K242960 · Boston Scientific Corporation · OTM · Nov 15, 2024 · General, Plastic Surgery
Device Facts
Record ID
K242960
Device Name
AdVance™ XP Male Sling System (720163-03)
Applicant
Boston Scientific Corporation
Product Code
OTM · General, Plastic Surgery
Decision Date
Nov 15, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AdVance™ XP Male Sling System is intended for the treatment of male stress urinary incontinence (SUI) by the placement of a suburethral sling.
Device Story
AdVance XP Male Sling System is a sterile, single-pack surgical kit for treating male stress urinary incontinence (SUI). The system includes a mesh sling with tissue anchors, needle passers, a retractor ring, and stay hooks. Used in a hospital/healthcare facility setting by a physician, the device functions by repositioning the urethra to restore continence. The mesh sling is implanted suburethrally to provide support. The procedure involves using needle passers to assist in the placement of the sling. The device is a permanent implant. Clinical benefit is the reduction or resolution of SUI symptoms through mechanical urethral support.
Clinical Evidence
Bench testing only. Performance evaluated using non-aged and accelerated-aged (37 months) samples. Evaluations included mesh thickness, weave characteristics, pore size, density, tensile strength, and burst strength per FDA guidance for surgical mesh. Biocompatibility confirmed per ISO 10993-1 for permanent implant contact.
Technological Characteristics
Suburethral mesh implant with tissue anchors. Materials evaluated per ISO 10993-1. Sterilization via ethylene oxide (EO) to SAL 10^-6 per ISO 11135. Bench testing performed per FDA Surgical Mesh guidance (1999).
Indications for Use
Indicated for treatment of male stress urinary incontinence (SUI) via suburethral sling placement.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 15, 2024
Boston Scientific Corporation Natalie Asprey Senior Regulatory Affairs Specialist 100 Boston Scientific Wav Marlborough, Massachusetts 01752
Re: K242960
Trade/Device Name: AdVance™ XP Male Sling System (720163-03) Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: II Product Code: OTM Dated: September 25, 2024 Received: September 25, 2024
Dear Natalie Asprey:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}------------------------------------------------
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica K. Nguyen -S
Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
Submission Number (if known)
K242960
Device Name
AdVance™ XP Male Sling System (720163-03)
Indications for Use (Describe)
The AdVance™ XP Male Sling System is intended for the treatment of male stress urinary incontinence (SUI) by the placement of a suburethral sling.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Urology 100 Boston Scientific Way Marlborough, MA 01752
# 510(k) Summary
#### Submitter
Boston Scientific Corporation Urology Division 100 Boston Scientific Way Marlborough, MA 01752 USA
Primary Contact: Natalie Asprey Senior Regulatory Affairs Specialist Email: Natalie.Asprey@bsci.com Date Prepared: 25 September 2024
Alternate Contact: Kristine Higgins Regulatory Affairs Director Email: Kristine.Higgins@bsci.com Telephone: 952-930-6475
#### Proposed Device K242960
Trade Name: AdVance™ XP Male Sling System Model Number: 720163-03 Regulatory Class: II Product Code: OTM Product Code Name: Mesh, Surgical, For Stress Urinary Incontinence, Male Common/Usual/Classification Name: Surgical Mesh Classification Number: 21 CFR 878.3300
#### Predicate Device K211847
Trade Name: AdVance™ XP Male Sling System Model Number: 720163-03 Regulatory Class: II Product Code: OTM Product Code Name: Mesh, Surgical, For Stress Urinary Incontinence, Male Common/Usual/Classification Name: Surgical Mesh Classification Number: 21 CFR 878.3300 Manufacturer: Boston Scientific Corporation
{5}------------------------------------------------
## Device Description
The AdVance XP Male Sling System is provided as a sterile, single-pack kit composed of the following primary components: Male Sling, Needle Passers (2), Retractor Ring, and Stay Hooks. The subject of this premarket notification is around the Male Sling component of the AdVance XP Male Sling System, specifically the implantable portion of the Male Sling assembly, which is the mesh sling and the tissue anchors. There are no modifications to the Needle Passers, or Retractor Ring and Stay Hooks (which form the Retraction System) as a part of this submission. Note that the Retraction System are comprised of purchased, off-the-shelf components cleared through K791665.
