K243620 · GE Medical Systems Ultrasound and Primary Care Diagnostics · IYN · Feb 11, 2025 · Radiology
Device Facts
Record ID
K243620
Device Name
Vivid iq
Applicant
GE Medical Systems Ultrasound and Primary Care Diagnostics
Product Code
IYN · Radiology
Decision Date
Feb 11, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
AI/ML, Pediatric
Indications for Use
The Vivid iq is high-performance compact diagnostic ultrasound system designed for Cardiovascular and Shared Services. It is intended for use by qualified and trained Healthcare professionals for Ultrasound imaging, measurement, display and analysis of the human body and fluid. Vivid iq clinical applications include: Fetal/Obstetrics, Abdominal (includes GYN), Pediatric, Small Organ (includes breast, testes, thyroid), Neonatal Cephalic, Adult Cephalic, Cardiac (includes Adult and Pediatric), Peripheral Vascular, Musculoskeletal Conventional. Musculoskeletal Superficial. Urology (Including prostate), Transcranial, Transrectal, Transvaginal, Transesophageal, Interventional Guidance (including Biopsy, Vascular Access), Thoracic/Pleural, Intraoperative(Vascular), Intracardiac and Intraluminal. Modes of operation include: B. M. PW Doppler, CW Doppler, Color Doppler, Color M, Power Doppler, Harmonic Imaging, Real-Time (RT) 3D Mode (4D), Coded Pulse and Combined modes: B/M, B/Color M, B/PWD, B/CWD, B/Color/PWD, B/Color/CWD, B/Power/PWD. The device is intended for use in an indoor hospital environment including echo lab, other hospital settings, operating room, Cath lab and EP lab, in private medical offices, and in limited settings outside of professional Healthcare facilities.
Device Story
Vivid iq is a compact, mobile diagnostic ultrasound system for cardiovascular and shared service imaging. It processes ultrasound signals from various electronic array transducers (linear, curved, phased, matrix, dual, CW, 3D, and ICE catheters) to generate real-time images and measurements. Operated by trained healthcare professionals in hospitals, clinics, or point-of-care settings, the system features a console with a touch panel and mobile cart with a charge box for portable use. It supports B, M, Doppler, and 4D modes. Output is displayed on-screen for clinical assessment, with options for DICOM storage, wireless connectivity, and video streaming to navigation systems (e.g., Carto 3). The device aids clinicians in diagnosis and interventional guidance, benefiting patients through high-quality tissue/blood flow visualization and efficient scanning workflows.
Clinical Evidence
No clinical studies were required to support substantial equivalence. The device relies on bench testing, including acoustic output, biocompatibility, cleaning/disinfection effectiveness, and thermal/electrical/electromagnetic safety testing in conformance with standards such as IEC 60601-1, IEC 60601-2-37, and ISO 10993-1.
Technological Characteristics
Compact ultrasound console with color LCD touch panel and mobile cart. Supports linear, curved, phased, matrix, dual, CW, 3D, and ICE transducers. Connectivity via LAN/wireless (USB adapter). Complies with AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-1, and NEMA PS 3.1-3.20 (DICOM). Features include Clarity+ (noise/speckle reduction) and Sleep mode for power management.
Indications for Use
Indicated for ultrasound imaging, measurement, and analysis of human body and fluid in patients requiring cardiovascular or shared service diagnostics. Applications include fetal/OB, abdominal, pediatric, small organ, neonatal/adult cephalic, cardiac, peripheral vascular, musculoskeletal, urology, transcranial, transrectal, transvaginal, transesophageal, interventional guidance, thoracic/pleural, intraoperative, intracardiac, and intraluminal imaging.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
February 11, 2025
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below.
GE Medical Systems Ultrasound and Primary Care Diagnostics, LLC Lee Bush Regulatory Affairs Director 3200 N Grandview Blvd WAUKESHA WI 53188
Re: K243620
Trade/Device Name: Vivid iq Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: November 22, 2024 Received: November 22, 2024
Dear Lee Bush:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
Page
3
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Marjan Nabili -S for
Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K243620
Device Name
Vivid ig
Indications for Use (Describe)
The Vivid ig is high-performance compact diagnostic ultrasound system designed for Cardiovascular and Shared Services. It is intended for use by qualified and trained Healthcare professionals for Ultrasound imaging, measurement, display and analysis of the human body and fluid.
Vivid in clinical applications include: Fetal/Obstetrics, Abdominal (includes GYN), Pediatric, Small Organ (includes breast, testes, thyroid), Neonatal Cephalic, Cardiac (includes Adult and Pediatric), Peripheral Vascular, Musculoskeletal Conventional, Musculoskeletal Superficial, Urology (Including prostate), Transcranial, Transvaginal, Transesophageal, Interventional Guidance (including Biopsy, Vascular access), Thoracic/Pleural, Intraoperative (Vascular), Intracardiac and Intraluminal.
