Clinical specimens from HSCT and SOT transplant recipients
Clinical specimens from transplant patients were used in a method comparison study to demonstrate substantial equivalence between the modified Alinity m CMV assay (alternative DNA polymerase) and the predicate device.
Hematopoietic stem cell transplant (HSCT) and solid organ transplant (SOT) recipients; Sample Size: 141 clinical samples; Number of Sites: 1 (internal testing site)
On-market Alinity m CMV assay (original DNA polymerase)
Deming regression analysis (correlation coefficient, slope, intercept) and systematic difference at selected viral load levels
Indications for Use
The Alinity m CMV assay is an in vitro polymerase chain reaction (PCR) assay for use with the automated Alinity m System to quantitate cytomegalovirus (CMV) DNA in human EDTA plasma. The Alinity m CMV assay is intended for use as an aid in the management of Hematopoietic Stem Cell Transplant and Solid Organ Transplant patients who are undergoing anti-cytomegalovirus therapy. The Alinity m CMV assay can be used to assess virological response to anti-cytomegalovirus therapy. The results from the Alinity m CMV test must be interpreted within the context of all relevant clinical and laboratory findings. The Alinity m CMV test is not intended as a screening test for the presence of CMV DNA in blood or blood products.
Device Story
In vitro quantitative PCR assay for CMV DNA detection in human EDTA plasma; utilizes automated Alinity m System. Process: sample preparation via magnetic microparticle technology; PCR assembly; real-time fluorescence amplification/detection of CMV genomic sequences (UL34 and UL80.5 regions). System automatically calculates viral load from calibration curve. Used in clinical laboratory settings by trained personnel. Output: CMV concentration in IU/mL or Log IU/mL. Aids clinicians in managing transplant patients on anti-CMV therapy by assessing virological response. Modification involves use of alternative DNA polymerase.
Clinical Evidence
Bench testing only. Precision and reproducibility studies (N=270 and N=120 replicates) confirmed performance across the quantitation range (30 to 100,000,000 IU/mL). Method comparison against the predicate using 141 clinical samples showed high correlation (r=0.983, slope 1.01) and mean bias ≤ 0.25 Log IU/mL. No clinical prospective/retrospective studies were required.
Technological Characteristics
Quantitative real-time PCR assay. Components: AMP Kit (lyophilized PCR reagents), CTRL Kit, CAL Kit. Uses magnetic microparticle technology for nucleic acid extraction. Automated on Alinity m System. Reagents stored at 2°C to 8°C (AMP) or -25°C to -15°C (CTRL/CAL). Software-controlled automated processing.
Indications for Use
Indicated for the quantitation of CMV DNA in human EDTA plasma to aid in the management of Hematopoietic Stem Cell Transplant and Solid Organ Transplant patients undergoing anti-cytomegalovirus therapy and to assess virological response to therapy.
Regulatory Classification
Identification
A quantitative cytomegalovirus (CMV) nucleic acid test for transplant patient management is identified as a device intended for prescription use in the detection of CMV and as an aid in the management of transplant patients to measure CMV deoxyribonucleic acid (DNA) levels in human plasma and/or whole blood using specified specimen processing, amplification, and detection instrumentation. The test is intended for use as an aid in the management of transplant patients with active CMV infection or at risk for developing CMV infection. The test results are intended to be interpreted by qualified healthcare professionals in conjunction with other relevant clinical and laboratory findings.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The labeling required under § 809.10(b) of this chapter must include:
(i) A prominent statement that the device is not intended for use as a donor screening test for the presence of CMV DNA in blood or blood products.
(ii) Limitations, which must be updated to reflect current clinical practice. The limitations must include, but are not limited to, statements that indicate:
(A) Test results are to be interpreted by qualified licensed healthcare professionals in conjunction with clinical signs and symptoms and other relevant laboratory results;
(B) Negative test results do not preclude CMV infection or tissue invasive CMV disease, and that CMV test results must not be the sole basis for patient management decisions.
(iii) A detailed explanation of the interpretation of results and acceptance criteria must be provided and include specific warnings regarding the potential for variability in CMV viral load measurement when samples are measured by different devices. Warnings must include the following statement, where applicable: “Due to the potential for variability in CMV viral load measurements across different CMV assays, it is recommended that the same device be used for the quantitation of CMV viral load when managing CMV infection in individual patients.”
