K243084 · Techsomed Medical Technologies , Ltd. · QZL · Dec 27, 2024 · Radiology
Device Facts
Record ID
K243084
Device Name
BioTraceIO Precision (2.0)
Applicant
Techsomed Medical Technologies , Ltd.
Product Code
QZL · Radiology
Decision Date
Dec 27, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.2052
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
BioTraceIO Precision is intended to provide physicians with adjunctive information in their clinical assessment of ablation zone created by liver tissue ablation, as part of their overall post-procedure clinical assessment. BioTraceIO Precision generates and depicts a map (BioTrace Map or BTM) post-procedure, that correlates with image findings seen with Contrast-enhanced Computed Tomography (CECT) obtained at 24hours post treatment. The information is provided in the 2D ultrasound plane. This is the only plane and location displayed. No imaging of other portions of the ablation zone is available. During the ablation procedure BioTraceIO Precision overlays the reference ablation zone (RAZ) provided by the ablation device manufacturer on the ultrasound image. BioTraceIO Precision is indicated for use in patients undergoing radiofrequency (RF) or microwave (MW) liver ablation procedures. BioTraceIO Precision is not intended for standalone prediction or for diagnostic purposes. BioTraceIO Precision does not support the use of multiple needles, either simultaneously or consecutively. Manual ablation target contours and margins as defined by the user are not intended for diagnosis, to predict ablation volumes or predict ablation success. The physician should not rely on BioTraceIO Precisions about patient management post treatment nor should BioTraceIO Precision serve as a substitute for any other assessment method, e.g., CT scans.
Device Story
BioTraceIO Precision is a standalone software application used in clinical operating environments by trained operators. It processes streamed 2D ultrasound images during liver ablation (RF or MW). In 'Online Mode', it overlays a Reference Ablation Zone (RAZ) onto the ultrasound image based on manufacturer technical parameters. In 'Offline Mode', it generates a BioTrace Map (BTM) 15 minutes post-procedure, correlating with 24-hour post-treatment CECT findings. The device provides adjunctive information to assist physicians in post-procedure clinical assessment; it does not replace CT scans or serve as a diagnostic tool. The system includes manual contouring tools for ablation targets and margins. By visualizing the expected ablation zone and correlating it with CECT, the device aims to support clinical decision-making regarding patient management post-treatment.
Clinical Evidence
Bench testing only. Validation compared optimized BTM to predicate BTM against 24-hour post-ablation CECT ground truth. Mean DICE coefficient for optimized BTM was 84.7 [95% CI: 83.07, 86.49] vs 85.5 [95% CI: 83.63, 87.43] for predicate (p=0.2408). Optimized BTM showed significantly higher DICE compared to T=0 CECT (p<0.0001).
Indicated for patients undergoing radiofrequency (RF) or microwave (MW) liver ablation procedures. Not for standalone prediction or diagnostic purposes. Does not support multiple needles.
Regulatory Classification
Identification
Post-ablation tissue response prediction software is an image processing software device intended to aid physicians with adjunctive information in their clinical assessment of the ablation zone following a tissue ablation procedure. This device uses information extracted from medical images along with other clinical data to predict the ablation zone post treatment. BioTraceIO Lite is indicated for use in patients undergoing radiofrequency (RF) or microwave (MW) liver ablation procedures to provide physicians with adjunctive information in their clinical assessment of the ablation zone created by liver tissue ablation, as part of their overall post-procedure clinical assessment.
Special Controls
In combination with the general controls of the FD&C Act, the post-ablation tissue response prediction software is subject to the following special controls:
- (1) Clinical performance testing must demonstrate the device performs as intended under anticipated conditions of use and evaluate the following:
- Ability to identify and visualize the ablation zone seen on images post treatment; (i) and
- Accuracy in predicting the ablation zone post treatment. (ii)
- (2) Software verification and validation must demonstrate device and algorithm functionality as informed by hazard analysis. Software documentation must include a detailed description of algorithm inputs and outputs, and any limitations of the algorithm.
