K243078 · Osseofuse International, Inc. · DZE · Oct 10, 2025 · Dental
Device Facts
Record ID
K243078
Device Name
HexaPLUS S OneDrill Implant System
Applicant
Osseofuse International, Inc.
Product Code
DZE · Dental
Decision Date
Oct 10, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The HexaPLUS S OneDrill Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cement-retained, screw-retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one and two stage surgical procedures. The system is intended for delayed loading.
Device Story
Dental implant system comprising titanium alloy (Ti 6Al 4V ELI Gr.23) fixtures and various abutments; surgically placed in maxilla or mandible bone. Fixtures feature SLA surface treatment and tapered body design with internal hexagon connection. System used by dental surgeons in clinical settings for tooth replacement. Inputs include surgical site preparation; device provides structural foundation for prosthetic restorations. Output is mechanical support for dental prosthetics. Healthcare providers use device to restore oral function and aesthetics. Benefits include stable anchorage for fixed or removable dental restorations.
Clinical Evidence
Bench testing only. Fatigue testing performed per ISO 14801. Biocompatibility, sterilization, and shelf-life validated via predicate device (K181157). Surface cleanliness verified via SEM and EDS. Bacterial endotoxins tested per USP <85>, USP <161>, and ANSI/AAMI ST72:2019. Non-clinical MRI assessment conducted for displacement force and torque based on scientific literature.
Indicated for partially or fully edentulous adult patients requiring single or multiple-unit dental restorations, including cement-retained, screw-retained, or overdenture restorations, and fixed bridgework support. Suitable for one or two-stage surgical procedures with delayed loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 10, 2025
Osseofuse International Inc.
% Priscilla Chung
Regulatory Affairs Consultant
LK Consulting Group USA, Inc.
18881 Von Karman Ave STE 160
Irvine, California 92612
Re: K243078
Trade/Device Name: HexaPLUS S OneDrill Implant System
Regulation Number: 21 CFR 872.3640
Regulation Name: Endosseous Dental Implant
Regulatory Class: Class II
Product Code: DZE, NHA
Dated: February 13, 2025
Received: September 12, 2025
Dear Priscilla Chung:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243078 - Priscilla Chung
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K243078 - Priscilla Chung
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Andrew I. Steen -S
Andrew I. Steen
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K243078 | |
| Device Name HexaPLUS S OneDrill Implant System | |
| Indications for Use (Describe) The HexaPLUS S OneDrill Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cement-retained, screw-retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one and two stage surgical procedures. The system is intended for delayed loading. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Summary
(K243078)
This summary of 510(K) is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: 10/9/2025
1. Submitter/Applicant
Osseofuse International Inc.
6170 West Desert Inn Rd. Suite B
Las Vegas, NV 89146
Tel: +1-888-446-9995
Email: support@osseofuse.com
2. U.S Agent/Contact Person
Priscilla Chung
LK Consulting Group USA, Inc.
18881 Von Karman Ave STE 160, Irvine CA 92612
Phone: 714-202-5789 Fax: 714-409-3357
Email: juhee.c@lkconsultinggroup.com
3. Device
- Trade Name: HexaPLUS S OneDrill Implant System
- Common Name: Dental Implant System
- Classification: Class II
- Classification regulation: 21 CFR 872.3640
- Product Code: DZE, NHA
4. Predicate Devices:
Primary Predicate Device:
- Conical Plus Implant System (K181157) by OsseoFuse International, Inc.
Reference Devices:
- Osseofuse Dental Implant System (K131748) by OsseoFuse International, Inc.
- Osseofuse Dental Implant System (K110577) by Dynamic Innovations Inc.
510(k) Submission
510(k) summary, 1 / 9 page
{5}
510(k) Submission
510(k) summary, 2 / 9 page
## 5. Description:
The HexaPLUS S OneDrill Implant System is a dental implant system made of Titanium 6AL 4V ELI Gr.23 alloy intended to be surgically placed in the bone of the upper or lower jaw arches. The system is similar to other commercially available products based on the indications for use, the technology used, the material compositions and performance characteristics. The surface of the implants are treated by SLA method. Implants with 4.50mm and 5.25mm body diameter are provided in lengths of 8.5, 10, 11.5, 13.0, 14.5, and 16 mm. Implants with 6.5mm and 7.5mm body diameter are provided in lengths of 8.5, 10, 11.5, 13.0, and 14.5.
The system includes the following components.
