Retrospective clinical CT images were used to train the deep-learning model and to conduct a standalone performance evaluation of the OncoStudio device.
Adult patients; 58% male, 42% female; diverse ethnic backgrounds; Sample Size: 310 CT images; Number of Sites: 3
Not applicable for this study
Mean Dice coefficient (DSC) of structures for anatomical regions (Head & Neck, Thorax, Abdomen, and Pelvis)
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Automatic contouring of organs at risk
Deep learning
—
—
Training: 2,128 images (731 from Yonsei Severance Hospital, 194 from OneMedNet, 1203 from University Hospital Basel).
3 (radiation oncologists)
Standalone performance test: 310 CT images (140 from Yonsei Severance Hospital, 121 from OneMedNet, 49 from University Hospital Basel).
3 (radiation oncologists)
Indications for Use
OncoStudio provides deep-learning-based automatic contouring to organs at risk in DICOM-RT format from CT images. This software could be used as an initial contouring for the clinicians to be confirmed by the radiation oncology department for treatment planning or other professions where a segmented mask of organs is needed.
Device Story
OncoStudio is a standalone software application that processes CT images to perform automatic segmentation of organs at risk (OARs). Using deep-learning algorithms, the device generates DICOM-RT structure sets. It is intended for use by radiation oncology departments and other medical professionals to provide initial contours, which must be reviewed, edited, and confirmed by a clinician before use in treatment planning. The software supports manual contouring and standard DICOM image management functions (receive, transmit, store, retrieve, display). By automating the initial contouring process, the device aims to assist clinicians in radiation therapy treatment planning workflows, potentially improving efficiency in clinical practice.
Clinical Evidence
Bench testing only. Performance evaluated on 310 independent CT images (140 Korea, 121 USA, 49 Europe). Ground truth established by three radiation oncologists following RTOG and clinical guidelines. Performance metrics included mean Dice Similarity Coefficient (DSC) across anatomical regions (Head & Neck, Thorax, Abdomen, Pelvis). Training data (2,438 total) included diverse ethnicities and CT manufacturers (Siemens, Philips, Toshiba, GE).
Indicated for adult patients requiring automatic contouring of organs at risk from CT images for radiation oncology treatment planning or other clinical workflows requiring segmented organ masks. Covers Head & Neck, Thorax, Abdomen, and Pelvis regions.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo is a stylized depiction of a human figure. The FDA acronym is in a blue square, and the words "U.S. FOOD & DRUG ADMINISTRATION" are in blue text to the right of the square.
February 24, 2025
OncoSoft. Co., Ltd. Boram Kim RA/QA Manager 37, Myeongmul-gil, Seodaemun-gu SEOUL, 03776 KOREA, SOUTH
Re: K242994
Trade/Device Name: OncoStudio (OS-01) Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: QKB Dated: January 23, 2025 Received: January 23, 2025
Dear Boram Kim:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Locon Weidner
Lora D. Weidner, Ph.D. Assistant Director Radiation Therapy Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K242994
Device Name
OncoStudio (OS-01)
Indications for Use (Describe)
OncoStudio provides deep-learning-based automatic contouring to organs at risk in DICOM-RT format from CT images. This software could be used as an initial contouring for the clinicians to be confirmed by the radiation oncology department for treatment planning or other professions where a segmented mask of organs is needed.
