K243230 · Pearl, Inc. · QIH · May 9, 2025 · Radiology
Device Facts
Record ID
K243230
Device Name
Second Opinion® BLE
Applicant
Pearl, Inc.
Product Code
QIH · Radiology
Decision Date
May 9, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K243230 · May 9, 2025
Second Opinion® BLE
Pearl, Inc.
Retrospective clinical dental radiographs (bitewing and periapical)
Retrospective clinical images were used to validate the performance of the Second Opinion® BLE software in measuring interproximal bone levels. The study assessed precision and recall against a ground truth established by dental professionals.
Second Opinion BLE Clinical Validation; Retrospective performance assessment
Patients 12 years of age or older requiring bitewing and periapical radiographs; Sample Size: 396 subjects (229 bitewing and 167 periapical radiographs)
Not applicable for this study
Precision and recall of interproximal bone level measurements; mean absolute difference in mm
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Interproximal bone level measurements
Neural network-based computer vision algorithms
Precision > 82%, Recall > 82%, Mean absolute difference < 1.5 mm
Precision: 87% (periapical), 87% (bitewing); Recall: 87% (periapical), 91% (bitewing); Mean absolute difference: 0.45 mm (periapical), 0.86 mm (bitewing)
—
—
Retrospective clinical subject data from 229 bitewing and 167 periapical radiographs from 396 subjects
Second Opinion® BLE is a radiological automated image processing software device intended to identify and display bone level measurements in bitewing and periapical radiographs. It should not be used in lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis. It is designed to aid dental health professionals to review bitewing and periapical radiographs of permanent teeth in patients 12 years of age or older as a concurrent and second reader.
Device Story
Second Opinion® BLE is a radiological automated image processing software; aids dental professionals in reviewing bitewing and periapical radiographs. Input: digital intraoral radiographs. Processing: neural network-based computer vision algorithms detect interproximal bone levels (CEJ-bonecrest). Output: linear bone level measurements displayed as graphical overlays on the original radiograph. Used in dental clinics; operated by dentists/dental hygienists. Workflow: images sent via API to ML modules; metadata generated; UI renders detections as overlays. Clinicians toggle highlights to review potential conditions. Benefit: provides automated second-reader assistance for bone height assessment; improves consistency and efficiency in radiographic review; does not replace full clinical evaluation.
Clinical Evidence
Standalone performance assessment using 396 subjects (229 bitewing, 167 periapical radiographs). Three licensed dentists labeled ground truth; two radiologists adjudicated discrepancies. Results: Periapical images (87% precision, 87% recall); Bitewing images (87% precision, 91% recall). Mean absolute difference: 0.45 mm (periapical) and 0.86 mm (bitewing). All metrics met pre-specified acceptance criteria (>82% precision/recall; <1.5 mm mean absolute difference; p<0.05).
Technological Characteristics
Software-only device; runs on Windows OS. Utilizes neural network-based computer vision algorithms developed from open-source models using supervised machine learning. Cloud-based processing environment. Displays linear bone level measurements as graphical overlays on radiographs. Moderate level of software concern. No hardware, biocompatibility, or sterility requirements.
Indications for Use
Indicated for dental health professionals to review bitewing and periapical radiographs of permanent teeth in patients 12 years of age or older as a concurrent and second reader for bone level measurements.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 9, 2025
Pearl Inc.
% Ashley Brown
Official Correspondent
2515 Benedict Canyon Dr.
BEVERLY HILLS, CA 90210
Re: K243230
Trade/Device Name: Second Opinion® BLE
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH
Dated: April 3, 2025
Received: April 3, 2025
Dear Ashley Brown:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243230 - Ashley Brown
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K243230 - Ashley Brown
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K243230 | |
| Device Name Second Opinion® BLE | |
| Indications for Use (Describe) Second Opinion® BLE is a radiological automated image processing software device intended to identify and display bone level measurements in bitewing and periapical radiographs. It should not be used in lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis. It is designed to aid dental health professionals to review bitewing and periapical radiographs of permanent teeth in patients 12 years of age or older as a concurrent and second reader. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov “An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.” | |
{4}
Pearl
K243230
# 510(K) SUMMARY
## 1. Submitter's Identification
Pearl Inc.
