K242926 · Xbody Hungary Kft. · NGX · May 30, 2025 · Physical Medicine
Device Facts
Record ID
K242926
Device Name
XBody Go USA, XBody Pro USA
Applicant
Xbody Hungary Kft.
Product Code
NGX · Physical Medicine
Decision Date
May 30, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Software as a Medical Device, Therapeutic
Indications for Use
The XBODY Go USA and XBODY Pro USA are machines with electronic muscle stimulation based on EMS technology. The devices are specifically designed as an addition to other sports and for training muscles. They must be used only for healthy muscles and clients, not for rehabilitation purposes. Both models are intended to stimulate healthy muscles in order to improve or facilitate muscle performance. Neither model is intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. None of the XBODY Go USA or XBODY Pro USA training programs are designed for injured or ailing muscles and their use on such muscles is contraindicated. For both models, electrical impulses allow the triggering of action potentials on motoneurons of motor nerves (excitations). These excitations of motoneurons are transmitted to the muscle fibers via the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, total session duration), different types of muscle work can be imposed on the stimulated muscles. The various types of muscle work that the XBODY Go USA and XBODY Pro USA can impose on the stimulated muscles are able to improve or facilitate muscle performance. Both models may therefore be considered a technique of muscle training.
Device Story
Device uses electrical muscle stimulation (EMS) to trigger action potentials on motoneurons, generating mechanical muscle fiber responses. Input parameters (pulse frequency, contraction/rest duration, session duration) are controlled by a registered trainer via a touchscreen tablet (Microsoft Surface series) running specialized software. Output is delivered via wireless stimulation unit (Actiwear G2) to electrodes integrated into training suits (TS 2.1, TS 3.0, or DrySuit). Used in home healthcare environments; requires trained operator. Output affects muscle performance through imposed muscle work. Benefits include muscle training and performance improvement. System includes safety circuits for overload, no-load, and battery monitoring.
Clinical Evidence
Bench testing only. Validation included electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), usability (IEC 62366-1), and performance testing of the new DrySuit compared to predicate training suits. Biocompatibility testing (ISO 10993) performed on applied parts.
Technological Characteristics
Powered muscle stimulator; battery-operated (Li-ion 4x 3.7V). Symmetric biphasic rectangular waveform; frequency 1-150Hz; pulse width 50-500usec. Max output 20.8V/41.6mA at 500Ω. Wireless connectivity (BT/BLE) between tablet and Actiwear G2 unit. Materials: Plastic housing, various textiles (polyamide, polyester, elastane, acrylic). Safety: Overload/no-load trip, watchdog monitoring. Software: IEC 62304 compliant.
Indications for Use
Indicated for healthy individuals to stimulate muscles to improve or facilitate muscle performance as an addition to sports and training. Contraindicated for use on injured or ailing muscles and for any medical therapy or treatment of diseases or conditions.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 30, 2025
XBody Hungary Kft.
% Bhoomika Joyappa
Senior Regulatory Consultant
Medical Device Academy, Inc.
345 Lincoln Hill Road
Shrewsbury, Vermont 05738
Re: K242926
Trade/Device Name: XBody Go USA, XBody Pro USA
Regulation Number: 21 CFR 890.5850
Regulation Name: Powered Muscle Stimulator
Regulatory Class: Class II
Product Code: NGX
Dated: September 20, 2024
Received: September 24, 2024
Dear Bhoomika Joyappa:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}
K242926 - Bhoomika Joyappa
Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K242926 - Bhoomika Joyappa
Page 3
Sincerely,
Tushar Bansal -S
for Heather Dean, PhD
Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K242926
Device Name
XBody Go USA, XBody Pro USA
Indications for Use (Describe)
The XBODY Go USA and XBODY Pro USA are machines with electronic muscle stimulation based on EMS technology. The devices are specifically designed as an addition to other sports and for training muscles. They must be used only for healthy muscles and clients, not for rehabilitation purposes.
Both models are intended to stimulate healthy muscles in order to improve or facilitate muscle performance. Neither model is intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. None of the XBODY Go USA or XBODY Pro USA training programs are designed for injured or ailing muscles and their use on such muscles is contraindicated.
For both models, electrical impulses allow the triggering of action potentials on motoneurons of motor nerves (excitations). These excitations of motoneurons are transmitted to the muscle fibers via the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, total session duration), different types of muscle work can be imposed on the stimulated muscles.
The various types of muscle work that the XBODY Go USA and XBODY Pro USA can impose on the stimulated muscles are able to improve or facilitate muscle performance. Both models may therefore be considered a technique of muscle training.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”
{4}
XBody Hungary Kft.
