NITRILE EXAMINATION GLOVES (Tested For Use With Chemotherapy Drugs)
K242812 · Nb Medical Company Limited · LZA · Oct 12, 2024 · General Hospital
Device Facts
Record ID
K242812
Device Name
NITRILE EXAMINATION GLOVES (Tested For Use With Chemotherapy Drugs)
Applicant
Nb Medical Company Limited
Product Code
LZA · General Hospital
Decision Date
Oct 12, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
NITRILE EXAMINATION GLOVES (Tested For Use With Chemotherapy Drugs) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05(2023) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
Device Story
Nitrile examination gloves; non-sterile; disposable; ambidextrous; blue color; sizes S/M/L/XL. Used by healthcare examiners in clinical settings to prevent cross-contamination. Tested for permeation resistance against specific chemotherapy drugs per ASTM D6978-05(2023). Provides barrier protection; chemotherapy drug permeation data included in labeling to inform clinical decision-making regarding safe handling of hazardous drugs. Not for use with Carmustine or ThioTEPA due to low permeation times.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Material: Nitrile (NBR). Powder-free. Ambidextrous. Non-sterile. Single-use. Dimensions and physical properties (tensile strength, elongation) meet ASTM D6319-19. Permeation resistance tested per ASTM D6978-05(2023). Biocompatibility tested per ISO 10993-5, 10993-10, and 10993-11.
Indications for Use
Indicated for use as a non-sterile, disposable, ambidextrous patient examination glove worn by healthcare examiners to prevent cross-contamination between patient and examiner. Suitable for use with specific chemotherapy drugs as tested per ASTM D6978-05(2023).
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 12, 2024
Nb Medical Company Limited % Wendy Huang Technical Manager Wenzhou ThiWe Business Consulting Co., Ltd. Room 1203, Building C, Hua'ou Jiayuan, No.50 Tangjiaqiao South Road, Longwan District Wenzhou, Zhejiang 325000 China
Re: K242812
Trade/Device Name: NITRILE EXAMINATION GLOVES (Tested For Use With Chemotherapy Drugs) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, OPJ Dated: August 12, 2024 Received: September 18, 2024
Dear Wendy Huang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bifeng Qian -S
Bifeng Qian, M.D, Ph.D. Assistant Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K242812
Device Name NITRILE EXAMINATION GLOVES (Tested For Use With Chemotherapy Drugs)
#### Indications for Use (Describe)
NITRILE EXAMINATION GLOVES (Tested For Use With Chemotherapy Drugs) is a non-sterile disposable device intended for medical purposes that is worn on the exammer's hands or finger to prevent contamination between patient and examiner. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05(2023) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes, and can be used with the device.
Cyclophosphamide (20.0 mg/ml) Doxorubicin HCl (2.0 mg/ml) Etoposide (20.0 mg/ml) Fluorouracil (50.0 mg/ml) Paclitaxel (6.0 mg/ml) Cisplatin (1.0 mg/ml) Dacarbazine (10.0 mg/ml)
The following chemotherapy drugs have low permeation times: Carmustine (3.3 mg/ml): 15 minutes ThioTEPA (10.0 mg/ml): 97 minutes
Warning: Not for Use with: Carmustine, ThioTEPA
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary
## K242812
This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.
## 1.0 Submitter Information
| · Company Name: | NB MEDICAL COMPANY LIMITED |
|-----------------|---------------------------------------------------------------------------------------------|
| · Address: | Tay An Industrial Cluster, Tien Hai Town, Tien Hai District,<br>Thai Binh Province, Vietnam |
| · Phone Number: | +84-934-388877 |
| · Contact: | VU THI VINH |
# 2.0 Designated Submission Correspondent
| · Company Name: | Wenzhou ThiWe Business Consulting Co., Ltd. | |
|-----------------|-------------------------------------------------------------------------------------------------------------------------------|--|
| · Address: | Room 1203, Building C, Hua'ou Jiayuan, No.50 Tangjiaqiao<br>South Road, Longwan District, Wenzhou, Zhejiang, 325000,<br>CHINA | |
| · Phone Number: | 86-577-88201260 | |
| · Contact: | Ms. Wendy Huang | |
| · Email: | thiwe_reg@outlook.com | |
# 3.0 Date of Preparation: October 3, 2024
## 4.0 Device Information
・Trade/Device Name: NITRILE EXAMINATION GLOVES (Tested For Use With Chemotherapy Drugs)
・Model(s): S, M, L, XL (Blue)
・ Common Name: Patient Examination Gloves (Tested For Use With Chemotherapy Drugs)
・510(k) Number: K242812
{5}------------------------------------------------
#### 5.0 Classification
- · Device: Polymer Patient Examination Glove & Medical Glove, Specialty & Medical Gloves With Chemotherapy Labeling Claims - Test For Use With Chemotherapy Drugs
- · Regulation Description: Non-powdered patient examination glove.
