Goldtrace FSE is indicated for patients requiring continuous electronic intrapartum monitoring of the fetal heart rate during labor and delivery.
Device Story
Goldtrace FSE is a sterile, single-use fetal scalp electrode used in delivery wards by trained personnel. It consists of a gold-plated stainless steel spiral needle electrode, a reference plate, and a guide/drive tube assembly. The device attaches to the fetal presenting part after amniotic membrane rupture to capture fetal ECG signals. It connects to a compatible fetal monitor (e.g., Philips Avalon FM30) via a reusable legplate adaptor cable. The device acts as a passive conduit for the fetal ECG signal; it performs no signal conversion, analysis, or modification. The monitor performs all signal processing, analysis, and quantification. By providing a continuous, high-quality fetal ECG signal, the device assists clinicians in monitoring fetal heart rate during labor, facilitating clinical decision-making to ensure fetal well-being.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993), electrical safety (IEC 60601-1, IEC 60601-1-2, ANSI/AAMI EC12, ANSI/AAMI EC53), sterility validation (ISO 11135), shelf-life/package integrity (ASTM F1980, ASTM D4169), and mechanical testing. Comparative performance testing demonstrated equivalent signal transmission capability to the predicate device when used with a compatible fetal monitor.
Technological Characteristics
Materials: SS316 stainless steel with 10-25 µm gold-plated surface. Sensing: Spiral needle electrode with reference plate. Connectivity: Wired connection to external fetal monitor via legplate adaptor. Sterilization: Ethylene Oxide (EO). Standards: ISO 10993-1, ISO 11135, IEC 60601-1, IEC 60601-1-2, ASTM D4169, ASTM F1980.
Indications for Use
Indicated for patients requiring continuous electronic intrapartum monitoring of fetal heart rate during labor and delivery, specifically for fetuses and mothers after rupture of amniotic membranes.
Regulatory Classification
Identification
A fetal scalp circular (spiral) electrode and applicator is a device used to obtain a fetal electrocardiogram during labor and delivery. It establishes electrical contact between fetal skin and an external monitoring device by a shallow subcutaneous puncture of fetal scalp tissue with a curved needle or needles. This generic type of device includes nonreusable spiral electrodes and reusable circular electrodes.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 17, 2025
Neoventa Medical AB Johan Sundberg Chief Quality and Regulatory Officer Norra Ågatan 32 Mölndal, Västra Götaland SE-431 35 SWEDEN
Re: K243291
> Trade/Device Name: Goldtrace Fetal Spiral Electrode (FSE) (CNS000004) Regulation Number: 21 CFR 884.2675 Regulation Name: Fetal Scalp Circular (Spiral) Electrode and Applicator Regulatory Class: II Product Code: HGP Dated: October 18, 2024 Received: October 18, 2024
Dear Johan Sundberg:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}------------------------------------------------
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Monica D. Garcia -S
Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K243291
Device Name
Goldtrace Fetal Spiral Electrode (FSE) (CNS000004)
Indications for Use (Describe)
Goldtrace FSE is indicated for patients requiring continuous electronic intrapartum monitoring of the fetal heart rate during labor and delivery.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Neoventa Medical. The logo consists of the word "neoventa" in a teal sans-serif font, followed by the tagline "A better start in life" in a smaller font. To the right of the text is a teal graphic of a mother and child in an embrace. The text "Page 1 of 4" is in the upper left corner of the image.
