Masimo W1™ and the integrated Masimo W1 Module are intended to record, store and transfer single-channel electrocardiogram (ECG) rhythms. The Masimo W1 also displays ECG rhythms, and the Masimo W1 Module detects the presence of atrial fibrillation. The Masimo W1 and the integrated Masimo W1 Module are intended for use by healthcare professionals, patients with known or suspected heart conditions, and health-conscious individuals. Masimo W1 Module ECG software is an over-the-counter (OTC) software that operates on the Masimo W1 Module that can be used with compatible watches (e.g., Masimo W1). The software is intended to create, record, store, transfer, and display a single channel electrocardiogram (ECG) for informational use only in adults 22 years and older. It supports the classification of either atrial fibrillation (AFib) or sinus rhythm with the intention of aiding heart rhythm identification; it is not intended to replace traditional methods of diagnosis or treatment. The software is not intended for users with other known arrhythmias and users should not interpret or take clinical action based on the device output without consultation of a qualified healthcare professional. The Masimo W1™ and the integrated Masimo W1 Module are also intended for the spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). The Masimo W1 and the integrated Masimo W1 Module are indicated for adults in hospitals, clinics, long-term care facilities, and homes.
Device Story
Masimo W1 is a wrist-worn watch integrating Masimo W1 Module for physiological signal detection. Input: single-channel ECG via wrist-to-finger contact; optical PPG signals via wrist-based LEDs. Operation: module processes electrical signals for heart rate and rhythm classification (AFib vs. Sinus Rhythm) and optical signals for SpO2/pulse rate using Masimo SET technology. Output: ECG waveforms, rhythm classification, SpO2, and pulse rate displayed on watch screen; data transmitted via Bluetooth to smartphone app for user/caregiver tracking. Used in clinical and home settings by patients or healthcare professionals. Output aids heart rhythm identification; clinical decision-making requires professional consultation. Benefits: enables convenient, non-invasive monitoring of heart rhythm and oxygen saturation.
Clinical Evidence
Prospective clinical validation on adult subjects across 4 sites. AFib classification performance: 99.3% sensitivity [96.3%, 100%], 100% specificity [97.8%, 100%], 100% PPV [97.5%, 100%], 5.0% unclassified rate, 1.7% noise rate. Qualitative waveform assessment showed 98% agreement with gold-standard 12-lead ECG (Lead I). Quantitative analysis confirmed QRS amplitude/width similarity to reference.
Technological Characteristics
Wrist-worn watch with integrated module. Sensing: electrical (ECG) and optical (PPG/SpO2). Connectivity: Bluetooth to smartphone app. Power: internal rechargeable Li-Ion battery. Safety: IEC 60601-1, IEC 60601-1-2, CF applied part, IP24 ingress protection. Software: AFib classification algorithm based on 'irregularly irregular' rhythm detection.
Indications for Use
Indicated for adults 22+ years in hospitals, clinics, long-term care, and homes. Used by healthcare professionals, patients with known/suspected heart conditions, and health-conscious individuals for recording/displaying single-channel ECG and spot-checking SpO2/pulse rate. Supports classification of AFib or sinus rhythm. Not for users with other known arrhythmias; not for diagnostic/treatment replacement.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 3, 2025
Masimo Corporation
Sindura Penubarthi
Associate Director, Regulatory Affairs
52 Discovery
Irvine, California 92618
Re: K243305
Trade/Device Name: Masimo W1
Regulation Number: 21 CFR 870.2340
Regulation Name: Electrocardiograph
Regulatory Class: Class II
Product Code: DPS, DQA, DXH, QDA
Dated: February 25, 2025
Received: February 25, 2025
Dear Sindura Penubarthi:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243305 - Sindura Penubarthi
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K243305 - Sindura Penubarthi
Page 3
Sincerely,
Jennifer W. Shih -S
Jennifer Kozen
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243305
Device Name
Masimo W1
Indications for Use (Describe)
Masimo W1™ and the integrated Masimo W1 Module are intended to record, store and transfer single-channel electrocardiogram (ECG) rhythms. The Masimo W1 also displays ECG rhythms, and the Masimo W1 Module detects the presence of atrial fibrillation. The Masimo W1 and the integrated Masimo W1 Module are intended for use by healthcare professionals, patients with known or suspected heart conditions, and health-conscious individuals.
