K243436 · Basis Medical · DQY · Feb 28, 2025 · Cardiovascular
Device Facts
Record ID
K243436
Device Name
Seclusion Catheter
Applicant
Basis Medical
Product Code
DQY · Cardiovascular
Decision Date
Feb 28, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Seclusion Catheter is indicated for localized infusion and aspiration of various therapeutic and diagnostic agents in isolated segments of the peripheral venous vasculature. The Seclusion Catheter may be repositioned for multiple treatments with the same patient. The Seclusion Catheter is not indicated for use in the neurovasculature.
Device Story
Seclusion Catheter is a percutaneous, single-use, four-lumen catheter designed for localized infusion and aspiration of therapeutic/diagnostic agents within isolated peripheral venous segments. Device features two balloons independently inflatable with air to create a 7 cm treatment area. Operated by clinicians in a clinical setting; catheter is inserted percutaneously and positioned to isolate a specific vessel segment. Control unit allows for balloon inflation/deflation, infusion, and aspiration. By isolating the target segment, the device enables targeted delivery and removal of agents, potentially improving therapeutic efficacy while minimizing systemic exposure. Device is EtO sterilized.
Clinical Evidence
No clinical data. Evidence consists of bench testing including visual/dimensional, cycle/simulated use, physical characteristics, functional, usability, and packaging testing. Biocompatibility testing performed per ISO 10993-1 for external communicating device with circulating blood contact (<24 hours).
Technological Characteristics
Four-lumen percutaneous catheter; two air-inflatable balloons (4-12 mm diameter); 7 cm treatment region; 7Fr sheath compatible; 0.014"/0.018" guidewire compatible; EtO sterilization; no coating; no integrated sensors.
Indications for Use
Indicated for localized infusion and aspiration of therapeutic/diagnostic agents in isolated peripheral venous vasculature segments. Not for neurovasculature use.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
February 28, 2025
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below that. The logo is simple and professional, and it is easily recognizable.
Basis Medical Grace Powers Founder/ Principal Consultant Powers Regulatory Consulting 2451 Cumberland Blvd SE Suite 3740 Atlanta, Georgia 30339
Re: K243436
Trade/Device Name: Seclusion Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II Product Code: DOY Dated: November 4, 2024 Received: November 5, 2024
Dear Grace Powers:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/2/Picture/3 description: The image shows a digital signature. The signature is for "LYDIA S. GLAW -S". The date of the signature is February 28, 2025, and the time is 14:55:23 -05'00'.
Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K243436
Device Name Seclusion Catheter
Indications for Use (Describe)
The Secusion Catheter is indicated for localized infusion and aspiration of various therapeutic agents in isolated segments of the peripheral venous vasculature. The Seclusion Catheter may be repositioned for multiple treatments with the same patient. The Seclusion Catheter is not indicated for use in the neurovasculature.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------|
| <span style="font-size:16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span style="font-size:16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the Seclusion Catheter Traditional 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on August 12, 2005.
- Sponsor: Basis Medical 4080 McGinnis Ferry Road Building 300, Suite 304 Alpharetta, GA 30005 Tel: 404-308-0071 Submission Contact: Grace Powers, FRAPS, MS, MBA, RAC Founder/Principal Consultant Powers Regulatory Consulting Tel: 404-931-8730
Submission Date: November 4, 2024
## Subject Device:
| Trade Name: | Seclusion Catheter |
|----------------------------|-----------------------|
| Classification Regulation: | 21 CFR § 870.1250 |
| Class: | 2 |
| Common Name: | Percutaneous Catheter |
| Panel: | Cardiovascular |
| Product Code: | DQY |
| | |
Predicate Device:
Legally marketed device to which substantial equivalence is claimed: Occlusion Perfusion Catheter (K153488)
## Device Description
The Seclusion Catheter is indicated for localized infusion and aspiration of various therapeutic and diagnostic agents in isolated segments of the peripheral venous vasculature. The Seclusion Catheter may be repositioned for multiple treatments with the same patient. A photograph of the Seclusion Catheter is shown in the figures below.
{5}------------------------------------------------
## 510(k) Summary
Image /page/5/Picture/2 description: The image shows a medical device called the Setclusion Catheter with inflated balloons. The catheter is a long, thin tube with two small, transparent balloons attached to it. The catheter is connected to a white control unit with the words "BASIS MEDICAL" printed on it. The control unit has a dial that can be turned to different settings, including "DISTAL," "CLOSED," and "PROXIMAL."
## Figure 1: The Seclusion Catheter (Balloons inflated
The device is EtO sterilized and is a single-use device.
There is a device-specific guidance that is applicable to the subject device: Peripheral Percutaneous Transluminal Angioplasty (PTA) and Specialty Catheters - Premarket Notification (510(k)) Submissions.
