This device is used for the endoscopic removal of stones in the biliary system and foreign bodies.
Device Story
Extraction Basket is a sterile, single-use endoscopic accessory for biliary stone and foreign body removal. Type 1 device operates via duodenoscope channel, crossing duodenal papilla into bile duct, guided by X-ray imaging. Type 2 device operates via choledochoscope channel, utilizing direct visual guidance. Both types feature basket assembly, sheath, and handle; operator advances basket to target, opens it to grasp stone/foreign body, and withdraws catheter. Used in hospital settings by physicians. Output is physical retrieval of biliary obstructions. Benefits include minimally invasive stone removal, reducing need for more invasive surgery.
Clinical Evidence
No clinical data or animal studies included. Evidence consists of bench testing: biocompatibility (ISO 10993-1), dimension, connection force, simulated-use, stone capture/durability, injection, luer, and X-ray detectability testing. Shelf-life and packaging integrity validated per ASTM F1980-21 and ISO 11607-1/2. Sterilization validated per ISO 11135.
Technological Characteristics
Single-use biliary extraction basket; metal wire construction. Type 1 (7Fr) and Type 2 (3Fr) configurations. EO sterilized (SAL 10^6). Shelf life: 1 year (Type 1), 3 years (Type 2). Compatible with endoscopic working channels. No software or electronic components.
Indications for Use
Indicated for endoscopic removal of stones in the biliary system and foreign bodies in patients requiring such procedures.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
Web II Memory Extraction Basket or Memory II Double Lumen Extraction Basket Memory 5 Fr. Soft Wire Baskets or Memory Helical Stone Extractor Memory Eight Wire Baskets Fusion Wire Guided Extraction Basket Non-Lithotripsy Extraction Basket (K171969)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
December 16, 2024
Micro-Tech (Nanjing) Co., Ltd. Sally He Regional RA Manager No.10 Gaoke Third Rd Nanjing National Hi-Tech, Industrial Development Zone Nanjing, Jiangsu 210032 China
Re: K243471
Trade/Device Name: Extraction Basket Regulation Number: 21 CFR 876.5010 Regulation Name: Biliary Catheter And Accessories Regulatory Class: Class II Product Code: LQR Dated: November 8, 2024 Received: November 8, 2024
Dear Sally He:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Anthony Lee -S
Anthony Lee, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
### K243471
Device Name
Extraction Basket
Indications for Use (Describe)
This device is used for the endoscopic removal of stones in the biliary system and foreign bodies.
Type of Use (Select one or both, as applicable)
> | Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional font. The letters are connected and appear to be angled slightly, giving the logo a dynamic look. A small registered trademark symbol is located in the upper right corner of the logo.
### 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92. The assigned 510(k) Number:
1. Date of Preparation: 2024-11-27
#### Sponsor Identification 2.
# Micro-Tech (Nanjing) Co., Ltd.
No.10 Gaoke Third Road, Nanjing National Hi-Tech, Industrial Development Zone, Nanjing, Jiangsu Province, PRC
Establishment Registration Number: 3004837686
Contact Person: Sally He
Position: Regional RA Manager
Tel: +86-25-58646395
Fax: +86-25-58350006
Email: ra.micro-tech@outlook.com
#### 3. Identification of Proposed Device
Trade Name: Extraction Basket
## Regulatory Information
Classification Name: Biliary Catheter And Accessories Classification: 2 Product Code: LQR Regulation Number: 21 CFR 876.5010 Review Panel: Gastroenterology/Urology
#### Identification of Predicate Device 4.
510(k) Number: K171969 Product Name: Web II Memory Extraction Basket or Memory II Double Lumen Extraction Basket Memory 5 Fr. Soft Wire Baskets or Memory Helical Stone Extractor Memory Eight Wire Baskets Fusion Wire Guided Extraction Basket Non-Lithotripsy Extraction Basket Manufacturer: Wilson-Cook Medical
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a blue and white logo with the letters "MT" in a stylized, three-dimensional font. The letters are interconnected, with the "M" on the left and the "T" on the right. The logo has a registered trademark symbol, the letter R inside a circle, in the upper right corner.