Masimo W1 Module ECG software is an over-the-counter (OTC) software that operates on the Masimo W1 Module that can be used with compatible watches (e.g., Masimo W1). The software is intended to create, record, store, transfer, and display a single channel electrocardiogram (ECG) for informational use only in adults 22 years and older. It supports the classification of either atrial fibrillation (AFib) or sinus rhythm with the intention of aiding heart rhythm identification; it is not intended to replace traditional methods of diagnosis or treatment. The software is not intended for users with other known arrhythmias and users should not interpret or take clinical action based on the device output without consultation of a qualified healthcare professional.
The Masimo W1™ and the integrated Masimo W1 Module are also intended for the spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). The Masimo W1 and the integrated Masimo W1 Module are indicated for adults in hospitals, clinics, long-term care facilities, and homes.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
M
MASIMO CORPORATION
52 Discovery
Irvine, CA 92618
# 510(k) Summary
| Submitter and Address of Manufacturing Facility: | Masimo Corporation 52 Discovery Irvine, CA 92618 Phone: (949) 297-7000 FAX: (949) 297-7592 |
| --- | --- |
| Date: | April 2nd 2025 |
| Contact: | Sindura Penubarthi Associate Director, Regulatory Affairs Masimo Corporation Phone: (949) 396-4041 |
| Trade Name: | Masimo W1 |
| Common Name: | Electrocardiograph |
| Classification Regulation/ Product Code: | 21 CFR 870.2340, Class II/DPS |
| Additional Product Code: | DQA, DXH, QDA |
| Establishment Registration Number: | 3011353843 |
| Reason for Premarket Notification: | Update to add Atrial Fibrillation (AFib) Classification Feature |
| Primary Predicate: | K240229 – Masimo W1 |
| Secondary Predicate: | K201456 - Withing Scan Monitor |
| Tertiary Predicate | K201168 – Samsung ECG Monitor |
# 1 Device Description
The Masimo W1 is a watch that incorporates the Masimo W1 Module, which is the device that is responsible for the physiological signal detection and algorithm used to support the different parameters. The Masimo W1 Module incorporates spot check ECG functionality and Masimo SET Pulse Oximetry technology so that it can provide both ECG and Masimo SET pulse oximetry parameters. As part of this submission, Masimo is requesting clearance for an automated atrial fibrillation "AFib" Classification Feature that is used to analyze the single channel ECG waveform.
The parameter outputs from the Masimo W1 Module are communicated and displayed on the watch screen so that the data can be viewed and recorded. Masimo W1 also supports wireless communication of monitored data to a compatible smart device application. The sharing of the parameter data to the applications allows the users to see and track their data using their smart phones. Smart phone applications can also help to share information to caregivers and healthcare professionals.