- · Deep learning contouring from Head & Neck, Thorax, Abdomen, and Pelvis
- · Generates DICOM-RT structure of contoured objects
- · Manual Contouring
- · Receive, transmit, store, retrieve, display, and process medical images and DICOM objects
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
[As Required by 21 CFR 807.92]
#### 1. Date Prepared [21 CFR 807.92(a)(a)]
January 23, 2025
#### 2. Submitter's Information [21 CFR 807.92(a)(1)]
- Name of Manufacturer: OncoSoft Co., Ltd. 37, Myeongmul-gil, Seodaemun-gu, Seoul, Republic of Korea ● Address: (03776) ● Contact Name: Boram Kim ● Telephone No.: +82-10-6305-7428 ● Email Address: kbrstar@oncosoft.io
- 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
| 510(k) Number | K242994 |
|-----------------------------|------------------------------------------------|
| Trade/Device/Model Name | OncoStudio / OS-01 |
| Device Classification Name | Medical Image Management and Processing System |
| Regulation Number | 21 CFR 892.2050 |
| Classification Product Code | QKB |
| Device Class | Class II |
| 510(k) Review Panel | Radiology |
{5}------------------------------------------------
### 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]
The identified predicate device within this submission is shown as follow;
### Predicate Device
| 510(k) Number | K230685 |
|-----------------------------|------------------------------------------------|
| Trade/Device/Model Name | AutoContour Model RADAC V3 |
| Device Classification Name | Medical Image Management And Processing System |
| Regulation Number | 21 CFR 892.2050 |
| Classification Product Code | OKB |
| Device Class | Class II |
| 510(k) Review Panel | Radiology |
#### Reference Device
| 510(k) Number | K232899 |
|-----------------------------|------------------------------------------------|
| Trade/Device/Model Name | AI-Rad Companion Organs RT |
| Device Classification Name | Medical Image Management And Processing System |
| Regulation Number | 21 CFR 892.2050 |
| Classification Product Code | OKB |
| Device Class | Class II |
| 510(k) Review Panel | Radiology |
These predicate devices have not been subject to a design-related recall
{6}------------------------------------------------
#### 5. Description of the Device [21 CFR 807.92(a)(4)]
OncoStudio is a standalone software that provides deep-learning-based automatic contouring to organs at risk in DICOM-RT format from CT images. This software could be used as an initial contouring for the clinicians to be confirmed by the radiation oncology department for treatment planning or other professions where a segmented mask of organs is needed.
- Deep learning contouring from Head & Neck, Thorax, Abdomen, and Pelvis
- · Generates DICOM-RT structure of contoured objects
- Manual Contouring
- · Receive, transmit, store, retrieve, display, and process medical images and DICOM objects
It also has the following general functions:
- Patient management;
- · Review of processed images;
- Open and Save of files.
#### 6. Indications for use [21 CFR 807.92(a)(5)]
OncoStudio provides deep-learning-based automatic contouring to organs at risk in DICOM-RT format from CT images. This software could be used as an initial contouring for the clinicians to be confirmed by the radiation oncology department for treatment planning or other professions where a segmented mask of organs is needed.
- Deep learning contouring from Head & Neck, Thorax, Abdomen, and Pelvis
- · Generates DICOM-RT structure of contoured objects
- Manual Contouring
- · Receive, transmit, store, retrieve, display, and process medical images and DICOM objects
{7}------------------------------------------------
