K243452 · Ziehm-Orthoscan, Inc. · OXO · Jan 14, 2025 · Radiology
Device Facts
Record ID
K243452
Device Name
Orthoscan VERSA Mini C-Arm
Applicant
Ziehm-Orthoscan, Inc.
Product Code
OXO · Radiology
Decision Date
Jan 14, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Orthoscan VERSA Mini C-arm X-ray system is designed to provide physicians with general fluoroscopic visualization, using pulsed or continuous fluoroscopy, of a patient including but not limited to, diagnostic, surgical, and critical emergency care procedures for patients of all ages including pediatric populations when imaging limbs/extremities, shoulders, at locations including but not limited to, hospitals, ambulatory surgery, emergency, traumatology, orthopedic, critical care, or physician office environments.
Device Story
Mobile fluoroscopic mini C-arm system; provides real-time X-ray visualization of extremities/shoulders. Inputs: X-ray radiation passing through patient anatomy; captured by flat panel detector. Transformation: System processes X-ray signals into digital fluoroscopic images; displays on integrated workstation monitor. Operation: Used by physicians/surgeons in OR, clinic, or emergency settings. Output: Live fluoroscopic images; used for intraoperative guidance, diagnostic assessment, and surgical monitoring. Benefits: Enables minimally invasive procedures; provides high-quality imaging with optimized dose. Modifications: Upgraded Linux OS (Ubuntu 20.04), improved generator power management, alternate radiation shielding, and updated wireless footswitch communication (code-hopping).
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including system performance, imaging performance, and usability testing. Radiologist assessment of images from anthropomorphic/anatomical phantoms confirmed equivalent or improved image quality compared to predicate at similar dose rates.
Technological Characteristics
Mobile fluoroscopic X-ray system; flat panel detector; pulsed/continuous fluoroscopy. Upgraded Linux OS (Ubuntu 20.04). Generator PCB for power efficiency. Alternate radiation shielding material. Wireless footswitch (code-hopping protocol). Compliant with IEC 60601-1 Ed 3.2 and IEC 60601-2-54. DICOM-compatible wired/wireless networking.
Indications for Use
Indicated for general fluoroscopic visualization of patients of all ages, including pediatric populations, during diagnostic, surgical, and critical emergency care procedures involving limbs, extremities, and shoulders in clinical settings such as hospitals, ambulatory surgery centers, and physician offices.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 14, 2025
Ziehm-Orthoscan, Inc. % Kevin Bridgman VP of Regulatory Affairs and Quality Assurance 14555 N. 82nd St. SCOTTSDALE, AZ 85260
Re: K243452
Trade/Device Name: Orthoscan VERSA Mini C-Arm Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II Product Code: OXO, JAA, MQB Dated: November 21, 2024 Received: November 21, 2024
Dear Kevin Bridgman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K243452
Device Name Orthoscan VERSA Mini C-Arm
#### Indications for Use (Describe)
The Orthoscan VERSA Mini C-arm X-ray system is designed to provide physicians with general fluoroscopic visualization, using pulsed or continuous fluoroscopy, of a patient including but not limited to, diagnostic, surgical, and critical emergency care procedures for patients of all ages including pediations when imaging limbs/extremities, shoulders, at locations including but not limited to, hospitals, ambulatory surgency, traumatology, orthopedic, critical care, or physician office environments.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <span style="font-style: normal;"> <span style="font-weight: normal;"> <span style="text-decoration: none;"> <span style="vertical-align: baseline;"> </span> </span> </span> </span> </span> </span> <span style="font-family: Wingdings;"> <span style="font-size: 10pt;"> <span style="font-style: normal;"> <span style="font-weight: normal;"> <span style="text-decoration: none;"> <span style="vertical-align: baseline;"> </span> </span> </span> </span> </span> </span> <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <span style="font-style: normal;"> <span style="font-weight: normal;"> <span style="text-decoration: none;"> <span style="vertical-align: baseline;"> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> </span> </span> </span> </span> | <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <span style="font-style: normal;"> <span style="font-weight: normal;"> <span style="text-decoration: none;"> <span style="vertical-align: baseline;"> </span> </span> </span> </span> </span> </span> <span style="font-family: Wingdings;"> <span style="font-size: 10pt;"> <span style="font-style: normal;"> <span style="font-weight: normal;"> <span style="text-decoration: none;"> <span style="vertical-align: baseline;"> </span> </span> </span> </span> </span> </span> <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <span style="font-style: normal;"> <span style="font-weight: normal;"> <span style="text-decoration: none;"> <span style="vertical-align: baseline;"> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> </span> </span> </span> </span> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Traditional 510(k) Summary Traditional 510 (k) Premarket Notification Submission - Ziehm-Orthoscan. Inc. VERSA Mini C-Arm
