K242898 · Shanghai Jianzhong Medical Packaging Co., Ltd. · FRG · Feb 21, 2025 · General Hospital
Device Facts
Record ID
K242898
Device Name
Sterilization Pouches and Reels
Applicant
Shanghai Jianzhong Medical Packaging Co., Ltd.
Product Code
FRG · General Hospital
Decision Date
Feb 21, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The Sterilization Pouches and Reels is intended to provide health care workers with an effective method to enclosed devices in a single or double pouch configuration by steam or ethylene oxide sterilization. The recommended sterilization cycle parameters are as follows: a) Steam sterilization at 132°C (270°F) for 4 minutes; Drying time of 20 minutes. b) Ethylene Oxide (EO) with a concentration of 800 mg/L at 55 ℃ (131 °F) and 40%-80% relative humidity for 6 hours. Aeration time of 7 days at 20℃ (68 °F). The Sterilization Pouches and Reels is composed of medical compound film, maintains the sterility of the enclosed devices for up to 12 months post Steam or EO gas sterilization, and before sterilization has a maximum shelf life of 5 years from the date of manufacture. The pouch's external chemical indicators are designed to indicate to the user that the pouch has undergone either a steam or EO Sterilization indicators will change color according to the sterilization method (Steam, Ethylene oxide). The color of steam sterilization indicator changes from Pink before sterilization to Cocoa after sterilization. The color of Ethylene oxide sterilization indicator changes from Blue before sterilization to Golden Brown after sterilization. The maximum validated pouch load is 1.87 lbs. (850g); The maximum validated reel load is 1.87 lbs. (850 g)
Device Story
Device consists of medical compound film pouches and reels used by healthcare workers to contain medical instruments during steam or EO sterilization. Input: medical devices placed inside pouch/reel. Operation: heat-sealed by user; subjected to validated steam (132°C) or EO (800 mg/L) cycles. Output: sterilized, sealed package maintaining sterility for 12 months. External chemical indicators (printed on film) change color (Pink-to-Cocoa for steam; Blue-to-Golden Brown for EO) to confirm exposure to sterilization process. Benefits: provides microbial barrier for sterile storage of instruments. Used in clinical settings; operator is healthcare worker. Output confirms process exposure, not sterility; used alongside biological indicators.
Clinical Evidence
No clinical data. Bench testing only, including seal strength (ASTM F88), leak detection (ASTM F1929, F2096), tensile properties (ISO 1924-2), thickness (ASTM F2251), microbial barrier (DIN 58953-6), and biocompatibility (ISO 10993 series).
Technological Characteristics
Materials: medical grade paper and medical compound film. Sensing: chemical indicator bars (ISO 11140-1). Dimensions: 72 variations of pouches and reels. Sterilization: steam (132°C) and EO (800 mg/L). Biocompatibility: non-cytotoxic, non-irritative, non-sensitizing (ISO 10993).
Indications for Use
Indicated for healthcare workers to enclose medical devices for terminal sterilization via steam or ethylene oxide (EO) processes. Suitable for single or double pouch configurations.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format. The logo is simple and professional, and it is easily recognizable.
February 21, 2025
Shanghai Jianzhong Medical Packaging Co., Ltd. Xinhua Wei Medical Device Regulatory Affairs Specialist Bldg 16, 789# Puxing Road, Minhang District Shanghai. 201114 China
Re: K242898
Trade/Device Name: Sterilization Pouches and Reels Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG, JOJ Dated: September 24, 2024 Received: January 24, 2025
Dear Xinhua Wei:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Digitally signed by Stephen A. Stephen A. Anisko -S Date: 2025.02.21 Anisko -S 15:53:10 -05'00'
for: Christopher K. Dugard, M.S. Division Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K242898
Device Name Sterilization Pouches and Reels
#### Indications for Use (Describe)
The Sterilization Pouches and Reels is intended to provide health care workers with an effective method to enclosed devices in a single or double pouch configuration by steam or ethylene oxide sterilization. The recommended sterilization cycle parameters are as follows:
a) Steam sterilization at 132°C (270°F) for 4 minutes; Drying time of 20 minutes.
b) Ethylene Oxide (EO) with a concentration of 800 mg/L at 55 ℃ (131 °F) and 40%-80% relative humidity for 6 hours. Aeration time of 7 days at 20℃ (68 °F).