#### ર. Indications for Use
This device is used for the endoscopic removal of stones in the biliary system and foreign bodies.
#### 6. Device Description
The proposed device is a sterile, single-use accessory to be used with endoscopic, intended to be used for the removal of stones in the biliary system and foreign bodies. The proposed device includes two types, one type (Hereinafter referred to Type 1) is working through the channel provided by duodenoscopy, cross the duodenal papilla and insert into the bile duct, with the assistance of Imaging field of view formed by X-ray, capture and remove stones out of bile duct. Another one (Hereinafter referred to Type 2) is working through the channel provided by choledochoscope, after the choledochoscope was inserted into the bile duct successfully, inserted the extraction basket into the bile duct by the channel of choledochoscope, capture and remove stones out of bile duct by the direct field view of choledochoscope.
The proposed device is EO sterilized to achieve the Sterility Assurance Level (SAL) of 106 and placed in a sterility maintenance package to ensure a shelf life of 1 year for Type 1 and 3 years for Type 2.
#### 7. Comparison of Technological Characteristics
The Extraction Basket incorporates substantially equivalent device's intended use, using environment, design, packaging fundamental technology, principles of operation, manufacturing processes including sterilization process, configuration, main material as those featured in the predicate device Memory Eight Wire Baskets cleared under K171969. Only the dimension and shelf life of proposed device are different from predicate device, which have been demonstrated not influence the safety and effectiveness of proposed device.
| Item | Proposed Device<br>Extraction Basket<br>(Type 1 ) | Proposed Device<br>Extraction Basket<br>(Type 2) | Predicate Device<br>(K171969) | Remark |
|---------------------|-------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|--------|
| Product Code | LQR | LQR | LQR | SE |
| Regulation No. | 21 CFR 876.5010 | 21 CFR 876.5010 | 21 CFR 876.5010 | SE |
| Class | II | II | II | SE |
| Indications for Use | This device is used for the<br>endoscopic removal of<br>stones in the biliary<br>system and foreign bodies. | This device is used for the<br>endoscopic removal of<br>stones in the biliary<br>system and foreign<br>bodies. | Endoscopic removal of<br>stones in the biliary<br>system and foreign bodies. | SE |
| Single Use | Yes | Yes | Yes | SE |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional font. The letters are interconnected and appear to be made of blocks, giving them a geometric look. A circled "R" symbol, indicating a registered trademark, is positioned in the upper right corner of the logo.
| | 510(k) summary<br>N |
|--|---------------------|
|--|---------------------|
| Item | Proposed Device<br>Extraction Basket<br>(Type 1 ) | Proposed Device<br>Extraction Basket<br>(Type 2) | Predicate Device<br>(K171969) | Remark | |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|---------|
| Supplied in Sterile | Yes | Yes | Yes | SE | |
| Configuration | Consists of Basket<br>Assembly, Sheath, Handle<br>assembly. | Consists of Basket<br>Assembly, Sheath, Handle<br>assembly. | Consists of Basket<br>Assembly, Sheath, Handle<br>assembly. | SE | |
| Main Material | Metal wires | Metal wires | Metal wires | SE | |
| Using Environment | Hospital | Hospital | Hospital | SE | |
| Dimensions | Basket<br>Diameter | 10mm, 15 mm, 20 mm, 25<br>mm, 30 mm | 10mm, 15 mm, 20 mm,<br>25 mm, 30 mm | 15mm,20 mm, 25 mm, 30<br>mm | Similar |
| | Basket<br>Length | 30 mm,40 mm, 45mm,50<br>mm,55mm, 60 mm, 70<br>mm | 30 mm,40 mm, 50 mm,<br>60 mm, 70 mm | 35mm,40 mm, 50 mm,60<br>mm | Similar |
| | Outer Tube<br>Diameter | 7Fr | 3Fr | 5.5Fr,7Fr,8Fr,10Fr | Similar |
| | Compatible<br>Endoscopy<br>Working<br>Channel | 7Fr:≥3.7 mm | 3Fr:≥1.1 mm | 5.5Fr:≥2.0 mm, 7Fr:≥2.8<br>mm<br>8Fr:≥3.2 mm, 10Fr:≥4.2<br>mm | Similar |
| | Working<br>Length | 2500 mm,2000 mm | 2900 mm,2000 mm,1600<br>mm,1200 mm,1000<br>mm,800 mm,600 mm | 2000 mm | Similar |
