← Product Code [KCT](/productcode/KCT) · K242834

# System 9 Sterile Battery Container (K242834)

_Stryker Instruments · KCT · Jan 16, 2025 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K242834

## Device Facts

- **Applicant:** Stryker Instruments
- **Product Code:** [KCT](/productcode/KCT.md)
- **Decision Date:** Jan 16, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6850
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The Stryker System 9 Sterile Battery Container is a reusable rigid sterilization container intended to allow for sterilization, by hospitals, surgery centers, and healthcare facilities, of the enclosed, compatible Stryker Sterile Battery Packs and maintain sterility of the enclosed device until used. The container is validated for use with compatible battery packs in low temperature hydrogen peroxide sterilization cycles.

## Device Story

System 9 Sterile Battery Container is a reusable rigid container for sterilizing Stryker battery packs via low-temperature hydrogen peroxide cycles. Device consists of a Radel (PPSU) base and stainless-steel lid with filter retention plate and single-use polypropylene filters. User loads battery pack into base, attaches lid, secures latching mechanism, and applies tamper-evident lock. Container maintains sterile barrier for up to 45 days post-sterilization. Used in healthcare facilities; operated by clinical staff. Output is a sterilized, protected battery pack ready for clinical use. Sterilization verification performed via indicator strips. Design allows battery charging while maintaining sterile barrier. Benefits include safe, efficient sterilization and storage of power components for surgical equipment.

## Clinical Evidence

No clinical data. Bench testing only. Performance validated via sterilization efficacy (ISO 14937, ISO 22441), toxicological risk assessment (ISO 10993-17), cleaning validation (AAMI ST98), maintenance of sterility (AAMI ST77), simulated use design validation, and human factors engineering (IEC 62366-1).

## Technological Characteristics

Rigid, reusable container. Base: Radel (PPSU). Lid: Stainless steel. Filter: Single-use polypropylene. Sterilization: Low-temperature hydrogen peroxide (STERRAD/Steris V-PRO cycles). Dimensions: Small (3.2"H x 4.2"W x 12.9"L), Large (3.9"H x 4.2"W x 12.9"L). Standards: ISO 14937, ISO 22441, ISO 10993-17, ISO 17664-1, AAMI TIR 12, AAMI ST98, AAMI ST77, IEC 62366-1.

## Regulatory Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

## Predicate Devices

- Aesculap Aicon™ Container ([K214041](/device/K214041.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

January 16, 2025

Stryker Instruments Megan Guilbault Staff Regulatory Affairs Specialist 1941 Stryker Way Portage, Michigan 49002

Re: K242834

Trade/Device Name: System 9 Sterile Battery Container Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: September 19, 2024 Received: September 19, 2024

Dear Megan Guilbault:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Stephen A. Anisko -S

Digitally signed by Stephen A. Anisko -S Date: 2025.01.16 20:37:15 -05'00'

for: Christopher K. Dugard Assistant Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

Submission Number (if known)

K242834

Device Name

System 9 Sterile Battery Container

Indications for Use (Describe)

The Stryker System 9 Sterile Battery Container is a reusable rigid sterilization container intended to allow for sterilization, by hospitals, surgery centers, and healthcare facilities, of the enclosed, compatible Stryker Sterile Battery Packs and maintain sterility of the enclosed device until used. The container is validated for use with compatible battery packs in low temperature hydrogen peroxide sterilization cycles.

Validated Low Temperature Hydrogen Peroxide Sterilization Cycles: Load Configuration Cycles

System 9 Sterile Battery Container Bundle, Small: STERRAD NX Standard cycle STERRAD 100NX Standard cycle STERRAD 100S short cycle Steris V-PRO maX Lumen cycle Steris V-PRO maX 2 Lumen cycle Steris V-PRO 60 Lumen cycle Steris V-PRO s2 Lumen cycle

System 9 Sterile Battery Container Bundle, Large: STERRAD NX Standard cycle STERRAD 100NX Standard cycle STERRAD 100S short cycle Steris V-PRO maX Lumen cycle

Steris V-PRO maX 2 Lumen cycle

Steris V-PRO 60 Lumen cycle

Steris V-PRO s2 Lumen cycle

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

> Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{4}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

# 1.0 Submitter Information

## 1.1 This Premarket Notification is submitted by:

Stryker Instruments 1941 Stryker Way Portage MI 49002 USA

## 1.2 Contact Information

| Contact name:       | Megan Guilbault             |
|---------------------|-----------------------------|
| Cellular Telephone: | (401) 241-6152              |
| Email:              | megan.guilbault@stryker.com |
| Date Prepared:      | January 15, 2025            |

