Peer-reviewed clinical literature (van Westrhenen et al., 2023); In-home patient monitoring data; Caregiver seizure diaries; Video recordings of nocturnal events
The sponsor utilized a phase 4, multicenter, in-home clinical study to demonstrate the device's sensitivity and false alarm rate for detecting nocturnal epileptic major motor seizures in a pediatric population (ages 4-16) in their natural home environment.
In-home monitoring; Pediatric epilepsy; Nocturnal seizure detection; Phase 4 study
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Multimodal nocturnal seizure detection in children with epilepsy: A prospective, multicenter, long-term, in-home trial (van Westrhenen et al., 2023); Phase 4, multicenter, prospective, video-controlled, in-home study; Follow-up/Duration: 2310 nights (28,173 hours) of data
53 children aged 4-16 years with epilepsy living at home; Sample Size: 53; Number of Sites: Multicenter
Not applicable for this study
Sensitivity for nocturnal epileptic major motor seizures; False Alarm Rate (FAR)
van Westrhenen et al. 2023 (Clinical Performance Study): 53 children aged 4-16 years
>1 (trial nurses)
Indications for Use
The NightWatch+ US is a prescription only device that is indicated for use as an adjunct to seizure monitoring of children age 4 till 16 diagnosed with epilepsy having Nocturnal Epileptic Major Motor Seizures which includes tonic-clonic (TC), tonic (if clustered or prolonged >30 seconds), hyperkinetic and TC-like seizures, in home or residential facilities during periods of rest. The Sensor of the device is worn on the upper arm and measures heart rate and motion data to detect patterns that may be associated with nocturnal epileptic motor seizures in patients with epilepsy. When a seizure event is detected by the Sensor of the NightWatch+ US, it sends a command to the paired wireless alarm station of the NightWatch+ US that is programmed to initiate an alarm to a designated caregiver. The system records and stores data from seizure events. The data can be viewed by the user in a cloud based data portal. The NightWatch+ US is not intended to diagnose specific seizure types.
Device Story
Device consists of upper-arm wearable sensor and wireless alarm station. Sensor uses PPG for heart rate and accelerometer for motion data. Microprocessor runs detection algorithm to identify patterns associated with nocturnal epileptic major motor seizures. Upon detection, sensor sends command to alarm station, which triggers audible alarm and blinking LED for caregiver notification. Data is stored and transmitted via ethernet to cloud-based portal for review. Used in home or residential settings during rest. Assists caregivers in monitoring seizure frequency; does not diagnose specific seizure types.
Clinical Evidence
Prospective, multicenter, video-controlled in-home study (van Westrhenen et al. 2023) of 53 children (ages 4-16). Analyzed 2,310 nights (28,173 hours) and 552 seizures. Primary endpoint: sensitivity for nocturnal epileptic major motor seizures. Results: 89% overall sensitivity (94% for TC, 53% for tonic, 83% for hypermotor, 91% for TC-like). False alarm rate (FAR) was 0.06/h. Comparator: predicate device performance data.
Technological Characteristics
Wearable sensor with PPG heart rate and accelerometer. Communicates wirelessly to alarm station. Battery-powered (rechargeable Li-ion). Compliant with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), and ISO 10993 (biocompatibility). Data connectivity via ethernet to cloud portal. No ionizing radiation. Not sterilized.
Indications for Use
Indicated for children age 4-16 diagnosed with epilepsy having Nocturnal Epileptic Major Motor Seizures (tonic-clonic, tonic, hyperkinetic, TC-like) in home or residential facilities during rest. Prescription only.
Regulatory Classification
Identification
A non-electroencephalogram (non-EEG) physiological signal based seizure monitoring system is a noninvasive prescription device that collects physiological signals other than EEG to identify physiological signals that may be associated with a seizure.