The Masimo W1 specifications are provided in Table 1-1 below:
| Table 1-1 - Specifications | |
| --- | --- |
| Feature | Specifications |
| Continuous Display of Parameter Data | Yes |
| Performance Specifications | |
| SpO2, No Motion/ Low Perfusion (70-100%) | 2% Arms adults |
K243305
{5}
MASIMO CORPORATION
52 Discovery
Irvine, CA 92618
# 510(k) Summary
| Table 1-1 - Specifications | |
| --- | --- |
| Feature | Specifications |
| Pulse Rate (25-240 bpm) | 3 bpm Arms adults |
| Heart Rate (25-240 bpm) | ≤5 bpm adults |
| ECG Rhythm Classifications | Atrial Fibrillation, Sinus Rhythm |
| Electrical Specifications | |
| Battery | Internal Rechargeable Li-Ion |
| Mechanical Specifications | |
| Size | 40 mm (1.57”) |
| Display Type | Touchscreen |
| Weight | 54 g (including watchband) |
| Environmental Specifications | |
| Operating Temperature | 0 to 35 °C (32 to 95°F) |
| Operating Humidity | 10% to 95% RH (non-condensing) |
| Storage/Transport Temperature | -20°C to 60°C (-4°F to 140°F) |
| Storage/Transport Humidity | 10% to 95% RH (non-condensing) |
| Classification per IEC 60601-1 | |
| Electrical Safety | IEC 60601-1 |
| EMC | IEC 60601-1-2 |
| Electrical Isolation Type | Internally Powered |
| Applied Part Type | CF Applied Part |
| Ingress Protection | IP24 |
| Mode of Operation | Continuous |
# 2 Indications for Use
Masimo W1™ and the integrated Masimo W1 Module are intended to record, store and transfer single-channel electrocardiogram (ECG) rhythms. The Masimo W1 also displays ECG rhythms, and the Masimo W1 Module detects the presence of atrial fibrillation. The Masimo W1 and the integrated Masimo W1 Module are intended for use by healthcare professionals, patients with known or suspected heart conditions, and health-conscious individuals.
Masimo W1 Module ECG software is an over-the-counter (OTC) software that operates on the Masimo W1 Module that can be used with compatible watches (e.g., Masimo W1). The software is intended to create, record, store, transfer, and display a single channel electrocardiogram (ECG) for informational use only in adults 22 years and older. It supports the classification of either atrial fibrillation (AFib) or sinus rhythm with the intention of aiding heart rhythm identification; it is not intended to replace traditional methods of diagnosis or treatment. The software is not intended for users with other known arrhythmias and users should not interpret or take clinical action based on the device output without consultation of a qualified healthcare professional.
The Masimo W1™ and the integrated Masimo W1 Module are also intended for the spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). The Masimo W1 and the integrated Masimo W1 Module are indicated for adults in hospitals, clinics, long-term care facilities, and homes.
K243305
Page 2 of 13
{6}
MASIMO CORPORATION
52 Discovery
Irvine, CA 92618
# 510(k) Summary
## 3 Technological Characteristics
### Principle of Operation
There were no changes made to the principles of operation from the previously cleared features of the Masimo W1 under K240229. However, as part of this submission, an AFib Classification Feature is added that expands upon the principles of operation for the ECG feature.
### Electrocardiogram (ECG)
Same as under K240229, the ECG feature on Masimo W1 relies on the principle that the electrical signals can be detected as the heart contracts and relaxes during each cardiac cycle allowing for the estimation of heart rate (HR). The depolarization and repolarization of the heart muscle creates electrical signals that are propagated for recording on the skin surface at the wrist.
The Masimo W1 Module ECG software can also analyze these electrical signals for signs of AFib. To detect AFib, the software relies on the understanding that AFib rhythms are defined as “irregularly irregular” heartbeat patterns.
### Pulse oximetry-based parameters
The Masimo W1 still relies on Masimo SET pulse oximetry, which relies on the Beer-Lambert law and the following principles of pulse oximetry to provide estimates of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate:
- Oxyhemoglobin (oxygenated blood) and deoxyhemoglobin (non-oxygenated blood) differ in their absorption of red and infrared light (spectrophotometry).
- The amount of arterial blood in tissue changes with your pulse (photoplethysmography). Therefore, the amount of light absorbed by the varying quantities of arterial blood changes as well.
### Mechanism of Action for Achieving the Intended Effect
There were no changes to the mechanism of action for the Masimo W1 from the previous clearance under K240229.
