CROMA Electrosurgical Generator (PRD-EMR-050)

K242983 · Creo Medical, Ltd. · GEI · Feb 5, 2025 · General, Plastic Surgery

Device Facts

Record IDK242983
Device NameCROMA Electrosurgical Generator (PRD-EMR-050)
ApplicantCreo Medical, Ltd.
Product CodeGEI · General, Plastic Surgery
Decision DateFeb 5, 2025
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The CROMA Electrosurgical Generator is intended to provide Radiofrequency (RF) and microwave (MW) energy for cutting, coagulation, and ablation of soft tissue, when used in conjunction with compatible electrosurgical instruments and accessories. The Creo Medical Electrosurgical System is not intended for use in cardiac procedures. The Reusable Interface Cable is for connection only of a compatible Creo Medical Instrument (Surgical Accessory) to the CROMA Electrosurgical Generator to deliver Radiofrequency (RF) and/or Microwave (MW) energy for the cutting, coagulation and ablation of tissue via endoscopic access.

Device Story

Tabletop, mains-powered electrosurgical generator; provides bipolar radiofrequency (RF) and microwave (MW) energy for soft tissue cutting, coagulation, and ablation. Used in clinical settings by physicians; connects to compatible Creo Medical instruments via interface cable. Device transforms electrical mains input into controlled RF (400 kHz) and MW (5.8 GHz) outputs. Updated front panel features LCD screen (replacing VFD) and modified keypad membrane. Healthcare providers operate device via front panel controls; output delivered through instruments to target tissue. Benefits include precise tissue management during endoscopic procedures. Device is non-networked.

Clinical Evidence

No clinical data. Bench testing only. Includes electrical safety, EMC, software verification/validation (IEC 62304), and mechanical/usability testing (power, initialization, accessory connection, UI assessment).

Technological Characteristics

Tabletop ESU; Class I, Type CF patient circuit. Energy: Bipolar RF (400 kHz) and MW (5.8 GHz). LCD interface. Mains powered (100-240V). Complies with IEC 60601-1, 60601-2-2, 60601-2-6, and 60601-1-2. Software developed per IEC 62304 (enhanced documentation level).