XBody Go USA, XBody Pro USA 510(k) Submission
Premarket Notification [510(k) Summary]
K242926
A. General Information
Sponsor's Name: XBody Hungary Kft.
Sponsor's Contact Person: Orsolya Balog (Regulatory Compliance)
Address: György Aladár u. 35-39 Budapest, Hungary 1125
orsolya.balog@XBodyworld.com
www.XBodyworld.com
Submission Contact Person: Bhoomika Joyappa
Senior Regulatory Consultant
Address: Medical Device Academy, Inc
345 Lincoln Hill Rd,
Shrewsbury, VT 05738
bhoomika@fdaestar.com
Date Prepared: 5/28/2025
B. Device
Trade Names of Companion Products: XBody Go USA, XBody Pro USA
Common Name: Powered Muscle Stimulator
Product Classification Code: NGX
Regulatory Class: 2
Device Classification Name: Stimulator, Muscle, Powered, For Muscle Conditioning
Regulation Name: Powered Muscle Stimulator
Regulation Number: 21 CFR 890.5850
C. Identification of Legally Marketed Predicate Device
Predicate Device: XBody Go USA, XBody Pro USA
Manufacturer: XBody Hungary Kft
Predicate 510(k) Number: K221200
{5}
XBody Hungary Kft.
XBody Go USA, XBody Pro USA 510(k) Submission
## D. Description of the Devices
The XBody Go USA and the XBody Pro USA generate electronic muscle stimulation based on electrical muscle stimulation (EMS) technology. The devices are designed as additions to other sports and for training muscles. They are not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. They are intended for use only by persons with healthy muscles, not for rehabilitation purposes.
## E. Indication for Use
The XBody Go USA and XBody Pro USA are machines with electronic muscle stimulation based on EMS technology. The devices are specifically designed as an addition to other sports and for training muscles. They must be used for only healthy muscles and clients, not for rehabilitation purposes.
Both models are intended to stimulate healthy muscles in order to improve or facilitate muscle performance. Neither model is intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. None of the XBody Go USA or XBody Pro USA training programs are designed for injured or ailing muscles and its use on such muscles is contraindicated.
For both models, electrical impulses allow the triggering of action potentials on motoneurons of motor nerves (excitations). These excitations of motoneurons are transmitted to the muscle fibers via the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, total session duration), different types of muscle work can be imposed on the stimulated muscles.
The various types of muscle work that the XBody Go USA and XBody Pro USA can impose on the stimulated muscles are able to improve or facilitate muscle performance. Both models may therefore be considered a technique of muscle training.
## F. Technological Characteristics
Compared to the predicate devices, the XBody Go USA and XBody Pro USA are the same in indication for use, intended use performance, design, dimensions, and materials as the predicate devices. The new devices meet the same standards for safety and performance as the predicate devices.
The differences cited in the comparison tables between subject and predicate devices do not affect safety and performance of the subject devices when compared for equivalence to the predicate devices.
| Parameter / application | Subject Device
XBody Go USA
and XBody Pro USA | Predicate Device
XBody Go USA
and XBody Pro USA
(K221200) | Assessment of substantial equivalence between Subject and Predicate Devices |
| --- | --- | --- | --- |
| 510k Number | NA | K221200 | NA |
{6}
XBody Hungary Kft.
XBody Go USA , XBody Pro USA 510(k) Submission
| Submitter | XBody Hungary Kft | XBody Hungary Kft | Same |
| --- | --- | --- | --- |
| Common or Usual Name | Powered muscle stimulator. | Powered muscle stimulator. | Same |
| Product code | NGX | NGX | Same |
| Classification Name | Stimulator, Muscle, Powered, For Muscle Conditioning | Stimulator, Muscle, Powered, For Muscle Conditioning | Same |
| Product Classification | Class II | Class II | Same |
| Review Panel | Physical Medicine | Physical Medicine | Same |
| Regulated voltage | Yes | Yes | Same |
| Regulation Name | 21 CFR 890.5850 | 21 CFR 890.5850 | Same |
| Type of use | Prescription Use | Prescription Use | Same |
| Indications for use | The XBody Go USA and XBody Pro USA are machines with electronic muscle stimulation based on EMS technology. The devices are specifically designed as an addition to other sports and for training muscles. They must be used for only healthy muscles and people (clients), not for rehabilitation purposes.
The XBody Go USA and XBody Pro USA are intended to stimulate healthy muscles in order to improve or facilitate muscle performance The XBody Go USA and XBody Pro USA are not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. None of the XBody Go USA or XBody Pro USA training programs are designed for injured or ailing muscles and its use on such muscles is contraindicated. | The XBody Go USA and XBody Pro USA are machines with electronic muscle stimulation based on EMS technology. The devices are specifically designed as an addition to other sports and for training muscles. They must be used for only healthy muscles and people (clients), not for rehabilitation purposes.
