Parkell Pit and Fissure Sealant

K243254 · Parkell, Inc. · EBC · Aug 8, 2025 · Dental

Device Facts

Record IDK243254
Device NameParkell Pit and Fissure Sealant
ApplicantParkell, Inc.
Product CodeEBC · Dental
Decision DateAug 8, 2025
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 872.3765
Device ClassClass 2

Indications for Use

The Parkell Pit and Fissure Sealant is indicated for prophylactic sealing of pits and fissures as well as for micro-restorative procedures for composite restorations.

Device Story

Light-cured, flowable, resin-based sealant; used for prophylactic sealing of pits and fissures and micro-restorative procedures. Contains filling agents that release and recharge fluoride and calcium ions. Intended for use by licensed healthcare professionals in a clinical setting. Applied via syringe with applicator tip; cured using standard power (600 mW/cm²) light source for 20 seconds. Provides physical barrier to prevent dental caries; ion release supports tooth health. No sterilization required.

Clinical Evidence

Bench testing only. No clinical data performed. Physical property testing confirmed compliance with established pass/fail criteria for bond strength (≥10 MPa), compressive strength (≥200 MPa), flexural strength (≥90 MPa), depth of cure (1.5-2.5 mm), shrinkage (≤8%), and viscosity (1-5 Pa.s).

Technological Characteristics

Light-cured, resin-based, low-viscosity paste. Contains monomers, fillers, photoinitiator/coactivators, calcium donor, inhibitor, fluorescence agent, and colorant. Delivery via 1.2 mL syringe. Curing time 20 seconds at 600 mW/cm². Non-sterile.

Indications for Use

Indicated for prophylactic sealing of pits and fissures and micro-restorative procedures for composite restorations in dental patients.

