SonoTAP and SonoTAP II

K243525 · PAJUNK GmbH Medizintechnologie · BSP · Mar 24, 2025 · Anesthesiology

Device Facts

Record IDK243525
Device NameSonoTAP and SonoTAP II
ApplicantPAJUNK GmbH Medizintechnologie
Product CodeBSP · Anesthesiology
Decision DateMar 24, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5150
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The SonoTAP and SonoTAP II anaesthesia conduction needles enhanced for ultrasound visibility are intended for the transient delivery of anesthetics to provide regional anesthesia and analgesia.

Device Story

SonoTAP and SonoTAP II are single-use anesthesia conduction needles; designed for transient delivery of anesthetics for regional anesthesia/analgesia. Needles feature ultrasound-enhancing 'CornerStone' reflectors; stainless steel tubing; and either Luer (ISO 80369-7) or NRFit (ISO 80369-6) connectors. Used in clinical settings under ultrasound guidance by healthcare professionals. Device input is manual positioning by clinician; output is delivery of anesthetic agent. Benefits include improved needle visibility under ultrasound for accurate placement. Subject device modifications include additional sterilization facility (HA2 Medizintechnik) and alternative packaging materials (medical paper).

Clinical Evidence

Bench testing only. No clinical data presented. Testing included ISO 7864 (cleanliness, acidity/alkalinity, extractable metals, tube/point integrity, hub-to-needle bond strength), ISO 9626 (stiffness, breakage, corrosion resistance), and ISO 80369 series (leakage, stress cracking, axial load). Hub-to-needle bond strength met acceptance criteria (104.9N and 196.01N vs 69N requirement). Sterilization validated to SAL 10^-6.

Technological Characteristics

Materials: Stainless steel 1.4301 tubing; Copolyester (Tritan) or ABS hubs; Silicone coating. Features: CornerStone ultrasound reflectors; Luer or NRFit connectors. Dimensions: 18G-24G, 40mm-150mm length. Sterilization: Ethylene Oxide (SAL 10^-6). Packaging: Tyvek/foil or medical paper/foil. Manufacturing: Glued or directly molded hub-to-needle bonding.

Indications for Use

Indicated for patients requiring regional anesthesia and analgesia via transient delivery of anesthetics under ultrasound guidance. Not for intrathecal use.

