FUJIFILM Processor EP-8000; FUJIFILM Endoscope Model EG-860R

K243260 · Fujifilm Corporation · FET · Apr 11, 2025 · Gastroenterology, Urology

Device Facts

Record IDK243260
Device NameFUJIFILM Processor EP-8000; FUJIFILM Endoscope Model EG-860R
ApplicantFujifilm Corporation
Product CodeFET · Gastroenterology, Urology
Decision DateApr 11, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Processor EP-8000 - The EP-8000 is an endoscopic processor with an integrated light source that is intended to provide illumination, process electronic signals transmitted from a video endoscope and enable image recording. - This product can be used in combination with compatible medical endoscope, a monitor, a recorder and various peripherals. - It supplies air through the endoscope, for obtaining clear visualization and is used for endoscopic observation, diagnosis and treatment. - BLI (Blue Light Imaging), LCI (Linked Color Imaging), ACI (Amber-red Color Imaging) and FICE (Flexible spectral-Imaging Color Enhancement) are adjunctive tools for gastrointestinal endoscopic examinations which can be used to supplement Fujifilm white light endoscopy. BLI, LCI, ACI and FICE are not intended to replace histopathological sampling as a means of diagnosis. Endoscope Model EG-860R This product is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

Device Story

System comprises EP-8000 processor/light source and EG-860R video endoscope; used in clinical settings by physicians for upper GI tract observation, diagnosis, and treatment. Endoscope captures images via distal CMOS sensor; light delivered via fiber bundles from 4-LED source (violet, blue, green, amber). Processor converts electrical signals to video for monitor display. Features include WLI, BLI, LCI, FICE, and ACI (Amber-red Color Imaging) modes; ACI enhances red information/blood-like saturation. Multi-observation mode allows frame-by-frame switching between different imaging modes on main/sub-screens. Tablet connectivity allows remote operation. Output assists clinicians in identifying mucosal abnormalities; adjunctive to white light; not a replacement for histopathology. Benefits include improved visualization and diagnostic capability during endoscopic procedures.

Clinical Evidence

Bench testing only. Evaluated image performance (color reproduction, geometric distortion, resolution, DOF, ISO-SNR, dynamic range, intensity uniformity, FOV) and mechanical performance (force transmission, bending, suction, air/water volume). EP-8000 and EG-860R met all pre-defined acceptance criteria and demonstrated substantial equivalence to predicate devices.

Technological Characteristics

Steel-polycarbonate housing; 4-LED light source (violet, blue, green, amber); CMOS image sensor; fiber optic light guides. Connectivity: digital/analog monitor output, tablet interface, peripheral control. Power: 100-240V AC. Image processing modes: WLI, BLI, LCI, FICE, ACI. Multi-observation frame-switching capability.