The Sterilization Pouches and Reels is composed of medical compound film, maintains the sterility of the enclosed devices for up to 12 months post Steam or EO gas sterilization, and before sterilization has a maximum shelf life of 5 years from the date of manufacture. The pouch's external chemical indicators are designed to indicate to the user that the pouch has undergone either a steam or EO Sterilization indicators will change color according to the sterilization method (Steam, Ethylene oxide). The color of steam sterilization indicator changes from Pink before sterilization to Cocoa after sterilization. The color of Ethylene oxide sterilization indicator changes from Blue before sterilization to Golden Brown after sterilization. The maximum validated pouch load is 1.87 lbs. (850g); The maximum validated reel load is 1.87 lbs. (850 g)
The following tables (Table 1~5) identifies the size of subject device and its Maximum validated load. Table 1:Sterilization Pouches, self seal
| Size of device | | Content/Max Load (Ibs.) | | |
|----------------|------------------|-------------------------|---------|--------------|
| Inch | Metric | Metal | Plastic | Gauze/Linens |
| 2.25"×4" | 57× (100+30) mm | 0.004 | 0.0088 | 0.005 |
| 2.25"×2.75" | 57× (70+30) mm | 0.004 | 0.0088 | 0.004 |
| 2.75"×9" | 70× (230+30) mm | 0.020 | 0.04 | 0.02 |
| 3.5"×6.5" | 90× (135+30) mm | 0.016 | 0.032 | 0.016 |
| 3.5"×9" | 90× (230+30) mm | 0.024 | 0.12 | 0.024 |
| 3.5"×10.5" | 90× (270+30) mm | 0.032 | 0.15 | 0.032 |
| 3.5"×22" | 90× (560+30) mm | 0.072 | 0.32 | 0.072 |
| 5"×15" | 127× (380+30) mm | 0.062 | 0.30 | 0.062 |
| 5.25"×10" | 135× (254+29) mm | 0.045 | 0.22 | 0.045 |
| 5.25"×10" | 135× (255+30) mm | 0.045 | 0.22 | 0.045 |
| 5.25"×12" | 135× (305+30) mm | 0.052 | 0.26 | 0.05 |
| 5.25"×13" | 140× (330+30) mm | 0.072 | 0.36 | 0.07 |
| 6"×9" | 150× (230+30) mm | 0.055 | 0.27 | 0.05 |
| 7.5"×13" | 190× (330+30) mm | 0.102 | 0.48 | 0.10 |
| 8"×12.5" | 200× (320+30) mm | 0.12 | 0.53 | 0.10 |
| 10"×14.5" | 250× (370+30) mm | 0.21 | 0.65 | 0.19 |
| 10.6"×17" | 270× (400+30) mm | 0.38 | 0.78 | 0.22 |
| 10.5"×17" | 270× (430+30) mm | 0.42 | 0.82 | 0.24 |
| 12"×15" | 300× (380+30) mm | 0.52 | 0.90 | 0.23 |
| 12"×16.5" | 300× (420+30) mm | 0.58 | 1.02 | 0.25 |
| 12"×18" | 300× (460+30) mm | 0.68 | 1.10 | 0.28 |
| 14"×19" | 355× (480+30) mm | 0.92 | 1.10 | 0.34 |
{4}------------------------------------------------
| 15"×25" | 380× (635+30) mm | 1.30 | 1.10 | 0.50 |
|---------|------------------|------|------|------|
| 20"×22" | 500× (560+40) mm | 1.87 | 1.10 | 0.60 |
Table 2 Sterilization Pouches, flat, heat seal
| Size of device | |
|----------------|-----------|
| Inch | Metric |
| 3.5"×22" | 90×560mm |
| 4"×12" | 100×300mm |
| 5"×15" | 127×380mm |
| 6"×10" | 150×254mm |
| 6"×22" | 150×560mm |
| 7.5"×13" | 190×330mm |
| 8"×20" | 200×500mm |
| 12"×15" | 300×380mm |
| 12"×18" | 300×460mm |
| 16.5"×22.5" | 420×570mm |