The XBody Go USA and XBody Pro USA are intended to stimulate healthy muscles in order to improve or facilitate muscle performance The XBody Go USA and XBody Pro USA are not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. None of the XBody Go USA or XBody Pro USA training programs are designed for injured or ailing muscles and its use on such muscles is contraindicated. | Same |
3
{7}
XBody Hungary Kft.
XBody Go USA , XBody Pro USA 510(k) Submission
| | The XBody electrical impulses allow the triggering of action potentials on motoneurons of motor nerves (excitations). These excitations of motoneurons are transmitted to the muscle fibers via the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, total session duration), different types of muscle work can be imposed on the stimulated muscles.
The various types of muscle work that the XBody Go USA or XBody Pro USA can impose on the stimulated muscles are able to improve or facilitate muscle performance. The XBody Go USA and the XBody Pro USA may therefore be considered as a technique of muscle training. | The XBody electrical impulses allow the triggering of action potentials on motoneurons of motor nerves (excitations). These excitations of motoneurons are transmitted to the muscle fibers via the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, total session duration), different types of muscle work can be imposed on the stimulated muscles.
The various types of muscle work that the XBody Go USA or XBody Pro USA can impose on the stimulated muscles are able to improve or facilitate muscle performance. The XBody Go USA and the XBody Pro USA may therefore be considered as a technique of muscle training. | |
| --- | --- | --- | --- |
| Portability/ Mobile Use | The device is portable with ease. It is not a mobile device; its intended environment is indoors. Its intended use requires a qualified and trained operator. | The device is portable with ease. It is not a mobile device; its intended environment is indoors. Its intended use requires a qualified and trained operator. | Same |
| User interface | The device can be controlled using the graphical windows appearing on the touchscreen of the external control unit (Microsoft Surface Go 2, Microsoft Surface Go 3 or Microsoft Surface Go 4 for XBody Go USA / Microsoft Surface Pro 7, Microsoft Surface Pro 9 or Microsoft Surface Pro 10 for XBody Pro USA). On the training screen where stimulation controls can be used the START/ STOP buttons are large and easily controllable. Stimulation controls for adjusting channel intensities, and all other stimulation parameters are clearly visible and easily controllable. Channel identification is supported with big pictures showing the selected muscle groups. When the stimulation is on, the STOP button is always visible and accessible. | The device can be controlled using the graphical windows appearing on the touchscreen of the external control unit (Microsoft Surface Go 2 for XBody Go USA / Microsoft Surface Pro 7 for XBody Pro USA). On the training screen where stimulation controls can be used the START/ STOP buttons are large and easily controllable. Stimulation controls for adjusting channel intensities, and all other stimulation parameters are clearly visible and easily controllable. Channel identification is supported with big pictures showing the selected muscle groups. When the stimulation is on, the STOP button is always visible and accessible. | Similar. Subject devices have more options for control unit devices. The control units are equipped with the same specialized software. |
{8}
XBody Hungary Kft.
XBody Go USA , XBody Pro USA 510(k) Submission
| Menu / Settings | Easy-to-use multi-choice menu for registered and certified trainers to customize training parameters and stimulation Programs. | Easy-to-use multi-choice menu for registered and certified trainers to customize training parameters and stimulation Programs. | Same |
| --- | --- | --- | --- |
| Operator | To operate the devices the trainer must complete an XBody US EMS Trainer Course. The certification data received at the end of the course is required when XBody registers trainers in the device database. Only registered trainers can start training stimulation Programs using a passcode. | To operate the devices the trainer must complete an XBody US EMS Trainer Course. The certification data received at the end of the course is required when XBody registers trainers in the device database. Only registered trainers can start training stimulation Programs using a passcode. | Same |
| Display | GO: <10.5” LCD display on the XBody Go USA tablet>
PRO: <12.3” or 13” LCD display on the XBody Pro USA tablet > | GO: <10.5” LCD display on the XBody Go USA tablet>
PRO: <12.3” LCD display on the XBody Pro USA tablet > | Similar. The XBody Pro USA subject device has optional control units equipped with slightly bigger displays. |
| Statistical Functions | Training data (trainer, client, date, duration). Client related data. Number of training sessions (today, yesterday, this week, this month, total). | Training data (trainer, client, date, duration). Client related data. Number of training sessions (today, yesterday, this week, this month, total). | Same |
| Output specifications | Max Output Voltage = 20.8V @500Ω | Max Output Voltage = 20.8V @500Ω | Same |
| | Max Output Current = 41.6mA @500Ω | Max Output Current = 41.6mA @500Ω | Same |
| | Max Phase Charge = 16.64μC@500Ω | Max Phase Charge = 16.64μC@500Ω | Same |
| | Max Current Density =0.65mA/cm2 @500Ω | Max Current Density =0.65mA/cm2 @500Ω | Same |
| | Max Power Density =3.46mW/cm2 @500Ω | Max Power Density =3.46mW/cm2 @500Ω | Same |
| Net Charge (μC per pulse) | 0 @500Ω (each phase uses symmetric waveform) | 0 @500Ω (each phase uses symmetric waveform) | Same |
| Shape | Rectangular | Rectangular | Same |
{9}
XBody Hungary Kft.
