The Silicone Urethral Catheter is passed through the urethra during urinary catheterization and into the bladder to drain urine, or for routine post-operative drainage and bladder irrigation. It is intended for use by male, female and pediatric patients (2 years old to less than 12 years old).
Device Story
Silicone Urethral Catheter is a sterile, single-use, indwelling device inserted through the urethra into the bladder. It consists of a shaft, drainage funnel, inflation funnel, optional irrigation funnel, balloon, and valve. The balloon is inflated with sterile liquid to anchor the catheter in the bladder. Used in healthcare facilities by clinicians for urine drainage or bladder irrigation. The device is available in various configurations (2-way, 3-way, Tiemann tip) and sizes for pediatric and adult patients. Indwelling time is limited to 30 days. The device facilitates patient management by providing a reliable pathway for bladder drainage or irrigation, reducing the need for repeated catheterization.
Clinical Evidence
No clinical or animal testing was performed. Substantial equivalence is supported by bench performance testing (including balloon integrity, flow rate, tensile strength, and kink stability) and biocompatibility testing (cytotoxicity, sensitization, irritation, systemic toxicity, implantation, and pyrogenicity) conducted in accordance with ISO 10993-1 and relevant FDA guidance.
Technological Characteristics
Materials: Silicone. Sensing/Actuation: Mechanical balloon inflation. Dimensions: 6-24Fr (2-way), 16-24Fr (3-way). Connectivity: None. Sterilization: Ethylene Oxide (ISO 11135:2014). Standards: ASTM F623-19, ISO 20696:2018, ISO 10993-1.
Indications for Use
Indicated for male, female, and pediatric patients (2-12 years old) requiring urinary catheterization for urine drainage, post-operative drainage, or bladder irrigation.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 16, 2025
Guangdong Ecan Medical Co., Ltd.
Summer Wang
RA Engineer
Building 1, No. 222, Xindu Road, Chengjiao Street,
Conghua District
Guangzhou, Guangdong 510920
CHINA
Re: K243011
Trade/Device Name: Silicone Urethral Catheter (Silicone Urethral Catheter)
Regulation Number: 21 CFR 876.5130
Regulation Name: Urological Catheter and accessories
Regulatory Class: II
Product Code: EZL
Dated: May 23, 2025
Received: May 23, 2025
Dear Summer Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243011 - Summer Wang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K243011 - Summer Wang
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica K. Nguyen -S
Jessica K. Nguyen, Ph.D.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K243011
Device Name
Silicone Urethral Catheter
Indications for Use (Describe)
The Silicone Urethral Catheter is passed through the urethra during urinary catheterization and into the bladder to drain urine, or for routine post-operative drainage and bladder irrigation. It is intended for use by male, female and pediatric patients (2 years old to less than 12 years old).
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{4}
K243011
Page 1 of 5
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
Date of Preparation: May 23, 2025
## 1. Applicant:
510(k) Owner’s Name: Guangdong Ecan Medical Co., Ltd.
Address: Building 1, No. 222, Xindu Road, Chengjiao Street, Conghua District, Guangzhou City, Guangdong Province, 510920, P.R. China
Contact Person: Liuyuan Han
Title: General Manager
Email: hanliuyuan@ecanmed.com
Tel: +86 020-37501359
Fax: +86 020-37501359
## 2. Submission Correspondent:
Contact Person: Summer Wang
Email: 1135770505@qq.com
Tel: +86-20-84558599
Fax: +86-20-84558599
## 3. Device Information
Device Name: Silicone Urethral Catheter
Common Name: Urethral Catheter, Foley Catheter
Regulation Number: 21 CFR 876.5130
Classification Name: Urological Catheter and Accessories
Review Panel: Gastroenterology/Urology
Classification: II
Product Code: EZL
## 4. Predicate and Reference Device Identification
Predicate: K233094 — Wellead® Latex Foley Catheter
Predicate device was not subjected to any design related recall.
## 5. Indications for Use
The Silicone Urethral Catheter is passed through the urethra during urinary catheterization and into the bladder to drain urine, or for routine post-operative drainage and bladder irrigation. It is intended for use by male, female and pediatric patients (2 years old to less than 12 years old).
## 6. Device Description
The Silicone Urethral Catheter is inserted through the urethra during urinary catheterization and goes into the bladder for urine drainage or irrigation. It consists of shaft, drainage funnel,
Traditional 510(k) for Disposable Endotracheal Tube
{5}
K243011
Page 2 of 5
inflation funnel, irrigation funnel (if present), balloon and valve. The balloon of the catheter can be inflated and fixed the catheter in place after being injected with a sterile liquid.