Regulatory Classification

Identification

An anesthesia conduction needle is a device used to inject local anesthetics into a patient to provide regional anesthesia.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. March 24, 2025 PAJUNK GmbH Medizintechnologie Chiara Meyer Specialist Regulatory Affairs Karl-Hall-Str. 1 Pajunkstr. 1 Geisingen, BW 78187 Germany Re: K243525 Trade/Device Name: SonoTAP and SonoTAP II Regulation Number: 21 CFR 868.5150 Regulation Name: Anesthesia Conduction Needle Regulatory Class: Class II Product Code: BSP Dated: February 25, 2025 Received: February 25, 2025 Dear Chiara Meyer: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory {2}------------------------------------------------ assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Bradley Q. Quinn -S Bradley Quinn Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use Submission Number (if known) K243525 Device Name ## SonoTAP and SonoTAP II Indications for Use (Describe) The SonoTAP and SonoTAP II anaesthesia conduction needles enhanced for ultrasound visibility are intended for the transient delivery of anesthetics to provide regional anesthesia and analgesia. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the word "PAJUNK" in a bold, sans-serif font. The letters are a dark green color, and there is a registered trademark symbol to the right of the letter "K". The letter "P" has a white horizontal line going through it. ## USA # 08 510(k) Summary # Registrations USA 510(k) Summary as required by 21 CFR 807.92(c). Date of Preparation: 2025-01-14 # Document Control Number: K243525 | 510(k) owner: | PAJUNK® GmbH Medizintechnologie | |----------------------------------------------|---------------------------------------------| | | Karl-Hall-Str. 1/ Pajunkstr. 1 | | | 78187 Geisingen | | | Baden-Wuerttemberg, Germany | | | Fon: +49(0)7704-9291-586 | | | Fax: +49(0)7704-9291-605 | | | Establishment Registration Number: | | | 9611612 | | Submitter Information/ production site: | PAJUNK® GmbH Medizintechnologie | | | Karl-Hall-Str. 1/ Pajunkstr. 1 | | | 78187 Geisingen | | | Baden-Wuerttemberg, Germany | | | Fon: +49(0)7704-9291-586 | | | Fax: +49(0)7704-9291-605 | | | Establishment Registration Number: | | | 9611612 | | Manufacturer/ Submitter Contact: | Christian G. H. Quass | | | Director Regulatory Affairs, Safety Officer | | | Fon: +49(0)7704-9291-586 | | | Fax: +49(0)7704-9291-602 | | | E-Mail: christian.quass@pajunk.com | | Alternative Manufacturer/ Submitter Contact: | Chiara Meyer | | | Regulatory Affairs Specialist | | | Fon: +49(0)7704-9291-627 | | | Fax: +49(0)7704-9291-602 | | | E-Mail: chiara.meyer@pajunk.com | {5}------------------------------------------------ # 510(k) Premarket Notification Submission Image /page/5/Picture/2 description: The image shows the word "PAJUNK" in a bold, sans-serif font. The color of the text is a dark green. There is a registered trademark symbol to the right of the letter K. | USA Contact: | PAJUNK MEDICAL SYSTEMS<br>4575 Marconi Dr.<br>Alpharetta, GA 30005<br>Registration Number: 3004076349<br>FEI Number*: 3004076349 | |---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------| | US-Official Correspondent | Marco Wohnig<br>President<br>PAJUNK Medical Systems<br>4575 Marconi Dr.<br>Alpharetta, GA 30005<br>Phone: 001-770-4936832<br>E-Mail: Marco.Wohnig@pajunk-usa.com | | Contract Sterilizer: | Sterigenics Germany GmbH<br>Kasteler straße 45<br>65203 Wiesbaden<br>Germany, Hessen<br>Establishment Registration Number: 3002807090 | | 2nd Contract Sterilizer: | HA2 MEDIZINTECHNIK GMBH<br>Am Bahndamm 11<br>Halberstadt Saxony-Anhalt, DE 38820<br>Registration Number: 3009039068<br>FEI Number*: 3009039068 | # Device Name and Classification | Device Name: | SonoTAP und SonoTAP II | |------------------------------------|------------------------------| | Premarket Notification Number: | K253525 | | Classification Name: | Anesthesia Conduction Needle | | Classification Reference: | 21 CFR § 868.5150 | | Product Code: | BSP | | Establishment Registration Number: | 9611612 | | Regulatory Class: | II | | Panel: | Anesthesiology | {6}------------------------------------------------ # Premarket Notification Submission Image /page/6/Picture/2 description: The image shows the word "PAJUNK" in a bold, sans-serif font. The color of the text is