The SLT Select Fiber Delivery Systems and Contact Tips are intended to be used in general surgery as well as multiple surgical specialties such as Urology, Gynecology, ENT, Head and Neck, Gastroenterology, Neurosurgery, Thorasic surgery, Pulmonology, Plastic surgery and Orthopedics. The fiber delivery systems and tips achieve the precise tissue effects of incision, excision, vaporization of tissue. The universal SMA-905 connector allows this family of fiber delivery systems to be used with any laser system of 532 to 1064 nm wavelength which accepts the SMA-905connector. Laser must operate at a wavelength between 532 and 1064 nm. Laser must operate in Continuous Wave (CW) or Quasi Continuous Wave (Quasi CW) mode. Laser must have a numerical aperture of 0.35 or less. Laser must accept universal SMA-905 connector, SLT proprietary connector, or an appropriate adapter into its fiber launch connector. Note: These Fiber Delivery Systems are cleared for the particular indications of the laser system to which they are connected. During use these products are exposed to biohazardous substances such as blood and other bodily fluids. They should be disposed of per your facility's procedures for biohazardous waste.
Device Story
Fiber delivery system acts as transmission path for 532-1064 nm wavelength laser energy from external laser generator to surgical site; utilizes optical fiber as propagation medium. System includes various contact tips and fiber models (e.g., SSRH, TCRH, FEF, CFE, LAL, ER, GR, MTR series) with universal SMA-905 connectors. Used in clinical settings by surgeons to achieve tissue incision, excision, vaporization, and coagulation. Device is passive; requires connection to compatible laser system (CW or Quasi-CW mode, NA ≤ 0.35). Output is laser energy delivered to tissue; clinical decision-making relies on surgeon's control of laser parameters and fiber positioning. Benefits include precise tissue interaction across multiple surgical specialties.
Clinical Evidence
No clinical data. Bench testing performed to demonstrate substantial equivalence, including biocompatibility testing (in vitro cytotoxicity, skin sensitization, intracutaneous reactivity, pyrogen, acute systemic toxicity) per ISO 10993 standards.
Technological Characteristics
Optical fiber delivery system with SMA-905 connectors; compatible with 532-1064 nm laser wavelengths. Materials biocompatibility tested per ISO 10993-5, 10993-10, 10993-23, and 10993-11. Passive transmission device; no internal software or energy source.
Indications for Use
Indicated for patients undergoing surgical procedures in general surgery, Urology, Gynecology, ENT, Head and Neck, Gastroenterology, Neurosurgery, Thoracic surgery, Pulmonology, Plastic surgery, and Orthopedics requiring tissue incision, excision, vaporization, or coagulation.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Submission Summary (Full Text)
{0}------------------------------------------------
December 22, 2024
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
JiangXi Medex Technology Co.,Ltd % Owen He Consultant Microkn Medical Technology Service (Shanghai) Co., Ltd. Room 901,No.889, Pinglu Road, Jing'an District Shanghai. China
Re: K243121
Trade/Device Name: SLT Select Fiber Delivery System and Contact Tips (models SSRH 8-SMA; TCRH 7-SMA; FEF 2.2-SMA; CFE 0.6-SMA; SSRH 8Z-SMA; SSRH 4-SMA; SSRH 6-SMA; SSRH 9-SMA; SSRH 10-SMA; SSRH 11-SMA; SSRH 7L-SMA; TCRH 7Z-SMA; FEF 1.8-SMA; CFE 0.55-SMA; CFE 0.36-SMA; LAL 550-SMA; LAL 550Z-SMA; LAL 365-SMA; LAL 365Z-SMA; ER2; ERP2; GR2; GRP2; ER4; ERP4; GR4; GRP4; ER6; ERP6; GR6; GRP6; ER8; ERP8; GR8; GRP8; ER10; ERP10; GR10; GRP10; ER12; ERP12; GR12; GRP12; MD2.5; MD3.5; HP1.0; MTR1.5; MTR3.5; MTRG1.5; MTRG2.5; MTRG3.5; MT1.5; MT3.5; DF2; MTRGL3.5; MTRL6; MTRL7; MTRL8; MTRL10; R3ERP2; R4ERP2; D1ERP4; D3ERP4; D3GRP4; D3GRP10; D3GRP14; D3MTRG4; D3MTRG20; SERP4; SGRP4; SMD3.5; SMTRG3.5).
