K243065 · Caption Health, Inc. · QJU · Jan 15, 2025 · Radiology
Device Facts
Record ID
K243065
Device Name
Cardiac Guidance
Applicant
Caption Health, Inc.
Product Code
QJU · Radiology
Decision Date
Jan 15, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2100
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, PCCP
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Cardiac ultrasound image acquisition
—
—
—
—
—
—
—
Indications for Use
The Cardiac Guidance software is intended to assist medical professionals in the acquisition of cardiac ultrasound images. The Cardiac Guidance software is an accessory to compatible general purpose diagnostic ultrasound systems. The Cardiac Guidance software is indicated for use in two-dimensional transthoracic echocardiography (2D-TTE) for adult patients, specifically in the acquisition of the following standard views: Parasternal Long-Axis (PLAX), Parasternal Short-Axis at the Aortic Valve (PSAX-AV), Parasternal Short-Axis at the Mitral Valve (PSAX-MV), Parasternal Short-Axis at the Papillary Muscle (PSAX-PM), Apical 4-Chamber (AP4), Apical 5-Chamber (AP5), Apical 2-Chamber (AP2), Apical 3-Chamber (AP3), Subcostal 4-Chamber (SubC4), and Subcostal Inferior Vena Cava (SC-IVC).
Device Story
Cardiac Guidance is an AI-powered, software-only accessory for diagnostic ultrasound systems; assists medical professionals in acquiring standard 2D-TTE images. Input: real-time ultrasound video stream. Operation: emulates skilled sonographer expertise via AI algorithms to provide real-time feedback. Features: Quality Meter (advises on diagnostic quality); Prescriptive Guidance (directs transducer manipulation); Auto-Capture (triggers recording when diagnostic quality is reached); Save Best Clip (retrospectively saves highest quality clip). Used in clinical settings; provides real-time visual/audio cues to guide probe positioning, pressure, and angulation. Output: optimized diagnostic-quality ultrasound clips. Benefits: enables consistent, high-quality image acquisition, reducing user variability and improving diagnostic accuracy.
Clinical Evidence
No clinical data provided. Safety and performance are supported by software verification and validation testing, including system-level validation and usability testing to ensure adherence to pre-defined acceptance criteria for modified features.
Technological Characteristics
Software-only radiological computer-assisted acquisition guidance system. Operates as an accessory to general-purpose diagnostic ultrasound systems. Features AI-based real-time guidance, quality assessment, and automated image capture. Supports various screen sizes and mobile platforms. Connectivity: integrates with compatible ultrasound systems. Software includes algorithm thresholds, averaging logic, transfer functions, and refresh rate optimizations.
Indications for Use
Indicated for adult patients undergoing 2D-TTE for acquisition of standard cardiac views (PLAX, PSAX-AV, PSAX-MV, PSAX-PM, AP4, AP5, AP2, AP3, SubC4, SC-IVC).
Regulatory Classification
Identification
A radiological acquisition and/or optimization guidance system is a device that is intended to aid in the acquisition and/or optimization of images and/or diagnostic signals. The device interfaces with the acquisition system, analyzes its output, and provides guidance and/or feedback to the operator for improving image and/or signal quality.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed, technical device description, including a detailed description of the impact of any software and hardware on the device's functions, the associated capabilities and limitations of each part, and the associated inputs and outputs.
(ii) A detailed, technical report on the non-clinical performance testing of the subject device in the intended use environments, using relevant consensus standards when applicable.
(iii) A detailed report on the clinical performance testing, obtained from either clinical testing, accepted virtual/physical systems designed to capture clinical variability, comparison to a closely-related device with established clinical performance, or other sources that are justified appropriately. The choice of the method must be justified given the risk of the device and the general acceptance of the test methods. The report must include the following:
(A) A thorough description of the testing protocol(s).
(B) A thorough, quantitative evaluation of the diagnostic utility and quality of images/data acquired, or optimized, using the device.
(C) A thorough, quantitative evaluation of the performance in a representative user population and patient population, under anticipated conditions and environments of use.
(D) A thorough discussion on the generalizability of the clinical performance testing results.
(E) A thorough discussion on use-related risk analysis/human factors data.