a dark green. A registered trademark symbol is located to the right of the letter "K". | Sterilization Method: | Ethylene Oxide<br>disposable device, supplied sterile to the end user and non-sterile<br>intended to be sterilized prior to use to re-packagers/ medical device<br>manufacturers | |---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contract Sterilizer: | Sterigenics Germany GmbH<br>Kasteler straße 45<br>65203 Wiesbaden<br>Germany, Hessen<br>Establishment Registration Number: 3002807090 | | Additional Contract Sterilizer: | HA2 MEDIZINTECHNIK GMBH<br>Am Bahndamm 11<br>Halberstadt Saxony-Anhalt, DE 38820<br>Registration Number: 3009039068<br>FEI Number*: 3009039068 | # Changes to Device: Predicate Device | Device Name: | SonoTAP, Tuohy SONO | |------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Premarket Notification Number: | K113207 | | Classification Name: | Anesthesia Conduction Needle | | Classification Reference: | 21 CFR § 868.5150 | | Product Code: | BSP | | Establishment Registration Number: | 9611612 | | Regulatory Class: | II | | Panel: | Anesthesiology | | Sterilization Method: | Ethylene Oxide<br>disposable device, supplied sterile to the end user and non-sterile<br>intended to be sterilized prior to use to re-packagers/ medical device<br>manufacturers | | Contract Sterilizer: | Sterigenics Germany GmbH<br>Kasteler straße 45<br>65203 Wiesbaden<br>Germany, Hessen<br>Establishment Registration Number: 3002807090 | {7}------------------------------------------------ Image /page/7/Picture/1 description: The image shows the word "PAJUNK" in a bold, sans-serif font. The color of the text is a dark green. A registered trademark symbol is located to the right of the letter "K". # Changes to Device: Reference Device | Device Name: | SonoBlock, SonoBlock II | |------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Premarket Notification Number: | K241954 | | Classification Name: | Anesthesia Conduction Needle | | Classification Reference: | 21 CFR § 868.5150 | | Product Code: | BSP | | Establishment Registration Number: | 9611612 | | Regulatory Class: | II | | Panel: | Anesthesiology | | Sterilization Method: | Ethylene Oxide<br>disposable device, supplied sterile to the end user and non-sterile<br>intended to be sterilized prior to use to re-packagers/ medical device<br>manufacturers | | Contract Sterilizer: | Sterigenics Germany GmbH<br>Kasteler straße 45<br>65203 Wiesbaden<br>Germany, Hessen<br>Establishment Registration Number: 3002807090 | | Additional Contract Sterilizer: | HA2 MEDIZINTECHNIK GMBH<br>Am Bahndamm 11<br>Halberstadt Saxony-Anhalt, DE 38820<br>Registration Number: 3009039068<br>FEI Number*: 3009039068 | {8}------------------------------------------------ Image /page/8/Picture/1 description: The image shows the word "PAJUNK" in a bold, sans-serif font. The letters are a dark green color, and there is a registered trademark symbol to the right of the letter "K". The background is white. # 1 Narrative Device Description The subject device, the SonoTAP II needles are single-use anaesthesia conducting needles intended for the transient delivery of anesthetics to provide regional anesthesia and analgesia. The SonoTAP and SonoTAP II needles are intended for use under ultrasound guidance. Visibility under ultrasound is enhanced by CornerStone reflectors as cleared by FDA in several Premarket Notification Submissions. The SonoTAP and SonoTAP II needles is equipped with an injection tube. The distal connection of the tube is either equipped with a LUER Connector according to ISO 80369-7 or a NRFit-Connector according to ISO 80369-6. Image /page/8/Picture/5 description: The image shows a medical device, specifically an epidural catheter. The catheter is a thin, flexible tube with a connector at one end and markings along its length. The markings are evenly spaced and likely indicate depth or distance for insertion. The catheter is coiled at the end. Figure 1 SonoTAP Image /page/8/Picture/7 description: The image shows a medical device with a clear tube connected to a yellow connector on one end and a white and black device with a needle on the other end. The clear tube is coiled and appears to be made of plastic. The needle has markings on it, possibly indicating depth or measurement. Figure 2 SonoTAP II The device in focus is a