← Product Code [HRS](/productcode/HRS) · K243357

# CastleLoc Pectus Bar System (K243357)

_L&K BIOMED Co., Ltd. · HRS · Feb 27, 2025 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K243357

## Device Facts

- **Applicant:** L&K BIOMED Co., Ltd.
- **Product Code:** [HRS](/productcode/HRS.md)
- **Decision Date:** Feb 27, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The CastleLoc Pectus Bar System is indicated for the treatment of Pectus Excavatum and other anterior chest deformities. It is intended to be used in pediatric (children and adolescents) and adult populations.

## Device Story

Thoracic support system for surgical correction of pectus excavatum; utilizes minimally invasive techniques to elevate ribs and sternum. System comprises various sizes of straight and curved bars, stabilizers, nuts, and claw fixators. Implantation typically lasts 2-3 years. Surgeon selects appropriate bar size based on patient body size. Device provides structural support to thoracic wall to correct chest deformity; benefits patient through anatomical restoration. Used in surgical settings by surgeons.

## Clinical Evidence

Bench testing only. Mechanical performance verified via static and dynamic 4-point bending and vertical tests per ASTM F382-17. Results compared to Park's Pectus System (K191057) showed similar or superior mechanical performance, meeting all acceptance criteria.

## Technological Characteristics

Materials: Ti-6Al-4V (ASTM F136). Components: Straight/curved bars, stabilizers, nuts, claw fixators. Mechanical performance testing per ASTM F382-17. No software or electronic components.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Park's Pectus System ([K191057](/device/K191057.md))
- Pectus Blu Support Bar System ([K241709](/device/K241709.md))
- KLS Martin Pure Pectus System ([K221938](/device/K221938.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the Department of Health & Human Services seal. To the right of the seal is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 27, 2025

L&K BIOMED Co., Ltd. Kihyang Kim Vice Chairman #101, 201, 202 16-25, Dongbaekjungang-ro 16 beon-gil Giheung-gu Yongin-si, Gyeonggi-do 17015 South Korea

Re: K243357

Trade/Device Name: CastleLoc Pectus Bar System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS Dated: January 24, 2025 Received: January 27, 2025

Dear Kihyang Kim:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Thomas Mcnamara -S

For: Christopher Ferreira, MS Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

Submission Number (if known)

K243357

Device Name

CastleLoc Pectus Bar System

Indications for Use (Describe)

The CastleLoc Pectus Bar System is indicated for the treatment of Pectus Excavatum and other anterior chest deformities. It is intended to be used in pediatric (children and adolescents) and adult populations.

Type of Use (Select one or both, as applicable)

> | Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

| 510(k) #: K243357                  | 510(k) Summary                                                                                               | Prepared on: 2025-02-27 |
|------------------------------------|--------------------------------------------------------------------------------------------------------------|-------------------------|
| Contact Details                    |                                                                                                              | 21 CFR 807.92(a)(1)     |
| Applicant Name                     | L&K BIOMED Co., Ltd.                                                                                         |                         |
| Applicant Address                  | #101, 201, 202 16-25, Dongbaekjungang-ro 16 beon-gil Giheung-gu,<br>Yongin-si Gyeonggi-do 17015 Korea, South |                         |
| Applicant Contact Telephone        | 82-10-5477-0325                                                                                              |                         |
| Applicant Contact                  | Ms. Kihyang Kim                                                                                              |                         |
| Applicant Contact Email            | khkim@Inkbiomed.com                                                                                          |                         |
| Device Name                        |                                                                                                              | 21 CFR 807.92(a)(2)     |
| Device Trade Name                  | CastleLoc Pectus Bar System                                                                                  |                         |
| Common Name                        | Single/multiple component metallic bone fixation appliances and<br>accessories                               |                         |
| Classification Name                | Plate, Fixation, Bone                                                                                        |                         |
| Regulation Number                  | 888.3030                                                                                                     |                         |
| Product Code(s)                    | HRS                                                                                                          |                         |
| Legally Marketed Predicate Devices |                                                                                                              | 21 CFR 807.92(a)(3)     |
| Predicate #                        | Predicate Trade Name (Primary Predicate is listed first)                                                     | Product Code            |
| K191057                            | Park's Pectus System                                                                                         | HRS                     |
| K241709                            | Pectus Blu Support Bar System                                                                                | HRS                     |
| K221938                            | KLS Martin Pure Pectus System                                                                                | HRS                     |
| Device Description Summary         |                                                                                                              | 21 CFR 807.92(a)(4)     |

The CastleLoc Pectus Bar System is a thoracic support product that repairs the thoracic wall, using minimally invasive surgical techniques to elevate the ribs and sternum to correct a type of chest wall deformity called pectus excavatum. Recommended implantation time is 2~3 years but may vary based on surgeon preference and patient.

This system includes various sizes of straight and curved CastleLoc Pectus Bars. The apropriate bar is selected based on suitability for the patient body size. The CastleLoc Pectus Bar, and Castleloc Pectus Stabilizer, Nut and Castleloc Pectus Claw Fixator for fixing the Castleloc Pectus Bar are made of Ti-6Al-4V (ASTM F136).

# Intended Use/Indications for Use

The CastleLoc Pectus Bar System is indicated for the treatment of Pectus Excavatum and other anterior chest deformities. It is intended to be used in pediatric (children and adolescents) and adult populations.

# Indications for Use Comparison

21 CFR 807.92(a)(5)

{5}------------------------------------------------

The subject CastleLoc Pectus Bar System have been found to be substantially equivalent to the predicate devices with respect to technological characteristics, performance, and intended use. The information provided within this premarket notification supports substantial equivalence of the subject CastleLoc Pectus Bar System to the predicate devices.

# Technological Comparison

### 21 CFR 807.92(a)(6)

The CastleLoc Pectus Bar System has undergone a comprehensive battery of non-clinical performance testing, to include static and dynamic 4-point bending test and vertical test. Based on the test result the CastlLloc Pectus Bar System is suitable for its intended use and the test results also supports substantial equivalence of the CastleLoc Pectus Bar System to the primary predicate Park's Pectus System (K191057)

The design feature and indications for use of the subject devices are substantially equivalent to the predicate devices. Primary Predicate Device: Park's Pectus System (K191057) Additional Predicate Device: Pectus Blu Support Bar System (K241709) KLS Martin Pure Pectus System (K221938)

The subject and predicate devices are substantially equivalent in the areas of materials, design, indications for use, intended use and operational principles.

#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

These tests verified the mechanical performance of the CastleLoc Pectus Bar system according to ASTM F382-17 standards. Side-by-side comparison of the CastleLoc-Pectus Bar system test results for the Park's Pectus System (K191057) was conducted. The mechanical performance test results of the CastleLoc-Pectus Bar system were shown to be similar or higher than the predicate device, Park's Pectus System (KI 91057), and meets the specified acceptance criteria.

Therefore, this test report support that the CastLloc-Pectus Bar system has substantially equivalent mechanical performance as compared to the predicate device, and support that the subject device is at least as safe and effective as the predicate. It is concluded that the information provided in this 510(k) submission supports substantial equivalence

---

**Source:** [https://fda-staging.innolitics.com/device/K243357](https://fda-staging.innolitics.com/device/K243357)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