### 7. Technological Characteristics (Equivalence to Predicate Device) [21 CFR 807.92(a)(6)]
There are no significant differences in the technological characteristics of these devices compared to the predicate devices which adversely affect safety or effectiveness. Provided below is a table summarizing and comparing the technological characteristics of the OncoStudio and the predicate devices:
#### [Table 1. Comparison of Proposed Device to Predicate Device and Reference Device]
| | Subject Device | Predicate Device | Reference Device1 | |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Item | OncoStudio | AutoContour Model RADAC<br>V3 | AI-Rad Companion<br>Organs RT | SE<br>Note |
| Regulation<br>Name | Medical Image Management<br>And Processing System | Medical Image<br>Management And<br>Processing System | Medical Image<br>Management And<br>Processing System | - |
| Regulation<br>Number | 21 CFR 892.2050 | 21 CFR 892.2050 | 21 CFR 892.2050 | - |
| Product<br>Code | QKB | QKB | QKB | - |
| Class | II | II | II | - |
| 510k<br>Number | K242994 | K230685 | K232899 | - |
| Indication<br>for Use | OncoStudio provides deep-<br>learning-based automatic<br>contouring to organs at risk<br>in DICOM-RT format from CT<br>images. This software could<br>be used as an initial<br>contouring for the clinicians<br>to be confirmed by the<br>radiation oncology<br>department for treatment<br>planning or other professions<br>where a segmented mask of<br>organs is needed.<br>• Deep learning<br>contouring from Head &<br>Neck, Thorax, Abdomen,<br>and Pelvis<br>• Generates DICOM-RT<br>structure of contoured<br>objects<br>• Manual Contouring<br>• Receive, transmit, store | AutoContour is intended to<br>assist radiation treatment<br>planners in contouring and<br>reviewing structures within<br>medical images in<br>preparation for radiation<br>therapy treatment<br>planning | AI-Rad Companion<br>Organs RT is a post-<br>processing software<br>intended to<br>automatically<br>contour DICOM CT<br>and MR predefined<br>structures using<br>deep-learning-based<br>algorithms. Contours<br>that are generated<br>by AI-Rad<br>Companion Organs<br>RT may be used as<br>input for clinical<br>workflows including<br>external beam<br>radiation therapy<br>treatment planning.<br>AI-Rad Companion<br>Organs RT must be<br>used in conjunction<br>with appropriate | Same |
{8}------------------------------------------------
# K242994
| | retrieve, display, and process<br>medical images and DICOM<br>objects | | | | software such as<br>Treatment Planning<br>Systems and<br>Interactive<br>Contouring<br>applications, to<br>review, edit, and<br>accept contours<br>generated by AI-Rad<br>Companion Organs<br>RT. The output of AI-<br>Rad Companion<br>Organs RT are<br>intended to be used<br>by trained medical<br>professionals. The<br>software is not<br>intended to<br>automatically detect<br>or contour lesions. | |
|----------------------------------|-----------------------------------------------------------------------|--------------------|------------------|-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| | • A_Aorta | • LN_Sclav_L_RTOG | • A_Aorta | • LN_Ax_L | (166 OARs) | |
| | • A_Carotid_L | • LN_Sclav_R_ESTRO | • A_Aorta_Asc | • LN_Ax_L1_L | Full list of OAR is not provided | |
| | • A_Carotid_R | • LN_Sclav_R_RTOG | • A_Aorta_Dsc | • LN_Ax_L1_R | | |
| | • A_Coronary_R | • Larynx | • A_LAD | • LN_Ax_L2_L | | |
| | • A_Iliac_L | • Lens_L | • A_Pulmonary | • LN_Ax_L2_L3_L | | |
| | • A_Iliac_R | • Lens_R | • Bladder | • LN_Ax_L2_L3_R | | |
| | • A_LAD | • Liver | • Bladder_F | • LN_Ax_L2_R | | |
| | • A_Subclavian_L | • Lobe_Temporal_L | • Bone_Ilium_L | • LN_Ax_L3_L | | |
| | • A_Subclavian_R | • Lobe_Temporal_R | • Bone_Ilium_R | • LN_Ax_L3_R | | |
| | • Anus | • Lung_L | • Bone_Mandible | • LN_Ax_R | | |
| | • Atrium_L | • Lung_LLL | • Bone_Pelvic | • LN_IMN_L | | |
| Regions of<br>Interest(R<br>OIs) | • Atrium_R | • Lung_LUL | • Bone_Skull | • LN_IMN_R | | |
