The ZZIREN™ Orogastric Tube is indicated for use in gastric and bariatric surgical procedures for gastric aspiration and lavage, to serve as a size guide, and to test for leaks.
Device Story
ZZIREN™ Orogastric Tube is a non-sterile, single-use device for gastric/bariatric surgery. It consists of a 37-inch flexible PVC insertion tube (32, 36, or 40 Fr), a Y-extension with a suction-regulating valve, and a pinch clamp. The device connects to an external vacuum source for gastric aspiration and lavage. The integrated mechanical suction-regulating valve limits negative pressure to <175 mmHg and provides an audible signal if pressure exceeds ~160 mmHg, alerting the surgeon to adjust the external vacuum source. The tube serves as a sizing guide and allows for leak testing of staple lines. Used in the OR by surgeons/anesthesiologists, the device assists in gastric decompression and intraoperative assessment of staple line integrity, potentially reducing the risk of visceral injury or inadvertent stapling through its specific stiffness and compression characteristics.
Clinical Evidence
Bench testing only. Tests included insertion tube kink, stiffness, and compression tests, pinch clamp performance, accessory connector compatibility, and air syringe compatibility for leak testing. Dimensional analysis confirmed adherence to design specifications. No clinical data was required or provided.
Indicated for patients undergoing gastric and bariatric surgical procedures requiring gastric aspiration, lavage, size guidance, or leak testing.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 10, 2025
Tools for Surgery, LLC Arnold Leiboff President 8 Technology Drive, #100 East Setauket, New York 11733
### Re: K242901
Trade/Device Name: ZZIREN™ Orogastric Tube; ZZIREN™ SGT Orogastric Tube - 32 Fr (ZZ-SGT-32); ZZIRENTM SGT Orogastric Tube - 36 Fr (ZZ-SGT-36); ZZIREN™ SGT Orogastric Tube - 40 Fr (ZZ-SGT-40); ZZIREN™ GBT Orogastric Tube - 32 Fr (ZZ-GBT-32): ZZIREN™ GBT Orogastric Tube - 36 Fr (ZZ-GBT-36): ZZIRENTM GBT Orogastric Tube - 40 Fr (ZZ-GBT-40) Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal Tube And Accessories Regulatory Class: Class II Product Code: KNT Dated: September 23, 2024 Received: December 11, 2024
Dear Arnold Leiboff:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/2/Picture/3 description: The image shows the name "Anthony Lee -S" in a large, bold, sans-serif font. The text is black against a white background. The letters are evenly spaced and easy to read.
Anthony Lee, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
| Submission Number (if known) | |
|------------------------------|--|
|------------------------------|--|
K242901
| Device Name |
|--------------------------------------------------|
| ZZIREN™ Orogastric Tube; |
| ZZIREN™ SGT Orogastric Tube - 32 Fr (ZZ-SGT-32); |
| ZZIREN™ SGT Orogastric Tube - 36 Fr (ZZ-SGT-36); |
| ZZIREN™ SGT Orogastric Tube - 40 Fr (ZZ-SGT-40); |
| ZZIREN™ GBT Orogastric Tube - 32 Fr (ZZ-GBT-32); |
| ZZIREN™ GBT Orogastric Tube - 36 Fr (ZZ-GBT-36); |
| ZZIREN™ GBT Orogastric Tube - 40 Fr (ZZ-GBT-40) |
#### Indications for Use (Describe)
The ZZIREN™ Orogastric Tube is indicated for use in gastric and bariatric surgical procedures for gastric aspiration and lavage, to serve as a size guide, and to test for leaks.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| 510(k) #: K242901 | 510(k) Summary | Prepared on: 2025-01-06 | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|--|--|
| Contact Details | | 21 CFR 807.92(a)(1) | | |
| Applicant Name | Tools for Surgery, LLC | | | |
| Applicant Address | 8 Technology Drive, #100 East Setauket NY 11733 United States | | | |
| Applicant Contact Telephone | (631)745-2179 | | | |
| Applicant Contact | Dr. Arnold Leiboff | | | |
| Applicant Contact Email | aleiboff(@toolsforsurgery.com | | | |
| Device Name | | 21 CFR 807.92(a)(2) | | |
| Device Trade Name | ZZIREN™ Orogastric Tube;<br>ZZIREN™ SGT Orogastric Tube - 32 Fr (ZZ-SGT-32);<br>ZZIREN™ SGT Orogastric Tube - 36 Fr (ZZ-SGT-36);<br>ZZIREN™ SGT Orogastric Tube - 40 Fr (ZZ-SGT-40);<br>ZZIREN™ GBT Orogastric Tube - 32 Fr (ZZ-GBT-32);<br>ZZIREN™ GBT Orogastric Tube - 36 Fr (ZZ-GBT-36);<br>ZZIREN™ GBT Orogastric Tube - 40 Fr (ZZ-GBT-40) | | | |
| Common Name | Gastrointestinal tube and accessories | | | |
| Classification Name | Tubes, Gastrointestinal (And Accessories) | | | |
| Requlation Number | 876.5980 | | | |
| Product Code(s) | KNT | | | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) | | |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | | |
| K220218 | SIREN SGT ™ | KNT | | |
| K234145 | Boehringer Laboratories ViSiGi® 3D | KNT | | |
| Device Description Summary<br>21 CFR 807.92(a)(4) | | | | |
| The ZZIREN™ Orogastric Tube is a non-sterile, single use device comprising a 37-inch-long insertion tube, a Y extension containing a<br>suction regulating valve, and a pinch clamp. An accessory straight connector is included. The insertion is made of<br>flexible polyvinylchloride (PVC). The suction requlating valve limits negative pressure applied to the stomach or intestine to less than<br>175 mmHg, produces an audible signal when the suction pressure exceeds ~160 mmHg, and is identical in all models. All six models<br>have multiple apertures in proximity to a rounded distal tip. Three of the models, ZZ-SGT-36, ZZ-SGT-40 have side-holes<br>extending 13 cm from the tip, and three models, ZZ-GBT-36, and ZZ-GBT-40, have side-holes extending 6 cm from the tip.<br>The insertion tubes have diameters of 32, 36, and 40 Fr and humeric markings to indicate depth of insertion. | | | | |
| Intended Use/Indications for Use<br>21 CFR 807.92(a)(5) | | | | |
The ZZIREN™ Orogastric Tube is indicated for use in gastric and bariatric surgical procedures for gastric aspiration and lavage, to serve as a size guide, and to test for leaks.