## Table 3 Sterilization Pouches, gusseted, heat seal
| Size of device | |
|----------------|---------------|
| Inch | Metric |
| 4"×12" | 100×50×300mm |
| 4"×14" | 100×50×360mm |
| 6"×16" | 150×50×400mm |
| 6"×18" | 150×50×460mm |
| 8"×20" | 200×55×500mm |
| 8"×16" | 200×55×400mm |
| 8"×19" | 250×65×480mm |
| 12"×22" | 300×80×550mm |
| 16.5"×22.5" | 420×100×570mm |
## Table 4 Sterilization Reels, flat, heat seal
| | Size of device |
|--|----------------|
|--|----------------|
| Inch | Metric |
|-------|--------------|
| 3" | 75mm×30.5m |
| 4" | 100mm×30.5m |
| 6" | 150mm×30.5m |
| 8" | 200mm×30.5m |
| 10" | 250mm×30.5m |
| 2" | 50mm × 200m |
| 2.15" | 55mm × 200m |
| 3" | 75mm × 200m |
| 4" | 100mm × 200m |
| 4.7" | 120mm × 200m |
| 6" | 150mm × 200m |
| 8" | 200mm × 200m |
| 10" | 250mm×200m |
| 12" | 300mm × 200m |
| 14" | 350mm × 200m |
| 16" | 400mm × 200m |
| 18" | 450mm × 200m |
| 20" | 500mm × 200m |
| | Table 5 Sterilization Reels, gusseted, heat seal | | | | |
|--|--------------------------------------------------|--|--|--|--|
|--|--------------------------------------------------|--|--|--|--|
| Content/Max Load (Ibs.) | Metal | Plastic | Gauze/Linens |
|-------------------------|-------|---------|--------------|
| | 0.005 | 0.01 | 0.020 |
| | 0.004 | 0.0088 | 0.0176 |
| | 0.07 | 0.20 | 0.055 |
| | 0.06 | 0.19 | 0.05 |
| | 0.12 | 0.32 | 0.10 |
| | 0.10 | 0.25 | 0.09 |
| | 0.12 | 0.53 | 0.10 |
| | 0.44 | 0.54 | 0.38 |
| | 0.58 | 0.74 | 0.52 |
| | 1.87 | 1.10 | 0.60 |
| Content/Max Load (Ibs.) | | |
|-------------------------|---------|--------------|
| Metal | Plastic | Gauze/Linens |
| 0.004 | 0.0088 | 0.0176 |
| 0.02 | 0.04 | 0.025 |
| 0.05 | 0.12 | 0.04 |
| 0.06 | 0.18 | 0.06 |
| 0.12 | 0.53 | 0.10 |
| 0.12 | 0.48 | 0.10 |
| 0.30 | 0.74 | 0.26 |
| 1.15 | 0.98 | 0.40 |
| 1.87 | 1.10 | 0.60 |
| Content/Max Load (Ibs.) | | |
|-------------------------|---------|--------------|
| Metal | Plastic | Gauze/Linens |
| 0.02 | 0.04 | 0.02 |
| 0.025 | 0.15 | 0.032 |
| 0.06 | 0.27 | 0.07 |
| 0.10 | 0.53 | 0.10 |
| 0.12 | 0.53 | 0.10 |
| 0.004 | 0.0088 | 0.004 |
| 0.004 | 0.0088 | 0.005 |
| 0.02 | 0.04 | 0.002 |
| 0.025 | 0.08 | 0.035 |
| 0.04 | 0.10 | 0.04 |
| 0.06 | 0.20 | 0.06 |
| 0.12 | 0.38 | 0.08 |
| 0.12 | 0.53 | 0.10 |
| 0.52 | 0.53 | 0.25 |
| 0.92 | 1.05 | 0.34 |
| 1.08 | 1.07 | 0.52 |
| 1.30 | 1.10 | 0.58 |
| 1.87 | 1.10 | 0.60 |
{5}------------------------------------------------
| Size of device | | Content/Max Load (Ibs.) | | |
|----------------|--------------|-------------------------|---------|--------------|
| Inch | Metric | Metal | Plastic | Gauze/Linens |
| 4.15" | 55mm × 100m | 0.004 | 0.0088 | 0.004 |
| 3" | 75mm × 100m | 0.02 | 0.04 | 0.04 |
| 4" | 100mm × 100m | 0.05 | 0.15 | 0.08 |
| 6" | 150mm × 100m | 0.07 | 0.20 | 0.10 |
| 8" | 200mm x 100m | 0.12 | 0.53 | 0.10 |
| 10" | 250mm x 100m | 0.12 | 0.53 | 0.10 |
| 12" | 300mm × 100m | 1.15 | 0.98 | 0.40 |
| 14" | 350mm x 100m | 1.35 | 1.0 | 0.50 |
| 16" | 400mm × 100m | 1.65 | 1.03 | 0.58 |