The Masimo W1 still achieves its intended effect through the integrated Masimo W1 Module that provides both the electrical and optical sensing technology and parameter algorithms. The Masimo W1 Module is integrated into the Masimo W1 watch so that the sensing components contact the skin on the wrist. The watch is provided with an electrical sensing pad that is contacted by fingers on the opposing hand. Contacting the electrical sensing pad allows for recording of the ECG signal when the ECG feature is activated on the watch touchscreen. Once the ECG feature is activated, it detects and calculates the heart rate and determines the rhythm classification (e.g., Sinus Rhythm, Atrial Fibrillation).
K243305
Page 3 of 13
{7}
MAsIMO
MASIMO CORPORATION
52 Discovery
Irvine, CA 92618
# 510(k) Summary
The Masimo SET pulse oximetry-based parameters are supported by the Masimo W1 Module’s optical sensing components located on the bottom of the module that contacts the skin on the wrist. The Masimo W1 Module continuously detects and processes the optical signals that change with the transmission of LED light into the wrist tissue. The Masimo W1 Module utilizes multiple wavelengths of light and advanced signal processing techniques to isolate the arterial signal from other static factors (e.g., skin pigment) to establish the ratio used in the estimation of the SpO2. The pulse rate is determined by the periodic changes in the photoplethysmograph (PPG).
The monitored parameters are then continuously updated and displayed on the watch so that it can be viewed, recorded and/or transferred. Masimo W1 is also provided with Bluetooth connectivity to support the wireless communication of the monitored data to a smart device application for supplemental display. The use of the Masimo W1 watch can be discontinued by taking off the watch or deactivating the continuous parameters from the watch’s touchscreen.
# 4 Discussion of Similarities and Differences Between Primary Predicate and Subject Device
Similarities and Differences between Predicate and Subject Device
The subject device, Masimo W1 with AFib Classification Feature, and the predicate device, Masimo W1 (K240229), have the following key similarities:
- Both devices have similar intended use.
- Both devices rely on the same principles of operation.
The subject device, Masimo W1 with AFib Classification Feature, and the predicate device, Masimo W1 (K240229), have the following key difference:
- The subject device is provided with an AFib Classification Feature.
The key difference between the subject device and predicate device is the addition of the AFib Classification Feature. The intended use of this feature is substantially equivalent to the predicate devices. To support there are no technological characteristic differences that raise different questions of safety and effectiveness, clinical performance data of the AFib classification feature is provided as part of this submission. The test results support the substantial equivalence of the subject device.
K243305
Page 4 of 13
{8}
MASIMO CORPORATION
52 Discovery
Irvine, CA 92618
# 510(k) Summary
| Feature 510(k) Number | Masimo W1 Subject Device | Masimo W1 Primary Predicate (K240229) | Withing Scan Monitor Secondary Predicate (K201456) | Samsung ECG Monitor Tertiary Predicate (K201168) | Comparison to Primary Predicate |
| --- | --- | --- | --- | --- | --- |
| General Information | | | | | |
| Classification | Class II, Electrocardiograph | Class II, Electrocardiograph | Class II, Electrocardiograph | Class II, Electrocardiograph software for Over-The-Counter Use | Same. |
| Regulation/Product Code | 21 CFR 870.2340, Class II/ DPS | 21 CFR 870.2340, Class II/ DPS | 21 CFR 870.2340, Class II/ DPS | 21 CFR 870.2345, Class II/ QDA | Same. |
| Additional Product Code(s) | DQA
DXH
QDA | DQA
DXH | DQA
DXH | QDA | The subject device includes an additional product code QDA to support atrial fibrillation classification feature.
The addition of the product code is the same as the tertiary predicate device that has the same intended use |
| Intended Use | Masimo W1™ and the integrated Masimo W1 module are intended to record, store and transfer single-channel electrocardiogram (ECG) rhythms | The Masimo W1 and the integrated Masimo W1 module records, stores, transfers, and displays the single-channel electrocardiogram (ECG). | The Scan Monitor is intended to record, store and transfer single-channel electrocardiogram (ECG) rhythms. | The app is intended to create, record, store, transfer, and display a single channel electrocardiogram (ECG). | Same. |
| Indications for Use | Masimo W1 also displays ECG rhythms, and the Masimo W1 Module detects the presence of atrial fibrillation. | The Masimo W1 and the integrated Masimo W1 module records, stores, transfers, and displays the single-channel | The Scan Monitor also displays ECG rhythms and detects the presence of atrial fibrillation (when the monitor | The Samsung ECG Monitor Application is an over-the-counter (OTC) software-only, mobile medical | Different. The subject device is also indicated for an atrial fibrillation classification feature.