Indications for Use

Indicated for cutting, coagulation, and ablation of soft tissue using compatible electrosurgical instruments via endoscopic access. Not for use in cardiac procedures.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue. February 5, 2025 Creo Medical Ltd Diane Davis RAQA Product Lead Unit 2 Beaufort Park Beaufort Park Way Chepstow, NP16 5UH United Kingdom Re: K242983 Trade/Device Name: CROMA Electrosurgical Generator (PRD-EMR-050) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI, NEY, KNS Dated: January 6, 2025 Received: January 6, 2025 Dear Diane Davis: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory {2}------------------------------------------------ assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Digitally signed by James James H. Jang H. Jang H. Jang -S Date: 2025.02.05 13:05:38 -2 -05'00' For Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K242983 Device Name CROMA Electrosurgical Generator (PRD-EMR-050) Indications for Use (Describe) The CROMA Electrosurgical Generator is intended to provide Radiofrequency (RF) and microwave (MW) energy for cutting, coagulation, and ablation of soft tissue, when used in conjunction with compatible electrosurgical instruments and accessories. The Creo Medical Electrosurgical System is not intended for use in cardiac procedures. The Reusable Interface Cable is for connection only of a compatible Creo Medical Instrument (Surgical Accessory) to the CROMA Electrosurgical Generator to deliver Radiofrequency (RF) and/or Microwave (MW) energy for the cutting, coagulation and ablation of tissue via endoscopic access. X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # K242983 510(k) Summary ### CROMA Electrosurgical Generator Date Prepared: February 4, 2025 #### CONTACT DETAILS | Applicant Name: | Creo Medical Ltd. | |------------------------------|----------------------------------------------------------------| | Applicant Address: | Unit 2 Beaufort Park, Beaufort Park Way, Chepstow NP16 5UH, UK | | Applicant Contact Telephone: | +44 7899 306000 | | Applicant Contact: | Diane Davis | | Applicant Contact Email: | diane.davis@creomedical.com | #### DEVICE NAME | Device Trade Name | CROMA Electrosurgical Generator (PRD-EMR-050) | |---------------------|----------------------------------------------------------------| | Common Name | Electrosurgical cutting and coagulation device and accessories | | Classification Name | Electrosurgical, Cutting & Coagulation & Accessories | | Regulation Number | 878.4400 | | Product Code(s) | GEI, NEY, KNS | #### LEGALLY MARKETED PREDICATE DEVICES K230328 Speedboat Flush SB1 Instrument with Croma Electrosurgical Generator, Product Code KNS K223138 AB1 Electrosurgical Instrument, Creo Electrosurgical System, Product Code NEY #### DEVICE DESCRIPTION SUMMARY The CROMA Electrosurgical Generator is an Electrosurgical unit (ESU) to be used with compatible electrosurgical instruments and accessories. The CROMA Electrosurgical Generator is a tabletop, non-network connected, mains powered ESU. It comprises two distinct energy sources for the independent generation of bipolar radiofrequency (RF) and microwave (MW) energies, intended to supply electrosurgical power for cutting, and ablation of soft tissues, when connected to compatible instruments via the interface cable. The CROMA Electrosurgical Generator incorporates proprietary software developed by Creo Medical Ltd. The CROMA Electrosurgical Generator subject of this submission incorporates a new front panel, which has been updated from a Vacuum Fluorescent Display (VFD) in the predicate device to a Liquid Crystal Display (LCD) screen in the subject device. The subject and predicate devices have identical principles of operations: - Supply suitable Creo Medical instruments with appropriate power for their surgical uses at a MW frequency of 5.8 GHz for the purpose of coagulation, cauterisation and/or ablation. - . Supply suitable Creo Medical instruments with appropriate power for their surgical uses at a nominal RF of 400 kHz for the purpose of cutting. #### INDICATIONS FOR USE STATEMENT The CROMA Electrosurgical Generator is intended to provide Radiofrequency (RF) and microwave (MW) energy for cutting, coagulation, and ablation of soft tissue, when used in conjunction with compatible electrosurgical instruments and accessories. The Creo Medical Electrosurgical System is not intended for use in cardiac