The Pulse Oximeter FS20P is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adolescent, child and infant patients in hospitals, hospital-type facilities and homecare. The device is not intended for continuous monitoring, use during motion or use with low perfusion. The device is intended for reuse. The device is wearing on fingertips while using. The Pulse Oximeter FS20C and FS10C is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult patients in hospitals, hospital-type facilities and homecare. The device is not intended for continuous monitoring, use during motion or use with low perfusion. The device is intended for reuse. The device is wearing on fingertips while using.
Device Story
Handheld, battery-powered pulse oximeter; models FS20P, FS20C, FS10C. Device uses transmittance optical sensor (red 660nm, infrared 905nm) to measure SpO2 and pulse rate via fingertip. Signal collection/processing module (MCU) calculates values; results displayed on OLED/LED screen. Used in hospitals, hospital-type facilities, and homecare by clinicians or patients for spot-checking. Not for continuous monitoring, motion, or low perfusion. Provides immediate physiological data to assist in clinical assessment of oxygenation status.
Clinical Evidence
Clinical studies conducted per ISO 80601-2-61. FS20P study: 10 healthy adult volunteers (ages 22-30, Fitzpatrick 1-6), ARMS 1.70% (70-100% range). FS20C study: 13 healthy adult volunteers (ages 22-30, Fitzpatrick 1-6), ARMS 1.71% (70-100% range). Comparator: arterial blood CO-Oximetry. Bench testing included IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, and ISO 80601-2-61.
Technological Characteristics
Battery-powered (2xAAA). Transmittance optical sensor (660nm/905nm). Materials: ABS (enclosure/button), PC (lens), Silica gel (fingertip cushion). Connectivity: None. Sterilization: Not specified. Software: Embedded MCU for signal processing. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 80601-2-61.
Indications for Use
Indicated for spot-checking SpO2 and pulse rate. FS20P: adolescent, child, and infant patients. FS20C/FS10C: adult patients. Settings: hospitals, hospital-type facilities, homecare. Contraindications: continuous monitoring, use during motion, use with low perfusion.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 2, 2025
Hunan Accurate Bio-Medical Technology Co., Ltd. % Jie Yang Consultant Chonconn Medical Device Consulting Co., Ltd. Room 504, Block C, No. 1029 Nanhai Avenue, Nanshan District Shenzhen, Guangdong 518067 China
Re: K243049
Trade/Device Name: Pulse Oximeter (FS20P); Pulse Oximeter (FS20C); Pulse Oximeter (FS10C) Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DOA Dated: September 27, 2024 Received: December 6, 2024
Dear Jie Yang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bradley Q. Quinn -S
Bradley Quinn Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
### K243049
Device Name
Pulse Oximeter (FS20P); Pulse Oximeter (FS20C); Pulse Oximeter (FS10C)
Indications for Use (Describe)
The Pulse Oximeter FS20P is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adolescent, child and infant patients in hospitals, hospital-type facilities and homecare. The device is not intended for continuous monitoring, use during motion or use with low perfusion. The device is intended for reuse. The device is wearing on fingertips while using.
The Pulse Oximeter FS20C and FS10C is a handheld non-invasive device intended for spotchecking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult patients in hospitals, hospital-type facilities and homecare. The device is not intended for continuous monitoring, use during motion or use with low perfusion. The device is intended for reuse. The device is wearing on fingertips while using.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) Summary
## Prepared in accordance with the requirements of 21 CFR Part 807.92
## Prepared Date: 2024/09/27
#### 1. Submission sponsor
Name: Hunan Accurate Bio-Medical Technology Co., Ltd. Address: Accurate Industrial Park, No.108, Zhixian Road, Xuelian Community, Xueshi Street of Yuelu District, 410208 Changsha, Hunan Province, PEOPLE'S REPUBLIC OF China Contact person: Zhang Li Title: General Manager E-mail: regulation@accbiomed.com Tel: +86 731-8559853
#### 2. Submission correspondent
Name: Chonconn Medical Device Consulting Co., Ltd.
