Bioceramic Root Canal Sealer and Repair Materials (i-MTA BP)
K242934 · Longly Biotechnology (Wuhan) Co., Ltd. · KIF · Dec 12, 2024 · Dental
Device Facts
Record ID
K242934
Device Name
Bioceramic Root Canal Sealer and Repair Materials (i-MTA BP)
Applicant
Longly Biotechnology (Wuhan) Co., Ltd.
Product Code
KIF · Dental
Decision Date
Dec 12, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3820
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
i-MTA BP Bioceramic Root Canal Sealer and Repair Material is developed for permanent root canal sealer and repair and surgical applications. - Repair of Root Perforation - Repair of Root Resorption - Root End Filling - Apexification - Pulp Capping
Device Story
i-MTA BP is a ready-to-use, premixed bioceramic paste for permanent root canal sealing and surgical endodontic repair. Composed of calcium silicate, zirconium oxide, tantalum pentoxide, and calcium sulfate; material is insoluble, radiopaque, and aluminum-free. Device is supplied in a preloaded syringe for direct application by dentists/endodontists in a clinical setting. Upon contact with moisture in dentin, the paste sets and hardens without shrinkage. The material acts as a physical barrier and sealing agent within the root canal system or surgical site. Benefits include biocompatibility, effective sealing of root structures, and promotion of healing in endodontic applications.
Clinical Evidence
No clinical data. Bench testing only. Performance testing per ISO 6876:2012 demonstrated: setting time 2.5h (req: 2-6h), solubility 2.80% (req: <3%), radiopacity >3mm Al, size change -0.50% (req: <1.0%), and pH 12.29 (req: 11.5-13.5). Biocompatibility testing included cytotoxicity, sensitization, irritation, acute/subchronic systemic toxicity, genotoxicity (TK gene mutation, bacterial reverse mutation), implantation, pulp capping, and endodontic usage tests, all meeting ISO 10993/7405 requirements.
Technological Characteristics
Premixed bioceramic paste in preloaded syringe. Composition: calcium silicate, zirconium oxide, tantalum pentoxide, calcium sulfate. Setting principle: moisture-activated hardening. Standards: ISO 6876:2012 (performance), ISO 10993-1, ISO 7405 (biocompatibility). No electronic components or software.
Indications for Use
Indicated for permanent root canal sealing, root perforation repair, root resorption repair, root end filling, apexification, and pulp capping in patients requiring endodontic treatment.
Regulatory Classification
Identification
A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.
Predicate Devices
Bioceramic Root Repair Material (C-Root BP) (K240365)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
December 12, 2024
Longly Biotechnology (Wuhan) Co., Ltd. % Shirely Kong RA Beijing Xinranyicheng Medicine & Technology Co., Ltd. A-15B01, Langqin International Building, No.168, Guang'anmen Outer Street, Xicheng District Beijing, Beijing 100053 China
Re: K242934
Trade/Device Name: Bioceramic Root Canal Sealer and Repair Materials (i-MTA BP) Regulation Number: 21 CFR 872.3820 Regulation Name: Root Canal Filling Resin Regulatory Class: Class II Product Code: KIF. Dated: September 14, 2024 Received: September 25, 2024
Dear Shirely Kong:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/2/Picture/3 description: The image contains the name "Bobak Shirmohammadi" stacked on top of each other. Below the name is the character string "-S". To the left of the name is a faded image of the letters "FDA".
For Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K242934
Device Name
Bioceramic Root Canal Sealer and Repair Materials
Indications for Use (Describe)
i-MTA BP Bioceramic Root Canal Sealer and Repair Material is developed for permanent root canal sealer and repair and surgical applications.
- Repair of Root Perforation
- Repair of Root Resorption
- Root End Filling
- Apexification
- Pulp Capping
| Type of Use (Select one or both, as applicable) | | |
|-------------------------------------------------|--|--|
| | | |
|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary K242934
This summary of 510(k) substantial equivalence is being submitted in accordance with requirements of 21 CFR 807.92.