# Submitter: Ziehm-Orthoscan, Inc. 14555 N 82nd St. Scottsdale, AZ 85260 Phone: (480) 503-8010 Fax: (480) 503-8011 Primary Contact: Kevin Bridgman; Kevin.bridgman@ziehm-orthoscan.com Secondary Contact: TuAnh Ngo; tuanh.ngo@ziehm-orthoscan.com Date: October 1, 2024
In accordance with the requirements of 21 CFR §807.92 the following Traditional 510(k) summary of information is provided:
| Device Trade Name: | Orthoscan VERSA Mini C-Arm |
|-----------------------|-----------------------------------------------------------------|
| 510(k) Number: | K243452 |
| Common/Usual Names: | Fluoroscopic X-Ray System, Mobile Mini Mobile C-arm, Mini C-arm |
| Device Class: | Class II |
| Classification(s): | 21CFR 892.1650 |
| Classification Names: | Image-intensified fluoroscopic x-ray system |
| Product Code: | OXO, JAA, MQB |
| Primary Predicate: | Orthoscan TAU Mini C-Arm |
|-----------------------|---------------------------------------------|
| 510(k) Number: | K213113 |
| Classification(s): | 21CFR 892.1650 |
| Device Class: | Class II |
| Classification Names: | Image-intensified fluoroscopic x-ray system |
| Product Codes: | OXO, JAA, MQB |
| Secondary Predicate: | Orthoscan Mobile DI Mini C-Arm |
|-----------------------|---------------------------------------------|
| 510(k) Number: | (K113708) |
| Classification(s): | 21CFR 892.1650 |
| Device Class: | Class II |
| Classification Names: | Image-intensified fluoroscopic x-ray system |
| Product Codes: | OXO, JAA, MQB |
# General Description:
The proposed modifications to Ziehm-Orthoscan, Inc. VERSA Mini C-Arm series (which we will refer to internally and in this submittal as Orthoscan VERSA, for distinction from predicate Orthoscan TAU and Orthoscan Mobile DI) retain identical function as the predicate TAU Mini C-arm (K2131130) and predicate Mobile DI (K113708) as a mobile fluoroscopic mini C-arm system that provides fluoroscopic images of patients of all ages during diagnostic, treatment and surgical procedures involving anatomical regions such as but not limited to that of extremities, limbs, shoulders and knees. The system consists of C-arm support attached to the image workstation.
The changes to the Orthoscan VERSA Mini C-arm X-ray system represent a modification of our presently legally marketed device Orthoscan TAU Mini C-Arm K213113 and Orthoscan Mobile DI Mini C-arm (K113708). The proposed modifications to the predicate encompass the implementation of a LINUX based operating system upgrade from Ubuntu version 16.04 to Ubuntu version 20.04 and related software revisions, modification to the mechanical design to further facilitate desk top use, revisions to
{5}------------------------------------------------
generator printed circuit board to improve power management efficiency, implementation of an alternate generator radiation shielding material to reduce environmental impact of lead and an update to wireless footswitch communication protocol.
The proposed device replicates the features and functions of the predicate devices without impacting image clarity or dose levels.
For both the predicate and proposed device, the following are unchanged: C-arm support of flat panel detector, generator and x-ray controls, mechanical connections, balancing, locking, rotations, work-station platform, monitor display and main user interface controls, touch screen interface, selectable imaging, Xray technique control, entry of patient information, wired or wireless footswitch operation, interface connection panel and DICOM fixed wire and wireless network interfaces.
#### Indications for Use:
The Orthoscan VERSA Mini C-arm X-ray system is designed to provide physicians with general fluoroscopic visualization, using pulsed or continuous fluoroscopy, of a patient including but not limited to, diagnostic, surgical, and critical emergency care procedures for patients of all ages including pediatric populations when imaging limbs/extremities, shoulders, at locations including but not limited to, hospitals, ambulatory surgery, emergency, traumatology, orthopedic, critical care, or physician office environments.
# Indications for Use Comparison:
The proposed modifications to Ziehm-Orthoscan, Inc. VERSA Mini C-Arm series do not change the indications for use. The indications for use are the same as the predicate device TAU (K213113) and Mobile DI (K113708) Mini C-Arms.
#### Technology:
The changes to the proposed device Orthoscan VERSA Mini C-arm X-ray systems represent a modification of our presently legally marketed device Orthoscan TAU Mini C-Arm (K213113) and Orthoscan Mobile DI (K113708). With the introduction of a Linux based operating system (OS) upgrade from Ubuntu 16.04 to Ubuntu 20.04, an open source highly secure OS, cybersecurity controls are improved, image quality is maintained or improved, and a capability is created for future expansions of the proposed VERSA device. Modification to the predicate Orthoscan Mobile DI model 1000-0005 (K113708) mechanical design further facilitates desk top use by relocating the image processing printed circuit card, power manager circuit card and related hardware and cabling behind the display electronics within the monitor display assembly. Revisions to generator printed circuit board improve power management efficiency and implementation of an alternate generator radiation shielding material reduces environmental impact of lead. Updating the communication protocol of wireless footswitch from fixed pairing method to code-hopping method maintains state of the art wireless communications and perpetuate free operator movement with fewer cables in the way.