| | • Autochthon_L | • Lung_R | • Bone_Sternum | • LN_IMN_RC_L | | Differ |
| | • Autochthon_R | • Lung_RLL | • Bowel | • LN_IMN_RC_R | | ent |
| | • Bag_Bowel | • Lung_RML | • Bowel_Bag | • LN_Inguinofem_L | | |
| | • Bladder | • Lung_RUL | • Bowel_Large | • LN_Inguinofem_R | | |
| | • Bone_Mandible | • OpticChiasm | • Bowel_Small | • LN_Neck_IA | | |
| | • Bowel_Large | • OpticNrv_L | • BrachialPlex_L | • LN_Neck_IB-V_L | | |
| | • Bowel_Small | • OpticNrv_R | • BrachialPlex_R | • LN_Neck_IB-V_R | | |
| | • BrachialPlex_L | • Pancreas | • Brain | • LN_Neck_II_L | | |
| | • BrachialPlex_R | • Parotid_L | • Brainstem | • LN_Neck_II_R | | |
| | • Brachiocephalic_Trunk | • Parotid_R | • Breast_L | • LN_Neck_II-IV_L | | |
| | • Brain | • Pharynx | • Breast_R | • LN_Neck_II-IV_R | | |
| | • Brainstem | • Pituitary | • Bronchus | • LN_Neck_II-V_L | | |
| | • Breast_L | • Prostate | • BuccalMucosa | • LN_Neck_II-V_R | | |
| | • Breast_R | • Rectum | • Carina | • LN_Neck_III_L | | |
{9}------------------------------------------------
## K242994
| • Bronchus_L | • Rib01_L | • CaudaEquina | • LN_Neck_III_R |
|---------------------|-----------------|-------------------|----------------------|
| • Bronchus_R | • Rib01_R | • Cavity_Oral | • LN_Neck_IV_L |
| • CaudaEquina | • Rib02_L | • Cavity_Oral_Ext | • LN_Neck_IV_R |
| • Cavity_Oral | • Rib02_R | • Chestwall_L | • LN_Neck_V_L |
| • Clavicle_L | • Rib03_L | • Chestwall_OAR | • LN_Neck_V_R |
| • Clavicle_R | • Rib03_R | • Chestwall_R | • LN_Neck_VIA |
| • Cochlea_L | • Rib04_L | • Chestwall_RC_L | • LN_Neck_VIIA_L |
| • Cochlea_R | • Rib04_R | • Chestwall_RC_R | • LN_Neck_VIIA_R |
| • Colon | • Rib05_L | • Cochlea_L | • LN_Neck_VIIB_L |
| • Kidney_Cortex_L | • Rib05_R | • Cochlea_R | • LN_Neck_VIIB_R |
| • Kidney_Cortex_R | • Rib06_L | • Colon_Sigmoid | • LN_Paraaortic |
| • Costal_Cartilages | • Rib06_R | • Cornea_L | • LN_Pelvics |
| • Duodenum | • Rib07_L | • Cornea_R | • LN_Pelvic_NRG |
| • Esophagus | • Rib07_R | • Duodenum | • LN_Sclav_L |
| • Eye_L | • Rib08_L | • Ear_Internal_L | • LN_Sclav_R |
| • Eye_R | • Rib08_R | • Ear_Internal_R | • LN_Sclav_RADCOMP_L |
| • Femur_Head_L | • Rib09_L | • Esophagus | • LN_Sclav_RADCOMP_R |
| • Femur_Head_R | • Rib09_R | • External | • Lobe_Temporal_L |
| • Femur_L | • Rib10_L | • Eye_L | • Lobe_Temporal_R |
| • Femur_R | • Rib10_R | • Eye_R | • Lung_L |
| • Gallbladder | • Rib11_L | • Femur_Head_L | • Lung_R |
| • Glnd_Adrenal_L | • Rib11_R | • Femur_Head_R | • Macula_L |
| • Glnd_Adrenal_R | • Rib12_L | • Femur_L | • Macula_R |
| • Glnd_Submand_L | • Rib12_R | • Femur_R | • Marrow_Ilium_L |
| • Glnd_Submand_R | • Sacrum | • Femur_RTOG_L | • Marrow_Ilium_R |
| • Glnd_Thyroid | • Scapula_L | • Femur_RTOG_R | • Musc_Constrict |
| • Gluteus_Maximus_L | • Scapula_R | • GallBladder | • Nipple_L |
| • Gluteus_Maximus_R | • Colon_Sigmoid | • Genitals_F | • Nipple_R |
| • Gluteus_Medius_L | • Skull | • Genitals_M | • OpticChiasm |
| • Gluteus_Medius_R | • SpinalCord | • Glnd_Lacrimal_L | • OpticNrv_L |
| • Gluteus_Minimus_L | • Spleen | • Glnd_Lacrimal_R | • OpticNrv_R |
| • Gluteus_Minimus_R | • Sternum | • Glnd_Submand_L | • Pancreas |
| • Heart | • Stomach | • Glnd_Submand_R | • Parotid_L |
| • Hip_L | • Trachea | • Glnd_Thyroid | • Parotid_R |
| • Hip_R | • VB_C1 | • HDR_Cylinder | • PenileBulb |
| • Hippocampus_L | • VB_C2 | • Heart | • Pericardium |
| • Hippocampus_R | • VB_C3 | • Hippocampus_L | • Pituitary |
| • Humerus_L | • VB_C4 | • Hippocampus_R | • Prostate |
| • Humerus_R | • VB_C5 | • Humerus_L | • Rectum |
| • Iliopsoas_L | • VB_C6 | • Humerus_R | • Rectum_F |
| • Iliopsoas_R | • VB_C7 | • Kidney_L | • Retina_L |
| • Joint_TM_L | • VB_L1 | • Kidney_R | • Retina_R |
| • Joint_TM_R | • VB_L2 | • Kidney_Outer_L | • Rib |
