K242984 · Ortho Development Corp. · JWH · Oct 25, 2024 · Orthopedic
Device Facts
Record ID
K242984
Device Name
BKS Revision System
Applicant
Ortho Development Corp.
Product Code
JWH · Orthopedic
Decision Date
Oct 25, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
1. Loss of joint configuration and joint function. 2. Osteoarthritis of the knee joint. 3. Rheumatoid arthritis of the knee joint. 4. Post-traumatic arthritis of the knee joint. 5. Valgus, varus, or flexion deformities of the knee joint. 6. Revision procedures where other treatments or devices have failed. The BKS, BKS TriMax, and BKS Revision are intended for total knee arthroplasty procedures. The BKS is indicated in the salvage of previously failed surgical attempts where bone loss does not require the use of augments or stem extensions and where collateral ligaments may be relied upon for medial/lateral (M/L) stability. BKS TriMax is intended in the salvage of previously failed surgical attempts where bone loss does not require the use of augments or stem extensions, where collateral ligaments may be relied upon for medial/lateral stability, where postoperative flexion up to 150° may be desirable, and the patient meets all indications requirements. The BKS Revision system is indicated in the salvage of previously failed surgical attempts where bone loss may require the use of augments, sleeves or stem extensions. The BKS Revision system is indicated for cemented use only. The porous-coated metaphyseal sleeves are intended for cemented and uncemented applications. The BKS Revision CK Tibial Insert is indicated for use with the BKS Revision Modular femoral and Modular, Offiset, or Sleeve tibial components where collateral ligaments may not be relied upon for medial/lateral (M/L) stability.
Device Story
BKS Revision System is a knee joint prosthesis used in total knee arthroplasty and revision procedures. The subject device, a Sleeve Junction Box, acts as a mechanical interface component within the modular femoral assembly. It utilizes a locking taper interface to secure the A/P position of the femoral component relative to the revision sleeves. Used in clinical settings by orthopedic surgeons, the device facilitates reconstruction of the knee joint in patients with bone loss or failed prior surgeries. The system provides structural support and stability to the joint. The device is implanted during surgery; it does not involve electronic inputs, software, or AI. Benefits include restoration of joint function and stability in complex revision cases.
Clinical Evidence
Bench testing only. Evaluation included geometric analysis, axial fatigue strength per ASTM F1814, disassembly strength, assembly evaluation, and packaging evaluation. No clinical data provided.
Technological Characteristics
Knee joint patellofemorotibial semi-constrained cemented prosthesis. Components include a Sleeve Junction Box with a locking taper interface. Materials and manufacturing processes are identical to the predicate. Sterilization and packaging remain unchanged. Mechanical design utilizes a locking mechanism for A/P positioning.
Indications for Use
Indicated for patients requiring total knee arthroplasty due to loss of joint function, osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, or joint deformities (valgus, varus, flexion). Also indicated for revision procedures following failed surgical attempts, including cases requiring augments, sleeves, or stem extensions.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 25, 2024
Ortho Development Corp. Drew Weaver Director of Regulatory Affairs 12187 S. Business Park Drive Draper, Utah 84020
Re: K242984
Trade/Device Name: BKS Revision System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH, MBH Dated: October 25, 2024 Received: September 26, 2024
Dear Drew Weaver:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Peter G. Allen -S Allen-S
Digitally signed by Peter G.
Date: 2024.10.25 15:56:31
-04'00'
For: Lixin Liu, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K242984
Device Name BKS Revision System
Indications for Use (Describe)
- 1. Loss of joint configuration and joint function.
- 2. Osteoarthritis of the knee joint.
- 3. Rheumatoid arthritis of the knee joint.
- 4. Post-traumatic arthritis of the knee joint.
- 5. Valgus, varus, or flexion deformities of the knee joint.
- 6. Revision procedures where other treatments or devices have failed.
The BKS, BKS TriMax, and BKS Revision are intended for total knee arthroplasty procedures.
The BKS is indicated in the salvage of previously failed surgical attempts where bone loss does not require the use of augments or stem extensions and where collateral ligaments may be relied upon for medial/lateral (M/L) stability.
BKS TriMax is intended in the salvage of previously failed surgical attempts where bone loss does not require the use of augments or stem extensions, where collateral ligaments may be relied upon for medial/lateral stability, where postoperative flexion up to 150° may be desirable, and the patient meets all indications requirements.
The BKS Revision system is indicated in the salvage of previously failed surgical attempts where bone loss may require the use of augments, sleeves or stem extensions.
The BKS Revision system is indicated for cemented use only. The porous-coated metaphyseal sleeves are intended for cemented and uncemented applications.
The BKS Revision CK Tibial Insert is indicated for use with the BKS Revision Modular femoral and Modular, Offiset, or Sleeve tibial components where collateral ligaments may not be relied upon for medial/lateral (M/L) stability.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for "Ortho Development". The logo consists of a purple circle with gray lines emanating from the top left, followed by the word "ORTHO" in purple block letters. Below "ORTHO" is the word "DEVELOPMENT" in a smaller, lighter font, with a registered trademark symbol.