### 1.3 Device Name

### Table 1: Subject Device Information

| Subject Device Information  |                                                                     |
|-----------------------------|---------------------------------------------------------------------|
| Trade/Proprietary Name      | System 9 Sterile Battery Container                                  |
| Common Name                 | Sterilization Container Wrap                                        |
| Classification              | Class II                                                            |
| Classification Product Code | KCT                                                                 |
| Classification Name         | Sterilization Wrap Containers, Trays, Cassettes & Other Accessories |
| Classification Regulation   | 21 CFR 880.6850                                                     |
| Review Panel                | General Hospital                                                    |

### 1.4 Predicate Device

| Table 2: Predicate Device Information |               |              |                |
|---------------------------------------|---------------|--------------|----------------|
| Predicate Device Trade Name           | 510(k) Number | Product Code | Manufacturer   |
| Aesculap Aicon™ Container             | K214041       | KCT          | Aesculap, Inc. |

## 1.5 Device Description

The Stryker System 9 Sterile Battery Container is a rigid, reusable device intended to enclose compatible Stryker sterile battery packs for hydrogen peroxide sterilization. The container consists of a stainless-steel perforated lid with a stainless-steel filter retention plate and a plastic (Radel®) base. The Sterile Battery container utilizes single use polypropylene filters and indicator cards. The container base is designed to exclusively fit onto the Stryker sterile battery charger to allow the sterile battery pack to charge while maintaining a sterile barrier. The Stryker System 9 Sterile Battery Container encompasses one distinct configuration in two sizes, small and large.

{6}------------------------------------------------

# 1.6 Indications for Use

The Stryker System 9 Sterile Battery Container is a reusable rigid sterilization container intended to allow for sterilization, by hospitals, surgery centers, and healthcare facilities, of the enclosed, compatible Stryker Sterile Battery Packs and maintain sterility of the enclosed device until used. The container is validated for use with compatible battery packs in low temperature hydrogen peroxide sterilization cycles.

| Load Configuration                                     | Cycles                         |
|--------------------------------------------------------|--------------------------------|
| System 9 Sterile Battery<br>Container Bundle,<br>Small | STERRAD NX Standard cycle      |
|                                                        | STERRAD 100NX Standard cycle   |
|                                                        | STERRAD 100S short cycle       |
|                                                        | Steris V-PRO maX Lumen cycle   |
|                                                        | Steris V-PRO maX 2 Lumen cycle |
|                                                        | Steris V-PRO 60 Lumen cycle    |
|                                                        | Steris V-PRO s2 Lumen cycle    |
| System 9 Sterile Battery<br>Container Bundle,<br>Large | STERRAD NX Standard cycle      |
|                                                        | STERRAD 100NX Standard cycle   |
|                                                        | STERRAD 100S short cycle       |
|                                                        | Steris V-PRO maX Lumen cycle   |
|                                                        | Steris V-PRO maX 2 Lumen cycle |
|                                                        | Steris V-PRO 60 Lumen cycle    |
|                                                        | Steris V-PRO s2 Lumen cycle    |

Validated Low Temperature Hydrogen Peroxide Sterilization Cycles:

# 1.7 Comparison of Technological Characteristics

A comparison of the technological characteristics of the subject devices included in the scope of this Traditional 510(k) is included in Table 3 below.

