K243299 · Siemens Medical Solutions USA, Inc. · IYN · Nov 15, 2024 · Radiology
Device Facts
Record ID
K243299
Device Name
ACUSON Redwood Diagnostic Ultrasound System
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
IYN · Radiology
Decision Date
Nov 15, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Ultrasound-Derived Fat Fraction (UDFF)
—
Pearson correlation coefficient p-Value < 0.05; Test-retest ICC > 0.75; Exam time < 60 seconds; Absolute mean difference < 5 percentage points; Bland-Altman limits of agreement within ±15 percentage points; AUROC > 0.80
Pearson correlation coefficient = 0.98 (p=0.792); ICC = 0.98; Acquisition time < 22 seconds; Mean absolute difference = 1.59 percentage points; Bland-Altman limits of agreement = -3.1 to 6.3 percentage points; AUROC > 0.85
—
—
Study 1: 30 subjects (correlation and reliability); Study 2: 28 subjects (agreement with MRI-PDFF and clinical accuracy)
—
Indications for Use
The ACUSON Redwood ultrasound imaging system is intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Fetal, Abdominal, Pediatric, Small Parts, OB/GYN (useful for visualization of the ovaries, follicles, uterus, and other pelvic structures), Adult, Pediatric and Neonatal Cardiac, Pelvic, Neonatal Cephalic, Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, Transcranial, and Peripheral Vascular. The system supports the Ultrasound-Derived Fat Fraction (UDFF) measurement tool to report an index that can be useful as an aid to a physician managing adult patients with hepatic steatosis. The system also provides the ability to measure anatomical structures: fetal, abdominal, pediatric, small organ, neonatal cephalic, adult cephalic, cardiac (adult, pediatric and neonatal), transesophageal, transrectal, transvaginal, peripheral vessel, musculoskeletal (conventional), musculoskeletal (superficial), and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
Device Story
Mobile, software-controlled diagnostic ultrasound system; acquires harmonic ultrasound echo data; displays in B-Mode, M-Mode, Doppler (PW, CW, Color, Tissue, Amplitude), and 3D/4D modes. Operated by trained healthcare professionals in clinical settings. Features include Auto IMT (Intima-Media Thickness) and Virtual Touch UDFF (Ultrasound-Derived Fat Fraction) for hepatic steatosis management. System provides anatomical measurements and calculation packages used adjunctively by physicians for clinical diagnosis. Output displayed on flat panel; aids clinical decision-making through visualization and quantitative indices.
Clinical Evidence
Bench testing confirmed safety and performance. UDFF feature validated in two studies: Study 1 (N=30) compared subject device to reference device, showing Pearson correlation of 0.98 (p=0.792) and ICC of 0.98. Study 2 (N=28) compared UDFF to MRI-PDFF, showing Bland-Altman limits of agreement between -3.1 to 6.3 percentage points, mean absolute difference of 1.59 percentage points, and AUROC > 0.85.
Technological Characteristics
Mobile ultrasound system; harmonic imaging; supports multiple transducers (e.g., 9C2, 5V1, 8V3). Complies with IEC 60601-1, IEC 60601-2-37, IEC 62359, and ISO 10993. Software-controlled; features include Auto IMT and Virtual Touch UDFF. Connectivity via standard ultrasound interfaces.
Indications for Use
Indicated for fetal, abdominal, pediatric, small parts, OB/GYN, cardiac (adult/pediatric/neonatal), pelvic, cephalic (neonatal/adult), vascular, musculoskeletal, and transcranial imaging. UDFF tool indicated for adult patients with hepatic steatosis.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
November 15, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. Food & Drug Administration".
Siemens Medical Solutions USA, Inc. % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K243299
Trade/Device Name: ACUSON Redwood Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX, OIJ Dated: October 18, 2024 Received: October 18, 2024
Dear Prithul Bom:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# YANNA S. KANG -S
Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
# K243299
Device Name
ACUSON Redwood Diagnostic Ultrasound System
#### ndications for Use (Describe)
The ACUSON Redwood ultrasound imaging system is intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Fetal, Abdominal, Pediatric, Small Parts, OB/GYN (useful for visualization of the ovaries, follicles, uterus, and other pelvic structures), Adult, Pediatric and Neonatal Cardiac, Pelvic, Neonatal Cephalic, Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, Transcranial, and Peripheral Vascular.
The system supports the Ultrasound-Derived Fat Fraction (UDFF) measurement tool to report an index that can be useful as an aid to a physician managing adult patients with hepatic steatosis.
