Diagnostic X-ray System

K243411 · Nanjing Perlove Medical Equipment Co., Ltd. · OWB · May 19, 2025 · Radiology

Device Facts

Record IDK243411
Device NameDiagnostic X-ray System
ApplicantNanjing Perlove Medical Equipment Co., Ltd.
Product CodeOWB · Radiology
Decision DateMay 19, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1650
Device ClassClass 2
AttributesPediatric

Indications for Use

The Diagnostic X-ray System is intended to be used and operated by: adequately trained, qualified and authorized health care professionals who have full understanding of the safety information and emergency procedures as well as the capabilities and functions of the device. The device is used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the device. The device is to be used in health care facilities both inside and outside the operating room, sterile as well as non-sterile environment in a variety of procedures.

Device Story

Mobile C-arm fluoroscopic X-ray system for real-time imaging in clinical/surgical settings. Input: X-ray radiation attenuated by patient anatomy. Operation: X-ray source assembly emits radiation; flat-panel detector converts X-rays to digital signals; image processing workstation displays images on C-arm monitor. C-arm allows multi-axis positioning for various angles. Operated by trained healthcare professionals. Output: Real-time fluoroscopic images used for guidance during procedures. Benefits: Enables visualization of internal structures during interventional/surgical tasks, facilitating clinical decision-making.

Clinical Evidence

No clinical study included. Bench testing performed on detector imaging metrics. Clinical images of multiple body parts (radiography and fluoroscopy) were acquired and assessed by a qualified radiologist to demonstrate acceptable imaging performance.

Technological Characteristics

Mobile C-arm fluoroscopic system. Components: X-ray source, collimator, Thales Pixium3030S flat-panel detector (amorphous silicon, Cesium Iodide scintillator, 12"x12"), image processing workstation. Connectivity: Digital image capture/manipulation. Standards: IEC 60601-1-2, IEC 60601-2-54, IEC 60601-1-3, IEC 60601-2-43, ISO 10993-1. Software: Embedded and workstation-based.

Indications for Use

Indicated for radiological guidance and visualization during diagnostic, interventional, and surgical procedures for all patients except neonates (birth to one month).

