K242809 · Medical Graphics Corporation · BTY · Oct 17, 2024 · Anesthesiology
Device Facts
Record ID
K242809
Device Name
Ascent Cardiorespiratory Diagnostic Software
Applicant
Medical Graphics Corporation
Product Code
BTY · Anesthesiology
Decision Date
Oct 17, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1890
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric, 3rd-Party Reviewed
Indications for Use
Ascent Cardiorespiratory Diagnostic Software is intended to be used for measurements, data collection and analysis of lung function (PFT) parameters, and cardiopulmonary testing (CPET) parameters, aiding in the diagnosis of related conditions. All the measurements are performed via a mouthpiece or a mask. The results of the test can be viewed on-line with the help of a computer screen and can be printed after the test. The test results can be saved for further referral or report generation purposes. For use of the Bronchial Challenge option, the medical director of the laboratory, physician, or person appropriately trained to treat acute bronchoconstriction, including appropriate use of resuscitation equipment, must be close enough to respond quickly to an emergency. The product can be utilized for patients from 4 years old and older as long as they can cooperate in the performance -- no special limit to patient's sex or height. Measurements will be performed under the direction of a physician in a hospital environment, physician's office or similar settings.
Device Story
Ascent is a stand-alone software application used with MGC diagnostic hardware to collect, analyze, and display cardiorespiratory data. Inputs include breath-by-breath gas samples (O2/CO2) via mouthpiece/mask, spirometry maneuvers, and peripheral data (e.g., pulse oximetry via Bluetooth). The software calculates lung volumes, diffusing capacity, VO2, VCO2, and cardiac output (via Fick method). Used in hospitals/clinics by trained professionals under physician oversight. The device dashboard displays hardware status, calibration, and test progress. Output is presented on a computer screen for physician interpretation to aid in diagnosing pulmonary/cardiac conditions. Benefits include standardized diagnostic reporting and support for exercise testing (CPET, HAST, 6MWT) to assess functional capacity and treatment response.
Clinical Evidence
Bench testing only. Validation included system-level performance testing (FEV1, MVV, FRC, SVC, DLCO, VA, TGV, VO2, VCO2, VE) and safety/hazard analysis. Software verified against ISO 14971 and IEC 62304 standards. Performance evaluated against ATS/ERS/ACCP guidelines for spirometry, lung volumes, DLCO, bronchial challenge, and CPET. No clinical trial data required.
Technological Characteristics
Stand-alone Windows desktop software. Interfaces with USB-connected diagnostic hardware (spirometers, plethysmographs, metabolic carts). Uses Nafion tubes for gas drying. Connectivity includes USB, Bluetooth (for pulse oximetry), and networked database/printing. Calibration via 3L syringe, integrated gas calibration, and deadweight/water manometers. Software lifecycle per IEC 62304.
Indications for Use
Indicated for patients 4 years and older capable of cooperating with testing. Used for measurement, data collection, and analysis of pulmonary function (PFT) and cardiopulmonary exercise (CPET) parameters to aid in diagnosing conditions like COPD, asthma, exercise intolerance, and heart failure. Requires physician direction.
Regulatory Classification
Identification
A predictive pulmonary-function value calculator is a device used to calculate normal pulmonary-function values based on empirical equations.
{0}------------------------------------------------
Build Correspondence
Image /page/0/Picture/2 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 17, 2024
Medical Graphics Corporation % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K242809
Trade/Device Name: Ascent Cardiorespiratory Diagnostic Software Regulation Number: 21 CFR 868.1890 Regulation Name: Predictive Pulmonary-Function Value Calculator Regulatory Class: Class II Product Code: BTY, BZC Dated: September 17, 2024 Received: September 17, 2024
Dear Prithul Bom:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page
2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Binoy J.
Mathews -S
Digitally signed by Binoy J.
Mathews -S
Date: 2024.10.17 16:18:59
-04'00'
For
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K242809
Device Name Ascent Cardiorespiratory Diagnostic Software
#### Indications for Use (Describe)
Ascent Cardiorespiratory Diagnostic Software is intended to be used for measurements, data collection and analysis of lung function (PFT) parameters, and cardiopulmonary testing (CPET) parameters, aiding in the diagnosis of related conditions.