The system also provides the ability to measure anatomical structures: fetal, abdominal, pediatric, small organ, neonatal cephalic, adult cephalic, cardiac (adult, pediatric and neonatal), transesophageal, transvaginal, peripheral vessel, musculoskeletal (conventional), musculoskeletal (superficial), and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
# Operating Modes
- · 2D-mode
- 2D-mode with Harmonic imaging
- · Color flow Doppler
- Color (velocity)
- Power (energy)
- Doppler Tissue Imaging
- · Pulsed Wave Doppler
- Pulsed Wave Doppler Tissue Imaging
- High Pulsed Repetition Frequency Pulsed Wave Doppler
- · Continuous Wave Doppler
- Steerable Continuous Wave Doppler for phased array transducers
- Auxiliary Continuous Wave Doppler for pencil transducers
- · M-mode
- M-mode with Harmonic imaging
- Anatomical M-mode
- · Volume Imaging
Combined Modes
- · 2D-mode with Color
- · 2D-mode with power
- · 2D/Doppler
- · 2D/Doppler with Color
- · 2D/Doppler with power
- · 2D/M-mode
{4}------------------------------------------------
#### · 2D/M-mode with color
· 2D/Anatomical M-mode
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
#### 510(k) Summary
#### K243299
| Date: | November 15, 2024 | | | | |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------|----------------|--------------|--|--|
| 1. Sponsor: | Siemens Medical Solutions USA, Inc.<br>Ultrasound Division<br>22010 South East 51st Street<br>Issaquah, Washington 98029 | | | | |
| Contact Person: | Shilpa Rapaka<br>Phone: 512-913-1053 | | | | |
| Secondary Contact: | Emily Woo<br>Phone: 82-10-4697-2683 | | | | |
| 2. Device Name: | ACUSON Redwood Diagnostic Ultrasound System | | | | |
| Common Name:<br>Classification: | Diagnostic Ultrasound System with Accessories<br>Regulatory Class: II<br>Classification Panel: Radiology | | | | |
| | Classification Name | 21 CFR Section | Product Code | | |
| | Ultrasonic Pulsed Doppler Imaging System | 892.1550 | IYN | | |
| | Ultrasonic Pulsed Echo Imaging System | 892.1560 | IYO | | |
| | Diagnostic Ultrasound Transducer | 892.1570 | ITX | | |
| | Biopsy Needle Guide Kit | 892.1560 | OIJ | | |
| Manufacturing Site: | Siemens Healthcare s.r.o<br>Panattoni Park Kosice Airport ul.<br>Andreja Kvasa 5040 17, Kosice-Barca, Slovakia | | | | |
#### 3. Legally Marketed Predicate Device and Reference Device
The ACUSON Redwood Diagnostic Ultrasound System is multi-purpose diagnostic ultrasound system with proprietary software and optional accessories and is substantially equivalent to the company's own Ultrasound devices, the ACUSON Redwood (K210743) which is the primary predicate device and the ACUSON Sequoia (K232145) as the reference device.
{6}------------------------------------------------
# 4. Device Description
The ACUSON Redwood Diagnostic Ultrasound System is multi-purpose, mobile, softwarecontrolled, diagnostic ultrasound system with an on-screen display for thermal and mechanical indices related to potential bio-effect mechanisms. Its function is to acquire harmonic ultrasound echo data and display it in B-Mode, M-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Color M mode, Doppler Tissue Image mode, Amplitude Doppler Mode, Combination modes, Harmonic Imaging and 3D Imaging modes, and 4D imaging modes on a flat panel display for diagnostic ultrasound imaging.
#### 5. Intended Use/Indications for Use
The ACUSON Redwood ultrasound imaging system is intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Fetal, Abdominal, Pediatric, Small Parts, OB/GYN (useful for visualization of the ovaries, follicles, uterus, and other pelvic structures), Adult, Pediatric and Neonatal Cardiac, Pelvic, Neonatal Cephalic, Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, Transcranial, and Peripheral Vascular.
The system supports the Ultrasound-Derived Fat Fraction (UDFF) measurement tool to report an index that can be useful as an aid to a physician managing adult patients with hepatic steatosis.
The system also provides the ability to measure anatomical structures: fetal, abdominal, pediatric, small organ, neonatal cephalic, adult cephalic, cardiac (adult, pediatric and neonatal), transesophageal, transrectal, transvaginal, peripheral vessel, musculoskeletal (conventional), musculoskeletal (superficial), and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diaqnosis purposes.
Operating Modes
- 2D-mode .
- । 2D-mode with Harmonic imaging
- Color flow Doppler .
