da Vinci SP SureForm 45 Staplers and Reloads (SP1098)
K243596 · Intuitive Surgical, Inc. · GAG · Mar 10, 2025 · General Hospital
Device Facts
Record ID
K243596
Device Name
da Vinci SP SureForm 45 Staplers and Reloads (SP1098)
Applicant
Intuitive Surgical, Inc.
Product Code
GAG · General Hospital
Decision Date
Mar 10, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4740
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Intuitive Surgical da Vinci SP SureForm 45 Stapler, SP SureForm 45 Curved-Tip Stapler, and SureForm 45 Reloads and other accessories are intended to be used with the da Vinci SP Surgical System for resection, transection, and/or creation of anastomoses in Urologic, Thoracic, and Colorectal surgery.
Device Story
The da Vinci SP SureForm 45 Staplers are disposable, fully wristed articulating instruments designed for use with the da Vinci SP Surgical System (SP1098). The device accepts single-use reload cartridges containing multiple staggered rows of implantable staples and a stainless-steel knife. During surgery, the system drives the instrument to clamp tissue, fire staples, and transect tissue along the staple line. The device utilizes SmartFire and ForceFire features to manage clamping and firing forces based on motor torque, gear ratios, and friction parameters specific to the SP Patient Side Cart. Operated by a surgeon via the da Vinci SP console, the system provides real-time feedback and control. The device facilitates minimally invasive procedures, potentially reducing trauma and improving surgical outcomes through precise, computer-controlled tissue manipulation.
Clinical Evidence
No human clinical data provided. Evidence consists of bench testing (physical, mechanical, electrical, user interface, reliability) and acute animal validation studies. Animal studies included 5 canine and 8 porcine models for staple line performance/hemostasis, ex-vivo porcine tissue for leak onset pressure, and 1 canine, 3 porcine, and 2 cadaveric models for design validation. Results demonstrated acceptable pass rates for transection, tissue approximation, hemostasis, and staple formation, meeting all acceptance criteria.
Technological Characteristics
Disposable, fully wristed articulating surgical stapler. Compatible with da Vinci SP Surgical System. Uses stainless-steel knife and staggered staple rows. Sterilization: Ethylene Oxide (EO). Software-controlled via SP Patient Side Cart instrument drives (motor torque, gear ratios, friction). Connectivity: System-integrated. Materials: Patient-contacting materials similar to predicate.
Indications for Use
Indicated for patients undergoing Urologic, Thoracic, or Colorectal surgery requiring resection, transection, or creation of anastomoses using the da Vinci SP Surgical System.
Regulatory Classification
Identification
Surgical stapler for external use: A specialized prescription device used to deliver compatible staples to skin during surgery. Surgical stapler for internal use: A specialized prescription device used to deliver compatible staples to internal tissues during surgery for resection, transection, and creating anastomoses.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(i) Performance testing must demonstrate that the stapler, when used with compatible staples, performs as intended under anticipated conditions of use. Performance testing must include the following:
(A) Evaluation of staple formation characteristics in the maximum and minimum tissue thicknesses for each staple type;
(B) For manual staplers only, measurement of the worst-case deployment pressures on stapler firing force;
(C) Measurement of staple line strength;
(D) Confirmation of staple line integrity; and
(E) In vivo confirmation of staple line hemostasis.
(ii) For powered staplers only, appropriate analysis/testing must demonstrate the electromagnetic compatibility and electrical, thermal, and mechanical safety of the device.
(iii) For powered staplers only, appropriate software verification, validation, and hazard analysis must be performed.
(iv) Human factors testing must demonstrate that the clinician can correctly select and safely use the device, as identified in the labeling, based on reading the directions for use.
(v) The elements of the device that may contact the patient must be demonstrated to be biocompatible.
(vi) Performance data must demonstrate the sterility of the device.
(vii) Validation of cleaning and sterilization instructions must demonstrate that any reusable device components can be safely and effectively reprocessed per the recommended cleaning and sterilization protocol in the labeling.
(viii) Performance data must support the shelf life of the device by demonstrating continued device functionality, sterility, and package integrity over the identified shelf life.
(ix) Labeling of the device must include the following:
(A) Unless data demonstrates the safety of doing so, contraindications must be identified regarding use of the device on tissues for which the risk of stapling outweighs any reasonably foreseeable benefit due to known complications, including the stapling of tissues that are necrotic, friable, or have altered integrity.