The indication for the AFib feature of the subject device is similar to the secondary |
K243305
Page 5 of 13
{9}
MASIMO CORPORATION
52 Discovery
Irvine, CA 92618
# 510(k) Summary
| Feature 510(k) Number | Masimo W1 Subject Device | Masimo W1 Primary Predicate (K240229) | Withing Scan Monitor Secondary Predicate (K201456) | Samsung ECG Monitor Tertiary Predicate (K201168) | Comparison to Primary Predicate |
| --- | --- | --- | --- | --- | --- |
| | The Masimo W1 and the integrated Masimo W1 Module are intended for use by healthcare professionals, patients with known or suspected heart conditions, and health-conscious individuals.
Masimo W1 Module ECG software is an over-the-counter (OTC) software that operates on the Masimo W1 Module that can be used with compatible watches (e.g., Masimo W1). The software is intended to create, record, store, transfer, and display a single channel electrocardiogram (ECG) for informational use only in adults 22 years and older. It supports the classification of either atrial fibrillation (AFib) or sinus rhythm with the intention of aiding heart rhythm identification; it is not intended to replace traditional methods of | ECG for the manual interpretation of heart rate. It is worn on the wrist and also provides other continuous parameters technologies (e.g., pulse oximetry). The Masimo W1™ and the integrated Masimo W1 module are also intended for the spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). The Masimo W1 and the integrated Masimo W1 Module are indicated for adults in hospitals, clinics, long-term care facilities, and homes. | is prescribed or used under the care of a physician). The Scan Monitor is intended for use by healthcare professionals, patients with known or suspected heart conditions and health-conscious individuals. | application operating on a compatible Samsung Galaxy Watch and iPhone. The app is intended to create, record, store, transfer, and display a single channel electrocardiogram (ECG), similar to a Lead I ECG, for informational use only in adults 22 years and older. Classifiable traces are labeled by the app as either atrial fibrillation (AFib) or sinus rhythm with the intention of aiding heart rhythm identification; it is not intended to replace traditional methods of diagnosis or treatment. The app is not intended for users with other | and tertiary predicate devices under the same intended use. |
K243305
Page 6 of 13
{10}
M
MASIMO CORPORATION
52 Discovery
Irvine, CA 92618
510(k) Summary
| Feature 510(k) Number | Masimo W1 Subject Device | Masimo W1 Primary Predicate (K240229) | Withing Scan Monitor Secondary Predicate (K201456) | Samsung ECG Monitor Tertiary Predicate (K201168) | Comparison to Primary Predicate |
| --- | --- | --- | --- | --- | --- |
| | diagnosis or treatment. The software is not intended for users with other known arrhythmias and users should not interpret or take clinical action based on the device output without consultation of a qualified healthcare professional.
The Masimo W1™ and the integrated Masimo W1 module are also intended for the spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR).