### Fixtures
- Size: 4.50mm (Dia.) x 8.5/10.0/11.5/13.0/14.5/16.0mm (L)
- 5.25mm (Dia.) x 8.5/10.0/11.5/13.0/14.5/16.0mm (L)
- 6.50mm (Dia.) x 8.5/10.0/11.5/13.0/14.5
- 7.50mm (Dia.) x 8.5/10.0/11.5/13.0/14.5
### Abutments
- Healing Abutment
- One-Step Abutment
- Final Cement Abutment
- Straight Abutment
- Angled Abutment
- Temporary Abutment
- Ball Abutment
### Cover Screw
- Size: 3.45mm (Dia.) x 4.9
### Abutment Screw
- Size: 2.30mm (Dia.) x 8.25
### Multi Angle Abutment Screw
- Size: 2.30mm (Dia.) x 8.20
## 6. Indication for use:
The HexaPLUS S OneDrill Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cement-retained, screw-retained, or overdenture restorations, and
{6}
terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one and two stage surgical procedures. The system is intended for delayed loading.
## 7. Performance Data
- Fatigue testing was performed on the angled abutments according to ISO 14801.
- Sterilization, Shelf-life validation and biocompatibility are leveraged from the predicate device (K181157).
- SEM and EDS testing was conducted on the subject devices to verify that there is no residual on the implants after the surface treatment.
- Bacterial Endotoxins Test (BET) is performed using a kinetic Limulus Amebocyte Lysate (LAL) assay with a Portable Test System (PTS), in accordance with USP <85>, USP <161>, and ANSI/AAMI ST72:2019.
- A non-clinical worst-case MRI assessment was conducted for the HexaPLUS S OneDrill Implant System to evaluate its behavior in the MRI environment. The evaluation was based on scientific rationale and published literature (e.g., Terry O. Woods, Jana Delfino, & Sunder Rajan (2019) Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices. Journal of Testing and Evaluation 49.2, 783-795) and considered the complete system, including all compatible implant bodies, dental abutments, and fixation screws, as well as their material compositions. The rationale addressed relevant parameters in accordance with the FDA guidance “Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment”, specifically evaluating magnetically induced displacement force and torque.
## 8. Basis for Substantial Equivalence
The HexaPLUS S OneDrill Implant System is substantially equivalent to the predicate device, Conical Plus Implant System (K81157) made by our company.
Comparison Chart
| | Subject Device | Predicate Device | Reference Device |
| --- | --- | --- | --- |
| 510(K) Number | K243078 | K181157 | K110577 |
| Device Name | HexaPLUS S OneDrill Implant System | Conical Plus Implant System | OsseoFuse Dental Implant System |
| 510k Applicant | OsseoFuse International, Inc. | OsseoFuse International, Inc. | Dynamic Innovations Inc. |
| Indications for Use | The HexaPLUS S OneDrill Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cement-retained, screw-retained, or overdenture | The Conical Plus Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cement-retained, screw-retained, or overdenture | The OsseoFuse Dental Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw-retained, |
510(k) Submission
510(k) summary, 3 / 9 page
{7}
| | restorations, and terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one and two stage surgical procedures. The system is intended for delayed loading. | restorations, and terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one and two stage surgical procedures. The systems is intended for delayed loading. | or overderiture restorations, and terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one and two stage surgical procedures and not dedicated for immediate loading. This system is intended for delayed loading. |
| --- | --- | --- | --- |