XBody Go USA , XBody Pro USA 510(k) Submission
| Number of Output channels | 6 output channels, but maximum 12 independently regulated outputs in case of Training Suit (TS) 2.1 and TS 3.0. 5 output channels, but maximum 10 independently regulated outputs in case of DrySuit | 6 output channels, but 12 independently regulated outputs | Similar The subject devices have the option to use the DrySuit which has only 10 channels. |
| --- | --- | --- | --- |
| Waveform | Symmetric biphasic | Symmetric biphasic | Same |
| Burst mode - Pulses per burst | 100 * 10 = 1000 | 100 * 10 = 1000 | Same |
| Burst mode - Bursts per second | 0.1 | 0.1 | Same |
| Burst mode - Burst duration (seconds) | 11 | 11 | Same |
| Burst mode - ON Time (seconds) | 10 | 10 | Same |
| Burst mode - OFF Time (seconds) | 1 | 1 | Same |
| Burst mode - Duty Cycle: Line (d)/(Line (d)+ Line (e))* | 0.91 | 0.91 | Same |
| Additional Features (specify, if applicable) | N/A | N/A | Same |
| Output frequency | 1-150Hz | 1-150Hz | Same |
| Positive pulse width | 50-500usec | 50-500usec | Same |
| Negative pulse width | 50-500usec | 50-500usec | Same |
| Power source – Battery | Li-ion 4x 3.7V (3.4 Ah) | Li-ion 4x 3.7V (3.4 Ah) | Same |
6
{10}
XBody Hungary Kft.
XBody Go USA , XBody Pro USA 510(k) Submission
| Size of the electrodes | Predefined electrode sizes inside the training suit described in User Manual | Predefined electrode sizes inside the training suit described in User Manual | Same |
| --- | --- | --- | --- |
| Safety circuits | Overload trip detects short-circuit, No-load trip detects circuit break, battery voltage monitoring, hardware error detection at startup, and watchdog monitoring. | Overload trip detects short-circuit, No-load trip detects circuit break, battery voltage monitoring, hardware error detection at startup, and watchdog monitoring. | Same |
| Plugs | The XBody Go USA and XBody Pro USA tablets and the XBody Actiwear G2 wireless stimulation unit are connected wirelessly. The XBody Actiwear G2 and the XBody Training Suit are connected with magnetic connectors. The internal cable of the training suit connects to snap fasteners in the suit to which detachable electrodes are attached via waterproof connections. | The XBody Go USA and XBody Pro USA tablets and the XBody Actiwear G2 wireless stimulation unit are connected wirelessly. The XBody Actiwear G2 and the XBody Training Suit are connected with magnetic connectors. The internal cable of the training suit connects to snap fasteners in the suit to which detachable electrodes are attached via waterproof connections. | Same |
| Lead Wires – Cables | TS2.1: PVC coated ultra-flexible LIFY 0,50 mm2 (256 x 0,05 mm) cables and LiYV 0,56 mm2 (7 x 0,32 mm) in the training suit. Cables are compliant with protected lead wire and patient cable safety requirements
TS3.0: Suit cable: Ultra-flexible Microminiature & Miniature PVC Insulated Lead Wire Textile material (yarn): Polyester
DS: PFA coated stainless steel filaments cored yarn (1.0mm~1.05mm) 275F*6, 316L | TS2.1: PVC coated ultra-flexible LIFY 0,50 mm2 (256 x 0,05 mm) cables and LiYV 0,56 mm2 (7 x 0,32 mm) in the training suit. Cables are compliant with protected lead wire and patient cable safety requirements
TS3.0: Suit cable: Ultra-flexible Microminiature & Miniature PVC Insulated Lead Wire Textile material (yarn): Polyester | Similar
The subject devices have the option to use the DrySuit with its own lead wire |
7
{11}
XBody Hungary Kft.
