← Product Code [OHT](/productcode/OHT) · K243030

# Diamond Air+ (UI04A, UI04B, UI04C, UI04M, UI04G, UI04W) (K243030)

_Ulike Co., Ltd. · OHT · Oct 25, 2024 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K243030

## Device Facts

- **Applicant:** Ulike Co., Ltd.
- **Product Code:** [OHT](/productcode/OHT.md)
- **Decision Date:** Oct 25, 2024
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Diamond Air+ (Model: UI04A, UI04B, UI04C,UI04M, UI04G, UI04W) is an overthe-counter device intended for removal of unwanted body and/or facial hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

## Device Story

Diamond Air+ is an over-the-counter, light-based hair removal device for home use. It utilizes Intense Pulsed Light (IPL) technology to heat hair roots and inhibit regrowth. The device features a Xenon lamp as the light source and includes a skin proximity sensor to ensure safe contact before pulse emission; it cannot be triggered without proper skin application. Users operate the device to treat areas such as underarms, bikini line, arms, and legs. It offers 5 levels of output energy. The device provides a non-invasive method for long-term hair reduction, allowing patients to manage unwanted hair at home.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on identical technological characteristics and specifications to the predicate device.

## Technological Characteristics

Intense Pulsed Light (IPL) technology; Xenon arc flashlamp energy source; 550-1200nm wavelength spectrum; 6 J/cm2 maximum fluence; 5 energy levels; skin proximity sensor for safety; dimensions 60x38x169.86mm; over-the-counter home-use device.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Diamond Air+(UI04A, UI04B, UI04C) ([K221553](/device/K221553.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 25, 2024

Ulike Co., Ltd. Blue Yang Registration Director 2, Myeongdong, 6-gil, Jung-gu, Seoul, Korea Seoul. Korea, South

## Re: K243030

Trade/Device Name: Diamond Air+ (UI04A, UI04B, UI04C, UI04M, UI04G, UI04W) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: September 27, 2024 Received: September 27, 2024

Dear Blue Yang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{1}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by TANISHA TANISHA L. HITHE -S L. HITHE -S 20:07:43 -04'00'

Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

Submission Number (if known)

K243030

Device Name

Diamond Air+ (UI04A, UI04B, UI04C, UI04C, UI04M, UI04G, UI04W)

#### Indications for Use (Describe)

Diamond Air+ is an over-the-counter device intended for removal of unwanted body and/or facial hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the longterm, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X | Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary - K243030

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92. Prepared 10/24/2024.

## 1. Submitter's Information

| Sponsor        |                                             |
|----------------|---------------------------------------------|
| Company Name   | Ulike Co., Ltd.                             |
| Address        | 2, Myeongdong, 6-gil, Jung-gu, Seoul, Korea |
| Contact Person | Ms. Blue Yang                               |
| Title          | Regulatory Director                         |
| Email          | blue@ulike.com                              |

## 2. Subject Device Information

| Trade Name          | Diamond Air+ (UI04A, UI04B, UI04C, UI04M, UI04G,<br>UI04W)           |
|---------------------|----------------------------------------------------------------------|
| Classification Name | Laser Surgical Instruments for Use in General and<br>Plastic Surgery |
| Review Panel        | General & Plastic Surgery                                            |
| Product Code        | OHT                                                                  |
| Regulation Class    | 2                                                                    |
| Regulation Number   | 878.4810                                                             |

## 3. Predicate Device Information

Predicate device I :

| Predicate Device | I                                 |
|------------------|-----------------------------------|
| Sponsor          | Ulike Co., Ltd.                   |
| Device Name      | Diamond Air+(UI04A, UI04B, UI04C) |

{5}------------------------------------------------

| 510(k) Number        | K221553  |
|----------------------|----------|
| Product Code         | OHT      |
| Regulation<br>Number | 878.4810 |
| Regulation Class     | 2        |

#### 4. Device Description

Diamond Air+, Model: UI04A, UI04B, UI04C, UI04M, UI04G, UI04W is a light-based device for long-term hair removal. It is intended for the removal of unwanted hair and permanent reduction in hair regrowth. Ideal body areas include the underarms, bikini line, arms and legs. The device used the IPL technology with lower energy level, including 5 Levels of output energy. Intense Pulsed light technology can achieve long-term hair removal results at a fraction of the energy level used in other light -based hair removal equipment.

The size of the device is about 60*38*169.86mm (W x D x H). The device incorporates Intense Pulse Light (IPL) technology. The purpose of the light is to heat the root where the hair grows.

The device contains a Xenon Lamp and a skin proximity sensor to detect appropriate skin contact. If the Diamond Air+ is not properly applied to the treatment area (in full contact with the skin), the device cannot be triggered a pulse emitting.

## 5. Intended Use / Indications for Use

The Diamond Air+ (Model: UI04A, UI04B, UI04C,UI04M, UI04G, UI04W) is an overthe-counter device intended for removal of unwanted body and/or facial hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

{6}------------------------------------------------

#### 9. Comparison to predicate device and conclusion

The technological characteristics, features, specifications, materials, and intended use of the Diamond Air+, model: UI04A, UI04B, UI04C, UI04M, UI04G, UI04W are identical to the cited predicate. The sole difference between the subject and predicate devices are the colors of the device housing.

The Diamond Air+ relies on the same technology as both predicate device: Intense Pulsed Light (IPL). Their energy source is the same: Xenon Arc Flashlamp.

The safety and efficacy of IPL treatment for hair reduction are governed by the following parameters:

· Wavelength of the emitted light (spectrum): Defines the interaction with specific chromophores (the part of the molecule responsible for its color) such as melanin, hemoglobin and water. The Diamond Air+ and the predicate device utilize the same spectrum(550-1200nm).

· Fluence/flux - defines the energy per area (e.g. joules per cm²) for the treatment. The Diamond Air+ and the predicate device deliver the same maximum energy (6 J/cm2).

#### Conclusion:

The subject device "Diamond Air+, Model: UI04A, UI04B, UI04C, UI04M, UI04G, UI04W" is substantial equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K243030](https://fda-staging.innolitics.com/device/K243030)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).