QuickRad

K242838 · Aikenist Technologies Pvt, Ltd. · LLZ · Feb 21, 2025 · Radiology

Device Facts

Record IDK242838
Device NameQuickRad
ApplicantAikenist Technologies Pvt, Ltd.
Product CodeLLZ · Radiology
Decision DateFeb 21, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesAI/ML, Software as a Medical Device

Indications for Use

QuickRad is a web-based PACS intended for use in radiology centers, diagnostic labs, and hospitals. It is designed to receive, store, transmit, process, and display medical images and associated data from various DICOM-compliant sources, including CT scanners, MRI systems, ultrasound machines, and X-ray devices. The system enhances communication and accessibility by allowing distributed access to images and data across computer networks. QuickRad also offers optional integration with FDA-cleared third-party Al models, enabling the visualization of Al-generated outputs. The system displays these outputs "as-is," with the safety and effectiveness of the Al models being covered under the original third-party manufacturer's regulatory clearance.Mammographic images may only be interpreted using a monitor that meets the technical specifications identified by the FDA. The system is designed for utilization by proficient and certified medical practitioners, including physicians, radiologists, and medical technicians.

Device Story

QuickRad is a web-based PACS for radiology workflow management; receives, stores, transmits, processes, and displays DICOM images from CT, MRI, ultrasound, and X-ray sources. Operates via web browser (Chrome, Edge, Firefox) on Windows, macOS, or Ubuntu; supports on-premise or cloud hosting. Provides 3D visualization tools (MPR, MIP), measurement tools (length, angle, ROI), and report editing. Integrates with FDA-cleared third-party AI models (e.g., Qure AI) to display AI-generated outputs "as-is" alongside original images. Used by radiologists and medical technicians in clinical settings to facilitate image review, diagnostic interpretation, and report generation. Enhances clinical efficiency and accessibility through distributed network access. Security features include HTTPS, AES 256 encryption, and multi-factor authentication.

Clinical Evidence

Bench testing only. No clinical data. Performance validated via usability testing (IEC 62366-1:2015), VAPT cybersecurity assessment, and software verification/validation. Measurement accuracy study compared QuickRad against FlexView Diagnostic (K233226) using 12 X-ray images for angles (0°-180°) and CT/MR images for length/ROI/reference lines; results demonstrated equivalence and clinical reliability.

Technological Characteristics

Web-based PACS; supports DICOM standards (NEMA PS 3.1-3.20). Software environment: Windows 10, web browsers (Chrome, Edge, Firefox). Security: HTTPS, AES 256 encryption, MFA. Visualization: 3D (MPR, MIP), measurement tools. Standards: ISO 13485:2016, IEC 62304:2015, IEC 62366-1:2015, ISO 14971:2019.

