Minnesota Medical Technologies Fecal Incontinence Insert (My Miracle)
Applicant
Minnesota Medical Technologies
Product Code
PBP · Gastroenterology, Urology
Decision Date
Jul 22, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The fecal incontinence insert is indicated for the management of accidental bowel leakage (ABL) due to bowel incontinence. The rectal insert is designed for self-insertion to seal and help prevent the involuntary leakage of stool from the rectum.
Device Story
Device is a single-use, soft, liquid-filled silicone rectal insert with a plastic applicator. Patient self-inserts device via anus into anal canal/rectum; device conforms to anorectal junction to seal and prevent involuntary stool leakage. Applicator is removed and discarded after placement. Used by patients in home setting to manage ABL. Output is physical sealing of the anal canal. Benefits include reduction in ABL episodes, improving patient quality of life by providing mechanical control of fecal matter.
Clinical Evidence
Prospective, open-label, single-arm, non-randomized study (n=124 enrolled, 58 in effectiveness analysis). Primary endpoint: relative percentage change in ABL episodes. 75.9% of subjects (95% CI: 62.8%, 86.1%) achieved ≥50% reduction in ABL (p < .0001). Safety profile: 15 AEs reported; no device-related serious AEs or malfunctions. Anoscopy performed on subset post-treatment showed no insert-related trauma.
Technological Characteristics
Materials: liquid-filled silicone insert, plastic applicator. Design: bar-bell shape with internal/external retention features and central shaft. Sizes: Standard and Large. Provided non-sterile. Biocompatibility: tested per ISO 10993-1, -3, -5, -10, -11. Mechanical testing: strength, detachment, puncture, buckling, bending, and simulated use sealing function.
Indications for Use
Indicated for management of accidental bowel leakage (ABL) due to bowel incontinence in adult patients (18+) who have failed conservative medical or biofeedback therapy.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 22, 2025
Minnesota Medical Technologies
% Gregory Sachs
Consultant
Sachs & Associates, Inc.
16 Crescent Lane
North Oaks, Minnesota 55127
Re: K243367
Trade/Device Name: Minnesota Medical Technologies Fecal Incontinence Insert (My Miracle)
Regulation Number: 21 CFR 876.5980
Regulation Name: Gastrointestinal Tube and Accessories
Regulatory Class: Class II
Product Code: PBP
Dated: June 17, 2025
Received: June 17, 2025
Dear Gregory Sachs:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243367 - Gregory Sachs
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K243367 - Gregory Sachs
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# STEPHANIE COLE -S
for Anthony C. Lee, Ph.D., M.B.A.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243367
Device Name
Minnesota Medical Technologies Fecal Incontinence Insert (My Miracle)
Indications for Use (Describe)
The fecal incontinence insert is indicated for the management of accidental bowel leakage (ABL) due to bowel incontinence. The rectal insert is designed for self-insertion to seal and help prevent the involuntary leakage of stool from the rectum.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
K243367
Page 1 of 6
# 510(k) Summary
Minnesota Medical Technologies Fecal Incontinence Insert
## I. SUBMITTER
Minnesota Medical Technologies
2446 Henry Road NW, Stewartville, MN 55976
Contact person: Robert Anglin, VP, Quality & Regulatory
Phone: 507-533-0366
Fax: 507-533-0388
Date prepared: October 28, 2024
## II. DEVICE
Name of the device: Minnesota Medical Technologies Fecal Incontinence Insert
Common name: Rectal Insert
Regulation number: 21 CFR 876.5980
Regulation name: Gastrointestinal tube and accessories
Regulatory Class: II
Product Code: PBP
## III. PREDICATE DEVICE
Renew Medical Inc., Renew Insert, K122003
Reference device: Eclipse System DEN140020
## IV. DEVICE DESCRIPTION
The device is a simple rectal insert which is used to control accidental bowel leakage in adult patients. The single use device is inserted via the anus into the anal canal and rectum. The device engages with and conforms to the anorectal junction and the anal canal to control the accidental leakage of fecal matter.
The device consists of two components: a soft, pre-lubricated, liquid-filled silicone Insert device, and a plastic applicator. The Insert device is soft and flexible which allows engagement with the anorectal junction and the anal canal. The applicator is preassembled inside the Insert device to facilitate insertion of the device and is removed and discarded after the device is in place. The device is available in two sizes, Standard and Large. The device is provided preassembled and non-sterile in individual single-use packages.
## V. INDICATION FOR USE
{5}
K243367
Page 2 of 6
The Fecal Incontinence Insert is indicated for the management of Accidental Bowel Leakage (ABL) due to bowel incontinence. The rectal insert is designed for self-insertion to seal and help prevent the involuntary leakage of stool from the rectum.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
At a high level, the subject and predicate device are based on the following same technological elements:
| • Product Code
• Device Classification Name
• 21 CFR Regulation Number
• Indications for Use
• Principle of Operation
• Patient Population
• Anatomical Site
• Location Where Used
• Placement/Human Factors
• Single Use Only
• Provided Non-sterile
• Packaging
• Assembly
• Applicator | • Diameter
• Shaft Length
• Strength Test
• Detachment Test
• Puncture Test
• Buckling Test
• Bending Test
• Packaging Integrity
• Instructions For Use
• Labeling
• Warning, Precautions
• Shelf Life
• Applicator Material
• Biocompatibility Test Results |
| --- | --- |
The following technological differences exist between the subject and predicate device:
| • Anorectal Junction Sealing Feature
• Shaft Configuration
• External Retaining Feature at Anus
• Sizes (different in name only)
• External Retainer Diameter
• Simulated Use Testing Results | • Contraindications
• Disk or Bulb Material
• Shaft Material
• External Retainer Material
• Patient Lube |
| --- | --- |
## VII. NON-CLINICAL PERFORMANCE DATA
### Bench Testing
The following bench performance data was provided in support of the substantial equivalence.