| • Kidney_L | • VB_L3 | | • Rib_L |
{10}------------------------------------------------
# K242994
| | • Kidney_R | • VB_L4 | • Kidney_Outer_R | • Rib_R | | |
|---------------------|--------------------------------|------------------------|------------------|------------------------------------------------------------------------------------------------------------------------|----------------------------|---------------|
| | • LN_Ax_L1_L | • VB_L5 | • Larynx | • SeminalVes | | |
| | • LN_Ax_L1_R | • VB_S1 | • Larynx_Glottic | • SpinalCanal | | |
| | • LN_Ax_L2_L | • VB_T01 | • Larynx_NRG | • SpinalCord | | |
| | • LN_Ax_L2_R | • VB_T02 | • Larynx_SG | • Spleen | | |
| | • LN_Ax_L3_L | • VB_T03 | • Lens_L | • Stomach | | |
| | • LN_Ax_L3_R | • VB_T04 | • Lens_R | • Trachea | | |
| | • LN_IMN_L | • VB_T05 | • Lips | • UteroCervix | | |
| | • LN_IMN_R | • VB_T06 | • Liver | • V_Venacava_I | | |
| | • LN_Neck_IA | • VB_T07 | | • V_Venacava_S | | |
| | • LN_Neck_IB_L | • VB_T08 | | • VB | | |
| | • LN_Neck_IB_R | • VB_T09 | | • VB_C1 | | |
| | • LN_Neck_III_L | • VB_T10 | | • VB_C2 | | |
| | • LN_Neck_III_R | • VB_T11 | | • VB_C3 | | |
| | • LN_Neck_II_L | • VB_T12 | | • VB_C4 | | |
| | • LN_Neck_II_R | • V_Brachioceph_L | | • VB_C5 | | |
| | • LN_Neck_IVA_L | • V_Brachioceph_R | | • VB_C6 | | |
| | • LN_Neck_IVA_R | • V_Iliac_L | | • VB_C7 | | |
| | • LN_Neck_IVB_L | • V_Iliac_R | | • VB_L1 | | |
| | • LN_Neck_IVB_R | • V_Portal_And_Splenic | | • VB_L2 | | |
| | • LN_Neck_VA_L | • V_Pulmonary | | • VB_L3 | | |
| | • LN_Neck_VA_R | • V_Venacava_I | | • VB_L4 | | |
| | • LN_Neck_VBC_L | • V_Venacava_S | | • VB_L5 | | |
| | • LN_Neck_VBC_R | • Ventricle_L | | • VB_T01 | | |
| | • LN_Sclav_L_ESTRO | • Ventricle_R | | • VB_T02 | | |
| | | | | • VB_T03 | | |
| | | | | • VB_T04 | | |
| | | | | • VB_T05 | | |
| | | | | • VB_T06 | | |
| | | | | • VB_T07 | | |
| | | | | • VB_T08 | | |
| | | | | • VB_T09 | | |
| | | | | • VB_T10 | | |
| | | | | • VB_T11 | | |
| | | | | • VB_T12 | | |
| | | | | Windows based .NET<br>front-end application that | | |
| Operating<br>System | Local deployment on<br>Windows | | | also serves as agent<br>Uploader supporting<br>Microsoft Windows 10 (64-<br>bit) and Microsoft<br>Windows Server 2016. | Edge & Cloud<br>Deployment | Differ<br>ent |
| | | | | Cloud-based Server based<br>automatic contouring<br>application compatible | | |
C
{11}------------------------------------------------
| | | with Linux. Windows<br>python-based automatic<br>contouring application<br>supporting Microsoft<br>Windows 10 (64-bit) and<br>Microsoft Windows Server<br>2016. | | |
|---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|
| Image<br>Format | DICOM | DICOM | DICOM | Same |
| General<br>Functions | 1) Deep learning contouring<br>from Head & Neck, Thorax,<br>Abdomen, and Pelvis<br>2) Generates DICOM-RT<br>structure of contoured<br>objects<br>3) Manual Contouring<br>4) Receive, transmit, store,<br>retrieve, display, and process<br>medical images and DICOM<br>objects | 1) Automatically contour<br>various structures of<br>interest for radiation<br>therapy treatment<br>planning<br>2) Allow the user to review<br>and modify the resulting<br>contours<br>3) Generate DICOM-<br>compliant structure set<br>data the can be imported<br>into a radiation therapy<br>treatment planning system | CT or MR series of<br>images serve as<br>input for AI-Rad<br>Companion Organs<br>RT and are acquired<br>as part of a typical<br>scanner acquisition.<br>Once processed by<br>the AI algorithms,<br>generated contours<br>in DICOM-RTSTRUCT<br>format are reviewed<br>in a confirmation<br>window, allowing<br>clinical user to<br>confirm or reject the<br>contours before<br>sending to the target<br>system. Optionally,<br>the user may select<br>to directly transfer<br>the contours to a<br>configurable DICOM<br>node | Same |