procedures. {5}------------------------------------------------ The Reusable Interface Cable is for connection only of a compatible Creo Medical Instrument (Surgical Accessory) to the CROMA Electrosurgical Generator to deliver Radiofrequency (RF) and/or Microwave (MW) energy for the cutting, coagulation and ablation of tissue via endoscopic access. #### INDICATIONS FOR USE COMPARISON The subject device has the same Indications for Use as the predicate for use statement for the subject device has been reworded for readability and clarity by removing the specific reference of the compatible instruments. Instead, compatible instruments are provided in a table form in the Instructions for Use, right under the Intended Use Statement. Therefore, this formatting change does not raise concerns of safety and effectiveness. #### TECHNOLOGICAL COMPARISON The subject and predicate devices have the following same technological characteristics: - . Energy inputs - Energy outputs - Principles of operations - . Functionalities of the control console (front panel), including push buttons, output sockets for interface cable connection, symbols, LED lights and audio tones in compliance with IEC 60601 requirements. The subject device has been modified to modernise the visual of the user interface and upgrade the screen from a Vacuum Fluorescent Display (VFD) to a Liquid Crystal Display (LCD). As a result, the following technological differences exist between the subject and predicate devices: - . Front panel screen changed from VFD to LCD - Front panel keypad membrane updated (aesthetics, features location) - . Software changed to support the new LCD screen, as well as include other minor software updates The differences in the subject device technological characteristics do not raise concerns of safety and effectiveness. | Characteristic | Subject<br>CROMA Electrosurgical Generator<br>PRD-EMR-050 | Primary Predicate<br>CROMA Electrosurgical Generator<br>7-EMR-050 (K230328) | Secondary Predicate<br>CROMA Electrosurgical Generator<br>7-EMR-050 (K223138) | |---------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Classification: | Class II | SAME | SAME | | Regulation: | 21 CFR 878.4400 | 21 CFR 876.4300 | SAME | | Product Code: | GEI – Electrosurgical, Cutting & Coagulation & Accessories | & KNS – Unit, Electrosurgical, Endoscopic | NEY – System, Ablation, Microwave And Accessories | | Intended Use /<br>Indications For<br>Use CROMA<br>Electrosurgical<br>Generator: | The CROMA Electrosurgical Generator is intended to provide Radiofrequency (RF) and Microwave (MW) energy for cutting, coagulation, and ablation of soft tissue, when used in conjunction with compatible electrosurgical instruments and accessories. | The Electrosurgical Generator provides microwave (MW) energy to compatible Creo Medical instruments AB1 MicroBlate Flex and NP1 MicroBlate Fine instruments, intended for coagulation (ablation) of soft tissue | The Electrosurgical Generator provides microwave (MW) energy to compatible Creo Medical instruments AB1 MicroBlate Flex and NP1 MicroBlate Fine instruments, intended for coagulation (ablation) of soft tissue | | | | The Electrosurgical Generator provides microwave (MW) energy to compatible Creo Medical instruments HS1 SlypSeal | The Electrosurgical Generator provides microwave (MW) energy to compatible Creo Medical instruments HS1 SlypSeal | | Characteristic | Subject | Primary Predicate | Secondary Predicate | | | CROMA Electrosurgical Generator<br>PRD-EMR-050 | CROMA Electrosurgical Generator<br>7-EMR-050 (K230328) | CROMA Electrosurgical Generator<br>7-EMR-050 (K223138) | | | The Creo Medical Electrosurgical<br>System is not intended for use in<br>cardiac procedures. | Instrument, intended for<br>coagulation (hemostasis and<br>cauterization) of soft tissue. | Instrument, intended for<br>coagulation (hemostasis and<br>cauterization) of soft tissue. | | | | The Electrosurgical Generator<br>provides microwave (MW) and<br>radiofrequency (RF) energy to<br>compatible Creo Medical<br>Speedboat Instruments, intended<br>for coagulation (hemostasis and<br>cauterization) and cutting soft<br>tissue. | The Electrosurgical Generator<br>provides microwave (MW) and<br>radiofrequency (RF) energy to<br>compatible Creo Medical<br>Speedboat Instruments, intended<br>for coagulation (hemostasis and<br>cauterization) and cutting soft<br>tissue. | | | | The Electrosurgical System is not<br>intended for use in cardiac<br>procedures. | The Electrosurgical System is not<br>intended for use in cardiac<br>procedures. | | Display: | Liquid Crystal display (LCD)<br>supporting Alphanumeric, Chinese<br>and Japanese characters. | Vacuum fluorescent display (VFD)<br>supporting Alphanumeric<br>characters. | Vacuum fluorescent display (VFD)<br>supporting Alphanumeric<br>characters. | | Mains power<br>input: | Input Voltage: 100 V - 120 V / 220<br>V - 240 V. Maximum Input Power:<br>450 W.