The device is made of silicone and available in different types (e.g., male, female, Tiemann, 2-way, 3-way etc.) and sizes for adult and pediatric use.
The device is for single use and provided with sterile. The indwelling time is no more than 30 days.
Table of Variants
| Type | Size |
| --- | --- |
| 2-way Pediatric | 6Fr, 8 Fr, 10 Fr |
| 2-way Female | 12Fr, 14 Fr, 16 Fr, 18 Fr, 20 Fr, 22 Fr, 24 Fr |
| 2-way Male (Standard) | 12Fr, 14 Fr, 16 Fr, 18 Fr, 20 Fr, 22 Fr, 24 Fr |
| 2-way Tiemann | 16 Fr, 18 Fr, 20 Fr, 22 Fr, 24 Fr |
| 3-way Male (Standard) | 16 Fr, 18 Fr, 20 Fr, 22 Fr, 24 Fr |
# 7. Substantial Equivalence—Comparison to Predicate Devices
A side by side comparison of the proposed device and the predicate device are provided below.
Table 1 - Comparison Table
| Comparison between proposed device and predicate device | | | |
| --- | --- | --- | --- |
| Comparison Items | Proposed Device | Predicate Device | Discussion of Differences |
| Device Name | Silicone Urethral Catheter | Wellead® Latex Foley Catheter | --- |
| 510k Number | --- | K233094 | --- |
| Product Code | EZL | EZL | Same |
| Regulation Number | 21 CFR 876.5130 | 21 CFR 876.5130 | Same |
| Regulatory Class | Class II | Class II | Same |
| Classification Name | Urological Catheter and Accessories | Urological Catheter and Accessories | Same |
| Indications for Use | The Silicone Urethral Catheter is passed through the urethra during urinary catheterization and into the bladder to drain urine, or for routine post-operative drainage and bladder irrigation. It is intended for use by male, female and pediatric patients (2 years old to less than 12 years old). | Latex Foley Catheter is inserted into the bladder through urethra and are indicated for routine urine drainage or for routine post-operative drainage and bladder irrigation. | Same.
Both of them are intended to be inserted through the urethra to the bladder for urine drainage. |
| Patient Population | Pediatric, Male and Female | Pediatric, Male and Female | Same |
| Device Type (Lumen) | 2-way
3-way | 2-way
3-way | Same |
| Patient Contact Materials | Shaft: Silicone
Tip: Silicone
Balloon: Silicone | Shaft: Latex(silicone coated)
Tip: Latex
Balloon: Latex | ※ Different |
| Catheter French Size (Fr) | 2-way: 6-24Fr
3-way: 16-24Fr | 6-30Fr | Same.
The proposed device's sizes are included within the size range of the predicate device. |
| Balloon | 2-way: Yes | 2-way: Yes | Same |
Traditional 510(k) for Disposable Endotracheal Tube
{6}
K243011
Page 3 of 5
| | 3-way: Yes | 3-way: Yes | |
| --- | --- | --- | --- |
| Balloon Volume | 3mL | 3mL | Same.
The proposed device's balloon volumes are included within the range of the predicate device. |
| | 3-5mL | 3-5mL | |
| | 5mL | 5mL | |
| | 5-10mL | 5-10mL | |
| | 10mL | 10mL | |
| | 5-15mL | 5-15mL | |
| | 15-20mL | 15mL | |
| | 20-30mL | 15-30mL | |
| | 20mL | 20mL | |
| | 30mL | 30mL | |
| Tip Shape | Straight round and Tiemann (Coude tip) | Straight and Coude | Same |
| Stylet in Pediatric Type | Yes (for pediatric sizes) | Yes (for pediatric sizes) | Same |
| | Not for adult sizes | Not for adult sizes | |
| Duration of Use | Less than 30 days | less than 30 days | Same |
| Use Environment | Health care facilities | Health care facilities | Same |
| Target Population | Pediatric, Male and Female | Pediatric, Male and Female | Same |
| Sterilization Method | EtO Sterilized | EtO Sterilized | Same |
| Packaging | Peel pack comprises paper and film;
Corrugated board outer case. | Peel pack comprises paper and film;
Corrugated board outer case. | Same |
| Use | Single Use, Disposable | Single Use, Disposable | Same |
| Prescription Use | Yes | Yes | Same |
| Biocompatibility | Cytotoxicity
Sensitization
Irritation (intracutaneous reactivity)
Acute systemic toxicity
Subacute toxicity
Implantation effects
Pyrogenicity
Conforms to ISO 10993-1 | Conforms to ISO 10993-1 | Same |
## Similarities Between Proposed and Predicate Device
The proposed Silicone Urethral Catheters have the same technological characteristics as the predicate device (K233094 — Wellead® Latex Foley Catheter). The proposed and predicate device are based on the following same technological elements:
- Same intended use
- Same Indications for Use
- Same operating principals
- Same patient population
- Same Environment of Use
- Composed of biocompatible materials
- Provided sterile for single-use
- Ethylene Oxide Sterilized
## Differences Between Proposed and Predicate Device
### Materials
The proposed device is made from Silicone material while the predicate device is made from Latex (silicone coated). Although the materials are not identical, but they both are
Traditional 510(k) for Disposable Endotracheal Tube
{7}
K243011
Page 4 of 5
manufactured from materials that meet all the requirements of biocompatibility, the materials in contact were tested as per ISO 10993-1.
