The Fix2Lock is intended use for fixation of soft tissue to bone, using suture, in the following procedure; shoulder, foot/ankle, knee, hand/wrist and elbow. Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, digital tendon transfers, Mid-foot reconstruction Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints for all digits, digital tendon transfers Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction
Device Story
Fix2Lock is an absorbable bone fixation screw system used in orthopedic surgery to secure soft tissues (ligaments, tendons, articular capsules) to bone. The system comprises an anchor (biocomposite or PEEK), non-absorbable UHMWPE suture, needle, driver shaft, and handle. Operated by orthopedic surgeons in clinical settings, the device is implanted to provide mechanical fixation during tissue healing. The surgeon uses the driver to insert the anchor into bone, securing the suture to the soft tissue. This restores anatomical attachment, facilitating tissue repair and potentially improving patient functional outcomes. The device is single-use and provided sterile.
Clinical Evidence
Bench testing only. Performance verified against ASTM F543 (metallic bone screws) and ASTM F2502 (absorbable plates/screws). Material compliance verified against ASTM F2026 (PEEK) and ASTM F2848 (UHMWPE). Sterilization, shelf-life, and packaging validated per ISO 11135, ISO 11607, and ASTM F1980 standards.
Technological Characteristics
Anchor materials: PLGA+β-TCP (biocomposite) and PEEK. Suture material: UHMWPE. Mechanical performance per ASTM F543 and ASTM F2502. Sterilization: Ethylene oxide (EtO) per ISO 11135. Single-use, sterile, 3-year shelf life. Dimensions: 3.75-6.5mm diameter, 12.0-14.0mm length.
Indications for Use
Indicated for soft tissue-to-bone fixation using suture in orthopedic procedures of the shoulder, foot/ankle, knee, hand/wrist, and elbow, including ligament, tendon, and articular capsule repairs.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
December 3, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
OSTEONIC Co., Ltd. % Lee Sanglok Manager WISE COMPANY Inc. #1107, #1108, 204 Gasandigital 1-ro, Geumcheon-gu Seoul. 08502 Korea, South
Re: K243467
Trade/Device Name: Fix2Lock (Biocomposite medial, Biocomposite lateral, Biocombi Self Punching) Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: MAI, MBI Dated: November 8, 2024 Received: November 8, 2024
Dear Lee Sanglok:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# CHRISTOPHER FERREIRA -S
Christopher Ferreira, MS Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K243467
Device Name
Fix2Lock (Biocomposite medial, Biocomposite lateral, Biocombi Self Punching)
Indications for Use (Describe)
The Fix2Lock is intended use for fixation of soft tissue to bone, using suture, in the following procedure:
shoulder, foot/ankle, knee, hand/wrist and elbow.
Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsulolabral Reconstruction Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, digital tendon transfers, Mid-foot reconstruction Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints for all digits, digital tendon transfers Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
ne-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Date of Submission: 2024.11.08
### 01. Applicant
OSTEONIC Co., Ltd.
405Ho, 505-2Ho, 505-3Ho, 508Ho, 603Ho, 902Ho, 1004Ho, 1005Ho, 1202Ho, 1206Ho, 38 Digital-ro 29-gil, Guro-gu, Seoul, Korea, 102Ho, 103Ho, 24 Digital-ro 27-gil, Guro-gu, Seoul, Korea, 1012Ho, 1013Ho, 272, Digital-Ro, Guro-gu, Seoul, Korea TEL: +82-2-6902-8400 FAX: +82-2-6902-8401 Email: dakham@osteonic.com
#### 02. Submission Correspondent
Sanglok, Lee Wise COMPANY Inc. 1107Ho, 1108Ho, 204 Gasan digital 1-ro, Geumcheon-gu, Seoul, Korea TEL: +82 70 8812 3619 / +82 2 831 3615 FAX: +82 50 4031 3619 Email: info@wisecompany.org
#### 03. Subject Device Identification
Trade Name: Fix2Lock (Biocomposite medial, Biocomposite lateral, Biocombi Self Punching) Common Name: Bioabsorbable bone anchor Classification Name: fastener, fixation, biodegradable, soft tissue Classification Product Code: MAI Subsequent Product Code: MBI Panel: Orthopedic Regulation Number: 21 CFR 888.3030 Device Class: II
#### 04. Predicate device
510(k) Number: K202806, K231326 Device Name: Fix2Lock (Biocomposite medial, Biocomposite lateral, Biocombi Self Punching) Manufacturer: Osteonic Co., Ltd.