| Algorithm | Deep Learning | Deep Learning | Deep Learning | Same |
| Compatibl<br>e Modality | CT Images. DICOM<br>RTSTRUCT for output | CT or MR input for<br>contouring or<br>registration/fusion. PET/CT<br>input for<br>registration/fusion only.<br>DICOM RTSTRUCT for<br>output | CT & MR Images | Equiv<br>alent |
| Segmentat<br>ion of<br>Organ | Head & Neck, Thorax,<br>Abdomen, Pelvis | Head and Neck, Thorax,<br>Abdomen, Pelvis | Head & Neck,<br>Thorax, Abdomen &<br>Pelvis Head & Neck<br>lymph nodes | Same |
| Workflow | Automatically processes input<br>image data contour organs | Automatically contour<br>various structures of | AI-Rad Companion<br>Organs RT | Same |
| | and DICOM sends generated<br>RT Structure set | interest for radiation<br>therapy treatment<br>planning using machine<br>learning based contouring | automatically<br>processes input<br>image data and<br>sends the results as<br>DICOM-RT Structure<br>Sets to a user-<br>configurable target<br>node | |
| Compatibl<br>e Scanner<br>Models | No Limitation on scanner<br>model, DICOM 3.0<br>compliance required. | No Limitation on scanner<br>model, DICOM 3.0<br>compliance required. | No Limitation on<br>scanner model,<br>DICOM 3.0<br>compliance required. | Same |
| Compatibl<br>e<br>Treatment<br>Planning<br>System | No Limitation on TPS model,<br>DICOM 3.0 compliance<br>required. | No Limitation on TPS<br>model, DICOM 3.0<br>compliance required. | No Limitation on TPS<br>model, DICOM 3.0<br>compliance required. | Same |
| Patient<br>Population | Adult only | Adult only | Adult only | Same |
{12}------------------------------------------------
The intended use of the predicate device and the subject device are equivalent. Both devices are intended to aid users to contour the body structure using artificial intelligence algorithm that can be used as an initial contouring for the clinicians to be confirmed by the radiation oncology department for treatment planning or other professions where a segmented mask of organs is needed.
A detailed comparison shows the subject device is substantially equivalent in indications for use, image format, general functions, algorithm, segmentation of organs, workflow, compatible scanner models, compatible treatment planning system and patient population to the predicate device. The differences between the subject and the predicate devices do not raise any new questions regarding safety and effectiveness.
#### 8. Non-Clinical Test summary
The following data were provided in support of the substantial equivalence determination:
- 1) Software Validation
The OncoStudio contains basic document level of concern software was designed and developed according to a software development process and was verified and validated. Software information is provided in accordance with FDA guidance:
- "Content of Premarket Submissions for Device Software Functions," dated June 14, 2023. .
- 2) Performance characteristics
{13}------------------------------------------------
We collected training data from mainly three datasets of source : OneMedNet, Yonsei Severance Hospital, and University Hospital Basel, Switzerland. OneMedNet is a purchased set of CT data, mainly comprised of U.S.A. population. Yonsei Severance Hospital is located in South Korea, and we collected mainly Eastern population data from this source. The Basel data is known as TotalSegmentator dataset, which is open for public at this moment.
The collected data comprises a total of 2,438 dataset, consisting of 315 datasets from the US, 871 from Korea, and 1,252 from Europe.
The data was constructed with various ethnics (White, Black, Asian, Hispanic, Latino, African, American, etc.), and the training model can be obtained by performing generalization without differences according to ethnicity.