2515 Benedict Canyon Dr.
Beverly Hills, CA, 90210
USA
(239) 450-8829
Contact Person: Ashley Brown Position: Director of Regulatory Affairs
Date Summary Prepared: May 9th, 2025
## 2. Trade Name of the Device
Second Opinion® BLE
## 3. Common or Usual Name
Radiological Automated Image Processing System
## 4. Classification Name, Regulatory Classification & Product Code
Classification Name: Medical Image Management and Processing System
Regulatory Classification: 21CFR 892.2050, Class II
Product Code: QIH (Radiological Automated Image Processing System)
## 5. Predicate Information
Predicate device: The primary proposed predicate device for clearance of Second Opinion® BLE is Overjet Dental Assist by Overjet Inc., cleared in (K210187, May 19, 2021), a semi-automated software device classified as a Class II device pursuant to 21 CFR §892.2050 Medical image management and processing system under product code LLZ.
Overjet Dental Assist is a radiological semi-automated image processing software device intended to aid dental professionals in the measurements of mesial and distal bone levels associated with each tooth from bitewing and periapical radiographs.
The Overjet Dental Assist device has the same technology (AI software) and intended use as automated image processing to aid dental professionals in the measurement and visualization of bone levels associated with bitewing and periapical radiographs.
A reference device for the clearance of Second Opinion ® BLE is Video Perio Assist, a radiological semi-automated image processing software device classified as a
Hallopearl.com
{5}
Pearl
K243230
Class II device pursuant to 21 CFR §892.2050 Medical image management and processing system under product code QIH. The reference device was cleared under K223296 on February 6 2023. The reference device is to justify the device's safety and effectiveness by providing benchmarks for performance testing.
The device is a radiological semi-automated image processing software device intended to aid dental professionals in the measurements and visualization of mesial and distal bone levels associated with each tooth from bitewing and periapical radiographs. Measurements are made available as linear distances or relative percentages.
The subject device is substantially equivalent to the predicate as well as the reference device as the overall intended use and nature of the software remains the same. The difference with the subject device is the addition of the bone level measurement feature.
# 6. Device Description
Second Opinion BLE is a radiological automated image processing software device intended to identify and display bone level measurements in bitewing and periapical radiographs. It should not be used in lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis.
It is designed to aid dental health professionals to review bitewing and periapical radiographs of permanent teeth in patients 12 years of age or older as a concurrent and second reader.
Second Opinion BLE consists of three parts:
- Application Programming Interface ("API")
- Machine Learning Modules ("ML Modules")
- Client User Interface (UI) ("Client")
The processing sequence for an image is as follows:
1. Images are sent for processing via the API
2. The API routes images to the ML modules
3. The ML modules produce detection output
4. The UI renders the detection output
The API serves as a conduit for passing imagery and metadata between the user interface and the machine learning modules. The API sends imagery to the machine learning modules for processing and subsequently receives metadata generated by the machine learning modules which is passed to the interface for rendering.
Second Opinion BLE uses machine learning to detect bone level measurements. Images received by the ML modules are processed yielding detections which are represented as metadata. The final output is made accessible to the API for the purpose of sending to the UI for visualization. Detected bone level measurements are displayed as linear overlays
Hellopearl.com
{6}
Karl
K243230
atop the original radiograph which indicate to the practitioner which regions contain which detected potential conditions that may require clinical review. The clinician can toggle over the image to highlight a potential condition for viewing.
## 7. Indications for Use
Second Opinion® BLE is a radiological automated image processing software device intended to identify and display bone level measurements in bitewing and periapical radiographs. It should not be used in lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis.
It is designed to aid dental health professionals to review bitewing and periapical radiographs of permanent teeth in patients 12 years of age or older as a concurrent and second reader.
## 8. Summary of Substantial Equivalence:
The predicate device as well as the reference device and subject device are similar devices in the following ways:
1) Intended use: All devices are intended to be used to aid dental clinicians in their review of bone level measurements in bitewing and periapical radiographs of permanent teeth.