| 18" | 450mm x 100m | 1.87 | 1.10 | 0.60 |
| 20" | 500mm x 100m | 1.87 | 1.10 | 0.60 |
#### Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{6}------------------------------------------------
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.The
assigned 510(k) Number: K242898
- 1. Date of Preparation: February 20, 2025
- 2. Sponsor Identification
### Shanghai Jianzhong Medical Packaging Co., Ltd.
Bldg 16, 789# Puxing Road, Minhang District, Shanghai 201114, P.R. China Registration Number: 3005030944. Contact Person: Xinhua Wei Position: Medical Device Regulatory Affairs Specialist Tel: +86-21-54315666-5201 Fax: +86-21-54315801 Email: weixinhua@mpackchina.com
- 3. Identification of Proposed Device
Trade Name: Sterilization Pouches and Reels
Common Name: Sterilization Pouches and Reels
Regulatory Information
Classification Name: Wrap, Sterilization
Classification: II;
Product Code: FRG;
Subsequent Product Code: JOJ;
Regulation Number: 21 CFR 880.6850
Review Panel: General Hospital;
- 4. Identification of Predicate Device 510(k) Number: K221875 Product Name: Sterilization Pouch and Roll Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II
{7}------------------------------------------------
Product Code: FRG, Subsequent Product Code: JOJ Dated: January 17, 2023 Received: January 17, 2023
#### 5. Device Description
The Sterilization Pouches and Reels is composed of medical compound film. It is intended to be used to contain medical devices to be terminally sterilized by Steam or EO sterilization process. The recommended sterilization cycle parameters are as follows:
- Steam sterilization at 132°C (270°F) for 4 minutes; Drying time of 20 minutes. a)
- b) Ethylene Oxide (EO) with a concentration of 800 mg/L at 55 °C (131 °F) and 40%-80% relative humidity for 6 hours. Aeration time of 7 days at 20℃ (68 °F).
The medical devices are inserted into the Pouches/Reels, and then sterilized for the EO or Steam Sterilization Process. The heat-sealed Pouches/Reels are heat sealed prior to sterilization processing. After completion of the sterilization process, the Pouches/Reels maintain sterility of the enclosed medical devices until the seal is opened. The Sterilization Pouches and Reels maintains the sterility of the enclosed devices for up to 12 months post Steam or EO gas sterilization, and before sterilization has a maximum shelf life of 5 years from the date of manufacture.
The Sterilization Pouches and Reels is printed with chemical indicator bars that changed from Pink to Cocoa(Steam) or Blue to Golden Brown (EO) when exposed to steam vapor or EO gas during process. The steam and EO chemical indicator offer an addition way to verify processing in the sterilization cycle. The chemical indicator should be used in addition to, not in place of, the biological indicator. The steam and EO chemical indicators do not signify sterilization; they only indicate that the indicators have been exposed to the Steam or EO.