XBody Go USA , XBody Pro USA 510(k) Submission
| Conductivity of the Electrodes | When using the TS 2.1 or TS 3.0, the client has two options: wearing an XBody cotton underwear (biocompatibility certified), or wearing the DryWear with lightweight thin undergarments that do not obstruct the connection between the electrodes and the skin on the glutes. When using the cotton underwear, the electrodes are contained in cotton covers which must be watered using normal tap water to create conductive media. The cotton textiles hold enough water to provide conductivity during the training. The electrodes are washable and can be disinfected, as described in User Manual.
When using the XBody DryWear with undergarments, the electrodes of the TS2.1 or TS3.0 are fixed in the training suit with Velcro attachments, instead of cotton covers. Conductive media in this case is the DryWear. The XBody DrySuit must be worn with lightweight thin undergarments that do not obstruct the connection between the electrodes and the skin on the glutes. The XBody DrySuit transmits electrical stimulation to the surface of the skin without the use of water. | The client must wear an XBody cotton underwear (biocompatibility certified). The electrodes are contained in cotton covers which must be watered using normal tap water to create conductive media. The cotton textiles hold enough water to Provide conductivity during the training. The electrodes are washable and can be disinfected, as described in User Manual. | Similar |
| --- | --- | --- | --- |
| Soldering of the Printed Circuit Boards | In accordance with the ROHS directive, no lead solder material used. | In accordance with the ROHS directive, no lead solder material used. | Same |
| Placement of the electrodes | The electrodes are located at fixed positions in the training suit ensuring proper placement. | The electrodes are located at fixed positions in the training suit ensuring proper placement. | Same |
| Reusable pads | Yes | Yes | Same |
| Number of programs | XBody Go USA: 5
XBody Pro USA: 6 | XBody Go USA: 5
XBody Pro USA: 6 | Same |
| Treatment duration | 1 min to 60 min maximum | 1 min to 60 min maximum | Same |
| Environment(s) of use | Home healthcare environment, according to IEC 60601-1-11 | Home healthcare environment, according to IEC 60601-1-11 | Same |
| Pulse duration | 1-10 s | 1-10 s | Same |
8
{12}
XBody Hungary Kft.
XBody Go USA , XBody Pro USA 510(k) Submission
| **Compliance standards** | IEC 60529:1989+ A2:2013+C1:2019
IEC 60601-1:2005+ A1:2012+A2:2020
IEC 60601-1-2:2014 +A1:2020
IEC 60601-1-6:2010 +A1:2013+A2:2020 | IEC 60529:1989+ A2:2013+C1:2019
IEC 60601-1:2005+ A1:2012+A2:2020
IEC 60601-1-2:2014 +A1:2020
IEC 60601-1-6:2010 +A1:2013+A2:2020 | Same |
| --- | --- | --- | --- |
9
{13}
XBody Hungary Kft.
XBody Go USA , XBody Pro USA 510(k) Submission
| | IEC 60601-1-11:2015 +A1:2020
IEC 60601-2-10:2012 +A1:2016
IEC 62304:2006 +A1:2015
IEC 62366-1:2015 +A1:2020 | IEC 60601-1-11:2015 +A1:2020
IEC 60601-2-10:2012 +A1:2016
IEC 62304:2006 +A1:2015
IEC 62366-1:2015 +A1:2020 | |
| --- | --- | --- | --- |
| Compliance with 21 CFR 898? | YES | YES | Same |
| Automatic No-Load Trip | YES | YES | Same |
| Method of Line Current Isolation | N/A (battery-operated device) | N/A (battery-operated device) | Same |
| Patient leakage current – normal condition (μA) | N/A (battery-operated device) | N/A (battery-operated device) | Same |
| Patient leakage current – single fault condition (μA) | N/A (battery-operated device) | N/A (battery-operated device) | Same |
| Synchronous or alternating? | N/A (battery-operated device) | N/A (battery-operated device) | Same |
| Method of Channel Isolation | Hardware and timing | Hardware and timing | Same |
| Regulated Current or Regulated Voltage? | Regulated current and voltage | Regulated current and voltage | Same |
| Software/Firmware/Microprocessor Control? | Yes | Yes | Same |
| Automatic Over current Trip? Yes/no | Yes | Yes | Same |
| Patient Override Control? Yes/no | Yes | Yes | Same |
| Patient Override Control method | Disconnect suit connectors | Disconnect suit connectors | Same |
| Indicator Display | Yes | Yes | Same |
| Number of Output Modes | One output mode, but with varying stimulation frequency and duty cycle ranges | One output mode, but with varying stimulation frequency and duty cycle ranges | Same |
| Total Dimensions (in.) [W x H x D] | 9.65 X 6.9 X 0.33 (tablet)
1.96 X 1.96 X 0.98 (Actiwear G2) | 9.65 X 6.9 X 0.33 (tablet)
1.96 X 1.96 X 0.98 (Actiwear G2) | Same |
| Housing Materials and Construction | Plastic | Plastic | Same |
10
{14}
XBody Hungary Kft.
