This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained staff as an aid in diagnosis, treatment and radiation therapy planning as well as for diagnostic and therapeutic interventions. This CT system can be used for low dose lung cancer screening in high risk populations*. * As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
Device Story
NAEOTOM Alpha/Alpha.Pro/Alpha.Prime are CT scanner systems using photon-counting detector technology (QuantaMax) to generate cross-sectional images. Inputs include X-ray transmission data; system transforms these via computer reconstruction into DICOM images. Used in professional healthcare facilities by physicians, radiologists, or technologists. Output images aid in diagnosis, treatment, and radiation therapy planning (e.g., Brachytherapy, Proton Therapy, Surgery). Features include dual-source (Alpha/Alpha.Pro) or single-source (Alpha.Prime) configurations, VECTRON X-ray tubes, and SOMARIS/10 syngo CT VB20 software. Supports advanced workflows like FAST 3D Camera positioning, myNeedle Guide for interventions, and respiratory gating. Output affects clinical decision-making by providing high-resolution anatomical data for therapy planning and diagnostic assessment. Benefits include improved image quality, reduced radiation dose, and streamlined clinical workflows.
Clinical Evidence
Bench testing only. Performance verified through phantom tests, integration/functional testing, and image quality evaluations. Testing confirmed accuracy of FAST 3D camera features, respiratory gating (Direct i4D, DirectBreathhold), and radiotherapy planning protocols. Low-dose lung cancer screening performance was compared to predicates via technical parameter analysis, demonstrating substantial equivalence.
Indicated for patients requiring cross-sectional imaging via CT reconstruction for diagnosis, treatment, radiation therapy planning, and diagnostic/therapeutic interventions. Includes low-dose lung cancer screening for high-risk populations as defined by professional medical societies.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue and "ADMINISTRATION" in a smaller font size below.
February 12, 2025
Siemens Medical Solutions USA, Inc. % Tabitha Estes Regulatory Affairs Professional 810 Innovation Drive KNOXVILLE, TN 37932
Re: K243523
Trade/Device Name: NAEOTOM Alpha.Peak/ NAEOTOM Alpha; NAEOTOM Alpha.Pro; NAEOTOM Alpha.Prime Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: JAK Dated: November 14, 2024 Received: November 14, 2024
Dear Tabitha Estes:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K243523
Device Name
NAEOTOM Alpha.Peak/ NAEOTOM Alpha; NAEOTOM Alpha.Pro; NAEOTOM Alpha.Prime
Indications for Use (Describe)
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.
The images delivered by the system can be used by a trained staff as an aid in diagnosis, treatment and radiation therapy planning as well as for diagnostic and therapeutic interventions.
This CT system can be used for low dose lung cancer screening in high risk populations*.
* As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is an orange graphic of a cluster of circles.
# K243523
# 510(k) Summary for
## NAEOTOM CT Scanner Systems
## with software version SOMARIS/10 syngo CT VB20
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
### I. Submitter
Siemens Medical Solutions USA, Inc. 810 Innovation Drive Knoxville, TN 37932 Establishment Registration Number: 1034973
Importer/Distributor Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355 Establishment Registration Number: 2240869
Location of Manufacturing Site Siemens Healthineers AG Siemensstr. 1 -OR- Rittigfeld 1 D-91301 Forchheim, Germany Establishment Registration Number: 3004977335
Note: Descriptions in this submission use the short company name Siemens. Brand name on all products is Siemens Healthineers.
Submitter Contact Person: Tabitha Estes Regulatory Affairs Siemens Medical Solutions USA, Inc. (865) 804-4553 (work cell) tabitha.estes@Siemens-healthineers.com
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" are several orange dots of varying sizes.