The Masimo W1 and the integrated Masimo W1 Module are indicated for adults in hospitals, clinics, long-term care facilities, and homes. | | | known arrhythmias and users should not interpret or take clinical action based on the device output without consultation of a qualified healthcare professional. | |
| Technological Characteristics | | | | | |
| Principles of Operation | The ECG feature on Masimo W1 relies on the principle that the electrical signals can be detected as the heart contracts and | The ECG feature on Masimo W1 still relies on the principle that the electrical signals can be | The iECG software analyzes the patterns in the ECG waveform to determine different | Unavailable. | Same. |
| | electrical signals. The ECG feature on Masimo W1 is not a reliable indicator of the electrical signals in the ECG. | electrical signals. The ECG feature on Masimo W1 is not a reliable indicator of the electrical signals in the ECG. | electrical signals. The ECG feature on Masimo W1 is not a reliable indicator of the electrical signals in the ECG. | | |
| Specifications | The following are the characteristics of the system: The following are the characteristics of the system: The output is recorded and recorded in the ECG, and the output is recorded in the ECG. The output is recorded in the ECG, and the output is recorded in the ECG. The output is recorded in the ECG. The output is recorded in the ECG. The output is recorded in the ECG. The output is recorded in the ECG. The output is recorded in the ECG. The output is recorded in the ECG. The ECG is recorded in the ECG. The ECG is recorded in the ECG. The ECG is recorded in the ECG. The ECG is recorded in the ECG. The ECG is recorded in the ECG. The ECG is recorded in the ECG. The ECG is recorded in the ECG. The ECG is recorded in the ECG. The ECG is recorded in the ECG. The ECG is recorded in the ECG. The ECG is | | | | |
K243305
{11}
MASIMO CORPORATION
52 Discovery
Irvine, CA 92618
# 510(k) Summary
| Feature 510(k) Number | Masimo W1 Subject Device | Masimo W1 Primary Predicate (K240229) | Withing Scan Monitor Secondary Predicate (K201456) | Samsung ECG Monitor Tertiary Predicate (K201168) | Comparison to Primary Predicate |
| --- | --- | --- | --- | --- | --- |
| | relaxes during each cardiac cycle allowing for the estimation of heart rate (HR). The depolarization and repolarization of the heart muscle creates electrical signals that are propagated for recording on the skin surface at the wrist.
The Masimo W1 Module ECG software can also analyze these electrical signals for signs of AFib. To detect AFib, the software relies on the understanding that AFib rhythms are defined as “irregularly irregular” heartbeat patterns.
The Masimo W1 still relies on Masimo SET pulse oximetry, which relies on the Beer-Lambert law and the following principles of pulse oximetry to provide estimates of functional oxygen saturation of | detected as different parts of the heart contract and relax during a cardiac cycle allowing the detection of heart activity and for the estimation of heart rate (HR). The change in polarization of the heart muscles creates the electrical signals that are propagated so that they can be detected at the skin surface of the wrist.
The Masimo SET pulse oximeter technology relies on the Beer-Lambert law and the following principles of pulse oximetry:
• Oxyhemoglobin (oxygenated blood) and | types of known heart activity (e.g., Heart Rate, Atrial Fibrillation).
Scan Monitor pulse oximetry technology relies on the differential absorption by blood of red (660nm) and infrared light (940nm) which relies on the Beer-Lambert law for the principles of pulse oximetry. | | |
K243305
Page 8 of 13
{12}
MASIMO CORPORATION
52 Discovery
Irvine, CA 92618
# 510(k) Summary
| Feature 510(k) Number | Masimo W1 Subject Device | Masimo W1 Primary Predicate (K240229) | Withing Scan Monitor Secondary Predicate (K201456) | Samsung ECG Monitor Tertiary Predicate (K201168) | Comparison to Primary Predicate |
| --- | --- | --- | --- | --- | --- |
| | arterial hemoglobin (SpO2) and pulse rate:
• Oxyhemoglobin (oxygenated blood) and deoxyhemoglobin (non-oxygenated blood) differ in their absorption of red and infrared light (spectrophotometry).
• The amount of arterial blood in tissue changes with your pulse (photoplethysmography). Therefore, the amount of light absorbed by the varying quantities of arterial blood changes as well. | deoxyhemoglobin (non-oxygenated blood) differ in their absorption of red and infrared light (spectrophotometry).