- (3) Labeling must include:
- (i) A detailed description of the user workflow; and
- A detailed summary of the clinical performance testing, including test methods, (ii) dataset characteristics, imaging modality/equipment, anatomical region, patient population specifying type and pathology of target tissue, target locations and sizes, and results.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services USA logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 27, 2024
TechsoMed Medical Technologies Ltd. Dalia Dickman Head of Regulatory Affairs Meir Weisgal 2 REHOVOT, 7654055 ISRAEL
Re: K243084
Trade/Device Name: BioTraceIO Precision (2.0) Regulation Number: 21 CFR 892.2052 Regulation Name: Post-Ablation Tissue Response Prediction Software Regulatory Class: Class II Product Code: QZL Dated: December 2, 2024 Received: December 2, 2024
Dear Dalia Dickman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Digitally signed by Shahram for Shahram Vaezy -S Vaezy -S Date: 2024.12.27 13:58:03 -05'00'
Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K243084
Device Name BioTraceIO Precision (2.0)
#### Indications for Use (Describe)
BioTraceIO Precision is intended to provide physicians with adjunctive information in their clinical assessment of ablation zone created by liver tissue ablation, as part of their overall post-procedure clinical assessment. BioTracelO Precision generates and depicts a map (BioTrace Map or BTM) post-procedure, that correlates with image findings seen with Contrast-enhanced Computed Tomography (CECT) obtained at 24hours post treatment. The information is provided in the 2D ultrasound plane. This is the only plane and location displayed. No imaging of other portions of the ablation zone is available.
During the ablation procedure BioTraceIO Precision overlays the reference ablation zone (RAZ) provided by the ablation device manufacturer on the ultrasound image.
BioTracelO Precision is indicated for use in patients undergoing radiofrequency (RF) or microwave (MW) liver ablation procedures.
BioTraceIO Precision is not intended for standalone prediction or for diagnostic purposes. BioTraceIO Precision does not support the use of multiple needles, either simultaneously or consecutively.
Manual ablation target contours and margins as defined by the user are not intended for diagnosis, to predict ablation volumes or predict ablation success.
The physician should not rely on BioTraceIO Precisions about patient management post treatment nor should BioTraceIO Precision serve as a substitute for any other assessment method, e.g., CT scans.
| Type of Use (Select one or both, as applicable) | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| <table style="border:none;"><tr><td><span style="font-size:16px;"> ☒ </span> Prescription Use (Part 21 CFR 801 Subpart D)</td></tr><tr><td><span style="font-size:16px;"> ☐ </span> Over-The-Counter Use (21 CFR 801 Subpart C)</td></tr></table> | <span style="font-size:16px;"> ☒ </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:16px;"> ☐ </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
| <span style="font-size:16px;"> ☒ </span> Prescription Use (Part 21 CFR 801 Subpart D) | | |
| <span style="font-size:16px;"> ☐ </span> Over-The-Counter Use (21 CFR 801 Subpart C) | | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(K) SUMMARY
# K243084
### TechsoMed's BioTracelO Precision
## Submitter
TechsoMed Medical Technologies. LTD. Meir Weisgal 2 Rehovot Israel
Phone: +972545595951
Contact Person: Dalia Dickman, PhD.
Date Prepared: September 26, 2024
Name of Device: BioTraceIO Precision (2.0)
Common or Usual Name: BioTracelO Precision
Classification Name: Post-Ablation Tissue Response Prediction Software (CFR 892.2052)
Regulatory Class: Class II
Product Code: QZL
#### Predicate Device
| 510(k) Number | DEN230020 |
|----------------------|---------------------------------------------------|
| Trade Name | BioTraceIO Precision |
| Manufacturer | TechsoMed Medical Technologies. LTD. |
| Device Name | BioTraceIO Lite |
| Regulation Number | 892.2052 |
| Regulation Name | Post-Ablation Tissue Response Prediction Software |
| Regulatory Class | Class II |
| Primary Product Code | QZL |
#### Reference Device
| 510(k) Number | K240773 |
|-------------------|------------------------------------------------|
| Trade Name | BioTraceIO Vision |
| Manufacturer | TechsoMed Medical Technologies. LTD. |
| Device Name | VisAble.IO |
| Regulation Number | 892.2050 |
| Regulation Name | Medical Image Management and Processing System |
| Regulatory Class | Class II |
| Product Code | QTZ, QIH, LLZ |
{5}------------------------------------------------
### Device Description
BioTracelO Precision is a stand-alone software application with tools and features designed to assist users in their clinical assessment of ablation zone created by liver tissue ablation, as part of their overall post-procedure clinical assessment.