(iv) A detailed explanation of the principles of operation and procedures for assay performance.
(2) Design verification and validation must include the following:
(i) Detailed documentation of the device description, including all parts that make up the device, reagents required for use with the CMV assay but not provided, an explanation of the methodology, design of the primer/probe sequences, rationale for the selected gene target, and specifications for amplicon size, guanine-cytosine content, and degree of nucleic acid sequence conservation. The design and nature of all primary, secondary, and tertiary quantitation standards used for calibration must also be described.
(ii) A detailed description of the impact of any software, including software applications and hardware-based devices that incorporate software, on the device's function.
(iii) Documentation and characterization of all critical reagents (
*e.g.,* determination of the identity, supplier, purity, and stability) and protocols for maintaining product integrity throughout its labeled shelf life.(iv) Stability data for reagents provided with the device and indicated specimen types, in addition to the basis for the stability acceptance criteria at all time points chosen across the spectrum of the device's indicated life cycle, which must include a time point at the end of shelf life.
(v) All stability protocols, including acceptance criteria.
(vi) Final lot release criteria, along with documentation of an appropriate justification that lots released at the extremes of the specifications will meet the claimed analytical and clinical performance characteristics as well as the stability claims.
(vii) Risk analysis and documentation demonstrating how risk control measures are implemented to address device system hazards, such as Failure Modes Effects Analysis and/or Hazard Analysis. This documentation must include a detailed description of a protocol (including all procedures and methods) for the continuous monitoring, identification, and handling of genetic mutations and/or novel CMV stains (
*e.g.,* regular review of published literature and annual in silico analysis of target sequences to detect possible primer or probe mismatches). All results of this protocol, including any findings, must be documented.(viii) Analytical performance testing that includes:
(A) Detailed documentation of the following analytical performance studies: Limit of detection, upper and lower limits of quantitation, inclusivity, precision, reproducibility, interference, cross reactivity, carryover, quality control, specimen stability studies, and additional studies as applicable to specimen type and intended use for the device.
(B) Identification of the CMV strains selected for use in analytical studies, which must be representative of clinically relevant circulating strains.
(C) Inclusivity study results obtained with a variety of CMV genotypes as applicable to the specific assay target and supplemented by in silico analysis.
(D) Reproducibility studies that include the testing of three independent production lots.
(E) Documentation of calibration to a standardized reference material that FDA has determined is appropriate for the quantification of CMV DNA (
*e.g.,* a recognized consensus standard).(F) Documentation of traceability performed each time a new lot of the standardized reference material to which the device is traceable is released, or when the field transitions to a new standardized reference material.
(ix) Clinical performance testing that includes:
(A) Detailed documentation of device performance data from either a method comparison study with a comparator that FDA has determined is appropriate, or results from a prospective clinical study demonstrating clinical validity of the device.
(B) Data from patient samples, with an acceptable number of the CMV positive samples containing an analyte concentration near the lower limit of quantitation and any clinically relevant decision points.
(C) The method comparison study must include predefined maximum acceptable differences between the test and comparator method across all primary outcome measures in the clinical study protocol.
(D) The final release test results for each lot used in the clinical study.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY AND INSTRUMENT
## I Background Information:
A 510(k) Number
K243485
B Applicant
Abbott Molecular Inc.
C Proprietary and Established Names
Alinity m CMV
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| PAB | Class II | 21 CFR 866.3180 - Quantitative Cytomegalovirus Nucleic Acid Tests For Transplant Patient Management | MI - Microbiology |
## II Submission/Device Overview:
A Purpose for Submission:
To obtain clearance for the modification of a previously approved device, Alinity m CMV assay, which allows the use of an alternative DNA polymerase.
B Measurand:
Cytomegalovirus (CMV) DNA.