- 6. Indication for Use Statement
The Sterilization Pouches and Reels is intended to provide health care workers with an effective method to enclosed devices in a single or double pouch configuration by steam or ethylene oxide sterilization. The recommended sterilization cycle parameters are as follows:
a) Steam sterilization at 132°C (270°F) for 4 minutes; Drying time of 20 minutes.
b) Ethylene Oxide (EQ) with a concentration of 800 mg/L at 55 ℃ (131 °F) and 40%-80% relative humidity for 6 hours. Aeration time of 7 days at 20°C (68 °F).
The Sterilization Pouches and Reels is composed of medical compound film, maintains the sterility of the enclosed devices for up to 12 months post Steam or EO gas sterilization, and before sterilization has a maximum shelf life of 5 years from the date of manufacture. The pouch's external chemical indicators are designed to indicate to the user that the pouch has undergone either a steam or EO Sterilization process. Sterilization indicators will change color according to the sterilization method (Steam, Ethylene oxide). The color of steam sterilization indicator changes from Pink before sterilization to Cocoa after sterilization. The color of
{8}------------------------------------------------
Ethylene oxide sterilization indicator changes from Blue before sterilization to Golden Brown after sterilization. The maximum validated pouch load is 1.87 lbs. (850g); The maximum validated reel load is 1.87 lbs. (850 g)
| Item | Size of Device | | Content/Max Load(Ibs) | | |
|----------------------------------------------------------------|----------------|----------------|-----------------------|---------|--------------|
| | Inch | Metric | Metal | Plastic | Gauze/Linens |
| Sterilization<br>Pouches,<br>self-seal | 2.25"x4" | 57x(100+30)mm | 0.004 | 0.0088 | 0.005 |
| | 2.25"x2.75" | 57x(70+30)mm | 0.004 | 0.0088 | 0.004 |
| | 2.75"x9" | 70x(230+30)mm | 0.020 | 0.04 | 0.02 |
| | 3.5"x6.5" | 90x(135+30)mm | 0.016 | 0.032 | 0.016 |
| | 3.5"x9" | 90x(230+30)mm | 0.024 | 0.12 | 0.024 |
| | 3.5"x10.5" | 90x(270+30)mm | 0.032 | 0.15 | 0.032 |
| | 3.5"x22" | 90x(560+30)mm | 0.072 | 0.32 | 0.072 |
| | 5"x15" | 127x(380+30)mm | 0.062 | 0.30 | 0.062 |
| | 5.25"x10" | 135x(254+29)mm | 0.045 | 0.22 | 0.045 |
| | 5.25"x10" | 135x(255+30)mm | 0.045 | 0.22 | 0.045 |
| | 5.25"x12" | 135x(305+30)mm | 0.052 | 0.26 | 0.05 |
| | 5.25"x13" | 140x(330+30)mm | 0.072 | 0.36 | 0.07 |
| es,<br>self-<br>seal | 6"x9" | 150x(230+30)mm | 0.055 | 0.27 | 0.05 |
| | 7.5"x13" | 190x(330+30)mm | 0.102 | 0.48 | 0.10 |
| | 8"x12.5" | 200x(320+30)mm | 0.12 | 0.53 | 0.10 |
| | 10"x14.5" | 250x(370+30)mm | 0.21 | 0.65 | 0.19 |
| | 10.6"x17" | 270x(400+30)mm | 0.38 | 0.78 | 0.22 |
| | 10.5"x17" | 270x(430+30)mm | 0.42 | 0.82 | 0.24 |
| | 12"x15" | 300x(380+30)mm | 0.52 | 0.90 | 0.23 |
| | 12"x16.5" | 300x(420+30)mm | 0.58 | 1.02 | 0.25 |
| | 12"x18" | 300x(460+30)mm | 0.68 | 1.10 | 0.28 |
| | 14"x19" | 355x(480+30)mm | 0.92 | 1.10 | 0.34 |
| | 15"x25" | 380x(635+30)mm | 1.30 | 1.10 | 0.50 |
| | 20"x22" | 500x(560+40)mm | 1.87 | 1.10 | 0.60 |
| Sterilization<br>Pouches,<br>flat,<br>heat<br>seal | 3.5"x22" | 90x560mm | 0.005 | 0.01 | 0.020 |