K243291
## 510(k) Summary - Goldtrace Fetal Spiral Electrode (K243291)
#### I. SUBMITTER
Neoventa Medical AB Norra Agatan 32 SE-431 35 Mölndal, Sweden Phone: +46(0)31 758 32 00 Fax: +46(0)31 758 32 99 Date Summary Prepared: January 17, 2025
Contact Person: Johan Sundberg Chief Quality and Regulatory Officer Email: johan.sundberg@neoventa.com
#### II. DEVICE
Device Trade Name: Goldtrace Fetal Spiral Electrode (Goldtrace FSE), CNS000004 Device Common Name: Fetal Spiral Electrode (FSE) Regulation number: 21 CFR 884.2675 Regulation Name: Fetal scalp circular (spiral) electrode and applicator Regulatory Class: II Product code: HGP (electrode, circular (spiral), scalp and applicator) Classification Panel: Obstetrics/Gynecology
#### III. PREDICATE DEVICE
Device Name: Fetal Spiral Electrode Manufacturer: Clinical Innovations, Inc 510(k) Number: K030691
The predicate device has not been subject to a design-related recall.
#### IV. DEVICE DESCRIPTION
Goldtrace FSE is a sterile, single use device transmitting the fetal ECG signal for continuous electronic intrapartum monitoring.
The Goldtrace FSE is intended for use in fetuses and mothers after rupture of the amniotic membranes requiring continuous electronic intrapartum monitoring of the fetal heart rate or analysis of the fetal ECG waveform. Goldtrace FSE is intended to be used by medically trained personnel at delivery wards.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Neoventa Medical. The logo includes the company name in a stylized font, with the tagline "A better start in life" underneath. To the right of the name is an image of a mother and child.
The device consists of gold-plated stainless steel spiral needle electrode that attaches to the fetal presenting part. The device includes a reference plate to measure the reference voltage level in the amniotic fluid. A guide tube and drive tube are used for application to the fetal presenting part. The subject device is connected to a compatible monitor (e.g., Philips Avalon FM30) via Philips DECG Reusable Legplate adaptor cable to provide a fetal ECG signal to the monitor. The subject device does not perform any conversion, analysis, quantification, or modifications to the fetal ECG.
The Goldtrace FSE materials conform to FDA recognized consensus standard for medical use (ISO 10993-1: 2018) and the device is sterilized using Ethylene Oxide sterilization (EO).
- V. INDICATIONS FOR USE
Goldtrace FSE is indicated for patients requiring continuous electronic intrapartum monitoring of the fetal heart rate during labor and delivery.
#### COMPARISON OF INTENDED USE AND TECHNOLOGICAL CHARACTERISTICS VI. WITH THE PREDICATE DEVICE
The subject device and predicate device are intended to be used on patients during labor requiring electronic intrapartum monitoring of the fetal heart rate by means of a spiral electrode.
Following table provides comparison of the intended use and technological characteristics of the subject device with the predicate device:
| Specification | Subject Device (K243291) | Predicate Device (K030691) |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Intended Use | Goldtrace FSE is indicated for<br>patients requiring continuous<br>electronic intrapartum monitoring of<br>the fetal heart rate during labor and<br>delivery. | For patients requiring fetal heart<br>rate monitoring during labor. |
| Principle of Operation | Transmit fetal ECG signal for<br>recording, analysis and<br>quantification of the monitor | Transmit fetal ECG signal for<br>recording, analysis, and<br>quantification of the monitor |
| Electrode Design | Single spiral needle with twisted<br>two-wire concept. Guide tube and<br>drive tube used for application. | Single spiral needle with twisted<br>two-wire concept. Guide tube and<br>drive tube used for application. |
| FSE components | Spiral needle, Reference Plate,<br>Connector, Guide Tube, Drive Tube,<br>Drive Tube Grip, Wires | Spiral needle, Reference Plate,<br>Connector, Guide Tube, Drive<br>Tube, Drive Tube Grip, Wires |
| Materials used for<br>spiral electrode | SS316 stainless steel for electrodes<br>with 10-25 µm gold-plated surface | Stainless steel for electrodes |
| Spiral needle diameter | Diameter: 4.7mm (0.185 inches) ±<br>0.13mm (0.005 inches) (O.D.)<br>Outside Diameter: 0.55 mm (0.021<br>inches) ± 0.05 mm (0.002 inches) | Not publicly available |
| Hub | Outside diameter:<br>0.219 inches ± 0.005 inches (O.D.) | Not publicly available |
| Wires | Length: 60 cm (23.6 inches) ± 2 cm | Not publicly available |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the document number K243291 and page number 3 of 4. Below this is the logo for Neoventa, which is written in a teal color. Underneath the logo is the text "A better start in life".
