← Product Code [FDS](/productcode/FDS) · K243512

# FUJIFILM Endoscope Model EG-840N; FUJIFILM Endoscope Model EG-840T; FUJIFILM Endoscope Model EG-840TP (K243512)

_Fujifilm Corporation · FDS · Jun 10, 2025 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K243512

## Device Facts

- **Applicant:** Fujifilm Corporation
- **Product Code:** [FDS](/productcode/FDS.md)
- **Decision Date:** Jun 10, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

FUJIFILM Endoscope Models EG-840T and EG-840TP is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum. FUJIFILM Endoscope Model EG-840N is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum. This product can be inserted orally or nasally. Never use this product for any other purposes.

## Device Story

Flexible endoscopes (EG-840N, EG-840T, EG-840TP) used for upper gastrointestinal tract visualization, diagnosis, and treatment. Device features insertion tube with bending section, forceps channel, and distal tip containing CMOS image sensor, objective lens, and light guide. Operates by connecting to external light source and video processor; light travels through fiber bundles to illuminate target tissue; reflected light forms image on CMOS sensor; electrical signals transmitted to processor for monitor display. Operated by clinicians in clinical settings. Provides real-time visualization to guide diagnostic or therapeutic procedures. Benefits include minimally invasive access to upper digestive tract for observation and intervention.

## Clinical Evidence

Bench testing only. Performance validated against specifications including field of view, bending capability, resolution, suction rate, and illumination. Biocompatibility evaluated per ISO 10993-1, -5, -10. Electrical safety tested per ANSI/AAMI ES 60601-1 and IEC 60601 series. Reprocessing validated per FDA guidance.

## Technological Characteristics

Flexible endoscope with CMOS image sensor, fiber optic light guides, and mechanical bending section. Materials evaluated per ISO 10993. Connectivity via scope connector to external light source/processor. Software validated per IEC 62304. Standards: ISO 8600-1, -3, -4; ANSI/AAMI ES 60601-1; IEC 60601-1-2, -1-6, -2-18.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- FUJIFILM Endoscope Model EG-840N; FUJIFILM Endoscope Model EG-840TP ([K240142](/device/K240142.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

June 10, 2025

FUJIFILM Corporation
% Chaitrali Kulkarni
Sr. Regulatory Affairs Specialist
Fujifilm Healthcare Americas Corporation
81 Hartwell Ave. Suite 300
Lexington, Massachusetts 02421

Re: K243512
Trade/Device Name: FUJIFILM Endoscope Model EG-840N; FUJIFILM Endoscope Model EG-840T; FUJIFILM Endoscope Model EG-840TP
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope And Accessories
Regulatory Class: Class II
Product Code: FDS
Dated: May 9, 2025
Received: May 9, 2025

Dear Chaitrali Kulkarni:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K243512 - Chaitrali Kulkarni
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K243512 - Chaitrali Kulkarni
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shanil P. Haugen -S

Shanil P. Haugen, Ph.D.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K243512  |   |
|  Device Name FUJIFILM Endoscope Model EG-840N; FUJIFILM Endoscope Model EG-840T; FUJIFILM Endoscope Model EG-840TP  |   |
|  Indications for Use (Describe) FUJIFILM Endoscope Models EG-840T and EG-840TP is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum. FUJIFILM Endoscope Model EG-840N is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum. This product can be inserted orally or nasally. Never use this product for any other purposes.  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

K243512 Page 1 of 3

|  510(k) #: K243512 | 510(k) Summary | Prepared on: 2025-06-05  |
| --- | --- | --- |
|  Contact Details |   | 21 CFR 807.92(a)(1)  |
|  Applicant Name | FUJIFILM Corporation |   |
|  Applicant Address | 798 Miyanodai Kaisei-Machil Ashigara Kami-Gun KANAGAWA 258-8538 Japan |   |
|  Applicant Contact Telephone | 704-517-4886 |   |
|  Applicant Contact | Ms. Chaitrali Kulkarni |   |
|  Applicant Contact Email | hcusregulatoryaffairs@fujifilm.com |   |
|  Correspondent Name | Fujifilm Healthcare Americas Corporation |   |
|  Correspondent Address | 81 Hartwell Ave. Suite 300 Lexington MA 02421 United States |   |
|  Correspondent Contact Telephone | 704-517-4886 |   |
|  Correspondent Contact | Ms. Chaitrali Kulkarni |   |
|  Correspondent Contact Email | hcusregulatoryaffairs@fujifilm.com |   |
|  Device Name |   | 21 CFR 807.92(a)(2)  |
|  Device Trade Name | FUJIFILM Endoscope Model EG-840N; FUJIFILM Endoscope Model EG-840T; FUJIFILM Endoscope Model EG-840TP |   |
|  Common Name | Endoscope and accessories |   |
|  Classification Name | Gastroscope And Accessories, Flexible/Rigid |   |
|  Regulation Number | 876.1500 |   |
|  Product Code(s) | FDS |   |
|  Legally Marketed Predicate Devices |   | 21 CFR 807.92(a)(3)  |
|  Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code  |
|  K240142 | FUJIFILM Endoscope Model EG-840N; FUJIFILM Endoscope Model EG-840TP | FDS  |
|  Device Description Summary |   | 21 CFR 807.92(a)(4)  |
|  A. Intended Use/indications for Use  |   |   |
|  FUJIFILM Endoscope Model EG-840T, EG-840TP is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.  |   |   |
|  FUJIFILM Endoscope Model EG-840N is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum. This product can be inserted orally or nasally. Never use this product for any other purposes.  |   |   |

{5}

K243512 Page 2 of 3

## B. Technological Characteristics

The insertion portion of the device has a mechanism (hereinafter “the bending portion”) which bends the tip from right to left and up and down, and a flexible tube (hereinafter “the flexible portion”) consists of the bending portion and operating portion with a knob which controls the bending portion. The forceps channel which runs through the operating portion to the tip is arranged inside the insertion portion for inserting the surgical instrument. The insertion portion of the endoscopes comes into contact with the mucosal membrane.

