templates. | The application also offers the possibility to define customized segmentation templates. |
| Automatic Segmentation – Resection Cavity | Not available | A resection cavity object can be segmented automatically based on ultrasound image data.
Similar to other anatomical or tumor objects, the user can review, adjust or reject the resection cavity object. |
PAGE 10 OF 19
{14}
BRAINLAB
| Topic/ Feature | Primary Predicate Device (Brainlab Elements Contouring 4.5 (K223106)) | Subject Device (Brainlab Elements Contouring 5.0) |
| --- | --- | --- |
| Manual Creation and Refinement of Objects | The application provides tools for manual creation and refinement of segmentation objects:
- SmartShaper
- Brush 3D / Erase 3D
- Brush 2D / Erase 2D
- Smart Brush | The application provides tools for manual creation and refinement of segmentation objects:
- SmartShaper
- Brush 3D / Erase 3D
- Brush 2D / Erase 2D
- Smart Brush |
| Multi-Modal Drawing | SmartBrush can be used for multi-modal drawing, i.e. two images are used for drawing in a multi-modal view layout. | SmartBrush can be used for multi-modal drawing, i.e. two images are used for drawing in the newly introduced blending view when Blending Mode is active. |
| Manual Creation of Objects: Threshold Segmentation | The application provides a Threshold Segmentation tool to create objects. Threshold Segmentation supports MR, CT, PET, US and Contrast Clearance Analysis (CCA) data.
The size of a geometric ROI during Threshold Segmentation can be adjusted. | The application provides a Threshold Segmentation tool to create objects. Threshold Segmentation supports MR, CT, PET, US, Contrast Clearance Analysis (CCA) and RGB data.
The size of a geometric ROI during Threshold Segmentation can be adjusted and it can be rotated. |
| Object Manipulation | The application provides tools for the manipulation of objects:
- Copy
- Margins
- Smoothing
- Mirroring
- Shifting and Rotation
- Splitting
- Automatic Object Fitting | The application provides tools for the manipulation of objects:
- Copy
- Margins
- Smoothing
- Mirroring
- Shifting and Rotation
- Splitting
- Automatic Object Fitting |
| Logical Operations | The application provides tools for logical operations on objects:
- Union
- Subtract
- Intersect | The application provides tools for logical operations on objects:
- Union
- Subtract
- Intersect |
| Object Review | The application provides functionality to control the review state of objects. In case of leaving the application with unreviewed semi-automatically or automatically created objects, a message is displayed. | The application provides functionality to control the review state of objects. In case of leaving the application with unreviewed semi-automatically or automatically created objects, a message is displayed. |
| Volumetric Report | It is possible to create a printable report for an object in a platform independent file | It is possible to create a printable report for an object in a platform independent file format, giving information about the object's volume, its diameter according to the |
PAGE 11 OF 19
{15}
BRAINLAB
| Topic/ Feature | Primary Predicate Device (Brainlab Elements Contouring 4.5 (K223106)) | Subject Device (Brainlab Elements Contouring 5.0) |
| --- | --- | --- |
| | format, giving information about the object's volume, its diameter according to the Macdonald criteria and its response evaluation criteria (RECIST). If multiple objects are selected and a volumetric report is created, it contains information about all objects. | Macdonald criteria and its response evaluation criteria (RECIST). If multiple objects are selected and a volumetric report is created, it contains information about all objects. |
| Drawing Tools for the Spine SRS Use Case | The application calculates a clinical target volume (CTV) and a cropped organ at risk (OAR) object based on a manually outlined Gross Tumor (GTV). CTV and OAR objects can be created upon button click, based on a guideline published by Cox et al. (2012) when adding a GTV outline. In a user session, multiple combinations of GTV, CTV and OAR can be created. | The application calculates a clinical target volume (CTV) and a cropped organ at risk (OAR) object based on a manually outlined Gross Tumor (GTV). CTV and OAR objects can be created upon button click, based on a guideline published by Cox et al. (2012) when adding a GTV outline. In a user session, multiple combinations of GTV, CTV and OAR can be created. |
| Drawing Tools for the Angio Use Case | The application offers outlining features for cerebrovascular diseases. These features shall be supported by utilization of angiographic image data (i. e. 2D DSA) and projective fusions written by Image Fusion Angio. | The application offers outlining features for cerebrovascular diseases. These features shall be supported by utilization of angiographic image data (i. e. 2D DSA) and projective fusions written by Image Fusion Angio. |
# Fibertracking 3.0
| Topic/ Feature | Primary Predicate Device (Brainlab Elements Fibertracking 2.0 (K223106)) | Subject Device (Brainlab Elements Fibertracking 3.0) |
| --- | --- | --- |
| Indications for use | Brainlab Elements Fibertracking is an application for the processing and visualization of cranial white matter tracts based on Diffusion Tensor Imaging (DTI) data for use in treatment planning procedures. | Brainlab Elements Fibertracking is an application for the processing and visualization of cranial white matter tracts based on diffusion weighted imaging (DWI) data for use in treatment planning procedures. It is not intended for diagnostic purposes. Brainlab Elements Fibertracking |
PAGE 12 OF 19
{16}
BRAINLAB
| Topic/ Feature | Primary Predicate Device (Brainlab Elements Fibertracking 2.0 (K223106)) | Subject Device (Brainlab Elements Fibertracking 3.0) |
| --- | --- | --- |
| | The device itself does not have clinical indications. | is indicated for planning of cranial surgical treatments and preplanning of cranial radiotherapy treatments. |
| Operator profile | The intended users are medical professionals. Typical users are:
- Image Guided Surgery (IGS):
Neurosurgeons, Ear-Nose-Throat (ENT) surgeons and Cranio-Maxillofacial (CMF) surgeons including their assistants
Radiotherapy (RT): medical professionals who perform radiation treatment planning (medical physicists, radiation oncologists, dosimetrists, physicians, etc.) | The intended users are medical professionals. Typical users are:
- Image Guided Surgery (IGS): Surgeons trained in the areas of neurosurgery including their assistants.