{5}------------------------------------------------
# Indications for Use Comparison
## 21 CFR 807.92(a)(5)
The ZZIREN™ Orogastric Tube is indicated for use in gastric surgical procedures for gastric aspiration and lavage, to serve as a sizing guide, and to test for leaks. The Inclusion of "to test for ZZIREN™ Orogastric Tube differs from the indication stated for the primary predicate device in Submission K220218. "To test for leaks" is included in the indications for use of the secondary reference ViSiGi® 3D, which was cleared per Special 510(k) K234145. Use of ZZREN™ Orogastric Tube for leak testing allows the surgeon to more efficiently assess staple line integrity intraoperatively and does not impact the procedures overall safety and effectiveness or the intended use of the device to assist in completion of gastric and bariatric surgery.
## Technological Comparison
The proposed Tools for Surgery™ ZZIREN™ Orogastric Tubes, models ZZ-SGT-30, ZZ-GBT-32, ZZ-GBT-36, and ZZ-GBT-40 have technical characteristics similar to that of the primary predicate device, Tools for Surgery™ SIREN SGT™ gastrointestinal tube. All devices are orogastric tubes comprised of biocompatible elastomeric materials.
The proposed devices differ from the predicate device, SIREN SGT™ model SGT-36, by having a Y extension in place of three separate parts (Y connector, extension tube, and suction connector), by having the additional insertion tube calibers of 32 and 40 Fr, and by having 32, 48, 72, or 108 distal apertures (side-holes) instead of the 18 or 36 distal apertures specified in the predicate's 510(k) submission (K220218). (The ViSiGi 3Dº model 5236 [510(k) no. K130483], the antecedent predicate device for K220218, has 108 side holes.)
The Y extension is a single, integral, molded unit having a forward stem to join with the insertion tube, a first arm to contain the suction regulating valve, and a second arm having a segment around which the pinch clamp is disposed, and a rearward expansion to receive a straight connector accessory that is attachable to medical suction tubing. All proposed models have similar Y extensions but different insertion tube calibers or number of apertures. All models employ the same pirmary predicate device to interrupt suction. All models, proposed and precical index markings to help the surgeon and anesthesiologist gauge the depth of insertion and the position of the tube within the upper gastrointestinal tract.
All models, proposed and predicate, connect to a vacuum source in order to deliver neqative pressure to the gastion pressure delivered by the vacuum source is adjustable by means of an external vacuum source suction regulator. All models, proposed and primary predicate, incorporate an identical integral suction regulating valve, which serves as additional means for limiting negative pressure applied to the gastric mucosa. This is a pressure relief valve, which opens at 103 to 155 mm Hg and limits the pressure in the insertion tube and stomach to less than negative 175 mm Hq when full suction (~360 mm Hq) is applied. When negative pressure within the insertion tube of any model, proposed or primary predicate, exceeds 160 mm Hg an audible signal produced by mechanical means by the suction regulating valve informs the operator that excessive suction is being applied, and prompts the operator to lower the suction pressure by manually adjusting the vacuum source suction regulator.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
The stiffness, compressibility, and propensity to kink of the insertion tubes of the proposed device are compared to the those the primary and secondary predicates. The pinction of the proposed device is compared to the function of the same part on the primary predicate. Accessory tubing connectors are shown to fit and function well. The proposed device is shown to be compatible with a device for leak testing.
The INSERTION TUBE KINK TEST data shows that the six models of the proposed ZZIREN™Orogastric Tube are substantially equivalent in their propensity to kink to the predicate SIREN SGT™ and are therefore as unlikely to interfere with gastric decompression, lavage, and stapling. The INSERTION TUBE STIFFNESS TEST shows that all models of the proposed device are equally or less stiff than the insertion tubes of their corresponding predicate device and are therefore equally or less likely to cause visceral injury. The INSERTION TUBE COMPRESSION TEST shows that all models of the proposed device are equally or less likely to be inadvertently stapled during gastric surgery than the corresponding predicate device. The PINCH CLAMP TEST shows that the performance of the PINCH CLAMP on the proposed ZZIREN™ Orogastric Tube is substantially equivalent to its performance on the predicate SIREN SGT™. The CONNECTOR TESTS show that both accessory straight tubing connectors attach to the proposed device easily and reversibly in a secure manner and provide a reliable connection to medical suction tubing. The AlR SYRINGE COMPATIBILITY TEST shows that the pariativ Air Syringe are compatible and may be safely used together for leak testing. Dimensional Analysis Tests of all models of the proposed ZZIREN™ device show that all models are made to design specifications.
## 21 CFR 807.92(a)(6)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.