C Type of Test:
Quantitative in vitro nucleic acid amplification test.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
III Intended Use/Indications for Use:
A Intended Use(s):
The Alinity m CMV assay is an in vitro polymerase chain reaction (PCR) assay for use with the automated Alinity m System to quantitate cytomegalovirus (CMV) DNA in human EDTA plasma. The Alinity m CMV assay is intended for use as an aid in the management of Hematopoietic Stem Cell Transplant and Solid Organ Transplant patients who are undergoing anti-cytomegalovirus therapy. The Alinity m CMV assay can be used to assess virological response to anti-cytomegalovirus therapy.
The results from the Alinity m CMV test must be interpreted within the context of all relevant clinical and laboratory findings. The Alinity m CMV test is not intended as a screening test for the presence of CMV DNA in blood or blood products.
B Indication(s) for Use:
See Intended Use above.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
D Special Instrument Requirements:
For use with the Alinity m System
IV Device/System Characteristics:
A Device Description:
The Alinity m CMV assay is an in vitro polymerase chain reaction (PCR) assay for use with the automated Alinity m System to quantitate cytomegalovirus (CMV) DNA in human EDTA plasma. The Alinity m CMV assay is intended for use as an aid in the management of Hematopoietic Stem Cell Transplant (HSCT) and Solid Organ Transplant (SOT) patients who are undergoing anti-cytomegalovirus therapy. The Alinity m CMV assay can be used to assess virological response to anti-cytomegalovirus therapy.
The Alinity m CMV assay requires 3 separate assay specific kits:
- Alinity m CMV AMP Kit (List No. 09N46-095) consisting of 2 types of multi-well assay trays. The amplification trays (AMP TRAY 1) contain lyophilized, unit-dose PCR amplification/detection reagents and lyophilized, unit-dose internal control (IC) with proteinase K in separate wells, and the activation trays (ACT TRAY 2) contain liquid unit-dose activation reagent. The intended storage condition for the Alinity m CMV AMP Kit is 2°C to 8°C.
- Alinity m CMV CTRL Kit (List No. 09N46-085) consisting of negative controls, low-positive controls and high-positive controls, each supplied as liquid in single-use tubes. The intended storage condition for the Alinity m CMV CTRL Kit is –25°C to –15°C.
K243485 - Page 2 of 12
{2}
- Alinity m CMV CAL Kit (List No. 09N46-075) consisting of 2 calibrator levels, each supplied as liquid in single-use tubes. The intended storage condition for the Alinity m CMV CAL Kit is –25°C to –15°C.
## B Principle of Operation:
The Alinity m CMV assay utilizes real-time polymerase chain reaction (PCR) to amplify and detect CMV genomic DNA sequences that have been extracted from human plasma specimens. The steps of the Alinity m CMV assay consist of sample preparation, PCR assembly, amplification/detection, and result calculation and reporting. All steps of the Alinity m CMV assay procedure are executed automatically by the Alinity m System. Manual dilutions may be performed for low-volume specimens to meet the minimum volume requirement.
The Alinity m System is designed to be a random access analyzer that can perform the Alinity m CMV assay in parallel with other Alinity m assays on the same instrument.
CMV DNA from human plasma is extracted using the Alinity m Sample Prep Kit 2, Proteinase K, Alinity m Lysis Solution, and Alinity m Diluent Solution. The Alinity m System employs magnetic microparticle technology to facilitate nucleic acid capture, wash, and elution. The resulting purified DNA is then combined with liquid unit-dose Alinity m CMV activation reagent and lyophilized unit-dose Alinity m CMV amplification/detection reagents and transferred into a reaction vessel. Alinity m Vapor Barrier Solution is then added to the reaction vessel which is then transferred to an amplification/detection unit for PCR amplification, and real-time fluorescence detection of CMV.
At the beginning of the Alinity m CMV sample preparation process, a lyophilized unit-dose internal control (IC) on the AMP Tray is rehydrated by the Alinity m System and delivered into each sample preparation reaction. The IC is then processed through the entire sample preparation and real-time PCR procedure along with the specimens, calibrators, and controls to demonstrate proper sample processing and validity. The Alinity m CMV amplification/detection reagents consist of enzymes, primers, probes, and activation reagents that enable polymerization, and detection.
A CMV calibration curve is required for determination of CMV DNA concentration. Two levels of calibrators are processed through sample preparation and PCR to generate the calibration curve. The concentration of CMV DNA in specimens and controls is then calculated from the stored calibration curve.