Indications for Use

Indicated for visualization, diagnosis, and endoscopic treatment of the upper digestive tract (esophagus, stomach, duodenum) in patients undergoing gastrointestinal endoscopic examinations.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION April 11, 2025 Fujifilm Corporation % Chaitrali Kulkarni Sr. Regulatory Affairs Specialist FUJIFILM Healthcare Americas Corporation 81 Hartwell Ave, Suite 300 Lexington, Massachusetts 02421 Re: K243260 Trade/Device Name: FUJIFILM Processor EP-8000; FUJIFILM Endoscope Model EG-860R Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: FET, NTN, NWB, PEA, FDS Dated: March 12, 2025 Received: March 12, 2025 Dear Chaitrali Kulkarni: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K243260 - Chaitrali Kulkarni Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K243260 - Chaitrali Kulkarni Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Sivakami Venkatachalam -S for Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K243260 Device Name FUJIFILM Processor EP-8000; FUJIFILM Endoscope Model EG-860R Indications for Use (Describe) a. Processor EP-8000 - The EP-8000 is an endoscopic processor with an integrated light source that is intended to provide illumination, process electronic signals transmitted from a video endoscope and enable image recording. - This product can be used in combination with compatible medical endoscope, a monitor, a recorder and various peripherals. - It supplies air through the endoscope, for obtaining clear visualization and is used for endoscopic observation, diagnosis and treatment. - BLI (Blue Light Imaging), LCI (Linked Color Imaging), ACI (Amber-red Color Imaging) and FICE (Flexible spectral-Imaging Color Enhancement) are adjunctive tools for gastrointestinal endoscopic examinations which can be used to supplement Fujifilm white light endoscopy. BLI, LCI, ACI and FICE are not intended to replace histopathological sampling as a means of diagnosis. b. Endoscope Model EG-860R This product is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov “An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.” {4} 510(k) #: K243260 510(k) Summary Prepared on: 2025-03-26 # Contact Details 21 CFR 807.92(a)(1) | Applicant Name | Fujifilm Corporation | | --- | --- | | Applicant Address | 798 MIYANODAI KAISEI-MACHI ASHIGARAKAMI-GUN KANAGAWA 258-8538 Japan | | Applicant Contact Telephone | 704-517-4886 | | Applicant Contact | Ms. Chaitrali Kulkarni | | Applicant Contact Email | hcusregulatoryaffairs@fujifilm.com | | Correspondent Name | FUJIFILM Healthcare Americas Corporation | | Correspondent Address | 81 Hartwell Ave, Suite 300, Lexington, MA 0242, United States | | Correspondent Contact Telephone | 704-517-4886 | | Correspondent Contact | Ms. Chaitrali Kulkarni | | Correspondent Contact Email | hcusregulatoryaffairs@fujifilm.com | ## Device Name 21 CFR 807.92(a)(2) | Device Trade Name | FUJIFILM Processor EP-8000; FUJIFILM Endoscope Model EG-860R | | --- | --- | | Common Name | Endoscopic video imaging system/component, gastroenterology-urology | | Classification Name | Endoscope and accessories | | Regulation Number | 876.1500 | | Product Code(s) | FET, NTN, NWB, PEA, FDS | ## Legally Marketed Predicate Devices 21 CFR 807.92(a)(3) | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | | --- | --- | --- | | K163675 | Processor VP-7000, Light Source BL-7000 | FET | | K172916 | FUJIFILM Endoscope Model EG-760R | FDS | | K240142 | FUJIFILM Endoscope Model EG-840N, EG-840T | FDS | ## Device Description Summary 21 CFR 807.92(a)(4) A. Intended Use/indications for Use a. Processor EP-8000 {5} K243260 Page 2 of 5 - The EP-8000 is an endoscopic processor with an integrated light source that is intended to provide illumination, process electronic signals transmitted from a video endoscope and enable image recording. - This product can be used in combination with compatible medical endoscope, a monitor, a recorder and various peripherals. - It supplies air through the endoscope, for obtaining clear visualization and is used for endoscopic observation, diagnosis and treatment. - BLI (Blue Light Imaging), LCI (Linked Color Imaging), ACI (Amber-red Color Imaging) and FICE (Flexible spectral-Imaging Color Enhancement) are adjunctive tools for gastrointestinal endoscopic examinations which can be used to supplement Fujifilm white light endoscopy. BLI, LCI, ACI and FICE are not intended to replace histopathological sampling as a means of diagnosis. b. Endoscope Model EG-860R This product is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum. B. Technological characteristics a. Processor EP-8000 FUJIFILM Video Processor EP-8000 is intended to provide illumination, process electronic signals transmitted from a video endoscope and enable image recording. FUJIFILM Video Processor EP-8000 relays the image from an endoscope to a video monitor. The projection can be either analog or digital at the user's preference. The processor employs fiber bundles to transmit light