K243600 · Forus Health Pvt.Ltd · HKI · Aug 14, 2025 · Ophthalmic
Device Facts
Record ID
K243600
Device Name
3nethra neo HD FA; 3nethra neo HD
Applicant
Forus Health Pvt.Ltd
Product Code
HKI · Ophthalmic
Decision Date
Aug 14, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1120
Device Class
Class 2
Attributes
Pediatric
Indications for Use
3nethra neo HD FA: 3nethra neo HD FA is used as a wide-field retinal imaging digital camera for photo documentation of ocular diseases that manifest in infants. The device acquires only images and does not provide any pathological analysis or diagnosis for treatment. 3nethra neo HD: 3nethra neo HD is used as a wide-field retinal imaging digital camera for photo documentation of ocular diseases that manifest in infants. The device acquires only images and does not provide any pathological analysis or diagnosis for treatment.
Device Story
Handheld wide-angle fundus imaging system for neonatal screening; acquires, displays, stores, and transmits images of posterior and anterior eye surfaces. Operates in contact with cornea; uses continuous white light (requires pupil dilation). 3nethra neo HD FA model includes blue light source and software-controlled motorized green filters for fundus fluorescein angiography (FFA). Used by clinicians for documentation of retinopathy of prematurity (ROP) and other ocular conditions. Output consists of digital images (PNG/JPEG/DCM) for clinician review to assist in evaluation; device provides no automated diagnosis or analysis. Benefits include improved documentation of neonatal ocular health.
Clinical Evidence
Clinical observational study compared FA image quality between 3nethra neo HD FA and RetCam 3 reference device. Results demonstrated comparable image quality, supporting substantial equivalence in safety and effectiveness for neonatal fundus imaging.
Technological Characteristics
Handheld wide-field fundus camera; white/blue LED light sources; 150° field of view. Power: 100-240V AC. Connectivity: image export to PNG/JPEG/DCM. Software-controlled motorized blue filter for FA mode. Safety standards: IEC 60601-1, IEC 60601-1-2, ISO 15004-2, ANSI Z80.36. Patient-contact parts identical to predicate.
Indications for Use
Indicated for wide-field retinal imaging and photo documentation of ocular diseases in infants. Not for pathological analysis or diagnosis.
Regulatory Classification
Identification
An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.
Special Controls
*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 14, 2025
Forus Health Pvt.ltd.,
Rj Venkataramanan
Cmo & Ro
No. 8, 27th Cross Rd, Banashankari Stage II, Banashankari
Bengaluru, Bengaluru 560070
India
Re: K243600
Trade/Device Name: 3nethra neo HD FA; 3nethra neo HD
Regulation Number: 21 CFR 886.1120
Regulation Name: Ophthalmic Camera
Regulatory Class: Class II
Product Code: HKI
Dated: November 21, 2024
Received: November 21, 2024
Dear Rj Venkataramanan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243600 - Rj Venkataramanan
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K243600 - Rj Venkataramanan
Page 3
Sincerely,
Elvin Y. Ng -S
Elvin Ng
Assistant Director
DHT1A: Division of Ophthalmic Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243600
Device Name
3nethra neo HD FA;
3nethra neo HD
Indications for Use (Describe)
3nethra neo HD FA:
3nethra neo HD FA is used as a wide-field retinal imaging digital camera for photo documentation of ocular diseases that manifest in infants. The device acquires only images and does not provide any pathological analysis or diagnosis for treatment.