| | 4"x12" | 100x300mm | 0.004 | 0.0088 | 0.0176 |
| | 5"x15" | 127x380mm | 0.07 | 0.20 | 0.055 |
| | 6"x10" | 150x254mm | 0.06 | 0.19 | 0.05 |
| | 6"x22" | 150x560mm | 0.12 | 0.32 | 0.10 |
| | 7.5"x13" | 190x330mm | 0.10 | 0.25 | 0.09 |
| | 8"x20" | 200x500mm | 0.12 | 0.53 | 0.10 |
| | 12"x15" | 300x380mm | 0.44 | 0.54 | 0.38 |
| | 12"x18" | 300x460mm | 0.58 | 0.74 | 0.52 |
| | 16.5"x22.5" | 420x570mm | 1.87 | 1.10 | 0.60 |
| | 4"x12" | 100x50x300mm | 0.004 | 0.0088 | 0.0176 |
| | 4"x14" | 100x50x360mm | 0.02 | 0.04 | 0.025 |
| Sterilization<br>Pouch<br>es,<br>gusset<br>ed,<br>heat<br>seal | 6"x16" | 150x50x400mm | 0.05 | 0.12 | 0.04 |
| | 6"x18" | 150x50x460mm | 0.06 | 0.18 | 0.06 |
| | 8"x20" | 200x55x500mm | 0.12 | 0.53 | 0.10 |
| | 8"x16" | 200x55x400mm | 0.12 | 0.48 | 0.10 |
| | 8"x19" | 250x65x480mm | 0.30 | 0.74 | 0.26 |
| | 12"x22" | 300x80x550mm | 1.15 | 0.98 | 0.40 |
| | 16.5"x22.5" | 420x100x570mm | 1.87 | 1.10 | 0.60 |
| | 3" | 75mmx30.5m | 0.02 | 0.04 | 0.02 |
| | 4" | 100mmx30.5m | 0.025 | 0.15 | 0.032 |
| | 6" | 150mmx30.5m | 0.06 | 0.27 | 0.07 |
| | 8" | 200mmx30.5m | 0.10 | 0.53 | 0.10 |
| | 10" | 250mmx30.5m | 0.12 | 0.53 | 0.10 |
| | 2" | 50mm x 200m | 0.004 | 0.0088 | 0.004 |
| | 2.15" | 55mm x 200m | 0.004 | 0.0088 | 0.005 |
| Sterilization<br>Reels,<br>flat,<br>heat<br>seal | 3" | 75mm x 200m | 0.02 | 0.04 | 0.002 |
| | 4" | 100mm x 200m | 0.025 | 0.08 | 0.035 |
| | 4.7" | 120mm x 200m | 0.04 | 0.10 | 0.04 |
| | 6" | 150mm x 200m | 0.06 | 0.20 | 0.06 |
| | 8" | 200mm x 200m | 0.12 | 0.38 | 0.08 |
| | 10" | 250mmx200m | 0.12 | 0.53 | 0.10 |
| | 12" | 300mm x 200m | 0.52 | 0.53 | 0.25 |
| | 14" | 350mm x 200m | 0.92 | 1.05 | 0.34 |
| | 16" | 400mm x 200m | 1.08 | 1.07 | 0.52 |
| | 18" | 450mm x 200m | 1.30 | 1.10 | 0.58 |
| | 20" | 500mm x 200m | 1.87 | 1.10 | 0.60 |
| | 4.15" | 55mm x 100m | 0.004 | 0.0088 | 0.004 |
| Sterilization<br>Reels,<br>gusset<br>ed, | 3" | 75mm x 100m | 0.02 | 0.04 | 0.04 |
| | 4" | 100mm x 100m | 0.05 | 0.15 | 0.08 |
| | 6" | 150mm x 100m | 0.07 | 0.20 | 0.10 |
| | 8" | 200mm x 100m | 0.12 | 0.53 | 0.10 |
| heat<br>seal | 10" | 250mm x 100m | 0.12 | 0.53 | 0.10 |
| | 12" | 300mm x 100m | 1.15 | 0.98 | 0.40 |
| | 14" | 350mm x 100m | 1.35 | 1.0 | 0.50 |
| | 16" | 400mm x 100m | 1.65 | 1.03 | 0.58 |
| | 18" | 450mm x 100m | 1.87 | 1.10 | 0.60 |
| | 20" | 500mm x 100m | 1.87 | 1.10 | 0.60 |
(Model and Dimension and Content/Max. Load)
{9}------------------------------------------------
{10}------------------------------------------------
#### 7. Non-Clinical Test Conclusion
Performance testing and Biocompatibility assessment were conducted to verify that the subject device met all design specifications. The test results demonstrated that the subject device meet the following standards (included but not fully described):
- > ASTM F88/F88M-2015, Standard Test Method for Seal Strength of Flexible Barrier Materials.(Sterility)
- > ASTM F1929-2015 Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration
- > ASTM F 2096-2011 (2019) Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test)
- A ISO 1924-2:2008 Paper and board -- Determination of tensile properties.