(iv) A detailed protocol that describes, in the event of a future change, the level of change in the device technical specifications or indications for use at which the change or changes could significantly affect the safety or effectiveness of the device and the risks posed by these changes. The assessment metrics, acceptance criteria, and analytical methods used for the performance testing of changes that are within the scope of the protocol must be included.
(v) Documentation of an appropriate training program, including instructions on how to acquire and process quality images and video clips, and a report on usability testing demonstrating the effectiveness of that training program on user performance, including acquiring and processing quality images.
(2) The labeling required under § 801.109(c) of this chapter must include:
(i) A detailed description of the device, including information on all required and/or compatible parts.
(ii) A detailed description of the patient population for which the device is indicated for use.
(iii) A detailed description of the intended user population, and the recommended user training.
(iv) Detailed instructions for use, including the information provided in the training program used to meet the requirements of paragraph (b)(1)(iv) of this section.
(v) A warning that the images and data acquired using the device are to be interpreted only by qualified medical professionals.
(vi) A detailed summary of the reports required under paragraphs (b)(1)(ii) and (iii) of this section.
(vii) A statement on upholding the As Low As Reasonably Achievable (ALARA) principle with a discussion on the associated device controls/options.
{0}------------------------------------------------
January 15, 2025
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Caption Health, Inc. Tahir Rizvi Sr. Director of Regulatory Affairs 4 West 4th Avenue Suite 215 San Mateo, California 94402
Re: K243065
Trade/Device Name: Cardiac Guidance Regulation Number: 21 CFR 892.2100 Regulation Name: Radiological Acquisition And/Or Optimization Guidance System Regulatory Class: Class II Product Code: QJU Dated: December 11, 2024 Received: December 16, 2024
Dear Tahir Rizvi:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an
{1}------------------------------------------------
established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rele"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
{2}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# YANNA S. KANG -S
Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K243065
Device Name
Cardiac Guidance
#### Indications for Use (Describe)
The Cardiac Guidance software is intended to assist medical professionals in the acquisition of cardiac ultrasound images. Cardiac Guidance software is an accessory to compatible general purpose diagnostic ultrasound systems.
The Cardiac Guidance software is indicated for use in two-dimensional transthoracic echocardiography (2D-TTE) for adult patients, specifically in the acquisition of the following standard views: Parasternal Long-Axis (PLAX), Parasternal Short-Axis at the Aortic Valve (PSAX-AV), Parasternal Short-Axis at the Mitral Valve (PSAX-MV). Parasternal Short-Axis at the Papillary Muscle (PSAX-PM), Apical 4-Chamber (AP4), Apical 5-Chamber (AP5), Apical 2-Chamber (AP2), Apical 3-Chamber (AP3), Subcostal 4-Chamber (SubC4), and Subcostal Inferior Vena Cava (SC-IVC).
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logos of GE HealthCare and Caption Health. The GE HealthCare logo is on the top, and the Caption Health logo is on the bottom. The GE HealthCare logo consists of a purple circular design with the letters "ge" inside, followed by the text "GE HealthCare" in purple. The Caption Health logo consists of the text "Caption Health" in black.
Image /page/4/Picture/1 description: The image shows the alphanumeric string "K243065" in a bold, sans-serif font. The characters are all uppercase and evenly spaced. The string appears to be a code or identifier.
## 510(k) Summary
Name of Device: Cardiac Guidance Classification Name: Image Acquisition And/Or Optimization Guided By Artificial Intelligence Regulatory Class: II Product Code: QJU Regulation: 21 CFR 892.2100 Predicate Device: Caption Guidance Device Description:
The Cardiac Guidance software is a radiological computer-assisted acquisition guidance system that provides real-time guidance during echocardiography to assist the user capture anatomically correct images representing standard 2D echocardiographic diagnostic views and orientations. This Al-powered, software-only device emulates the expertise of skilled sonographers.
Cardiac Guidance is comprised of several different features that, combined, provide expert guidance to the user. These include:
- Quality Meter: The real-time feedback from the Quality Meter advises the user on the expected ● diagnostic quality of the resulting clip, such that the user can make decisions to further optimize the quality, for example by following the prescriptive guidance feature below.