Modes of operation include: B, M, PW Doppler, CW Doppler, Color Doppler, Color M, Power Doppler, Harmonic Imaging, Real-Time (RT) 3D Mode (4D), Coded Pulse and Combined modes: B/M, B/Color M, B/PWD, B/Color/PWD, B/Color/CWD, B/Power/PWD.
The device is intended for use in an indoor hospital environment including echo lab, other hospital settings, operating room, Cath lab and EP lab, in private medical offices, and in limited settings outside of professional Healthcare facilities.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for GE HealthCare. The logo consists of a purple circular emblem with a stylized "GE" monogram inside. To the right of the emblem is the text "GE HealthCare" in a simple, sans-serif font, also in purple. The logo is clean and modern, reflecting the company's focus on health and technology.
| K243620 | 510(k) Summary |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | In accordance with 21 CFR 807.92 the following summary of information is provided:<br>Nov 22, 2024 |
| Submitter: | GE Medical Systems Ultrasound and Primary care Diagnostics, LLC<br>3200 N Grandview Blvd<br>Waukesha, WI 53188, USA |
| Manufacturer: | GE Medical Systems (China) Co., Ltd.<br>No.19, ChangJiang Road, WuXi National Hi-Tech Dev. Zone,<br>214028 Jiangsu China |
| Primary Contact Person: | Lee Bush<br>Regulatory Affairs Director<br>GE HealthCare<br>T:(262) 309-9429 |
| Alternate Contact Person: | Yuan Wang<br>Senior Lead Specialist, Regulatory<br>Affairs<br>GE HealthCare<br>T: +86 13816996425 |
| Device Trade Name: | Vivid iq |
| Common/Usual Name: | Diagnostic Ultrasound System |
| Classification Names: | Class II |
| Product Code: | IYN (primary), IYO, ITX (secondary)<br>Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550, 90-IYN;<br>Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO;<br>Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX |
| Primary Predicate Device: | K221148 Vivid iq |
| Reference Device(s): | K232186 LOGIQ e<br>K220619 Vivid S70N/S60N<br>K211524 LOGIQ Fortis |
| Classification Names: | Class II |
| Product Code: | IYN (primary), IYO, ITX (secondary)<br>Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550, 90-IYN;<br>Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO;<br>Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the GE HealthCare logo. The logo consists of a circular emblem on the left and the text "GE HealthCare" on the right. The emblem is a stylized version of the letters "GE" inside of a circle. The text and emblem are purple.
Device Description: The proposed Vivid ig system is a general-purpose, Track 3, diagnostic ultrasound device, primarily intended for cardiovascular diagnostic use and shared service imaging. It is an ultrasound imaging & analysis system, consisting of a compact console with control panel including a track pad, color LCD Touch Panel that includes an on-screen alphanumeric keyboard. The system also has standard height-adjustable new ergonomic mobile cart for comfortable standing and sitting positions. The Charge Box in the Mobile Cart provides Vivid iq up to 4 hours scanning time without power supply.
> There are options for image storage, USB wireless connectivity, cardiac signal input for cardiac gating and output capabilities to printing devices. Vivid ig utilizes a variety of electronic array transducers operating in linear, curved, sector/phased array, matrix array, or dual array format, including dedicated CW transducers and real time 3D transducer. The system can also be used with compatible ICE transducers.
> The system includes electronics for transmit and receive of ultrasound data, ultrasound signal processing, software computing, hardware for image storage, hard copy printing, and network access to the facility through both LAN and wireless (supported by use of a wireless LAN USB-adapter) connection.
- Intended Use: The Vivid iq is high-performance compact diagnostic ultrasound system designed for Cardiovascular and Shared Services. It is intended for use by qualified and trained Healthcare professionals for Ultrasound imaging, measurement, display and analysis of the human body and fluid. Vivid iq clinical applications include: Fetal/Obstetrics, Abdominal (includes GYN), Pediatric, Small Organ (includes breast, testes, thyroid), Neonatal Cephalic, Adult Cephalic, Cardiac (includes Adult and Pediatric), Peripheral Vascular, Musculoskeletal Conventional. Musculoskeletal Superficial. Urology (Including prostate), Transcranial, Transrectal, Transvaginal, Transesophageal, Interventional Guidance (including Biopsy, Vascular Access), Thoracic/Pleural, Intraoperative(Vascular), Intracardiac and Intraluminal. Modes of operation include: B. M. PW Doppler, CW Doppler, Color Doppler, Color M, Power Doppler, Harmonic Imaging, Real-Time (RT) 3D Mode (4D), Coded Pulse and Combined modes: B/M, B/Color M, B/PWD, B/CWD, B/Color/PWD, B/Color/CWD, B/Power/PWD. The device is intended for use in an indoor hospital environment including echo lab, other hospital settings, operating room, Cath lab and EP lab, in private medical offices, and in limited settings outside of professional Healthcare facilities.
- The Vivid ig employs the same fundamental scientific technology as its Technology: predicate device(s).
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for GE HealthCare. The logo consists of the GE monogram enclosed in a circle on the left, followed by the words "GE HealthCare" in a sans-serif font. The color of the logo and text is a shade of purple.