Assay controls are tested at or above an established minimum frequency to help ensure that instrument and reagent performance remains satisfactory. During each control event, a negative control, a low positive control, and a high positive control are processed through sample preparation and PCR procedures that are identical to those used for specimens.
## Calculation
Quantitative viral load results are reported for patient specimens with CMV viral concentrations within the assay's quantitation range. The concentration of CMV DNA in a specimen is calculated from the calibration curve by the system software. The Alinity m System reports the results in International Units as IU/mL or Log [IU/mL].
K243485 - Page 3 of 12
{3}
K243485 - Page 4 of 12
# Interpretation of results
| Result | Interpretation | Interpretation Additional Information |
| --- | --- | --- |
| Not Detected | CMV DNA not detected | |
| <LLoQ | CMV DNA detected but not quantified | CMV DNA concentration is below the Lower Limit of Quantitation (LLoQ) of the assay |
| LLoQ to ≤ULoQ | CMV DNA detected and quantified | CMV DNA concentration is within the linear range of the assay (≥LLoQ to ≤ULoQ) |
| >ULoQ | CMV DNA detected | CMV DNA concentration is above the Upper Limit of Quantification (ULoQ) of the assay. |
# C Instrument Description Information:
1. Instrument Name:
Alinity m System
2. Specimen Identification:
Cytomegalovirus DNA in Human EDTA plasma
3. Specimen Sampling and Handling:
Plasma specimens may be tested for viral load determination. For the Alinity m CMV assay, only use collection tubes as described: K₂ EDTA, K₃ EDTA, Plasma Preparation Tube (PPT)
4. Calibration:
Alinity m CMV CAL Kit consisting of 2 calibrator levels, each supplied as liquid in single-use tubes.
5. Quality Control:
Alinity m CMV CTRL Kit consisting of negative controls, low-positive controls, and high-positive controls, each supplied as liquid in single-use tubes.
# V Substantial Equivalence Information:
{4}
A Predicate Device Name(s):
Alinity m CMV
B Predicate 510(k) Number(s):
P210022
C Comparison with Predicate(s):
| Device & Predicate Device(s): | P210022 | K243485 |
| --- | --- | --- |
| Device Trade Name | Alinity m CMV | Same |
| Regulation No./Product Code | 21 CFR 814 / PAB | 21 CFR 866.3180 / PAB |
| Device Class | Class III | Class II |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The Alinity m CMV assay is an in vitro polymerase chain reaction (PCR) assay for use with the automated Alinity m System quantitate cytomegalovirus (CMV) DNA in human EDTA plasma. The Alinity m CMV assay is intended for use as an aid in the management of Hematopoietic Stem Cell Transplant and Solid Organ Transplant patients who are undergoing anti-cytomegalovirus therapy. The Alinity m CMV assay can be used to assess virological response to anti-cytomegalovirus therapy. The results from the Alinity m CMV test must be interpreted within the context of all relevant clinical and laboratory findings. The Alinity m CMV test is not intended as a screening test for the presence of CMV DNA in blood or blood products. | Same |
| Assay Type | Quantitative | Same |
| Assay Targets | Conserved regions of CMV genome (UL34 and UL80.5) | Same |
| Specimen Types | EDTA Plasma | Same |
| Specimen Collection and Transport | K2 EDTA, K3 EDTA, Plasma Preparation Tube (PPT) Plasma may be stored in primary tubes (with or without gel) and secondary tubes | Same |
| Principles of the Procedure | See above for description of principles of procedure | Same |
K243485 - Page 5 of 12
{5}
| Instrument and System Components | Alinity m System: Fully integrated and automated diagnostics analyzer which utilizes real-time PCR technology | Same |
| --- | --- | --- |
| Sample Preparation Instrument Components | Automated liquid handling and robotic manipulation platform | Same |
| Assay Controls | • Negative Control
• Low Positive Control
• High Positive Control
• Internal Control (IC) | Same |
| Results Reporting | • Not Detected
• Detected < LLoQ
• Detected > ULoQ
• Concentration from LLoQ to ≤ ULoQ, reported in IU/mL or Log IU/mL
For low volume specimens tested with the Specimen Dilution Procedure, quantitative results generated on the Alinity m System represent the CMV DNA concentration in the specimen prior to dilution. | Same |
| Reagents and Storage Conditions | • Alinity m CMV AMP Kit: 2°C to 8°C
• Alinity m CMV CTRL Kit: –25°C to –15°C
• Alinity m CMV CAL Kit: –25°C to –15°C | Same |
| General Device Characteristic Differences | | |
| Amplification Enzymes | Original DNA Polymerase is the enzyme used for DNA amplification. | Original DNA Polymerase or alternative DNA Polymerase is the enzyme used for DNA amplification. |
VI Standards/Guidance Documents Referenced:
Special Control (89 FR 77448), labeling requirements under 21 CFR 809.10(b), the information for special controls pertaining to cytomegalovirus (CMV) quantitative nucleic acid test intended for transplant patient management.