from four LED lamps (Violet, Blue, Green, and Amber), with a total power of 79.2W lamps, to the body cavity. The EP-8000, like the VP-7000 and BL-7000, has additional image processing options called BLI, BLI-bright, and LCI that provide endoscopic assistance for white light imaging (WLI). There is also an additional image processing option called "ACI" (Amber-red Color Imaging). ACI is an image processing function that simultaneously emphasizes the brightness and color difference of red information in endoscopic images and serves as an adjunct to white light imaging (WLI). Compared to WLI mode, ACI relatively increases the ratio of amber red light and decreases the ratio of violet light. Relatively high-saturation red information such as blood-like red in the image signal digitized by the camera unit is enhanced by signal processing. The EP-8000 also has a Multi Observation option that allows endoscopic images to be displayed in the main screen area and sub-screen area by switching image processing options at every frame. This allows each image frame to be displayed in the main screen area and sub-screen area 1 with a different combination of image processing options applied [WLI+(LCI), LCI+(WLI), BLI+(WLI), WLI+(BLI)]. The device is AC operated at a power setting of 100-240V/50-60Hz/ 3.0-1.5A. The processor is housed in a steel-polycarbonate case measuring 395×210×515mm b. Endoscope Model EG-860R The insertion portion of the device has a mechanism (hereinafter "the bending portion") which bends the tip from right to left and up and down, and a flexible tube (hereinafter "the flexible portion") consists of the bending portion and operating portion with a knob which controls the bending portion. The forceps channel which runs through the operating portion to the tip is arranged inside the insertion portion for inserting the surgical instrument. The insertion portion of the endoscopes comes into contact with the mucosal membrane. The tip of the insertion portion is called the "Distal end" which contains the Imaging section, Distal cap, Objective lens, Air/water nozzle, Water jet nozzle, Instrument channel outlet, Objective lens, and Light guide. The bending portion is controlled by knobs on the control portion/operation section to angulate the distal end to certain angles. The Flexible portion refers to the long insertion area between the Bending portion and the Control portion (a part of Non-insertion portion). This portion contains light guides (glass fiber bundles), air/water channels, a forceps/suction channel, a CMOS image sensor, and cabling. The glass fiber bundles allow light to travel through the endoscope to illuminate the body cavity, thereby providing enough light to the CMOS image sensor to capture an image and display the image on a monitor. The forceps channel is used to introduce biopsy forceps and other endoscopic accessories, as well as providing suction. The control portion/operating section provides a grip to grasp the endoscopes and contains mechanical parts to operate the endoscopes. This section includes a Forceps inlet, which allows endoscope accessories to be introduced. The Scope connector connects the endoscopes to the light source and video processor, respectively. C. Principles of Operation {6} K243260 Page 3 of 5 a. Processor EP-8000 The subject device, FUJIFILM Video Processor EP-8000, combines the principles of operation of the predicate devices, FUJIFILM Video Processor VP-7000 and Light Source BL-7000. EP-8000 is intended to provide illumination, process electronic signals transmitted from a video endoscope during the endoscopic observation, diagnosis, treatment, and image recording. After connecting to the light source and processor, the devices guide the light through the fiber bundles situated inside the insertion portion of the device. The light that emits from the distal end of the insertion portion reflects from the target region and forms an image on the CMOS image sensor through a group of object lenses placed at the distal end of the devices. An electric signal from the image sensor is transmitted to the video processor connected to the device. The video processor converts the electric signal into a video signal and displays an image on a monitor on a monitor. b. EG-860R The Endoscope Model EG-860R function on the same principles of operation as the predicate device. EG-860R is endoscopes for observation, diagnosis, and treatment of the upper digestive tract. After connecting to the light source, the devices guide the light through the fiber bundles situated inside the insertion portion of the device. The light that emits from the distal end of the insertion portion reflects from the target region and forms an image on the CMOS image sensor through a group of object lenses placed at the distal end of the devices. An electric signal from the image sensor is transmitted to the video processor connected to the device. The video processor converts the electric signal into a video signal and