Indications for Use

Indicated for use in radiology centers, diagnostic labs, and hospitals by proficient medical practitioners (physicians, radiologists, medical technicians) to receive, store, transmit, process, and display DICOM-compliant medical images (CT, MRI, ultrasound, X-ray). Mammographic images require FDA-compliant monitors.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION". Aikenist Technologies Pvt Ltd % David Dills Regulatory Consultant David Dills 5450 Ariva Way #302 Lakeland, Florida 33812 February 21, 2025 Re: K242838 Trade/Device Name: QuickRad Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: LLZ Dated: September 19, 2024 Received: January 23, 2025 Dear David Dills: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Re"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See {2}------------------------------------------------ the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Jessica Lamb Jessica Lamb Assistant Director Imaging Software Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K242838 Device Name QuickRad Indications for Use (Describe) QuickRad is a web-based PACS intended for use in radiology centers, diagnostic labs, and hospitals. It is designed to receive, store, transmit, process, and display medical images and associated data from various DICOM-compliant sources, including CT scanners, MRI systems, ultrasound machines, and X-ray devices. The system enhances communication and accessibility by allowing distributed access to images and data across computer networks. QuickRad also offers optional integration with FDA-cleared third-party Al models, enabling the visualization of Al-generated outputs. The system displays these outputs "as-is," with the safety and effectiveness of the Al models being covered under the original third-party manufacturer's regulatory clearance.Mammographic images may only be interpreted using a monitor that meets the technical specifications identified by the FDA. The system is designed for utilization by proficient and certified medical practitioners, including physicians, radiologists, and medical technicians. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for Alkenist. The logo consists of the letters "AI" inside of a circle that is partially colored in yellow and blue. To the right of the circle is the word "Alkenist" in a sans-serif font. ## 510(k) Summary [AS REQUIRED BY 21CFR807.92] # I. APPLICANT INFORMATION | Submitter's Name | Aikenist Technologies Pvt. Ltd. | | |--------------------------|-------------------------------------------------------------------------------------|--| | Submitter's Address | 007, Pushpanjali, 1st Cross, 1st Main Road, Chamarajpet,<br>Bangalore 560018, India | | | Name of Contact Person | Ashwin Amarapuram Chandramouly | | | Designation | Director and CEO | | | Contact Number | +91 9845320314 | | | Contact E-mail | ashwin@aikenist.com | | | Date of Summary Prepared | 22 Jan 2025 | | # II. DEVICE DETAILS | Device Trade Name | QuickRad | |-------------------------------------|------------------------------------------------| | Regulation Name/ Device Common Name | Medical image management and processing system | | Regulation Number | 21 CFR 892.2050 | | Device Class | Class 2 | | Product Code | LLZ | ## III. PREDICATE DEVICE DETAILS | Device Trade Name | Augmento | |--------------------------|-----------------------------------------| | Device Manufacturer Name | DEEPTEK MEDICAL IMAGING PRIVATE LIMITED | | 510(k) Number | K222781 | | Regulation Number | 21 CFR 892.2050 | | Device Class | Class 2 | | Product Code | LLZ | Image /page/4/Picture/10 description: The image shows a black silhouette of a factory. The factory has three triangular peaks of the same height. To the right of the peaks is a tall rectangular chimney. {5}------------------------------------------------ ## IV. DEVICE DESCRIPTION QuickRad is a web-based PACS designed to optimize radiology workflows and enhance clinical efficiency. It allows seamless receipt, management, and processing of DICOM images from various modalities, including CT, MRI, ultrasound, X-ray, and more. ### Key Features and Functions: - Comprehensive Radiology Management: QuickRad provides end-to-end functionality for the storage, transmission, processing, and visualization of medical images, ensuring seamless collaboration across healthcare networks. - Zero Footprint Viewer (ZFP): QuickRad's ZFP viewer offers cross-platform accessibility without the need for additional native software installation. It is compatible with macOS, Windows, Ubuntu and Chrome, Edge, Firefox browsers. - . Enhanced Viewer Options: QuickRad ensures that all scans are displayed with true-to-source fidelity, offering optional FDA-cleared viewer integrations that enhance diagnostic accuracy. - . Advanced 3D Viewer: QuickRad's 3D viewer empowers radiologists with