CLSI EP05-A3: Clinical Laboratory Standards Institute. Evaluation of the Precision of Quantitative Measurement Procedures; Approved Guideline – Third Edition. CLSI Guideline EP05-A3, CLSI: Wayne, PA, 2014.
K243485 - Page 6 of 12
{6}
CLSI EP35: Clinical Laboratory Standards Institute. Assessment of Equivalence or Suitability of Specimen Types for Medical Laboratory Measurement Procedures; Approved Guideline – First Edition. CLSI Guideline EP35. CLSI: Wayne, PA: 2019.
CLSI EP09c: Clinical and Laboratory Standards Institute. Measurement Procedure Comparison and Bias Estimation Using Patient Sample – Third Edition. CLSI Guideline EP09c. Wayne, PA: CLSI; 2018.
CLSI EP06: Clinical Laboratory Standards Institute. Evaluation of Linearity of Quantitative Measurement Procedures; Approved Guideline – Second Edition. CLSI Guideline EP06, CLSI: Wayne, PA, 2020.
CLSI EP17-A2: Clinical Laboratory Standards Institute. Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition. CLSI Guideline EP17-A2, CLSI: Wayne, PA, 2012.
## VII Performance Characteristics (if/when applicable):
## A Analytical Performance:
### 1. Precision/Reproducibility:
**Precision:**
Precision of Alinity m CMV assay with the alternative DNA polymerase enzyme was determined by testing 8 panel members with CMV concentrations that span the targeted quantitation range of the assay (30 to $10^{8}$ IU/mL [1.48 to 8.00 Log IU/mL]). Panel members with multiple targeted concentration levels were prepared with clinical sample (1.48 to 2.7 Log IU/mL), cultured virus (4.0 to 5.0 IU/mL) or plasmid DNA (7.0 to 8.0 IU/mL). The quantitation of the panel members was traceable to international standard material for Human Cytomegalovirus (CMV) DNA (code 09/162). Each panel member was tested in 3 replicates, twice each day for 15 days on 1 Alinity m System operated by 3 operators using 3 lots of CMV AMP kit, for a total of 270 replicates per panel member. The study design and analysis were derived from recommendations in CLSI Guideline EP05-A3 $3^{\text{rd}}$ Edition.
The precision study results are shown in Table 1.
Table 1. Precision Analysis (All lots)
| Panel Member | N | Mean Conc. (Log IU/mL) | Within-Run | | Between-Run | | Between-Day | | Within-Laboratory^{a} | | Between-Lot | | Total^{b} | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 01 | 270 | 1.52 | 0.26 | 17.2 | 0.06 | 3.7 | 0.00 | 0.0 | 0.27 | 17.6 | 0.02 | 1.3 | 0.27 | 17.6 |
| 02 | 270 | 1.70 | 0.16 | 9.6 | 0.02 | 0.9 | 0.03 | 1.5 | 0.17 | 9.7 | 0.01 | 0.4 | 0.17 | 9.8 |
| 03 | 270 | 2.13 | 0.21 | 9.8 | 0.05 | 2.4 | 0.00 | 0.0 | 0.22 | 10.1 | 0.00 | 0.2 | 0.22 | 10.1 |
| 04 | 270 | 2.83 | 0.05 | 1.8 | 0.02 | 0.9 | 0.00 | 0.0 | 0.06 | 2.0 | 0.00 | 0.1 | 0.06 | 2.0 |
| 05 | 270 | 4.04 | 0.04 | 1.0 | 0.03 | 0.7 | 0.00 | 0.0 | 0.05 | 1.3 | 0.02 | 0.4 | 0.05 | 1.3 |
| 06 | 270 | 5.03 | 0.03 | 0.7 | 0.03 | 0.5 | 0.00 | 0.0 | 0.04 | 0.8 | 0.01 | 0.3 | 0.04 | 0.9 |
| 07 | 270 | 7.19 | 0.04 | 0.5 | 0.02 | 0.3 | 0.00 | 0.0 | 0.04 | 0.6 | 0.02 | 0.3 | 0.05 | 0.7 |
| 08 | 270 | 8.15 | 0.05 | 0.6 | 0.00 | 0.0 | 0.01 | 0.1 | 0.05 | 0.6 | 0.03 | 0.3 | 0.05 | 0.7 |
<a>^{a}</a> Within-Laboratory includes Within-Run, Between-Run, and Between-Day Components.