- Color (velocity) ー
- Power (energy) -
- Doppler Tissue Imaging
- Pulsed Wave Doppler .
- Pulsed Wave Doppler Tissue Imaging —
- High Pulsed Repetition Frequency Pulsed Wave Doppler -
- . Continuous Wave Doppler
- Steerable Continuous Wave Doppler for phased array transducers -
- Auxiliary Continuous Wave Doppler for pencil transducers ー
- M-mode
- M-mode with Harmonic imaging l
- l Anatomical M-mode
- Volume Imaqinq
#### Combined Modes
- 2D-mode with Color .
{7}------------------------------------------------
- 2D-mode with power
- . 2D/Doppler
- . 2D/Doppler with Color
- 2D/Doppler with power
- 2D/M-mode ●
- 2D/M-mode with color .
- 2D/Anatomical M-mode
# 6. Substantially Equivalent Devices and Summary of Technological Characteristics
The ACUSON Redwood Diagnostic Ultrasound System is substantially equivalent to the company's own Ultrasound devices that are already cleared for distribution in the United States. The predicate and reference devices of ACUSON Redwood are ACUSON Redwood (K210743) and ACUSON Sequoia (K232145) with regards to intended use, indications for use, technological characteristics (transducers, accessories, software features), safety, and effectiveness.
The submission device differs from the predicate/reference devices as follows:
- . The addition of 9C2 transducer, which is substantially equivalent to 9C2 transducer previously cleared on the ACUSON Sequoia (K232145).
- . The addition of Auto IMT (Intima-Media Thickness) feature, which is substantially equivalent to Auto IMT feature previously cleared on the ACUSON Sequoia (K232145).
- . The addition of Virtual Touch UDFF (Ultrasound-Derived Fat Fraction) feature, which is substantially equivalent to Virtual Touch UDFF feature previously cleared on the ACUSON Sequoia (K232145).
- . The addition of TCD (Transcranial doppler) Imaging, which is substantially equivalent to TCI previously cleared on the ACUSON Sequoia (K232145).
- . The modified ACUSON Redwood Ultrasound System updated the indications for use statement (Intended use) to add Virtual Touch UDFF (Ultrasound-Derived Fat Fraction) indication and Transcranial clinical application for TCD which were previously cleared on the reference device, ACUSON Sequoia (K232145) where transcranial application was covered by Adult Cephalic.
- . The addition of needle quides for 9C2 transducer and 9VE4 transducer, which are substantially equivalent to needle guides for 9C2 transducer and 9VE4 transducer previously cleared on the ACUSON Sequoia (K232145).
- . The addition of Adult Cephalic and Abdominal clinical applications on 5V1 transducer and Neonatal Cephalic clinical application on 8V3 transducer, which is substantially equivalent to clinical applications of 5V1, 4V1 and 8V3 transducer previously cleared on the ACUSON Sequoia (K232145).
All other hardware and software features of the ACUSON Redwood Diagnostic Ultrasound devices remain unchanged. The foundation of the ACUSON Redwood (this submission) is the ACUSON Redwood (K210743) hardware. ACUSON Redwood (this submission) reuses software developed for ACUSON Redwood (K210743) and adds features such as 9C2 transducer, Auto IMT, Virtual Touch UDFF (Ultrasound-Derived Fat Fraction), TCD from ACUSON Sequoia (K232145).
{8}------------------------------------------------
#### 7. Nonclinical Data
The subject devices have been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal. electrical, electromagnetic and mechanical safety and have been found to conform to applicable medical device safety standards. The systems comply with the following voluntary standards:
- I IEC 62359: Edition 2.1 2017-09. Ultrasonics - Field characterization - Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields / Combines IEC 62359 (2010-10) and AMD 1 (2017-09)
- AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 . (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005)
- י IEC 60601-1:2005/A1(2012), Medical electric equipment – Part 1: General requirements for basic safety and essential performance / This document and its separate amendments continue to be valid together with the consolidated version
- . IEC 60601-1-2 Edition 4.0 2014-02, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- I IEC 60601-2-18: Edition 3.0 2009-08, Medical electrical equipment - Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment
- . IEC 60601-2-37 Edition 2.1 2015, Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
- ISO 10993-1 Fifth edition 2018-08, Biological evaluation of medical devices Part 1: I Evaluation and testing within a risk management process
- . IEC 60601-1-6:2010+A1:2013+A2:2020 Medical Electrical Equipment Part 1-6, General Requirements for Basic Safety and Essential Performance- Collateral standard: Usability
- I ANSI AAMI ISO 14971: Medical devices- Applications of risk management to medical devices, 2019
- . FDA Ultrasound Guidance document, "Marketing Clearance of Diagnostic Ultrasound Systems and Transducers," issued in February 2023 (https://www.fda.gov/media/71100/download) for determining the measurement accuracy
{9}------------------------------------------------
# 8. Clinical Data
Summary of Clinical Tests
Since the ACUSON Redwood Diagnostic Ultrasound Systems use the same technology and principles as existing predicate device, clinical studies were not required to support substantial equivalence.
