HyHub™ and HyHub™ Duo vial access devices are indicated for patients 17 years of age or older to allow a drug to be transferred from vials without using a needle, as prescribed, in a home environment or clinical setting.
Device Story
HyHub™ and HyHub™ Duo are single-use, disposable, non-pyrogenic vial access devices (VAD) designed for needleless transfer of medicinal liquids from vials. The device features an infusion tray with docking stations (two for HyHub™ Duo, four for HyHub™) that hold dual vial units. It utilizes spikes to pierce vials and withdraw drug products, with hydrophobic air filters facilitating venting to enable withdrawal without manual air injection. The device connects to standard syringes, infusion pumps, and infusion sets. Used in home or clinical environments, it is operated by patients or healthcare providers to support continuous drug delivery. By enabling sequential or simultaneous access to multiple vials, the device simplifies the infusion process, potentially improving patient convenience and adherence to prescribed therapy.
Clinical Evidence
Bench testing only. Testing included leak, particulate, Luer compatibility, stopper fragmentation, sterile packaging, flow, residual/injectable volume, and human factors validation. Biocompatibility was assessed per ISO 10993-1 for prolonged exposure, including cytotoxicity, sensitization, reactivity, pyrogenicity, systemic toxicity, hemolysis, chemical characterization, and drug compatibility.
Technological Characteristics
Materials include PVC (non-DEHP). Features spikes for vial piercing, hydrophobic air filters for venting, and flexible tubing. Supports up to 2 or 4 source containers. Connectivity via standard Luer connections to pumps/infusion sets. Sterilized via Gamma radiation (ISO 11137).
Indications for Use
Indicated for patients 17 years of age or older to allow needleless drug transfer from vials in home or clinical settings.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 18, 2025
Takeda Pharmaceuticals
Avanti Sawant
Associate Director Global Regulatory Affairs, Device PDT
500 Kendall Street
Cambridge, Massachusetts 02142
Re: K243404
Trade/Device Name: HyHub™ and HyHub™ Duo Vial Access Devices
Regulation Number: 21 CFR 880.5440
Regulation Name: Intravascular Administration Set
Regulatory Class: Class II
Product Code: LHI
Dated: June 20, 2025
Received: June 20, 2025
Dear Avanti Sawant:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243404 - Avanti Sawant
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K243404 - Avanti Sawant
Page 3
Sincerely,
David Wolloscheck -S
David Wolloscheck, Ph.D.
Assistant Director
DHT3C: Division of Drug Delivery and
General Hospital Devices, and
Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K243404 | |
| Device Name HyHub™ and HyHub™ Duo Vial Access Device | |
| Indications for Use (Describe) HyHub™ and HyHub™ Duo vial access devices are indicated for patients 17 years of age or older to allow a drug to be transferred from vials without using a needle, as prescribed, in a home environment or clinical setting. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Takeda
Page 1
# K243404 – 510(K) SUMMARY
## DATE SUBMITTED
18 Jul 2025
## Submitter
Takeda Pharmaceuticals U.S.A.,
Inc. 500 Kendall Street
Cambridge, MA 02142
## Contact
Avanti Sawant
Associate Director Global Regulatory Affairs CMC & Device PDT
Avanti.sawant@takeda.com
617-943-1107
## Device Information
Trade name: HYHUB™ and HYHUB™ DUO Vial Access Devices
Configurations: Vial Access Device with 2 docking stations
Vial Access Device with 4 docking stations
Common Name: I.V. Fluid Transfer Set
Classification Regulation: 880.5440
Classification Name: Intravascular administration set
Device Classification: Class II
Product Code: LHI
## Predicate Device
Trade Name: Fleboset MULTIPLE
510K#: K040456
Product Code: LHI
Manufacturer: LABORATORIOS GRIFOLS, S.A.
{5}
Takeda
Page 2
# Device Description Summary
The HYHUB™ and HYHUB™ DUO vial access devices (VAD) are a stand-alone, single-use, disposable, non-pyrogenic, gamma sterilized device, which are intended to support the infusion of two medicinal liquids, as prescribed, in a home environment or clinical setting. The VAD is designed to accommodate up to two (2) or four (4) dual vial units (DVU) to be docked onto the VAD infusion tray which allows the transfer of medicinal liquids in a sequential, needleless manner using standard connections for syringes, applicable pumps, and infusion sets.