- · Review Panel: General Hospital
- · Product Code: LZA, LZC, OPJ
- · Submission Type: 510(k)
- · Regulation Number: 21 CFR 880.6250
- · Device Class: 1
### 6.0 Predicate Device Information
- · Device Name: NITRILE EXAMINATION GLOVES
- ・510(k) Number: K220343
- · Manufacturer: NB Medical Co., LTD (i.e. NB MEDICAL COMPANY LIMITED)
### 7.0 Device Description
NITRILE EXAMINATION GLOVES (Tested For Use With Chemotherapy Drugs) is a Class I, patient examination gloves bearing the product codes LZA, LZC, OPJ (21CFR880.6250). They meet all the current specifications listed under the ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application and also complies with requirements for Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs as per D6978-05(2023). They are made from Nitrile (NBR). These gloves are blue in color and are powder free. The product is non-sterile, ambidextrous and single use only. NITRILE EXAMINATION GLOVES (Tested For Use With Chemotherapy Drugs) are offered in different sizes: S/M/L/XL.
{6}------------------------------------------------
#### 8.0 Indications for Use
NITRILE EXAMINATION GLOVES (Tested For Use With Chemotherapy Drugs) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05(2023) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes, and can be used with the device.
Cyclophosphamide (20.0 mg/ml) Doxorubicin HCl (2.0 mg/ml) Etoposide (20.0 mg/ml) Fluorouracil (50.0 mg/ml) Paclitaxel (6.0 mg/ml) Cisplatin (1.0 mg/ml) Dacarbazine (10.0 mg/ml)
The following chemotherapy drugs have low permeation times: Carmustine (3.3 mg/ml): 15 minutes ThioTEPA (10.0 mg/ml): 97 minutes
Warning: Not for Use with: Carmustine, ThioTEPA
# 9.0 Comparison of Intended Use and Technological Characteristics with the Predicate Device
The table below compares the intended use and technological characteristics of the subject and predicate device.
## Table 1: Comparison Table for Subject and Predicate Device
{7}------------------------------------------------
| Item | Subject Device<br>(K242812) | Predicate Device<br>(K220343) | Comparison |
|------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| General Comparison | | | |
| Product Code | LZA, LZC, OPJ | LZA | Different<br>Analysis 1 |
| Regulation No. | 21 CFR880.6250 | 21 CFR880.6250 | Same |
| Class | 1 | 1 | Same |
| Intended<br>Use/Indications for<br>Use | NITRILE<br>EXAMINATION<br>GLOVES (Tested For<br>Use With<br>Chemotherapy Drugs)<br>is a non-sterile<br>disposable device<br>intended for medical<br>purposes that is worn<br>on the examiner's<br>hands or finger to<br>prevent contamination<br>between patient and<br>examiner. These gloves<br>were tested for use with<br>Chemotherapy Drugs<br>as per ASTM<br>D6978-05(2023)<br>Standard Practice for<br>Assessment of Medical<br>Gloves to Permeation<br>by Chemotherapy | NITRILE<br>EXAMINATION<br>GLOVES is a<br>non-sterile disposable<br>device intended for<br>medical purposes that is<br>worn on the examiner's<br>hands or finger to<br>prevent contamination<br>between patient and<br>examiner. | Different<br>Analysis 1 |
| | Drugs. | | |
| Material | Nitrile | Nitrile | Same |
| Powder Free | Yes | Yes | Same |
| Design Feature | Ambidextrous | Ambidextrous | Same |
| Color | Blue | White/Cobalt<br>Blue/Black/Blue | Different<br>Analysis 2 |
| Size | S/M/L/XL | S/M/L/XL | Same |
| Sterility | Non-Sterile | Non-Sterile | Same |
| Use | Single Use | Single Use | Same |
| Rx Only or OTC | OTC | OTC | Same |
| Chemotherapy<br>Drugs Claim | The following<br>chemotherapy drugs<br>and concentration had<br>NO breakthrough<br>detected up to 240<br>minutes, and can be<br>used with the device.<br>Cyclophosphamide<br>(20.0 mg/ml)<br>Doxorubicin HCl (2.0<br>mg/ml)<br>Etoposide (20.0 mg/ml)<br>Fluorouracil (50.0<br>mg/ml)<br>Paclitaxel (6.0 mg/ml)<br>Cisplatin (1.0 mg/ml)<br>Dacarbazine (10.0<br>mg/ml) | No chemotherapy drugs<br>claim | Different<br>Analysis 1 |
| | The following<br>chemotherapy drugs<br>have low permeation<br>times:<br>Carmustine (3.3 mg/ml): 15 minutes<br>ThioTEPA (10.0 mg/ml): 97 minutes<br><br>Warning: Not for Use with: Carmustine, ThioTEPA | | |