<a>^{b}</a> Total includes Within-Run, Between-Run, Between-Day, and Between-Lot Components.
K243485 - Page 7 of 12
{7}
# Reproducibility:
Reproducibility of Alinity m CMV assay with the alternative DNA polymerase enzyme was determined by testing 9-member reproducibility panel. Eight positive panel members with multiple targeted concentration levels were prepared with clinical sample (1.0 to 2.0 Log IU/mL), cultured virus (3.0 to 5.0 Log IU/mL), and or plasmid DNA (> 5.0 Log IU/mL). The quantitation of the panel members was traceable to international standard material for CMV DNA (code 09/162).
Each panel member was tested using 1 reagent lot, 1 lot of calibrators and controls, 3 external sites with 1 instrument per site, 4 replicates per panel per run, 5 non-consecutive days, 2 runs per day. The study design contained a total of 120 replicates across all runs for each panel member tested. The study design and analysis were derived from recommendations in CLSI Guideline EP05-A3 $3^{\mathrm{rd}}$ Edition.
The reproducibility results are summarized in Table 2.
Table 2. Reproducibility for Positive Panel Members (Log IU/mL)
| Panel Member | N | Mean Conc. (Log IU/mL) | Within-Run | | Between-Run | | Between-Day | | Within-Laboratorya | | Between-Site | | Totalb | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 117 | 7.87 | 0.05 | 0.7 | 0.02 | 0.3 | 0.00 | 0.0 | 0.06 | 0.7 | 0.09 | 1.1 | 0.11 | 1.4 |
| 2 | 117 | 6.85 | 0.05 | 0.7 | 0.01 | 0.2 | 0.01 | 0.2 | 0.05 | 0.7 | 0.07 | 1.0 | 0.08 | 1.2 |
| 3 | 119 | 5.11 | 0.05 | 1.0 | 0.00 | 0.0 | 0.01 | 0.3 | 0.05 | 1.1 | 0.03 | 0.5 | 0.06 | 1.2 |
| 4 | 119 | 4.03 | 0.04 | 1.0 | 0.02 | 0.4 | 0.00 | 0.1 | 0.04 | 1.1 | 0.04 | 1.0 | 0.06 | 1.5 |
| 5 | 117 | 3.09 | 0.06 | 1.9 | 0.02 | 0.7 | 0.01 | 0.2 | 0.06 | 2.1 | 0.05 | 1.7 | 0.08 | 2.7 |
| 6 | 118 | 1.87 | 0.16 | 8.5 | 0.03 | 1.7 | 0.00 | 0.0 | 0.16 | 8.7 | 0.09 | 4.9 | 0.19 | 10.0 |
| 7 | 116 | 1.44 | 0.25 | 17.4 | 0.11 | 7.5 | 0.00 | 0.0 | 0.27 | 19.0 | 0.13 | 8.7 | 0.30 | 20.9 |
| 8 | 111 | 1.20 | 0.32 | 26.6 | 0.10 | 8.2 | 0.00 | 0.0 | 0.33 | 27.9 | 0.17 | 14.4 | 0.38 | 31.4 |
a Within-Laboratory includes Within-Run, Between-Run, and Between-Day Variance Components.
b Total (Overall) includes Within-Run, Between-Run, Between-Day, and Between-Site Variance Components.