cutting-edge image analysis tools, including: - o Smooth stack scrolling, zooming, panning, and reset options. - O Measurement tools for length, angles, and annotations. - Advanced visualization techniques such as Multiplanar Reconstruction (MPR) and Maximum Intensity O Projection (MIP). - O PET scan mode with customizable contrast settings, annotation capabilities, and snapshot tools for enhanced diagnostic insights. - . Smart Reporting Editor: QuickRad includes a modality-specific report editor that simplifies report creation. Pre-built templates for various imaging studies ensure consistent report generation with voiceenabled editing features. - . AI Integration: QuickRad offers optional integration with FDA-cleared third-party AI models, allowing radiologists to review AI-generated results directly within the platform. The AI output is displayed alongside original images for confirmation by qualified medical professionals. - DICOM Compliance: Fully compliant with DICOM standards, QuickRad guarantees secure and interoperable ● image and data exchange across systems, ensuring compatibility with a wide range of imaging devices. - . Secure Data Transmission: It adheres to strict security protocols, utilizing HTTPS, AES 256 encryption, and multi-factor authentication (MFA) for secure access. ## V. INDICATIONS FOR USE QuickRad is an advanced PACS solution, available for on-premise or cloud hosting. It receives medical images and data from various DICOM-compliant origins, such as CT, MR scanners, ultrasound systems, and X-ray machines. These images and data undergo storage, transmission, processing, and display within the system, fostering smooth communication and accessibility across distributed locations on computer networks. It provides optional integration with FDA-cleared 3rd party AI models. The solution only supports the visualization of outputs of 3rd party AI models "as-is". The safety and effectiveness of the 3rd party model are covered under the original 3rd party manufacturer's regulatory clearance. Mammographic images may only be interpreted using a monitor that meets the technical specifications identified by the FDA. The system is designed for utilization by proficient and certified medical practitioners, including physicians, radiologists, and medical technicians. {6}------------------------------------------------ # VI.COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATEDEVICE ### Table 1. Summary of Technological Characteristics Comparison of DeepTek and QuickRad. | SI.<br>No | Features | Predicate Device-Augmento | Proposed Device- QuickRad | Remarks | |-----------|---------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------| | | | General information | | | | 1. | 510(k) Number | K222781 | K242838 | - | | 2. | Manufacturer | DeepTek Medical Imaging<br>Private Limited | Aikenist Technologies<br>Private Limited | - | | 5. | Classification | Class II | Class II | Same | | 6. | Regulation number | 21 CFR 892.2050 | 21 CFR 892.2050 | Same | | 7. | Product Code | LLZ | LLZ | Same | | 8. | Indications For<br>Use | Augmento is a web-based PACS<br>and radiology workflow<br>management solution.<br>It accepts digital images and data<br>from various DICOM-compliant<br>sources (i.e. CT scanners, MR<br>scanners, ultrasound systems, RF<br>Units, PET Units, computed &<br>digital radiographic devices,<br>secondary capture devices,<br>imaging gateways, and other<br>imaging sources). Images and data<br>can be stored, communicated,<br>processed, and displayed within<br>the system and/or across computer<br>networks at distributed locations.<br>Only preprocessed DICOM "for<br>presentation" images can be<br>interpreted for primary image<br>diagnosis in mammography. Lossy<br>compressed images and digitized<br>film screens of mammographic<br>images must not be reviewed for<br>primary image interpretations.<br>Mammographic images may only<br>be interpreted using a monitor that<br>meets the technical specifications<br>identified by the FDA. This system<br>is meant to be used by trained and<br>qualified medical professionals | QuickRad is an advanced PACS<br>solution, available for on-premise<br>or cloud hosting. It receives<br>medical images and data from<br>various DICOM-compliant<br>origins, such as CT, MR scanners,<br>ultrasound systems, and X-ray<br>machines. These images and data<br>undergo storage, transmission,<br>processing, and display within the<br>system, fostering smooth<br>communication and accessibility<br>across distributed locations on<br>computer networks. It provides<br>optional integration with FDA-<br>cleared 3rd party AI models. The<br>solution only supports the<br>visualization of outputs of 3rd<br>party Al models "as-is". The<br>safety and effectiveness of the 3rd<br>party model are covered under the<br>original 3rd party manufacturer's<br>regulatory clearance.