K242704 · Synapse Biomedical, Inc. · GZI · Jun 5, 2025 · Neurology
Device Facts
Record ID
K242704
Device Name
Synchrony (20-3000)
Applicant
Synapse Biomedical, Inc.
Product Code
GZI · Neurology
Decision Date
Jun 5, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Synchrony is intended for use in adult patients (22 years and older) whose paralysis or paresis is caused by upper motor neuron injury (e.g., stroke, spinal cord injury, etc.) by prescription use only.
Device Story
Synchrony is a functional electrical stimulation (FES) and powered muscle stimulator system for hand paresis. Components include a stimulator, sensor, electrodes, motor point probe, and iOS clinician app. Therapies include Contralaterally Controlled Electrical Stimulation (CCFES), cyclical NMES, and Functional Task Practice (FTP). CCFES uses sensor data from the unimpaired hand to synchronize stimulation with natural motor patterns. The sensor uses a Class I laser to measure hand opening/closure. Clinicians program therapies via the app; patients perform sessions at home. Stimulation is delivered via up to four electrodes using balanced biphasic pulses. The motor point probe assists clinicians in identifying optimal electrode placement. The system provides personalized therapeutic interventions to improve hand functionality and muscle re-education.
Clinical Evidence
No clinical data were reviewed in this submission. Evidence is based on bench testing, including electrical safety, electromagnetic compatibility, usability, biocompatibility, and wireless coexistence.
Technological Characteristics
System includes stimulator, sensor, and motor point probe. Enclosure: ABS Cycoloy, Polycarbonate. Stimulation: Balanced biphasic, rectangular waveform, 32-37 Hz, 1-250 μsec pulse width. Connectivity: Bluetooth Low Energy (BLE). Power: Wireless charging pad. Safety: IP22 ingress protection, automatic overload trip, automatic shut-off. Standards: IEC 60601-1, IEC 60601-2-10, ISO 10993-1, ASTM F899-20 (stainless steel probe).
Indications for Use
Indicated for Functional Electrical Stimulation (FES) to improve hand function and active range of motion in adults (22+) with hemiplegia due to stroke or upper limb paralysis due to C5 spinal cord injury. Also indicated for NeuroMuscular Electrical Stimulation (NMES) for maintenance/increase of hand range of motion, prevention/retardation of disuse atrophy, increase in local blood circulation, reduction of muscle spasm, and re-education of muscles.
Regulatory Classification
Identification
An external functional neuromuscular stimulator is an electrical stimulator that uses external electrodes for stimulating muscles in the leg and ankle of partially paralyzed patients (e.g., after stroke) to provide flexion of the foot and thus improve the patient's gait.
Predicate Devices
NESS H200 Wireless Rehabilitation System with optional Intelli-connect (K123636)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 5, 2025
Synapse Biomedical, Inc.
Jason Fiest
Vice President - Quality and Regulatory
300 Artino Street
Oberlin, Ohio 44074
Re: K242704
Trade/Device Name: Synchrony (20-3000)
Regulation Number: 21 CFR 882.5810
Regulation Name: External Functional Neuromuscular Stimulator
Regulatory Class: Class II
Product Code: GZI, IPF, KQX
Dated: May 6, 2025
Received: May 6, 2025
Dear Jason Fiest:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K242704 - Jason Fiest
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K242704 - Jason Fiest
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Lauren E. Woodard -S
for Amber Ballard, PhD
Assistant Director
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K242704 | |
| Device Name Synchrony® (20-3000) | |
| Indications for Use (Describe) The Synchrony (20-3000) is an electrical stimulation device indicated for the following uses: Functional Electrical Stimulation (FES). • Improvement of hand function and active range of motion in patients with hemiplegia due to stroke or upper limb paralysis due to C5 spinal cord injury. NeuroMuscular Electrical Stimulation (NMES). • Maintenance and/or increase of hand range of motion. • Prevention and/or retardation of disuse atrophy. • Increase in local blood circulation. • Reduction of muscle spasm. • Re-education of muscles. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Synapse Biomedical Incorporated
# Submitter Information:
Name: Jason Fiest
Address: 300 Artino Street, Oberlin, OH 44074
Contact person: Jason Fiest – Vice President – Quality and Regulatory
Telephone number: +1.440.774.2488 x138
Date the summary was prepared: 6/4/2025
Device Common Name: External neuromuscular stimulator
Device trade/proprietary name: Synchrony® 20-3000
Device Classification: Class 2
Product Code: GZI, IPF, and KQX
Regulation number: 882.5810 External functional neuromuscular stimulator
Predicate devices:
K123636 - NESS H200® Wireless Rehabilitation System with optional Intelli-connect
K960457- EMS-2C Med Labs Inc.