• The amount of arterial blood in tissue changes with your pulse (photoplethysmography). Therefore, the amount of light absorbed by the varying quantities of arterial blood changes as well. | | | |
| Supported Measured Pulse Oximetry | SpO2, PR | SpO2, PR | SpO2, PR | N/A | Same. |
K243305
Page 9 of 13
{13}
MASIMO
MASIMO CORPORATION
52 Discovery
Irvine, CA 92618
510(k) Summary
| Feature 510(k) Number | Masimo W1 Subject Device | Masimo W1 Primary Predicate (K240229) | Withing Scan Monitor Secondary Predicate (K201456) | Samsung ECG Monitor Tertiary Predicate (K201168) | Comparison to Primary Predicate |
| --- | --- | --- | --- | --- | --- |
| Parameters | | | | | |
| Supported Measured ECG Parameters | HR | HR | HR, HRV | HR | Same. |
| Supported Rhythm Classifications | Atrial Fibrillation, Sinus Rhythm | N/A | Atrial Fibrillation, Sinus Rhythm | Atrial Fibrillation, Sinus Rhythm | Different. The subject device provides an Atrial Fibrillation classification feature similar to the secondary and tertiary predicate devices.
Testing is provided to support the substantial equivalence. |
| Supported Calculated features | Pi | Pi | N/A | N/A | Same. |
| Lead | A single channel ECG, similar to Lead 1 ECG. | A single channel ECG, similar to Lead 1 ECG. | A single channel ECG, similar to Lead 1 ECG. | A single channel ECG, similar to Lead 1 ECG. | Same. |
| Performance specifications | | | | | |
| SpO2, (70-100%) | 2%, adults (No Motion/Low Perfusion) | 2%, adults (No Motion/ Low Perfusion) | 3%, adults (No Motion) | N/A | Same. |
| Pulse Rate (25-240 bpm) | 3 bpm | 3 bpm | Not supported | N/A | Same. |
| Heart Rate (25-240 bpm) | 5 bpm | 5 bpm | Not Known | N/A | Same. |
| Rhythm Classification | Atrial Fibrillation Sensitivity: 99.3% | N/A | Atrial Fibrillation Sensitivity: 96.3% | Atrial Fibrillation Sensitivity: 98.1% | AFib feature is being added. The performance has been evaluated to be |
K243305
Page 10 of 13
{14}
MASIMO
MASIMO CORPORATION
52 Discovery
Irvine, CA 92618
510(k) Summary
| Feature 510(k) Number | Masimo W1 Subject Device | Masimo W1 Primary Predicate (K240229) | Withing Scan Monitor Secondary Predicate (K201456) | Samsung ECG Monitor Tertiary Predicate (K201168) | Comparison to Primary Predicate |
| --- | --- | --- | --- | --- | --- |
| Performance | Sinus Rhythm Specificity: 100% | | Sinus Rhythm Specificity: 100% | Sinus Rhythm Specificity: 100% | substantially equivalent to the secondary and tertiary predicates. |
| Electrical Specifications | | | | | |
| Battery | Internal Rechargeable | Internal Rechargeable | Internal Rechargeable | N/A | Same. |
| Mechanical Specifications | | | | | |
| Watch Face Size | 40 mm (1.57") | 40 mm (1.57") | 38 mm or 42 mm | N/A | Same. |
| Weight | 54 g (including watchband) | 54 g (including watchband) | 58 g or 83g | N/A | Same. |
| Environmental Specifications | | | | | |
| Operating Temperature | 0 to 35 °C (32 to 95°F) | 0 to 35 °C (32 to 95°F) | -10 to 45 °C (14 to 113°F) | N/A | Same. |
| Operating Humidity | 10% to 95%, non-condensing | 10% to 95%, non-condensing | Not Known | N/A | Same. |
| Storage/Transport Temperature | -20 to 60°C (-4 to 140°F) | -20 to 60°C (-4 to 140°F) | -20 to 85°C (4 to 185°F) | N/A | Same. |
| Storage/Transport Humidity | 10% to 95%, non-condensing | 10% to 95%, non-condensing | Not Known | N/A | Same. |
| Classification per IEC 60601-1 | | | | | |
| Electrical Safety | IEC 60601-1 | IEC 60601-1 | IEC 60601-1 | N/A | Same. |
| EMC | IEC 60601-1-2 | IEC 60601-1-2 | IEC 60601-1-2 | N/A | Same. |
| Ingress Protection | IP24 | IP24 | Not Known. | N/A | Same. |
| Mode of Operation | Continuous | Continuous | Continuous | N/A | Same. |
K243305
{15}
MAsIMO
MASIMO CORPORATION
52 Discovery
Irvine, CA 92618
# 510(k) Summary
## 5 Performance Data
As part of this submission, clinical and non-clinical testing was included to support addition of the Atrial Fibrillation Classification Feature.