3nethra neo HD:
3nethra neo HD is used as a wide-field retinal imaging digital camera for photo documentation of ocular diseases that manifest in infants. The device acquires only images and does not provide any pathological analysis or diagnosis for treatment.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”
{4}
Forus Health
Original Premarket 510(k) Notification
510(k) Summary
# 510(k) SUMMARY
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92
# 1. General Information
| Applicant | Forus Health Pvt. Ltd |
| --- | --- |
| Applicant’s Address | No.8, 27^{th} Cross Rd, Banashankari Stage II, Banashankari, Bengaluru, Karnataka, INDIA 560070 |
| Phone Contact | +91 80 69439999 |
| FAX | +91 80 6943 9943 |
| Contact Person & Email | RJ Venkataramanan
quality@forushealth.com |
| Date Prepared | 14-08-2025 |
| Common Name | Ophthalmic Camera |
| Trade Name and Model | 3nethra neo HD FA
3nethra neo HD |
| Device Classification | 2 |
| Product Code | HKI |
| Classification Name | Camera, Ophthalmic, Ac-powered |
| Regulation Number | 886.1120 |
| Classification Panel | Ophthalmic |
| Predicate Device | 3nethra neo (K183059) |
Page 1 of 6
K243600
{5}
Forus Health
Original Premarket 510(k) Notification
510(k) Summary
## 2. Device Description
### 3nethra neo HD FA:
The 3nethra neo-HD FA is a handheld wide-angle fundus imaging system for neonatal screening. It is designed to acquire, display, store and transmit images of the posterior and anterior surfaces of human eye. The images assist clinicians in the evaluation and documentation of visual health in retinopathy of prematurity (ROP) and other problems. It operates in contact with the cornea of the eye under test. 3nethra neo HD FA uses continuous white light for operation and hence need pupil dilation. It also equipped with blue light source and green filters for fundus fluorescein angiography (FFA).
### 3nethra neo HD:
The 3nethra neo-HD is a handheld wide-angle fundus imaging system for neonatal screening. It is designed to acquire, display, store and transmit images of the posterior and anterior surfaces of human eye. The images assist clinicians in the evaluation and documentation of visual health in retinopathy of prematurity (ROP) and other problems. It operates in contact with the cornea of the eye under test. 3nethra neo HD uses continuous white light for operation and hence need pupil dilation. The 3nethra neo HD is a lower end variant of the 3nethra neo HD FA without the Fluorescein Angiography feature.
## 3. Indication for Use
### 3nethra neo HD FA:
3nethra neo HD FA is used as a wide-field retinal imaging digital camera for photo documentation of ocular diseases that manifest in infants. The device acquires only images and does not provide any pathological analysis or diagnosis for treatment.
### 3nethra neo HD:
3nethra neo HD is used as a wide-field retinal imaging digital camera for photo documentation of ocular diseases that manifest in infants. The device acquires only images and does not provide any pathological analysis or diagnosis for treatment.
K243600
{6}
Forus Health
Original Premarket 510(k) Notification
510(k) Summary
# 4. Substantial Equivalence Summary and Predicate Device
Based on our evaluation of substantial equivalence, The design of 3nethra neo HD FA/ 3nethra neo HD is substantially equivalent to the design of the predicate device. The methods of operation is based on well-established methods of use currently available in the market.
## Primary Predicate
Device Name: 3nethra neo (K183059)
Device Class: Class II
Product Code: HKI
Applicant: Forus Health Pvt.Ltd.