Special Controls
In combination with the general controls of the FD&C Act, the Non-EEG physiological signal based seizure monitoring system is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) The technical parameters of the device, hardware and software, must be fully characterized and include the following information:
(i) Hardware specifications must be provided. Appropriate verification, validation, and hazard analysis must be performed.
(ii) Software, including any proprietary algorithm(s) used by the device to achieve its intended use, must be described in detail in the Software Requirements Specification (SRS) and Software Design Specification (SDS). Appropriate software verification, validation, and hazard analysis must be performed.
(2) The patient-contacting components of the device must be demonstrated to be biocompatible.
(3) The device must be designed and tested for electrical, thermal, and mechanical safety and electromagnetic compatibility (EMC).
(4) Clinical performance testing must demonstrate the ability of the device to function as an assessment aid for monitoring for seizure-related activity in the intended population and for the intended use setting. Performance measurements must include positive percent agreement and false alarm rate.
(5) Training must be provided for intended users that includes information regarding the proper use of the device and factors that may affect the collection of the physiologic data.
(6) The labeling must include health care professional labeling and patient-caregiver labeling. The health care professional and the patient-caregiver labeling must include the following information:
(i) A detailed summary of the clinical performance testing, including any adverse events and complications.
(ii) Any instructions technicians and clinicians should convey to patients and caregivers regarding the proper use of the device and factors that may affect the collection of the physiologic data.
(iii) Instructions to technicians and clinicians regarding how to set the device threshold to achieve the intended performance of the device.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 12, 2025
LivAssured BV
Jolanda Oorthuizen
Quality Assurance and Regulatory Affairs Manager
Schipholweg 103
2316 XC
Leiden, Netherlands
Re: K243199
Trade/Device Name: NightWatch+ US
Regulation Number: 21 CFR 882.1580
Regulation Name: Non-Electroencephalogram (EEG) Physiological Signal Based Seizure Monitoring System
Regulatory Class: Class II
Product Code: POS
Dated: June 11, 2025
Received: June 11, 2025
Dear Jolanda Oorthuizen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243199 - Jolanda Oorthuizen
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K243199 - Jolanda Oorthuizen
Page 3
Sincerely,
Jay R. Gupta -S
Jay Gupta
Assistant Director
DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243199
Device Name
NightWatch+ US
Indications for Use (Describe)
The NightWatch+ US is a prescription only device that is indicated for use as an adjunct to seizure monitoring of children age 4 till 16 diagnosed with epilepsy having Nocturnal Epileptic Major Motor Seizures which includes tonic-clonic (TC), tonic (if clustered or prolonged >30 seconds), hyperkinetic and TC-like seizures, in home or residential facilities during periods of rest. The Sensor of the device is worn on the upper arm and measures heart rate and motion data to detect patterns that may be associated with nocturnal epileptic motor seizures in patients with epilepsy. When a seizure event is detected by the Sensor of the NightWatch+ US, it sends a command to the paired wireless alarm station of the NightWatch+ US that is programmed to initiate an alarm to a designated caregiver. The system records and stores data from seizure events. The data can be viewed by the user in a cloud based data portal. The NightWatch+ US is not intended to diagnose specific seizure types.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”
{4}
510(k) Summary
NightWatch+ US


LiVAssured
{5}
LivAssured
NightWatch
# 510(K) SUMMARY
The content of this 510(k) summary is provided in conformance with 21 CFR § 807.92.
This summary was prepared on the 10th of July 2025.