# II. Device Name and Classification
| Product name | Trade name |
|---------------------|-------------------------------------|
| NAEOTOM Alpha | NAEOTOM Alpha.Peak<br>NAEOTOM Alpha |
| NAEOTOM Alpha.Pro | NAEOTOM Alpha.Pro |
| NAEOTOM Alpha.Prime | NAEOTOM Alpha.Prime |
| Classification Name: | Computed Tomography X-ray System |
|-----------------------|----------------------------------|
| Classification Panel: | Radiology |
| Regulation Number: | 21 CFR §892.1750 |
| Device Class: | Class II |
| Product Code: | JAK |
# III. Predicate Device
| Primary Predicate Device: | |
|---------------------------|------------------------------------------------------------------------------------|
| Trade Name: | NAEOTOM Alpha |
| 510(k) Number: | K233657 |
| Clearance Date: | March 28, 2024 |
| Classification Name: | Computed Tomography X-ray System |
| Classification Panel: | Radiology |
| Regulation Number: | 21 CFR §892.1750 |
| Device Class: | Class II |
| Product Code: | JAK |
| Recall Information: | All predicate device recalls have been considered in the subject<br>device design. |
| Secondary Predicate Device: | |
|-----------------------------|------------------------------------------------------------------------------------|
| Trade Name: | SOMATOM X.ceed |
| 510(k) Number: | K233650 |
| Clearance Date: | March 26, 2024 |
| Classification Name: | Computed Tomography X-ray System |
| Classification Panel: | Radiology |
| Regulation Number: | 21 CFR §892.1750 |
| Device Class: | Class II |
| Product Code: | JAK |
| Recall Information: | All predicate device recalls have been considered in the subject<br>device design. |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a cluster of orange dots.
| Reference Predicate Device: | |
|-----------------------------|------------------------------------------------------------------------------------|
| Trade Name: | SOMATOM go. Platform |
| 510(k) Number: | K233650 |
| Clearance Date: | March 26, 2024 |
| Classification Name: | Computed Tomography X-ray System |
| Classification Panel: | Radiology |
| Regulation Number: | 21 CFR §892.1750 |
| Device Class: | Class II |
| Product Code: | JAK |
| Recall Information: | All predicate device recalls have been considered in the subject<br>device design. |
Note: K233650 was a bundle 510(k) including the SOMATOM go. Platform (SOMATOM go.Now, SOMATOM go.Up, SOMATOM go.All, SOMATOM go.Top, SOMATOM go.Sim and SOMATOM go.Open Pro) and the SOMATOM X. Platform (SOMATOM X.cite and SOMATOM X.ceed).
## IV. Device Description
Siemens intends to market a new Single source CT system NAEOTOM Alpha.Prime and a new Dual Source CT system NAEOTOM Alpha.Pro based on the SOMARIS/10 software platform.
Siemens also intends to market updated software version, SOMARIS/10 syngo CT VB20, for the new NAEOTOM CT scanner systems and for the NAEOTOM Alpha cleared in K233657 (clearance date March 28th, 2024). The updated software version SOMARIS/10 syngo CT VB20 incorporates mainly features for radiotherapy planning support.
With the new software version SOMARIS/10 syngo CT VB20, the trade name of NAEOTOM Alpha cleared in K233657 has been changed to NAEOTOM Alpha.Peak for ex-factory systems. Systems already installed will receive the software update without change of the trade name. The system label keeps the product name NAEOTOM Alpha. System label information and UDI are not changed and remain the same as the original submitted.
For simplicity, the product name of NAEOTOM Alpha will be used throughout this submission instead of the trade name NAEOTOM Alpha.Peak.
The subject devices NAEOTOM Alpha (trade name ex-factory CT systems: NAEOTOM Alpha.Peak) and NAEOTOM Alpha.Pro with software version SOMARIS/10 syngo CT VB20 are Computed Tomography X-ray systems which feature two continuously rotating tube-detector systems, denominated as Aand B-systems respectively (dual source NAEOTOM CT scanner system).
The subject device NAEOTOM Alpha.Prime with software version SOMARIS/10 syngo CT VB20 is a Computed Tomography X-ray system which features one continuously rotating tube-detector systems, denominated as A-system (single source NAEOTOM Alpha CT scanner system).
The detectors' function is based on photon-counting technology.
In this submission, the above-mentioned CT scanner systems are jointly referred to as subject devices by "NAEOTOM CT scanner systems".
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange. To the right of the word "Healthineers" is a graphic of orange dots arranged in a circular pattern.
The NAEOTOM CT scanner systems with SOMARIS/10 syngo CT VB20 produce CT images in DICOM format, which can be used by trained staff for post-processing applications commercially distributed by Siemens and other vendors. The CT images can be used by a trained staff as an aid in diagnosis, treatment and radiation therapy planning as well as for diagnostic and therapeutic interventions. The radiation therapy planning support includes, but not limited to, Brachytherapy, Particle including Proton Therapy, External Beam Radiation Therapy, Surgery. The computer system delivered with the CT scanner is able to run optional post-processing applications.
