Reusable Temperature Probe (T1306, T2306, T3306, T4306); Disposable Temperature Probe (T5106, T6106)
K243000 · Shenzhen Medke Technology Co., Ltd. · FLL · Jun 20, 2025 · General Hospital
Device Facts
Record ID
K243000
Device Name
Reusable Temperature Probe (T1306, T2306, T3306, T4306); Disposable Temperature Probe (T5106, T6106)
Applicant
Shenzhen Medke Technology Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Jun 20, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Temperature Probes are intended to be used for monitoring temperature. The temperature probes are reusable or for single patient use and designed for use with Mindray monitor Model PM-8000 and other monitors compatible with YSI 400 series temperature probes. These devices are used by qualified medical professional only.
Device Story
Temperature probes consist of a connector and a thermistor; designed for use with Mindray PM-8000 or YSI 400-series compatible monitors. Operates via resistance-based thermistor sensing; resistance changes linearly with temperature. Used in clinical settings by qualified medical professionals for continuous temperature monitoring. Output is displayed on a compatible patient monitor to assist clinicians in patient temperature management. Reusable models (T1306, T2306, T3306, T4306) and disposable models (T5106, T6106) are available for skin or body cavity (esophageal/rectal) sites.
Clinical Evidence
No clinical data. Bench testing only. Verification included biocompatibility (cytotoxicity, sensitization, irritation, oral/rectal irritation) per ISO 10993-1 and performance testing per IEC 60601-1, IEC 60601-1-2, and ISO 80601-2-56.
Technological Characteristics
Thermistor-based sensing; 2.25KΩ resistance at 25°C. Reusable materials: PVC, copper nickel-plated, epoxy AB glue. Disposable materials: PVC, foam, aluminum foil. Non-sterile. Continual operation mode. Compatible with YSI 400 series interfaces.
Indications for Use
Indicated for monitoring patient body temperature in healthcare facilities. Suitable for adult and pediatric populations. Used for skin, esophageal, or rectal temperature measurement.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Unimed Temperature Probe (Unimed Skin Temperature Probe, Unimed General Purpose Temperature Probe) (K121427)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 20, 2025
Shenzhen Medke Technology Co., Ltd.
% Mrs. Jie Yang
Consultant
Chonconn Medical Device Consulting Co., Ltd.
Room 504, Block C, No. 1029 Nanhai Avenue, Nanshan District
Shenzhen, Guangdong 518067
China
Re: K243000
Trade/Device Name: Reusable Temperature Probe (T1306, T2306, T3306, T4306); Disposable Temperature Probe (T5106, T6106)
Regulation Number: 21 CFR 880.2910
Regulation Name: Clinical Electronic Thermometer
Regulatory Class: Class II
Product Code: FLL
Dated: September 26, 2024
Received: May 23, 2025
Dear Jie Yang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243000 - Jie Yang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K243000 - Jie Yang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
PORSCHE P. BENNETT -S
Porsche Bennett
For David Wolloscheck, Ph.D.
Assistant Director
DHT3C: Division of Drug Delivery and
General Hospital Devices, and
Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243000
Device Name
Reusable Temperature Probe(T1306, T2306, T3306, T4306);
Disposable Temperature Probe(T5106, T6106)
Indications for Use (Describe)
Temperature Probes are intended to be used for monitoring temperature. The temperature probes are reusable or for single patient use and designed for use with Mindray monitor Model PM-8000 and other monitors compatible with YSI 400 series temperature probes.
These devices are used by qualified medical professional only.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
510(K) Summary- K243000
Prepared in accordance with the requirements of 21 CFR Part 807.92
Prepared Date: 2025/06/20
1. Submission sponsor
Name: Shenzhen Medke Technology Co., Ltd.
Address: 4/F, Bldg. A1, Anle Ind. Zone, Hangcheng RD., Baoan District,518126, Shenzhen, China
Contact person: Li Wei Tao
Title: CEO
E-mail: tate@medke.com
Tel: 0755-23463462
2. Submission correspondent
Name: Chonconn Medical Device Consulting Co., Ltd.
