27G x 1/2 TW K-Pack Surshield Needle (KN-S2713RBT)
K243309 · Terumo Europe N.V. · FMI · May 29, 2025 · General Hospital
Device Facts
Record ID
K243309
Device Name
27G x 1/2 TW K-Pack Surshield Needle (KN-S2713RBT)
Applicant
Terumo Europe N.V.
Product Code
FMI · General Hospital
Decision Date
May 29, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The 27G x 1/2" TW K-Pack Surshield Needle is a hypodermic needle with a passive sharps protection feature that covers the cannula immediately and permanently after use; and is intended for use in combination with a (pre-)filled syringe for subcutaneous and intramuscular injection.
Device Story
The 27G x 1/2" TW K-Pack Surshield Needle is a single-use, manual hypodermic needle featuring a passive sharps protection mechanism. The device consists of a stainless-steel cannula attached to a polypropylene hub, designed to interface with standard male luer syringe nozzles. Upon completion of an injection, the integrated safety feature automatically and permanently covers the needle cannula to mitigate accidental needle stick injuries. The device is supplied sterile, processed via ethylene oxide, and packaged in a hard plastic container. It is intended for use by healthcare professionals in clinical settings for subcutaneous or intramuscular medication delivery. The primary modification from the predicate is the use of a thin-wall (TW) cannula, which increases the inner diameter while maintaining the same outer diameter, facilitating improved flow characteristics for the injected substance.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including performance verification per ISO 7864, ISO 9626, ISO 80369-7, ISO 6009, and ISO 23908, as well as biocompatibility, sterility, and packaging validation.
Technological Characteristics
Materials: Stainless steel cannula, polypropylene hub/basecover/guide/case/cap, MABS sheath, POM ring, polycarbonate stopper, silicone lubricant, acrylic glue. Standards: ISO 7864 (needle), ISO 9626 (tubing), ISO 80369-7 (luer connectors), ISO 23908 (sharps protection), ISO 11607 (packaging). Sterilization: Ethylene oxide (SAL 10^-6). Form factor: 27G x 1/2" thin-wall needle with passive safety mechanism.
Indications for Use
Indicated for subcutaneous and intramuscular injection in patients requiring hypodermic needle delivery via (pre-)filled syringe.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 29, 2025
Terumo Europe N.V.
Liesbeth Decoster
Regulatory Affairs Manager
Interleuvenlaan 40
Leuven, 3001, Belgium
Re: K243309
Trade/Device Name: 27G x 1/2" TW K-Pack Surshield Needle (KN-S2713RBT)
Regulation Number: 21 CFR 880.5570
Regulation Name: Hypodermic Single Lumen Needle
Regulatory Class: Class II
Product Code: FMI
Dated: March 4, 2025
Received: May 2, 2025
Dear Liesbeth Decoster:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243309 - Liesbeth Decoster
Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K243309 - Liesbeth Decoster
Page 3
Sincerely,
Shruti N. Mistry -S
Shruti Mistry
Assistant Director
DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K243309 | |
| Device Name 27G x 1/2" TW K-Pack Surshield Needle (KN-S2713RBT) | |
| Indications for Use (Describe) The 27G x 1/2" TW K-Pack Surshield Needle is a hypodermic needle with a passive sharps protection feature that covers the cannula immediately and permanently after use; and is intended for use in combination with a (pre-)filled syringe for subcutaneous and intramuscular injection. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
TERUMO
510(k) SUMMARY
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92
1. Submitter Information (807.92(a)(1))
Prepared for:
TERUMO EUROPE N.V.
Interleuvenlaan 40,
3001 Leuven,
BELGIUM
Prepared by/Contact person:
Mrs. L. Decoster – Regulatory Affairs Manager
Tel. (+32) 16 38 13 02
Fax (+32) 16 40 02 49
Date prepared:
September 2024
2. Device Name (807.92(a)(2))
Proprietary Name: 27G x ½" TW K-Pack Surshield Needle
Common Name: Hypodermic Needle
Classification Name: Hypodermic Single Lumen Needle
Classification Panel: General Hospital
Regulation: 21CFR, Section §880.5570
Product Code: FMI
Classification: Class II
3. Predicate Devices (807.92(a)(3))
The legally marketed device(s) to which substantial equivalence is claimed:
- 27G x ½" K-Pack Surshield Needle (K111797) manufactured by Terumo Europe N.V.
