The VSH101 is designed to record, transmit, and store single channel electrocardiogram (ECG) data via Bluetooth communication to compatible Bluetooth enabled devices. The device is intended for use by healthcare professionals, individuals with known or suspected cardiac conditions, and health-conscious users. The ECG data serves as a supplementary source of patient information and is not intended for automated analysis.
Device Story
VitalSigns 1-Lead Holter (VSH101) is an ambulatory ECG monitoring system consisting of a chest-worn electrode patch (VSP101), an ECG recorder host (VSH101), and a mobile application (VSHealth). The device captures single-channel ECG signals via the electrode patch, which are transmitted via Bluetooth Low Energy to the VSHealth app for display, recording, and storage. The system supports up to 24 hours of continuous recording in home or clinical environments. It is operated by healthcare professionals or patients. The output is raw ECG data intended for manual interpretation by a clinician to supplement patient information; the device does not perform automated analysis. The system provides a supplementary diagnostic tool for cardiac monitoring.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included biocompatibility (cytotoxicity, irritation, sensitization), electrical safety (IEC 60601-1, IEC 60601-1-11), electromagnetic compatibility (IEC 60601-1-2), performance testing (IEC 60601-2-47, ANSI/AAMI EC12), and software/cybersecurity validation.
Technological Characteristics
Single-channel ECG recorder; Bluetooth Low Energy (2402–2480 MHz) connectivity. Electrode patch uses 3M Single-Coated Medical Tan Elastic Nonwoven Tape and AmGel AG600 Series Sensing Hydrogels. Powered by 3.7V/160mAh rechargeable lithium battery. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-47, and ANSI/AAMI EC12. Software validated per IEC 62304.
Indications for Use
Indicated for patients 21 years of age or older with known or suspected cardiac conditions. Contraindicated for patients with chronic disease, polyester allergies, pregnancy, breastfeeding, infants/children, patients with pacemakers or ICDs, and patients undergoing defibrillation or AED rescue.
Regulatory Classification
Identification
A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
June 17, 2025
VitalSigns Technology Co., Ltd.
Leon Chuang
Chief of Technology Officer
3F., No. 153, Ziqiang S. Rd., Zhubei City
Hsinchu County, 302042
Taiwan
Re: K243003
Trade/Device Name: VitalSigns 1-Lead Holter (VSH101)
Regulation Number: 21 CFR 870.2920
Regulation Name: Telephone electrocardiograph transmitter and receiver
Regulatory Class: Class II
Product Code: DXH
Dated: May 23, 2025
Received: May 23, 2025
Dear Leon Chuang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K243003 - Leon Chuang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K243003 - Leon Chuang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer W. Shih -S
Jennifer Kozen
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243003
Device Name
VitalSigns 1-Lead Holter (VSH101)
Indications for Use (Describe)
The VSH101 is designed to record, transmit, and store single channel electrocardiogram (ECG) data via Bluetooth communication to compatible Bluetooth enabled devices. The device is intended for use by healthcare professionals, individuals with known or suspected cardiac conditions, and health-conscious users. The ECG data serves as a supplementary source of patient information and is not intended for automated analysis.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
VitalSigns
K243003
# 510(K) Summary
## 1. Submitter
| 510(K) Sponsor | VitalSigns Technology Co., Ltd |
| --- | --- |
| Address | 3F., No. 153, Ziqiang S. Rd., Zhubei City, Hsinchu County 302042, Taiwan (R.O.C.) |
| Applicant | JIH-LIANG, JUANG |
| Contact Information | TEL: +886-3-6682091
Mail: jl_juang@vsigntek.com |
| Correspondent Person | JIH-LIANG, JUANG |
| Contact Information | TEL: +886-3-6682091
Mail: jl_juang@vsigntek.com |
| Date | 2025/05/23 |
## 2. Subject device
| Device Name | VitalSigns 1-Lead Holter |
| --- | --- |
| Regulation Name | Telephone electrocardiograph transmitter and receiver |
| Regulation Number | 21 CFR870.2920 |
| Product Code | DXH |
| 510(k) review | Cardiovascular |
| Regulatory Class | Class II |
## 3. Predicate device
| Device Name | eCordum™ Cardiac Monitor |
| --- | --- |
| Regulation Name | Telephone electrocardiograph transmitter and receiver |
| Regulation Number | 21 CFR870.2920 |
| Product Code | DXH |
| 510(k) review | Cardiovascular |
| Regulatory Class | Class II |
| 510(k) number | K193296 |
{5}
VitalSigns
K243003
# 4. Device Description

ECG recorder host

Bluetooth

iOS APP VSHealth
Figure-1 Illustration of VitalSigns 1-Lead Holter
VitalSigns 1-Lead Holter is an ambulatory and Bluetooth-based wireless communication ECG measurement solution designed to allow users to record, store, transmit, and display single-channel ECG data.