Radiotherapy (RT): medical professionals who perform radiation treatment planning (medical physicists, radiation oncologists, dosimetrists, physicians, etc.). |
| Patient population | In general, there are no demographic, regional or cultural limitations for patients. It is up to the user to decide if the system shall be used to assist a certain procedure. | There are no demographic, regional or cultural limitations for patients. |
| Conditions of use | The system shall be used in a hospital office environment or rooms appropriate for surgical interventions or radiotherapy planning. | The system shall be used in a hospital office environment or rooms appropriate for surgical interventions or radiotherapy planning. |
| Computer hardware requirements | Brainlab Elements can be used on hardware that fulfills the defined minimum requirements:
- Operating System: Windows 8.1 64bit
- Minimum 4 logical cores
- Minimum RAM: 6 GB
- Graphics: Direct X compatible
- Display Resolution: 1920 x 1080 (Full HD) | IGS Workstation:
- Operating system: Windows 10
- minimum 4 physical cores
- RAM: 8 GB
- Graphics: DirectX 12 compatible
- Display resolution: 1920 x 1080 (Full HD)
IGS Server / Virtual Machine:
- Operating System: Windows Server 2016
- RAM 16GB
RT Workstation:
- Operating System: Windows 10
- minimum 6 physical cores
- RAM: 48GB
- Graphics card: DirectX 12
- Display resolution: 1920 x 1080 (Full HD)
RT Server / Virtual Machine:
- Operating System: Windows Server 2016
- minimum 12 physical cores
- RAM: 64GB (+16GB per additional user) |
| Operating system | - Windows Server 2012 R2 / 2016 / 2019
- Windows 8.1
- Windows 10 | - Windows Server 2016
- Windows Server 2019
- Windows Server 2022
- Windows 10 |
PAGE 13 OF 19
{17}
BRAINLAB
| Topic/ Feature | Primary Predicate Device (Brainlab Elements Fibertracking 2.0 (K223106)) | Subject Device (Brainlab Elements Fibertracking 3.0) |
| --- | --- | --- |
| Preprocessing | Automatic DTI Preprocessing (by using component DTI Preprocessing Performer 2.0):
☐ Eddy Current Correction,
☐ Motion Correction,
☐ B-Vector Realignment,
☐ Denoising,
☐ Regularization | Automatic DWI Preprocessing (by using component DWI Performer 3.0):
☐ Eddy Current Correction,
☐ Motion Correction,
☐ B-Vector Realignment,
☐ Denoising,
☐ Regularization
☐ Creation of ODF field for HARDI data |
| Define seed regions of interest | Application offers the following possibilities to define seed ROIs for Fibertracking (ROI based Fibertracking):
☐ Manual 3D Object (as include or exclude ROI)
☐ Existing 3D Object(s), e.g. by using ROI templates
☐ “Interactive Tracking” to use a dynamic ROI to generate fiber bundles on the fly | Application offers the following possibilities to define seed ROIs for Fibertracking (ROI based Fibertracking):
☐ Manual 3D Object (as include or exclude ROI)
☐ Existing 3D Object(s), e.g. by using ROI templates
☐ “Interactive Tracking” to use a dynamic ROI to generate fiber bundles on the fly |
| Algorithm configuration | Application offers three sliders to influence the algorithm's stopping criteria:
☐ Minimum FA Threshold (default: 0.2)
☐ Minimum Length (default: 80 mm)
☐ Maximum Angulation (default: 20°)
See also topic “Toolarea”. | Application offers sliders to influence the algorithm's stopping criteria:
Deterministic tracking using DTI:
☐ Refinement (remove outliers, default: 10%)
☐ Minimum FA Threshold (default: 0.2)
☐ Maximum Angulation (default: 20°)
Probabilistic tracking using CSD:
☐ Refinement (remove outliers, default: 10%)
☐ Amplitude (default: 0.1)
☐ Maximum Angulation (default: 20°) |
| Fibertracking algorithm | • Deterministic Tracking Algorithm based on Combined Single/Dual Tensor Tracking
Create 3D Objects from Fiber bundles | • Deterministic Tracking Algorithm based on Combined Single/Dual Tensor Tracking
• Probabilistic tracking algorithm
• Constrained Spherical Deconvolution tracking
Create 3D Objects from Fiber bundles
**Reference Device K212397**
**StealthStation S8 Cranial v2.0:**
StealthStation S8 Cranial v2.0 offers a |
PAGE 14 OF 19
{18}
BRAINLAB
| Topic/ Feature | Primary Predicate Device (Brainlab Elements Fibertracking 2.0 (K223106)) | Subject Device (Brainlab Elements Fibertracking 3.0) |
| --- | --- | --- |
| | | similar algorithm like the new probabilistic CSD tracking algorithm introduced with Fibertracking 3.0. |
| User Interface: Tool Area, description | ☐ Fiber bundles are accessible via toggle control and manageable via In-App DataSelection.