4. Reason for 510(k) Submission
This premarket notification [510(k)] is being submitted for the 27G x ½" TW K-Pack Surshield Needle to provide supporting information that the proposed device is safe and effective and substantially equivalent to the following devices:
- 27G x ½" K-Pack Surshield Needle (K111797) manufactured by Terumo Europe N.V.
The change proposed in this submission is the 27G x ½" TW K-Pack Surshield Needle which has an increased cannula's inner diameter and therefore different cannula wall thickness (thin wall vs regular wall) compared to the 27G x ½" K-Pack Surshield Needle (K111797).
{5}
TERUMO
5. Device Description (807.92(a)(4))
## Principle of Operation Technology
The 27G x ½" TW K-Pack Surshield Needle is operated manually or by manual process.
## Design/Construction
The 27G x ½" TW K-Pack Surshield Needle is a hypodermic single lumen needle, for single use consisting of a stainless-steel tube that is sharpened at one end and at the other end joined to a female luer connector (hub) made of polypropylene designed to be connected with a male luer connector (nozzle) of a hypodermic syringe. Furthermore, the needle has a passive sharps protection feature that covers the cannula immediately and permanently after use to prevent the risk of accidental needle stick injuries. The needle is non-toxic, non-pyrogenic and sterilized by ethylene oxide. Its operation is manual. The 27G x ½" TW K-Pack Surshield Needle is packed in a hard plastic container (cap and case) made of polypropylene and sealed with a label and sterilized by ethylene oxide.
## Specifications
The following table shows the product codes, needle gauge and needle length.
Table 1 - Product Specifications
| Product Code | Needle Gauge | Needle Length | Needle bevel | Cannula wall |
| --- | --- | --- | --- | --- |
| KN-S2713RBT | 27G (0.4 mm) | ½" (12 mm) | Regular bevel | Thin Wall (TW) |
6. Indications for Use (807.92(a)(5))
The 27G x ½" TW K-Pack Surshield Needle is a hypodermic needle with a passive sharps protection feature that covers the cannula immediately and permanently after use; and is intended for use in combination with a (pre-)filled syringe for subcutaneous and intramuscular injection.
Note: Indications for use of the subject device are the same as for the following predicate device:
- 27G x ½" K-Pack Surshield Needle (K111797) manufactured by Terumo Europe N.V.
7. Substantial Equivalence Comparison (807.92(a)(6))
The 27G x ½" TW K-Pack Surshield Needle, the subject of this 510(k), is substantially equivalent to the following predicate devices:
- 27G x ½" K-Pack Surshield Needle (K111797) manufactured by Terumo Europe N.V.
The similarities and differences are summarized below.
{6}
TERUMO
Table 2 - Intended Use/Indications for Use
| Characteristics | Subject Device: 27G x ½” TW K-Pack Surshield Needle | Predicate device: 27G x ½” K-Pack Surshield Needle (K111797) | Comments |
| --- | --- | --- | --- |
| Indications for Use | The 27G x ½” TW K-Pack Surshield Needle is a hypodermic needle with a passive sharps protection feature that covers the cannula immediately and permanently after use; and is intended for use in combination with a (pre-)filled syringe for subcutaneous and intramuscular injection. | The 27G x ½” K-Pack Surshield Needle is a hypodermic needle with a passive sharps protection feature that covers the cannula immediately and permanently after use; and is intended for use in combination with a (pre-)filled syringe for subcutaneous and intramuscular injection. | Same indications for use |
| Intended Use | The K-Pack Surshield Needle is for general application – for treatment (subcutaneous or intramuscular injection). | The K-Pack Surshield Needle is for general application – for treatment (subcutaneous or intramuscular injection). | Same intended use |
| Prescription or OTC (over the counter) | Prescription | Prescription | Same |
{7}
TERUMO
Table 3 - Technological characteristics
| Characteristics | Subject Device: 27G x ½" TW K-Pack Surshield Needle | Predicate device: 27G x ½" K-Pack Surshield Needle (111797) | Comments |
| --- | --- | --- | --- |
| Manufacturer | Terumo Europe N.V. | Terumo Europe N.V. | Same manufacturer |