BioTracelO Precision is a software application that uses a proprietary computational algorithm to analyze ultrasound images captured during liver ablation treatment (microwave ablation [MWA] or radiofrequency ablation [RFA]), as depicted in standard abdominal ultrasound imaging.
The streamed ultrasound images are captured and analyzed by the BioTracelO algorithm. providing a visual display of the expected ablation zone, namely the Reference Ablation Zone (RAZ), calculated based on technical parameters provided by the ablation manufacturer datasheet. The RAZ is displayed during the procedure (Online Mode). BioTracelO Precision also provides a
visual display of the estimated ablation zone correlative to the 24-hour CECT, namely the BioTrace Map (BTM), after the completion of the procedure (Offline Mode).
#### Intended Use / Indications for Use
BioTracelO Precision is intended to provide physicians with adjunctive information in their clinical assessment of ablation zone created by liver tissue ablation, as part of their overall post-procedure clinical assessment.
BioTraceIO Precision generates and depicts a map (BioTrace Map or BTM) post-procedure, that correlates with image findings seen with Centrast-enhanced Computed Tomography (CECT) obtained at 24hours post treatment. The information is provided in the 2D ultrasound plane. This is the only plane and location displayed. No imaging of other portions of the ablation zone is available.
During the ablation procedure BioTracelO Precision overlays the reference ablation zone (RAZ) provided by the ablation device manufacturer on the ultrasound image. BioTraceIO Precision is indicated for use in patients undergoing radiofrequency (RF) or microwave (MW) liver ablation procedures.
BioTracelO Precision is not intended for standalone prediction or for diagnostic purposes. BioTracelO Precision does not support the use of multiple needles, either simultaneously or consecutively.
Manual ablation target contours and margins as defined by the user are not intended for diagnosis, to predict ablation volumes or predict ablation success.
The physician should not rely on BioTraceIO Precision BTM alone in decisions about patient management post treatment nor should BioTracelO Precision serve as a substitute for any other assessment method, e.g., CT scans.
#### Summary of Technological Characteristics
Both the subject and predicate device are stand-alone software applications that use a proprietary computational algorithm to analyze ultrasound images captured during liver ablation treatment (microwave ablation [MWA] or radiofrequency ablation [RFA]), as depicted in standard abdominal ultrasound imaging.
The streamed ultrasound images are captured and analyzed by the BioTracelO algorithm, providing a visual display of the expected ablation zone, namely the Reference Ablation Zone (RAZ), calculated based on technical parameters provided by the ablation manufacturer datasheet. The RAZ is displayed during the procedure (Online Mode). BioTracelO Precision also provides a visual display of the estimated ablation zone correlative to the 24-hour CECT, namely the BioTrace
{6}------------------------------------------------
Map (BTM), after the completion of the procedure (Offline Mode).
The following technological differences exist between the subject and the predicate device:
The primary differences between devices are: (1) BTM algorithm processing time has been optimized to reduce delay for the user, and in turn is displayed in the offline mode 15 minutes after the ablation procedure has been completed (2) the RAZ algorithm has been adapted to have the same implementation used for BTM processing for the purpose of algorithm optimization and easier maintenance (3) user interface (UI) modifications have been made to improve the user experience
(4) cybersecurity improvements have been made (5) technical modifications have been made such as name change, file configuration change and architecture change.
A table comparing the key features of the subject and predicate devices is provided below.