## Performance Bench Testing
Bench testing for the Masimo W1 was included in this submission to support the addition of the Atrial Fibrillation Classification Feature.
## Biocompatibility Testing
As there were no changes made to the patient contacting materials of the subject device from the previous clearance under K240229, no new biocompatibility testing was included in this submission.
## Electromagnetic Compatibility, Electrical Safety, Environmental, Mechanical and Cleaning
As there were no hardware changes made to the subject device from the previous clearance under K240229, no new electrical safety, environmental, mechanical, and cleaning testing was included as part of this submission.
## Software Verification and Validation Testing
Software verification and validation testing were conducted and provided as part of this submission as recommended by the FDA Guidance, “Content of Premarket Submissions for Software Device Software Functions”, June 2023 to support the addition of the AFib detection feature.
The Masimo W1 software was found to fit the category of products that would require Basic Documentation Level because the failure or latent flaw of the device software function would not present a hazardous situation with a probable risk of death or serious injury to either a patient, user of the device or others in the environment of use prior to the implementation of the risk controls.
## Wireless Testing
As there were no changes made to the subject device from the previous clearance under K240229, no new wireless testing was included as part of this submission.
## Cybersecurity Testing
As there were no changes made to the subject device wireless communication from the previous clearance under K240229, thus no new cybersecurity testing was included as part of this submission.
## Human Factors and Usability Testing
Human factors testing demonstrated the acceptability of the human factors and usability risks.
K243305
Page 12 of 13
{16}
MAsIMO
MASIMO CORPORATION
52 Discovery
Irvine, CA 92618
# 510(k) Summary
## Clinical Testing
To support the performance of the Masimo W1 with Atrial Fibrillation Classification Feature, clinical validation testing is provided as part of this submission.
## Masimo W1 ECG AFib Classification feature Performance
Prospective clinical testing was conducted to validate the AFib Classification Feature on adult subjects from 4 different sites. The testing included subjects with wide range of age and BMI.
The breakdown of the performance results, including the 95% Confidence Interval, are provided below in Table 5-1.
| Table 5-1 | |
| --- | --- |
| Performance Metrics | Performance Result |
| Sensitivity | 99.3% [96.3%,100%] |
| Specificity | 100% [97.8%, 100%] |
| PPV | 100% [97.5%, 100%] |
| Unclassified Rate | 5.0% |
| Noise Rate | 1.7% |
## Masimo W1 ECG Waveform Quality Analysis
To assess the quality of Masimo W1 ECG waveform in adults, the ECG waveform was accessed qualitatively and quantitatively by comparing similarity between Masimo W1 and gold-standard 12 lead ECG as reference. The qualitative assessment found a 98% [95% CI: 96% -98%] agreement by qualified clinicians in the analysis of the Masimo W1 ECG waveforms to that of the reference Lead I of a gold-standard 12-lead ECG reference.
The quantitative analysis showed key ECG features (QRS amplitude and QRS width) of the Masimo W1 ECG waveforms were similar to Lead I of a gold-standard 12-lead ECG reference.
## 6 Conclusion
The data supported the substantial equivalence of the subject device.
K243305
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.