All the measurements are performed via a mouthpiece or a mask.
The results of the test can be viewed on-line with the help of a computer screen and can be printed after the test. The test results can be saved for further referral or report generation purposes.
For use of the Bronchial Challenge option, the medical director of the laboratory, physician, or person appropriately trained to treat acute bronchoconstriction, including appropriate use of resuscitation equipment, must be close enough to respond quickly to an emergency.
The product can be utilized for patients from 4 years old and older as long as they can cooperate in the performance -- no special limit to patient's sex or height.
Measurements will be performed under the direction of a physician in a hospital environment, physician's office or similar settings.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size: 10pt;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span>□</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for MGC Diagnostics. The logo consists of a gray circle with three vertical lines inside, resembling a pause symbol. Below the symbol, the text "MGC DIAGNOSTICS" is displayed, with "MGC" in yellow and "DIAGNOSTICS" in gray, followed by the registered trademark symbol.
MGC Diagnostics Corporation through its subsidiary Medical Graphics Corporation 350 Oak Grove Parkway St. Paul, Minnesota USA, 55127 -8599
# 510(k) Summary
October 17, 2024
#### Submitted by:
MGC Diagnostics Corporation Through its subsidiary Medical Graphics Corporation 350 Oak Grove Parkway St. Paul, MN 55127-8599 USA 651-484-4874
#### Contact Person:
Amy Fowler Regulatory Counsel for MGC Diagnostics Pathmaker FDA Law, P.L.L.C. 1415 Lilac Drive N Minneapolis, MN 55422 Phone: (612) 356-9653 Email: fowler@pathmakerlaw.com
Proprietary or Trade Name: Ascent™ Cardiorespiratory Diagnostic Software
#### Available Model Number: v2.1
Common Name: Cardiorespiratory Diagnostic Software
#### Classification Names:
Calculator, Predicted Values, Pulmonary Function, 21 CFR 868.1890, BTY Pulmonary Function Data Calculator, 21 CFR 868.1880, BZC
Predicate Devices: K181912, Ascent™ Cardiorespiratory Diagnostic Software K133925, Vyntus SentrySuite Product Line
#### Intended Use/Indications for Use:
Ascent Cardiorespiratory Diagnostic Software is intended to be used for measurements, data collection and analysis of lung function (PFT) parameters, and cardiopulmonary testing (CPET) parameters, aiding in the diagnosis of related conditions.
All the measurements are performed via a mouthpiece or a mask.
{5}------------------------------------------------
The results of the test can be viewed on-line with the help of a computer screen and can be printed after the test. The test results can be saved for further referral or report generation purposes.
For use of the Bronchial Challenge option, the medical director of the laboratory, physician, or person appropriately trained to treat acute bronchoconstriction, including appropriate use of resuscitation equipment, must be close enough to respond quickly to an emergency.
The product can be utilized for patients from 4 vears old and older as long as they can cooperate in the performance -- no special limit to patient's sex or height.
Measurements will be performed under the direction of a physician in a hospital environment, physician's office or similar settings.
#### Device Description:
Ascent™ Cardiorespiratory Diagnostic Software ("Ascent") is a stand-alone software application which can be used with several hardware devices in the Medical Graphics Corporation product line.
The core purpose of the software for measurement, data collection and analysis of testing in patients who may be suffering from pulmonary illnesses like chronic obstructive pulmonary disease (COPD), asthma, exercise intolerance, heart failure and/or cardiorespiratory concerns where diagnosis and prognosis needs to be determined.
In conjunction with diagnostic hardware, Ascent is used to collect data pertaining to the patient's degree of obstruction, lung volumes, and diffusing capacity. It is also used to present the collected lung diagnostic information so that it can be checked for quality and interpreted by a qualified physician, often a pulmonologist or cardiologist.
All the measurements are performed via a mouthpiece or a face mask.