<br>Mammographic images may only<br>be interpreted using a monitor that<br>meets the technical specifications<br>identified by the FDA.<br>The system is designed for<br>utilization by proficient and | Same | | | physicians, radiologists, nurses, and medical technicians. | certified medical practitioners, including physicians, radiologists, and medical technicians. | | | | 9. | Prescription or OTC | Prescription | Same | | | | Specifications | | | | | 10. | Modalities | Various image sources | Same | | | 11. | Web browser software | Google Chrome, Mozilla, and Edge | Same | | | 12. | Resolution | 32-bit Color Display & 1920x1080 | Same | | | 13. | Image Storage | Yes | Same | | | 14. | Software environment | OS: Windows 10 | Same | | | | Functions | | | | | 15. | Main Functions | Log In Worklist - Search Filter Worklist - Open image Work list- study List Worklist - Report Worklist - Series Viewer - View exam Viewer - Control View window Viewer - view mode (real resolution) Viewer – View mode (Highlight) Viewer - Stacking Viewer – Changing the layout Viewer – Comparative study Viewer – Preset filter Viewer - Zoom Viewer – Panning Viewer - Invert Image Viewer – Viewing mode (Normal/ Image/ Stack/ Custom/ Annotation) Viewer - Comparative study Viewer - Rotation MIP/MPR Reconstruction Viewer – Reference line | Log In Worklist - Search Filter Worklist - Open image Work list- study List Worklist - Report Worklist - Series Viewer - View exam Viewer - Thumbnail view of the series Viewer-Changing the layout Viewer-Comparative study Viewer - Preset filter Viewer - Zoom Viewer - Panning Viewer - Invert Image Viewer – Viewing mode (Normal/Stack/ Custom/ Annotation/ synchronized screening) Viewer – Comparative study Viewer - Rotation Viewer- MIP/MPR Reconstruction Viewer – Reference line Viewer-Measurement- rectangular, ellipse, angle/cobb angle Viewer – Cine/movie mode Viewer- Overlaying/ Hide or display overlay Viewer- Magnify | Similar¹ | | | | Viewer – Measure Viewer – Inverting imagecolor Viewer – Cine Viewer- Overlaying. | Viewer- crop | | | 16. | 3D Cursor | Yes | Yes | Same | | 17. | Optional<br>Integration of<br>FDA-cleared<br>3rd party AI<br>models | Yes | Yes | Same | | 18. | Operation feature | Web environment-based<br>PACS.<br>Viewing and handling DICOM<br>medical images. Review and<br>report study located on a<br>server. | Web environment-based<br>PACS.<br>Viewing and handling DICOM<br>medical images. Review, modify,<br>and approve the study located in a<br>server. | Same | {7}------------------------------------------------ {8}------------------------------------------------ 1Deep Tek's Augmento viewer offers a sharpening feature for enhanced image clarity, a specific function that QuickRad's viewer does not include. QuickRad and DeepTek share extensive similarities in their core functionalities. Both systems are designed to manage radiology workflows effectively, supporting various modalities, resolution, image storage, and adhering to DICOM standards. They provide essential features such as study worklists, and software environments, making them comparable in many aspects. Integration with AI models in QuickRad is optional, allowing users the flexibility to incorporate AI capabilities as needed. QuickRad supports the integration of the following FDA-cleared AI models from Qure AI, offering seamless functionality while displaying outputs "as-is": - qXR: Chest X-ray interpretation (FDA 510(k) clearance: K231149) ● - qCT LN Quant: Advanced lung nodule quantification for chest CT scans (FDA 510(k) clearance: K231149) . - qER: AI for triaging intracranial hemorrhage and other conditions in CT scans (FDA 510(k) clearance: K212624) These integrations ensure compliance with regulatory standards while enhancing diagnostic capabilities. A key difference is that DeepTek's Augmento viewer includes a sharpening feature to enhance image clarity, which is not available in QuickRad's viewer. Despite this distinction, QuickRad maintains substantial equivalence to DeepTek in terms of overall performance and functionality. QuickRad offers comparable capabilities with added flexibility and adaptability, catering to diverse user needs. The absence of the sharpening feature in QuickRad does not affectiveness and safety. QuickRad delivers robust performance with its flexible and adaptable platform, aligning well with DeepTek's offerings. # VII. NON-CLINICAL PERFORMANCE DATA The non-clinical performance activities include: - . Risk Management & Cybersecurity - Device Hazard Analysis as per ISO 14971:2019 O - Vulnerability Assessment & Penetration Testing (VAPT) based on OWASP Top 10. o - . Usability Testing - Formative and Summative Usability Testing per IEC 62366-1:2015 o - Software Verification & Validation - Unit Testing based on Software Requirements Specifications (SRS) O {9}------------------------------------------------ - Integration Testing for interaction between software components o - System Testing covering end-to-end functionality O - Validation based on User Requirements Specifications (URS) o - . Measurement Study - Establishing equivalence with FDA-cleared viewer O - . 