## Indications for Use
The Seclusion Catheter is indicated for localized infusion and aspiration of various therapeutic and diagnostic agents in isolated segments of the peripheral venous vasculature. The Seclusion Catheter may be repositioned for multiple treatments with the same patient. The Seclusion™ Catheter is not indicated for use in the neurovasculature.
## Technological Characteristics
The Seclusion Catheter has similar technological characteristics as the predicate device.
| - | Subject Device:<br>Seclusion Catheter | Predicate Device:<br>Occlusion Perfusion Catheter (OPC) |
|---------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation Number | 21 CFR 870.1250 | 21 CFR 870.1250 |
| Device Classification Name | Percutaneous Catheter | Percutaneous Catheter |
| FDA Product Code | DQY | DQY |
| | Subject Device:<br>Seclusion Catheter | Predicate Device:<br>Occlusion Perfusion Catheter (OPC) |
| Classification | Class 2 | Class 2 |
| Indications for Use | The Seclusion™ Catheter is a percutaneous catheter that is indicated for localized infusion and aspiration of various therapeutic and diagnostic agents in isolated segments of the peripheral venous vasculature.<br><br>The Seclusion Catheter may be repositioned for multiple treatments with the same patient. | The Occlusion Perfusion Catheter™ is indicated for localized infusion or irrigation of various diagnostic and therapeutic agents into the peripheral vasculature. The Occlusion Perfusion Catheter™ may be repositioned for multiple treatments with the same patient. It is not indicated for use in neurovasculature. |
| Prescription or<br>OTC? | Prescription | Prescription |
| Fluids Delivered | Various therapeutic and diagnostic agents | Various diagnostic and therapeutic agents |
| Vasculature | Venous peripheral vasculature | Peripheral vasculature |
| Lumen and Tip | Four (4) lumen catheter with rounded tip. | Five (5) lumen catheters with rounded tip. |
| Number of<br>Balloons | Two (2) balloons that can be independently inflated to create the treatment area. | Two (2) balloons that can be simultaneously inflated to create the treatment area and one (1) center balloon to occupy space in the treatment area. |
| Balloon<br>Diameter | 4 mm to 12 mm | 3mm to 10mm |
| Balloon<br>inflation | Air | Saline |
| Treatment<br>region length | The treatment area between balloons is 7 cm. | The treatment chamber is between balloons. There are multiple sizes available from 3cm to 15cm. |
| Maximum<br>Pressure/ Rated<br>Balloon<br>Pressure | Rated burst volume (RBV) no less than 1.5x the volume required to reach the maximum balloon diameter. Additionally, the minimum burst volume is no less than 6 mL ambient air when using a 3 mL syringe to inflate the balloons. | 4 atm |
| Coating | No catheter coating. | No catheter coating. |
| Guidewire<br>compatibility | 0.014" and 0.018" guidewire compatibility (via the aspiration lumen for stiffness if needed). The guidewire does not pass through the tip. | 0.014" guidewire compatibility. There is a lumen used solely for the guidewire. The guidewire does pass through the tip. |
| Sheath<br>Compatibility | Compatible with a 7Fr sheath. | Unknown |
| | Subject Device:<br>Seclusion Catheter | Predicate Device:<br>Occlusion Perfusion Catheter (OPC) |
| Number and<br>types of ports | Three (3) ports:<br>1: Balloon inflation/deflation<br>2: Aspiration<br>3: Infusion | Five (5) ports:<br>1: Fluid Inflow (Infusion)<br>2: Fluid Outflow (Aspiration)<br>3: Occlusion Balloon inflation/ deflation<br>4: Center Balloon Inflation/deflation<br>5: Guidewire |
| Sensor<br>technology | None | The OPC has pressure sensing technology. |
| Sterilization | Ethylene Oxide (EO) | Ethylene Oxide (EO) |
## Subject Device compared to the Predicate Device
{6}------------------------------------------------
{7}------------------------------------------------
## 510(k) Summary
## Performance Data
The Seclusion Catheter was subjected to ship testing, packaging testing and functional testing. The testing was based off the FDA guidance document listed below that is applicable to the subject device.
- Peripheral Percutaneous Transluminal Angioplasty (PTA) and Specialty Catheters Premarket ● Notification (510(k)) Submissions: Guidance for Industry and Food and Drug Administration Staff issued on April 14, 2023.
The following testing was conducted:
- Visual and Dimensional Testing
- Cycle and Simulated Use Testing
- Physical Characteristics Testing
- Functional Testing
- Usability Testing
- Packaging Testing
- Biocompatibility testing per ISO 10993-1 for an external communicating device with circulating blood contact for a limited duration (less than 24 hours)
## Conclusion
The Seclusion Catheter is substantially equivalent to the legally marketed predicate device as demonstrated by the same intended use, similar technologies and performance data, and does not raise different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.