| Materials | Cannula – Stainless Steel
Hub – Polypropylene/ Colour Masterbatch
Basecover – Polypropylene/ Colour Masterbatch
Sheath – MehtylMethacrylate Acrylonitrile Butadiene Styrene (MABS)
Ring – Polyoxymethylene (POM)/ Colour Masterbatch/ Slipping Agent
Stopper – Polycarbonate
Guide – Polypropylene
Spring – Stainless Steel
Glue – Acrylic Glue (UV cured)
Lubricant – Silicone oil
Case – Polypropylene
Cap – Polypropylene
Label – Self-Adhesive Paper Label | Cannula – Stainless Steel
Hub – Polypropylene/ Colour Masterbatch
Basecover – Polypropylene/ Colour Masterbatch
Sheath – MehtylMethacrylate Acrylonitrile Butadiene Styrene (MABS)
Ring – Polyoxymethylene (POM)/ Colour Masterbatch/ Slipping Agent
Stopper – Polycarbonate
Guide – Polypropylene
Spring – Stainless Steel
Glue – Acrylic Glue (UV cured)
Lubricant – Silicone oil
Case – Polypropylene
Cap - Polypropylene
Label – Self-Adhesive Paper Label | Same materials |
{8}
TERUMO
| Characteristics | Subject Device: 27G x ½" TW K-Pack Surshield Needle | Predicate device: 27G x ½" K-Pack Surshield Needle (111797) | Comments |
| --- | --- | --- | --- |
| Design/ Constructions | The 27G x ½" TW K-Pack Surshield Needle is a Hypodermic single lumen needle, for single use consisting of a stainless-steel tube that is sharpened at one end and at the other end joined to a female luer connector (hub) made of polypropylene designed to be connected with a male luer connector (nozzle) of a hypodermic syringe. Furthermore, the needle has a passive sharps protection feature that covers the cannula immediately and permanently after use. The 27G x ½" TW K-Pack Surshield Needle is packed in a hard plastic container (cap and case) made of polypropylene and sealed with a label. | The 27G x ½" K-Pack Surshield Needle is a Hypodermic single lumen needle, for single use consisting of a stainless-steel tube that is sharpened at one end and at the other end joined to a female luer connector (hub) made of polypropylene designed to be connected with a male luer connector (nozzle) of a hypodermic syringe. Furthermore, the needle has a passive sharps protection feature that covers the cannula immediately and permanently after use. The 27G x ½" K-Pack Surshield Needle is packed in a hard plastic container (cap and case) made of polypropylene and sealed with a label. | Same design/construction
Both needles are designed in accordance with ISO 7864 and ISO 80369-7. |
| Specifications | 27G x ½" (0.4 x 12 mm)
Thin Wall
Regular Bevel
Outside diameter
Min. 0.40 mm – Max. 0.42 mm
Inner diameter
Min. 0.241 mm – Max. 0.281 mm | 27G x ½" (0.4 x 12 mm)
Regular Wall
Regular Bevel
Outside diameter
Min. 0.40 mm – Max. 0.42 mm
Inner diameter
Min. 0.20 mm – Max. 0.24 mm | Increased cannula’s inner diameter and therefore different cannula wall thickness (thin wall vs regular wall)
Cannula specifications are in accordance with ISO 9626 |
| Principle of Operation | Manual | Manual | Same operation principle |
{9}
TERUMO
| Unit packaging | Hard plastic container consisting of cap and case | Hard plastic container consisting of cap and case | Same packaging system validated in accordance with ISO 11607-1 and ISO 11607-2 |
| --- | --- | --- | --- |
| Characteristics | Subject Device: 27G x ½” TW K-Pack Surshield Needle | Predicate device: 27G x ½” K-Pack Surshield Needle (111797) | Comments |
| --- | --- | --- | --- |
| Sterilization | EO to SAL 10⁻⁶ | EO to SAL 10⁻⁶ | Same sterilization validation process in accordance with ISO 11135
Bacterial endotoxin limits and EO residual limits (ISO 10993-7) are the same |
| Shelf life | 5 years | 5 years | Same shelf life |
{10}
TERUMO
8. Substantial Equivalence Comparison (807.92(a)(6))
There are no differences in the Intended Use and Indications for Use and the differences in the technological characteristics of the subject device to the predicate device are supported by the below performance verification:
- Performance verification is evaluated according to the following standards: ISO 7864 and ISO 9626. The same standards were utilized for the predicate device 27G x ½" K-Pack II Needle (K111797) to demonstrate performance.