## Submitter's Information
Company Name LivAssured BV
Company Address Schipholweg 103
2316 XC, Leiden
The Netherlands
Contact Person Jolanda Oorthuizen
Telephone +3185 0601252
Email Jolanda@nightwatch.nl
## Subject Medical Device
Device Name NightWatch+ US
Regulation Name Non-electroencephalogram (EEG) physiological signal-based seizure monitoring system
Regulation Number 21 CFR 882.1580
Device Class Class II
Product Code POS
Product Code Name Physiological Signal-Based Seizure Monitoring System
## Legally Marketed Predicate Device
510(k) Number 181861
Device Name Embrace
Regulation Name Non-electroencephalogram (EEG) physiological signal-based seizure monitoring system
Regulation Number 21 CFR 882.1580
Device Class Class II
Product Code POS
Product Code Name Physiological Signal-Based Seizure Monitoring System
Page 2 of 12
{6}
LivAssured
NightWatch
# Device Description
NightWatch+ US consists of a sensor worn during sleep on the biceps of the upper arm and an alarm station. The sensor consists of a heart rate sensor using PPG (photoplethysmography), a ACC (Accelerometry) movement sensor, and a microprocessor. The microprocessor processes the data from the sensors using a detection algorithm which detects if the sensor readings match pre-programmed parameters that are associated with nocturnal epileptic major motor seizures. When a nocturnal epileptic major motor seizures is detected, the seizure alarm is triggered and transferred from the sensor to the accompanying alarm station which alarms the caregiver by a sound and a blinking LED light. The seizure alarms data can be transferred from the alarm station, using an ethernet connection, to a database in the cloud to be viewed a web-based interface called NightWatch Portal to be able to monitor seizure frequency overtime.



# Indication for use
The NightWatch+ US is a prescription only device that is indicated for use as an adjunct to seizure monitoring of children age 4 till 16 diagnosed with epilepsy having Nocturnal Epileptic Major Motor Seizures which includes tonic-clonic (TC), tonic (if clustered or prolonged >30 seconds), hyperkinetic and TC-like seizures, in home or residential facilities during periods of rest. The Sensor of the device is worn on the upper arm and measures heart rate and motion data to detect patterns that may be associated with nocturnal epileptic motor seizures in patients with epilepsy. When a seizure event is detected by the Sensor of the NightWatch+ US, it sends a command to the paired wireless alarm station of the NightWatch+ US that is programmed to initiate an alarm to a designated caregiver. The system records and stores data from seizure events. The data can be viewed by the user in a cloud based data portal. The NightWatch+ US is not intended to diagnose specific seizure types.
# Substantial Equivalence
A comparison of NightWatch+ US to the predicate device is provided in the table below
#
Page 3 of 12
{7}
LivAssured
NightWatch
| Attribute | NightWatch+ US
(subject device) | Embrace
KIB1861 (predicate device) | Comparison |
| --- | --- | --- | --- |
| Intended Use | Non-EEG physiological signal based seizure monitoring system. | Non-EEG physiological signal based seizure monitoring system. | EQUIVALENT
Both devices have the same intended use. |
| Indications for Use | The NightWatch+ US is a prescription only device that is indicated for use as an adjunct to seizure monitoring of children age 4 till 16 diagnosed with epilepsy having Nocturnal Epileptic Major Motor Seizures which includes tonic-clonic (TC), tonic (if clustered or prolonged >30 seconds), hyperkinetic and TC-like seizures, in home or residential facilities during periods of rest. The Sensor of the device is worn on the upper arm and measures hearth rate and motion data to detect patterns that may be associated with nocturnal epileptic motor seizures in patients with epilepsy. When a seizure event is detected by the Sensor of the NightWatch+ US, it sends a command to the paired wireless alarm station of the NightWatch+ US that is programmed to initiate an alarm to a designated caregiver. The system records and stores data from seizure events. The data can be viewed by the user in a cloud based data portal. The NightWatch+ US is not intended to diagnose specific seizure types. | The Embrace is a prescription only device that is indicated for use as an adjunct to seizure monitoring of adults and children age 6 and up in home or healthcare facilities during periods of rest. The device is worn on the wrist and senses Electrodermal Activity (EDA) and motion data to detect patterns that may be associated with generalized tonic-clonic seizures in patients with epilepsy or at risk of having epilepsy. When a seizure event is detected, Embrace sends a command to a paired wireless device that is programmed to initiate an alert to a designated caregiver. The System records and stores data from Accelerometer, EDA, and Temperature sensors for subsequent review by a trained healthcare professional. | Patient Population:
EQUIVALENT
The devices are indicated for use on patients with epilepsy. Both NightWatch+ US and Embrace are indicated for pediatric population. NightWatch+ US is indicated for use in children age 4 till 16 years and Embrace is indicated for children 6 years and up. The difference in age groups does not raise any new safety or performance concerns.
