K243521 · BioHorizons Implant Systems, Inc. · NHA · May 1, 2025 · Dental
Device Facts
Record ID
K243521
Device Name
Conical Ti Base abutments
Applicant
BioHorizons Implant Systems, Inc.
Product Code
NHA · Dental
Decision Date
May 1, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
BioHorizons conical dental prosthetic components connected to the endosseous dental implants are intended for use as an aid in prosthetic rehabilitations of the maxillary or mandibular arch to provide support for prosthetic restorations. All digitally designed abutments for use with Conical Ti Base abutments are to be sent to a BioHorizons validated milling center for manufacture or to be designed and manufactured according to digital dentistry workflow. The workflow system integrates multiple components of the digital dentistry workflow: scan files from Intra-Oral Scanners, CAD software, CAM software, milling machine and associated tooling and accessories.
Device Story
Conical Ti Base abutments are two-piece dental prosthetic components consisting of a pre-manufactured Ti-6Al-4V titanium base and a CAD/CAM-designed zirconia superstructure. The device is used by clinicians to restore dental implants. The workflow involves intra-oral scanning, CAD design (3Shape Abutment Designer), and CAM milling (imes-icore CORiTEC 150i Pro). The zirconia superstructure is bonded to the titanium base using dual-cure resin cement (3M RelyX Unicem 2). The final assembly is screwed onto the dental implant. The device supports single or multi-unit restorations. It is provided non-sterile and requires sterilization at the dental clinic before placement. The system allows for customized abutment fabrication via a digital dentistry workflow, benefiting patients by providing patient-specific prosthetic support.
Clinical Evidence
Bench testing only. No clinical data included. Performance evaluation included mechanical testing, biological safety, steam sterilization, and MRI compatibility, relying on predicate data. New testing supported the CAD/CAM digital dentistry workflow, including surface characterization (SEM/EDX) of the titanium base to ensure no residual blast media and verification that the digital design process produces abutments equivalent to the validated milling center workflow.
Indicated for patients requiring prosthetic rehabilitation of the maxillary or mandibular arch via endosseous dental implants to provide support for prosthetic restorations.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
BioHorizons Implant Systems Inc.
Jared Cooper
Sr. Director, Regulatory Affairs
2300 Riverchase Center
Birmingham, Alabama 35244
May 1, 2025
Re: K243521
Trade/Device Name: Conical Ti Base abutments
Regulation Number: 21 CFR 872.3630
Regulation Name: Endosseous Dental Implant Abutment
Regulatory Class: Class II
Product Code: NHA, PNP
Dated: January 31, 2025
Received: February 3, 2025
Dear Jared Cooper:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243521 - Jared Cooper
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K243521 - Jared Cooper
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Andrew I. Steen -S
Andrew I. Steen
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K243521 | |
| Device Name Conical Ti Base abutments | |
| Indications for Use (Describe) BioHorizons conical dental prosthetic components connected to the endosseous dental implants are intended for use as an aid in prosthetic rehabilitations of the maxillary or mandibular arch to provide support for prosthetic restorations. All digitally designed abutments for use with Conical Ti Base abutments are to be sent to a BioHorizons validated milling center for manufacture or to be designed and manufactured according to digital dentistry workflow. The workflow system integrates multiple components of the digital dentistry workflow: scan files from Intra-Oral Scanners, CAD software, CAM software, milling machine and associated tooling and accessories. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Premarket Notification
Conical Ti Base abutments
Page 1 of 7
510(k) Summary – K243521
BioHorizons Implant Systems Inc.
Conical Ti Base abutments
April 30, 2025
## ADMINISTRATIVE INFORMATION
Manufacturer Name: BioHorizons Implant Systems Inc.