☐ ROIs are accessible in sidebar for drag&drop to fiber bundle.
☐ ROI-Brush accessible via toolarea.
☐ Other ROIs directly accessible via Sidebar.
☐ Parameter sliders area accessible in toolarea.
☐ “Refresh” accessible via toolarea.
☐ “Erase” accessible via toolarea
☐ “Objects Exclude” via Sidebar | ☐ Fiber bundles are accessible and manageable in sidebar control
☐ ROIs are accessible in sidebar for drag&drop to fiber bundle and selectable for editing.
☐ ROI-Brush accessible via toolarea.
☐ Other ROIs directly accessible via Sidebar.
☐ Parameter sliders area accessible in toolarea.
☐ “Refresh” accessible via sidebar.
☐ “Erase” accessible via toolarea.
☐ “Objects Exclude” via Sidebar
☐ Algorithm (DTI or CSD) can be selected in toolarea
☐ Measurement of coordinates and FA values available via toolarea |
BOLD MRI Mapping 1.0
| Topic/ Feature | Primary Predicate Device (Brainlab Elements BOLD MRI Mapping (K223106)) | Subject Device (Brainlab Elements BOLD MRI Mapping 1.0) |
| --- | --- | --- |
| Indications for use | Brainlab Elements BOLD MRI Mapping provides tools to analyze blood oxygen level dependent data (BOLD MRI Data) to visualize the activation signal and correlation between brain areas.
The device itself does not have clinical indications | Brainlab Elements BOLD MRI Mapping provides tools to analyze blood oxygen level dependent data (BOLD MRI Data) to visualize the activation signal. It is not intended for diagnostic purposes.
Brainlab Elements BOLD MRI Mapping is indicated for planning of cranial surgical treatments. |
| Operator profile | The intended users are medical professionals. Typical users are: | The intended users are medical professionals. Typical users are: |
PAGE 15 OF 19
{19}
BRAINLAB
| Topic/ Feature | Primary Predicate Device (Brainlab Elements BOLD MRI Mapping (K223106)) | Subject Device (Brainlab Elements BOLD MRI Mapping 1.0) |
| --- | --- | --- |
| | - Image Guided Surgery (IGS):
Neurosurgeons, Ear-Nose-Throat (ENT) surgeons and CranioMaxillofacial (CMF) surgeons including their assistants
Radiotherapy (RT): medical professionals who perform radiation treatment planning (medical physicists, radiation oncologists, dosimetrists, physicians, etc.) | Image Guided Surgery (IGS):
Surgeons trained in the area of neurosurgery including their assistants |
| Computer hardware requirements | Brainlab Elements can be used on hardware that fulfills the defined minimum requirements:
- Operating System: Windows 8.1 64bit
- Minimum 4 logical cores
- Minimum RAM: 6 GB
- Graphics: Direct X compatible
Display Resolution: 1920 x 1080 (Full HD) | IGS Workstation:
- Operating system: Windows 10
- minimum 4 physical cores
- RAM: 8 GB
- Graphics: DirectX 12 compatible
- Display resolution: 1920 x 1080 (Full HD)
IGS Server / Virtual Machine:
- Operating System: Windows Server 2016
- RAM 16GB |
| Operating system | - Windows Server 2012 R2 / 2016 / 2019
- Windows 8.1
Windows 10 | - Windows Server 2016
- Windows Server 2019
- Windows Server 2022
Windows 10 |
PAGE 16 OF 19
{20}
BRAINLAB
## 4. Performance Data
The Subject Device has been verified and validated according to Brainlab processes for product design and development. A high level explanation of the testing provided in this submission for the Subject Device is provided below.