Healthcare Environment:
EQUIVALENT
Both the subject and predicate devices are indicated for use in the home. Embrace is also indicated for use in healthcare settings. The more limited healthcare environment for the subject device does not raise any new safety or performance concerns.
Way of notification caregiver:
EQUIVALENT
Both do have a paired wireless device that is programmed to initiate an alert to a designated caregiver. In case of Nightwatch+ US there is a separate alarm station, in case of Embrace an app is used. This difference does not raise any new safety or performance concerns. as both devices are compliant to the FCC requirements of CFR no. 47, Part 15
Seizure Type:
EQUIVALENT
Both devices are indicated to provide an alert to caregivers when they each detect Tonic-Clonic seizures. Additionally NightWatch+ US is intended to alert for tonic (if cluster or prolonged >30 seconds), |
Page 4 of 12
{8}
LivAssured
NightWatch
| Attribute | NightWatch+ US
(subject device) | Embrace
K181861 (predicate device) | Comparison |
| --- | --- | --- | --- |
| | | | hyperkinetic and TC-like seizures. This does not raise any new safety or performance concerns as NightWatch+ was extensively validated in a phase 4 peer reviewed scientific clinical study for these seizure types. Both devices monitor physiological signals while the patient is at rest. |
| | | | Over The Counter or RX
EQUIVALENT
The subject device and the predicate are both prescription devices |
| Classification | Class II 21 CFR 882.1580 Non-EEG physiological signal-based seizure monitoring system | Class II 21 CFR 882.1580 Non-EEG physiological signal-based seizure monitoring system | EQUIVALENT |
| Product Code | POS | POS | EQUIVALENT |
| Sensor application | The Sensor of the NightWatch+ US is worn on the upper arm with an elastic strap | The embrace watch is worn on the wrist with an elastic strap. | EQUIVALENT
Both devices are worn using an elastic strap on the upper limb. |
| Materials | | | |
| Materials and Biocompatibility | All patient contacting parts are tested to applicable tests in EN ISO 10993-1:2018 (Cytotoxicity – ISO 10993-5:2009, Sensitization – ISO 10993- 10:2021, and Skin Irritation – ISO 10993-23:2021). | All patient contacting parts are tested to applicable tests in ISO 10993 (Cytotoxicity – ISO 10993-5, Sensitization – ISO 10993-10, and Skin Irritation – ISO 10993-10). | EQUIVALENT
Both devices have patient contacting parts that come into contact with intact skin for a duration that may exceed 24 hours. Both devices meet the requirements of ISO 10993. See chapter 4.3 non clinical testing. |
| Technical Specification | | | |
| Sensor technology | Utilizes an Photoplethysmogram (PPG) sensor to acquire Heart rate and an accelerometer sensor to acquire movement data. | Utilizes an electrodermal sensor to acquire Electrodermal Activity, an accelerometer sensor to acquire movement data and a temperature sensor. | EQUIVALENT
The difference in type of sensors does not raise different questions of safety and effectiveness. PPG sensors and accelerometer sensors are widely used in many kinds of consumer technology and are concerned generally safe. |
| Data communication | The NightWatch+ US Sensor communicates wirelessly to the NightWatch+ US Alarm station which alerts the | The Embrace device communicates wirelessly to a smartphone | EQUIVALENT
Both systems utilize wireless communication between the sensor and a alarm station or |
Page 5 of 12
{9}
LivAssured
NightWatch
| Attribute | NightWatch+ US
(subject device) | Embrace
KIB1861 (predicate device) | Comparison |
| --- | --- | --- | --- |
| | healthcare provider or caregiver with an audible alarm and lights
The NightWatch+ US is compliant to FCC Rules: Code of Federal Regulations (CFR) no. 47, Part 15 | application, which alerts the healthcare provider or caregiver in one or more ways (phone call, text message, etc.).