#### Pulmonary Function Tests
The following pulmonary function tests will be supported:
- 1) Forced Vital Capacity (FVC)
- Slow Vital Capacity (SVC) 2)
- 3) Maximum Voluntary Ventilation (MVV)
- 4) Plethysmography (Pleth)
- 5) Functional Residual Capacity (FRC)
- FRC Nitrogen Washout vs FRC Helium Rebreathing a)
- Diffusing Capacity (DLCO), Single breath Real Time Diffusion and single-breath gas ୧) chromatography
- 7) Maximum Inspiratory/Expiratory Pressure (MIP/MEP)
- 8) Six Minute Walk (6MWT) – software supports recording of results only, does not conduct test
- 9) Challenge Testing (Bronchial Provocation (BRP))
{6}------------------------------------------------
### Cardiopulmonary Exercise Tests (CPET)
For CPET testing, the software will be testing the patient's metabolic response while exercising or resting by collecting breath data. The gas samples are dried by Nafion tubes and analyzed for O2 and CO2 content. From the breathing volume and the differences between inspiratory and expiratory O2 and CO2 concentrations, the oxygen uptake and the CO2 production (VO2 and VCO2) are calculated by the software.
The following cardiopulmonary exercise tests will be supported by the Ascent software:
- 1) Breath by breath Cardiopulmonary Exercise Test (CPET
- a. Test can utilize Bike, Ergometer or other exercise device (i.e., Rowing Machine)
- b. Measuring high/low FiO2, end tidal, respiratory drive
- 2) Indirect Calorimetry
- a. Resting Test
- b. Face Tent Test
- 3) High/Low FiO2 (HAST)
- 4) Exercise Flow Volume Loops
- 5) Cardiac Output
- 6) Six Minute Walk (6MWT)
- a. The software supports a 6MWT by accepting test result data via Bluetooth functionality from a Nonin WristOx 3150 Pulse Oximeter.
#### Hardware interaction
An MGCD hardware device can be connected to the computer running Ascent™ Cardiorespiratory Diagnostic Software via a USB port. Communication with hardware can be managed by the application. The software can start or stop a patient test that is administered by the hardware device. The hardware device will output test results into the software where they can be managed and displayed to the user.
Ascent™ Cardiorespiratory Diagnostic Software will display a "device dashboard" on the home screen that will allow the user to view the status of any connected MGC device. The dashboard indicates the status of the selected device, the calibration status of the device, and the current device warm-up time. The Dashboard also allows the user to start/stop the vacuum pump if the device is so-equipped.
#### User Information
The software is intended to be used by trained professionals within the intended environments. And to present the collected lung diagnostic information to a qualified physician, (usually a pulmonologist or cardiologist), fordiagnostic interpretation. Users of the software are expected to receive training on how to work with the software prior to using the software for diagnostic purposes.
#### Intended Operational Environment
The software is intended to be used in hospitals, clinics and other pulmonary function and cardiorespiratory diagnostic environments. The software can also be used in any environment that has qualified physician oversight.
The software is intended to be installed as a desktop application on a Microsoft Windows based personal computer.
{7}------------------------------------------------
#### Supported Devices
The software is compatible with the following devices, which have been independently cleared:
| Device | Manufacturer | Device Class | Product<br>Code | FDA 510(k)<br>Number |
|------------------------------------------------------------|---------------------------------|--------------|-----------------|----------------------|
| Platinum Elite™ Series body<br>plethysmograph (RTD and GC) | Medical Graphics<br>Corporation | 2 | CCM | K943259 |
| CPFS/D USB™ spirometer | Medical Graphics<br>Corporation | 2 | BTY | K861486 |
| Ultima Series™ cardiorespiratory<br>diagnostic systems | Medical Graphics<br>Corporation | 2 | BZL | K061731 |
| Pulse Oximeter | Nonin | 2 | DQA | K080255 |
| Pulse Oximeter | Nonin | 2 | DQA | K040589 |
| Pulse Oximeter | Nonin | 2 | DQA | K102350 |
| NIBP | Suntech Medical | 2 | DXN | K122401 |
| Treadmill | Full Vision | 1 | IOL | N/A1 |
| Treadmill | Woodway | 1 | IOL | N/A1 |
| Treadmill | Medisoft RAM<br>Italia | 1 | IOL | N/A1 |
| Treadmill | Lode | 1 | IOL | N/A1 |
| Treadmill | H/P/Cosmos | 1 | IOL | N/A1 |
| Treadmill | Ergoline | 1 | IOL | N/A1 |
| Ergometer | Lode | 2 | ISD | K851097 |
| Ergometer | Lode | 2 | DXN | K060820 |
| Ergometer | Lode | 2 | MUD | K060820 |
| Ergometer | Ergoline | 2 | ISD | 510(k) Exempt |
| ECG | Philips Healthcare | 2 | DQK | K121638 |
| ECG | Baxter Healthcare | 2 | DPS | K152944 |
| Heart Rate | Medtronic<br>(formally Zephyr) | 2 | MHX, DRX | K113045 |
² – Noted Class 1 devices are exempted from premarket notification requirements.