We allocated 310 data as training dataset, which is more than one-tenth of the total 2,438 dataset, and used the remaining data(2128) for training.
For evaluation, we created a test dataset that was not involved in any kind of training process. The splitting was performed at the patient level to ensure that images from the same patient were not present in more than one dataset. A comprehensive audit was conducted to confirm the integrity of the data-splitting process and ensure that no patient overlap occurred between datasets.
Ground truth seqmentations were established by three radiation oncologists following international clinical quidelines.
#### a) Ground Truthing
The ground truth annotations for the dataset of Yonsei Severance Hospital(Korea) and OneMedNet(U.S) were established by three different radiation oncologists with 3-20 years of clinical practice following RTOG and clinical quidelines using manual annotation. The radiation oncologists included associate professor, assistant professor, and radiation oncologist resident from two institutions (Yonsei Cancer Center, Samsung Seoul Hospital)
- . Ground Truthing process
- First, the 1 radiation oncologist manually delineated the organs
- Second, seqmentation results generated by 1 radiation oncologist are sequentially edited and confirmed by 2 radiation oncologists. In this editing process, the first radiation oncologist makes corrections, and the corrected results are received and finalized by another radiation oncologist.
{14}------------------------------------------------
In case of University Hospital Basel (Europe) dataset is public data comprising 104 anatomical structures. A total of 1,368 CT images were randomly sampled from the years 2012, 2016, and 2020 from the University Hospital Basel through picture archiving and communication system (PACS). The Nora Imaging Platform was used for manual segmentation and further refinement of generated segmentations for ground truth. Segmentation was supervised by two physicians with 3 (M.S.) and 6 years (H.B.) of experience in body imaging, respectively.
#### b) Training
Out of a total of 2,438 images, 2,128 were allocated as training data. The allocated training data consists of 731 images from Yonsei Severance Hospital(Republic of Korea), 194 images from OneMedNet(U.S.A), and 1203 images from University Hospital Basel(Switzerland).
The training datasets consist of 62% of Contrast CT and 38% of Non-Contrast CT. The study population comprises 62% males and 38% females, with 22% under 49 years old, 47% aged 50-70 years, and 31% over 70 years old.
The data was constructed with various ethnics (White, Black, Asian, Hispanic, Latino, African, American, etc.), and the result can be obtained by performing generalization without performance differences according to ethnicity.
The acquired data encompasses CT manufacturers such as GE (2%), Siemens (72%), Philips (9%), Toshiba (11%), and unknown manufacturer(6%)
#### c) Segmentation Performance Test
A standalone performance test was conducted to compare the contouring capabilities of OncoStudio.
The dataset used in this test comprises a total of 310 CT images, with 140, 121, and 49 images collected from Yonsei Severance Hospital (Republic of Korea), OneMedNet (U.S.A.), and University Hospital Basel (Switzerland), respectively, each meeting the established inclusion criteria. All data used during the standalone performance evaluation was composed independently of product development training.
The images consist of 54% of Contrast CT and 46% of Non-Contrast CT. The study population comprises 58% males and 42% females, with 13% under 49 years old, 54% aged 50-70 years, and 28% over 70 years old.
The data was constructed with various ethnics (White, Black, Asian, Hispanic, Latino, African, American, etc.), and the result can be obtained by performing generalization without performance differences according to ethnicity.
{15}------------------------------------------------
The acquired data encompasses CT manufacturers such as GE (2%), Siemens (56%), Philips (14%), Toshiba (22%), and unknown manufacturer(6%)
Ground truth segmentations were established by three radiation oncologists following international clinical guidelines.
For the structures being compared, the mean Dice coefficient (DSC) of structures for each anatomical region (Head & Neck, Thorax, Abdomen, and Pelvis) should meet the established criteria.
3) Cybersecurity
· "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions", on September 27, 2023
#### 9. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92]
There are no significant differences between the subject, predicate and reference devices, K230685 and K232899 that would adversely affect the use of the product. It is substantially equivalent to the predicate device in indications for use and technology characteristics.
#### 10. Conclusion [21 CFR 807.92(b)(3)]
In according with the Federal Food & Drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification, concludes that the OncoStudio is substantially equivalent in safety and effectiveness to the predicate device as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.