Only trained and qualified users, certified in accordance with country-specific regulations, are authorized to operate the system. For example, physicians, radiologists, or technologists. The user must have the necessary U.S. qualifications in order to diagnose or treat the patient with the use of the images delivered by the system.
The platform software for the NAEOTOM CT scanner systems is syngo CT VB20 (SOMARIS/10 syngo CT VB20). It is a command-based program used for patient management, data management, X-ray scan control, image reconstruction, and image archive/evaluation. The software platform provides plugin software interfaces that allow for the use of specific commercially available post-processing software algorithms in an unmodified form from the cleared stand-alone post-processing version.
Software version syngo CT VB20 (SOMARIS/10 syngo CT VB20) is a modified software version of the primary predicate device NAEOTOM Alpha, syngo CT VB10 (SOMARIS/10 syngo CT VB10) cleared in K233657.
Software version SOMARIS/10 syngo CT VB20 will be offered ex-factory and as optional upgrade for the existing NAEOTOM Alpha systems.
The bundle approach is feasible for this submission since the subject devices have similar technological characteristics, software operating platform, and supported software characteristics. All subject devices will support previously cleared software and hardware features in addition to the applicable modifications as described within this submission. The intended use remains unchanged compared to the predicate devices.
## V. Indications for Use
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.
The images delivered by the system can be used by a trained staff as an aid in diagnosis, treatment and radiation therapy planning as well as for diagnostic and therapeutic interventions.
This CT system can be used for low dose lung cancer screening in high risk populations*.
* As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a cluster of orange dots that are arranged in a circular pattern.
## VI. Indications for Use Comparison
#### Subject Device Indications for Use:
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.
The images delivered by the system can be used by a trained staff as an aid in diagnosis, treatment and radiation therapy planning as well as for diagnostic and therapeutic interventions.
This CT system can be used for low dose lung cancer screening in high risk populations*.
* As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
#### Primary Predicate Device Indications for Use:
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.
The images delivered by the system can be used by a trained staff as an aid in diagnosis and treatment as well as for diagnostic and therapeutic interventions.
This CT system can be used for low dose lung cancer screening in high risk populations*.
* As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
#### Secondary Predicate Device Indications for Use:
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.
The images delivered by the system can be used by a trained staff as an aid in diagnosis, treatment and radiation therapy planning as well as for diagnostic and therapeutic interventions.
This CT system can be used for low dose lung cancer screening in high risk populations.*
*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
#### Comparison:
The subject device Indications for Use is the exact same as the secondary predicate device Indications for Use (K233650).
Compared to the primary predicate device, the indications for Use of the subject devices have been extended with the phrase "radiation therapy planning" in the sentence "The images delivered by the system can be used by a trained staff as an aid in diagnosis, treatment and radiation therapy planning as well as for diagnostic and therapeutic interventions.". The extension of the indications for use reflects the support for radiotherapy planning introduced with the software version syngo CT VB20 of the subject devices.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern.
None of the intended uses include computed tomography as the principal means of guidance in invasive procedures (involving the introduction of a device, such as a needle or a catheter into the body of the patient).
The subject devices NAEOTOM CT scanner systems are not the principal means of guidance, because the CT Systems do not guide the invasive procedures, the needle orientation and the needle advance and handling are always done under the physician's control.
## VII. Comparison of Technological Characteristics with the Predicate Device
Supported by the subject device, SOMARIS/10 syngo CT VB20 software version is a further development of the SOMARIS/10 syngo CT VB10 software version which is cleared in K233657.
The subject devices NAEOTOM CT scanner systems with SOMARIS/10 syngo CT VB20 software version provides the same technological characteristics in terms of materials, energy source, and control mechanisms when compared to the predicate devices. The software features of the NAEOTOM CT scanner systems have been modified or improved in comparison to the predicate devices to support enhanced device functionality compared to the predicate devices.
The new syngo CT VB20 software reuses all unmodified software features of the legacy software syngo CT VB10 cleared in K233657. Additionally, no features present in the predicate device are descoped.
Software version SOMARIS/10 syngo CT VB20 is designed to reuse hardware independent extended functionalities and GO technologies provided by Siemens cleared software applications.