- Prescriptive Guidance: The prescriptive guidance feature in Cardiac Guidance provides direction to the user to emulate how a sonographer would manipulate the transducer to acquire the optimal view.
- Auto-Capture: The Cardiac Guidance Auto-Capture feature triggers an automatic capture of a clip when the quality is predicted to be diagnostic, emulating the way in which a sonographer knows when an image is of sufficient quality to be diagnostic and records it.
- Save Best Clip: This feature continually assesses clip quality while the user is scanning and, in the event that the user is not able to obtain a clip sufficient for Auto-Capture, the software allows the user to retrospectively record the highest quality clip obtained so far, mimicking the choice a sonographer might make when recording an exam.
#### Intended Use/ Indications for Use:
No differences exist between the subject device of this PCCP and the predicate device with respect to intended use or indications for use. The following summary of intended use/ indications for use is reproduced here for reference:
The Cardiac Guidance software is intended to assist medical professionals in the acquisition of cardiac ultrasound images. The Cardiac Guidance software is an accessory to compatible general purpose diagnostic ultrasound systems.
{5}------------------------------------------------
The Cardiac Guidance software is indicated for use in two-dimensional transthoracic echocardiography (2D- TTE) for adult patients, specifically in the acquisition of the following standard views: Parasternal Long-Axis (PLAX), Parasternal Short-Axis at the Aortic Valve (PSAX-AV), Parasternal Short-Axis at the Mitral Valve (PSAX-MV), Parasternal Short-Axis at the Papillary Muscle (PSAX-PM), Apical 4-Chamber (AP4), Apical 5-Chamber (AP5), Apical 2- Chamber (AP2), Apical 3-Chamber (AP3), Subcostal 4-Chamber (SubC4), and Subcostal Inferior Vena Cava (SC-IVC).
#### Summary of Technological Characteristics compared to the previously-cleared PCCP:
The updated PCCP expands upon the existing list of changes from the most recent 510(k) clearance (K201992).
#### Performance Data and Performance Testing
Safety and performance of the Cardiac Guidance software will be evaluated and verified in accordance with software specifications and applicable performance standards through software verification and validation testing outlined in the submission. Compared to previously-cleared PCCP, two new tests have been incorporated to support validation of the modified device.
The following Table briefly compares the modifications between the previously-cleared PCCP and the subject PCCP of this submission.
| Modification | K201992 PCCP | Subject PCCP |
|-----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Real-Time Guidance<br>for Additional 2D TTE<br>Views | Added cardiac views to standard<br>echocardiography views, focusing<br>on various anatomical regions<br>(e.g., PLAX, PSAX, Apical,<br>Subcostal, Suprasternal). | Expands the list of views in previously<br>authorized PCCP by adding additional<br>standard cardiac subset views. These<br>additional cardiac ultrasound views are<br>focused on enhanced diagnostic imaging and<br>coverage of specific anatomical regions. |
| Labeling compatibility<br>of new compatible<br>ultrasound systems<br>and UI Enhancement | Ensured software compatibility<br>with ultrasound systems of similar<br>screen sizes. | Builds on compatibility with additional UI/UX<br>improvements, incl. support for various screen<br>sizes such as mobile platforms. |
| Non-expert Validation | Not included. | Adds standalone test protocol to enable<br>validation of modified device performance by<br>the intended user groups, ensuring<br>equivalency to the original device based on<br>predefined clinical endpoints. |
| Human Factors Test | Not included. | Introduces Human Factors Testing to evaluate<br>user interactions and ensure critical tasks<br>remain unaffected by UI/UX modifications. |
The revised PCCP does not introduce new or different questions of safety and effectiveness compared to the predicate device. Accordingly, the Cardiac Guidance software is determined to be substantially equivalent.
{6}------------------------------------------------
#### Summary of planned modifications, testing methods, validation activities and performance requirements
This 510(k) submission outlines future algorithm improvements under an updated Predetermined Change Control Plan (PCCP) and seek authorization of this updated PCCP. The plan details anticipated software modifications, methods and validation criteria that will be utilized to implement those modifications in a controlled manner and ensure the subject device remains as safe and effective as the predicate. Anticipated software modifications are discussed in the main submission and are summarized in the table below, as well as a modification protocol describing the verification activities that will support these planned modifications. The modification protocol incorporates impact assessment considerations and specifies requirements for data management, including data sources, collection, storage, and sequestration, as well as documentation and data segregation/re-use practices.