displays an image on a monitor. Intended Use/Indications for Use 21 CFR 807.92(a)(5) a. Processor EP-8000 - The EP-8000 is an endoscopic processor with an integrated light source that is intended to provide illumination, process electronic signals transmitted from a video endoscope and enable image recording. - This product can be used in combination with compatible medical endoscope, a monitor, a recorder and various peripherals. - It supplies air through the endoscope, for obtaining clear visualization and is used for endoscopic observation, diagnosis and treatment. - BLI (Blue Light Imaging), LCI (Linked Color Imaging), ACI (Amber-red Color Imaging) and FICE (Flexible spectral-Imaging Color Enhancement) are adjunctive tools for gastrointestinal endoscopic examinations which can be used to supplement Fujifilm white light endoscopy. BLI, LCI, ACI and FICE are not intended to replace histopathological sampling as a means of diagnosis. b. Endoscope Model EG-860R This product is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum. Indications for Use Comparison 21 CFR 807.92(a)(5) a. Processor EP-8000 This product is an endoscopic processor with an integrated light source that is intended to provide illumination, process electronic signals transmitted from a video endoscope and enable image recording. This product can be used in combination with compatible medical endoscope, a monitor, a recorder and various peripherals. It supplies air through the endoscope, for obtaining clear visualization and is used for endoscopic observation, diagnosis and treatment. These indications for use are nearly identical to those of the predicate device Processor VP-7000 and Light source BL-7000(K163675), except for the addition of language providing clarification about the use of the ACI feature of the device. The company makes no ACI-specific claims in the indications for use or the labeling. These minor additions do not affect the diagnostic use of the device relative to the predicates, as they merely provide clarification around the use of ACI with the system. b. Endoscope Model EG-860R The indication for use for the proposed device and the predicate devices are same. Technological Comparison 21 CFR 807.92(a)(6) a. Processor EP-8000 The subject and predicate devices share the same mode of operation, number of LED lamps used for light transmission and intended {7} K243260 Page 4 of 5 use. The subject device is unchanged from the Predicate device's BL-7000, which uses the same four LED colors: violet, blue, green, and amber. The subject device differs from the predicate devices VP-7000 and BL-7000 in terms of technological characteristics. A summary of major differences between the subject device EP-8000 and the predicate devices VP-7000 and BL-7000 is provided as follows: - The subject device is comprised of one unit that provides both video processing and sourcing functions. The predicate devices separate the two functions into two separate. Risk analysis showed that having one unit for both video and sourcing function does not affect the safety of efficacy of the subject device. - The power rating is modified from 0.8-0.5A(VP-7000), 1.2-0.7A(BL-7000) to 2.0-1.1A. IEC testing was performed on the subject device, the difference in power rating does not increase the safety or efficacy of the subject device. - The subject device is compatible with the following. 【Gastrointestinal endoscope】800 System endoscopies,700 System endoscopies,600 System endoscopies,500 System endoscopies. The difference in the compatible endoscope systems does not affect the safety or efficacy of the subject device, the compatible scopes are listed on the Operation Manual for EP-8000. - The subject device is compatible with the DK-8000E keyboard, whereas the predicate devices are compatible with the DK-7000E keyboard. The difference in compatible keyboard does not affect the safety or efficacy of the subject device, the compatible keyboard is provided with EP-8000. - User modes changes from FICE, BLI, BLI-bright and LCI to FICE, BLI, BLI-bright, LCI and ACI. Color Performance testing was performed on EP-8000, the addition of ACI user mode does not increase the safety or efficacy of the subject device. - The subject device does not offer analog output via Y/C and analog input. Not having the analog output via Y/C or analog input does not affect the safety or efficacy of the subject device because the features were taken out due to increase of digitization of external devices and the main function of the device is not changed. - The subject device has a Peripheral device control function. The peripheral device control function is a function that allows the processor device for endoscopes to centrally control the connected peripheral devices, but since it is the same as the control that is originally performed by each peripheral device, it does not lead to any new procedures. Therefore, it is not a change that affects the safety or effectiveness of the processor device for