The Embrace is compliant to FCC Rules: Code of Federal Regulations (CFR) no. 47, Part 15 | app, and use technologies to alert the caregiver like audible alarms.
Both devices are compliant to the FCC requirements of CFR no. 47, Part 15. See chapter 4.3 non-clinical performance testing. |
| Data storage | Data of the device and seizure frequency can be stored on a cloud based platform for later viewing. | Data of the device and seizure frequency is stored on a cloud based platform for later viewing. | EQUIVALENT
For both devices data of the device and seizure frequency can be stored on a cloud based platform for later viewing. |
| Energy Source Wearable | Battery | Battery | EQUIVALENT
The Energy Source of the Wearable for both devices is a battery |
| Battery Type | Rechargeable Lithium-Ion | Rechargeable Lithium-Ion | EQUIVALENT
For both devices the battery in the wearable is a Rechargeable Lithium-Ion |
| Algorithm | Uses algorithms to analyze PPG and accelerometer data to detect patterns in the data that may be associated with tonic-clonic, tonic (if cluster or prolonged), myoclonic (if clustered) or hyperkinetic seizures | Uses algorithms to analyze EDA and accelerometer data to detect patterns in the data that may be associated with generalized tonic-clonic seizures | EQUIVALENT
The difference in the algorithms do not raise different questions of safety and effectiveness. The same questions of safety and effectiveness regarding performance of the algorithms (Sensitivity and False Alarm Rate) apply to all seizure detection devices. |
| Electrical Safety | Electrical Safety testing performed to IEC 60601-1
Electromagnetic compatibility testing performed to IEC 60601-1-2 | Electrical Safety testing performed to IEC 60601-1
Electromagnetic compatibility testing performed to IEC 60601-1-2 | EQUIVALENT
Both devices underwent Electrical Safety testing performed to IEC 60601-1 and Electromagnetic compatibility testing performed to IEC 60601-1-2 |
| Thermal Safety | Not applicable. The device does not generate any localized heat | Not applicable. The device does not generate any localized heat | EQUIVALENT
Not applicable. Both devices do not generate any localized heat |
| Radiation Safety | Not applicable. Device does not use any ionizing radiation | Not applicable. Device does not use any ionizing radiation | EQUIVALENT
Not applicable. Both devices do not use any ionizing radiation |
| Sterility | Not applicable. Device is not | Not applicable. Device is | EQUIVALENT |
Page 6 of 12
{10}
LivAssured
NightWatch
| Attribute | NightWatch+ US
(subject device) | Embrace
K181861 (predicate device) | Comparison |
| --- | --- | --- | --- |
| | sterilized | not sterilized | Not applicable. Both devices are not sterilized. |
| Packaging method | No special packaging requirements | No special packaging requirements | EQUIVALENT
No special packaging requirements |
Table 1: Substantial equivalence comparison
## Performance testing
## Non-Clinical testing
The following non-clinical testing was conducted which support a determination of substantial equivalence to the predicate. The non-clinical tests included:
| Standards | Description | Certificate or laboratory test report | FDA recognition number |
| --- | --- | --- | --- |
| IEC 60601-1:2005+A1:2012+A2:2020 | Medical electrical equipment - Part 1: General requirements for basic safety and essential performance | TD-NWP-85: Certificate SIQ SI-10555 | 19-49 |
