The LumiRex Bronchoscope is a single-use flexible endoscope designed to be used with Contra Healthcare Solutions video processors, endotherapy accessories, and ancillary equipment for endoscopy and endoscopic procedures within the airways and the tracheobronchial tree in a hospital environment.
Device Story
LumiRex Bronchoscope is a single-use, flexible, video-based endoscope for airway/tracheobronchial tree visualization. Device inserts through mouth or nose; features integrated light source and camera. Includes working channel for passage of instruments (biopsy forceps, brushes, suction). Operates via wired or wireless adapters connected to a Video Processor Unit. Wireless mode utilizes a built-in battery for improved maneuverability. Healthcare providers view real-time video on a monitor connected to the processor; images can be saved to USB media. Automated white balance simplifies user interface. Device facilitates endoscopic procedures, aiding clinical decision-making through direct visualization. Benefits include enhanced maneuverability (smaller bend radius) and flexible connectivity options.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including sterilization validation (ISO 11135), biocompatibility (ISO 10993), electrical safety/EMC (IEC 60601 series), and performance testing (ISO 8600 series). Wireless latency and image quality (MTF, SNR, field of view) were verified against standards and compared to the predicate device.
Technological Characteristics
Single-use flexible endoscope. Materials evaluated per ISO 10993. Sterilization: Ethylene Oxide (ISO 11135). Connectivity: Wired or wireless (battery-powered) adapters. Imaging: Integrated camera/LED. Standards: ISO 80369-7 (connectors), ISO 8600-1/3/4 (endoscope requirements), IEC 60601-1/1-2 (electrical/EMC), IEC 62471 (photobiological safety). Software: IEC 62304 compliant.
Indications for Use
Indicated for use in patients requiring endoscopic procedures within the airways and tracheobronchial tree. Intended for use in a hospital environment by healthcare professionals.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 30, 2025
Contra Healthcare Solutions, LLC
John O'Conner
President
105 Lincoln Avenue
Butler, Pennsylvania 16001
Re: K243541
Trade/Device Name: LumiRex Bronchoscope
Regulation Number: 21 CFR 874.4680
Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories
Regulatory Class: Class II
Product Code: EOQ
Dated: March 31, 2025
Received: March 31, 2025
Dear John O'Conner:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243541 - John O'Conner
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K243541 - John O'Conner
Page 3
Sincerely,
Joyce C. Lin -S
for Shu-Chen Peng, Ph.D.
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
{3}
LumiRex Bronchoscope
Page 12 of 90
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K243541 | ? |
| Please provide the device trade name(s). | | ? |
| LumiRex Bronchoscope | | |
| Please provide your Indications for Use below. | | ? |
| The LumiRex Bronchoscope is a single-use flexible endoscope designed to be used with Contra Healthcare Solutions video processors, endotherapy accessories, and ancillary equipment for endoscopy and endoscopic procedures within the airways and the tracheobronchial tree in a hospital environment. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(k) Summary
1. Submitter
Contra Healthcare Solutions, LLC
105 Lincoln Avenue Butler PA 16001 United States
724-285-6324
Applicant Contact: Mr. John O'Conner
jconnor@alpinsurgical.com
2. Device Name
Trade Name: LumiRex Bronchoscope
Common Name: Bronchoscope (flexible or rigid) and accessories
Classification Name: Bronchoscope (Flexible or Rigid)
Regulation Number: 874.4680
Product Code: EOQ
3. Predicate Device
Trade Name: Vathin Video Bronchoscope System
510(K) Number: K191828
Product Code: EOQ
4. Device Description Summary
The LumiRex Single-Use Bronchoscope is a flexible video-based medical device designed for insertion through the mouth or nose into the trachea and bronchial tubes. It comprises a long, thin tube equipped with a light source and a camera to provide real-time visualization of the airways. The bronchoscope also features a working channel that allows for the passage of various instruments, such as biopsy forceps, brushes, and suction devices, facilitating endoscopic procedures. The device is available in different diameters to accommodate a range of patient anatomies and clinical needs.