- A ASTM F 2251-13 Standard Test Method for Thickness Measurement of Flexible Packaging Material
- > ISO 11140-1:2014 Sterilization of health care products. Chemical indicators. General requirements
- > ISO 10993-1: Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process
- > ISO 10993-5:2009 Biological evaluation of medical devices- Part 5: Tests for in vitro cytotoxicity
- > ISO 10993-7:2008 Biological Evaluation of Medical Device- Part 7: Ethylene Oxide Sterilization Residuals
- > ISO 10993-10:2010 Biological evaluation of medical device- Part 10: Tests for irritation and skin sensitization
- > ISO 10993-12: Biological evaluation of medical devices Part 12: Sample preparation and reference materials
- > ISO 11135: 2014 Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices.
- > DIN58953-6-2010 Sterilization - Sterile supply - Part 6: Microbial barrier testing of packaging materials for medical devices which are to be sterilized.
- > ISO17665-1:2006 Sterilization of health care products-Moist heat-Part1: requirements for the development, validation and routine control of a sterilization process for medical devices.
{11}------------------------------------------------
#### 8. Technological Characteristic Comparison Table
| ITEM | Subject Device | Predicate Device K221875 | Comparison<br>analysis |
|-------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | FRG, JOJ | FRG, JOJ | same |
| Regulation No. | 21 CFR 880.6850 | 21 CFR 880.6850 | same |
| Class | II | II | same |
| Indication for Use | The Sterilization Pouches and Reels is intended to provide<br>health care workers with an effective method to enclosed<br>devices in a single or double pouch configuration by steam or<br>ethylene oxide sterilization. The recommended sterilization<br>cycle parameters are as follows:<br>a) Steam sterilization at 132°C (270°F) for 4 minutes;<br>Drying time of 20 minutes.<br>b) Ethylene Oxide (EO) with a concentration of 800 mg/L<br>at 55 °C (131 °F) and 40%-80% relative humidity for 6<br>hours. Aeration time of 7 days at 20℃ (68 °F).<br>The Sterilization Pouches and Reels is composed of medical<br>grade paper and medical compound film, maintains the sterility<br>of the enclosed devices for up to 12 months post Steam or EO<br>gas sterilization, and before sterilization has a maximum shelf<br>life of 5 years from the date of manufacture.<br>The pouch's external chemical indicators are designed to<br>indicate to the user that the pouch has undergone either a steam<br>or EO Sterilization process. Sterilization indicators will change<br>color according to the sterilization method (Steam, Ethylene<br>oxide). The color of steam sterilization indicator changes from<br>Pink before sterilization to Cocoa after sterilization. The color<br>of Ethylene oxide sterilization indicator changes from Blue<br>before sterilization to Golden Brown after sterilization.<br>The maximum validated pouch load is 1.87 lbs. (850g); The<br>maximum validated reel load is 1.87 lbs. (850 g) | The Sterilization Pouch and Roll are intended to provide health<br>care workers with an effective method to enclose devices<br>intended for sterilization in the Steam or via Ethylene Oxide<br>(EO). The recommended sterilization cycles are as follows:<br>• Steam Sterilization at 132°C<br>(270°F) for 4 minutes; Drying time of 20 minutes.<br>• Ethylene Oxide (EO) with a concentration of 800 mg/L at<br>55°C (131 °F) and 40% to 90% relative humidity for 6 hours.