| IEC 60601-1-2:2014+A1:2020 | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests | TD-NWP-81: Certificate SIQ SI-10670
TD-NWP-82: Test Report: T251-0707/23 | 19-36 |
| ETSI EN 301 489-1 V2.2.3:2019 | ElectroMagnetic Compatibility (EMC) standard for radio equipment and services; Part 1: Common technical requirements | TD-NWP-82: Test Report: T251-0707/23 | - |
| ETSI EN 301 489-6 V2.2.1:2019 | ElectroMagnetic Compatibility (EMC) Standard For Radio Equipment And Services; Part 6: Specific Conditions For Digital Enhanced Cordless Telecommunications (DECT) Equipment | TD-NWP-82: Test Report: T251-0707/23 | - |
| IEC 60601-1-6:2010+A1:2013+A2:2020 | Medical electrical equipment - Part 1-6: General requirements for basic safety and essential | TD-NWP-85: Certificate SIQ SI-10555 | 5-132 |
Page 7 of 12
{11}
LivAssured
NightWatch
| Standards | Description | Certificate or laboratory test report | FDA recognition number |
| --- | --- | --- | --- |
| | performance- Collateral standard: Usability | | |
| IEC 60601-1-8:2006+A1:2012+A2:2020 | Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems | TD-NWP-85: Certificate SIQ SI-10555 | 5-131 |
| IEC 60601-1-11:2015+A1:2020 | Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment | TD-NWP-85: Certificate SIQ SI-10555 | 19-38 |
| 47 CFR 1.1310 | Radiofrequency radiation exposure limits. | TD-NWUS-23: Test Report No: 1-6175-23-02-04_TR1-R01 | - |
| 47 CFR Part 15, Subpart B | Equipment authorization of unintentional radiators. | TD-NWUS-12: Certificate SIQ C251-0059/24 | - |
| 47 CFR Part 15, Subpart D | Unlicensed Personal Communications Service Devices | TD-NWUS-19: FCC ID: 2BGSP-NWP1
TD-NWUS-22: FCC ID: Y82-SC14S | - |
| ETSI EN 301 406 V2.2.2 (2016-09) | Digital Enhanced Cordless Telecommunications (DECT); Harmonised Standard for access to radio spectrum; Part 1: DECT, DECT Evolution and DECT ULE | TD-NWP-87: Test report: CETECOM nr: 1-6175_23-01-06 & CETECOM nr: 1-6175_23-01-07 | - |
| ISO 10993-1:2018 | Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process | TD-NWUS-10: 24-74-1 Biological Risk Assessment of NightWatch + US according to EN ISO 10993-1:2018 and FDA Guidance document "Use of | 2-258 |
Page 8 of 12
{12}
LivAssured
NightWatch
| Standards | Description | Certificate or laboratory test report | FDA recognition number |
| --- | --- | --- | --- |
| | | International Standard ISO 10993-1, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk Management Process" dated 8. September 2023 | |
| ISO 10993- 10:2021 | Biological evaluation of medical devices- Part 10: Tests for skin sensitization | TD-NWP-103: 23-109-2-SP.01 Sensitization | 2-296 |
| ISO 10993-5:2009 | Biological evaluation of medical devices- Part 5: Tests for in vitro cytotoxicity | TD-NWP-101: Test Report 23-54-1-SP.01 Cytotoxicity | 2-245 |
| ISO 10993-23:2021 | Biological evaluation of medical devices- Part 23: Tests for irritation | TD-NWP-104: Test Report: 23-109-3-SP.01 Irritation | 2-291 |
Table 2: Non-clinical testing of the subject device NightWatch+ US
## Clinical testing
NightWatch+ US is a US-targeted version of a NightWatch+, which is a class IIa medical device under the Medical Device Regulation 2017/745 in Europe. Its predecessor NightWatch has been on the European market since 2018. The seizure detection method, algorithms and components used for seizure detection are identical in both NightWatch, NightWatch+ and NightWatch+ US.