The Bronchoscope can be used wired or wireless through a Wired Scope Adapter or Wireless Scope Adapter. The adapters transmit images from the bronchoscope to the Video Processor Unit where the user can view the image from the bronchoscope, control the LED, and save and export images from the bronchoscope onto USB removable media.
5. Intended Use/ Indications for Use
The LumiRex Bronchoscope is a single-use flexible endoscope designed to be used with Contra Healthcare Solutions video processors, endotherapy accessories, and ancillary equipment for
{5}
endoscopy and endoscopic procedures within the airways and the tracheobronchial tree in a hospital environment.
## 6. Indications for Use Comparison
The subject device LumiRex Bronchoscope has the same intended use as the predicate device Vathin Video Bronchoscope System (K191828).
## 7. Technological Comparison
### Overview of Similarities
The subject device and the predicate device have same or similar optical capabilities and conical lock, mostly similar insertion tube characteristics, and some same or similar image characteristics. These similarities do not raise any questions in regard to the safety or performance of the subject device.
### Overview of Differences
The subject device allows the user to operate the device wirelessly and wired through respective adapters while the predicate device only has the option of wired use. The wired adapter of the subject device is operated in the same manner as the predicate device. The wireless adapter of the subject device has a built in battery powering the bronchoscope and allows better maneuverability and convenience during use for healthcare providers. The risks of wireless use has been assessed and found to acceptable. The functionality of the bronchoscope with wireless adapter has been validated.
The risks associated with wireless use, including latency performance testing to analyze the latency of the video feed transmission from the medical endoscope device to the medical monitor, have been assessed. The testing demonstrates that latency remains within an acceptance range. Notably, latency performance is comparable between the subject device's wired and wireless modes, although latency data for the predicate device was not tested.
The subject device has a smaller insertion tube bend radius, allowing better flexibility and maneuverability within a patient's respiratory tract. This does not affect the user's ability to carry out the intended use of the device or introduce any new risks.
The subject device has automated White balance and no image enhancement control, this simplifies the user interface and does not affect the user's ability to carry out the intended use of the device or introduce any new risks.
Comparative testing between the subject device and predicate device showed that the subject device has a wider field of view and larger depth of view.
{6}
The subject device has different video format inputs and output formats. However, these do not affect the user's ability to carry out the intended use of the device or introduce any new risks as the LumiRex Video Processor Unit takes the resultant output format and provides the user a display of bronchoscope video image and data storage in a manner similar to the predicate device.
The above differences between the subject and predicate device do not raise any questions to the safety and performance of the predicate device.
## 8. Non-Clinical and/or Clinical Tests Summary & Conclusions
Non-clinical validation testing was performed to demonstrate the safety and effectiveness of the device. The following tests were performed:
### Sterilization and Shelf-Life
Ethelyne Oxide Sterilization of LumiRex Single-Use Bronchoscope has been evaluated to ISO 11135:2014. Sterile Barrier was evaluated to ISO 11607-2:2019 and ASTM F1980 for accelerated aging methodology. The shelf life of the LumiRex Single-Use Bronchoscope has validated.