(B) Unless available information demonstrates that the specific warnings do not apply, the labeling must provide appropriate warnings regarding how to avoid known hazards associated with device use including:
(
*1* ) Avoidance of use of the stapler to staple tissue outside of the labeled limits for maximum and minimum tissue thickness;(
*2* ) Avoidance of obstructions to the creation of the staple line and the unintended stapling of other anatomic structures;(
*3* ) Avoidance of clamping and unclamping of delicate tissue structures to prevent tissue damage;(
*4* ) Avoidance of use of the stapler on the aorta;(
*5* ) Establishing proximal control of blood vessels prior to stapling where practical and methods of blood vessel control in the event of stapler failure;(
*6* ) Ensuring stapler compatibility with staples; and(
*7* ) Risks specifically associated with the crossing of staple lines.(C) Specific user instructions for proper device use including measures associated with the prevention of device malfunction, and evaluation of the appropriateness of the target tissue for stapling.
(D) List of staples with which the stapler has been demonstrated to be compatible.
(E) Identification of key performance parameters and technical characteristics of the stapler and the compatible staples needed for safe use of the device.
(F) Information regarding tissues on which the stapler is intended to be used.
(G) Identification of safety mechanisms of the stapler.
(H) Validated methods and instructions for reprocessing of any reusable device components.
(I) An expiration date/shelf life.
(x) Package labels must include critical information and technical characteristics necessary for proper device selection.
{0}------------------------------------------------
March 10, 2025
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services-USA. To the right, the FDA logo is displayed in blue, with the words "FDA U.S. FOOD & DRUG" stacked above the word "ADMINISTRATION".
Intuitive Surgical, Inc. Alifiya Jagmag Regulatory Affairs Project Manager 1266 Kifer Road Sunnyvale, California 94086
Re: K243596
Trade/Device Name: da Vinci SP SureForm 45 Staplers and Reloads (SP1098) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: NAY, GDW, GAG Dated: February 5, 2025 Received: February 5, 2025
Dear Alifiya Jagmag:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mark Trumbore Digitally signed by Mark
Trumbore -S Date: 2025.03.10 13:37:48 -5 -04'00' Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K243596
Device Name
da Vinci SP SureForm 45 Staplers and Reloads (SP1098)
Indications for Use (Describe)
The Intuitive Surgical da Vinci SP SureForm 45 Stapler, SP SureForm 45 Curved-Tip Stapler, and SureForm 45 Reloads and other accessories are intended to be used with the da Vinci SP Surgical System for resection, transection, and/or creation of anastomoses in Urologic, Thoracic, and Colorectal surgery.
Type of Use (Select one or both, as applicable)
> | Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary (21 CFR § 807.92)
#### I. SUBMITTER INFORMATION
| Submitter: | Intuitive Surgical, Inc.<br>1266 Kifer Road<br>Sunnyvale, CA 94086 |
|------------|-------------------------------------------------------------------------------------------------------------------|
| Contact: | Alifiya Jagmag<br>Regulatory Project Manager<br>Phone Number: 617-850-5688<br>Email: Alifiya.Jagmag@jintusurg.com |
Date Summary Prepared: March 7, 2025
#### II. SUBMITTER INFORMATION
| Trade Name: | da Vinci SP SureFormTM 45 Curved-Tip Stapler |
|----------------------|--------------------------------------------------------|
| | da Vinci SP SureFormTM 45 Stapler |
| | SureFormTM 45 Reload, Gray |
| | SureFormTM 45 Reload, White |
| | SureFormTM 45 Reload, Blue |
| | SureFormTM 45 Reload, Green |
| | SureFormTM 45 Reload, Black |
| Common Name: | Stapler, Surgical |
| | Staple, Implantable |
| | System, Surgical, Computer Controlled Instrument |
| Classification Name: | Surgical Stapler (21 CFR §878.4740) |
| | Implantable Staple (21 CFR §878.4750) |
| | Endoscope and Accessories (21 CFR §876.1500) |
| Regulatory Class: | Class II |
| Product Code: | NAY (System, Surgical, Computer Controlled Instrument) |
| | GDW (Staple, Implantable) |
| | GAG (Stapler, Surgical) |
| Submission Type: | Traditional 510(k) |
Image /page/4/Picture/8 description: The image shows the word "INTUITIVE" in blue font. There is a small blue circle above the letter "U". The font is sans-serif and the letters are evenly spaced. The image is simple and clean, with a white background.