Address: Room 504, Block C, No. 1029 Nanhai Avenue, Nanshan District, Shenzhen, Guangdong, P.R. China
Contact person: Yang Jie
E-mail: yangjie@chonconn.com
Tel: +86-755 33941160
3. Subject Device Information
| Trade/Device Name | Reusable Temperature Probe (T1306, T2306, T3306, T4306); Disposable Temperature Probe (T5106, T6106) |
| --- | --- |
| Common Name | Temperature Probe |
| Regulatory Class | Class II |
| Regulation & Name | 21CFR 880.2910 / Thermometer, electronic, clinical |
| Product Code | FLL |
| Submission type | Traditional 510(K) |
4. Predicate Device
| Predicate | Device name and model | K number | Manufacturer |
| --- | --- | --- | --- |
| Primary Predicate device | Unimed Temperature Probe (Unimed Skin Temperature Probe, Unimed General Purpose Temperature Probe) | K121427 | Unimed Medical Supplies Inc. |
| Reference device | ESOPHAGEAL RECTAL EMPERATURE PROBE ER400-9, ER400-12 | K863646 | RESPIRATORY SUPPORT PRODUCTS, INC. (owned by ICU Medical, Inc. now) |
1 / 7
{5}
2 / 7
# 5. Device Description
The subject devices are used for patient temperature measurement. The probes are reusable or disposable depending on models. These probes consist of a connector on the monitor end and a thermistor on the patient end. The working principle is resistance based on the metal conductor increases with temperature decrease, and the linear changes to the characteristics of the temperature measurement. The subject devices are designed to be used in healthcare facilities like hospital and compatible with a monitor of Mindray Model PM-8000 and other monitors compatible with YSI 400 series temperature probes.
The six models have two types of structure designs corresponding to reusable and disposable use which consists of different materials. The NTC of the six models are identical. Reusable models T1306, T2306, T3306, T4306 have a similar structure design with two different sensor shapes for different measure sites and consist of the same materials. Disposable models T5106 and T6106 have a similar structure design with two different sensor shapes for different measure sites and consist of the same materials.
| Model | Description |
| --- | --- |
| T1306 | Skin contact Temperature Probe, adult, reusable |
| T2306 | Body cavity Temperature Probe, Esophageal/Rectal, adult, reusable |
| T3306 | Skin contact Temperature Probe, pediatric, reusable |
| T4306 | Body cavity Temperature Probe, Esophageal/Rectal, pediatric, reusable |
| T5106 | Skin contact Temperature Probe, adult/ pediatric, disposable |
| T6106 | Body cavity Temperature Probe, Esophageal/Rectal, adult/ pediatric, disposable |
# 6. Intended use & Indication for use
Temperature Probes are intended to be used for monitoring temperature. The temperature probes are reusable or for single patient use and designed for use with Mindray monitor Model PM-8000 and other monitors compatible with YSI 400 series temperature probes.
These devices are used by qualified medical professional only.
# 7. Comparison to the Predicate Device
The following tabulation indicates the detailed differences between the subject devices and the predicate device.
| Item | Subject Device | Primary Predicate Device | Comparison |
| --- | --- | --- | --- |
| Trade name | Reusable Temperature Probe (T1306, T2306, T3306, T4306); Disposable Temperature Probe (T5106, T6106) | Unimed Temperature Probe | / |
| 510(k) submitter | Shenzhen Medke Technology Co., Ltd. | Unimed Medical Supplies Inc. | / |
| 510(k) Number | K243000 | K121427 | / |
{6}
| Models | Skin Temperature Probe: Reusable: T1306, T3306 Disposable: T5106 Body cavity Temperature Probe: | Unimed Skin Temperature Probe, Unimed General Purpose Temperature Probe | / |
| --- | --- | --- | --- |
| Item | Subject Device | Primary Predicate Device | Comparison |
| | Reusable: T2306, T4306 Disposable: T6106 | | |
| Classification Regulation | 880.2910 | 880.2910 | Same |
| Class | II | II | Same |
| Product Code | FLL | FLL | Same |
| Common Name | Temperature Probe | Temperature Probe | Same |
| Type of use | Prescription | Prescription | Same |
| Indication for Use | Temperature Probes are intended to be used for monitoring temperature. The temperature probes are reusable or for single patient use and designed for use with Mindray monitor Model PM-8000 and other monitors compatible with YSI 400 series temperature probes. These devices are used by qualified medical professional only. | Unimed Temperature Probes are intended to be used for monitoring temperature. The temperature probes are reusable and designed for use with monitors of Philips, Marquette, Mindray, Spacelabs, Siemens, Artema/S&W and other monitors compatible with YSI 400 series temperature probes. These devices are indicated for used by qualified medical personnel only. | Similar, note 1 |
| Operating Principle | Resistance of thermistor based on the metal conductor increases with temperature decrease, and the linear changes to the characteristics of the temperature measurement. | Resistance of thermistor based on the metal conductor increases with temperature decrease, and the linear changes to the characteristics of the temperature measurement. | Same |