XBody Go USA, XBody Pro USA 510(k) Submission
| Table 2 Comparison of Predicate and Subject Training Suits: XBody Training Suit 2.1, XBody Training Suit 3.0 and XBody DrySuit | | | |
| --- | --- | --- | --- |
| Training suits | XBody Training Suit 2.1 & Training Suit 3.0 Predicate training suits for XBody Go USA and XBody Pro USA (K221200) | XBody DrySuit | Assessment |
| Compatible product | XBody Go USA XBody Pro USA | XBody Go USA XBody Pro USA | Same |
| Parts | Vest, pants, optional bands | Jacket, pants | Same |
| Material | TS 2.1: 75% Polyamide, 25% Polyester (Same as TS 2.0) TS 3.0: 40% Polyamide, 30% Acrylic, 25% Polyester, 5% Elastane | 75% Polyester, 15% Spandex, 15% Elastane | Similar |
| Underwear required during use | Yes. Option 1: same underwear for XBody Newave USA (K190038). Underwear serves as barrier to direct contact with the body during use Option 2: XBody DryWear worn over lightweight thin undergarments that do not obstruct the connection between the electrodes and the skin on the glutes. XBody DryWear is new to this submission. | Yes. The XBody DrySuit must be worn over lightweight thin undergarments that do not obstruct the connection between the electrodes and the skin on the glutes. The XBody DrySuit contacts the skin of the client directly. | Different |
| Number of channels | 12 | 10 | Similar, but less available channels |
| Number of electrodes | 24 pieces per suit | 20 pieces per suit | Similar, but less available electrodes |
| Weight with electrodes and cables | TS2.1: 2.7 – 3.3 kg (6 – 7.3 lbs.) TS3.0: 1.8 - 2.4 kg (4 – 5.3 lbs.) | 1 kg (2.2 lbs.) | Similar |
| Available Sizes | TS2.1: XXXS, XS, S, M, L, XL, XXL TS3.0: 2, 3, 3-L, 4, 4-L, 5, 5-W, 6, 7, 8 | XXS/2, XS/3, S/4, M/5, L/6, XL/7, XXL/8, XXXL/9 | Similar |
{15}
XBody Hungary Kft.
XBody Go USA , XBody Pro USA 510(k) Submission
| Available Electrode Sizes and Shapes | TS2.1: 8 different electrode shapes
TS3.0: 10 different electrode shapes | 9 different electrodes | Similar |
| --- | --- | --- | --- |
Table 3 Similarities of XBody Go USA and XBody Pro USA and Comparison to Predicate Control Unit Parameters
{16}
XBody Hungary Kft.
XBody Go USA , XBody Pro USA 510(k) Submission
| Parameter | XBody Go USA (Subject Device) | XBody Pro USA (Subject Device) | Comparison of Subject Devices | Subject Devices vs. Predicate Devices |
| --- | --- | --- | --- | --- |
| Control Unit | Microsoft Surface | Microsoft Surface | Same | Same |
| Dimensions of Control Unit | 9.65” x 6.9” x 0.33” (245 mm x 175 mm x 8.3 mm) | 11.5” x 7.9” x 0.33” (292 mm x 201 mm x 8.5 mm) | Similar | Same |
| Wireless Stimulation Unit | XBody Actiwear G2 | XBody Actiwear G2 | Same | Same |
| Training Suit | Able to use XBody Training Suit 2.1, XBody Training Suit 3.0 and XBody DrySuit | Able to use XBody Training Suit 2.1, XBody Training Suit 3.0 and XBody DrySuit | Same | Similar. Predicate can only use XBody Training Suit 2.1 and XBody Training Suit 3.0 |
| Requirement for Underwear | XBody Cotton underwear or XBody DryWear worn over undergarments must be worn during use with XBody Training Suit 2.1 and XBody Training Suit 3.0. Undergarments must be worn with the XBody DrySuit. | XBody Cotton underwear or XBody DryWear worn over undergarments must be worn during use with XBody Training Suit 2.1 and XBody Training Suit 3.0. Undergarments must be worn with the XBody DrySuit. | Same | Similar. Predicate can only use XBody Training Suit 2.1 and XBody Training Suit 3.0 |
| Training Programs | Manual settings, Muscle Development, Endurance, Relax, Professional training programs | Manual settings, Muscle Development, Endurance, Relax, Professional, XBeat training programs | Similar | Similar |
| Professional Training Features | In person, Virtual trainer, Video editor | In person, Virtual trainer, Video editor | Same | Same |
| User Manual | Yes, content varies based on device design | Yes, content varies based on device design | Similar | Similar |
| Maximum number of Simultaneous Clients | 1-2 persons | 1-6 persons | Similar | Similar |
| Control Unit Storage | Tablet: 64 GB | Tablet: Solid-state drive: 128GB | Similar | Similar |
| Control Unit Display | 10.5-inch Tablet Screen | 12.3-inch Tablet Screen | Similar | Similar |
13
{17}
XBody Hungary Kft.