The clinical study used for the clearance of the NightWatch+ US for the indicated user group of Children age 4-16 years old is the study of van Westrhenen et al. 2023, a phase 4, multicenter, prospective, video-controlled, in-home study.
1. van Westrhenen, R. Lazeron, J. van Dijk, F. S. S. Leijten and R. D. Thijs, "Multimodal nocturnal seizure detection in children with epilepsy: A prospective, multicenter, long-term, in-home trial," Epilepsia, vol. 64, pp. 2137-2152, 2023.
In this study 53 children aged 4-16 years were included with $\geq 1$ weekly nocturnal epileptic major motor seizure, living at home. Performance for nocturnal epileptic major motor seizures was determined. The study sample size was determined based on an $80\%$ sensitivity estimate, and performance goals or acceptance criteria for other endpoints were not defined.
Page 9 of 12
{13}
LivAssured
NightWatch
Demographics:
| N=53 | |
| --- | --- |
| Sex | |
| Male, n (%) | 29 (55%) |
| Female, n (%) | 24 (45%) |
| Age, years, Mean (range) | 9.7±3.6 (4-16) |
| Learning disability | |
| Yes, n (%) | 36(68%) |
| No, n (%) | 17(32%) |
| Epilepsy etiology | |
| Structural, n (%) | 13 (25%) |
| Genetic, n (%) | 20 (38%) |
| infectious, n (%) | 1 (1%) |
| Metabolic, n (%) | 0 |
| Immune, n (%) | 0 |
| Unknown, n (%) | 19(36%) |
Results: 2310 nights (28173 h), including 552 nocturnal epileptic major motor seizures were analyzed.
Table 3: Demographics of participants in the van Westrhenen et al. 2023 study
| | Overall Nocturnal Epileptic Major motor | Tonic Clonic (TC) | Tonic (>30 sec) | Hypermotor | Other major (TC-like seizures) |
| --- | --- | --- | --- | --- | --- |
| nr of patients | 51 | 20 | 11 | 7 | 23 |
| nr of seizures | 552 | 204 | 30 | 48 | 270 |
| nr of seizures detected | 492 | 191 | 16 | 40 | 245 |
| PPA point estimate mean sensitivity | 90% | 98% | 71% | 58% | 87% |
| PPA 95% CI lower | 84% | 94% | 43% | 17% | 75% |
| PPA 95% CI upper | 96% | 100% | 100% | 99% | 100% |
| Overall sensitivity | 89% | 94% | 53% | 83% | 91% |
Table 4: Performance metrics, Sensitivity, of the van Westrhenen et al. 2023 study
| False alarms rate (FAR) total | |
| --- | --- |
| Total false alarms | 1642 |
| FAR/h mean | 0,07 |
| FAR/h 95% CI Lower | 0,04 |
| FAR/h 95% CI upper | 0,10 |
| Overall FAR/h | 0,06 |
Table 5: Performance metrics, FAR, of the van Westrhenen et al. 2023 study
In the following table a summary of the study results can be found and equivalence to the predicate device study results is discussed:
Page 10 of 12
{14}
LivAssured
NightWatch
| Reference Standard | Video | Video-EEG | Equivalence? |
| --- | --- | --- | --- |
| | | | EQUIVALENT, Video is an equally robust reference standards to Video-EEG provided the whole dataset is reviewed for inferring the reference standard as recognized by the International League Against Epilepsy (ILAE)^{2}
Trained trial nurses annotated all events (generated NightWatch alarms, video alarms, and caregivers' seizure diary) using the video recordings while blinded for alarm type and NightWatch sensor data (HR and movement). In case of discrepancies (when the recorded night was annotated by one nurse, but screened by another) or doubt, the trial nurses consulted one of the principal investigators (R.D.T., R.H.C.L.) for a final decision. The principal investigators (R.D.T., R.H.C.L.) double checked a random sample of 5% of the annotations
Additionally, trial nurses retrospectively analyzed video tracings with a previously validated automated video- based seizure detection algorithm. Trained trial nurses also fully screened the video of 5% of all nights for missed seizures; every video was screened by one nurse. In total, 1402 h (5%) of all recorded nights were randomly screened, ranging from half a night to four full nights per participant. |
| Data analysis method.