Determination of Substantial Equivalence:
#### Comparison to Predicates
The proposed Vivid iq is substantially equivalent to the predicate devices. The following is an overview of the differences between the Vivid iq and the predicate Vivid iq (K221148).
# Indications for Use:
The proposed Vivid iq and predicate Vivid iq (K21148) have similar clinical indications for use, however the proposed device Vivid iq is adding Vascular Access which has been cleared on reference device Vivid S70N/S60N (K220619).
# Transducers:
The proposed Vivid ig and predicate Vivid iq (K221148) transducers are similar, except for:
- Addition of two OEM Ultrasound Catheters (AcuNav Crystal Ultrasound Catheter (K233270) from SIEMENS Medical Solutions USA. Inc and SOUNDSTARTM CRYSTAL Ultrasound Catheter (K240050) from Biosense Webster Inc.) to the compatible accessory list. The clinical indications and modes on the proposed Vivid iq are the same as on the predicate device Vivid iq (K221148)
Software Features/Functionality:
- Sleep mode: similar to Standby mode on cleared LOGIQ e ● (K232186). It helps reduce battery consumption and allows for longer time between battery charges, enabling up to 22% reduced daily power consumption and 22% extended portable echo usage.
- Clarity +: similar to Advanced SRI-HD on cleared LOGIQ Fortis ● (K211524), it is a real-time image processing/filtering technique that suppresses noise and graininess (reduces speckles) while sharpening boundaries and small structures.
- DICOM Encapsulated PDF report: allows PDF reports generated on . the system to be transferred through DICOM data flows and stored in DICOM storage, in accordance with the DICOM standard.
- Remote Viewing: same as on cleared Vivid S70N(K220619), enables . streaming of the main monitor over the local network. This feature is not for diagnostic use.
- Addition of Carto 3 Communication Options (K231207) ●
- Digital Video Streaming: same as cleared on Vivid -
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the GE HealthCare logo. The logo consists of a purple circular emblem with a stylized "GE" inside. To the right of the emblem, the words "GE HealthCare" are written in purple, with a sans-serif font.
S70N(K220619), digital video streaming of 2D and CF modes to OEM (Original Equipment Manufacturer) Carto 3 EP Navigation System V8.0 in addition to previously available analog/VGA output.
- . One-click QuickApps shortcuts: allows users to maintain high-quality tissue and blood flow visualization during transition between high flow / low flow and superficial / deep scanning with only one-click on the main screen.
# Accessories:
- Additional ICE Cord-RS CRYSTAL, to connect with AcuNav ● Crystal Ultrasound Catheter (K233270) and SOUNDSTAR CRYSTAL Ultrasound Catheter (K240050)
- Additional wireless adapter Netgear A8000 ●
- Additional Transend DVD writer ●
# Summary of Non-Clinical Tests:
The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic and mechanical safety, and has been found to conform with applicable medical device safety standards. The proposed Vivid iq complies with voluntary standards:
- AAMI/ANSI ES60601-1, Medical Electrical Equipment Part 1: ● General Requirements for Safety, 2005/A2:2021
- . AAMI TIR69:2017/(R2020) Technical Information Report Risk management of radio-frequency wireless coexistence for medical devices and systems
- IEC 60601-1-2 Medical Electrical Equipment Part 1-2: General . Requirements for Basic Safety and Essential Performance -Collateral Standard: Electromagnetic Disturbance - Requirements and Tests, Edition 4.1, 2020
- IEC 60601-2-37, Medical Electrical Equipment Part 2-37: . Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment, Edition 2.1, 2015
- ISO 10993-1, Biological Evaluation of Medical Devices-Part 1: . Evaluation and Testing Within Risk Management Process, Fifth edition, 2018
- . ISO 14971. Application of risk management to medical devices. 2019
- NEMA PS 3.1 3.20, Digital Imaging and Communications in ● Medicine (DICOM) Set. (Radiology), 2022d
- IEC 62359, Ultrasonics - Field characterization - Test methods for the determination of thermal and mechanical indices related to
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the GE HealthCare logo. The logo consists of a circular emblem with a stylized "GE" inside, followed by the text "GE HealthCare". The emblem and text are both in a purple color.
medical diagnostic ultrasonic fields, Edition 2.1, 2017
The following quality assurance measures are applied to the development of the system:
- Risk Analysis
- Requirements Reviews
- Design Reviews ●
- Testing on unit level (Module verification) ●
- Integration testing (System verification) ●
- Performance testing (Verification & Validation) ●
- Safety testing (Verification) ●
Transducer materials and other patient contact materials are biocompatible.
#### Summary of Clinical Tests:
The subject of this premarket submission, Vivid iq, did not require clinical studies to support substantial equivalence.
#### Conclusion:
Based on the equipment design similarities, conformance to recognized performance standards, and performance testing, GE HealthCare considers the proposed Vivid ig to be as safe, effective, and performs in a substantially equivalent manner as the predicate Vivid iq (K221148).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.