{7}------------------------------------------------
# Substantial Equivalence Comparison Table
| | Subject Device | Predicate Device |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | BioTraceIO Precision (Ver 2.0) | BioTraceIO Precision<br>(previously commercially named<br>BioTraceIO Lite) |
| DeNovo/510(k)<br>number | K243084 | DEN230020 |
| Classification | Class II QZL | Class II QZL |
| Intended Use | A stand-alone software application<br>with tools and features designed to<br>assist users in their clinical<br>assessment of ablation zone<br>created by liver tissue ablation, as<br>part of their overall post-procedure<br>clinical assessment. | A stand-alone software application<br>with tools and features designed to<br>assist users in their clinical<br>assessment of ablation zone<br>created by liver tissue ablation, as<br>part of their overall post-procedure<br>clinical assessment. |
| Indications<br>for<br>Use | BioTraceIO Precision is intended<br>to provide physicians with<br>adjunctive information in their<br>clinical<br>assessment of ablation zone<br>created by liver tissue ablation, as<br>part of their overall post-procedure<br>clinical assessment.<br>BioTraceIO Precision generates<br>and depicts a map (BioTrace Map<br>or BTM) post-procedure, that<br>correlates with image findings<br>seen with Contrast-enhanced<br>Computed Tomography (CECT)<br>obtained at 24 hours post<br>treatment. The information is<br>provided in the 2D ultrasound<br>plane. This is the only plane<br>and location displayed. No imaging<br>of other portions of the ablation<br>zone is available.<br>During the ablation procedure<br>BioTraceIO Precision overlays the<br>reference ablation zone (RAZ)<br>provided by the ablation device<br>manufacturer on the ultrasound<br>image.<br>BioTraceIO Precision is indicated<br>for use in patients undergoing<br>radiofrequency (RF) or microwave<br>(MW) liver ablation procedures.<br>BioTraceIO Precision is not<br>intended for standalone prediction<br>or for diagnostic purposes.<br>BioTraceIO Precision does not<br>support the use of multiple | BioTraceIO Lite is intended to<br>provide physicians with adjunctive<br>information in their clinical<br>assessment of ablation zone<br>created by liver tissue ablation, as<br>part of their overall post-procedure<br>clinical assessment.<br>BioTraceIO Lite generates and<br>depicts a map (BioTrace Map or<br>BTM) post-procedure, that<br>correlates with image findings seen with<br>Contrast-enhanced Computed<br>Tomography (CECT) obtained at 24<br>hours post treatment. The<br>information is provided in the 2D<br>ultrasound plane. This is the only<br>plane<br>and location displayed. No imaging<br>of other portions of the ablation<br>zone is available.<br>During the ablation procedure<br>BioTraceIO Lite overlays the<br>reference ablation zone (RAZ)<br>provided<br>by the ablation device manufacturer<br>on the ultrasound image.<br>BioTraceIO Lite is indicated for use<br>in patients undergoing<br>radiofrequency (RF) or microwave<br>(MW)<br>liver ablation procedures.<br>BioTraceIO Lite is not intended for<br>standalone prediction or for<br>diagnostic purposes. |
| | needles, either simultaneously or<br>consecutively.<br>Manual ablation target contours<br>and margins as defined by the user<br>are not intended for diagnosis, to<br>predict ablation volumes or predict<br>ablation success.<br><br>The physician should not rely on<br>BioTraceIO Precision BTM alone<br>in decisions about patient<br>management<br>post treatment nor should<br>BioTraceIO Precision serve as a<br>substitute for any other<br>assessment method,<br>e.g., CT scans. | BioTraceIO Lite does not support<br>the use of multiple needles, either<br>simultaneously or consecutively.<br><br>The physician should not rely on<br>BioTraceIO Lite BTM alone in<br>decisions about patient<br>management<br>post treatment nor should<br>BioTraceIO Lite serve as a<br>substitute for any other assessment<br>method,<br>e.g., CT scans. |
| Primary<br>Differences | (1) the BTM algorithm processing<br>time has been optimized through<br>enhancements in algorithm<br>architecture to reduce delay for the<br>user, and in turn is displayed in the<br>offline mode 15 minutes after the<br>ablation procedure has been<br>completed (2) the RAZ algorithm<br>has been updated to utilize the<br>same architectural framework as<br>the BTM algorithm. This adaptation<br>facilitates optimization and<br>simplifies maintenance (3) user<br>interface (UI) modifications have<br>been made to improve the user<br>experience such as button layout<br>improvements and allow for<br>selection of a reference US frame<br>for needle marking (4)<br>cybersecurity improvements have<br>been made such as encryption of<br>PHI and installation responsibilities<br>(5) technical modifications have<br>been made such as name change,<br>report viewer file configuration<br>change and software framework<br>architecture change. | NA |
| User Population | Qualified trained operators | Qualified trained operators |
| Where used | BioTraceIO Precision is<br>designated for use in a clinical<br>operating environment with<br>compatible ultrasound and liver<br>tissue ablation devices. | BioTraceIO Precision is designated<br>for use in a clinical operating<br>environment with<br>compatible<br>ultrasound and liver tissue ablation<br>devices. |