### Software verification:
Software verification has been performed on software level, verifying the software requirements through integration tests as well as GUI tests and Unit tests. An incremental test strategy has been set up for the release candidates for changes with limited scope. In this case, an impact analysis of the modifications was performed and tests to be performed are identified and planned correspondingly.
### Usability Evaluation:
#### Image Fusion 5.0:
Formative and Summative Usability was carried out for Image Fusion 5.0. The main goal was the calculation of vMRI images that can support further neurosurgical procedures. Compared to Image Fusion 4.5, users can now additionally calculate a vMRI based on an intraoperative Ultrasound. The page layout and tools for verification of a fusion remain similar to predecessor versions.
#### Fibertracking 3.0:
The changes did not affect primary operating functions or any use related risks. No Summative Usability Evaluation was performed. Four rounds of formative evaluations of the changes were performed with a total of 23 users. Different rounds focused on various aspects such as general use of Fibertracking in a clinical workflow and experiences with the predecessor devices, new GUI components, review of the final concepts for the user interface.
### Accuracy Tests:
#### Image Fusion 5.0:
This test was to validate that Elements Virtual iMRI Cranial can be applied to cranial MR and intraoperative MR/CT/US image data related to image guided surgery approaches to compensate for surgery-related brain shift during resection. Tests were carried out using retrospective image data sets.
#### Fibertracking 3.0:
To validate if Fibertracking allows to visualize cranial white matter structures such as motoric, language and visual tracts based on state of the art approaches for Fibertracking as Constrained Spherical Deconvolution (CSD) and probabilistic tracking.
PAGE 17 OF 19
{21}
BRAINLAB
# AI/ML Performance Tests for Contouring 5.0:
To validate Elements AI Tumor Segmentation can be used to semi-automatically segment supported cranial tumors (metastases, meningiomas, cranial and paraspinal nerve tumors, gliomas, glioneuronal and neuronal tumors) on 3D medical Contrast Enhanced T1-weighted MR images powered by the Anatomical Patient Model. The approach used was quantitative validation - comparison to ground-truth annotations.
# Validation Summary:
Elements AI Tumor Segmentation algorithm was tested on de-identified 3D CE-T1 MR images from 412 patients (595 scans, 1878 annotations) from multiple clinical sites in the US and Europe. All data was acquired from adult patients with one or multiple contrast-enhancing tumors. The pool contained homogenous distribution by gender (female/male) and a diversity of ethnicity groups being White/Black/Latino/Asian with the highest representation. Most of data were obtained in patients who underwent stereotactic radiosurgery using a diverse acquisition of MR protocols (mainly containing 1.5T/3T MRI scans acquired in axial scan orientation). The test data pools also included control patients without cranial tumors. $\frac{1}{4}$ of the test pool corresponded to data from three independent sites in USA. The validation was conducted quantitatively by comparing the (manual) ground-truth segmentations, the so-called annotations with the respective automatically-created segmentations. The annotations involved external/independent annotator team including US radiologists and non US radiologists which followed a well-defined data curation process.
In accordance with best practices, measurement of recall and precision and object-specific DICE coefficient were performed to assess the similarity between both segmentations. Acceptance criteria were Dice $\geq 0.7$ , Recall $\geq 0.8$ and Precision $\geq 0.8$ for the lower bound of the respective $95\%$ confidence intervals. Successful validation has been completed based on images containing up to 30 cranial metastases, each showing a diameter of at least $3\mathrm{mm}$ , and images with primary cranial tumors that are at least $10\mathrm{mm}$ in diameter (for meningioma, cranial/paraspinal nerve tumors, gliomas, glioneuronal and neuronal tumors). See performance metrics in Table 1.
Table 1 Summary of test statistics
| Diagnostic Characteristics | Mean Dice | Mean Precision | Mean Recall |
| --- | --- | --- | --- |
| All | 0.75 | 0.86 | 0.85 |
| Metastases to the CNS | 0.74 | 0.85 | 0.84 |
{22}
BRAINLAB
| Meningiomas | 0.76 | 0.89 | 0.90 |
| --- | --- | --- | --- |
| Cranial and paraspinal nerve tumors | 0.89 | 0.97 | 0.97 |
| Gliomas and glio-/neuronal tumors | 0.81 | 0.95 | 0.85 |
# 5. Conclusion
The comparison of the Subject Device with the predicate device shows that Brainlab Elements 7.0 has similar functionality, intended use and technological characteristics as the predicate device. Based on the comparison to the predicate and the performance testing conducted, the Subject Device is considered substantially equivalent to the predicate device.
PAGE 19 OF 19
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.