9. Non Clinical Test (807.92(b)(1))
**Performance**
The design of the 27G x ½" TW K-Pack Surshield Needle is in accordance with the Design Control Requirements and recognized consensus standards that have been established for hypodermic needles under FDA product code FMI and 21CFR Section 880.5570:
ISO 7864:2016 "Sterile hypodermic needle for Single use"
ISO 9626:2016 "Stainless steel needle tubing for the manufacturing of medical devices"
ISO 80369-7:2021 "Small bore connectors for liquids & gases in healthcare applications – Part 7: Connectors for intravascular or hypodermic applications"
ISO 6009:2016 "Stainless steel needle tubing for the manufacturing of medical devices"
ISO 23908:2011 "Sharps injury protection – Requirements and test methods – Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling"
ISO 11607-1:2019 "Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems"
USP <788> Particulate matter in injections
USP <71> Sterility test
**Biocompatibility**
The 27G x ½" TW K-Pack Surshield Needles are categorized following the definitions in EN ISO 10993-1:2020 as external communicating devices that can contact tissue, or that can indirectly contact the blood path up to 24 hours (limited exposure).
Based on this classification and considering FDA Guidance document: Use of International Standard ISO-10993-1, "Biological Evaluation of Medical Devices Part-1: Evaluation and testing within a risk management process" the applicable biological endpoints are: cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, material-mediated pyrogenicity, bacterial endotoxins and haemocompatibility (material-induced haemolysis).
{11}
TERUMO
# Sterilization and shelf life
The sterility of the 27G x ½" TW K-Pack Surshield Needle is assured by using a validated sterilization method qualified in accordance with EN ISO 11135:2014 "Sterilization of Health Care Products – Ethylene oxide – Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices".
The physical validation of the sterilizer is conducted to verify the temperature and humidity in the sterilization load and the pressure in the sterilizer during the whole cycle.
The biological validation is performed in accordance with EN ISO 11135:2014 Annex B "Conservative determination of lethal rate of the sterilization process – Overkill approach" part B.1.2.b "Cycle calculations". This resulted in a holding time of 120 min for the sterilization cycle to assure a SAL of at least 10⁻⁶ according to the requirements of EN ISO 11135:2014.
The products can therefore be labelled sterile in accordance with EN 556-1:2001/AC1:2006 "Sterilization of medical devices - Requirements for medical devices to be labelled STERILE - Part 1: Requirements for terminally sterilized medical devices".
The levels of residual EO and ECH are defined in accordance with ISO 10993-7 'Biological evaluation of medical devices – Part 7: Ethylene Oxide sterilization residuals'.
The limits for the bacterial endotoxin testing LAL (Limulus Amebocyte Lysate) performed on each lot as part of the release criteria are aligned with the requirements described in USP <85> and <161>.
The shelf life of the 27G x ½" TW K-Pack Surshield Needle is 5 years.
10. Clinical Test (807.92(b)(2))
This 510(k) does not include data from clinical tests.
11. Conclusion (807.92(b)(3))
In summary, the 27G x ½" TW K-Pack Surshield Needle, manufactured by Terumo Europe N.V., being the subject of this 510(k), is substantially equivalent to its predicate device:
- 27G x ½" K-Pack Surshield Needle (K111797) manufactured by Terumo Europe N.V.
There are no differences in the Intended Use and Indications for Use and the differences in the technological characteristics, do not raise any new or different questions of safety or effectiveness when compared to the predicate device. The 27G x ½" TW K-Pack Surshield Needle is as safe and effective, and performs as well as the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.