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: July 20, 2024 Received: September 30, 2024
Dear Owen He:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination
{1}------------------------------------------------
product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part
{2}------------------------------------------------
3
803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
YAN FU -S
Digitally signed by YAN FU -S
Date: 2024.12.22 01:42:52 -05'00"
for Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K243121
#### Device Name
SLT Select Fiber Delivery System and Contact Tips (models SSRH 8-SMA: FEF 2.2-SMA: CFE 0.6-SMA: SSRH 8Z-SMA; SSRH 4-SMA; SSRH 6-SMA; SSRH 10-SMA; SSRH 11-SMA; SSRH 7L-SMA; TCRH 7Z-SMA; FEF 1.8-SMA; CFE 0.55-SMA; CFE 0.36-SMA; LAL 550-SMA; LAL 365-SMA; LAL 365Z-SMA; ER2; ERP2; GR2;
#### Indications for Use (Describe)
The SLT Select Fiber Delivery Systems and Contact Tips are intended to be used in general surgery as well as multiple surgical specialties such as Urology, Gynecology, ENT, Head and Neck, Gastroenterology, Neurosurgery, Thorasic surgery, Pulmonology, Plastic surgery and Orthopedics. The fiber delivery systems and tips achieve the precise tissue effects of incision, excision, vaporization of tissue. The universal SMA-905 connector allows this family of fiber delivery systems to be used with any laser system of 532 to 1064 nm wavelength which accepts the SMA-905connector.
·Laser must operate at a wavelength between 532 and 1064 nm.
* Laser must operate in Continuous Wave (CW) or Quasi Continuous Wave (Quasi CW) mode.
- · Laser must have a numerical aperture of 0.35 or less.
· Laser must accept universal SMA-905 connector, SLT proprietary connector, or an appropriate adapter into its fiber launch connector.
Note: These Fiber Delivery Systems are cleared for the particular indications of the laser system to which they are connected.
During use these products are exposed to biohazardous substances such as blood and other bodily fluids. They should be disposed of per your facility's procedures for biohazardous waste.
| Type of Use (Select one or both, as applicable) |
|----------------------------------------------------------------------------------|
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
510(k) Summary Prepared on: 2024-11-28 510(k) #: K243121 Contact Details 21 CFR 807.92(a)(1) JiangXi Medex Technology Co.,Ltd Applicant Name ChengXin Avenue, Xinfeng Industrial Area Xinfeng, Jiangxi CN 330104 Applicant Address Jiangxi 330104 China Applicant Contact Telephone +86 13530954829 Applicant Contact Ms. Zhang Kaixia Applicant Contact Email m055@jiangximedex.com Microkn Medical Technology Service (Shanghai) Co., Ltd. Correspondent Name Room 901,No.889, Pinglu Road, Jing'an District, Shanghai Shanghai Correspondent Address China +86 13795451145 Correspondent Contact Telephone Mr. Owen He Correspondent Contact Correspondent Contact Email fda@microkn.com 21 CFR 807.92(a)(2) Device Name SLT Select Fiber Delivery System and Contact Tips (Models: SSRH 8-SMA: TCRH 7-SMA: FEF 2.2-SMA: CFE 0.6-SMA: SSRH 8Z-SMA: SSRH 4-SMA: SSRH 6-SMA; SSRH 9-SMA; SSRH 10-SMA; SSRH 11-SMA; SSRH 7L-SMA; TCRH 7Z-SMA; FEF 1.8-SMA; CFE 0.55-SMA; CFE 0.36-SMA; LAL 550-SMA; LAL 550Z-SMA; LAL 365-SMA; LAL 365Z-SMA; ER2; ERP2; GR2; GRP2; ER4; ERP4; GR4; GRP4; ER6; ERP6; Device Trade Name GR6; GRP6; ER8; ERP8; GR8; GRP8; ER10; ERP10; GR10; GRP10; ER12; ERP12; GR12; GRP12; MD2.5; MD3.5; HP1.0; MTR1.5; MTR3.5; MTRG1.5; MTRG2.5; MTRG3.5; MT1.5; MT3.5; DF2; MTRGL3.5; MTRL6; MTRL7; MTRL8; MTRL10; R3ERP2; R4ERP2; D1ERP4; D3ERP4; D3GRP4; D3GRP10; D3GRP14; D3MTRG4; D3MTRG20; SERP4; SGRP4; SMD3.5; SMTRG3.5) Laser surgical instrument for use in general and plastic surgery and in Common Name dermatology Classification Name Powered Laser Surgical Instrument Regulation Number 878.4810 GEX Product Code(s) Legally Marketed Predicate Devices 21 CFR 807.92(a)(3)