Address: Room 504, Block C, No. 1029 Nanhai Avenue, Nanshan District, Shenzhen, Guangdong, P.R. China
Contact person: Yang Jie
E-mail: yangjie(@chonconn.com
Tel: +86-755 33941160
| Trade/Device Name | Pulse Oximeter |
|-------------------|---------------------------------|
| Model | FS20P, FS20C,FS10C |
| Common Name | Fingertip Pulse Oximeter |
| Regulatory Class | Class II |
| Classification | 21CFR 870.2700 / Oximeter / DQA |
| Submission type | Traditional 510(K) |
#### Subiect Device Information 3.
#### 4. Predicate Device
Manufacturer: Beijing Choice Electronic Technology Co., Ltd. Device name: Pulse Oximeter, Model: MD300C19 510(K) Number: K230172
#### 5. Device Description
The subject device Pulse Oximeter is a battery powered device, which can mainly detect and display the measured oxyhemoglobin saturation (SpO2) and pulse rate (PR) value. Place one fingertip into the photoelectric sensor for diagnosis and the pulse rate and oxygen saturation will appear on the display. The
{5}------------------------------------------------
device is normally applied to infants, children, adolescents and adult in hospitals, hospital-type facilities and homecare.
The subject device is composed of following components to achieve the above detection process: power supply module, detector and emitter LED, signal collection and process module (MCU), OLED/LED display screen, user interface and button control circuit.
#### 6. Intended use & Indication for use
The Pulse Oximeter FS20P is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adolescent, child and infant patients in hospitals, hospital-type facilities and homecare. The device is not intended for continuous monitoring, use during motion or use with low perfusion. The device is intended for reuse. The device is wearing on fingertips while using.
The Pulse Oximeter FS20C and FS10C is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult patients in hospital-type facilities and homecare. The device is not intended for continuous monitoring, use during motion or use with low perfusion. The device is intended for reuse. The device is wearing on fingertips while using.
| ITEM | Proposed Device<br>Pulse Oximeter<br>Model: FS20P,<br>FS20C,FS10C | Predicate Device<br>Pulse Oximeter<br>Model: MD300C19<br>K230172 | Comparison Result |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| Manufacture | Hunan Accurate Bio-Medical<br>Technology Co., Ltd. | Beijing Choice Electronic<br>Technology Co., Ltd. | Same |
| Indications for Use | The Pulse Oximeter FS20P is<br>a handheld non-invasive<br>device intended for spot-<br>checking of oxygen saturation<br>of arterial hemoglobin<br>(SpO2) and Pulse Rate of<br>adolescent, child and infant<br>patients in hospitals, hospital-<br>type facilities and<br>homecare. The device is not<br>intended for continuous<br>monitoring, use during<br>motion or use with low<br>perfusion. The device is | The Pulse Oximeter is a<br>handheld non-invasive device<br>intended for spot-checking of<br>oxygen saturation of arterial<br>hemoglobin (SpO2) and Pulse<br>Rate of adult, adolescent child<br>and infant patients in hospitals,<br>hospital-type facilities and<br>homecare. The device is not<br>intended for continuous<br>monitoring, use during motion<br>or use with low perfusion. The<br>device is intended for reuse. | Same |
| | | | |
| | intended for reuse. The<br>device is wearing on<br>fingertips while using.<br><br>The Pulse Oximeter FS20C<br>and FS10C is a handheld non-<br>invasive device intended for<br>spot-checking of oxygen<br>saturation of arterial<br>hemoglobin (SpO2) and Pulse<br>Rate of adult patients in<br>hospitals, hospital-type<br>facilities and homecare. The<br>device is not intended for<br>continuous monitoring, use<br>during motion or use with<br>low perfusion. The device is<br>intended for reuse. The<br>device is wearing on<br>fingertips while using. | The device is wearing on<br>fingertips while using. | |
| Population | FS20C and FS10C: adult<br>FS20P: adolescent, child and<br>infant patients. | adult, adolescent, child and<br>infant patients | Similar |
| Components | Power supply module,<br>detector and emitter LED,<br>signal collection and<br>processor module, display<br>module, user interface and<br>button control. | Power supply module, detector<br>and emitter LED, signal<br>collection and processor<br>module, display module, user<br>interface and button control. | Same |
| Type of SpO2<br>Sensor | Transmittance Optical Sensor | Transmittance Optical Sensor | Same |