## Reference Device
Device Name: RetCam 3 Ophthalmic Imaging System (K102859)
Device Class: Class II
Product Code: HKI
Applicant: Clarity Medical Systems
## Technological Similarities
| Technological Characteristics | Subject Device | | Primary Predicate | Reference Device | Equivalence |
| --- | --- | --- | --- | --- | --- |
| Device Name | 3nethra neo HD FA | 3nethra neo HD | 3nethra neo (K183059) | Retcam 3 Ophthalmic Imaging System (K102859) | N/A |
| | Design | | | | |
| Feature | White and blue | White and blue | White light module | White light module | Equivalent to the Primary Predicate. |
| | Camera | | | | |
| Field of view | 150 degrees | 150 degrees | 120 degrees | 30 to 130 degrees | Different, but does not add any new concerns regarding the |
K243600
{7}
Forus Health
Original Premarket 510(k) Notification
510(k) Summary
| | | | | | safety and efficacy of the device |
| --- | --- | --- | --- | --- | --- |
| | Light Source | | | | |
| Light source type | White and blue LED | White and blue LED | White light LED | White light LED | Equivalent to the Primary Predicate. |
| Light intensity control | Zero to maximum | Zero to maximum | Zero to maximum | Zero to maximum | Equivalent to the Primary Predicate. |
| External fixation light (if any) | None | None | None | None | N/A |
| | Electrical | | | | |
| Power Supply | AC 100-240 V, 50/60 Hz | AC 100-240 V, 50/60 Hz | AC 100-240 V, 50/60 Hz | AC 100-240 V, 50/60 Hz | Equivalent to the Primary Predicate. |
| | Operating Conditions | | | | |
| Insert filter | Software controlled Motorized blue filter | Software controlled Motorized blue filter | No insert filter needed | Insert filter needed for FA mode | Equivalent to the Reference Device |
| Imaging Format | PNG/jpeg (after export)/.DCM( after export) | PNG/jpeg (after export)/.DCM( after export) | PNG/jpeg (after export)/.DCM(after export) | TIF/JPEG/.AVI/.B MP | Equivalent to the Primary Predicate. |
| | Safety | | | | |
| Safety Compliance | IEC 60601-1 IEC 60601-1-2 | IEC 60601-1 IEC 60601-1-2 | IEC 60601-1 IEC 60601-1-2 | IEC 60601-1 IEC 60601-1-2 | Equivalent to the Primary Predicate. |
## Technological Difference
| Technological Characteristics | Subject Device | | Primary Predicate | Reference Device | Equivalence |
| --- | --- | --- | --- | --- | --- |
| Device Name | 3nethra neo HD FA | 3nethra neo HD | 3nethra neo (K183059) | Retcam 3 Ophthalmic Imaging System (K102859) | N/A |
| | Design | | | | |
| Feature | Fluorescence Angiography module | No Fluorescence Angiography module | No Fluorescence Angiography module | Fluorescence Angiography module | Equivalent to the Reference device |
K243600
{8}
Forus Health
Original Premarket 510(k) Notification
510(k) Summary
These technological differences do not alter the intended use of the device, nor do they affect the safety and effectiveness of the device relative to the primary predicate device.
# 5. Bench Test Summary
- Electrical safety and electromagnetic compatibility- Electrical safety and EMC testing were conducted on the 3nethra neo HD FA /3nethra neo HD in accordance with established protocols. The device complies with IEC 60601-1 for electrical safety and IEC 60601 standard for EMC.
- Light safety testing- The 3nethra neo HD FA /3nethra neo HD evaluated for compliance to standard ISO 15004-2- Ophthalmic Instruments - Fundamental requirements and test methods and ANSI Z80.36. The evaluation demonstrates compliance with Group 1 instrument requirements provided by the standards.
- Animal Testing:
Biocompatibility testing- The biocompatibility evaluation for the 3nethra neo HD FA /3nethra neo HD is not required as the part that comes in contact with the patient is identical to the Primary Predicate device. There is no difference in any characteristics of the part and is completely identical with the Primary Predicate. The primary predicate is our own device hence the biocompatibility assessment done during the Primary predicate holds good for the Subject Device.
# 6. Clinical Performance Data
- Image Quality:
A clinical observational study was performed to compare the quality of fluorescein angiography (FA) images produced by the 3nethra neo HD FA subject device and the RetCam 3 reference device (cleared under K182263). The primary objective of the study was to demonstrate that FA images produced by the 3nethra neo HD FA are substantially equivalent in safety and effectiveness to those produced by a cleared neonatal fundus camera reference device. The results of the image quality comparison study demonstrated that of all study images were found to be of comparable quality to those produced by the reference device supporting substantial equivalence of the subject device as safe and effective for its intended use.
K243600
{9}
Forus Health
Original Premarket 510(k) Notification
510(k) Summary
# 7. Conclusion on Bench tests and Clinical Performance Data
The bench tests and clinical performance data supports our conclusion of substantial equivalence of 3nethra neo HD FA /3nethra neo HD to the mentioned predicate devices.
K243600
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.