3rd Party AI Model Integration Interface Testing Non-clinical testing consisted of: - · QuickRad underwent a thorough hazard analysis, addressing both intentional risks in compliance with ISO 14971:2019. Strategies to mitigate these risks were implemented to ensure safety throughout the device lifecycle, and the hazard analysis and mitigation strategies were validated, confirming QuickRad's compliance with ISO 14971:2019. - · Periodic VAPT by third-party consultants were conducted to verify the adequacy of QuickRad's security controls, effectively mitigating potential cybersecurity risks. - · Usability testing, conducted in compliance with IEC 62366-1:2015, included both formative and summative evaluations. - · Unit testing verified core functional components were independently tested, ensuring QuickRad's units work as intended, with code correctness validated. Integration testing confirmed seamless interaction between software components and external systems, while system testing validated end-to-end functionality across all features. User Acceptance Testing verified QuickRad meets its intended use, ensuring adherence to the user requirement specifications documents through end-user validation. The test cases were addressfully passed the acceptance criteria. - · Integration testing for QuickRad validated seamless end functionality across its software components, including APIs, OTS/SOUP components detailed in the SBOM, independent libraries, and DICOM interfaces. System testing encompassed core functionalities such as user management. DICOM viewer operations, search capabilities, radiologist assignment workflows, report management, audit log tracking, error handling, and hospital management, ensuring robust performance and compliance with the system's defined specifications. - · An angle measurement study was conducted using 12 X-ray images (AP Pelvic, AP Knee, Lateral Ankle), evaluated independently by two experienced radiologists. Angles spanning the 0° to 180° range were measured using QuickRad's Viewer and benchmarked against the FDA-cleared FlexView Diagnostic DICOM Viewer (K233226). Statistical analyses, including equivalence testing and T-tests, demonstrated that QuickRad's angle measurement tool is substantially equivalent to FlexView Diagnostic and produced clinically reliable angle assessments. - · A study was conducted to validate measurement tools in QuickRad's Viewer. Length, rectangle, ellipse ROI, and probe measurements were evaluated on CT abdomen images, while reference line accuracy was assessed using MR spine and brain images. This comprehensive evaluation, encompassing diverse imaging modalities, confirmed that QuickRad delivers performance equivalent to the FDA-cleared FlexView Diagnostic DICOM Viewer (K233226), ensuring consistent, reliable, and clinically accurate measurements across the spectrum of diagnostic imaging. - · The integration interface between QuickRad and 3rd party AI models were tested to confirm that the integrity of the source DICOM file and output from Al models is unaltered and hence, does not get impacted when the source file is sent to the AI model for findings and the AI model sends the output to QuickRad through the interface. #### The standards applicable as per Recognized Consensus Standards are as follows: - · ISO 13485: 2016 Edition 3- Quality Management Systems for Medical Devices - · IEC 62304 Edition 1.1 2015-06 Consolidated Version Medical device software life cycle processes. - · IEC 62366-1:2015: Medical Devices 1: Application of usability engineering to medical devices. - · ISO 14971 Third Edition 2019-12 Medical Devices Application of risk management to medical devices. - · NEMA PS 3.1 3.20 2023e Digital Imaging and Communications in Medicine (DICOM) Set. {10}------------------------------------------------ - · IEC ISO 10918-1 First edition 1994-02-15 Information technology Digital compression and coding of continuous-tone still images: Requirements and guidelines [Including Technical Corrigendum 1 (2005)]. ## VIII. CONCLUSION The subject device, QuickRad, and the predicate device, DeepTek (K222781), have similar indications for use, technological characteristics, and principles of operation. After a careful evaluation of the technological features and non-clinical performance testing, it is evident that QuickRad is as safe and effective as DeepTek. Furthermore, QuickRad does not introduce any new concerns related to safety or effectiveness. Hence, QuickRad is considered substantially equivalent to DeepTek.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...