# Summary of Technological Characteristics:
The comparisons of the proposed modifications to Ziehm-Orthoscan, Inc. VERSA Mini C-Arm demonstrate that the scientific and technology characteristics indicate substantial equivalence to the predicate device Orthoscan TAU Mini C-Arm (K213113) and Orthoscan Mobile DI Mini C-arm (K113708).
# Adverse Effects on Health:
The proposed modified Ziehm-Orthoscan, Inc. VERSA Mini C-arm's potential radiation, mechanical, and electrical hazards are identified and analyzed as part of risk management and controlled by meeting the applicable CDRH 21CFR subchapter J performance requirements, Recognized Consensus Standards, designing and manufacturing under Ziehm-Orthoscan, Inc. Quality System, and system verification and validation testing ensure the device performs to the product specifications and its intended use. The adherence to these applicable regulations and certification to Recognized Consensus Standards that apply to this product provides the assurance of device safety and effectiveness.
{6}------------------------------------------------
# Non-Clinical Testing Summary
Verification and Validation, including hazard mitigations and usability testing executed in demonstrated system that met Design Input and user needs.
The device was tested by certified test laboratory resulting in device being certified compliant with 60601-1 ED 3.2 series, including IEC 60601-2-54. Further, the device met all applicable sections of 21 CFR Subchapter J performance standards.
The proposed modified Ziehm-Orthoscan, Inc. VERSA Mini C-Arm development occurred under our design control processes, software development processes, and overall quality management system. They included but are not limited to;
- Risk Analysis .
- . Required reviews
- . Design reviews
- Component testing
- . Integration testing
- . Performance testing
- . Safety testing
- . Product use and usability testing
Performance bench testing included non-clinical testing methods specific to quidance for submission of 510(k)s for Solid State X-Ray Imaging Devices (SSXI); demonstrating system and imaging performance. Non-clinical image and dose lab testing were employed. Anthropomorphic (PMMA material) phantoms and anatomical simulation phantoms were employed and images were taken by both the proposed VERSA and the predicate device Orthoscan TAU (K213113). Numerous image comparison sets were taken, and a Radiologist performed an assessment of individual images arranged in groups of image sets. His conclusion was that the image quality at same or similar patient dose rates will result in equivalent or slight improvement in patient care (images) for the proposed modified VERSA device over the predicate device. Therefore, Ziehm-Orthoscan, Inc. believes the VERSA Mini C-arm image quality, safety and effectiveness to be substantially equivalent to that of the predicate device Orthoscan TAU (K213113) and Orthoscan Mobile DI (K113708).
Usability testing included confirming the usability specifications of the proposed modified Ziehm-Orthoscan, Inc. VERSA Mini C-Arm, as well as the user interface of the device, especially the interfaces that have changed since the predicate devices. Usability testing concluded that there were no previously unknown use errors or hazardous situations and no unacceptable residual risks due to the changes in user interface from the predicate device Orthoscan TAU (K213113) and Orthoscan Mobile DI (K113708).
# Summary of Clinical Test Data:
Ziehm-Orthoscan, Inc. VERSA mobile fluoroscopic Mini C-arm system did not require live human clinical studies to support substantial equivalence in accordance with the related guidance documents, Premarket Assessment of Pediatric Medical Devices May 24, 2014 and Pediatric Information for X-ray Imaging Device Premarket Notifications Nov 28, 2017.
Therefore, Ziehm-Orthoscan, Inc. conducted a lab test image comparison study employing the use of anthropomorphic phantoms in establishing substantial equivalence based on the modifications to the proposed device and the bench data taken with Ziehm-Orthoscan, Inc. VERSA Mini C-Arm in comparison to the predicate Orthoscan TAU Mini C-Arm (K213113).
# Substantial Equivalence Conclusion:
The changes and differences of the proposed Ziehm-Orthoscan, Inc. VERSA Mini C-Arm described do not change the control mechanism, operating principle, energy type, or intended use found on the predicate device Orthoscan TAU Mini C-Arm (K213113) and Orthoscan Mobile DI Mini C-arm (K113708).
Ziehm-Orthoscan, Inc. considers the proposed modified VERSA Mini C-arm to be as safe, as effective, and performs substantially equivalent to the predicate device Orthoscan TAU Mini C-arm (K213113) and Orthoscan Mobile DI Mini C-arm (K113708) in accordance with its labeling.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.