## 1.0 Submitter's Information
Name: Longly Biotechnology (wuhan) Co., Ltd Address: No.02 of 2nd Floor, Building A2-2, Optics Valley Biomedical Park, Gaoxin Avenue No.858, East Lake High-tech Development Zone, Wuhan, China Tel: +86 - 13907166169 Contact: Yan Liang (Registration manager) Email: longlyyanliang466@163.com Date of Preparation: Sep.8,2024
## 2.0 Device Information
Trade Name: Bioceramic Root Canal Sealer and Repair Materials Model: i-MTA BP Common Name: Root canal filling resin Classification Name: Resin, Root Canal Filling Device Panel: Dental Classification Regulation: 21 CFR 872.3820 Device Classification: Class II Product Code: KIF
### 3.0 Identification of Predicate Device and Reference Device
| Information | Predicate Device | Reference Device |
|---------------------------|----------------------------------------------------|-----------------------------|
| 510(k) Number | K240365 | K092715 |
| Device Name | Bioceramic Root Repair Material<br>(C-Root BP) | IROOT BP PLUS |
| Common Name | Bioceramic Root Repair Material | Root Repair Material |
| Classification Name | Resin, Root Canal Filling | Resin, Root Canal Filling |
| Device Panel | Dental | Dental |
| Classification Regulation | 21 CFR 872.3820 | 21 CFR 872.3820 |
| Device Classification | Class II | Class II |
| Product Code | KIF | KIF |
| Manufacturer | Beijing C-Root Dental Medical<br>Devices Co., LTD. | Innovative BioCeramix, Inc. |
### 4.0 Device Description
{5}------------------------------------------------
i-MTA BP Bioceramic Root Canal Sealer and Repair Materials is a ready-to-use premixed bioceramic paste developed for permanent root canal sealer and repair and surgical applications. i-MTA BP is an insoluble, radiopaque and aluminum-free material based on a calcium silicate composition, which requires the presence of water to set and harden. i-MTA BP does not shrink during setting and demonstrates excellent physical properties. i-MTA BP is packaged in a preloaded syringe.
### 5.0 Indication for Use
i-MTA BP Bioceramic Root Canal Sealer and Repair Material is developed for permanent root canal sealer and repair and surgical applications.
- Repair of Root Perforation
- Repair of Root Resorption
- Root End Filling
- Apexification
- . Pulp Capping
#### 6.0 Comparison of Product Properties with the Predicate Device
The subject device Bioceramic Root Canal Sealer and Repair Materials(i-MTA BP) met all design specifications as was Substantially Equivalent (SE) to the predicated device.
{6}------------------------------------------------
510(k) Summary
| Comparison<br>Items | Subject Device | Predicate Device | Reference Device | Comparison |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|---------------------------------|
| K Number | K242934 | K240365 | K092715 | -- |
| Trade name | Bioceramic Root Canal Sealer and Repair<br>Materials | Bioceramic Root Repair Material | IROOT BP PLUS | -- |
| Model | i-MTA BP | C-Root BP | iRoot BP Plus | -- |
| Classification<br>name | Resin, Root Canal Filling | Resin, Root Canal Filling | Resin, Root Canal Filling | Same |
| Product code | KIF | KIF | KIF | Same |
| Class | II | II | II | Same |
| Intended<br>use/Indications<br>for Use | • Root End Filling<br>• Repair of Root Perforation<br>• Repair of Root Resorption<br>• Apexification<br>• Pulp Capping | • Repair of Root Perforation<br>• Repair of Root Resorption<br>• Root End Filling<br>• Apexification<br>• Pulp Capping | • Repair of Root Perforation<br>• Repair of Root Resorption<br>• Root End Filling<br>• Apexification<br>• Pulp Capping | Same |
| Targeted Site | Root canal | Root canal | Root canal | Same |
| Prescription<br>Use | Yes | Yes | Yes | Same |
| Chemical<br>Composition | Calcium silicate, zirconium oxide, tantalum<br>pentoxide, calcium sulfate and filler agents. | Zirconium oxide, strontium<br>silicate, calcium phosphates, calcium<br>hydroxide, tantalum oxide and filler<br>agents. | Tricalcium silicate, dicalcium silicate,<br>zirconium oxide, tantalum pentoxide,<br>calcium sulfate and filler agents. | Analysis 1 |
| Performance<br>Standard<br>Conformance | Standards or other<br>guidelines/ referential<br>used | Standards or other<br>guidelines/ referential<br>used | Standards or other<br>guidelines/ referential<br>used | Same as<br>predicated<br>device |
| | ISO 6876:2012 | ISO 6876:2012 | ISO 6876:2012 | |
| | ISO 6876:2012 | ISO 6876:2012 | ISO 6876:2012 | |
| | ISO 6876:2012 | ISO 6876:2012 | ISO 6876:2012 | |
| | Technical Specifications | Technical Requirements | Technical Requirements | |
| | Setting time | Setting time | Setting time | |
| | Solubility | Solubility | Solubility | |
| | Radio-opacity | Radio-opacity | Radio-opacity | |
| | Composition | Appearance | Appearance | |
| Comparison<br>Items | Subject Device | Predicate Device | Reference Device | Comparison |
| | Technical Specifications | | | |
| | Technical<br>Specifications=ISO6876:2<br>001 | Technical Requirements | Technical Requirements | |
| | Technical Specifications | | | |
| | Technical Specifications | Technical Requirements | Technical Requirements | |
| | Technical Specifications | | | |
| | | Dimension<br>changes after<br>setting | Dimension<br>changes after<br>setting | |
| | Content | | | |
| | Size change after<br>curing | | | |
| | pH | pH | pH | |
| | Acid-soluble<br>arsenic and lead<br>contents | | | |
| | Heavy metal | | | |
| Duration of<br>Contact | Permanent contact - contact exceeds 30d | Permanent contact - contact exceeds 30d | Permanent contact - contact exceeds 30d | Same |
| Biocompatibility | Comply with ISO 10993-1, ISO 7405 and<br>FDA Guidance | Comply with ISO 10993-1, ISO 7405 and<br>FDA Guidance | Comply with ISO 10993-1, ISO 7405 and<br>FDA Guidance | Same |
{7}------------------------------------------------
#### 510(k) Summary
{8}------------------------------------------------
# Analysis
- 1. The subject device Bioceramic Root Canal Sealer and Repair Materials(i-MTA BP) has the same intended use and comparable materials as the predicated device C-Root BP (K240365).