# Description of Subject Device:
The Synchrony (20-3000) system is a functional electrical stimulation (FES) device and a powered muscle stimulator. It is specifically designed to enhance hand functionality in patients with arm paresis through a range of customizable therapies. The system integrates five major components: the Stimulator, Sensor, Patient Interface (electrodes), Motor Point Probe, and Clinician App. The therapies are directed and programmed by a clinician through an iOS based application and then the patient may apply the therapy at their home with the Stimulator and Sensor.
The Synchrony system supports three primary therapeutic modes: Contralaterally Controlled Electrical Stimulation (CCFES), cyclical Neuromuscular Electrical Stimulation (CNMES), and Functional Task Practice (FTP). The CCFES mode uses sensor data from the patient's unimpaired side to control the timing, intensity, and movement of the paretic hand/arm, synchronizing stimulation with natural motor patterns. The cyclical NMES mode delivers repetitive stimulation to targeted muscles, aiding in muscle re-training, as described in the systematic review of research studies using cyclic NMES. The FTP mode allows for direct control of hand movements during task-oriented exercises, using real-time sensor feedback and stimulator output.
Stimulation is delivered through up to four electrodes placed on identified motor points, using symmetric balanced biphasic pulses. The system offers preset stimulation amplitudes of 20mA, 40mA, and 60mA, with a pulse width adjustable between 1-250 μsec, providing high precision with 1 μsec resolution. The device's enclosure is constructed from biocompatible ABS Cycoloy, ensuring durability and patient safety, and is rated IP22 for ingress protection.
Page 1 of 11
{5}
Synapse Biomedical Incorporated
The stimulator can operate independently or in conjunction with the Synchrony Sensor, depending on the therapy mode selected. The Sensor, which communicates wirelessly with the Stimulator via Bluetooth Low Energy (BLE), uses a Class I laser that is compliant with 21 CFR 1040.10 and 21 CFR 1040.11 to measure hand opening and closure distances, enabling accurate therapeutic adjustments. Both the Stimulator and Sensor are wirelessly charged using an FCC Part 15 compliant charging pad, enhancing ease of use and handling.
The Synchrony system is engineered for ease of use with user-friendly control interfaces for both patients and clinicians. It incorporates a range of safety mechanisms to ensure effective and safe operation, making it a versatile and reliable tool for improving hand functionality through personalized therapeutic interventions.
## Intended Use:
Synchrony is intended for use in adult patients (22 years and older) whose paralysis or paresis is caused by upper motor neuron injury (e.g., stroke, spinal cord injury, etc.) by prescription use only.
## Indications for Use:
The Synchrony System is an electrical stimulation device indicated for the following uses:
Functional Electrical Stimulation (FES).
- Improvement of hand function and active range of motion in patients with hemiplegia due to stroke or upper limb paralysis due to C5 spinal cord injury.
NeuroMuscular Electrical Stimulation (NMES).
- Maintenance and/or increase of hand range of motion.
- Prevention and/or retardation of disuse atrophy.
- Increase in local blood circulation.
- Reduction of muscle spasm.
- Re-education of muscles.
Page 2 of 11
{6}
Synapse Biomedical Incorporated
# Substantial Equivalence:
The Synchrony system has been compared with the predicate device: the H200 Wireless Hand Rehabilitation System with Optional Intelli-Connect Earpiece Triggering Device (K123636) by Bioness, Inc.
The Indications for Use of the Synchrony system and Bioness H200 are the same. Both, the Synchrony and H200 are battery operated stimulators. The Synchrony device has four potential output electrodes and one common return electrode. The Synchrony uses commercially available Surface electrode (Axelgaard Ultrastim Electrodes Model Numbers SN2020 and SN2040; FDA submission K130987) that can each be placed on any of the desired muscles of the hand or forearm.