{5}------------------------------------------------
#### III. PREDICATE DEVICE INFORMATION
Predicate Device: SureForm 45 Curved-Tip Stapler (K190999) SureForm 45 Stapler (K183224) SureForm 45 Reloads (K183224 & K190999)
#### IV. Device Description
The da Vinci SP SureForm 45 Curved Tip Stapler and da Vinci SP SureForm 45 Stapler, with the existing SureForm 45 Reloads (cleared via K183224 and K190999) are designed for use exclusively with compatible Intuitive da Vinci SP Surgical Systems (Model SP1098). It is intended for resection, transection and/or creation of anastomoses in surgery. The instrument achieves its intended use by placing multiple staggered rows of implantable staples in the target tissues (stapling) followed by cutting of the target tissue along the middle of the staple line (transection). The SP SureForm 45 Curved Tip Stapler and SP SureForm 45 Stapler Instruments are disposable, fully wristed articulating devices and are compatible with the existing, Intuitive manufactured SureForm 45 mm Gray, White, Blue, Green, and Black reloads (hereinafter referred to as SureForm 45 Reloads). The SP SureForm 45 Reloads consist of a single-use cartridge that contains multiple, staggered rows of implantable staples, and a stainless-steel knife. The specifications for the SureForm 45 Reloads are provided in Table 1.
| Attribute | SureForm 45 Reloads | | | | |
|-----------------------------------------------------------|----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| | Gray | White | Blue | Green | Black |
| No. of staple<br>rows and<br>staple line<br>configuration | 6 Staggered<br>staple rows<br>total;<br><br>3 on each<br>side of<br>transection; | 6 Staggered<br>staple rows<br>total;<br><br>3 on each<br>side of<br>transection; | 6 Staggered<br>staple rows<br>total;<br><br>3 on each<br>side of<br>transection; | 6 Staggered<br>staple rows<br>total;<br><br>3 on each<br>side of<br>transection; | 6 Staggered<br>staple rows<br>total;<br><br>3 on each<br>side of<br>transection; |
| No. of Staples | 66 staples<br>total | 66 staples<br>total | 66 staples<br>total | 66 staples<br>total | 66 staples<br>total |
## Table 1: SureForm 45 Reload Specifications
{6}------------------------------------------------
| Attribute | SureForm 45 Reloads | | | | |
|----------------------------------|---------------------|---------------------|--------------------|---------------------|---------------------|
| | Gray | White | Blue | Green | Black |
| Unformed<br>staple leg<br>length | 2.0 mm | 2.5 mm | 3.5 mm | 4.3 mm | 4.6 mm |
| Pictures | Image: Gray staple | Image: White staple | Image: Blue staple | Image: Green staple | Image: Black staple |
The SureForm 45 Reloads are single use devices and are shipped sterile to the surgeon with a retainer that protects the staples during shipping and transportation. The SureForm 45 Reloads are not compatible with any other Intuitive Surgical stapler instruments (the IS4000 Stapler 30 and 45 instruments and the IS3000 Stapler 45 instrument), and likewise, the IS4000 Stapler 30 and 45 reloads and IS3000 Stapler 45 reloads are not compatible with the SP SureForm 45 Stapler instruments.
## Intended Use
The SP SureForm 45 Staplers and Reloads are intended to resect, transect and/or create anastomoses in surgery.
#### Indications for Use
The Intuitive Surgical da Vinci SP SureForm 45 Stapler, SP SureForm 45 Curved-Tip Stapler, and SureForm 45 Reloads and other accessories are intended to be used with the da Vinci SP Surgical System for resection, transection, and/or creation of anastomoses in Urologic, Thoracic, and Colorectal surgery.
## Comparison of Technological Characteristics
The subject device, SP SureForm 45 Staplers and Reloads, and the predicate device, SureForm 45 Staplers and Reloads (cleared under K183224 & K190999) are regulated under the same regulation number, product code, and classification. They have the same intended use, sterility characteristics (EO sterilization), and principles of operation. The subject device, SP SureForm 45 Staplers and Reloads, and the predicate SureForm 45 Staplers and Reloads differ in some design attributes (the subject device has an instrument shaft and proximal housing that allows for the instrument to mate with and be driven by the da Vinci SP Surgical System Patient Side Cart (PSC) instrument drives), indications for use (the subject device is indicated for use with the da Vinci SP Surgical system in Urology, Thoracic and Colorectal surgeries), accessory (the subject SP SureForm 45 Staplers use the da Vinci SP Advanced Access Port system for insertion into the patient anatomy), software
{7}------------------------------------------------
(the subject device uses the SP Software control parameters taking into account for the mechanical characteristics of the SP Patient Side Cart Instrument drive and the subject SP SureForm 45 Staplers (e.g. motor torque constants, gear ratios, friction, etc.) in order to generate the clamping and firing forces and motion profiles that achieve the same performance outcome using the same SmartFire and ForceFire features as the predicate SureForm 45 Stapler) and patient-contacting materials (most same, one different).