2300 Riverchase Center
Birmingham, AL 35244
Telephone: +1 205-967-7880
Fax: +1 205-870-0304
Official Contact: Jared Cooper, Ph.D.,
Sr. Director, Regulatory Affairs
jcooper@biohorizons.com
## DEVICE NAME AND CLASSIFICATION
Trade/Proprietary Name: Conical Ti Base abutments
Common Names: Endosseous dental implant abutment
Regulation Number: 21 CFR 872.3630
Regulation Name: Endosseous dental implant abutment
Regulatory Class: Class 2
Product Code: NHA
Associated Product Code: PNP
Classification Panel: Dental
510(k) Submission Number: K243521
Reviewing Office: Office of Health Technology 1 (Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices)
Reviewing Division: Division of Health Technology 1B (Dental and ENT Devices)
## PREDICATE DEVICE INFORMATION
Primary Predicate Device: K240187, Tapered Pro Conical Implant System, BioHorizons Implant Systems Inc.
Secondary Predicate Device: K231307, Elos Accurate® Customized Abutment
{5}
510(k) Premarket Notification
Conical Ti Base abutments
Page 2 of 7
# INDICATIONS FOR USE STATEMENT
BioHorizons conical dental prosthetic components connected to the endosseous dental implants are intended for use as an aid in prosthetic rehabilitations of the maxillary or mandibular arch to provide support for prosthetic restorations.
All digitally designed abutments for use with Conical Ti Base abutments are to be sent to a BioHorizons validated milling center for manufacture or to be designed and manufactured according to digital dentistry workflow. The workflow system integrates multiple components of the digital dentistry workflow: scan files from Intra-Oral Scanners, CAD software, CAM software, milling machine and associated tooling and accessories.
# SUBJECT DEVICE DESCRIPTION
The BioHorizons Conical Ti Base abutments are prosthetic components and are intended for the restoration of BioHorizons dental implants within the specific indications of each implant system. All Conical Ti Base abutments are two-piece titanium-base type abutments with a pre-manufactured titanium base component cemented to a zirconia superstructure to create the final finished dental abutment. All the titanium bases are manufactured from Ti-6Al-4V titanium alloy per ASTM F136, Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy (UNS R56401) for Surgical Implant Applications and are for single patient use. The prosthetics are provided non-sterile as indicated on the label. The subject devices are permanent or long-term, tissue/bone implant devices intended for more than 30 days of patient contact. BioHorizons dental prosthetics are invasive and / or surgically invasive devices placed in healed or compromised oral sites in direct contact with bone and soft tissue.
The Conical Ti-Base abutments are provided in engaging and non-engaging designs and are designed for single-unit or multi-unit restoration with a CAD/CAM zirconia superstructure attached to the abutment. The titanium base components are available for narrow and regular platforms and with gingival heights of 0.8 mm and 2.0 mm. The titanium base component post has been manufactured with matte finish creating a roughened surface to accommodate cementing. All patient-specific custom abutment fabrication for Conical Ti Base abutments are by prescription on the order of the clinician. All zirconia superstructures for use with the subject device Conical Ti Base abutments are made at a BioHorizons validated milling center under FDA quality system regulations, as cleared in K240187, or according to a digital dentistry workflow (subject of this submission described in the section below). As cleared in K240187, zirconia material, sagemax NexxZr® zirconia (K130991) is used to fabricate the patient-specific zirconia superstructure or direct crown. The cement for bonding of superstructures or crowns is a dual cure cement, 3M™ RelyX™ Unicem 2 Automix (K022476). These cleared materials are identical to the proposed digital dentistry workflow. The pre-manufactured titanium base component are not to be modified and are only intended to permit a customized zirconia restoration.
{6}
510(k) Premarket Notification
Conical Ti Base abutments
Page 3 of 7
The design and fabrication of the zirconia superstructure will be conducted using a digital dentistry workflow requiring the use of the following equipment:
- Scanner: 3Shape Trios 5 intra-oral scanner.
- Design Software: 3Shape Abutment Designer Software, K151455.
- Zirconia Material: sagemax® NexxZr zirconia, K130991.
- CAM software: hyperDENT® Classic
- Milling machine: imes-icore® CORiTEC 150i Pro milling machine
- Cement: 3M™ RelyX™ Unicem 2 Automix Self-Adhesive Resin Cement, K022476.