Both devices use a Clinician application to program settings into the device for individual use by patients. Both devices also have a wireless input device that can be used to control the stimulation. The H200 device uses the wireless "Intelli-Connect" trigger switch to trigger a "pre-determined stimulation sequence at the H200W orthosis". The Synchrony device uses the wireless Sensor, that is essentially an application specific hand opening goniometer, to output pre-determined stimulation at a level related to the hand opening of the unaffected hand.
Since the Synchrony device doesn't have fixed positioning of the surface electrodes, the device includes a motor point probe for the clinician to use to locate the optimal electrode positioning for the desired muscle recruitment. This probe is a spring loaded stainless steel electrode disk that shuts off stimulation before the disk loses contact with the skin, thus avoiding a high current density stimulation as the probe is lifted from the skin location.
The EMS-2C from Med Labs Inc. is also included as a predicate device to enhance the substantial equivalence comparison for the Motor Point Probe (MPP) accessory of the Synchrony system. Like Synchrony, the EMS-2C is a powered muscular stimulator with the same product code (IPF). It was selected for its probe-shaped stimulator, which resembles the design and use of the Motor Point Probe.
Despite some minor differences in the technological variations between the Synchrony system and its predicate devices they do not raise new questions regarding safety and effectiveness. The Synchrony system complies with the IEC 60601-2-10 standard for basic safety and essential performance of nerve and muscle stimulators. The evaluation of the Synchrony system demonstrates that it is substantially equivalent to the predicate devices in terms of safety, effectiveness, and overall performance.
Page 3 of 11
{7}
Synapse Biomedical Incorporated
Summary table of technical characteristics vs. predicate devices:
| | Synchrony New Device | Bioness H200 Predicate | Med Labs EMS-2C Predicate^{5} | Difference / Rationale |
| --- | --- | --- | --- | --- |
| 1. 510(k) Number | K242704 | K123636 | K960457 | N/A |
| 2. Device Name, Model | Synchrony, 20-3000 | NESS H200 Wireless Hand Rehabilitation System | EMS-2C | N/A |
| 3. Manufacturer | Synapse Biomedical Inc. | Bioness Inc. | Med Labs, Inc. | N/A |
| 4. Classification Regulation(s) | 21 CFR 882.5810 | 21 CFR 882.5810 | 21 CFR 890.5850 | Same. No impact on Safety and Effectiveness |
| 5. Product Code | GZI: External Functional Neuromuscular Stimulator
IPF: Powered Muscle Stimulator
KQX: Goniometer
AC-Powered | GZI: External Functional Neuromuscular Stimulator
IPF: Powered Muscle Stimulator | IPF: Powered Muscle Stimulator | Same. No impact on Safety and Effectiveness |
| 6. Indications for Use: | The Synchrony System is an electrical stimulation device indicated for the following uses:
Functional Electrical Stimulation (FES).
o Improvement of hand function and active range of motion in patients with hemiplegia due to stroke or upper limb paralysis due to C5 spinal cord injury.
NeuroMuscular Electrical Stimulation (NMES). | The NESS H200® Wireless System is an electrical stimulation device indicated for the following uses:
Functional Electrical Stimulation (FES).
o Improvement of hand function and active range of motion in patients with hemiplegia due to stroke or upper limb paralysis due to C5 spinal cord injury.
NeuroMuscular Electrical Stimulation (NMES).
o Maintenance and/or | Not Publicly Available | |
Page 4 of 11
{8}
Synapse Biomedical Incorporated
| | o Maintenance and/or increase of hand range of motion.
o Prevention and/or retardation of disuse atrophy.
o Increase in local blood circulation.
o Reduction of muscle spasm.
o Re-education of muscles | increase of hand range of motion.
o Prevention and/or retardation of disuse atrophy.
o Increase in local blood circulation.
o Reduction of muscle spasm.