### Performance Data
Performance test data (bench and animal tests) for the SP SureForm 45 Staplers and Reloads demonstrate that the subject device is substantially equivalent to the predicate device and that the design output meets the design input requirements, user needs, and intended use. The testing is summarized below.
#### Design Verification (Bench):
The SP SureForm 45 Staplers and Reloads were subjected to full design verification testing included:
- Physical Specifications ●
- Mechanical Requirements ●
- Electrical Requirements ●
- User Interface Requirements
- Equipment Interface Requirements ●
- Reliability ●
- Packaging and labeling ●
The SP SureForm 45 Staplers and Reloads met all of the bench testing acceptance criteria, demonstrating that that the design output meets the design input requirements.
## Design Validation (Animal):
A series of acute validation studies were performed using simulated clinical models (animal) to evaluate the performance of the subject device. Acute clinical validation studies included Staple Line Performance, Leak Onset Testing, Buttress Compatibility, and Design Validation Testing. A side-by-side comparison between the subject and predicate device was performed in the Staple Line Performance and Leak Onset Testing to demonstrate substantial equivalence between the subject and predicate devices. Design Validation Testing demonstrated that the design outputs of the subject device fulfill the design input requirements and that user needs and intended uses are met.
The subject devices, SP SureForm 45 Staplers and Reloads, have similar technology and risk profiles and maintain identical tip geometry and utilize the same reloads as the predicate SureForm 45 Staplers and Reloads (cleared via K183224 and K190999). Hence animal survival studies were not conducted on the subject devices. The survival studies performed on predicate SureForm 45 Staplers and Reloads are used to support the safety of the subject SP SureForm 45 Staplers and Reloads.
{8}------------------------------------------------
A summary of the acute animal validation studies is provided in Table 2 below.
| Study Name | Study Purpose | Animal<br>Model | Study Outcome |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Acute Testing: | | | |
| Staple Line<br>Performance | Assess staple line<br>performance and staple<br>formation of the subject<br>device compared to the<br>predicate | 5 Canine<br>models<br>8 Porcine<br>models | The subject device met all<br>acceptance criteria and<br>exhibited acceptable pass<br>rates in the areas of<br>transection, tissue layer<br>approximation, hemostasis,<br>and staple formation |
| Leak Onset<br>Pressure | Assess leak onset pressure<br>performance on ex vivo<br>tissue of the subject device<br>as compared to the predicate<br>device. | Ex-vivo<br>porcine<br>tissue | All staple lines met all<br>acceptance criteria and<br>exhibited acceptable pass<br>rates in leak onset pressure<br>on thick (stomach) tissue<br>and thin (vein) tissue. |
| Design<br>Validation | Design validation testing of<br>the subject device was<br>performed in a clinical<br>laboratory setting closely<br>approximating an<br>intraoperative use situation. | 1 Canine<br>model<br>3 Porcine<br>models<br>2 Cadaveric<br>models | The subject device met all<br>acceptance criteria. |
## Table 2: Summary of Animal Validation Studies
Image /page/8/Picture/5 description: The image shows the word "INTUITIVE" in blue font. There is a small circle above the letter "U". The font is sans-serif and the letters are evenly spaced. The image is simple and clean, with a white background.
{9}------------------------------------------------
### Human Factors Evaluation:
Human Factors process conducted for the subject devices included the following activities:
- Known use-related issues for predicate devices similar to the subject devices O were analyzed using post-market data and the MAUDE database. All identified userelated issues that are relevant to the use of the subject devices were documented in the risk analvsis.
- A Comparative Task Analysis (CTA) was conducted to describe all aspects of the user- O device interaction, through the breakdown of steps into user tasks, and to provide an analysis of comparison to the predicate for each task.
- A Use-Related Risk Analysis (URRA) was conducted to identify all use-related risks for O each user task identified as New or Modified from the predicate in the CTA.
- Formative usability evaluations were conducted during the development process to O inform the device user interface design and confirm assessment of use-related risks.
- Summative usability validations were conducted on the subject SP SureForm 45 Staplers O Reloads and were found to be safe and effective for the intended users, uses, and use environments.
Summary: Based on the indications for use, technological characteristics, and performance data, the subject device, SP SureForm 45 Staplers and Reloads are substantially equivalent to the predicate devices, the SureForm 45 Staplers Reloads.
Image /page/9/Picture/10 description: The image shows the word "INTUITIVE" in blue font. There is a small circle above the letter "I". The font is sans-serif and the letters are evenly spaced. The word is centered in the image.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.