The intended use and fundamental scientific technology for the NAEOTOM Alpha remain unchanged from the predicate devices.
At a high level, the subject and predicate devices are based on the same subset of technological elements:
- Scanner Principle Whole body X-Ray Computed Tomography Scanner ●
- . System Acquisition - Continuously rotating tube detector system
- . Iterative Reconstruction – Support of various iterative reconstruction principles
- Workplaces – Support of workplaces that include reconstruction and image evaluation software
- . Patient table
- . Patient table foot switch for movement
- QuantaMax Detector
- Tin filtration technology
- Vectron X-ray Tube
- Power Generator
- Scan&GO
- . Mobile workflow (Tablet)
- Optional injector arm
- Optional support of CT guided intervention workflow (myNeedle Guide)
- Optional support of FAST 3D Camera operation for fast patient positioning workflow
- Scanner display and control functionality
- Remote Scan Control
- Long scan range
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a cluster of orange dots of varying sizes.
The subject devices NAEOTOM CT scanner systems with SOMARIS/10 syngo CT VB20 will support hardware and software modifications/further developments in comparison to the predicate devices as listed in the Table 2. The column "HW/ SW change" in the table contains reference to a hardware (HW) or/and software (SW) change. The columns under "Predicate devices" of the Table 2 indicate the clearance of the technological properties in the corresponding predicate device.
The modification overview in Table 2 uses the terms "new, modified, or enabled" to describe various technological characteristics of the subject devices in comparison to the primary and secondary predicate devices. The definition of the above-mentioned terms is provided in the following Table 1.
| Term | Definition |
|----------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| New | The feature is newly supported for Siemens CT scanners and the subject device |
| Modified | This feature is a modified form of a feature cleared within the primary and/or<br>secondary predicate devices |
| Enabled | This feature is currently supported by other cleared Siemens CT systems or<br>cleared Siemens stand-alone software applications. This feature will be<br>supported for the subject device with software version SOMARIS/10 syngo CT<br>VB20 and is substantially equivalent compared to the cleared version of the<br>primary/secondary predicate device. |
| n.a. | The feature is not supported by the subject device or by the primary and/or<br>secondary predicate device. |
Table 1: Overview of term definition.
The following Table 2 provides a modification overview of the subject devices at a high-level.
Table 2: High-level overview of key hardware (HW) and software (SW) modifications of the Dual source subject devices NAEOTOM Alpha (trade name ex-factory: NAEOTOM Alpha.Peak; trade name installed base: NAEOTOM Alpha) and
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a graphic of several orange dots arranged in a pattern.
NAEOTOM Alpha.Pro and the Single source subject device NAEOTOM Alpha.Prime with software version SOMARIS/10 syngo CT VB20 compared to the predicate devices.
| | Technological<br>property | HW/<br>SW<br>change | Subject devices | | | Predicate devices | |
|-----|-----------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|--------------------------------------|------------------------------------------|--------------------------------------------|---------------------------------------------------|----------------------------------------------------|
| | | | Dual Source | | Single<br>source | Dual Source | Single<br>source |
| | | | NAEOTOM<br>Alpha | NAEOTOM<br>Alpha.Pro | NAEOTOM<br>Alpha.Prime | NAEOTOM<br>Alpha | SOMATOM<br>X.ceed |
| | | | syngo CT<br>VB20 | syngo CT<br>VB20 | syngo CT<br>VB20 | syngo CT<br>VB10 | syngo CT<br>VB10 |
| | | | | | | (K233657) | (K233650) |
| 1. | Respiratory Motion<br>Management -<br>Interfaces for<br>respiratory gating:<br>- Varian RGSC<br>- Anzai<br>- Open Interface<br>- Open Online<br>Interface | HW | modified¹ | modified¹ | modified¹ | n.a. | cleared |