VitalSigns 1-Lead Holter consists of the following three components:
(1) VS Electrode Patch "VSP101"
(2) ECG Recorder Host "VSH101"
(3) iOS APP "VSHealth"
When the ECG Recorder Host "VSH101" is fully charged and connected wirelessly via Bluetooth to the iOS app "VSHealth", it can instantly obtain the user's ECG data. VSHealth assists in transmitting, displaying, recording, and storing ECG data.
After the "VSH101" is fully charged, the device, through the VSHealth app, allows users to continuously use and record ECG data for 24 hours in their daily routines, whether at home or in a work environment.
# 5. Intended Use/ Indication for Use
The VSH101 is designed to record, transmit, and store single channel electrocardiogram (ECG) data via Bluetooth communication to compatible Bluetooth enabled devices. The device is intended for use by healthcare professionals, individuals with known or suspected cardiac conditions, and health-conscious users. The ECG data serves as a supplementary source of patient information and is not intended for automated analysis.
Page 2 of 8
{6}
VitalSigns
K243003
# 6. Contraindications
VSH101 is contraindicated for those patients with the following conditions:
- Chronic Disease patients
- Allergic to polyester patient
- Pregnant women, breastfeeding women, infants or children
- Patients with implanted pacemakers or implantable cardioverter defibrillators (ICDs). The device should not be used in the presence of these implants.
- Patients during defibrillation or AED rescue. The device must be removed prior to such procedures.
Caution: The ECG data acquired by the subject device is a non-traditional lead intended only for manual interpretation.
# 7. Target Population
This product is intended for use in patients 21 years of age or older.
Page 3 of 8
{7}
K243003
Page 4 of 8
| Device Name | VitalSigns 1-Lead Holter | eCordum™ Cardiac Monitor | Substantial equivalence of difference determination |
| --- | --- | --- | --- |
| Applicant | VitalSigns Technology Co., Ltd | eCordum, Inc. | -- |
| 510(k) Number | K243003 | K193296 | -- |
| Regulation Number | 21 CFR 870.2920 | 21 CFR 870.2920 | SE |
| Product Code | DXH | DXH | SE |
| Classification | Class II | Class II | SE |
| Indication for use | The VSH101 is designed to record, transmit, and store single- channel electrocardiogram (ECG) data via Bluetooth communication to compatible Bluetooth enabled devices. The device is intended for use by healthcare professionals, individuals with known or suspected cardiac conditions, and health-conscious users. The ECG data serves as a supplementary source of patient information and is not intended for automated analysis. | The eCordum™ Cardiac Monitor is intended to record, transfer and store single- channel electrocardiogram (ECG) data via Bluetooth transmission to Bluetooth enabled devices. The monitor is intended for use by healthcare professionals, patients with known or suspected heart conditions and health-conscious individuals. The ECG data is intended to supplement other patient data and is not intended for automated analysis. The device has not been tested and it is not intended for pediatric use. | SE |
| Number of ECG | 1 channel | 1 channel | SE |
| Power source | Rechargeable Lithium battery pack 3.7V/160mAh/ 0.59Wh | Replaceable Coin Battery (CR2032) | Different
The battery has been tested and met its pre- defined criteria, and it does not raise new issues of safety and effectiveness. |
{8}
K243003
Page 5 of 8
| Intended use environment | Healthcare and home environments | Healthcare | Similar
Predicate device allows ambulatory use and by healthcare professionals. Although home use is not explicitly stated, the functional environment overlap supports similar use conditions |
| --- | --- | --- | --- |
| Communication Protocol | Bluetooth Low Energy (2402 – 2480 MHz) | Class II Bluetooth | Similar
transmission type |
| Data storage | 24 hours | Not provided | Different
There is no change in the safety and effectiveness of the device. Performance is equivalent to IEC 60601-2-47 for all devices. |