| Shelf life | 5 years | Not publicly available |
|------------|---------|------------------------|
| Sterile | Yes | Yes |
| Single use | Yes | Yes |
The subject and predicate device have similar indications for use statements and have the same intended use - continuous intrapartum monitoring of fetal ECG. The subject device (Goldtrace Fetal Spiral Electrode) has similar fundamental scientific technology to the predicate device (Fetal Scalp Electrode). The subject device differs from the predicate device in terms of device materials, dimensions, and electrical properties. The different technological features of the subject device do not raise different questions of safety and effectiveness as compared to the predicate device.
#### VII. SUMMARY OF NON-CLINICAL PERFORMANCE DATA
The following performance data are provided in support of the substantial equivalence determination.
## Biocompatibility
Biological evaluation of the subject device has been performed in accordance with the 2024 FDA guidance "Use of International standard ISO 10993-1, Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk management process" as well as ISO 10993- 1:2018 "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". Biocompatibility of the subject device is supported by performed tests of the following biological endpoints:
- . Cytotoxicity (ISO 10993-5:2009)
- . Sensitization (ISO 10993-10:2010)
- Irritation (ISO 10993-10:2010)
- Systemic Toxicity (ISO 10993-11:2006)
- Material Mediated Pyrogenicity (USP38-NF33:2015 <151> Pyrogen Test) .
The subject device was found to be non-cytotoxic, non-sensitizing, non-irritating, not systemically toxic, and non-pyrogenic (material mediated).
## Electrical Safety and Electromagnetic Compatibility
Testing was conducted in accordance with IEC 60601- 1:2020, Medical electrical equipment -Part 1: General requirements for basic safety and essential performance, and IEC 60601-1-2:2014 + AMD:2020, Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests, to support the electrical safety and electromagnetic compatibility of the subject device. Device was also tested (with deviations consistent with the device's intended use) according ANSI/AAMI EC12 EC12:2000/(R)2015 - Disposable ECG electrodes and ANSI/AAMI EC53 2013/(R)2020 - ECG trunk cables and patient leadwires.
## Sterility and Shelf-Life
Following testing was conducted to support sterility and shelf-life of the subject device:
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Neoventa Medical. The logo includes the text "neoventa" in a teal color, followed by the tagline "A better start in life" in a smaller font. To the right of the text is a stylized image of a mother and child in a teal color. The image also includes the text "K243291 Page 4 of 4" in the upper left corner.
- . Ethylene Oxide (EO) sterilization and validation testing per ISO 11135:2014 and the 2024 FDA guidance "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile Guidance for Industry and Food and Drug Administration Staff".
- . Simulated transportation and associated package integrity testing in accordance with ASTM D4169-22.
- Shelf-life testing (accelerated aging) per ASTM F1980:21, including package integrity ● testing (dye penetration test per ASTM F1929-15, seal strength testing per ASTM F88/F88M-21 and visual assessment per ASTM F1886/F1886M-16).
## Mechanical Testing
Testing to support structural and mechanical specifications of the subject device was conducted, including pull apart test for the Goldtrace FSE connector and Legplate and threepoint bend test for the drive tube and the guide tube. The bend tests demonstrated the material consistency over batches and that the material in the tubes was pliable within the load range.
## Comparative Performance Testing
A comparative test was performed between the Goldtrace FSE and the predicate device (K030691). The two devices are equivalent in electrical properties, in their ability to transmit fetal ECG signals and perform equally when used together with Philips Avalon FM30 fetal monitor.
#### CONCLUSION VIII.
The performance data described above demonstrate that the Goldtrace FSE is as safe and effective as the predicate device, Fetal Spiral Electrode (K030691) and supports determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.