#### Summary of Intended Use/Indications for Use of Device Compared to Predicate Device:
The intended use/indications for use statement for the subject device combines the intended use statements of both predicates. The addition of cardiopulmonary (CPET) testing is supported by the second predicate device. The intended patient population and intended user is the subject and predicate devices.
{8}------------------------------------------------
## Summary of Technological Characteristics of Device Compared to Predicate Device:
## Technological Characteristics Table
| | Subject Device<br>Ascent Cardiorespiratory Diagnostic<br>Software | Predicate Device<br>Ascent Cardiorespiratory<br>Diagnostic Software | Predicate Device<br>Vyntus/<br>SentrySuite Product Line | Comparison |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|
| 510(k) Number | K242809 | K181912 | K133925 | |
| Decision Date | TBD | April 12, 2019 | August 22, 2014 | |
| Manufacturer | Medical Graphics Corporation | Medical Graphics Corporation | CareFusion Germany | |
| Classification | Class II | Class II | Class II | Identical |
| Classification Name | 1) Calculator, Predicted Values,<br>Pulmonary Function; BTY<br>2) Pulmonary function data calculator,<br>BZC | 1) Calculator, Predicted Values,<br>Pulmonary Function; BTY<br>2) Pulmonary function data calculator,<br>BZC | Pulmonary function data<br>calculator, BZC | All use Pulmonary function<br>data calculator, BZC |
| Regulations | 868.1880<br>868.1890 | 868.1880<br>868.1890 | 868.1880 | Similar |
| Product Codes | BTY, BZC | BTY, BZC | BZC | Similar |
| Prescription Use? | Yes | Yes | Yes | Same |
| Intended Use<br>Statement | Ascent Cardiorespiratory Diagnostic<br>Software is intended to be used for<br>measurements, data collection and analysis<br>of lung function (PFT) parameters and<br>cardiopulmonary testing (CPET)<br>parameters, aiding in the diagnosis of<br>related conditions.<br>All the measurements are performed via a<br>mouthpiece or a mask.<br>The results of the test can be viewed on-<br>line with the help of a computer screen<br>and can be printed after the test. The test<br>results can be saved for further referral or<br>report generation purposes. | Ascent Cardiorespiratory Diagnostic<br>Software is intended to be used for<br>measurements, data collection and<br>analysis of lung function (PFT)<br>parameters, aiding in the diagnosis of<br>related conditions.<br>All the measurements are performed via a<br>mouthpiece or a mask. The results of the<br>test can be viewed on-line with the help of<br>a computer screen and can be printed<br>after the test. The test results can be saved<br>for further referral or report generation<br>purposes.<br>For use of the Bronchial Challenge option, | The Vyntus/SentrySuite<br>product line is intended to be<br>used for measurements, data<br>collection and analysis of lung<br>function (PFT) parameters and<br>cardio-pulmonary (CPET)<br>parameters, aiding in the<br>diagnosis of related conditions.<br>The results of the test can be<br>viewed online with the help of<br>a computer screen and can be<br>printed after the test. The test<br>results can be saved for future<br>reference or report generation | Same |
| | For use of the Bronchial Challenge option,<br>the medical director of the laboratory,<br>physician, or person appropriately trained<br>to treat acute bronchoconstriction,<br>including appropriate use of resuscitation<br>equipment, must be close enough to<br>respond quickly to an emergency.<br><br>The product can be utilized for patients<br>from 4 years old and older as long as they<br>can cooperate in the performance -- no<br>special limit to patient's sex or height.<br><br>Measurements will be performed<br>under the direction of a physician in a<br>hospital environment, physician's<br>office or similar settings. | physician, or person appropriately trained<br>to treat acute bronchoconstriction,<br>including appropriate use of resuscitation<br>equipment, must be close enough to<br>respond quickly to an emergency.