| 2. | Respiratory Motion<br>Management -<br>FAST 4D | SW | enabled | enabled | enabled | n.a. | cleared |
| 3. | Respiratory Motion<br>Management -<br>Direct i4D | SW | enabled | enabled | enabled | n.a. | cleared |
| 4. | Respiratory Motion<br>Management -<br>DirectBreathhold | SW | enabled | enabled | enabled | n.a. | cleared |
| 5. | CT View&GO<br>- Sim&GO | SW | modified² | modified² | modified² | cleared | cleared |
| 6. | Direct Laser<br>Steering | HW/<br>SW | enabled | enabled | enabled | n.a. | cleared |
| 7. | Recon&GO | SW | modified² | modified² | modified² | cleared | cleared |
| | Technological<br>property | HW/<br>SW<br>change | Subject devices | | | Predicate devices | |
| | | | NAEOTOM<br>Alpha<br>syngo CT<br>VB20 | NAEOTOM<br>Alpha.Pro<br>syngo CT<br>VB20 | NAEOTOM<br>Alpha.Prime<br>syngo CT<br>VB20 | NAEOTOM<br>Alpha<br>syngo CT<br>VB10<br>(K233657) | SOMATOM<br>X.ceed<br>syngo CT<br>VB10<br>(K233650) |
| | - RT planning | | | | | | |
| 8. | Oncology Exchange | SW | new | new | new | n.a. | n.a. |
| 9. | DirectSetupNotes | SW | enabled3 | enabled3 | enabled3 | n.a. | cleared |
| 10. | FAST 3D Camera/<br>FAST Integrated<br>Workflow | SW/<br>HW | modified4 | modified4 | modified4 | cleared | cleared |
| 11. | Flex 4D Spiral | SW | modified4 | modified4 | modified4 | cleared | cleared |
| 12. | FAST Planning | SW | modified4 | modified4 | modified4 | cleared | cleared |
| 13. | CARE Moodlight | HW | enabled5 | enabled5 | enabled5 | n.a. | n.a. |
| 14. | Eco Power Mode | HW/<br>SW | new | new | new | n.a. | n.a. |
| 15. | Power Shaver | SW/<br>HW | n.a. | n.a. | new | n.a. | n.a. |
| 16. | myExam<br>Companion –<br>myExam<br>Compass/<br>myExam Cockpit | SW | modified4 | modified4 | modified4 | cleared | cleared |
<sup>1</sup> The listed interfaces for respiratory gating are implemented unchanged in the subject devices compared to the secondary predicate device SOMATOM X.ceed with exception of Open Online Interface is not available on the predicate devices and the interface is new for this submission.
<sup>2</sup> In the subject devices NAEOTOM Alpha, NAEOTOM Alpha.Pro, and NAEOTOM Alpha.Prime with syngo CT VB20, the functionalities Recon&GO Inline Results and CT View&GO support the same functions as cleared in the secondary predicate device SOMATOM X.ceed but are modified compared to the primary predicate device NAEOTOM Alpha with software version syngo CT VB10.
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a cluster of orange dots arranged in a circular pattern.
A tabular summary of the comparable hardware and software properties between the subject devices NAEOTOM CT scanner systems with software version syngo CT VB20 and the
Siemens Healthineers AG
<sup>3</sup> DirectSetupNotes is cleared in K233650 for the SOMATOM go.Open Pro. K233650 is a bundle 510(k) including the SOMATOM go. Platform (SOMATOM go.Now, SOMATOM go.Up, SOMATOM go.All, SOMATOM go.Top, SOMATOM go.Sim and SOMATOM go.Open Pro) and the SOMATOM X.cite and SOMATOM X.cite and SOMATOM X.ceed).
<sup>4</sup> These features are modified compared to the same feature cleared in the primary and secondary predicate device.
<sup>5</sup> CARE Moodlight is cleared in K233650 for the SOMATOM go.Now, SOMATOM go.Top and SOMATOM go.All. K233650 is a bundle 510(k) including the SOMATOM go. Platform (SOMATOM go.Now, SOMATOM go.Up, SOMATOM go.All, SOMATOM go.Top, SOMATOM go.Sim and SOMATOM go.Open Pro) and the SOMATOM X. Platform (SOMATOM X.cite and SOMATOM X.ceed).
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a graphic of orange dots.
primary/secondary predicate devices are listed in the following tables (modifications are in gray shaded sections).
Table 3 compares the hardware properties of the dual source subject devices NAEOTOM Alpha (trade name ex-factory: NAEOTOM Alpha.Peak; trade name installed base: NAEOTOM Alpha) and NAEOTOM Alpha.Pro with the primary/secondary predicate devices.
Table 4 provides a comparison of the hardware properties between the single source subject device NAEOTOM Alpha.Prime and the primary/secondary predicate devices.