2) Technology characteristics: All devices employ computer vision and machine learning to output detections, use cloud-based environments to conduct processing, and display bone level measurements within a user interface with a graphical overlay over radiographs.
3) Safety: As both the candidate and predicate devices are software systems, neither pose a direct safety hazard to the patient. The primary hazards for both devices, subject and predicate, are potential inaccurate or misplaced bone lines/measurements which could result in a temporary non-serious injury. In the case of each device, users are not meant to rely solely on detection output for clinical decision-making.
4) Clinical Performance: All devices have undergone clinical studies which demonstrate statistically significant improvement in aided reader performance.
Table 1: Comparison of Second Opinion BLE with the predicate / reference devices.
Hellopearl.com
{7}
Pearl
K243230
| | Subject Device
Second Opinion®
BLE | Primary Predicate
Overjet Dental
Assist
K210187 | Reference Device
Videa Perio Assist
K223296 |
| --- | --- | --- | --- |
| Manufacturer | Pearl Inc. | Overjet, Inc. | VideaHealth, Inc. |
| Classification | 892.2050 | 892.2050 | 892.2050 |
| Product Code | QIH | LLZ | QIH |
| Class | Class II | Class II | Class II |
| Image Modality | Radiograph | Radiograph | Radiograph |
| Intended Use | Dental automated image processing software device to aid in bone level measurement dental radiograph review by HCP | Dental automated image processing software device to aid in bone level measurement dental radiograph review by HCP | Dental automated image processing software device to aid in bone level measurement dental radiograph review by HCP |
| Full IFU | Second Opinion® BLE is a radiological automated image processing software device intended to identify and display bone level measurements in bitewing and periapical radiographs. It should not be used in lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis.
It is designed to aid dental health professionals to review bitewing and periapical radiographs of permanent teeth in patients 12 years of age or older as a concurrent and second reader. | Overjet Dental Assist is a radiological semi-automated image processing software device intended to aid dental professionals in the measurements of mesial and distal bone levels associated with each tooth from bitewing and periapical radiographs. It should not be used in lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis. The system is to be used by trained professionals including, but not limited to, dentists and dental hygienists. | Videa Perio Assist is a radiological semi-automated image processing software device intended to aid dental professionals in the measurements and visualization of mesial and distal bone levels associated with each tooth from bitewing and periapical radiographs.
Measurements are made available as linear distances or relative percentages. It should not be used in-lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis. The system is to be used by trained professionals including, but not limited to, dentists and dental hygienists. |
| Intended body part | Dental | Dental | Dental |
Hellopearl.com
{8}
Pearl
K243230
| Technology | Utilizes computer vision neural network algorithms, developed from open-source models using supervised machine learning techniques | Utilizes computer vision neural network algorithms, developed from open-source models using supervised machine learning techniques | Utilizes computer vision neural network algorithms, developed from open-source models using supervised machine learning techniques |
| --- | --- | --- | --- |
| Device Description | Detection & display of linear bone level measurements | Detection & display of linear bone level measurements | Detection & display of linear bone level measurements |
# 9. Technological Comparison to Predicate and Reference Devices
The fundamental technological principle for both the candidate and predicate as well as reference device is the automatic detection and display of bone level measurements using machine learning.
The candidate and predicate / reference devices are technologically equivalent as follows:
- All are software devices designed to run on Windows operating systems.
- All devices are designed to process digital intraoral bitewing and periapical radiographs.
- All devices use neural network-based computer vision algorithms for bone level measurement display.
- All devices display measurements within the user interface with a graphical overlay on the radiograph.
- All devices produce near-instantaneous detection results.
- All devices are considered to be of "moderate" level of software concern.
- All devices passed all verification and validation testing requirements.
# 10. Assessment of Benefit-Risk, Safety and Effectiveness, and Substantial Equivalence to Predicate Device
Pearl demonstrated the benefits of the device through a standalone clinical study. The results of the study showed the system is both safe and effective for its intended use. When the probable benefits and probable risks of Second Opinion BLE are weighed against one another, the weight of benefits significantly exceeds that of risks. This judgment can be made based on review of the submitted materials showing that Second Opinion BLE meets the design verification and validation requirements. It is thus concluded that Second Opinion BLE can be considered safe and effective such that the device will aid users in the indicated user population in their radiographic review of bone heights.