<br><br>The product can be utilized for patients<br>from 4 years old and older as long as they<br>can cooperate in the performance<br>-- no special limit to patient's sex or<br>height.<br><br>Measurements will be performed<br>under the direction of a physician in a<br>hospital environment, physician's<br>office or similar settings. | purposes.<br><br>The products can be utilized<br>with patients age 4 years and<br>older as long as they can<br>cooperate in the performance<br>– no special limit to patient's<br>sex or height exists.<br><br>Measurements will be<br>performed under the direction<br>of a physician in a hospital<br>environment, physician's office<br>or similar setting (professional<br>healthcare facilities). | |
| Associated<br>Hardware | • Desktop or Notebook Computer<br>• Trolley or Desk<br>• Measurement Devices<br>- CPFS/D USB Spirometer<br>- Platinum Elite Series Body<br>- Plethysmograph<br>- Ultima Series<br>- Other optional items | • Desktop or Notebook Computer<br>• Trolley or Desk<br>• Measurement Devices<br>- CPFS/D USB Spirometer<br>- Platinum Elite Series Body<br>- Plethysmograph | • Desktop or Notebook<br>Computer<br>• Cart<br>• Measurement Devices<br>- Vyntus CPX unit<br>- Vyntus table top<br>power module<br>- Digital Volume<br>Transducer<br>- Ambient module<br>- USB cable<br>- Other optional items | Similar |
| Calibration Method | • Flow 3L Syringe<br>• O2/CO2 – integrated gas calibration<br>• Real Time DLCO - integrated gas<br>calibration<br>•Plethysmography – integrated<br>Deadweight and 50cc Cal Pump<br>•MIP/MEP – Integrated Deadweight<br>or water manometer | • Flow 3L Syringe<br>• O2/CO2 – integrated gas calibration<br>• Real Time DLCO – integrated gas<br>calibration<br>•Plethysmography - integrated<br>Deadweight and 50cc Cal Pump<br>•MIP/MEP – Integrated Deadweight | N/A | Similar |
| Bronchial Test | • Bronchial Test<br>O Bronchospasmolysis<br>O Bronchoprovocation | • Bronchial Test<br>O Bronchospasmolysis<br>O Bronchoprovocation | N/A | Identical to K181912 |
| Spirometry | • Spirometry (Slow and Forced) | • Spirometry (Slow and Forced) | • Spirometry (Slow and Forced) | Identical to K181912 |
| | | | | |
| | • Incentive Spirometry<br>• Flow/ Volume<br>• MVV | • Incentive Spirometry<br>• Flow/ Volume<br>• MVV | • MVV | |
| FRC Nitrogen<br>Washout | • FRC Nitrogen Washout | • FRC Nitrogen Washout | N/A | Identical to K181912 |
| Real Time Single<br>Breath Diffusion | • Real Time Single Breath Diffusion | • Real Time Single Breath Diffusion | N/A | Identical to K181912 |
| Diffusion Single<br>Breath CO/He | • Diffusion Single Breath CO/He | • Diffusion Single Breath CO/He | N/A | Identical to K181912 |
| Body<br>Plethysmography | • Body Plethysmography | • Body Plethysmography | N/A | Identical to K181912 |
| Respiratory Drive | • PI/PE Max (MIP/MEP) | • PI/PE Max (MIP/MEP) | N/A | Identical to K181912 |
| Breath-by-Breath | Breath-by-Breath | N/A | Breath-by-Breath | Identical to K133925 |
| Indirect Calorimetry | Indirect Calorimetry | N/A | Indirect Calorimetry | Identical to K133925 |
| Six Minute Walk<br>(6MWT) | Six Minute Walk (6MWT) | Six Minute Walk (6MWT) | N/A | Similar to K181912 |
| High Altitude<br>Simulation Test | High/Low FiO2 (HAST) | N/A | Breath-by-Breath | Similar to K133925 |
| Exercise Flow<br>Volume Loops | Exercise Flow Volume Loops | Spirometry | Breath-by-Breath | Similar to K133925 and<br>K181912 |
| Cardiac Output | Calculated value, derived from VO2 and<br>blood gases | N/A | Calculated value, derived<br>from VO2 and blood gases | Similar to K133925 |
| Software | | | | |
| Software device<br>only | Yes | Yes | No, this device has<br>software and hardware | Identical to K181912 |
| Software Network<br>Options | • Use as Workstation<br>• Data integration<br>• Database handling<br>• Networked database<br>• Offline mode<br>• Network printing | • Use as Workstation<br>• Data integration<br>• Database handling<br>• Networked database<br>• Offline mode<br>• Network printing | • Use as Workstation<br>• Use as Server<br>• Online Connection<br>• Vlink connection<br>• GTD connection<br>• Data integration<br>• Database handling<br>• JINET server | Identical to K181912 |