| Principles<br>of<br>Operation | The device enters the<br>human<br>body<br>through<br>endoscope, advance the<br>basket to the appropriate<br>position, slowly open the<br>basket, grasp the stones<br>or foreign bodies by pull<br>the basket over them,<br>withdraw<br>the<br>catheter<br>from the bile duct to<br>retrieve the stones<br>or<br>foreign bodies | The device enters the<br>human<br>body through<br>endoscope, advance the<br>basket to the appropriate<br>position, slowly open the<br>basket, grasp the stones<br>or foreign bodies by pull<br>the basket over them,<br>withdraw<br>the<br>catheter<br>from the bile duct to<br>retrieve the stones<br>or<br>foreign bodies | The device enters the<br>human<br>body through<br>endoscope, advance the<br>basket to the appropriate<br>position, slowly open the<br>basket, grasp the stones<br>or foreign bodies by pull<br>the basket over them,<br>withdraw<br>the<br>catheter<br>from the bile duct to<br>retrieve the stones<br>or<br>foreign bodies | SE | |
| Applicable Body Parts | Biliary Tract | Biliary Tract | Biliary Tract | SE | |
| Packaging | Single-use EO sterilized<br>pouch with one device per<br>pouch | Single-use EO sterilized<br>pouch with one device per<br>pouch | Single-use EO sterilized<br>pouch with one device per<br>pouch | SE | |
#### 8. Performance Data
The biocompatibility evaluation for Extraction Basket was conducted in accordance with ISO 10993-1: 2009 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a blue logo with the letters 'MT' in a stylized, three-dimensional design. The letters are blocky and connected, with the 'M' on the left and the 'T' on the right. A registered trademark symbol is located in the upper right corner of the logo. The logo has a clean and modern appearance.
510(k) summary
Management Process" and FDA's biocompatibility guidance, Use of International Standard ISO-10993-1, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a risk management process (issued on September 4, 2020.) the following tests were conducted:
- Cytotoxicity a)
- Sensitization b)
- Irritation c)
- Acute Systemic Toxicity d)
- Material Mediated Pyrogenicity e)
Performance testing was conducted to demonstrate the essential performance of the proposed device and confirmed that the proposed device works as intended with the compatible devices. The following tests are conducted:
Type 1 :
- > Dimension Testing;
- > Connection Force Testing:
- > Simulated-Use Testing;
- A Stone Capture and Durability Testing;
- > Injection Testing;
- > Luer Testing;
- > X-ray Detectability Testing;
Type 2 :
- ▲ Dimension Testing;
- A Connection Force Testing;
- A Simulated-Use Testing;
- A Stone Capture and Durability Testing.
Shelf-life testing and packaging integrity testing was conducted based on an accelerated aging test in accordance with ASTM F1980-21 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices and ISO 11607-1:2019: Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems and ISO 11607-2:2019: Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes. One-year accelerated aging test was performed to demonstrate the one-year stability in the shelf life.
Sterilization validation was carried out in accordance with ISO 11135:2014+A1:2018 "Sterilization of Health Care products - Ethylene Oxide - Part 1: Requirements for Development, Validation, and
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional font. The letters are connected and appear to be angled slightly, giving them a sense of depth. A small "R" in a circle, indicating a registered trademark, is positioned in the upper right corner of the logo.
K243471 Page 5 of 5
510(k) summary
Routine Control of Sterilization processes for Medical Devices".
#### 9. Animal Study
No animal study is included in this submission.
## 10. Clinical Study
No clinical study is included in this submission.
### 11. Substantially Equivalent (SE) Conclusion
Based on the indications for use, technological characteristics, safety and performance testing, Extraction Basket has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the currently cleared predicate device cleared under K171969.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.