The test methods specified in the PCCP establish substantial equivalence to the predicate device, and include sample size determination, analysis methods, and acceptance criteria. To ensure validation test datasets are representative of the intended use population, each will meet minimum demographic requirements. Caption Health shall communicate product changes made under the scope of this PCCP to end-users via customer and software update notifications, as well through updated product labeling.
| Modification | Rationale | Testing Methods |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Retraining/optimization/<br>modification of core algorithm(s) | These changes enhance model<br>robustness, reduce bias, improve<br>generalizability, and enable<br>Caption Health to respond<br>effectively real-world feedback. | Repeating verification tests and<br>the system level validation test to<br>ensure the pre-defined acceptance<br>criteria are met. |
| Real-time guidance for additional<br>2D TTE views: This modification<br>expands the list of views by adding<br>additional cardiac subset views | These changes enhance the<br>system's diagnostic capabilities by<br>providing more focused, specific<br>views that assist expert clinicians<br>in making more accurate cardiac<br>assessments, improving overall<br>usability. | Repeating verification tests and<br>two system level validation tests<br>including usability testing to<br>ensure the pre-defined acceptance<br>criteria are met for the additional<br>views. |
| Optimization of the core<br>algorithm(s) implementation in<br>the software such as optimization<br>of algorithm thresholds, averaging<br>logic, transfer functions,<br>frequency, and refresh rate | These optimizations improve the<br>performance of the algorithm,<br>making it more efficient and<br>accurate without changing its<br>fundamental functionality, ensure<br>consistency across different<br>devices and enhance the usability<br>of the device | Repeating relevant verification<br>test(s) and the system level<br>validation test to ensure the pre-<br>defined acceptance criteria are<br>met. |
| Addition of new types of<br>prescriptive guidance for existing<br>views such as patient positioning<br>and breathing guidance, guidance | These changes provide additional<br>real-time feedback to users<br>enhances system's utility and user<br>experience in obtaining diagnostic | Repeating relevant verification<br>tests and two system level<br>validation tests including usability |
A summary of these software modifications, rationale and test methods are presented in the table below:
{7}------------------------------------------------
| that combines two or more<br>physical probe movements,<br>pressure and probe sliding/angling<br>and addition of existing guidance<br>types to all existing views in the<br>device and future views | quality images. | testing to ensure the pre-defined<br>acceptance criteria are met. |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Labeling compatibility of software<br>with compatible ultrasound<br>systems for various screen sizes<br>including mobile platforms or/and<br>change in the user interface<br>features, such as addition of<br>Prescriptive Guidance audio and<br>configurability of guidance<br>features. | These changes enable support for<br>a broader range of devices and are<br>intended to enable an enhanced<br>user experience without<br>compromising critical tasks. | Repeating relevant verification<br>tests and the system level<br>validation test including usability<br>testing to ensure the pre-defined<br>acceptance criteria are met. |
### Conclusions:
The current iteration of the Cardiac Guidance software is as safe and effective as the previous iteration of such software. The Cardiac Guidance software is substantially equivalent in intended use, design, principles of operation, technological characteristics, and safety features to the predicate device. There are no different questions of safety and/or effectiveness introduced by the Caption Guidance software when used as intended. Thus, the Cardiac Guidance software is substantially equivalent. The activities proposed within the Modification Protocol will continue to reasonably ensure the safe and effectiveness of the product, and all modifications will maintain the device within stated intended use and indications for use.
The purpose of this 510(k) submission is to seek a focused review and clearance of an updated predetermined change control plan (PCCP) that outlines anticipated modifications to the methods that will be utilized to implement those modifications in a controlled manner while maintaining safety and efficacy. This current submission expands on the predetermined change control plan (PCCP) previously cleared under K201992, which was itself an update to the initial PCCP authorized as part of DEN190040.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.