Regulatory Classification

Identification

A pit and fissure sealant and conditioner is a device composed of resin, such as polymethylmethacrylate, intended for use primarily in young children to seal pit and fissure depressions (faults in the enamel) in the biting surfaces of teeth to prevent cavities.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION August 8, 2025 Parkell, Inc. % Keith Barritt Attorney Fish & Richardson P.C. 1000 Maine Avenue, S.W. Suite 1000 Washington, District of Columbia 20024 Re: K243254 Trade/Device Name: Parkell Pit and Fissure Sealant Regulation Number: 21 CFR 872.3765 Regulation Name: Pit And Fissure Sealant And Conditioner Regulatory Class: Class II Product Code: EBC Dated: July 8, 2025 Received: July 8, 2025 Dear Keith Barritt: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K243254 - Keith Barritt Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See {2} K243254 - Keith Barritt Page 3 the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, ![img-0.jpeg](img-0.jpeg) Bobak Shirmohammadi -S For Michael E. Adjodha, M.ChE., RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) k243254 Device Name Parkell Pit and Fissure Sealant Indications for Use (Describe) The Parkell Pit and Fissure Sealant is indicated for prophylactic sealing of pits and fissures as well as for micro-restorative procedures for composite restorations. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4} 510(k) Summary -K243254 Parkell, Inc. Parkell Pit and Fissure Sealant **Submitter** **(i) 510(k) Submitter** Parkell, Inc. 300 Executive Drive Edgewood, NJ 11717 FDA Registration No. 2411797 **(ii) 510(k) Submitter Contact** Keith A. Barritt Fish & Richardson P.C. 1000 Maine Avenue, S.W. Suite 1000 Washington, DC 20024 Phone: (202) 626-6433 Facsimile: (202) 783-2331 Email: barritt@fr.com **(iii) Preparation Date** August 8, 2025 **Device** Trade or Proprietary Name: Parkell Pit and Fissure Sealant Common Name: pit and fissure sealant and conditioner Classification Name: sealant, pit and fissure, and conditioner Product Code: EBC, 21 CFR 872.3765 Class: 2 {5} # Predicate Device The Parkell Pit and Fissure Sealant device (the “Device”) is substantially equivalent to Premier Dental Company Products’ “Premier Sealant” (K#161580, the “Predicate Device”). The Device is indicated for “prophylactic sealing of pits and fissures as well as for micro-restorative procedures.” The Indications for Use statement for the Predicate Device is nearly identical, namely “prophylactic sealing of pits and fissures. It may also be used for micro-restorative or initial layer of composite restorations.” The minor difference in wording is irrelevant and at most is narrowing, as the Indications for Use for the Device does not refer to “or initial layer” as does the Indications for Use for the Predicate Device. # Device Description The Device is a light-cured, flowable, resin-based sealant for prophylactic sealing of pits and fissures as well as for micro-restorative procedures. The Device is light-curable under standard power (600 mW/cm²) and contains filling agents which cause the Device to release and recharge beneficial ions (fluoride and calcium). The device is intended for use by licensed healthcare professionals only. The device does not come sterilized and is not intended to be sterilized prior to use. # Indications for Use The Device is indicated for prophylactic sealing of pits and fissures as well as for micro-restorative procedures for composite restorations. # Comparison of Technological Characteristics The Device has the same basic technological characteristics in terms of design, material, and chemical composition as the Predicate Device and is substantially equivalent thereto as shown below: {6} | Property | Parkell Device Pit and Fissure Sealant (Parkell, Inc.) K243254 | Predicate Device 510(k) no. K161580, BioCoat, filed as “Premier Sealant” (Premier Dental Company) | | --- | --- | --- | | Intended uses | Parkell Pit and Fissure Sealant is indicated for prophylactic sealing of pits and fissures as well as for micro-restorative procedures for composite restorations. | Premier Sealant, a prescription only medical device used for prophylactic sealing of pits and fissures. It may also be used for micro-restorative or “initial layer” of composite restorations | | Classification Product Code | EBC | EBC | | Regulation Number | 872.3765 | 872.3765 | | Principle of operation | Prophylactic sealing of pits and fissures | Prophylactic sealing of pits and fissures | | Material form | Low viscosity paste | Low viscosity paste | | Polymerization Method | Light-Cure | Light-Cure | | Curing Time | 20 sec | 20 sec | | Delivery System | 1.2 mL syringe | 1.2 mL syringe | | Bond Strength to Dry Enamel (MPa) Pass/Fail Criteria: ≥10 MPa | ≥10 MPa | ≥10 MPa | | Compressive Strength (MPa) Pass/Fail Criteria: ≥200MPa | ≥200MPa | ≥200MPa | | Flexural Strength (MPa) Pass/Fail Criteria: ≥90MPa | ≥90MPa | ≥90MPa | | Depth of cure (mm) Pass/Fail Criteria: 1.5-2.5mm | 1.5-2.5mm | 1.5-2.5mm | | Shrinkage (%) Pass/Fail Criteria: ≤8% | ≤8% | ≤8% | | Viscosity (initial) (Pa.S) Pass/Fail Criteria: 1-5Pa.s. | 1-5Pa.s | 1-5Pa.s | | Beneficial Ion Release and Recharge | Fluoride Calcium | Fluoride Calcium Phosphate | | Accessories | Applicator tip | Applicator tip | ## Shelf Life Testing The Device has been tested to establish a shelf life of two years. {7} # Material And Chemical Composition The Device does come into direct contact with the patient. Though the precise composition is confidential, the Device contains monomers, fillers, photoinitiator/coactivators, a calcium donor, inhibitor, fluorescence agent, and colorant, and is substantially equivalent to the Predicate Device in terms of ingredients and biocompatibility. # Performance Data Summary Non-clinical testing of the physical properties of the Device was conducted. There were no clinical tests performed for the Device. A comparison of the Device with the Predicate Device is shown in the table below: {8} | Property | Parkell Device Pit and Fissure Sealant (Parkell, Inc.) K243254 | Predicate Device 510(k) no. K161580, BioCoat, filed as “Premier Sealant” (Premier Dental Company) | | --- | --- | --- | | Intended uses | Parkell Pit and Fissure Sealant is indicated for prophylactic sealing of pits and fissures as well as for micro-restorative procedures for composite restorations. | Premier Sealant, a prescription only medical device used for prophylactic sealing of pits and fissures. It may also be used for micro-restorative or “initial layer” of composite restorations | | Classification Product Code | EBC | EBC | | Regulation Number | 872.3765 | 872.3765 | | Principle of operation | Prophylactic sealing of pits and fissures | Prophylactic sealing of pits and fissures | | Material form | Low viscosity paste | Low viscosity paste | | Polymerization Method | Light-Cure | Light-Cure | | Curing Time | 20 sec | 20 sec | | Delivery System | 1.2 mL syringe | 1.2 mL syringe | | Bond Strength to Dry Enamel (MPa) Pass/Fail Criteria: ≥10 MPa | ≥10 MPa | ≥10 MPa | | Compressive Strength (MPa) Pass/Fail Criteria: ≥200MPa | ≥200MPa | ≥200MPa | | Flexural Strength (MPa) Pass/Fail Criteria: ≥90MPa | ≥90MPa | ≥90MPa | | Depth of cure (mm) Pass/Fail Criteria: 1.5-2.5mm | 1.5-2.5mm | 1.5-2.5mm | | Shrinkage (%) Pass/Fail Criteria: ≤8% | ≤8% | ≤8% | | Viscosity (initial) (Pa.S) Pass/Fail Criteria 1-5Pa.s. | 1-5Pa.s | 1-5Pa.s | | Beneficial Ion Release and Recharge | Fluoride Calcium | Fluoride Calcium Phosphate | | Accessories | Applicator tip | Applicator tip | ## Conclusion Based on the non-clinical testing conducted of the physical properties of the Device in comparison to the Predicate Device identified above, and based on the biocompatibility of ingredients for the Device, Parkell concludes that the Device is substantially equivalent to the Predicate Device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...