Table 5 compares the software technological characteristics of the subject devices and the predicate devices.
Table 3: Hardware characteristics for the dual source subject devices NAEOTOM Alpha (trade name ex-factory: NAEOTOM Alpha.Peak; trade name installed base: NAEOTOM Alpha) and NAEOTOM Alpha.Pro with software version SOMARIS/10 syngo CT VB20 compared to the predicate devices.
| Hardware<br>property | Subject device<br>(Dual source NAEOTOM CT systems) | | Primary predicate<br>device | Secondary predicate<br>device |
|------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | NAEOTOM Alpha<br>SOMARIS/10 syngo<br>CT VB20 | NAEOTOM<br>Alpha.Pro<br>SOMARIS/10 syngo<br>CT VB20 | NAEOTOM Alpha<br>SOMARIS/10 syngo<br>CT VB10<br>(K233657) | SOMATOM X.ceed<br>SOMARIS/10 syngo<br>CT VB10<br>(K233650) |
| Scanner | whole body X-ray<br>computed<br>tomography (CT)<br>scanner | whole body X-ray<br>computed<br>tomography (CT)<br>scanner | whole body X-ray<br>computed<br>tomography (CT)<br>scanner | whole body X-ray<br>computed<br>tomography (CT)<br>scanner |
| Product<br>name | NAEOTOM Alpha | NAEOTOM<br>Alpha.Pro | NAEOTOM Alpha | SOMATOM X.ceed |
| Trade name | • Ex-factory CT<br>systems:<br>NAEOTOM<br>Alpha.Peak<br>• Installed base CT<br>systems:<br>NAEOTOM Alpha | • Ex-factory CT<br>systems:<br>NAEOTOM<br>Alpha.Pro<br>• Installed base CT<br>systems:<br>n.a., new CT system | NAEOTOM Alpha | SOMATOM X.ceed |
| Design | without myNeedle<br>Laser:<br>Image: CT scanner | without myNeedle<br>Laser:<br>Image: CT scanner | without myNeedle<br>Laser:<br>Image: CT scanner | without myNeedle<br>Laser:<br>Image: CT scanner<br>with myNeedle Laser: |
| Hardware<br>property | Subject device<br>(Dual source NAEOTOM CT systems) | | Primary predicate<br>device | Secondary predicate<br>device |
| | NAEOTOM Alpha<br>SOMARIS/10 syngo<br>CT VB20 | NAEOTOM<br>Alpha.Pro<br>SOMARIS/10 syngo<br>CT VB20 | NAEOTOM Alpha<br>SOMARIS/10 syngo<br>CT VB10<br>(K233657) | SOMATOM X.ceed<br>SOMARIS/10 syngo<br>CT VB10<br>(K233650) |
| | with myNeedle<br>Laser:<br>Image: CT scanner | with myNeedle<br>Laser:<br>Image: CT scanner | with myNeedle<br>Laser:<br>Image: CT scanner | Image: CT scanner |
| System<br>configuration | Dual Source | Dual Source | Dual Source | Single Source |
| Environment<br>of Use | Professional<br>Healthcare Facility | Professional<br>Healthcare Facility | Professional<br>Healthcare Facility | Professional<br>Healthcare Facility |
| Generator<br>max. power<br>(kW) | 2x 120 | 2x 120 | 2x 120 | 105 or 120 |
| Detector<br>technology | QuantaMax<br>direct conversion<br>with "Quantum<br>Technology" | QuantaMax<br>direct conversion<br>with "Quantum<br>Technology" | QuantaMax<br>direct conversion<br>with "Quantum<br>Technology" | StellarInfinity<br>UFC (Ultra Fast<br>Ceramic) |
| Detector<br>volume<br>coverage<br>(mm) | 2x 57.6 | 2x 38.4 | 2x 57.6 | 38.4 |
| Detector<br>physical rows | 2x 288 | 2x 192 | 2x 288 | 64 |
| Detector<br>slice width<br>(mm) | 0.2 | 0.2 | 0.2 | 0.6 |