Hellopearl.com
{9}
Karl
K243230
# 11. Discussion of Non-Clinical Tests Performed
The device is a software-only device, so most testable characteristics common to other device types, including Biocompatibility/Materials, Shelf Life/Sterility, Electromagnetic Compatibility and Electrical Safety, Magnetic Resonance (MR) Compatibility, are not applicable to this device.
# 12. Discussion of Clinical Tests Performed
Clinical evaluation of Second Opinion BLE was performed to clinically validate the inclusion of a new bone level measurement tool in Second Opinion. The software modification is intended to assist dentists with their manual comparison of radiographs over time by providing automatic estimated bone level measurements in the bitewing and periapical radiographs. The effectiveness of Second Opinion BLE was gauged in a standalone performance assessment to validate the clinical usefulness of the added bone level measurement tool in Second Opinion BLE in intraoral bitewing and periapical radiographs of the permanent teeth of patients 12 years of age or older.
Clinical testing of the Second Opinion BLE software was performed utilizing retrospective clinical subject data from 229 bitewing and 167 periapical radiographs from 396 subjects. The demographics of the data set was as follows:
| Subject Age | Count (%) |
| --- | --- |
| <30 | 63 (15.9) |
| 30-<40 | 54 (13.6) |
| 40-<50 | 79 (19.9) |
| 50-<60 | 71 (17.9) |
| 60-<70 | 73 (18.4) |
| 70+ | 56 (14.1) |
| Gender | Count (%) |
| --- | --- |
| Female | 160 (40.4) |
| Male | 190 (48.0) |
| Other | 44 (11.1) |
Hellopearl.com
{10}
Pearl
K243230
| Unknown | 2 (0.5) |
| --- | --- |
No information about ethnicity was available from the dental radiographs
Second Opinion BLE utilized three US licensed dentists to label interproximal bone levels. The three readers provided measurements per CEJ-bonecrest, measurement differences that were clinically significant required an adjudication. These divergent measurements were then adjudicated by two U.S. Dental Radiologists.
Second Opinion BLE demonstrated an 87% (95% CI: 85%, 89%) precision value and 87% (95% CI: 85%, 89%) recall value for measurement of interproximal bone levels in periapical images and a 87% (95% CI: 86%, 88%) precision and 91% (95% CI: 90%, 92%) recall in bitewing images. Second Opinion BLE met the prespecified acceptance criteria of > 82% for both precision & recall for bitewing and periapical images.
| Metric | Bitewing | Periapical |
| --- | --- | --- |
| Precision | 87% | 87% |
| Recall | 91% | 87% |
The mean interproximal bone level measurement of GT Second Opinion BLE and the mean absolute difference in millimeters (mm) by image type and overall are shown in the table below. The modeled mean absolute difference (95% CI) was 0.45 mm (0.37, 0.52) and 0.86 mm (0.83, 0.89) in periapical and bitewing images respectively. Second Opinion BLE met the pre-specified acceptance criteria of mean absolute difference < 1.5 mm in both bitewing and periapical images. In addition, the mean absolute difference remained statistically significant after the multiple comparisons adjustment as shown by all p-values < 0.05.
Based on the performance data and the application of general controls and special controls established for this device as published in the subject labeling, use of this device is unlikely to increase the rate of false negative or false positive diagnosis of dental pathologies as compared with the current clinical standard of practice. Further, possible misuse of the device does not present additional risks compared with the misuse of other types of radiological image processing devices.
Hellopearl.com
{11}
Pearl
K243230
# 13. Conclusions
Based on the information presented above, Second Opinion BLE and its primary predicate device, Overjet Dental Assist, are deemed to have similar intended uses as devices which aid in the detection & review of bone heights in dental radiographs. Second Opinion BLE’s clinical trial results demonstrate that the device effectively performs as well as the primary predicate device.
Second Opinion BLE raises no new or different questions of safety or effectiveness, performs in accordance with its specifications, meets user needs, meets the intended use and therefore was found substantially equivalent to the predicate devices.
Hellopearl.com
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.