| Labeling<br>Information | Single-use indication,<br>powder free, device color, device name,<br>glove size and quantity,<br>Non-Sterile, applicable<br>chemotherapy drugs<br>and the minimum<br>breakthrough time | Single-use indication,<br>powder free, device<br>color, device name,<br>glove size and quantity,<br>Non-Sterile<br>Different<br>Analysis 1 | |
| Technological Characteristic and Biocompatibility Comparison | | | |
| Dimensions -<br>Length | S: ≥220mm<br>M/L/XL: ≥230mm<br>Meet the requirement<br>of ASTM D6319-19 | S: ≥220mm<br>M/L/XL: ≥230mm<br>Meet the requirement<br>of ASTM D6319-19<br>Same | |
| Dimensions -<br>Width | S: 80±10 mm<br>M: 95±10 mm<br>L: 110±10 mm | S: 80±10 mm<br>M: 95±10 mm<br>L: 110±10 mm<br>Same | |
| | Meet the requirement<br>of ASTM D6319-19 | Meet the requirement<br>of ASTM D6319-19 | |
| Physical<br>Properties -<br>Tensile Strength | before aging ≥14 Mpa<br>after aging ≥14 Mpa | before aging ≥14 Mpa<br>after aging ≥14 Mpa | Same |
| | Meet the requirement<br>of ASTM D6319-19 | Meet the requirement<br>of ASTM D6319-19 | |
| | | | |
| Physical<br>Properties -<br>Ultimate<br>Elongation | before aging ≥500 %<br>after aging ≥400 % | before aging ≥500 %<br>after aging ≥400 % | Same |
| | Meet the requirement<br>of ASTM D6319-19 | Meet the requirement<br>of ASTM D6319-19 | |
| | | | |
| Thickness | Palm: ≥0.05 mm<br>Finger: ≥0.05 mm | Palm: ≥0.05 mm<br>Finger: ≥0.05 mm | Same |
| | Meet the requirement<br>of ASTM D6319-19 | Meet the requirement<br>of ASTM D6319-19 | |
| | | | |
| Powder Free<br>Residue | ≤2.0mg per glove | ≤2.0mg per glove | Same |
| | Meet the requirement<br>of ASTM D 6124<br>following ASTM<br>D6319-19 | Meet the requirement<br>of ASTM D 6124<br>following ASTM<br>D6319-19 | |
| | | | |
| Freedom from<br>holes | Be free from holes<br>when tested in<br>accordance with ASTM<br>D5151 following<br>ASTM D6319-19<br>AQL=2.5 | Be free from holes<br>when tested in<br>accordance with ASTM<br>D5151 following<br>ASTM D6319-19<br>AQL=2.5 | Same |
| | | | |
| | | | |
| Biocompatibility | ISO 10993-5<br>Under conditions of the<br>study, device extract is<br>cytotoxic | ISO 10993-5<br>Under conditions of the<br>study, device extract is<br>cytotoxic | Same |
| | | | |
| | | | |
| ISO 10993-10 | ISO 10993-10 | Same | |
| Under the conditions of<br>the study, not an irritant<br>or a sensitizer | Under the conditions of<br>the study, not an irritant<br>or a sensitizer | | |
| ISO 10993-11 | ISO 10993-11 | Same | |
| Under the condition of<br>acute systemic toxicity<br>test, the test article did<br>not show acute<br>systemic toxicity in<br>vivo. | Under the condition of<br>acute systemic toxicity<br>test, the test article did<br>not show acute<br>systemic toxicity in<br>vivo. | | |
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
Analysis 1: The subject device has different indications for use to the predicate device, which results in different product codes, that is, the subject device has extra applicable product codes compared with the predicate device. The major difference between the two devices is that the subject device is claimed for use with certain chemotherapy drugs, while the predicate device did not. However, such difference has been tested according to FDA recognized standard - ASTM D6978-05(2023) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs, which ensures that the subject device will be as safe as for use as the predicate device. Meanwhile, chemotherapy drugs and the minimum breakthrough time of subject device will be listed on labeling, so this different will not bring new concerns regarding safety and effectiveness.
Analysis 2: The subject device is offered only in a single color (Blue) whereas the predicate is offered in different colors (White/Cobalt Blue/Black/Blue), but the color of the subject device is included by those of the predicate device, so such difference will not bring questions regarding safety and effectiveness.
{12}------------------------------------------------
## 10.0 Summary of Non-Clinical Performance Testing
Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:
ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves
ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.
ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.
ASTM D6978-05(2023) Standard Practice for Assessment of Medical Gloves to
Permeation by Chemotherapy Drugs.
ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
ISO 10993-11:2017, Biological evaluation of medical devices - Part 11: Tests for systemic toxicity.
| Test Method | Purpose | Acceptance Criteria | Results |
|-------------|-----------------------------|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| ASTM D6319 | Physical<br>Dimensions Test | Length:<br>S: $\geq$ 220mm | Length:<br>S: $\geq$ 220mm |
| | | M/L/XL: $\geq$ 230mm | M/L/XL: $\geq$ 230mm |
| | | Width:<br>S: 80 $\pm$ 10 mm<br>M: 95 $\pm$ 10 mm<br>L: 110 $\pm$ 10 mm<br>XL: 120 $\pm$ 10 mm | Width:<br>S: 84-86/Pass<br>M: 96-102/Pass<br>L: 104-107/Pass<br>XL: 114-116/Pass |
| | | Thickness: | Thickness: |
Table 2: Summary of non-clinical performance testing
{13}------------------------------------------------
| | | Palm: ≥0.05 mm<br>Finger: ≥0.05 mm | Finger:<br>0.09-0.11/Pass<br><br>Palm:<br>0.07-0.08/Pass |
|--------------------------|-----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ASTM D5151<br>ASTM D6319 | Watertightness<br>Test for<br>Detection of<br>Holes | Meet the requirements<br>of ASTM D5151 AQL<br>2.5 | 0/125/Pass |
| ASTM D6124<br>ASTM D6319 | Powder Content | Meet the requirement<br>of ASTM D 6124<br>≤2.0mg | 0.16-0.19mg/Pass |
| ASTM D412<br>ASTM D6319 | Physical<br>Properties | Tensile Strength:<br>before aging ≥14 Mpa<br>after aging ≥14 Mpa<br>Ultimate Elongation:<br>before aging ≥500 %<br>after aging ≥400 % | Tensile Strength:<br>before aging:<br>20-24Mpa<br>after aging:<br>15-23Mpa<br>Ultimate<br>Elongation:<br>before aging:<br>501-598%<br>after aging:<br>471-561% |
| ASTM D6978 | Chemotherapy<br>drug claim | The chemotherapy<br>drugs and<br>concentration which<br>have NO breakthrough<br>detected up to 240 | The following<br>chemotherapy<br>drugs and<br>concentration had<br>NO breakthrough |
| | minutes can be<br>suggested for use with<br>the subject device. | detected up to 240<br>minutes, and can be<br>used with the<br>device.<br>Cyclophosphamide<br>(20.0 mg/ml)<br>Doxorubicin HCl<br>(2.0 mg/ml)<br>Etoposide (20.0<br>mg/ml)<br>Fluorouracil (50.0<br>mg/ml)<br>Paclitaxel (6.0<br>mg/ml)<br>Cisplatin (1.0<br>mg/ml)<br>Dacarbazine (10.0<br>mg/ml)<br>The following<br>chemotherapy<br>drugs have low<br>permeation times:<br>Carmustine (3.3<br>mg/ml): 15 minutes<br>ThioTEPA (10.0<br>mg/ml): 97 minutes<br>Warning: Not for | |
| | | | Use with: |
| | | | Carmustine, |
| | | | ThioTEPA |
| ISO 10993-5 | Cytotoxicity | toxicity | Under conditions<br>of the study, device<br>extract is cytotoxic. |
| ISO 10993-10 | Irritation | Non-irritating | Under the<br>conditions of the<br>study, not an<br>irritant. / Pass |
| ISO 10993-10 | Sensitization | Non-sensitizing | Under the<br>conditions of the<br>study, not an<br>sensitizer. / Pass |
| ISO 10993-11 | Acute Systemic<br>Toxicity | Non-acute systemic | Under the<br>conditions of the<br>study, did not show<br>acute systemic<br>toxicity in vivo. /<br>Pass |
{14}------------------------------------------------
{15}------------------------------------------------
# Table 3 Chemotherapy Permeation Testing Claim
| Chemotherapy Drug | Minimum Breakthrough Detection Time<br>(Minutes) |
|-------------------------------|--------------------------------------------------|
| Cyclophosphamide (20.0 mg/ml) | >240 |
| Doxorubicin HCl (2.0 mg/ml) | >240 |
| Etoposide (20.0 mg/ml) | >240 |
| Fluorouracil (50.0 mg/ml) | >240 |
| Paclitaxel (6.0 mg/ml) | >240 |
| Cisplatin (1.0 mg/ml) | >240 |
{16}------------------------------------------------
| Dacarbazine (10.0 mg/ml) | >240 |
|--------------------------|------|
| Carmustine (3.3 mg/ml) | 15 |
| ThioTEPA (10.0 mg/ml) | 97 |
# 11.0 Summary of Clinical Testing
Clinical testing is not needed for this device.
# 12.0 Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the subject device NITRILE EXAMINATION GLOVES (Tested For Use With Chemotherapy Drugs) is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K220343.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.