{7}
| Measure site | Skin, Esophageal and Rectal | Skin, Esophageal and Rectal | Same |
| --- | --- | --- | --- |
| Operation mode | Direct mode | Direct mode | Same |
| Usage | Reusable, Disposable | Reusable | Similar Note 1 |
| Item | Subject Device | Primary Predicate Device | Comparison |
| Measurement Range | 25-45°C | 25-45°C | Same |
| Accuracy | ±0.1°C | ±0.1°C | Same |
| Component | plug, cable and temperature sensing probe | plug, cable and temperature sensing probe | Same |
| Thermistor resistance | 2.25KΩ@25°C | 2.252KΩ@25°C | Same |
| Material | Materials of reusable probe: PVC, Copper nickel plated, Expoxy AB glue Materials of disposable probe: PVC, Foam, Aluminum foil | Not-provided | Similar Note 2 |
| Compatible Monitors | Mindray Model PM-8000, and other monitors compatible with YSI 400 series temperature probes. | Philips, Marquette, Mindray, Spacelabs, Siemens, Artema/S&W and other monitors compatible with YSI 400 series temperature probes. | Similar, Note 1 |
| Operating Environment | Temperature: +5 to+40°C Humidity: ≤80%(non-condensing) Atmospheric pressure: 86kPa~106kPa | Not-provided | Similar Note 3 |
| Storage Environment | Temperature range: -20°C to 55°C(-4 to 131°F) Humidity (Operating): ≤93% Hyperbaric Pressure: 86kPa~106kPa | Not-provided | Similar Note 3 |
| Electrical Safety | Comply with IEC 60601-1 | Comply with IEC 60601-1 | Same |
| Performance | ISO 80601-2-56 | EN 12470-4 | Similar, Note 4 |
4 / 7
{8}
| Biocompatibility | All the patient-contacting materials are evaluated by the biocompatibility standard ISO10993-5, ISO 10993-10, ISO 10993-23 | All the patient-contacting materials are evaluated by the biocompatibility standard ISO10993-5, ISO 10993-10. | Similar, Note 4 |
| --- | --- | --- | --- |
5 / 7
{9}
| Item | Subject Device | Primary Predicate Device | Comparison |
| --- | --- | --- | --- |
| Sterilization | Non-sterile | Non-sterile | Same |
| Operational Type | Continual | Continual | Same |
## Justification of differences:
### Note 1
The subject device claims different compatible patient monitor compared with predicate device. However, they all compatible with monitors compatible with YSI 400 series temperature probes. The compatibility has been verified by bench testing. Besides, the subject device includes both reusable and disposable models while the predicate only include reusable models. The reusable models are considered higher risk models and can be used to cover both reusable and disposable models of the subject devices. Therefore, the difference does not raise any new or different safety and effectiveness questions.
### Note 2
The materials information of the predicate is not available. However, the biocompatibility of the subject device has been evaluated per ISO 10993-1. The difference does not raise any new or different safety and effectiveness questions.
### Note 3
Operating and Storage condition difference: The subject device has been tested and the test results met the requirements of IEC 60601-1 and ISO 80601-2-56. The difference does not raise any new or different safety and effectiveness questions.
### Note 4
The subject device complies with the newest FDA recognized standard compared with the predicate device. The difference does not raise any new or different safety and effectiveness questions.
## 8. Non-clinical Data
The following performance data were provided in support of the substantial equivalence determination.
### Biocompatibility testing
The biocompatibility evaluation for the subject devices were conducted in accordance with the FDA Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The biocompatibility testing included the following tests:
- Cytotoxicity
- Sensitization
{10}
- Irritation
- Oral mucosa/ Rectal Irritation (Body cavity Temperature Probe only)
## Bench testing
Bench testing has been conducted to verify that the subject devices meet all design specifications which support the conclusion that it's Substantially Equivalent (SE) to the predicate devices. The testing results demonstrate that the targeted device complies with the following standards:
- IEC 60601-1-2005+CORR.1:2006+CORR.2:2007+A1:2012, Medical Electrical Equipment- Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- ISO 80601-2-56:2017+A1:2018 Medical electrical equipment – Particular requirements for the basic safety and essential performance of clinical thermometers for body temperature measurement.
## 9. Clinical study
No clinical study is needed for the subject device.
## 10. Conclusion
Substantial equivalence comparisons, performance testing and compliance with voluntary standards demonstrate that the differences between the subject and predicate device do not raise new questions of safety and effectiveness. The subject device is substantially equivalent to the predicate device.
7 / 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.