| Detector<br>DAS channel<br>no. | 2752 (A system)<br>1984 (B system) | 2752 (A system)<br>1984 (B system) | 2752 (A system)<br>1984 (B system) | 920 |
| Tube | VECTRON | VECTRON | VECTRON | VECTRON |
| Hardware<br>property | Subject device<br>(Dual source NAEOTOM CT systems) | | Primary predicate<br>device | Secondary predicate<br>device |
| | NAEOTOM Alpha<br>SOMARIS/10 syngo<br>CT VB20 | NAEOTOM<br>Alpha.Pro<br>SOMARIS/10 syngo<br>CT VB20 | NAEOTOM Alpha<br>SOMARIS/10 syngo<br>CT VB10<br>(K233657) | SOMATOM X.ceed<br>SOMARIS/10 syngo<br>CT VB10<br>(K233650) |
| technology | | | | |
| Tube<br>kV steps | 70, 90, 100, 120,<br>140, 150<br>(150 kV only<br>available on the<br>smaller tube-<br>detector system (B<br>system) and only in<br>combination with<br>the additional Sn<br>filter, 0.7 mm) | 70, 90, 100, 120,<br>140, 150<br>(150 kV only<br>available on the<br>smaller tube-<br>detector system (B<br>system) and only in<br>combination with<br>the additional Sn<br>filter, 0.7 mm) | 70, 90, 100, 120,<br>140, 150<br>(150 kV only<br>available on the<br>smaller tube-<br>detector system (B<br>system) and only in<br>combination with<br>the additional Sn<br>filter, 0.7 mm) | 70, 80, 90, 100, 110,<br>120, 130, 140, 150 |
| Tube<br>max. current<br>(mA) | 2x 1300 | 2x 1300 | 2x 1300 | 1200 mA (for 105<br>kW)<br>1300 mA (for 120<br>kW) |
| Tube<br>tube focus<br>(mm) | 0.4 x 0.5/8°<br>0.6 x 0.7/8°<br>0.8 x 1.1/8°<br>(for both tubes) | 0.4 x 0.5/8°<br>0.6 x 0.7/8°<br>0.8 x 1.1/8°<br>(for both tubes) | 0.4 x 0.5/8°<br>0.6 x 0.7/8°<br>0.8 x 1.1/8°<br>(for both tubes) | 0.4 x 0.5/8°<br>0.6 x 0.7/8°<br>0.8 x 1.1/8° |
| Tube<br>heat storage<br>capacity<br>(MHU) | higher than 30 | higher than 30 | higher than 30 | higher than 30 |
| Tube<br>cooling rate<br>(MHU/min) | 2.7 | 2.7 | 2.7 | 2.7 |
| Gantry<br>bore size<br>(cm) | 82 | 82 | 82 | 82 |
| Gantry<br>Scan FoV<br>(cm) | 50 | 50 | 50 | 50 |
| Gantry | 81.5 | 81.5 | 81.5 | 81.5 |
| Hardware<br>property | Subject device<br>(Dual source NAEOTOM CT systems) | | Primary predicate<br>device | Secondary predicate<br>device |
| | NAEOTOM Alpha<br>SOMARIS/10 syngo<br>CT VB20 | NAEOTOM<br>Alpha.Pro<br>SOMARIS/10 syngo<br>CT VB20 | NAEOTOM Alpha<br>SOMARIS/10 syngo<br>CT VB10<br>(K233657) | SOMATOM X.ceed<br>SOMARIS/10 syngo<br>CT VB10<br>(K233650) |
| Extended FoV<br>(cm) | | | | |
| Gantry<br>rotation time<br>(sec) | 0.25, 0.5, 1.0 | 0.25, 0.5, 1.0 | 0.25, 0.5, 1.0 | 0.25, 0.3, 0.5, 1.0 |
| Gantry<br>Tilt (degree) | N/A | N/A | N/A | +/- 25<br>+/-30 (only with<br>Vitus) |
| Maximum<br>temporal<br>resolution in<br>ECG gated or<br>triggered<br>examination<br>(ms) | mono- mono-<br>segment: 66<br>bi-segment: 33 | mono- mono-<br>segment: 66<br>bi-segment: 33 | mono-segment: 66<br>bi-segment: 33 | mono-segment: 125<br>bi-segment: 66 |
| Maximum<br>scan speed at<br>pitch<br>(mm/s at<br>pitch x) | 737 mm/s at pitch<br>3.2 | 492 mm/s at pitch<br>3.2 | 737 mm/s at pitch<br>3.2 | 261 at pitch 1.7…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.