## Environment of Use
Healthcare facility/Hospital
### Indications for Use
The AdVance XP Male Sling System is intended for the treatment of male stress urinary incontinence (SUI) by the placement of a suburethral sling.
## Operating Principle
The AdVance XP Male Sling is a suburethral mesh implant designed to treat male incontinence by repositioning the urethra. AdVance XP needle passers assist in the placement of the male sling. AdVance XP Male Sling System is an implanted, surgical device that penetrates inside the body. Its intended purpose is for the treatment of male stress urinary incontinence (SUI) by the placement of a suburethral sling.
### Comparison of Technological Characteristics with the Predicate Device
The proposed AdVance XP Male Sling System and the predicate device (K211847) have identical indications for use, operating principle, finished good device dimensions and features, environment of use, sterilization method, shelf life, Needle Passers, and Retraction System.
The primary differences between the proposed AdVance XP Male Sling System and the predicate device (K211847) are the material composition and the material wash process (including processing aids).
### Substantial Equivalence
The direct comparison of key characteristics, along with data presented in this premarket notification, demonstrates that the proposed AdVance XP Male Sling System is substantially equivalent to the predicate AdVance XP Male Sling System device, which was cleared by the FDA under premarket notification K211847.
#### Sterility
There were no changes in the sterilization process used for the AdVance XP Male Sling System, however an ethylene oxide (EO) sterilization qualification was performed to confirm that the proposed device continues to meet a SAL of 10-6 or better and in accordance with global BSC procedures and ISO 11135. Additionally, a microbiology evaluation was performed in accordance
{6}------------------------------------------------
with global BSC procedures, ISO 11737-1, and ANSI AAMI ST72 to ensure the proposed device continues to meet bioburden and endotoxin requirements.
#### Biocompatibility
A biological evaluation was performed per ISO 10993-1 Biological Evaluation of Medical Devices – Part 1 and FDA Guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" to demonstrate that the proposed AdVance XP Male Sling System device is biocompatible for its intended use. This biological evaluation included assessment of the materials of construction (including processing of the materials and processing aids), toxicological risk assessment, and biological testing.
Results of the material characterization, toxicological risk assessment, and biological testing support the biocompatibility of the device. The biological evaluation conducted in accordance with ISO 10993-1 and FDA Guidance (2023) confirms that the AdVance XP Male Sling System is biocompatible for its intended use per its ISO 10993-1 categorization as a Implant, Tissue/Bone, Permanent (> 30 day) contact device. The results of the biological evaluation of the device modifications support the proposed AdVance XP Male Sling System is substantially equivalent to the predicate (K211847).
### Performance Data
Performance testing was conducted on non-aged (T=0) samples and samples that were accelerated aged for 37 months (in order to support the 3 year device shelf life). 37 month real-time aging is ongoing. To demonstrate substantial equivalence of the AdVance XP Male Sling System to the predicate device (K211847), technological characteristics and performance criteria were evaluated using design verification bench testing. The FDA Guidance for the Preparation of Premarket Notification Application for a Surgical Mesh (issued on 02Mar1999) was utilized for preclinical mesh attribute evaluations. Mesh product characteristics such as mesh thickness, mesh weave characteristics, pore size, mesh density, tensile strength, and burst strength were evaluated through mesh material specification controls. Additional evaluations were performed through design testing on the final product assembly using internally developed product design specifications. The results of the performance testing demonstrate that the technological characteristics and performance criteria of the proposed AdVance XP Male Sling System are comparable to the predicate device (K211847) and functions as intended, equivalent to a device currently on the market for the same indications for use.
### Conclusion
Based on the indication for use, technological characteristics, and performance data it can be concluded that the proposed AdVance XP Male Sling System is substantially equivalent to the predicate device AdVance XP Male Sling System (K211847) and is appropriate for the indications for use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.