The tip of the insertion portion is called the “Distal end” which contains the Imaging section, Distal cap, Objective lens, Air/water nozzle, Water jet nozzle (Except EG-840N), Instrument channel outlet, and Light guide. The bending portion is controlled by knobs on the control portion/operation section to angulate the distal end to certain angles.

The Flexible portion refers to the long insertion area between the Bending portion and the Control portion (a part of Non-insertion portion). This portion contains light guides (glass fiber bundles), air/water channels, a forceps/suction channel, a CMOS image sensor, and cabling. The glass fiber bundles allow light to travel through the endoscope to illuminate the body cavity, thereby providing enough light to the CMOS image sensor to capture an image and display the image on a monitor. The forceps channel is used to introduce biopsy forceps and other endoscopic accessories, as well as providing suction.

The control portion/operating section provides a grip to grasp the endoscopes and contains mechanical parts to operate the endoscopes. This section includes a Forceps inlet, which allows endoscope accessories to be introduced. The Scope connector connects the endoscopes to the light source.

## C. Principles of Operation

The Endoscope Model EG-840N, EG-840T, and EG-840TP function on the same principles of operation as the predicate device. EG-840N, EG-840T, and EG-840TP are endoscopes for observation, diagnosis, and treatment of the upper digestive tract. After connecting to the light source and video processor, the devices guide the light through the fiber bundles situated inside the insertion portion of the device. The light that emits from the distal end of the insertion portion reflects from the target region and forms an image on the CMOS image sensor through a group of object lenses placed at the distal end of the devices. An electric signal from the image sensor is transmitted to the video processor connected to the device. The video processor converts the electric signal into a video signal and displays an image on a monitor.

|  **Intended Use/Indications for Use** | **21 CFR 807.92(a)(5)**  |
| --- | --- |
|  FUJIFILM Endoscope Models EG-840T and EG-840TP is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum. |   |
|  FUJIFILM Endoscope Model EG-840N is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum. This product can be inserted orally or nasally. Never use this product for any other purposes. |   |
|  **Indications for Use Comparison** | **21 CFR 807.92(a)(5)**  |
|  The indication for use for the proposed device and the predicate devices are same. |   |
|  **Technological Comparison** | **21 CFR 807.92(a)(6)**  |
|  The subject devices FUJIFILM Endoscope Model EG-840N, EG-840T and EG-840TP are substantially equivalent to the predicate devices based on the same intended use, indications for use, technological characteristics and materials other than the delayed reprocessing instructions and software version. |   |
|  The differences in the delayed reprocessing instructions raise no new issues of safety or effectiveness. Cleaning, Disinfection, and Sterilization (hereinafter “CDS”) validation testing data demonstrated that the subject devices are substantially equivalent in performance to the predicate devices. |   |
|  The difference in software version has been validated through Software & Cybersecurity evaluation testing. |   |
|  Thus, the subject devices FUJIFILM Endoscope Model EG-840N, EG-840T and EG-840TP are substantially equivalent to the predicate device, FUJIFILM Endoscope Model EG-840N, EG-840T and EG-840TP (K240142). |   |

{6}

K243512 Page 3 of 3

# Non-Clinical and/or Clinical Tests Summary &amp; Conclusions 21 CFR 807.92(b)

Electrical safety of the subject device was evaluated using following standards: ANSI/AAMI ES 60601-1:2012, IEC 60601-1-2:2020, IEC60601-1-6:2020 and IEC 60601-2-18:2009.

Biocompatibility of the subject device was evaluated using the following consensus standards: ISO 10993-1:2018, ISO 10993-5:2009, and ISO 10993-10:2010. Biocompatibility testing was conducted in accordance with the FDA guidance, Use of International Standards ISO 10993-1, "Biological evaluation of medical devices - Part 1 : Evaluation and testing within a risk management process", issued June 16,2016.

Endoscope specific testing was conducted using the following consensus standards: ISO 8600-1:2015, ISO 8600-3:2019, and ISO 8600-4:2014.

Software specific testing was conducted using the following consensus standards: IEC 62304:2015. The software validation activities were performed in accordance with the FDA Guidance, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, issued on May 11, 2005.

Validation of the cleaning, disinfection, and sterilization instructions was performed in accordance with FDA's guidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling," published March 17,2015.

The subject device met performance specifications in the following additional testing:

- Field of view
- Diameter of forceps channel
- Uneven illumination
- Bending capability
- Viewing direction
- Color reproducibility
- Rate of suction
- Resolution
- Air volume
- Working length
- LG output
- Water volume

The subject devices FUJIFILM Endoscope Model EG-840N, EG-840T and EG-840TP are substantially equivalent to the predicate devices based on the same intended use, indications for use, technological characteristics and materials other than the delayed reprocessing instructions and software version.

The differences in the delayed reprocessing instructions raise no new issues of safety or effectiveness. CDS validation testing data demonstrated that the subject devices are substantially equivalent in performance to the predicate devices. The difference in software version has been validated through Software &amp; Cybersecurity evaluation testing.

Thus, the subject devices FUJIFILM Endoscope Model EG-840N, EG-840T and EG-840TP are substantially equivalent to the predicate device, FUJIFILM Endoscope Model EG-840N, EG-840T and EG-840TP (K240142).

---

**Source:** [https://fda-staging.innolitics.com/device/K243512](https://fda-staging.innolitics.com/device/K243512)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