The designed superstructure is attached to the titanium base component by the use of an FDA-cleared cement, 3M™ RelyX™ Unicem 2 Automix Self-Adhesive Resin Cement, K022476. The resulting final prosthetic restoration is screwed to the dental implant. All subject abutments are single-use and provided non-sterile. Final restoration (which includes the corresponding screw) is intended to be sterilized at the dental clinic before it is placed in the patient. The final abutment is equivalent to the final abutment cleared in K240187.
## MATERIAL COMPOSITION
There have been no changes to the subject devices since previous clearance, K240187. All titanium subject device materials are Ti-6Al-4V per ASTM F136. The recommended zirconia superstructure materials are identical to the predicate materials in K240187. These materials are FDA cleared and have a long history of use in dental applications and no new device materials are being introduced in the subject submission.
## PERFORMANCE DATA
Non-clinical data including mechanical testing, biological safety, steam sterilization, and magnetic resonance testing relied upon information in the predicate submission since there were no changes since the previous clearance. New testing reviewed under this submission included supporting the CAD/CAM design and use within in a digital dentistry workflow and demonstrating controls of the digital dentistry workflow and that the final abutment is equivalent to the abutment produced by the validated milling center cleared in K240187.
Additionally, surface characterization using scanning electron microscopy (SEM) and elemental analysis using energy-dispersive X-ray spectroscopy (EDX) was conducted on the titanium base component with a matte post. The surface analysis demonstrated that no residual blast media was detected on the surface of the titanium base component using a post-blast cleaning process. No clinical data were included in this submission. BioHorizons assessed the output of the proposed digital dentistry workflow including the CAD/CAM design of the zirconia superstructure and
{7}
510(k) Premarket Notification
Conical Ti Base abutments
Page 4 of 7
determined the abutment angulation, diameter, height, wall thickness, tolerances, and part quality are equivalent to the validated milling center workflow cleared in K240187.
## EQUIVALENCE TO MARKETED DEVICES
The subject device is substantially equivalent in intended use, design, and technology to the primary predicate device and to the use within a digital dentistry workflow as the secondary predicate listed above. The substantial equivalence comparison table included below compares the Indications for Use Statements and the technological characteristics of the subject device, the primary predicate device, and the secondary predicate devices.
## Substantial Equivalence of Indication for Use Statement (IFUs)
The subject devices are similar in intended use to the primary predicate device cleared in K240187 and adds indications for digital dentistry workflows from the secondary predicate, K231307.
## Substantial Equivalence of Technological Characteristics
All subject devices are similar to or identical in design, materials and technological characteristics to corresponding abutments of the primary predicate device K240187.
The titanium base components are provided non-sterile, packaging consists of 1073B Tyvek® and 48ga polyethylene terephthalate (PET)/2mil low density polyethylene (LDPE) pouches. This packaging is identical to that used for corresponding products cleared in K240187.
Overall, the subject device has the following similarities to the predicate devices:
- have the same intended use,
- use the same operating principles,
- incorporate the same basic designs,
- incorporate the same materials, and
- have the same packaging and are cleaned and sterilized using the same materials and processes.
- have the same use within a digital dentistry workflow
The basis that the subject device is substantially equivalent to the predicate devices is summarized in the following table.
Substantial Equivalence – Conical Ti Base abutments
| Attributes | Subject Device | Primary Predicate Device | Predicate Device |
| --- | --- | --- | --- |
| Trade Name | Conical Ti Base abutments | K240187- Tapered Pro Conical Implant System (Conical Ti Base abutments) | K231307- Elos Accurate Customized Abutment |
| Reason for selection | Subject Device | Primary Predicate for Intended use, Design, technology, materials, manufacturing, packaging, and sterilization | Predicate for Digital Dentistry Workflow Indications and Directions for use |
| Indications for Use | BioHorizons conical dental prosthetic components connected to the endosseous dental implants | BioHorizons conical dental prosthetic components connected to the endosseous dental implants are | The Elos Accurate® Customized Abutments are intended for attaching to dental implants in order to provide basis for single or multiple tooth prosthetic |
{8}
510(k) Premarket Notification
Conical Ti Base abutments
Page 5 of 7
| | are intended for use as an aid in prosthetic rehabilitations of the maxillary or mandibular arch to provide support for prosthetic restorations.