| Principle of Operations | - Threaded Fixation
- Combined with various abutments and attachments | - Threaded Fixation
- Combined with various abutments and attachments | - Threaded Fixation
- Combined with various abutments and attachments |
| Design | • Implant Type: Bone Level Implant
• Connection Type: Internal Hexagon
• Neck Design: Straight walled neck with micro-thread
• Body Design: Tapered design | • Implant Type: Bone Level Implant
• Connection Type: Internal Hexagon
• Neck Design: Straight walled neck with micro-thread
• Body Design: Tapered design | • Implant Type: Bone Level Implant
• Connection Type: Internal Hexagon
• Neck Design: Straight walled neck with micro-thread
• Body Design: Tapered design |
| Appearance | | | |
| Implant Material | Ti 6Al 4V ELI, Gr.23 | Ti 6Al 4V ELI, Gr.23 | Ti 6Al 4V ELI, Gr.23 |
| Surface Treatment | SLA Treatment on the fixture body | SLA Treatment on the fixture body | RBM Treatment on the fixture body |
| Implant Sterile | Yes | Yes | Yes |
| Sterilization Method | Gamma | Gamma | Gamma |
| Implant Diameters | 4.5mm, 5.25mm, 6.5mm, 7.5mm | 3.75mm, 4.5mm, 5.25mm, 6.5mm, 7.5mm | 3.75mm, 4.1mm, 4.5mm, 5.25mm |
| Implant Lengths | 8.5 – 16.0 mm | 8.5 – 16.0 mm | 8.5 – 16.0 mm |
510(k) Submission
510(k) summary, 4 / 9 page
{8}
| Abutments | | | |
| --- | --- | --- | --- |
| 510(K) Number | | K243078 | K110577 |
| Device Name | | HexaPLUS S OneDrill Implant System | Hexa Plus System |
| 510k Applicant | | OsseoFuse International, Inc. | OsseoFuse International, Inc. |
| Healing Abutment | Appearance | | |
| | Implant Material | Ti 6Al 4V ELI Gr. 23. | Ti 6Al 4V ELI Gr. 23. |
| | Sterilization Method | Steam | Steam |
| | Diameters | 4.00mm, 4.70mm, 5.50mm | 4.00mm, 4.70mm, 5.50mm |
| | Cuff | 3.0mm, 5.0mm, 7.0mm | 3.0mm, 5.0mm, 7.0mm |
| | Lengths | 7.9mm, 9.9mm, 11.9mm | 7.9mm, 9.9mm, 11.9mm |
| | Appearance | | |
| | Implant Material | Ti 6Al 4V ELI Gr. 23. | Ti 6Al 4V ELI Gr. 23. |
| | Sterilization Method | Steam | Steam |
| | Diameters | 4.00mm, 4.70mm, 5.50mm | 4.00mm, 4.70mm, 5.50mm |
| | Cuff | 3.0mm, 5.0mm, 7.0mm | 3.0mm, 5.0mm, 7.0mm |
| | Lengths | 8.9mm, 10.9mm, 12.9mm | 7.9mm, 9.9mm, 11.9mm |
| | Appearance | | |
| | Implant Material | Ti 6Al 4V ELI Gr. 23. | Ti 6Al 4V ELI Gr. 23. |
| | Sterilization Method | Steam | Steam |
| | Diameters | 7.00 mm, 8.00 mm | 7.10 mm, 8.10 mm |
| | Cuff | 3.5 mm, 5.0 mm, 7.0 mm | 3.0 mm, 4.0 mm, 5.0 mm, 7.0 mm |
| | Lengths | 8.90 mm, 9.40 mm, 10.90 mm, 12.90 mm | 9.40 mm, 10.40 mm, 11.40 mm, |
510(k) Submission
510(k) summary, 5 / 9 page
{9}
| | | | 13.40 mm, 15.40 mm |
| --- | --- | --- | --- |
| | Appearance | | |
| | Implant Material | Ti 6Al 4V ELI Gr. 23. | Ti 6Al 4V ELI Gr. 23. |
| | Sterilization Method | Steam | Steam |
| | Diameters | 4.00 mm, 4.70 mm, 5.50 mm | 4.00mm, 4.70mm, 5.50mm |
| | Cuff | 3.0 mm, 5.0 mm, 7.0 mm | 3.0mm, 5.0mm, 7.0mm |
| | Lengths | 8.90 mm, 10.90 mm, 12.90 mm | 7.9mm, 9.9mm, 11.9mm |
| | Appearance | | |
| | Implant Material | Ti 6Al 4V ELI Gr. 23. | Ti 6Al 4V ELI Gr. 23. |
| | Sterilization Method | Steam | Steam |
| | Diameters | 7.00 mm, 8.00 mm | 7.10 mm, 8.10 mm |
| | Cuff | 3.5 mm, 5.0 mm, 7.0 mm | 3.0 mm, 4.0 mm, 5.0 mm, 7.0 mm |
| | Lengths | 8.90 mm, 9.40 mm, 10.90 mm, 12.90 mm | 9.40 mm, 10.40 mm, 11.40 mm, 13.40 mm, 15.40 mm |
| One-Step Abutment | Appearance | | |
| | Implant Material | Ti 6Al 4V ELI Gr. 23. | Ti 6Al 4V ELI Gr. 23. |
| | Sterilization Method | Steam | Steam |
| | Diameters | 4.5mm, 5.25mm, 6.00mm | 4.5mm, 5.25mm, 6.00mm |
| | Lengths | 11.6mm | 11.6mm |
510(k) Submission
510(k) summary, 6 / 9 page
{10}
| Final Cement Abutment | Appearance | | |