Patient-level analysis versus event-level analysis | Event-level analysis and Patient-level analysis | Event-level analysis | EQUIVALENT, Van Westrhenen et al. 2023 used patient-level analysis because it provides a more accurate reflection of device performance across individuals, rather than being skewed by outliers with many events. This approach is valid and meaningful due to the high number of events per participant, enabled by the prolonged in-home monitoring period. In contrast, the predicate device study included relatively few events among a small group of patients, making a robust patient-level analysis less feasible.
More-over, event-level analysis can distort overall performance metrics, especially when a large proportion of events originates from a small subset |
Page 11 of 12
{15}
LivAssured
NightWatch
| | Westrhnen et al. 20231 | K181861 Embrace clearance by FDA for Children 6 years and up | Equivalence? |
| --- | --- | --- | --- |
| | | | of participants. Nevertheless, both approaches were included in the publication of van Westrhnen et al. 2023. The difference in analysis method does not raise new concerns for performance or safety compared to the predicate device |
| Number of included children | 53 children (age 4-16 years) | 80 children (ages 6- 21 years) | EQUIVALENT, The difference in age group does not raise new concerns for performance or safety compared to the predicate device. Both studies include enough participants |
| Number of seizures detected Total for children | 33 children experienced in total 552 nocturnal epileptic major motor seizures of which 204 TC seizures | 17 children experienced in total 32 TC seizures | EQUIVALENT, No new concerns for performance or safety compared to the predicate device. Westrhnen et al. 2023 included more seizures than the study of the predicate device. |
| Number of hours data recorded | All 53 patients provided overall 2310 nights and 28173 hours of data | All 141 patients provided overall 9806 hours, with a median of 49.2 hours of data per patient. It is not clear how many of those hours recorded were of adults or children. | EQUIVALENT, No new concerns for performance or safety compared to the predicate device. Van Westrhnen et al. 2023 included significantly more hours of data than the study of the predicate device |
| Sensitivity Tonic-Clonic seizures Children | Overall 94% Mean 98% (95% CI: 94%- 100%) | Overall 96,9% Corrected Overall 91,5% (95% CI: 83%-95%) | EQUIVALENT, Both devices have a good sensitivity of >90% and equal false alarm rate. No new concerns for performance or safety compared to the predicate device. |
| Sensitivity Nocturnal epileptic major motor seizures | Overall 89% Mean 90% (95% CI: 84%-95%) | Not applicable | |
| False Alarm rate Children | Overall: 0,06/h Mean: 0,07/h (95% CI = 0.04-0.10/h) | Overall: 0,06/h (=1,35 per day/24h) Mean: 0,07/h= 1.63 per day/24h (95% CI: 0.04-0.10/h)= (95% CI: 1.06-2.28 per day/24h) | |
Table 6: Summary of the study results of van Westrhnen et al 2023. and equivalence discussion to the predicate device study results.
2. S. Beniczky, & P. Ryvlin., Standards for testing and clinical validation of seizure detection devices. Epilepsia, 59(S1), 9-13. 2018
# Conclusion
The clinical study of van Westrhnen et al. 2023 demonstrates the ability of the NightWatch+ US to function as an assessment aid for monitoring of seizure activity in the intended population of children 4-16 years old and for the intended use setting.
Based on the above discussion and enclosed sections regarding substantial equivalence to the predicate device, we conclude that NightWatch+ US is substantially equivalent to its predicate device, Empatica Embrace [K181861] and does not raise any new or different questions of safety or effectiveness.
#
Page 12 of 12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.