#### 05. Reference device
N/A
#### 06. Device Description
The Fix2Lock (Bioabsorbable bone anchor) is an absorbable bone fixation screw that fixes soft tissues such as ligament, tendon, and the articular capsules to bone, and is used in orthopedic surgery.
This product consists of an absorbable and/or non-absorbable suture and driver shaft and handle.
#### 07. Indication for use
The Fix2Lock is intended use for fixation of soft tissue to bone, using suture, in the following procedure; shoulder, foot/ankle, knee, hand/wrist and elbow.
Shoulder: Rotator Cuff Repair, Bankart Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, digital tendon transfers, Mid-foot reconstruction
{5}------------------------------------------------
Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis
Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of collateral
ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints for all digits, digital tendon transfers Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction
{6}------------------------------------------------
# 08. Non-Clinical Test Conclusion
Bench tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the subject device complies with the following standards:
# ▪ Material
- ASTM F2026: 2017 Standard specification for polyetheretherketone(PEEK) polymers for surgications - ASTM F2848: 2021 Standard Specification for Medical-Grade Ultra-High Molecular Weight Polyethylene Yarns and Surgical Implants
- ASTM F2502: 2017 Standard Specification and Test Methods for Absorbable Plates and Screws for Internal Fixation Implants
# ▪ Mechanical performance
- ASTM F543: 2013 Standard specification and test methods for metallic medical bone screws
- ASTM F2502: 2017 Standard Specification and Test Methods for Absorbable Plates and Screws for Internal Fixation Implants
#### ▪ Sterilization, shelf-life and packaging for sterile product
- ISO 11135:2014, Sterilization of health-care products Ethylene oxide Requirements for the development validation and routine control of a sterilization and routine control of a sterilization process for medical devices
- ISO 11138-1:2017, Sterilization of health care products Biological indicators Part 1: General requirements
- ISO 11138-2:2017, Sterilization of health care products Biological indicators Part 2: Biological indicators for ethylene oxide sterilization processes
- ISO 11140-1:2014, Sterilization of health care products -- Chemical indicators -- Part 1: General requirements
- ISO 11737-1 Third edition 2018-01 [Including AMD1:2021] Sterilization of health care products Microbiological methods - Part 1: Determination of a population of microorganisms on product [Including Amendment 1 (2021)]
- ISO 11737-2:2019 Sterilization of medical devices Microbiological methods Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process
- ISO 11607-1:2019 Packaging for terminally sterilized medical devices part 1: requirements for materials, sterile barrier systems and packaging system
- ISO 11607-2:2019 Packaging for terminally sterilized medical devices Part 2: Validation requirements for forming, sealing and assembly processes
- ASTM F1980:2016 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
- ASTM F88/F88M:2021 Standard test method for seal strength of flexible barrier materials.
- ASTM F1929:2015 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
- Bacterial Endotoxin
- USP <85> Bacterial Endotoxin Test
- USP <161> Medical Devices-Bacterial Endotoxin and Pyrogen Tests
- ISO 11737-3:2023, Sterilization of health care products Microbiological methods Part 3: Bacterial endotoxin testing
# 09. Comparison of Technological Similarities and Differences
#### Table 1: Substantial Equivalence Comparison
| Product Name | SUBJECT Device | Predicate Device<br>Fix2Lock(Biocomposite medial, lateral,<br>Biocombi Self Punching)(K231326,<br>K202806) | Equivalence<br>Discussion |
|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Product code | MAI | MAI | Same |
| Regulatory class | Class II | Class II | Same |
| Regulation Number | 21 CFR 888.3030 | 21 CFR 888.3030 | Same |