All digitally designed abutments for use with Conical Ti Base abutments are to be sent to a BioHorizons validated milling center for manufacture or to be designed and manufactured according to digital dentistry workflow. The workflow system integrates multiple components of the digital dentistry workflow: scan files from Intra-Oral Scanners, CAD software, CAM software, milling machine and associated tooling and accessories.
(Substantially equivalent to predicate) | intended for use as an aid in prosthetic rehabilitations of the maxillary or mandibular arch to provide support for prosthetic restorations.
All digitally designed abutments for use with Conical CAD/CAM Ti Blanks and Ti Bases are to be sent to a BioHorizons validated milling center for manufacture. | restorations. The Elos Accurate® Customized Abutment will be attached to a dental implant using the included Elos Prosthetic screw.
The Elos Accurate® Customized Abutments are compatible with the implant systems listed in Table 1: |
| --- | --- | --- | --- |
| Table 1: |
| Platform Compatibility | Platform diameter [mm] | Implant Body diameter [mm] |
| Zimmer Screw-vent 3.5 | Ø3.5 | Ø3.7/Ø4.1 |
| Zimmer Screw-vent 4.5 | Ø4.5 | Ø4.7 |
| Zimmer Screw-vent 5.7 | Ø5.7 | Ø6.0 |
| Biomet 3i Certain 3.4 | Ø3.4 | Ø3.25 |
| Biomet 3i Certain 4.1 | Ø4.1 | Ø4 |
| Biomet 3i Certain 5.0 | Ø5 | Ø5 |
| Biomet 3i Certain 6.0 | Ø6 | Ø6 |
| Straumann Standard RN | Ø4.8 | Ø3.3/Ø4.1/Ø4.8 |
| Straumann Standard WN | Ø6.5 | Ø4.8 |
| Neodent GM | Ø3.5/Ø4.5/Ø5.5/Ø 6.5 | Ø3.5/Ø3.75/Ø4/Ø4 .3/Ø 5/Ø 6/Ø7 |
| Hiossen ET Mini | Ø3.2/Ø3.5 | Ø3.2/Ø3.5 |
| Hiossen ET Regular | Ø4/Ø4.5/Ø5/Ø5.5/Ø6/Ø7 | Ø4/Ø4.5/Ø5/Ø5.5/Ø6/Ø7 |
| All digitally designed CAD/CAM customizations for the Elos Accurate® Customized Abutments are either intended to be sent and manufactured at a FDA registered Elos Medtech approved milling facility or to be designed and manufactured according to digital dentistry workflow. The workflow system integrates multiple components of the digital dentistry workflow: scan files from Intra-Oral Scanners, CAD software, CAM software, milling machine and associated tooling and accessories. |
| Stock Gingival Height, mm | 0.8, 2.0
(Identical Stock Gingival Height to primary predicate) | 0.8, 2.0 | N/A |
| Angulation | up to 20°
(Identical Angulation to primary predicate) | up to 20° | Up to 30° maximum |
| Connection | Engaging, Non-engaging
(Identical Connection to primary predicate) | Engaging, Non-engaging | Engaging, Non-engaging |
| Material (titanium base component) | Ti-6Al-4V Alloy
(Identical Material to primary predicate) | Ti-6Al-4V Alloy | Ti-6Al-4V Alloy |
| Surface Treatment (Ti Base) | Anodized, Blast
(Similar Surface Treatment to primary predicate) | Anodized, None | N/A |
| Method of Manufacturing (Pre-manufactured Ti Base) | Traditional manufacturing methods (machine, clean, anodize (if applicable), Blast (engaging), package, label.