{6}
- Strength Test
- Detachment Test
- Puncture Test
- Buckling Test
- Bending Test
- Sealing Function in Simulated Use
- Package Functional Testing
- Package Stability Testing
- Shelf Life Testing
## Biocompatibility Testing
The device was assessed in accordance with ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. The following tests were performed per GLP:
- ISO 10993-5, Cytotoxicity Using the ISO Elution Method
- ISO 10993-10, ISO Guinea Pig Maximization Sensitization Test
- ISO 10993-10, Rectal Irritation Study in Rabbits
- ISO 10993-11, Material Mediated Pyrogenicity
- ISO 10993-3, Bacterial Reverse Mutation Study
- ISO 10993-3, Mouse Lymphoma Assay
A biological risk assessment was performed to identify areas of concern to be addressed by literature review, clinical experience, and testing with consideration given to type of patient contact, potential hazards of the materials of construction, the history of clinical use, and testing of the materials of construction, biocompatibility and chemical characterization testing on the device, and other information available in the literature.
Based upon examination of the device materials and test results, use of the Minnesota Medical Technologies Fecal Incontinence Insert would not be expected to result in an adverse biological response when used as intended.
Based on the design verification performance, the Minnesota Medical Technologies Fecal Incontinence Insert was found to have a safety and effectiveness profile that is substantially equivalent to the predicate device.
## VIII. CLINICAL PERFORMANCE DATA
{7}
A prospective, open label, single-arm, non-randomized clinical study designed to establish the safety and effectiveness of the Minnesota Medical Technologies Fecal Incontinence Insert in subjects with fecal incontinence. The US clinical trial was performed with male and female patients, aged 18 and above, with a diagnosis of fecal incontinence and have also failed conservative medical therapy and biofeedback therapy as appropriate for fecal incontinence. The trial consisted of a 4-week baseline evaluation period, a 4-week fitting period, and a 4-week treatment period.
**Effectiveness of the insert** was evaluated by the primary effectiveness endpoint:
A relative percentage change in episodes of Accidental Bowel Leakage (ABL) determined by comparing treatment results to pretreatment results from the baseline period as measured by daily diary recordings.
The primary endpoint was satisfied with 44 of 58 subjects (75.9%, 95% CI: 62.8%, 86.1%) achieved a ≥ 50% reduction in ABL episodes demonstrating a highly significant device effect (p < .0001).
**Safety of the insert** was evaluated through safety endpoints which included adverse events, serious adverse events, unanticipated serious adverse events, device deficiencies, and device malfunctions.
The safety population included all 124 enrolled subjects (enrolled = signed informed consent prior to initial screening). There were 15 adverse events (AEs) reported in 12 subjects of which ten AEs were not related to the device, where one of these was serious. Three AEs were classified as possibly related to the device, one AE was classified as probably related to the device, and one AE was classified as definitely related to the device. The one AE classified as definitely related to the device was device over-insertion or device migration into the anal canal or rectum.
There were no device-related serious AEs, or unanticipated serious AEs. There were no device deficiencies or malfunctions.
After the treatment period, all individuals who withdrew from the study for reasons other than personal preference (i.e., for medical reasons) and a minimum of 50% of all subjects who completed the study, underwent an evaluation by anoscopy following the treatment period to assess for device related trauma. No patients necessitated an anoscopy due to new anorectal bleeding or moderate or severe pain noted in the bowel diary. Post-
{8}
treatment period anoscopy results do not show evidence of insert-related trauma.
Comparison of the insert to the predicate shows that the insert performs as well as or better than the legally marketed predicate device. The clinical trials for the proposed device and predicate device have similar design, compared a treatment period to a baseline period, and were performed on male and female adult patients with a diagnosis of fecal incontinence.
The clinical study results support that the Minnesota Medical Technologies Fecal Incontinence Insert is safe and effective when used for the intended purpose. The clinical performance (safety and effectiveness) of the insert support the determination of substantial equivalence of the insert to the predicate device.
## IX. SUBSTANTIAL EQUIVALENCE
The indication for use for the Minnesota Medical Technologies Fecal Incontinence Insert is substantially equivalent to the predicate. The technological characteristics of the Minnesota Medical Technologies Fecal Incontinence Insert are similar to the predicate device. Both the proposed device and the predicate device have a bar-bell design which incorporates both an internal and external retention feature which are connected by a central shaft. Both are inserted with a removable applicator. Both control accidental bowel leakage through engagement of the device with the anorectal junction.
The Minnesota Medical Technologies Fecal Incontinence Insert has the same intended use, patient population, and anatomical sites as well as similar technological characteristics as the predicate device. The differences in technological characteristics have been analyzed and addressed through testing and clinical evaluation. Any differences in the technological characteristics between the devices do not raise any new issues of safety or effectiveness.
The nonclinical and clinical testing demonstrate that the Minnesota Medical Technologies Fecal Incontinence Insert is as safe, as effective, and performs as well as or better than the predicate device. Thus, the Minnesota Medical Technologies Fecal Incontinence Insert is substantially equivalent to the predicate device.
{9}
X. CONCLUSION
Based on the same indications for use, technological characteristics, performance testing, and clinical evaluation, safety, and performance, the Minnesota Medical Technologies Fecal Incontinence Insert, is substantially equivalent to the Renew Medical, Inc. Renew Insert (K122003).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.