| --- | --- | --- | --- |
| | Implant Material | Ti 6Al 4V ELI Gr. 23. | Ti 6Al 4V ELI Gr. 23. |
| | Sterilization Method | Steam | Steam |
| | Diameters | 5.25mm, 6.0mm | 5.25mm, 6.0mm |
| | Cuff | 2.0mm, 4.0mm | 2.0mm, 4.0mm |
| | Lengths | 13.83mm, 15.83mm | 13.83mm, 15.83mm |
| Straight Abutment | Appearance | | |
| | Implant Material | Ti 6Al 4V ELI Gr. 23. | Ti 6Al 4V ELI Gr. 23. |
| | Sterilization Method | Steam | Steam |
| | Diameters | 3.75mm | 3.75mm |
| | Lengths | 13.83mm | 13.83mm |
| Angled Abutment | Appearance | | |
| | Implant Material | Ti 6Al 4V ELI Gr. 23. | Ti 6Al 4V ELI Gr. 23. |
| | Sterilization Method | Steam | Steam |
| | Diameters | 5.25 mm | 5.25 mm |
| | Cuff | 2.6 mm | 2.6 mm |
| | Lengths | 6.9 mm | 6.9mm |
| | Angulation | 17° | 17° |
| Temporary Abutment | Appearance | | |
510(k) Submission
510(k) summary, 7 / 9 page
{11}
| | Implant Material | Ti 6Al 4V ELI Gr. 23. | Ti 6Al 4V ELI Gr. 23. |
| --- | --- | --- | --- |
| | Sterilization Method | Steam | Steam |
| | Diameters | 4.75 mm | 4.75 mm |
| | Lengths | 12.45 mm, 13.0 mm | 13.0 mm |
| Ball Abutment | Appearance | | |
| | Implant Material | Ti 6Al 4V ELI Gr. 23. | Ti 6Al 4V ELI Gr. 23. |
| | Sterilization Method | Steam | Steam |
| | Diameters | 4.0 mm | 3.58 mm |
| | Cuff | 1.0 mm, 2.0mm, 3.00 mm, 4.0mm | 3.0 mm |
| | Lengths | 9.70 mm, 10.70mm, 11.70 mm, 12.70mm | 13.15 mm |
| Abutment Screw | Appearance | | |
| | Implant Material | Ti 6Al 4V ELI Gr. 23. | Ti 6Al 4V ELI Gr. 23. |
| | Sterilization Method | Steam | Steam |
| | Diameters | 2.30 mm | 2.30 mm |
| | Lengths | 8.20 mm | 8.20 mm |
| Multi Angle Abutment Screw | Appearance | | |
| | Implant Material | Ti 6Al 4V ELI Gr. 23. | Ti 6Al 4V ELI Gr. 23. |
| | Sterilization Method | Steam | Steam |
| | Diameters | 2.30 mm | 2.30 mm |
| | Lengths | 8.20 mm | 8.20 mm |
| Cover screw | | | |
| 510(K) Number | | - | K131748 |
| Device Name | | HexaPLUS S OneDrill Implant System | Osseofuse Dental Implant System |
| 510k Applicant | | OsseoFuse International, Inc. | OsseoFuse International, Inc. |
510(k) Submission
510(k) summary, 8 / 9 page
{12}
| Contract Manufacturer | KJ Meditech Co., Ltd. | KJ Meditech Co., Ltd. |
| --- | --- | --- |
| Appearance | | |
| Implant Material | Ti 6Al 4V ELI, Gr.23 | Ti 6Al 4V ELI, Gr.23 |
| Sterilization Method | Steam | Steam |
| Diameters | 3.45 mm | 3.45 mm |
| Lengths | 4.90 mm | 4.90 mm |
## Substantial Equivalence Discussion
The proposed subject device has the identical indications for use, technological characteristics, design and mode of operation as the above identified predicate device. The proposed device has the identical intended use as the predicate and is placed using the identical methodology as all selected predicate devices.
The fixture went through minor design change, so we conducted performance testing and the test result support that it does not raise a concern in safety and effectiveness. The abutments have not changed except ball abutments. But the changes in design and size are minor enough not to raise a concern in safety and effectiveness.
Based on the information submitted herein, we conclude that the subject device is substantially equivalent to the predicate device.
## 9. Conclusion
The new device and the predicate device are substantially equivalent in the areas of technical characteristics, raw material, and design. The new device does not introduce a fundamentally new scientific technology, and the validation activities demonstrate that the subject device is substantially equivalent to the predicate device.
510(k) Submission
510(k) summary, 9 / 9 page
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.