| Application<br>Site | Finger | Finger | Same |
| Measurement | Red (660 nm)<br>Infrared (905 nm) | Red (660 nm)<br>Infrared (905 nm) | Same |
| Display Type | OLED or LED | LED | Similar |
| Power<br>Supply | 2*AAA alkaline batteries | 2*AAA alkaline batteries | Same |
| Display Data | SPO2 , PR | SPO2 , PR | Same |
| SPO2 | FS20C,FS20P:<br>SpO2 Display Range:<br>0%~100%<br>Measurement range:<br>70~100%<br>Accuracy: 70%~100%: ±2%;<br>0~69% no definition<br>FS10C:<br>SpO2 Display Range:<br>0%~99%<br>Measurement range:<br>70~99%<br>Accuracy: 70%~99%: ±2%;<br>0~69% no definition | SpO2 Display Range:<br>0%~100%<br>Measurement range:<br>70%~100%<br>Accuracy: 70%~100%: ±2%;<br>0~69% no definition | Similar |
| PR | PR Display Range<br>25~250bpm<br>Measurement range:<br>25~250bpm<br>Resolution: 1bpm<br>Accuracy:±3bpm; | PR Display Range<br>0~255bpm<br>Measurement range:<br>30~250bpm<br>Resolution: 1bpm<br>Accuracy: 30~99bpm, ±2bpm;<br>100~250bpm, ±2% | Similar |
| Environment<br>Requirements | Operating Temperature:<br>5~40°C<br>Ambient Humidity:<br>10%~95%, no condensation<br>Storage/Transportation: -<br>20°C~60°C Ambient<br>Humidity: 10%~95% no<br>condensation | Operating Temperature:<br>0~40°C<br>Ambient Humidity: 15%~93%,<br>no condensation<br>Storage/Transportation: -<br>25°C~70°C Ambient Humidity:<br><=93% no condensation | Similar |
| Contacting<br>Material | Enclosure/Battery cover/<br>Button: ABS<br>Lens: PC<br>Fingertip cushion: Silica gel | Enclosure/Battery cover: ABS<br>Lens: PMMA<br>Button/Fingertip cushion:<br>Silica gel pad | Similar |
| Electrical<br>Safety | Conformed to IEC60601-1,<br>IEC 60601-1-11 | Conformed to IEC60601-1,<br>IEC 60601-1-11 | Same |
#### 7. Comparison to the Predicate Device
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
#### 8. Non-clinical Data
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility testing
The biocompatibility evaluation for the Pulse Oximeter was conducted in accordance with the FDA Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1. "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
## Bench testing
The Pulse Oximeter has been tested according to the following standards:
- IEC 60601-1-2005+CORR.1:2006+CORR.2:2007+A1:2012, Medical Electrical Equipment-. Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2:2014. Medical electrical equipment- Part 1-2: General requirements for basic . safety and essential performance- Collateral standard: Electromagnetic compatibility-Requirements and tests
- . IEC 60601-1-11 Edition 2.0 2015-01Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- ISO 80601-2-61: 2017 Medical Electrical Equipment Part 2-61: Particular Requirements for . Basic Safety and Essential Performance of Pulse Oximeter Equipment.
## Clinical data
Clinical studies were conducted to verify the accuracy of proposed device. The clinical studies were conducted per following standards:
- ISO 80601-2-61: 2017 Medical Electrical Equipment - Part 2-61: Particular Requirements for Basic Safety and Essential Performance of Pulse Oximeter Equipment.
- Pulse Oximeters-Premarket Notification Submissions: Guidance for Industry and Food and Drug Administration Staff
After Institutional Review Board (IRB) approval, 10 healthy female adult volunteer subjects (ages 22-30yr with skin tones of Fitzpatrick 1-6) were included in the study to evaluate the SpO2 accuracy performance of the FS20P Pulse Oximeter during stationary (non-motion) conditions over a wide range of arterial blood
{9}------------------------------------------------
oxygen saturation levels as compared to arterial blood CO-Oximetry.
The SpO2 accuracy performance results showed FS20P Pulse Oximeter to have an ARMS of 1.70% during steady state conditions over the range of 70-100%.
After Institutional Review Board (IRB) approval, 13 healthy adult volunteer subjects (ages 22-30yr with skin tones of Fitzpatrick 1-6) were included in the study to evaluate the SpO2 accuracy performance of the FS20C Pulse Oximeter during stationary (non-motion) conditions over a wide range of arterial blood oxygen saturation levels as compared to arterial blood CO-Oximetry.
The SpO2 accuracy performance results showed FS20C Pulse Oximeter to have an ARMS of 1.71% during steady state conditions over the range of 70-100%.
## 9. Conclusion
Performance testing and compliance with voluntary standards demonstrate that the proposed subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.