| Energy Used | None - software only application.<br>The software application does not<br>deliver or depend on energy<br>delivered to or from patients | None - software only application.<br>The software application does not<br>deliver or depend on energy<br>delivered to or from patients |
| Technological<br>Characteristics | BioTraceIO Precision is a software<br>application that uses a proprietary<br>computational algorithm to analyze<br>ultrasound images captured during | BioTraceIO Precision is a software<br>application that uses a proprietary<br>computational algorithm to analyze<br>ultrasound images captured during |
| | liver ablation treatment<br>(microwave ablation [MWA] or<br>radiofrequency ablation [RFA]), as depicted in<br>standard abdominal ultrasound<br>imaging.<br>The streamed ultrasound images<br>are captured and analyzed by the<br>BioTraceIO algorithm, providing a<br>visual display of the expected<br>ablation zone, namely the<br>Reference Ablation Zone (RAZ),<br>calculated based on technical<br>parameters provided by the<br>ablation manufacturer datasheet.<br>The RAZ is displayed, during the<br>procedure (Online Mode).<br>BioTracelO Precision also<br>provides a visual display of the<br>estimated ablation zone correlative<br>to the 24-hour CECT, namely the<br>BioTrace Map (BTM)<br>fifteen (15) minutes after the<br>completion of the procedure<br>(Offline Mode).<br>Color-coded display of scores<br>reflecting image quality and<br>reliability of the RAZ motion<br>compensation displayed in the UI<br>using a color-coded 5-point scale.<br>These scores are computed based<br>on objective metrics, and their<br>purpose is to ensure the correct<br>execution of the algorithms in the<br>BioTraceIO Precision system.<br>Manual contouring of ablation<br>target on the reference US frame<br>and display of the ablation margin<br>as defined by the user around the<br>contour. This feature was cleared in<br>K240773. | liver ablation treatment (microwave<br>ablation [MWA] or radiofrequency<br>ablation [RFA]), as depicted in<br>standard abdominal ultrasound<br>imaging.<br>The streamed ultrasound images<br>are captured and analyzed by the<br>BioTraceIO algorithm, providing a<br>visual display of the expected<br>ablation zone, namely the<br>Reference Ablation Zone (RAZ),<br>calculated based on technical<br>parameters provided by the ablation<br>manufacturer datasheet. The RAZ<br>is displayed, during the procedure<br>(Online Mode). BioTracelO<br>Precision also provides a visual<br>display of the estimated ablation<br>zone correlative to the 24-hour<br>CECT, namely the BioTrace Map<br>(BTM), thirty (30) minutes after the<br>completion of the procedure (Offline<br>Mode).<br>Generation of scores reflecting image<br>quality and reliability of the RAZ<br>motion compensation which runs as<br>background calculations within the<br>software and is not visible to the user.<br>These scores are computed based<br>on objective metrics, and their<br>purpose is to ensure the correct<br>execution of the algorithms in the<br>BioTraceIO Precision system.<br>NA |
| Design:<br>Data<br>Visualization | BTM on offline mode; RAZ on<br>online mode | BTM on offline mode; RAZ on<br>online mode |
| Design;<br>Data<br>Input | Used with a hardware streaming<br>device that is connected to an<br>ultrasound machine output, to<br>capture input ultrasound images. | Used with a hardware streaming<br>device that is connected to an<br>ultrasound machine output, to<br>capture input ultrasound images. |
| Design:<br>Online<br>Mode | In online mode, the BioTracelO<br>algorithm analyses the ultrasound<br>images, and calculates a<br>transformation matrix per frame.<br>The transformation matrix defines<br>the movement and orientation<br>change between two frames and is<br>used to transform the RAZ and<br>register it to the ultrasound image. | In online mode, the BioTracelO<br>algorithm analyses the ultrasound<br>images, and calculates a<br>transformation matrix per frame.<br>The transformation matrix defines<br>the movement and orientation<br>change between two frames and is<br>used to transform the RAZ and<br>register it to the ultrasound image. |
| | In offline mode, the BioTraceIO<br>algorithm calculates the BTM as it<br>is generated during the procedure. | In offline mode, the BioTraceIO<br>algorithm calculates the BTM as it is<br>generated during the procedure. |
| | BTM calculations are based on the<br>biological response of the tissue to<br>heating and its appearance on<br>ultrasound imaging. | BTM calculations are based on the<br>biological response of the tissue to<br>heating and its appearance on<br>ultrasound imaging. |
| Design: | In offline mode, the BioTraceIO<br>algorithm calculates the BTM as it<br>is generated during the procedure. | In offline mode, the BioTraceIO<br>algorithm calculates the BTM as it is<br>generated during the procedure. |
| Mode | BTM calculations are based on the<br>biological response of the tissue to<br>heating and its appearance on<br>ultrasound imaging. | BTM calculations are based on the<br>biological response of the tissue to<br>heating and its appearance on<br>ultrasound imaging. |
| Offline | | |
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
#### Performance Data
BioTraceIO Precision is validated and verified against its user needs and intended use by the successful execution of planned performance, functional and algorithmic testing included in this submission.