(Similar Method of Manufacturing to primary predicate) | Traditional manufacturing methods (machine, clean, anodize (if applicable), package, label. | N/A |
| Digital Design (Zirconia Superstructure) | 3Shape Trios 5 intra-oral scanner, Design Software: 3Shape | Design Software: 3Shape Abutment Designer Software, K151455 | 3Shape scanner (3Shape A/S), 3Shape Abutment Designer Software (3Shape A/S) - K151455 |
{9}
510(k) Premarket Notification
Conical Ti Base abutments
Page 6 of 7
| | Abutment Designer Software, K151455 | | |
| --- | --- | --- | --- |
| **Material (Zirconia Superstructure)** | sagemax® NexxZr zirconia (K130991) bonded using 3M™ RelyX™ Unicem 2 Automix Self-Adhesive Resin Cement (K022476) or similar
(Identical Superstructure to primary predicate) | sagemax® NexxZr zirconia (K130991) bonded using 3M™ RelyX™ Unicem 2 Automix Self-Adhesive Resin Cement (K022476) or similar | 3M™ ESPE™ Lava™ Plus High Translucency Zirconia |
| **Manufacturing (Zirconia Superstructure)** | Digital Dentistry Workflow or Validated Milling Center | Validated Milling Center | Digital Dentistry Workflow or Validated Milling Center |
| **Method of Manufacturing (Zirconia Superstructure)** | imes-icore® CORiTEC 150i Pro milling machine
(Identical Method of Manufacturing to primary predicate) | imes-icore® CORiTEC 150i Pro milling machine | CORiTEC milling unit (imes-icore) |
| **Design Limitations** | • 20° maximum post angulation.
• 0.4mm minimum wall thickness.
• 5mm maximum superstructure gingival margin height.
Note: The maximum gingival margin height is inclusive of the gingival height of the stock titanium base component. A CAD library shall have a less-than-5mm maximum gingival margin height design limitation if measured in the software from the gingival height of the stock titanium base component.
• Abutments with a superstructure post height less than 4.0mm are intended for multi-unit restorations only
• Screw channel:
• Engaging titanium base components:
• 25° maximum angled screw channel for 0.8mm margin height
• 15° maximum angled screw channel for 2.0mm margin height
• Non-engaging titanium base components: straight screw channel only.
Note: Minimal gingival margin height is established by the gingival height of the premanufactured titanium base component. The post-height of the abutment is measured above the total gingival margin height of the final patient-matched design.
(Similar Design Limitations to primary predicate) | • 20° maximum post angulation.
• 0.4mm minimum wall thickness.
• 5mm maximum superstructure gingival margin height.
• Abutments with a superstructure post height less than 4.0mm are intended for multi-unit restorations only
• Screw channel: 25° maximum angled screw channel for 0.8mm margin height, 15° maximum angled screw channel for 2.0mm margin height.
(Minimum gingival margin height is established by the gingival height of the stock Ti Base Abutment.)
NOTE: The post-height of the abutment is measured above the total gingival margin height of the final patient-matched design | The Elos Accurate library file has built-in design limitations and the user isn't allowed to exceed these limitations. The material thickness should not be less than 0.5 mm (screw hole to outer surface). The gingival height should not exceed 5 mm. The maximum angulation should not exceed 20°. The post height should not be less than 4 mm for single unit restorations. |
| **Usage** | Single patient, Single use
(Identical Usage to primary predicate) | Single patient, Single use | Single patient, Single use |
| **Sterility** | Non-Sterile, Steam sterilization
(Identical Sterility to primary predicate) | Non-Sterile, Steam sterilization | Non-Sterile, Steam sterilization |
| **Shelf-Life** | N/A | N/A | N/A |
{10}
510(k) Premarket Notification
Conical Ti Base abutments
Page 7 of 7
| | (Identical Shelf-Life to primary predicate) | | |
| --- | --- | --- | --- |
| MRI Compatibility | MR Conditional
(Identical MRI Compatibility to primary predicate) | MR Conditional | MR Conditional |
| Biocompatibility | Biocompatible per ISO 10993-1
(Identical Biocompatibility to primary predicate) | Biocompatible per ISO 10993-1 | Biocompatible per ISO 10993-1 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.