# 510(k) Summary of Safety & Effectiveness
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided:
| Name of the Sponsor: | Ortho Development® Corporation<br>12187 South Business Park Drive<br>Draper, Utah 84020 | |
|-------------------------------------------------|-----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Contact: | Name: | Drew Weaver |
| | Position: | Director of Regulatory Affairs |
| | Address: | 12187 S. Business Park Drive<br>Draper, UT 84020 USA |
| | Telephone: | (801) 553-9991 |
| | Email: | dweaver@orthodevelopment.com |
| Date Prepared: | October 25, 2024 | |
| Submission Type: | Special | |
| Proprietary Name: | BKS Revision System | |
| Common Name: | BKS Revision Sleeve Junction Box | |
| Product Code /<br>Classification: | JWH | Prosthesis, knee, patellofemorotibial, semi-<br>constrained, cemented, polymer/metal/polymer<br><br>21 CFR 888.3560: Knee joint patellofemorotibial<br>polymer/metal/polymer semi-constrained cemented<br>prosthesis. |
| Associated Product<br>Code(s) / Classification: | MBH | Prosthesis, knee, patello/femorotibial, semi-<br>constrained, uncemented, porous, coated,<br>polymer/metal/polymer<br><br>21 CFR 888.3565: Knee joint patellofemorotibial<br>metal/polymer porous-coated uncemented prosthesis. |
| Device Class: | Class 2 | |
| Primary Predicate<br>Device: | BKS® Revision Sleeves System (K181569) | |
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image shows the logo for "Ortho Development". The logo consists of a purple circle with three curved lines emanating from the top left, followed by the word "ORTHO" in purple, block letters. Below the word "ORTHO" is the word "DEVELOPMENT" in gray, smaller letters.
## 1.0 Device Description:
The purpose of this special 510(k) submission is to request clearance for BKS Revision Sleeve Junction Box of the BKS Revision System which is an extension of Ortho Development's previously cleared predicate BKS Revision Sleeves System (K181569).
The subject device incorporates the same locking taper interface for mating with the BKS Revision Sleeves and the same principle of operation for locking the A/P position relative to the BKS Revision Modular Femoral Component. The subject device is manufactured in the same manner from the same material. The subject device also employs the same packaging components and sterilization method.
Differences between the subject and predicate device include small adjustments to the envelope geometry, internal configuration, and driving feature for assembly during use. These changes have been made to optimize the Sleeve Junction Box for use specifically with femoral sleeves. The mechanics of the connection and locking mechanism are the same as the predicate device.
### 2.0 Indication for Use:
- 1. Loss of joint configuration and joint function.
- 2. Osteoarthritis of the knee joint.
- 3. Rheumatoid arthritis of the knee joint.
- 4. Post-traumatic arthritis of the knee joint.
- 5. Valgus, varus, or flexion deformities of the knee joint.
- 6. Revision procedures where other treatments or devices have failed.
The BKS, BKS TriMax, and BKS Revision are intended for total knee arthroplasty procedures.
The BKS is indicated in the salvage of previously failed surgical attempts where bone loss does not require the use of augments or stem extensions and where collateral ligaments may be relied upon for medial/lateral (M/L) stability.
BKS TriMax is intended in the salvage of previously failed surgical attempts where bone loss does not require the use of augments or stem extensions, where collateral ligaments may be relied upon for medial/lateral stability, where postoperative flexion up to 150° may be desirable, and the patient meets all indications/contraindications requirements.
The BKS Revision system is indicated in the salvage of previously failed surgical attempts where bone loss may require the use of augments, sleeves or stem extensions.
The BKS Revision system is indicated for cemented use only. The porous-coated metaphyseal sleeves are intended for cemented and uncemented applications.
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the logo for "Ortho Development". The logo consists of a purple circle with three gray arcs emanating from the left side, followed by the word "ORTHO" in purple, block letters. Below the word "ORTHO" is the word "DEVELOPMENT" in a smaller, gray font with a trademark symbol.
The BKS Revision CK Tibial Insert is indicated for use with the BKS Revision Modular femoral and Modular, Offset, or Sleeve tibial components where collateral ligaments may not be relied upon for medial/lateral (M/L) stability.
### 3.0 Comparison of Technological Characteristic:
The Sleeve Junction Box is technologically the same as the already cleared Tapered Junction Box that was a component of BKS Revision Sleeves (K181569). It is the same in terms of indications for use/intended use, basic design, device materials, and principles of operation.
Differences between the subject and predicate device include small adjustments to the envelope geometry, internal configuration, and driving feature for assembly during use. These changes have been made to optimize the Sleeve Junction Box for use specifically with femoral sleeves. The mechanics of the connection and locking mechanism are the same as the predicate device.
Labeling, packaging and sterilization do not change.
The modified device was evaluated using well established methods and recognized consensus standards to confirm that it meets the original design inputs of the predicate.
#### 4.0 Performance Data:
The modified device was evaluated using well established methods and recognized consensus standards to confirm that it meets the original design inputs of the predicate.
The evaluation includes: geometric analysis, axial fatigue strength per ASTM F1814, disassembly strength, assembly evaluation, and packaging evaluation.
These performance data demonstrate that the original design inputs of the predicate are met.
### 5.0 Substantial Equivalence Conclusion:
Verification and validation has been conducted according to design control procedures within a risk analysis framework to determine that the design outputs meet the design inputs using the same test criteria and methodology as the predicate K181569.
Based on similarities in indication for use/intended use, technological characteristic, basic design, device material, and principle of operation, the BKS Revision Sleeve Junction Box of the BKS Revision System is considered substantially equivalent to the previously cleared predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.