For the negative panel member the overall negative percent agreement with the expected result was $100.0\%$ (119/119) Table 3. All three external sites had $100.0\%$ negative percent agreement with the expected result.
Table 3. Reproducibility for Negative Panel Member
| Panel Member | Site | Total N | Positive | Negative | Negative Rate (%) | 95% CI |
| --- | --- | --- | --- | --- | --- | --- |
| 9 | all | 119 | 0 | 119 | 100% | (96.9-100%) |
# 2. Linearity:
Linearity was evaluated by testing 17 panel members that spanned the intended linear range of the assay (30 to 100,000,000 IU/mL), including a panel member below the expected Lower Limit of Quantification (LLoQ) at 15 IU/mL (1.18 Log IU/mL), and a Panel member exceeding the expected Upper Limit of Quantification (ULoQ) at 200,000,000 IU/mL (8.30 Log IU/mL). Panel members with multiple concentration levels were prepared with clinical sample (1.18 to 3.4 Log IU/mL), cultured virus (1.3 to 5.0 Log IU/mL) and plasmid DNA
K243485 - Page 8 of 12
{8}
(1.4 to 8.3 Log IU/mL). The quantitation of the panel members was traceable to international standard material for CMV DNA (code 09/162). The study was conducted using 1 lot of Alinity m CMV on 1 instrument with a minimum of 27 replicates for each panel member. Linearity study design and evaluation was based on the recommendations in CLSI EP06 2nd Edition.
Alinity m CMV was linear across the quantitation range from 30 to 100,000,000 IU/mL (1.48 to 8.0 Log IU/mL). Representative results for Alinity m CMV linearity performance are shown in Figure 1.

Figure 1. Alinity m CMV Linearity
3. Analytical Specificity/Interference:
Refer to P210022.
4. Assay Reportable Range:
Refer to P210022. The analytical measuring interval of the Alinity m CMV for plasma is from the lower limit of quantitation (LLoQ) of 30 IU/mL (1.48 Log IU/mL) to the upper limit of quantitation (ULoQ) of 100,000,000 IU/mL (8.0 Log IU/mL).
5. Traceability and Stability:
Traceability to international standard material:
Refer to P210022.
Reagent Stability:
The data submitted support the following storage conditions for the Alinity m CMV assay with alternative DNA polymerase.
K243485 - Page 9 of 12
{9}
Table 4. Alinity m CMV AMP kit stability
| Stability study | Expiration |
| --- | --- |
| Reagent On-board | 30 days |
| Reagent Shelf-life | 15 months |
6. Detection Limit:
The Limit of Detection (LoD) was assessed by testing dilutions of the international standard material for CMV DNA (code 09/162) prepared in CMV-negative human plasma. The three concentration panel members (20.0 IU/mL, 30.0 IU/mL, and 50.0 IU/mL) were selected based on expected LoD of the on-market Alinity m CMV. Testing for each CMV DNA concentration was performed with 3 lots of reagents across multiple days. LoD Study design and analysis follow CLSI EP17-A2 and demonstrate a detection rate of $\geq 95.0\%$ for samples at the claimed LoD of 30 IU/mL (1.48 Log IU/mL) and above.
7. Lower Limit of Quantitation (LLoQ):
LLoQ is defined as the lowest concentration at which CMV DNA is reliably quantitated within an acceptable total error. The LLoQ was evaluated based on the recommendation in CLSI EP17-A2, section 7. For each panel member from the CMV LoD study, Total Analytical Error (TAE) and Total Error in difference between two measurements (TE) were calculated for each lot and for all lots.
The results are presented in Table 5 and Table 6.