XBody Go USA , XBody Pro USA 510(k) Submission
| Control Unit Battery Life | Up to 9 hours of typical device usage | Up to 15 hours of typical device usage | Similar | Similar |
| --- | --- | --- | --- | --- |
14
{18}
XBody Hungary Kft.
XBody Go USA, XBody Pro USA 510(k) Submission
The following is a comparison of the similarities of the subject and predicate devices as listed in Table 3.
Control unit: For the predicate devices, the control devices were limited to the Microsoft Surface Go 2 for the XBody Go USA and the Microsoft Surface Pro 7 for the XBody Pro USA. For the subject devices, the available control devices include the Microsoft Surface Go 2, Microsoft Surface Go 3, and Microsoft Surface Go 4 for the XBody Go USA, and the Microsoft Surface Pro 7, Microsoft Surface Pro 9, and Microsoft Surface Pro 10 for the XBody Pro USA. This change does not impact safety or performance, as the newly introduced control units operate the same software developed by XBody Hungary Kft., have similar specifications, and exhibit identical performance.
Number of clients: Using the predicate or the subject device versions of the XBody Go USA, the same number of clients can train at a time. Using the predicate or the subject device versions of the XBody Pro USA, the same number of clients can train at a time.
Assessment: The subject devices and the predicate devices are the same in terms of intended use, performance, and design. Similarities derive from the introduction of the XBody DrySuit and DryWear. The new device meets the same standards for safety as the predicate device.
| Specification | Microsoft Surface Go 2 | Microsoft Surface Go 3 | Microsoft Surface Go 4 |
| --- | --- | --- | --- |
| Execution platform | Intel® Pentium® Gold Processor 4425Y 8th Gen | Intel® Pentium® Gold Processor 6500Y 10th Gen | Intel N200 |
| CPU Cores | 2 | 2 | 4 |
| RAM | 4 GB | 4 GB | 8 GB |
| Graphics | Intel® UHD Graphics 615 | Intel® UHD Graphics 615 | Intel UHD Graphics |
| Display | 10.5”, 1920 x 1280, 10-point multi-touch | 10.5”, 1920 x 1280, 10-point multi-touch | 10.5”, 1920 x 1280, 10-point multi-touch |
| Storage | eMMC drive: 64GB | eMMC drive: 64GB | UFS Drive: 64GB, 128GB, 256GB |
| Connections | 1 USB Type-C, HDMI over | 1 USB Type-C, HDMI over | 1 USB Type-C, HDMI over |
{19}
XBody Hungary Kft.
XBody Go USA, XBody Pro USA 510(k) Submission
| | docking adapter or Type-C | docking adapter or Type-C | docking adapter or Type-C |
| --- | --- | --- | --- |
| Wireless communication | Wi-Fi 6, BT/BLE 5 | Wi-Fi 6, BT/BLE 5 | Wi-Fi 6, BT/BLE 5.1 |
| Security | TPM 2.0, UEFI secure boot | TPM 2.0, UEFI secure boot | TPM 2.0, UEFI secure boot |
| OS support | Linux kernel supported | Linux kernel supported | Linux kernel supported |
| Construction | Handheld tablet with integrated battery | Handheld tablet with integrated battery | Handheld tablet with integrated battery |
| Specification | Microsoft Surface Pro 7 | Microsoft Surface Pro 9 | Microsoft Surface Pro 10 |
| --- | --- | --- | --- |
| Execution platform | Intel® i5-1035G4 10th Gen | Intel® i5-1245U 12th Gen | Intel Core Ultra 5 Processor 135U Intel Core Ultra 7 Processor 165U |
| CPU Cores | 4 | 4 | 12 |
| RAM | 8 GB | 8 GB | 8GB, 16GB, 32GB, 64GB |
| Graphics | Intel® Iris Plus Graphics 615 | Intel® Iris XE Graphics 615 | Intel Graphics |
| Display | 12.3”, 2736 x 1824, 10-point multi-touch | 13.0”, 2880 x 1920, 10-point multi-touch | 13.0”, 2880 x 1920, 10-point multi-touch |
| Storage | SSD drive: 128GB | SSD drive: 256GB | SSD drive: 256GB, 512GB, 1TB |
| Connections | 1 USB Type-C, HDMI over docking adapter or Type-C | 1 USB Type-C, HDMI over docking adapter or Type-C | 2 USB Type-C, HDMI over docking adapter or Type-C |
| Wireless communication | Wi-Fi 6, BT/BLE 5 | Wi-Fi 6E, BT/BLE 5.1 | Wi-Fi 6E, BT/BLE 5.3 |
| Security | TPM 2.0, UEFI secure boot | TPM 2.0, UEFI secure boot | TPM 2.0, UEFI secure boot |
| OS support | Linux kernel supported | Linux kernel supported | Linux kernel supported |
| Construction | Handheld tablet with integrated battery | Handheld tablet with integrated battery | Handheld tablet with integrated battery |
| Table 6 Standard Number and Title Used to Support XBody Go USA and XBody Pro USA Safety and Performance | |
| --- | --- |
| Standard Number | Applied Standards by Title |
{20}
XBody Hungary Kft.