<br>Aeration time of 7 days at 20℃).<br>The sterilization pouch and roll are made with medical grade<br>paper and medical compound film. The sterilization pouch and<br>roll maintains the sterility of the enclosed devices for up to 24<br>months post Steam or EO gas sterilization, and before<br>sterilization has a maximum shelf life of 5 years from the date<br>of manufacture. The pouch's external chemical ink indicators<br>are designed to indicate to the user that the pouch has<br>undergone either a steam or EO sterilization process.<br>Sterilization indicator will change color according to the<br>sterilization method (ethylene oxide, steam). Ethylene oxide<br>sterilization indicator color change from Pink before<br>sterilization to Yellow after sterilization. The steam<br>sterilization indicator color change from Blue before<br>sterilization to Dark Grey after sterilization. | Same |
| | | | K242898 |
| Material Composition | Composed of medical grade paper and medical compound film<br>and printed with a chemical indicator bar that changed from<br>Pink to Cocoa (Steam) or Blue to Golden Brown (EO) when<br>exposed to Steam vapor or EO gas during process.<br>"Sterilization Pouches, flat, self seal" type contains tape. | Composed of medical grade paper, medical compound film,<br>EO and Steam Process Indicator. | Similar. The<br>Pouches with tape<br>were validated for<br>their effectiveness<br>and safety. The test<br>results show that<br>the pouches are as<br>effective and safe<br>as predicate device. |
| Sterilization<br>Cycles | Steam sterilization at 132°C (270°F) for 4 minutes; Drying time<br>of 20 minutes.<br>Ethylene Oxide (EO) with a concentration of 800 mg/L at 55 °C<br>(131 °F) and 40%-80% relative humidity for 6 hours. Aeration<br>time of 7 days at 20°C (68 °F). | Steam Sterilization at 132°C (270°F) for 4 minutes; Drying time<br>of 20 minutes.<br>Ethylene Oxide (EO) with a concentration of 800 mg/L at 55°C<br>(131 °F) and 40% to 90% relative humidity for 6 hours. Aeration<br>time of 7 days at 20°C. | same |
| Dimensions | Length of Pouches ranges from 2.75" to 25";<br>Width of Pouches ranges from 2.25" to 20";<br>Length of Reels ranges from 30.5m (=1200.78") to 200m<br>(=7874.01");<br>Width of Reels ranges from 2" to 20";<br>There are 72 sizes in total.<br>Please refer to table of Comparison of Dimension for detailed<br>information. | Length of Pouches ranges from 3.5" to 17";<br>Width of Pouches ranges from 2.25" to 12";<br>Length of Reels (or Rolls) ranges from 100m (=3937") to<br>200m (=7874.01");<br>Width of Reels ranges from 2" to 17.7"<br>There are 32 sizes in total.<br>Please refer to table of Comparison of Dimension for detailed<br>information. | Similar |
| Sterilant Penetration Efficiency | It meets the requirement of SAL10-6 | It meets the requirement of SAL10-6 | same |
| EO、ECH residue | Referring to ISO 10993-7:2008, EO<4mg, ECH<9mg | Referring to ISO 10993-7:2008, EO<4mg, ECH<9mg | same |
| | Performance Testing | | |
| Chemical Indicator (CI)<br>Functionality and Endpoint | The sterilant penetrated through the pouch configuration and<br>rendered the CI color the endpoint color. | The sterilant penetrated through the pouch configuration and<br>rendered the CI color the endpoint color. | same |
| Device Design of Steam CI | It changes from Pink to Cocoa when exposed to Steam vapor.…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.