single use device and has an intended time of use up to 24 hours acc. EO-residuals acc. DIN EN ISO 10993-7. The standard time of use is less than 60 minutes. The SonoTAP and SonoTAP II cannulas are not for intrathecal use. The SonoTAP and SonoTAP II cannulas are produced at PAJUNK® GmbH Medizintechnologie in Geisingen, Germany. {9}------------------------------------------------ Image /page/9/Picture/1 description: The image shows the word "PAJUNK" in a bold, sans-serif font. The letters are all capitalized and colored in a dark green. A registered trademark symbol is located to the right of the letter K. # 2 Determination of Substantial Equivalence ## Intended Use Predicate Device (510(K) Predicate) The cannulas/ needles for anesthesia and analgesia enhanced for ultrasound visibility – Tuohy Sono, Sono, Sono TAP, Quincke Sono, Chiba Sono, SPROTTE® Sono and Crawford Sono - are intended for the transient delivery of anesthetics to provide regional anesthesia and analgesia or to facilitate placement of a catheter. ## Intended Use Subject Device The SonoTAP and SonoTAP II anaesthesia conduction needles enhanced for ultrasound visibility are intended for the transient delivery of anesthetics to provide regional anesthesia and analgesia. ## Discussion The Intended Use of the Predicate device and of the subject device is substantially equivalent. The Subject device as well as the predicate device offers localization via ultrasound as it is state of the art and well established. The Predicate Device additionally offers the facilitation of the placement of a catheter (just Tuohy Sono). Conclusion: substantially equivalent # 3 Technical Description ## Substantial Equivalence: Tabulatory Outline | Characteristi<br>CS | Subject Device<br>SonoTAP<br>Manufacturer: PAJUNK® GmbH<br>Medizintechnologie, Geisingen | Subject Device<br>SonoTAP II<br>Manufacturer: PAJUNK® GmbH<br>Medizintechnologie, Geisingen | Predicate Device K113207<br>SonoTAP<br>Manufacturer: PAJUNK® GmbH<br>Medizintechnologie, Geisingen | |--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------| | Picture | Image: SonoTAP needle | Image: SonoTAP II needle | Image: SonoTAP needle | | Needle<br>Tubing | Stainless Steel 1.4301 | Stainless Steel 1.4301 | Stainless Steel 1.4301 | | Tip | Facet; Tuohy | Facet | Facet; Tuohy | | Characteristics | Subject Device<br>SonoTAP<br>Manufacturer: PAJUNK® GmbH<br>Medizintechnologie, Geisingen | Subject Device<br>SonoTAP II<br>Manufacturer: PAJUNK® GmbH<br>Medizintechnologie, Geisingen | Predicate Device K113207<br>SonoTAP<br>Manufacturer: PAJUNK® GmbH<br>Medizintechnologie, Geisingen | | Hub | Copolyester (Tritan MX731)<br>(Indirect Patient Contact) | Lustran/Guardian ABS 348 white<br>012002 (No Patient Contact) | Copolyester (Tritan MX731)<br>(Indirect Patient Contact) | | Coating | Silicone (Dow Corning 360 Medical<br>Fluid 12500 cSt), Silicone thinner<br>(Dow Corning Q7-9180 Silicone<br>Fluid 1.0 cSt) | Silicone (Dow Corning 360 Medical<br>Fluid 12500 cSt), Silicone thinner<br>(Dow Corning Q7-9180 Silicone Fluid<br>1.0 cSt) | Silicone (Dow Corning 360 Medical<br>Fluid 12500 cSt), Silicone thinner<br>(Dow Corning Q7-9180 Silicone Fluid<br>1.0 cSt) | | US-<br>enhancement | Cornerstones | Cornerstones | Cornerstones | | Graduation | Tampapur TPU | Tampapur TPU | Tampapur TPU | | Bonding<br>Technology<br>Needle-to-<br>Hub | glued | Directly molded | glued | | Glue | (Epoxy adhesive - Araldite 2011),<br>alternatively: UV adhesive (Loctite<br>AA 3921) | n.a. | (Epoxy adhesive - Araldite 2011),<br>alternatively: UV adhesive (Loctite<br>AA 3921) | | Injection tube | PVC (MED7536), PC (HP4) | PVC (MED7536); PVC (RB1); PC<br>(HP4) | PVC (MED7536); PVC (RB1); PC<br>(HP4) | | Diameter | 18G-24G | 21G-24G | 21G-24G | | Length | 40mm-150mm | 40mm-150mm | 40mm-150mm | | Connectivity | ISO 80369-7 (LUER) | ISO 80369-6 (NRFit)<br>ISO 80369-7 (LUER) | ISO 80369-7 (LUER) | | Packaging | Tyvek, Foil, chromo-duplex board<br>Alternatively: Paper, Foil | Tyvek, Foil, chromo-duplex board<br>Alternatively: Paper, Foil | Tyvek, Foil, chromo-duplex board | | Biocompatibility | Complies with ISO10993-series as required by « Use of International Standard ISO 10993-1, "Biological evaluation<br>of medical devices - Part 1: Evaluation and testing within a risk management process" », Docket Number : FDA-201<br>3-D-0350 | | | | Sterilization<br>Method | Ethyleneoxide | Ethyleneoxide | Ethyleneoxide | | Sterility<br>Assurance<br>Level | SAL=10-6 | SAL=10-6 | SAL=10-6 | | Characteristi<br>cs | Subject Device<br>SonoTAP<br>Manufacturer: PAJUNK® GmbH<br>Medizintechnologie, Geisingen | Subject Device<br>SonoTAP II<br>Manufacturer: PAJUNK® GmbH<br>Medizintechnologie, Geisingen | Predicate Device K113207<br>SonoTAP<br>Manufacturer: PAJUNK® GmbH<br>Medizintechnologie, Geisingen | | Sterilization<br>Service<br>Provider | Sterigenics, Wiesbaden<br>HA2 Medizintechnik, Halberstadt | Sterigenics, Wiesbaden<br>HA2 Medizintechnik, Halberstadt | Sterigenics, Wiesbaden | | Shelf Life | 60 month from sterilization | 60 month from sterilization | 60 month from sterilization | {10}------------------------------------------------ # 510(k) # Premarket Notification Submission Image /page/10/Picture/2 description: The image shows the word "PAJUNK" in a bold, sans-serif font. The letters are all capitalized and colored in a dark green hue. A small, circled "R" symbol is present to the right of the letter "K", indicating a registered trademark. {11}------------------------------------------------ Image /page/11/Picture/1 description: The image shows the logo for PAJUNK. The logo is in a dark green color. The letters are in a sans-serif font and the logo has a registered trademark symbol on the upper right. # Premarket Notification Submission # 4 Substantial Equivalence Discussion The Predicate devices, as well as the subject devices are equipped with ultrasound enhancing cornerstones. Ultrasound position guidance is state of the art. The technique is safe and effective. The subject device can be either directly injection moulded or glued. Both manufacturing technologies are safe and the processes are running stable. Both technologies are standard technologies. The predicate devices, the SonoTAP needles, are packed with Tyvek and foil and sterilized with Ethyleneoxide at Sterigenics, Germany. The subject devices, the SonoTAP II needles, are packed with Tyvek and foil as well as optionally with Medical Paper and foil. Furthermore, the subject devices are sterilized at Sterigenics in Wiesbaden as well as optionally at HA2 Medizintechnik in Halberstadt. The main differences which are subject to this premarket submission are - the addition of a sterilization facility and a) - b) the addition of an alternative packaging material. In order to verify substantial equivalence of the sterilization validation validation according ISO 11135 has been performed and successfully accomplished. The sterility assurance level of 10-6 has been successfully validated at both facilities: Sterigenics and HA2. {12}------------------------------------------------ Image /page/12/Picture/1 description: The image shows the word "PAJUNK" in a bold, sans-serif font. The color of the text is a dark green. The letters are closely spaced together, creating a solid block of text. The background is white. ## Premarket Notification Submission In order to verify substantial equivalence of the packaging and shelf life tests have been performed. Both, medical paper and Tyvek have been validated and verified to allow sterilization using EtO and to maintain sterility after a shelf life of 60 month after sterilization. # 5 Performance Testing The needles subject to this premarket submission are subject to performance testing prior to release as well as after shelf life scenario of 5 years. There is no impact of the change subject to this submission (additional optional sterilization provider, additional optional packaging) on the performance of the needle as defined in the user requirements specification. Especially the hub to needle bonding requirements remain unaffacted. The processes are still valid and reliable. The test listed below are standard tests performed on a reqular basis as well as type tests performed after design transfer. Tests are performed to comply with the international standards listed below in the respectively most current version: ISO 9626 Stainless steel needle tubing for manufacture of medical devices ISO 7864 Sterile hypodermic needles for single use - Requirements and test methods ISO 80369-6 Small bore connectors for liguids and gases in healthcare applications - Part 6: Connectors for neuraxial applications