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
The subject device and the predicate devices are all software programs that are used with a variety of hardware systems to collect, display, manage and store patient data from pulmonary function tests and cardiorespiratory tests. Most of the test functionality is the same. There are just a few tests where there are minor differences between the subject device and one of the predicate devices:
#### Six Minute Walk Test (6MWT) 1.
The subject device performs an additional test which is not performed by the K133925 predicate but is similar to K181912. The 6MWT is typically performed to assess response to a medical or surgical intervention but may also be used to assess functional capacity of the subject and to help titrate oxygen as the patient walks. The object of the test is to have the subject walk as far as possible in 6 minutes. K181912 implements the 6MWT as simple lap counter allowing the medical professional to determine the total distance traveled. The subject device support input of pulse oximetry and heart rate data from an independent medical device into the test for real-time observation and subsequent reporting.
### 2. High Altitude Simulation Test (HAST)
HAST is a method of assessing a patient's ability to be exposed to conditions with a reduced partial pressure of oxygen, such as air travel or a high-altitude environment. A medical professional varies the inspired O2 concentration for subjects either at rest or during exercise. The mechanism which delivers this altered gas concentration to the patient is independent of the subject device. Measurement is accomplished via a breath-by-breath method, similar to K133925 in which both inspiratory and expiratory concentrations of oxygen (O2) and CO2 are measured.
#### 3. Exercise Flow Volume Loops (ExFVL)
ExFVL integrates a flow volume loop during an exercise gas exchange test. Both predicate devices support spirometry and K133925 supports breath-by-breath exercise testing. Subject device allows for a single maximal inhalation maneuver to be executed during the breath-by-breath exercise test. The resulting exercise flow volume loop can be overlayed over a flow volume loop collected during spirometry or the resting phase of the CPET test.
#### 4. Breath by Breath
Breath by breath is a comprehensive evaluation of the cardiovascular, pulmonary, and muscular systems during physical activity. The test typically involves exercising on a treadmill or stationary bike while breathing through a mouthpiece connected to a metabolic cart. Key measurements involve oxygen uptake (VO2), carbon dioxide production (VCO2), and ventilation. Subject device is similar to K133925 in the measurements of VO2, VCO2 ,and VE, however subject device includes additional derived values which are calculated values from literature.
#### 5. Indirect calorimetry
An indirect calorimetry test measures the amount of oxygen consumed and carbon dioxide produced by the body to estimate resting energy expenditure (REE). Subject device is equivalent in performance to predicate K133925.
### 6. Cardiac output
Cardiac output is a calculated parameter using the VO2 measurement by the subject device and blood gases which are manually entered by the clinician. Subject device measures VO2 in a substantially equivalent manner as K133925. This is a calculated parameter and well-established calculation from literature via FICK method.
{12}------------------------------------------------
#### Determination of Substantial Equivalence Based on Nonclinical Tests
Ascent Cardiorespiratory Diagnostic Software was extensively validated per medical device software standards and guidance. Testing results support that Ascent fulfills its intended use/indications for use and substantially equivalent to the predicate device.
#### Non-clinical Performance:
Software validation testing involved system level tests, performance tests and safety testing based on hazard analysis. Performance validation testing was done with the subject software device and recommended hardware devices working…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.