# Environment of Use
The HYHUB™ and HYHUB™ DUO vial access devices are used in clinical and home environments.
# Indications for Use
HYHUB™ and HYHUB™ DUO vial access devices are indicated for patients 17 years of age or older to allow a drug to be transferred from vials without using a needle, as prescribed, in a home environment or clinical setting.
# Summary of Technological Characteristics
The following technological characteristic of the HYHUB™ and HYHUB™ DUO vial access devices are substantially equivalent to those of the Predicate, Fleboset MULTIPLE:
- Use of a spike to pierce vials and withdraw drug product(s).
- Multiple spikes connected in series with flexible tubing segments.
- Spike “vent” function facilitated by hydrophobic air filter allowing for withdrawal from vial without having to inject air.
- Enables continuous delivery of drug product(s) from multiple source containers.
- Can be used in conjunction with an infusion pump.
{6}
Takeda
Page 3
# Summary of Comparison with Predicate Device
| Characteristics | HYHUB™ and HYHUB™ DUO Vial Access Device | Fleboset Multiple | Comparison |
| --- | --- | --- | --- |
| Indications for Use | HYHUB™ and HYHUB™ DUO vial access devices are indicated for patients 17 years of age or older to allow a drug to be transferred from vials without using a needle, as prescribed, in a home environment or clinical setting. | Fleboset Multiple is an ancillary device used as fluid pathway through which substances from six glass source flasks containing the same solution may be continuously delivered for I.V. administration when used in conjunction with a gravity or pump infusion set to channel the solution from the source containers to the infusion set. | Same |
| Sterilization Method | Gamma Radiation Validation per ISO-11137 | Ethylene Oxide Validation per ISO-11135 | Different |
| Discussion: Both predicate and subject devices were validated to the sterility assurance levels of 10^{-6}. The results demonstrated that the differences do not raise any new questions of safety and effectiveness for the subject device. | | | |
| Spike | Yes | Yes | Same |
| Fluid Path | Two | One | Different |
| Discussion: The predicate device contains six (6) spikes connected in series with flexible tubing segments, made up of Polyvinyl Chloride (PVC). Similarly, the subject device has two separate fluid paths which contain spikes connected in series with tubing segments, made of PVC (non-DEHP). The subject device fluid paths have undergone flow testing, residual volume testing and biocompatibility testing to support the intended use. The results demonstrated that the differences do not raise any new questions of safety and effectiveness for the subject device. | | | |
| Hydrophobic Air Filter | Yes | Yes | Same |
| Single use | Yes | Yes | Same |
| Pyrogenicity | Non-pyrogenic | Non-pyrogenic | Same |
| Source Containers | Up to 2 (HYHUB DUO)
Up to 4 (HYHUB) | Up to 6 | Different |
| Discussion: The predicate device allows withdrawal of drug product from six (6) source containers simultaneously. The subject device allows withdrawal of the drug product from up to two (2) or four (4) source containers simultaneously. Both predicate and subject device can be connected to an infusion pump. Pump compatibility with the VAD was assessed via flow testing and residual volume testing to support the intended use. The results demonstrated that the differences do not raise any new questions of safety and effectiveness for the subject device. | | | |
{7}
Takeda
Page 4
# Performance Bench Testing Summary
The following bench tests were performed in accordance with applicable FDA-recognized standards or internally validated protocols guided by a risk management process:
- Leak test
- Particulate test
- Luer connectors compatibility
- Stopper fragmentation test
- Sterile packaging test
- Flow test
- Residual/Injectable Volume test
- Human Factors Validation
# Biocompatibility Testing Summary
Biocompatibility evaluation was conducted in accordance with ISO 10993-1, considering the device's classification (externally communicating, blood path indirect, with prolonged exposure duration (>24 h to 30 d)). The following endpoints were assessed:
- Cytotoxicity
- Sensitization
- Intracutaneous reactivity
- Material mediated pyrogenicity
- Acute systemic toxicity
- Subacute Toxicity (Dual Route Repeated Exposure)
- Hemolysis test
- Chemical Characterization
- Drug Compatibility
# Conclusions
Based upon the comparison of the labeling, the technological and functional characteristics, the information provided in this submission – including results from risk assessments and testing activities related to verification, validation, performance, and usability/human factors – demonstrates that the VAD device functions as intended and is substantially equivalent to the predicate device. Therefore, Takeda concludes that the subject device, VAD, is substantially equivalent to the predicate device (K040456).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.