endoscopes. - The subject device has a Function to operate this product via a tablet with an operation extension program installed by connecting this product and a tablet. Risk analysis was performed, the addition of the tablet operation in the subject device, does not increase the safety or efficacy of the subject device. - The subject device also has a Multi Observation option that allows endoscopic images to be displayed in the main screen area and sub-screen area by switching image processing options at every frame. This allows each image frame to be displayed in the main screen area and sub-screen area 1 with a different combination of image processing options applied [WLI+(LCI), LCI+(WLI), BLI+(WLI), WLI+(BLI)]. The effect of the image frame option is unchanged and does not affect the safety and effectiveness of the image processing option as it only switches the image processing option frame by frame on the display. - The dimensions are modified from 390x110x485mm (for VP-7000) and 390x155x485mm BL-7000) to 395x210x515 mm. A change in the dimensions does not affect how the device is used, the difference in dimensions does not increase the safety or efficacy of the subject device. - The weight is modified from 9kg (for VP-7000) and 12kg (BL-7000) to 18kg. The change in the device weight does not affect how the device is used, the difference in the weight does not increase the safety or efficacy of the subject device. b. Endoscope Model EG-860R The product code for both predicate device and subject device is FDS. The difference in the F# of the objective lens in EG-860R compared to the predicate device does not affect the safety or efficacy of the subject device because this value is between Predicate and Reference devices. Although there is a difference in the resolution of EG-860R compared to the predicate device, the resolution of EG-860R is the same as the reference devices and does not increase the safety or efficacy of the subject device. The difference in the distal end diameter of EG-860R compared to the subject device does not affect the safety and efficacy of the subject device because this value is between Predicate and Reference devices. {8} K243260 Page 5 of 5 The difference in the angle rubber tube material and tube material of EG-860R compared to the predicate device does not affect the safety of efficacy. Biocompatibility testing was completed, no new risks were introduced. ## Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) The performance of Image and the performance of the WLI, FICE, BLI, BLI-bright, LCI and ACI imaging modes available in FUJIFILM Video Processor EP-8000 was evaluated according to the engineering requirements listed in this section. The test protocols are similar to those used to evaluate the performance of the predicate devices VP-7000 and BL-7000. The engineering requirements are identical to those assessed for the predicate devices. EP-8000 was evaluated for as follows. The image performance of EP-8000 was compared with the image performance of the predicate VP-7000 and BL-7000. Testing was conducted in combination with representative FUJIFILM gastroscopes. The following test reports are provided: (1) color reproduction, (2) Image Geometric Distortion, (3) Image Resolution Performance, (4) Depth of field (DOF) Performance Test, (5) ISO-SNR & Dynamic Range Performance, (6) Image intensity uniformity, (7) FIELD OF VIEW (FOV). For all imaging modes, EP-8000 demonstrated substantial equivalence to VP-7000 and BL-7000 in Image performance and color reproduction. Although there are minor differences between the proposed and predicate devices, these differences do not raise new or additional questions of safety or effectiveness of the proposed devices. Thus, the proposed device EP-8000 is substantially equivalent to the predicate device. The following performance for EG-860R was evaluated according to the engineering requirements listed in this section. The test protocols are similar to those used to evaluate the performance of the predicate devices. The engineering requirements are identical to those assessed for the predicate devices. In all cases, the devices met the pre-defined acceptance criteria for the test. (1) Image Geometric Distortion, (2) Image Resolution Performance, (3) Depth of field (DOF) Performance Test, (4) ISO-SNR & Dynamic Range Performance, (5) Image intensity uniformity, (6) Advanced Force Transmission, (7) Adaptive Bending, (8) Field of view, (9) Bending capability, (10) Rate of suction, (11) Working length, (12) Diameter of Forceps channel, (13) Viewing direction, (14) Resolution, (15) LG output, (16) Uneven illumination, (17) Color reproducibility, (18) Air volume, (19) Water volume Although there are minor differences between the proposed and predicate devices, these differences do not raise new or additional questions of safety or effectiveness of the proposed devices. Thus, the proposed device EG-860R is substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...