Therefore, the material difference does not raise different questions of safety and effectiveness than the predicate device.
As evidenced by the above table, both the proposed and the predicate device have same Indications for Use and technological characteristics. The minor difference in the device does not introduce new issues of safety and efficacy.
Therefore, the proposed Silicone Urethral Catheter is substantially equivalent to the predicate device (K233094 — Wellead® Latex Foley Catheter). The proposed device has the same classification information, same indications for use and technological characteristics as compared to the predicate device.
# 8. Summary of Non-Clinical Performance Testing
Non-clinical tests were conducted to verify that the proposed device met all design specifications as was same to the predicate device.
Device performance data was provided in support of the substantial equivalence determination.
# 1) Bench Testing
The following bench performance testing were conducted per ASTM F623-19, ISO 20696:2018 and the FDA guidance document, 'Conventional Foley Catheters - Performance Criteria for Safety and Performance Based Pathway'.
- Balloon Size and Shaft Size
- Flow Rate
- Balloon Integrity (Resistance to Rupture)
- Inflated Balloon Response to Traction
- Balloon Volume Maintenance
- Deflation Reliability (Failure to Deflate)
- Surface Finish (Appearance)
- Dimensions
- Strength
- Connector Security
- Balloon safety
- Corrosion resistance
- Kink stability
- Peak Tensile Force
- Stylet Pull-out Force
Testing datas and results are included in this submission, and demonstrated that the Silicone Urethral Catheter meets all the pre-determined testing and acceptance criteria.
# 2) Biocompatibility
Biocompatibility testing was conducted based upon ISO 10993-1 (2018): Biological evaluation of medical devices – Part 1: "Evaluation and testing within a risk management process" and FDA Guidance document (2023), "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management
Traditional 510(k) for Disposable Endotracheal Tube
{8}
K243011
Page 5 of 5
process" FDA". The Silicone Urethral Catheter is surface device in contact with mucosal tissue for prolonged exposure ( $>24$ h to $30\mathrm{d}$ ) and following tests were conducted in accordance with the relevant ISO 10993 standards:
- Cytotoxicity
- Intracutaneous
Skin Sensitization
- Acute Systemic Toxicity
- Subacute Toxicity
- Implantation
- Pyrogenicity
Biocompatibility testing reports are included in this submission, and demonstrated that the device components that are in contact with the patient are biocompatible. All evaluation acceptance criteria were met.
## 3) Sterilization and Shelf-Life
The Silicone Urethral Catheter was validated to be sterilized using ethylene oxide in accordance with ISO 11135:2014.
To support the claimed shelf life, the Silicone Urethral Catheter was subjected to bench performance testing, transportation testing and package integrity testing to demonstrate that the product and package would be undamaged throughout the product life and maintain device sterility.
## Conclusions Drawn from the Non-Clinical Testing
The results of these tests demonstrate that the device is as safe, as effective, and performs as well as the identified predicate and support a determination of substantial equivalence.
## 9. Clinical Testing and animal testing
Clinical and animal testing were not performed for Silicone Urethral Catheter as part of the premarket Notification requirements for this 510(k) submission and the subject of this premarket submission, Silicone Urethral Catheter, did not require clinical and animal studies to support substantial equivalence.
## 10. Conclusion
Based on the information presented in this submission, it can be concluded that the proposed device Silicone Urethral Catheter is substantially equivalent to the predicate device (K233094 — Wellead® Latex Foley Catheter).
Traditional 510(k) for Disposable Endotracheal Tube
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.