Predicate #
Predicate Trade Name (Primary Predicate is listed first)
Product Code
{5}------------------------------------------------
## SLT SELECT FIBER DELIVERY SYSTEM AND CONTACT TIPS
GEX
21 CFR 807.92(a)(4)
## Device Description Summary
The SLT select fiber delivery system and contact tips is the transmission path for 810mm, 980mm wavelength laser generator to the surgical site, with the optical fiber as the propagation medium, by transmitting the diseased area, to achieve the treatment of the disease.
The SLT select fiber delivery system and contact tips, except for the CFE series of needle-tip one-piece flexible fibers, are fitted with a metal connector at the metal connector is threaded to connect to the contact laser tread, The FEF series models can be used as non-contact lasers without a treatment head.
# Intended Use/Indications for Use
The SLT Select Fiber Delivery Systems and Contact Tips are intended to be used in general surgery as well as multiple surgical specialties such as Urology, Gynecology, ENT, Head and Neck, Gastroenterology, Neurosurgery, Pulmonology, Plastic surgery and Orthopedics. The fiber delivery systems and tips achieve the tsue effects of incision, vaporizationand coagulation of tisue. The universal SMA-905 connector allows this family of fiber delivery systems to be used with any laser system of 532 to 1064 nm wavelength which accepts the SMA-905connector.
·Laser must operate at a wavelength between 532 and 1064 nm.
- * Laser must operate in Continuous Wave (CW) or Quasi Continuous Wave (Quasi CW) mode.
·Laser must have a numerical aperture of 0.35 or less.
- Laser must accept universal SMA-905 connector, or an appropriate adapter into its fiber launch connector. Note: These Fiber Delivery Systems are cleared for use for the laser system to which they are connected. During use these products are exposed to biohazardous such as blood and other bodily fluids. They should be disposed of per your facility's procedures for biohazardous waste.
# Indications for Use Comparison
The indication for use is the same with predicate devices.
# Technological Comparison
The purposed device and the predicated device are same in product code, requlations for use, Patient Population, Transmission efficiency, The tensile strength of the ioint between the optical fiber transmission body and the connector, The optical fiber tensile strength, minimum bending radius, optical fiber bending fatique resistance, Divergence angle,
# Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
## Non-clinical Testing
The bench testing of SLT Select Fiber Delivery Systems and Contact Tips and non-clinical testing is performed on new devices is to demonstrate the substantial equivalence to the predicate devices. Tests setup and execution are performed in accordance with applicable standards. Results of the testing demonstrate the standards and matching the performance of new devices to the predicate devices. The following performance data were provided in support of the substantial equivalence determination.
## 21 CFR 807.92(a)(5)
21 CFR 807.92(a)(5)
## 21 CFR 807.92(a)(6)
{6}------------------------------------------------
### Biocompatibility test
ln vitro cytoxicity test, Skin sensitization, Intracutaneous reactivity, Pyrogen, Acute systemic toxicity were conducted according to ISO 10993-5:2009, ISO 10993-10:2021, ISO 10993-23:2021, ISO 10993-11:2017.
No clinical study is included in this submission.
The SL T Select Fiber Delivery Systems and Contact Tips are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.