| Viewing software platform | Mobile app | Mobile app | SE |
| Application/wear | Chest-based, single channel | Chest for ambulatory OTS electrodes | Similar
region |
| Electrodes | The VS Electrode Patch is composed of 3M™ Single-Coated Medical Tan Elastic Nonwoven Tape on Line and AmGel AG600 Series Sensing Hydrogels. Both materials are widely utilized in electrode applications. | Attachable standard ambulatory OTS electrodes with conductive gel. | Different
The VS Electrode Patch has been tested in accordance with ANSI/AAMI EC12, confirming compliance with established safety and performance standards. The patch provides stable conductivity, low impedance, and biocompatibility. |
{9}
K243003
Page 6 of 8
| **Compliance standard** | IEC 60601-1 | IEC 60601-1 | **Similar**
The predicate device does not reference ANSI/AAMI EC12 and IEC 60601-1-11. To address the required signal quality, evidence will be generated through the proposed performance verification protocol in this submission. |
| --- | --- | --- | --- |
| | IEC 60601-2-47
IEC 60601-1-11
IEC 10993-1
ANSI/AAMI EC12
FCC testing per part 15 | IEC 60601-1-2
IEC 60601-2-47
IEC 10993-1
FCC testing per part 15 | |
The VitalSigns 1-Lead Holter has been compared with a predicate device. The subject device has the same indications for use, similar technology/mechanism of action, and similar safety and performance as the predicate device.
Although there are some different specifications between two devices, the software validation, performance test and usability test have been completed to demonstrate that the differences between these parameters would not impact the safety and effectiveness of the subject device. The subject device has also undergone all safety and performance tests, and the results complied with the test requirements.
Therefore, the difference between the subject device and the predicate device did not raise any problem of substantial equivalence. The subject device is substantially equivalent to the predicate device in intended use, safety and performance claims.
Continued on the next page
{10}
VitalSigns
K243003
# 9. Summary of Non-Clinical Testing
The VitalSigns 1-Lead Holter meet all requirements for all design, biocompatibility, electrical, EMC safety and cybersecurity protection. The VitalSigns 1-Lead Holter pass all testing and supports the claims of substantial equivalence and safe operation.
The VitalSigns 1-Lead Holter was designed and tested for compliance with the applicable clauses of the following standards and USFDA guidance:
- ISO 10993-1
- IEC 60601-1
Note: VSH101 has been tested and complies with the requirements of Clause 8.5.5.2 – Energy reduction test.
- IEC 60601-1-2
- IEC 60601-2-47
- IEC 60601-1-11
- IEC 62304
- ISO 14971
- IEC 62366-1
- ANSI/AAMI EC 12
- IEEE/ANSI C63.27
- USFDA Guidance, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"
- USFDA Guidance, Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions.
- USFDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Device.
## Biocompatibility test
- Biological evaluation of medical device
- In vitro cytotoxicity
- Skin irritation
- Skin sensitization were conducted.
## Electromagnetic compatibility and electrical safety test
- IEC 60601-1 test
- IEC 60601-1-11 test
- Lithium battery safety test
- IEC 60601-1-2 test
## Performance testing
- IEC 60601-2-47 test
- Disposable ECG electrode test
## Software
- Recorder-VSH101 validation
- APP-VSHealth validation
- Cybersecurity-Cybersecurity management
## HFE/UE
## RF Wireless test
{11}
VitalSigns
## 10. Conclusion
The VitalSigns 1-Lead Holter is substantially equivalent with respect to safety and effectiveness to the legally marketed predicate device for its intended use. Minor differences between the VitalSigns 1-Lead Holter and the predicate device have no effect on safety or effectiveness, as established through various performance tests.
Page 8 of 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.