ISO 80369-7 Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications ISO 80369-20 Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methods ISO 9626 defines the quality and properties of the stainless steel tubing remains unaltered and therefore ongoing compliance with this standard remains unaffected by the change. There is no additional/ new risk of breakage or bending. ISO 80369-series defines the type of connector which also is not affected by the change. The technical equivalency of LUER and NRFit has been verified and cleared in several submissions before and therefore is in compliance. The test reports related to this standard series also do not directly contribute to demonstration of substantial equivalence in this submission. There is no additional risk in this regard. ISO 7864 defines properties of the needle. Those properties are not directly linked to the material. Section 4.12 of this standard is directly linked to the stability of the connection of the needle and the risk of detachment oft he needle from the hub. The {13}------------------------------------------------ ## 510(k) Image /page/13/Picture/1 description: The image shows the word "PAJUNK" in a bold, sans-serif font. The letters are all capitalized and colored in dark green. A small, circled "R" symbol is located to the right of the "K", indicating that the word is a registered trademark. The background is plain white. ## Premarket Notification Submission acceptance criteria have to be met non regarding the material oft he hub and the needle or the type of connection technology. So the test reports for hub-to-needle bondage are required to demonstrate substantial equivalence. The test reports attached to this e-mail subjecting worst case needles demonstrate compliance with the standard (69N required, an average of 104,9N and 196,01N met in result) as well as process stability when manufacturing the needles either with PC Tritan or ABS, either glued or directly moulded. ## 5.1 ISO 7864 The following sections have been tested: | section | Pass / Fail | |---------------------------------------|-------------| | 4.3 Cleanliness | Passed | | 4.4 Limits for acidity or alkalinity | Passed | | 4.5 Limits for extractable metals | Passed | | 4.10 Needle Tube | Passed | | 4.11 Needle point | Passed | | 4.12 Bond between hub and needle tube | Passed | | 4.13 Patency of lumen | Passed | ## 5.2 ISO 9626 #### The following sections have been tested: | section | Pass / Fail | |------------------------------------------|-------------| | 5.2 Surface finish and visual appearance | Passed | | 5.3 Cleanliness | Passed | | 5.4 Limits for acidity and alkalinity | Passed | | 5.5 Size designation | Passed | | 5.6 Dimensions | Passed | | 5.7 Sample size | Passed | | 5.8 Stiffness | Passed | | 5.9 Resistance to breakage | Passed | | 5.10 Resistance to corrosion | Passed | {14}------------------------------------------------ Image /page/14/Picture/2 description: The image shows the word "PAJUNK" in a bold, sans-serif font. The color of the text is a dark green. A registered trademark symbol is located to the right of the letter "K". ## 5.3 ISO 80369 The following sections of the 80369-6 have been tested: | section | | |---------|-----------------------| | 6.1 | Fluid Leakage | | 6.2 | Air Leakage | | 6.3 | Stress Cracking | | 6.4 | Separation Axial Load | | 6.5 | Unscrewing | | 6.6 | Overriding | ## The following sections of the 80369-7 have been tested: | section | | |---------|-----------------------| | 7.1 | Fluid Leakage | | 7.2 | Air Leakage | | 7.3 | Stress Cracking | | 7.4 | Separation Axial Load | | 7.5 | Unscrewing | | 7.6 | Overriding | # 6 Conclusion The comparison between the predicate device and the subject device of this submission focussed on the validated sterilization process and the shelf life testing demonstrates that the subject device can be optionally in addition packed using medical paper and can be optionally sterilized at the alternative sterilization service provider at an SAL of 10-6. Based on current in-process testing regularly conducted, safe performance of the SonoTAP II needle is demonstrated. The results demonstrate that the subject device is substantially equivalent and as safe, as effective, and performs as well as the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...