<br>Frequency: 50 / 60 Hz, Single phase. | SAME | SAME | | Protective<br>class (IEC<br>60601-1): | Equipment is Class I<br>Requires a protective ground<br>connection. | SAME | SAME | | Patient circuit,<br>as per IEC<br>60601-1: | Type CF | SAME | SAME | | Degree of IP<br>protection: | IPOO | SAME | SAME | | Energy output: | RF output: 400 Hz<br>MW output: 5.8 GHz | SAME | SAME | | Maximum<br>voltage: | RF: 460 Vpeak<br>MW: 79 Vpeak (50 Ohm load) | SAME | SAME | | RF output<br>modality: | Bipolar, single waveform | SAME | SAME | | RF output<br>operating<br>frequency: | 396.7 kHz ± 2.0 kHz | SAME | SAME | | Characteristic | Subject<br>CROMA Electrosurgical Generator<br>PRD-EMR-050 | Primary Predicate<br>CROMA Electrosurgical Generator<br>7-EMR-050 (K230328) | Secondary Predicate<br>CROMA Electrosurgical Generator<br>7-EMR-050 (K223138) | | RF output<br>capability<br>(Biomed<br>mode): | 200 W maximum into a rated load<br>of 400 Ohm | SAME | SAME | | RF output<br>capability<br>(treatment<br>mode): | 35 W (Max) into a rated load of 400 Ohm | SAME | SAME | | RF output<br>adjustment: | 15 W to 35 W at the rated load of<br>400 Ohm. 10 s ON, 30 s OFF, 1 hour | SAME | SAME | | MW output<br>modality: | Microwave, single waveform. | SAME | SAME | | MW output<br>operating<br>frequency: | 5.8 GHz ± 1 MHz | SAME | SAME | | MW output<br>capability<br>(treatment<br>mode): | 62 W maximum into a rated load of 50 Ohm | SAME | SAME | | MW output<br>power: | AB1 – 27 W (for use with Interface<br>Cable 2m only) NP1 – 62 W | AB1 - 25W NP1 - SAME | AB1 - 25W NP1 - SAME | | MW output<br>duration: | Defined in the generator IFU:<br>RS2 - 10 s ON, 1 s OFF x 3, then 268<br>s OFF HS1 – 10 s ON, 1 s OFF x 5,<br>then 246 s OFF SB1-L - 5 s ON, 1 s<br>OFF x 5, then 268 s OFF<br>AB1 – 180 s ON, 30 s OFF x 9, then<br>28 min 30 sec OFF NP1 - 120 s ON,<br>30 s OFF x 8, then 40 min OFF | Defined in the generator IFU: RS2 -<br>SAME<br>HS1 – SAME SB1-L – SAME AB1 –<br>SAME<br>NP1 – 120 s ON, 120 s OFF x 8, then<br>28 min OFF | Defined in the generator IFU: RS2 -<br>SAME<br>HS1 – SAME SB1-L – SAME AB1 –<br>SAME<br>NP1 – 120 s ON, 120 s OFF x 8, then<br>28 min OFF | {6}------------------------------------------------ {7}------------------------------------------------ #### NON-CLINICAL AND/OR CLINICAL TESTS SUMMARY #### Electrical safety and electromagnetic compatibility (EMC) Electrical safety and EMC testing was conducted on the subject device as part of the Creo Medical System. The system complies with IEC 60601-1 Ed. 3.2, IEC 60601-2-2 Ed. 6.0, IEC 60601-2-6 Ed. 2.1, and IEC 60601-1-2:2014. #### Software verification and validation testing The software for this device was developed, verified and validated in with IEC 62304 Ed. 1.1. The software for this device requires an enhanced documentation level, as a failure or flaw of the software could present a hazardous situation with a probable risk of death or serious injury to the patient or user of the device. #### Mechanical and Usability testing {8}------------------------------------------------ At a high level, the following mechanical and usability testing have been performed on the subject device: - . Device power on/off, initialisation, standby, timeout testing - . Settings, language, and accessory selection testing - . Accessory connection / reconnection testing - . Usability assessment of the changes to the user interface #### Animal & Clinical Studies Not Applicable. #### CONCLUSION Technological differences between subject and predicate devices do not raise new questions of safety or effectiveness. Based on the comparison of the indications for use, technological characteristics, and performance test results to the same of the predicate predicate, the subject CROMA Electrosurgical Generator has been shown to be substantially equivalent to the predicate devices for the requested indications for use.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...