| Intended use | The Fix2Lock is intended use for fixation<br>of soft tissue to bone, using suture, in the<br>following procedure;<br>shoulder, foot/ankle, knee, hand/wrist and<br>elbow.<br>Shoulder: Rotator Cuff Repair, Bankart<br>Repair, SLAP Lesion Repair, Biceps<br>Tenodesis, Acromio-Clavicular Separation. | The Fix2Lock is intended use for fixation<br>of soft tissue to bone, using suture, in the<br>following procedure;<br>shoulder, foot/ankle, knee, hand/wrist and<br>elbow.<br>Shoulder: Rotator Cuff Repair, Bankart<br>Repair, SLAP Lesion Repair, Biceps<br>Tenodesis, Acromio-Clavicular Separation. | Same |
{7}------------------------------------------------
| | Repair, Deltoid Repair, Capsular Shift or<br>Capsulolabral Reconstruction<br>Foot/Ankle: Lateral Stabilization, Medial<br>Stabilization, Achilles Tendon Repair,<br>Metatarsal Ligament Repair, Hallux<br>Valgus reconstruction, digital tendon<br>transfers, Mid-foot reconstruction<br>Knee: Medial Collateral Ligament Repair,<br>Lateral Collateral Ligament Repair, Patellar<br>Tendon Repair, Posterior Oblique<br>Ligament Repair, Iliotibial Band Tenodesis<br>Hand/Wrist: Scapholunate Ligament<br>Reconstruction, Carpal Ligament<br>Reconstruction, Repair/Reconstruction of<br>collateral<br>ligaments, Repair of Flexor and Extensor<br>Tendons at the PIP, DIP, and MCP joints<br>for all digits, digital tendon transfers<br>Elbow: Biceps Tendon Reattachment,<br>Ulnar or Radial Collateral Ligament<br>Reconstruction | Repair, Deltoid Repair, Capsular Shift or<br>Capsulolabral Reconstruction<br>Foot/Ankle: Lateral Stabilization, Medial<br>Stabilization, Achilles Tendon Repair,<br>Metatarsal Ligament Repair, Hallux<br>Valgus reconstruction, digital tendon<br>transfers, Mid-foot reconstruction<br>Knee: Medial Collateral Ligament Repair,<br>Lateral Collateral Ligament Repair, Patellar<br>Tendon Repair, Posterior Oblique<br>Ligament Repair, Iliotibial Band Tenodesis<br>Hand/Wrist: Scapholunate Ligament<br>Reconstruction, Carpal Ligament<br>Reconstruction, Repair/Reconstruction of<br>collateral<br>ligaments, Repair of Flexor and Extensor<br>Tendons at the PIP, DIP, and MCP joints<br>for all digits, digital tendon transfers<br>Elbow: Biceps Tendon Reattachment, Ulnar<br>or Radial Collateral Ligament<br>Reconstruction | |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Operating<br>Principles | Bone fixation screw that ties soft tissues<br>such as ligament, tendon, and the articular<br>capsules to bone. | Bone fixation screw that ties soft tissues<br>such as ligament, tendon, and the articular<br>capsules to bone. | Same |
| | | | |
| Material | Anchor:<br>-Biocomposite medial, lateral: PLGA+β-<br>TCP<br>-Biocombi self punching: PLGA+β-TCP,<br>PEEK<br>Suture: UHMWPE | Anchor:<br>-Biocomposite medial, lateral: PLGA+β-<br>TCP<br>-Biocombi self punching: PLGA+β-TCP,<br>PEEK<br>Suture: UHMWPE | Same |
| Structure | This product consists of an absorbable<br>and/or non-absorbable anchor, a non-absorbable suture, needle and driver shaft<br>and handle. | This product consists of an absorbable<br>and/or non-absorbable anchor, a non-absorbable suture and driver shaft and<br>handle. | Same |
| Product<br>Size | Anchor diameter: Ø 3.75 to 6.5 mm<br>Anchor length: 12.0 to 14.0mm | Anchor diameter: Ø 2.6 to 6.5 mm<br>Anchor length: 10.0 to 20.8mm | Similar |
| Sterilization | Sterile (EtO sterilization) | Sterile (EtO sterilization) | Same |
| Single Use/<br>Reuse | Single use | Single use | Same |
| Packaging | 1 EA / BOX | 1 EA / BOX | Same |
| Shelf-life | 3Years | 3Years | Same |
The Product Size of subject device is similar with the predicate device. It is not exactly same but the is within the range of unmodified device and the safety was evaluated as the performance bench test.
# 10. Substantially Equivalent Conclusion
Based on above, the subject device, Fix2Lock (Biocomposite medial, lateral, Biocombi Self punching), is determined to be Substantially Equivalent (SE) to the predicate device(Unmodified device), Fix2Lock(Biocomposite medial, lateral, Biocombi Sef pundhing), (K202806, K231326) in respect of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.