- 2. The chemical composition of the subject device Bioceramic Root Canal Sealer and Repair Materials is different with the predicated device C-Root BP (K240365).
The subject device uses calcium silicates whereas the predicated device uses strontium silicate. Calcium silicates and strontium silicates are all belong to silicate. The subject device uses calcium sulfate whereas the predicated device uses calcium phosphates and calcium hydroxide. They are all belong to calcium salt. They all can be set and harden upon the moisture in the dentin. They all play the same role in each's formula. Also, Calcium silicates have better biocompatibility. The subject device Bioceramic Root Canal Sealer and Repair Materials(i-MTA BP) has the same formulation/ processing/ non-sterilization/ principal of setting/ final form after setting as the
reference device iRoot BP Plus (K092715), and no other chemical composition has been added. All the raw materials of the subject device are all used in the reference device (K092715). Biocompatibility testing of the subject device and the reference device (K092715) performed meet ISO 10993 biocompatibility and FDA guidance requirements.
- 3. Also, the performance comparison testing including composition/appearance, setting time, size changes after curing, solubility, radiopacity, pH of 'Bioceramic Root Canal Sealer and Repair Materials' and C-Root BP provide the evidence that chemical and physical properties are substantially equivalent.
# 7.0 Summary of Testing
# Performance Testing - Bench
Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intended use. The following tests were performed on the device.
| Standards or other<br>guidelines/ referential<br>used | Performances | Requirements | Compliance<br>result |
|-------------------------------------------------------|----------------------------------------------|---------------------------------------------------------------------------------------------------|--------------------------------|
| ISO 6876:2012 | Setting time | 2h-6h | 2.5h |
| ISO 6876:2012 | Solubility | < 3% | 2.80% |
| ISO 6876:2012 | Radio-opacity | > 3mm Al | > 3mm Al |
| Technical Specifications | Composition | The product composition should<br>not contain any visible foreign<br>objects to the naked eye. | Compliant with<br>requirements |
| Technical Specifications | Content | Not less than 93% of the indicated<br>loading capacity. | Compliant with<br>requirements |
| Technical<br>Specifications=ISO6876<br>:2001 | Size change<br>after curing | < 1.0% | -0.50% |
| Technical Specifications | pH | 11.5-13.5 | 12.29 |
| Technical Specifications | Acid-soluble<br>arsenic and<br>lead contents | The maximum content of acid-<br>soluble arsenic is 2mg/kg, the<br>maximum content of acid-soluble | Compliant with<br>requirements |
{9}------------------------------------------------
| | | lead is 100mg/kg | |
|--------------------------|-------------|------------------|------------|
| Technical Specifications | Heavy metal | < 100 µg/g | < 100 µg/g |
Clinical testing was not required for this submission.
## Biocompatibility Testing
Biocompatibility testing was performed in accordance with the ISO 10993-1, ISO 7405 and FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management process." Testing included the following:
- Cytotoxicity (ISO 7405:2018)
- · Sensitization (ISO 10993-10:2021)
- Irritation (ISO 10993-10:2021)
- · Acute Systemic Toxicity (ISO 10993-11: 2017)
- · TK Gene Mutation Test (ISO 10993-3:2014)
- · Bacterial Reverse Mutation Test (ISO 10993-3:2014)
- · Subchronic Systemic Toxicity (ISO 10993-11: 2017)
- Implantation (ISO10993-6:2016)
- · Pulp Capping Testing (ISO7405:2018)
- Endodontic Usage Test (ISO 7405:2018)
All the results of biocompatibility testing demonstrate that the subject devices are complied with the biocompatibility requirements.
### Summary
Bioceramic Root Canal Sealer and Repair Materials has the same Indications for Use and the principle of operations as the predicate device and reference device. It is intended purpose as they are placed into the root canal as a root filling material which met the requirement according to ISO 6876. It has similar physical and biocompatible properties and demonstrates comparable performance specifications to the predicate devices.
# 8.0 Conclusion
The conclusions drawn from the comparison and analysis above demonstrate that the differences between the subject device and the predicated device C-Root BP are insignificant in terms of substantial equivalence. The subject device is substantially equivalent to the predicate device C-Root BP.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.