| Description                   | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Discussion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use                  | The Stryker System 9 Sterile Battery Container is a reusable rigid sterilization container                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | A device intended to be used to enclose another medical device that is to                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Similar                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Description                   | Subject Device<br>System 9 Sterile Battery<br>Container                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Predicate Device<br>Aesculap AICON TM<br>Container                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Discussion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                               | intended to allow for<br>sterilization, by hospitals,<br>surgery centers, and healthcare<br>facilities, of the enclosed,<br>compatible Stryker Sterile<br>Battery Packs and maintain<br>sterility of the enclosed device<br>until used.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | be sterilized by a health<br>care provider. It is<br>intended to allow<br>sterilization of the<br>enclosed medical device<br>and also to maintain<br>sterility of the enclosed<br>device until used.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Classification                | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Product<br>Code               | KCT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | KCT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Condition of<br>Use           | Professional Use Only,<br>Reusable, for use in a<br>healthcare facility                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Professional Use Only,<br>Reusable, for use in a<br>healthcare facility                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Patient<br>Population         | Not intended for direct patient<br>contact                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Not intended for direct<br>patient contact                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Operating<br>Principle        | The user loads Stryker Sterile<br>Battery Packs into the Sterile<br>Battery Container Base. The<br>Sterile Battery Container Lid<br>with installed filter and filter<br>retention plate is attached to the<br>Sterile Battery Container Base,<br>the external latching<br>mechanism is closed, and the<br>Tamper-Evident Lock is placed<br>on the outside latch. The lid,<br>filter, filter retention plate, and<br>base all generate a sterile<br>barrier. The loaded container<br>becomes sterile once it has been<br>run through a low-temperature<br>hydrogen peroxide sterilizer.<br>The indicator strips color-<br>change mechanism is used to<br>verify successful sterilization.<br>Once sterilized, the container<br>maintains sterility until opened,<br>up to 45 days. | The user places<br>instruments inside the<br>container in a manner to<br>allow proper sterilant<br>penetration. The<br>container lid is closed<br>securely with a locking<br>mechanism to ensure<br>airtight seal. The<br>container is placed in the<br>sterilization chamber<br>(under various<br>sterilization modalities).<br>After completion of<br>appropriate sterilization<br>cycle, verification of<br>sterilization is conducted<br>using indicator strips<br>color-change<br>mechanism. The<br>containers are then<br>stored in a clean, dry<br>environment to maintain<br>sterility and opened in a<br>sterile area when ready<br>to use. Once sterilized<br>the container maintains | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                               | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Discussion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Description                   | System 9 Sterile Battery<br>Container                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Aesculap AICON™<br>Container<br>sterility for 365 days or<br>until opened.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Sterilization<br>Method       | Low-temperature hydrogen<br>peroxide sterilization  STERRAD NX Standard cycle STERRAD 100NX Standard cycle STERRAD 100S Short cycle Steris V-PRO maX Lumen cycle Steris V-PRO maX 2 Lumen cycle Steris V-PRO 60 Lumen cycle Steris V-PRO s2 Lumen cycle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Low-temperature<br>hydrogen peroxide<br>sterilization  Steam Sterilization STERIS V-PRO maX/maX 2: Lumen,<br>Non-Lumen, Flex STERIS V-PRO 60:<br>Lumen, Non- Lumen, Flex STERRAD 100NX:<br>Standard, Express, Flex, Duo STERRAD NX:<br>Standard, Advanced STERRAD 100S STERIZONE VP4 EtO                                                                                                                                                                                                                                                                                                                                                                                                        | Similar                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Container<br>Base<br>Material | Radel® PPSU<br>Polyphenylsulfone                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Anodized Aluminum                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Different<br>The predicate<br>device is made<br>from material<br>that is<br>compatible with<br>a variety of<br>sterilization<br>methods. As the<br>subject device is<br>only being<br>cleared with<br>hydrogen<br>peroxide<br>sterilization,<br>Stryker has<br>chosen material<br>suitable for the<br>specific<br>sterilization<br>method and to<br>prevent<br>interference for<br>wireless charging                                                                                                                                                         |
|                               | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Discussion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Description                   | System 9 Sterile Battery<br>Container                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Aesculap AICONTM<br>Container                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Container<br>Lid Material     | Stainless Steel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Anodized Aluminum                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Different<br>The predicate<br>device is made<br>from material<br>that is<br>compatible with<br>a variety of<br>sterilization<br>methods. As the<br>subject device is<br>only being<br>cleared with<br>hydrogen<br>peroxide<br>sterilization,<br>Stryker has<br>chosen material<br>suitable for the<br>specific<br>sterilization<br>method.<br>Verification and<br>Validation<br>Testing confirms<br>that the subject<br>device is safe and<br>effective for its<br>intended use and<br>the differences<br>do not raise<br>different types of<br>questions of |
|                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | battery and<br>charger.<br>Verification and<br>Validation<br>Testing confirms<br>that the subject<br>device is safe and<br>effective for its<br>intended use and<br>the differences<br>do not raise<br>different types of<br>questions of<br>Safety and<br>Effectiveness.                                                                                                                                                                                                                                                                                    |
| Description                   | Subject Device<br>System 9 Sterile Battery<br>Container                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Predicate Device<br>Aesculap AICON™<br>Container                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Discussion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Filter<br>Material            | Polypropylene (Single Use)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Polypropylene (Single Use)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Maintenance<br>of Sterility   | 45 Days                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 365 Days                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Different<br>The predicate<br>device maintains<br>a sterile barrier<br>on the shelf for<br>365 days,<br>whereas the<br>subject maintains<br>a sterile barrier<br>for 45 days.<br>Testing did not<br>raise any<br>different types of<br>questions of<br>Safety and<br>Effectiveness.                                                                                                                                                                                                                                                                          |
| Device Size                   | Small: 3.2" H x 4.2" W x 12.9" L<br><br>Large: 3.9" H x 4.2" W x 12.9" L                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Full size: 4" H x 11 ¾" W x 23 5/8" L<br><br>Full size: 6" H x 11 ¾" W x 23 5/8" L<br><br>Full size: 8" H x 11 ¾" W x 23 5/8" L<br><br>Full size:10" H x 11 ¾" W x 23 5/8" L<br><br>Three-Quarter: 4" H x<br>11 ¾" W x 18 3/8" L<br><br>Three-Quarter: 6" H x<br>11 ¾" W x 18 3/8" L<br><br>Three-Quarter: 8" H x<br>11 ¾" W x 18 3/8" L<br><br>Three-Quarter:10" H x<br>11 ¾" W x 18 3/8" L                                                                                                                                                                                                                                                                                                    | Different<br>The predicate<br>offers a larger<br>range of sizes<br>while the subject<br>device is<br>designed to<br>specifically be<br>used with the<br>System 9 Sterile<br>Batteries, which<br>are available in<br>two sizes, small<br>and large.<br>Verification<br>testing confirms<br>that the subject<br>device is safe and<br>effective for its<br>intended use and<br>the differences<br>do not raise<br>different types of<br>questions of<br>Safety and<br>Effectiveness.                                                                           |
| Description                   | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Discussion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                               | System 9 Sterile Battery Container                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Aesculap AICON™ Container                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Half size: 4" H x 11 ¾" W x 11 7/8" L                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Half size: 6" H x 11 ¾" W x 11 7/8" L                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Half size: 8" H x 11 ¾" W x 11 7/8" L                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Half size: 10" H x 11 ¾" W x 11 7/8" L                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Container Design              | Solid Bottom                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Solid Bottom                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                               | Perforated Lid with filter retention plates and single use polypropylene filters                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Perforated Lid with filter retention plates and single use polypropylene filters                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                               | Gasket                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Gasket                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |

Table 3: Technological Comparison

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

# 1.8 Summary of Non-Clinical Testing

The intended use of the subject device and the predicate device are identical, and the technological characteristics are similar. The differences between the two devices do not raise any new or different types of questions of safety and effectiveness.

Risk management was conducted in accordance with ISO 14971.

A suite of non-clinical testing was performed in accordance with applicable standards. See table below:

|  |  |  |  | Table 4: Performance Testing |  |
|--|--|--|--|------------------------------|--|
|--|--|--|--|------------------------------|--|

| Testing                             | Standard in<br>compliance with   | Acceptance Criteria           | Results |
|-------------------------------------|----------------------------------|-------------------------------|---------|
| Sterilization<br>Efficacy           | ISO 14937:2009<br>ISO 22441:2022 | Confirms sterile efficacy     | Pass    |
| Toxicological<br>Risk<br>Assessment | ISO 10993-17                     | Confirms acceptable residuals | Pass    |

{12}------------------------------------------------

| Testing                               | Standard in<br>compliance with                            | Acceptance Criteria                                                                                                                                                                                                | Results |
|---------------------------------------|-----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Cleaning<br>Validation                | AAMI ST98:2022                                            | Cleaning instructions were validated and<br>demonstrated that the subject devices<br>could be visually and quantifiably cleaned                                                                                    | Pass    |
| Maintenance of<br>Sterility           | AAMI<br>ST77:2013/(R)2018                                 | Demonstrates that the subject device is<br>capable of maintaining product sterility for<br>the duration of the shelf life                                                                                          | Pass    |
| Simulated Use<br>Design<br>Validation | ISO 17664-1:2021<br>AAMI TIR 12:2020<br>ST77:2013/(R)2018 | Confirms that the device can undergo<br>repeated processing to support that the<br>device's<br>or<br>performance is maintained following a<br>minimum number of exposures to the<br>proposed sterilization methods | Pass    |
| Human Factors<br>Engineering          | IEC 62366-1 Ed 1.1 b<br>2020                              | Confirms safety and effectiveness of the<br>subject device for the intended users, uses,<br>and use environments                                                                                                   | Pass    |

Performance testing was performed in compliance with the following standards:

- ISO 14937:2009 Sterilization of health care products Low temperature . vaporized hydrogen peroxide - Requirements for the development, valid…

---

**Source:** [https://fda-staging.innolitics.com/device/K242834](https://fda-staging.innolitics.com/device/K242834)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