Table 5. Alinity m CMV TAE and TE (overall)
| Panel Member | Target concentration (IU/mL) | N | Target Concentration (Log IU/mL) | Mean (Log IU/mL) | Bias (Log IU/mL) | SD | TAE | TE |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| 1 | 20 | 139 | 1.30 | 1.30 | -0.00 | 0.31 | 0.62 | 0.88 |
| 2 | 30 | 139 | 1.48 | 1.42 | -0.06 | 0.27 | 0.59 | 0.75 |
| 3 | 50 | 144 | 1.70 | 1.74 | 0.04 | 0.17 | 0.38 | 0.48 |
Table 6. Alinity m CMV TAE and TE (by Lot)
| Panel Member | Target Concentration ( IU/mL) | Lot | N | Target Concentration (Log IU/mL) | Mean concentration (Log IU/mL) | Bias (Log IU/mL) | SD | TAE | TE |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| 1 | 20 | 1 | 46 | 1.30 | 1.30 | 0.00 | 0.26 | 0.52 | 0.74 |
| | | 2 | 48 | 1.30 | 1.26 | -0.04 | 0.36 | 0.76 | 1.01 |
| | | 3 | 45 | 1.30 | 1.34 | 0.04 | 0.30 | 0.64 | 0.86 |
| 2 | 30 | 1 | 45 | 1.48 | 1.36 | -0.12 | 0.29 | 0.71 | 0.83 |
| | | 2 | 47 | 1.48 | 1.41 | -0.07 | 0.27 | 0.62 | 0.77 |
| | | 3 | 47 | 1.48 | 1.49 | 0.01 | 0.21 | 0.43 | 0.60 |
| 3 | 50 | 1 | 48 | 1.70 | 1.76 | 0.06 | 0.15 | 0.37 | 0.43 |
| | | 2 | 48 | 1.70 | 1.76 | 0.06 | 0.19 | 0.44 | 0.54 |
| | | 3 | 48 | 1.70 | 1.69 | -0.01 | 0.16 | 0.33 | 0.46 |
The panel member with the target concentration at the assay's claimed LoQ of 30 IU/mL has a TAE and TE less than or equal to 1.00 Log IU/mL.
K243485 - Page 10 of 12
{10}
8. Accuracy (Instrument):
Accuracy was reviewed in P210022.
9. Carry-Over:
Carry-over was reviewed in P210022/S008.
## B Comparison Studies:
1. Method Comparison with Predicate Device:
The Alinity m CMV clinical specimen method comparison testing was conducted at one testing site (internal testing site) using 3 reagent lots of the Alinity m CMV with alternative and 1 lot of the on-market Alinity m CMV. The specimens were obtained from hematopoietic stem cell transplant (HSCT) and solid organ transplant (SOT) recipients. Samples were selected such that CMV levels in the samples span the measuring range of the Alinity m CMV assay. To ensure coverage of the measuring range, both neat clinical specimens and samples made with unique individual high-titer clinical specimens spiked into unique individual negative specimens were included. All samples were tested with both the on-market Alinity m CMV assay with the original DNA polymerase and the Alinity m CMV assay with the alternative DNA polymerase.
Figure 2 shows the Deming regression performed using the on-market Alinity m assay result as the independent variable and the respective Alinity m with alternative assay result as the dependent variable on 141 clinical samples. The Deming regression analysis results have correlation coefficient (r) of 0.983, a slope of 1.01 and intercept of 0.01, and the 95% confidence intervals (-0.13, 0.16) (Figure 2).

Figure 2. Deming Regression Plot – All data
K243485 - Page 11 of 12
{11}
Systematic Difference was calculated for selected viral load levels (2.70, 3.30, 4.00 Log IU/mL) along with the two-sided 95% confidence intervals (Table 7).
Table 7. Systematic Difference at Selected Viral Load Levels
| Target Viral Load Levels | Systematic Difference | 95% CI |
| --- | --- | --- |
| 2.70 Log IU/mL | 0.04 Log IU/mL | (-0.01, 0.09) |
| 3.30 Log IU/mL | 0.05 Log IU/mL | (0.01, 0.08) |
| 4.00 Log IU/mL | 0.05 Log IU/mL | (0.03, 0.08) |
The Alinity m CMV assay with alternative DNA polymerase has viral load values with a mean bias relative to the comparator assay of less than or equal to 0.25 Log IU/mL within the quantitation range.
2. Matrix Comparison:
N/A
C Clinical Studies:
1. Clinical Sensitivity:
N/A
2. Clinical Specificity:
N/A
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
N/A
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K243485 - Page 12 of 12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.