XBody Go USA , XBody Pro USA 510(k) Submission
| ANSI/AAMI 60601-1:2005/(R)2012 and A1:2012. | Medical electrical equipment - Part 1: General requirements for basic safety and essential performance |
| --- | --- |
| IEC 60601-1-2:2014+A1:2020 | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests |
| IEC 60601-1-6:2010+ A1:2013+A2:2020 | Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance -Collateral standard: Usability |
| IEC 60601-1-11:2015+A1:2020 | Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systemsused in the home healthcare environment |
| IEC 60601-2-10:2012+A1:2016 | Medical electrical equipment - Part 2-10: Particularrequirements for the basic safety and essential performance of nerve and muscle stimulators |
| IEC 62366-1:2015+A1:2020 | Medical devices - Part 1: Application of usability engineering to medical devices |
| IEC 62304:2006+A1:2015 | Medical device software - Software life cycle Processes |
| IEC 60529:1989+A2:2013+ C1:2019 | Degrees of Protection Provided by enclosures (IP Code) |
| EN ISO 14971:2019 | Medical devices - Application of risk management to medical devices. |
| ISO 10993-1 | Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process (Biocompatibility). |
| ISO 10993-5 | Biological evaluation of medical devices - part 5: testsfor in vitro cytotoxicity. |
| ISO 10993-10 | Biological evaluation of medical devices - part 10: tests for irritation and skin sensitization. |
| EN ISO 14971:2019 | Medical devices - Application of risk management to medical devices. |
| EN ISO 13485:2016 | Medical Devices – Quality Management Systems –requirements for regulatory purposes |
17
{21}
XBody Hungary Kft.
XBody Go USA , XBody Pro USA 510(k) Submission
| FDA Guidance | Food and Drug Administration Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices 2005 |
| --- | --- |
| FDA Guidance | Guidance for Industry, FDA Reviewers/Staff and Compliance Guidance Document for Powered Muscle Stimulator 510(k)s, June 9, 1999 |
## G. Conclusion of Substantial Equivalence
Electrical safety and electromagnetic compatibility of the XBody Go USA and XBody Pro USA devices have been validated through independent testing in accordance with international standards. This testing covered the XBody Actiwear G2 main unit, the control unit, the XBody Training Suit 2.1, the XBody Training Suit 3.0 (all four used in both the subject and predicate devices), and the XBody DrySuit, which is newly included in this submission.
The subject devices and the predicate devices have the same software programming except for minor functionalities. The software of the subject devices has been updated to include a feature that allows users to select the DrySuit among the available training suits. Independent electrical and electromagnetic compatibility testing, along with performance testing of the subject devices, confirms that these software differences are aimed at enhancing the user interface without impacting safety and effectiveness.
Performance testing of the XBody DrySuit demonstrated acceptable similarity in intended use to the XBody Training Suit 2.1 and the XBody Training Suit 3.0, both of which were previously cleared for the XBody Go USA and XBody Pro USA predicate devices.
The predicate devices' sole applied part is the cotton underwear, which was required to be worn under the XBody Training Suit 2.1 and the XBody Training Suit 3.0. The subject devices have optional applied parts, cotton underwear (same as the predicates') and the XBody DryWear, both of which can be interchangeably worn under the XBody Training Suit 2.1 and the XBody Training Suit 3.0. Additionally, the newly introduced applied parts, the XBody DrySuit and DryWear must be worn with undergarments that do not obstruct the connection between the electrodes and the skin on the glutes..
The Power source of the subject and predicate devices is the same rechargeable battery, supported by bench and usability studies.
Conclusion: XBody Go USA and XBody Pro USA are substantially equivalent to their predicate devices. Any differences that exist between the subject and predicate devices do not affect safety or performance.
END OF DOCUMENT.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.