Regulatory Classification

Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION May 19, 2025 Nanjing Perlove Medical Equipment Co., Ltd. % Jack Fang Official Correspondent APlus Healthcare Technology (Shanghai) Co., Ltd. Room 223, Building 17, JY-WISDOMBAY, Huqing Road 158, Baoshan District SHANGHAI, 200431 CHINA Re: K243411 Trade/Device Name: Diagnostic X-ray System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II Product Code: OWB Dated: November 1, 2024 Received: April 18, 2025 Dear Jack Fang: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1} K243411 - Jack Fang Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K243411 - Jack Fang Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, ![img-0.jpeg](img-0.jpeg) Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K243411 | | | Device Name Diagnostic X-ray System | | | Indications for Use (Describe) The Diagnostic X-ray System is intended to be used and operated by: adequately trained, qualified and authorized health care professionals who have full understanding of the safety information and emergency procedures as well as the capabilities and functions of the device. The device is used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the device. The device is to be used in health care facilities both inside and outside the operating room, sterile as well as non-sterile environment in a variety of procedures. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} K243411 510(k) Summary # 510(k) Summary ## I Submitter Nanjing Perlove Medical Equipment Co., Ltd. No. 97 and No. 99 Wangxi Road, Jiangning District, Nanjing, 211100 Jiangsu, P.R. China ## Contact person: Official Correspondent: Jack Fang APlus Healthcare Technology (Shanghai) Co., Ltd. Add: Room 223, Building 17, JY-WISDOMBAY, Huqing Road 158, Baoshan District, Shanghai, 200431 China Tel: +8613161499974 E-mail: jack.fang@ap-healthcare.com Preparation date: May 17th, 2025 ## II Proposed Device Trade Name of Device: Diagnostic X-ray System Models: PLX119C Regulation name: Image-intensified fluoroscopic x-ray system Regulation Number: 21 CFR 892.1650 Regulatory Class: Class II Product code: OWB Review Panel: Radiology ## III Predicate Device 510(k) Number: K222080 Product Code: OWB Classification: 21 CFR 892.1650 Trade Name: Garion Regulatory Class: II ## IV Device description The Diagnostic X-ray System is a mobile (within an imaging facility) general-purpose diagnostic fluoroscopic X-ray system that uses a C-arm and digital techniques for image capture, display and manipulation and is designed to be used in a variety of general-purpose applications requiring real-time fluoroscopic imaging capabilities. The Diagnostic X-ray System is consists of X-ray source assembly (combined type), collimator, flat-panel detector, image processing workstation, C-arm and mobile rack, Medical Image Workstation Software. ## Principles of Operation Based on the principle that different tissues and organs of the human body have different attenuation of X-ray, the X-ray tube (an evacuated tube with an anode and a {5} K243411 510(k) Summary cathode) emits X-rays that passing through the patient's body. The detector converts the X-rays into visible images that are displayed on the C-arm monitor. The C-shaped connecting element allows movement horizontally, vertically and around the swivel axes, so that X-ray images of the patient can be produced from almost any angle. # V Indications for use The Diagnostic X-ray System is intended to be used and operated by: adequately trained, qualified and authorized health care professionals who have full understanding of the safety information and emergency procedures as well as the capabilities and functions of the device. The device is used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the device. The device is to be used in health care facilities both inside and outside the operating room, sterile as well as non-sterile environment in a variety of procedures. # VI Comparison of technological characteristics with the predicate device The comparison between the overall specifications of the predicate device and the proposed device is shown in Table 1. Comparison between the Diagnostic X-ray System and the predicate device was conducted with respect to intended use, technological characteristics and principles of operations, providing more detailed information regarding the bases for the determination of substantial equivalence. Table 1 Comparison of technological characteristics with the predicate device | Items | Proposed Device Diagnostic X-ray System | Predicate Device Garion K222080 | | --- | --- | --- | | Indications for use | The Diagnostic X-ray System is intended to be used and operated by: adequately trained, qualified and authorized health care professionals who have full understanding of the safety information and emergency procedures as well as the capabilities and functions of the device. The device is used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the device. The device is to be used in health care facilities both inside and outside the operating room, sterile as well as non-sterile environment in a variety of procedures. | The Garion is intended to be used and operated by: adequately trained, qualified and authorized health care professionals who have full understanding of the safety information and emergency procedures as well as the capabilities and functions of the device. The device is used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the device. The device is to be used in health care facilities both inside and outside the operating room, sterile as well as non-sterile environment in a variety of procedures. | {6} K243411 510(k) Summary | X-ray Tube | Anode Type | Fixed Anode | Rotating Anode | | --- | --- | --- | --- | | | Anode target angle | 15° | 10° | | | Focal size | 0.6/1.8 | 0.3/0.6 | | Fluoroscopic Mode | kV range | 40-120 kV | 40-125 kV | | | mA range | 0.3-100 mA | 0.1-100 mA | | | Pulse Fluoro | YES | YES | | | ABS function | YES | YES | | Detector | Manufacture and Model | THALES Pixium3030S | IRAY's Mercu 0909F | | | Size | 12" x 12" | 9" x 9" | | | Scintillator | Cesium Iodide | Cesium Iodide | | | Detector type | amorphous silicon detector | amorphous silicon detector | | | Active detector size | 300 mm x300 mm | 228.6 mm x228.6 mm | | | Total pixel matrix | 1956x1956 | 1024x1024 | | | Pixel pitch | 154 μm | 205 μm | | | A/D Conversion | 16 bit | 16 bit | | | MTF @ 1.0 LP/mm | Typical 59% | Typical 64% | | | DQE @ 0 LP/min | Typical 77% | Typical 77% | | C arm | SID | 1000mm | 980mm | | | Range of C-arm Rail Rotation | ±180° | ±180° | | | Range of the Linear FR-arm Movement | 200mm | 200mm | | | Range of the Liner T-arm Movement | 400m | 400m | | | Range of swing-arm Rotation | ±15° | ±15° | | Collimator | Motor control / rotation | Motor control / rotation | Motor control / rotation | 3 / 5 {7} K243411 510(k) Summary ![img-1.jpeg](img-1.jpeg) ## VII Performance Data The following performance data were provided in support of the substantial equivalence determination. Electrical safety and electromagnetic compatibility (EMC) testing were conducted on the proposed device. The system complies with IEC 60601-1-2: 2014+AMD1:2020 / EN 60601-1-2: 2015+A1: 2021, IEC 60601-2-54: 2009+AMD2:2018 / EN 60601-2-54: 2009+A2:2019, IEC 60601-1-3:2008 +A1:2013+A2:2021/EN 60601-1-3:2008+A1:2013+A2:2021 and IEC 60601-2-43:2010+A1:2017+A2:2019/ EN 60601-2-43:2010+A1:2018+A2:2020. ## Non-Clinical Testing Bench testing of the detector imaging metrics was carried out and performance characteristics were measured and compared with the predicate. Clinical images of multiple body parts were taken using radiography and fluoroscopy, with motion, to demonstrate acceptable imaging performance of the Diagnostic X-ray System. The clinical images were reviewed and assessed by a qualified radiologist. ## Software - Verification and Validation Testing Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Software Contained in Medical Devices." The Embedded software and the workstation software for this device required basic level documentation, since a failure a failure or flaw of device software function(s) would not present a hazardous situation with a probable risk of death or serious injury, either to a patient, user of the device, or others in the environment of use. - Cybersecurity The cybersecurity documentation was provided according to FDA guidance Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions. 4 / 5 {8} K243411 510(k) Summary ## Usability Usability validation was performed and documentation was provided complies with IEC 60601-1-6 and IEC 62366-1. ## Biocompatibility The biocompatibility evaluation for the proposed device was conducted in accordance with the International Standard ISO 10993-1:2018 “Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process,” as recognized by FDA. ## VIII Clinical Testing No clinical study is included in this submission. ## IX Conclusion The proposed device has the same indications for use and has similar design features and technological characteristic as the predicate device. The differences between the predicate device and the proposed device do not raise new questions of safety or effectiveness. Performance testing data demonstrates that the proposed device is safety and effectiveness as the predicated device. Accordingly, the proposed device is substantially equivalent to the predicate device. 5 / 5
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...