#### UDFF Study Summary
Ultrasound-Derived Fat Fraction (UDFF) is a software tool that measures tissue in a selected region of interest and reports an index for assessing fat content in the liver. UDFF can be useful as an aid to a physician managing adult patients with hepatic steatosis.
The 5C1 transducer on the ACUSON Redwood Ultrasound System is compatible with the UDFF option. The UDFF feature in the ACUSON Redwood Ultrasound System (functionality, algorithms for UDFF measurements and calculations, implementation, etc.) is identical (remains unchanged) when compared to the UDFF feature in the reference device.
Two different studies were conducted to evaluate the performance of the UDFF feature on the ACUSON Redwood Ultrasound System. The demographic distribution of the study population in the two different studies are presented in the following table:
| Study 1 | Study 2 |
|------------------------|------------------------|
| N= 30 subjects | N= 28 subjects |
| Ethnicity/Country | Ethnicity/Country |
| USA = 15 | Korea = 28 |
| Korea = 15 | |
| Sex | Sex |
| Male = 15 | Male = 20 |
| Female = 15 | Female = 8 |
| Age (years) | Age (years) |
| Mean ±SD = 41.1 ± 18.5 | Mean ±SD = 43.8 ± 18.0 |
| Range = 6 - 76 | Range = 19 - 74 |
| BMI (kg/m²) | BMI (kg/m²) |
| Mean ± SD = 27.2 ± 6.8 | Mean ± SD = 27.2 ± 2.7 |
| Range = 11 - 44 | Range = 23 - 33 |
{10}------------------------------------------------
The study details and results are presented in the following table:
| Study # | Study Objective | Evaluation Criteria | Results |
|---------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | To verify the correlation<br>between the performance of<br>the UDFF feature on<br>ACUSON Redwood (subject<br>device) and the UDFF feature<br>on ACUSON Sequoia<br>(reference device) | Significance level p-Value<br>< 0.05 in a two-tailed test | Pearson Correlation<br>coefficient between<br>UDFF on ACUSON<br>Redwood and ACUSON<br>Sequoia was 0.98, and<br>the difference is not<br>statistically significant<br>(p=0.792 and t-statistic =<br>-0.263). |
| | To verify the clinical reliability<br>of the UDFF feature on the<br>ACUSON Redwood System | Test-retest ICC greater<br>than 0.75 | ICC = 0.98. The reliability<br>of the UDFF feature on<br>ACUSON Redwood is<br>considered excellent. |
| | To record the UDFF<br>measurement acquisition time | Exam time less than 60<br>seconds | All subjects acquisition<br>time was within 22<br>seconds. |
| 2 | To verify that UDFF of<br>ACUSON Redwood shows<br>good agreement with MRI-<br>PDFF | Absolute mean difference<br>less than 5 percentage<br>points and Bland-Altman<br>limits of agreement within<br>±15 percentage points | Redwood UDFF and<br>MRI's PDFF had Bland-<br>Altman limits of<br>agreement between -3.1<br>to 6.3 percentage points<br>with a mean absolute<br>difference of 1.59<br>percentage points. |
| | To verify the clinical accuracy<br>of the ACUSON Redwood<br>UDFF feature as compared<br>to MRI- PDFF | AUROC greater than 0.80 | Performance of UDFF on<br>ACUSON Redwood<br>showed strong<br>performance as<br>compared to MRI-PDFF<br>with AUROC > 0.85. |
#### 9. Summary
Intended uses and other key features of the subject device are consistent with traditional clinical practice and FDA guidelines. The design and development process of the manufacturer conforms to 21 CFR 820 Quality System Regulation and ISO 13485:2016 quality system standards. The product is designed to conform to applicable medical device safety standards and compliance is verified through independent evaluation with ongoing factory surveillance. Diagnostic ultrasound systems have accumulated a long history of safe and effective performance. Therefore, it is the opinion of Siemens Medical Solutions USA. Inc. that the ACUSON Redwood system is substantially equivalent with respect to safety and effectiveness to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.