The results of functional and algorithmic testing demonstrate that BioTracelO Precision meets the user needs and requirements of the device, which are considered to be substantially equivalent to those of the listed predicate device.
Alqorithmic testing was performed on the BTM alqorithm, to ensure that performance and accuracy was as expected. The validation analysis resulted in a mean DICE coefficient of 84.7 [95% C1: 83.07, 86.49] and 85.5 [95% Cl: 83.63, 87.43] for the optimized BTM and predicate BTM, respectively, as compared to ablation zone visualized on the 24-hour post ablation CECT (T=24 CECT). This result was tested using both paired t-test (p = 0.2408), as well as Wilcoxon Signed-Rank test (p = 0.11). Both methodologies yielded similar non-significant p-values, indicating no
significant differences in measurement of mean dice coefficients between the two algorithms. The optimized BTM also demonstrated a significantly (paired t-test p < 0.0001; Wilcoxon Signed-Rank p < 0.0001) higher mean DICE coefficient compared to T=0 CECT when both were compared to T=24 CECT, with a mean within-tumor difference of 7.9 [95% CJ: 4.2, 11.7) in favor of the optimized BTM.
These results correspond to the intended use of the device in correlating with the ablation zone visualization on T=24 CECT and supports its clinical utility.
The mean DICE coefficient for BioTracelO Precision BTM compared to T=24 CECT was not significantly lower than the mean DICE coefficient for BioTracelO Lite BTM versus T=24 CECT. This result highlights that BioTracelO Precision BTM demonstrates equivalent performance in correlation of ablation zone visualized on T=24 CECT compared to the predicate BTM.
Overall, the validation has successfully demonstrated that the optimized BTM demonstrates equivalent performance to the predicate BTM, and that it is safe and effective in estimating the area of ablation zone as visualized at 24-hour post procedure CECT scan.
Test planning was performed in accordance with standard testing procedures and quidelines as listed in internal development processes.
Verification and validation testing were carried out as per planned arrangements in the Project Test Plan and Phase Test Plan(s) to ensure that design outputs meet design inputs and that this version
{11}------------------------------------------------
of BioTracelO Precision meets the product acceptance criteria. These are in accordance with the company's Design Control process in compliance with 21 CFR Part 820.30, which included testing that fulfills the requirements of FDA "Guidance on Software Contained in Medical Devices" and adherence to the DICOM standard.
### Conclusions
The BioTracelO Precision (V2.0) is as safe and effective as the cleared BioTracelO Precision (DEN230020). The BioTracelO Precision has the same intended use, and principles of operation and similar technological characteristics as its predicate device. The minor differences in indications do not alter the intended use of the device and do not affect its safety and effectiveness when used as labeled. In addition, the minor technological differences between the BioTracelO Precision and its predicate device raise no new issues of safety or effectiveness. Algorithmic testing demonstrates that the BioTraceIO Precision (V 2.0) is as safe and effective as the cleared BioTracelO Precision (DEN230020). Thus, the BioTraceIO Precision is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.