### Reprocessing
Reprocessing of reusable components of the LumiRex Single-Use Bronchoscope has been evaluated in accordance with:
- Guidance for Industry and FDA Staff – Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling issued March 17, 2025
- AAMI TIR12:2020 Designing, testing, and labeling reusable medical devices for reprocessing health care facilities: A guide for medical device manufacturers
- AAMI/ANSI/ISO 11737-1:2018 Sterilization of health care products-Microbiological Methods Part 1: Determination of the Population Microorganisms on produce
- ANSI/AAMI ST79:2017 Comprehensive guide to steam sterilization and sterility assurance in health care facilities
- ANSI/AAMI/ISO 17665-1:2006/(R)2013 Sterilization of health care products – Moist heat – Part 1 Requirements for the development, validation, and routine control of a sterilization process for medical devices
- AAMI ST98:2022 Cleaning validation of health care products- Requirements for development and validation of a cleaning process for medical devices
### Biocompatibility
The biocompatibility evaluation for the LumiRex device was conducted in accordance with the FDA Guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical device – Part 1: Evaluation and testing within a risk management process" issued September 4, 2020, and International Standard 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," FDA recognition number 258 for:
- Cytotoxicity (10993-5:2009)
- Sensitization (10993-10:2021)
- Systemic Toxicity (10993-11:2017)
{7}
- Pyrogenicity (10993-11:2017)
- Irritation (10993-23:2021)
## Software
Software verification and validation testing has been conducted in compliance with IEC 62304 Edition 1.1 2015-06 (which is the principal normative standard applied), ANSI/AAMI/IEC ES60601-1:2005/(R)2012 and A1:2012 and the FDA guidance documents “General Principles of Software Validation; Final Guidance for Industry and FDA Staff” (11-Jan-2002) and “FDA Guidance for the Content of Premarket Submissions for Device Software Functions” (14-June-2023).
## Cybersecurity
The cybersecurity of LumiRex Bronchoscope system was evaluated in conformance to “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions” issued September 2023.
## Electrical Safety and Electromagnetic Compatibility
The LumiRex Single-Use Bronchoscope, Wired Scope Adapter, Wireless Scope Adapter, Wireless Scope Adapter Charging Cradle, and LumiRex Video Processor have undergone the following Electrical Safety and Electromagnetic Compatibility testing, complying with IEC 60601-1:2005/IEC 60601-1-2:2014+A1:2020/AMD2:2020, IEC 60601-1-2:2014+A1:2020, and IEC 60601-2-18 (3rd edition).
## Wireless
The latency of the LumiRex Bronchsope was tested, referring to "Radio Frequency Wireless Technology in Medical Devices – Guidance for Industry and Food and Drug Administration Staff".
## Bench Testing
- ISO 80369-7:2021 Small-bore Connectors for Liquids and Gases in Healthcare Applications
- ISO 8600-1:2015 Endoscopes – General Requirements
- ISO 8600-3:2019 Endoscopes – Determination of field of view and direction of view of endoscopes with optics
- ISO 8600-4:2014 Endoscopes – Determination of maximum width of insertion portion
- IEC 62471:2006 Photobiological Safety
- Camera Image Quality
- Spatial Resolution MTF (ISO 12233)
- Signal to noise Ratio (ISO 15739)
- Dynamic Range
- Depth of Field
- Geometric Distortion
- Image Intensity Uniformity
- Color performance
- System Functional Testing
{8}
- Performance characteristics such as leaking, irrigation, bending, articulating bending angle, endurance of bending section, instrument and accessory compatibility, and bending section radius.
## Verification in Differences of Design
Scope Adapter – Evaluation of the LumiRex Bronchoscope for Reprocessing, Software, Electrical Safety, Electromagnetic Compatibility, and System Functional Testing, FDA Guidance Document - Electromagnetic Compatibility (EMC) of Medical Devices (June 6, 2022) – Section J – Common Electromagnetic (EM) Emitters, and IEEE/ANSI C63.27-2021 demonstrated that the device in design of the subject device in use wired/wireless scope adapters met requirements in the referenced standards and/or matched the performance of the predicate device. The results met the requirements defined in the standard and/or matched the performance of the predicate device.
Insertion Tube – To verify the design of the maneuverable portion, performance characteristic testing and testing per ISO 8600-1:2015 was performed. The bending performance of the LumiRex Bronchoscope met the requirements defined in the standard.
Image and Optical Characteristics - To verify the image and optical performance of the LumiRex Bronchoscope, comparative testing was performed between the predicate and subject device for Camera Image Quality, and Determination of Field of View and Direction of View (ISO 8600-3:2019). The results met the requirements defined in the standard and demonstrated substantially equivalent performance to the predicate device.
## 9. Conclusions
The subject device LumiRex Bronchoscope is substantially equivalent to predicate device, Vathin Video Bronchoscope System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.