o Re-education of muscles | | |
| --- | --- | --- | --- | --- |
| 7. Power Source(s)
- Method of Line Current Isolation
- Patient Leakage Current
- Normal condition
- Single fault condition | Battery | Battery | Battery | Same. No impact on Safety and Effectiveness |
| 7. Number of Output Modes | 1 | - Not Publicly Available | 2 (EMS-2C has IDC (Interrupted Direct Current) and Pulse Mode) | Synchrony has one output mode, and the EMS can be toggled between two. No impact on safety or effectiveness. |
| 8. Number of Output Channels
- Synchronous or Alternating
- Method of Channel Isolation | 1 – 4
- Alternating
- Solid state relay | - Not Publicly Available
- Not Publicly Available
- Not Publicly Available | - Not Publicly Available
-Not Publicly Available
-Not Publicly Available | Synchrony has fewer output channels than the predicates. No impact on safety or effectiveness by having the same or fewer output channels. |
| 9. Regulated Current or Voltage | Regulated Current | Not Publicly Available | Not Publicly Available | Synchrony uses Regulated Current, Same as the predicate devices. No impact on safety or effectiveness |
| 10. Software / Firmware / Microprocessor Control | Yes | Not Publicly Available | Not Publicly Available | Same. No impact on safety or effectiveness |
Page 5 of 11
{9}
Synapse Biomedical Incorporated
| | Synchrony New Device | Bioness H200 Predicate | Med Labs EMS-2C Predicate^{5} | Difference / Rationale |
| --- | --- | --- | --- | --- |
| 11. Automatic Overload Trip | Yes | Not Publicly Available | Not Publicly Available | Same, Information for Predicate Devices is not publicly available |
| 12. Automatic No-Load Trip | No | Not Publicly Available | Not Publicly Available | Same. No impact on safety or effectiveness. |
| 13. Automatic Shut Off | Yes | Not Publicly Available | Not Publicly Available | Same. No impact on safety or effectiveness. |
| 14. Patient Override Control | Yes | Not Publicly Available | Not Publicly Available | Same. No impact on safety or effectiveness. |
| 15. Indicator Display
- On/Off Status
- Low Battery
- Voltage/Current Level | Yes
Yes
Yes | Not Publicly Available
Not Publicly Available
Not Publicly Available | Not Publicly Available
Not Publicly Available
Not Publicly Available | Synchrony App displays same or more data as the predicate devices. No impact on safety or effectiveness. |
| 16. Timer Range (minutes) | 1-51 minutes | N/A | N/A | N/A |
| 15. Compliance with 21 CFR 898 | Yes | N/A | N/A | Same. No impact on safety or effectiveness |
| 17. Weight | Stimulator: 2.2 Ounces
Sensor: 1.7 Ounces
Motor Point Probe: 2 Ounces | Not Publicly Available | Not Publicly Available | Same. No impact on safety or effectiveness. |
| 18. Dimensions (in. W x H x D) | Stimulator: Diameter = 2.8 inches, Depth = 0.9 inches
Sensor: Height = 3.8 inches, Width = 1.6 inches, Depth = 3.1 inches
Motor Point Probe: Diameter = 1 inch, Length = 6 inch | Not Publicly Available | Not Publicly Available | Same. No impact on Safety and Effectiveness |
Page 6 of 11
{10}
Synapse Biomedical Incorporated
| | Synchrony New Device | Bioness H200 Predicate | Med Labs EMS-2C Predicate^{5} | Difference / Rationale |
| --- | --- | --- | --- | --- |
| 19. Housing Materials and Construction | Synchrony Stimulator
Enclosure: Cycoloy-ABS, Polycarbonate Matte (overlay)
Sensor: Cycoloy-ABS, Polycarbonate Matte (overlay)
Motor Point Probe: Cycoloy-ABS | Not Publicly Available | Not Publicly Available. | Same. No impact on Safety and Effectiveness |
# Notes:
1. Power Source: For AC line-powered devices, specify line voltage and frequency, method of line current isolation, and measured patient leakage current; for battery powered devices, specify number, size and type of batteries.
2. Leakage Current: Patient leakage current should be measured under both normal conditions and under single fault conditions and should be no greater than 100 microamperes and 500 microamperes, respectively. The method of testing should be described, e.g., IEC 60601-1.
3. Output Modes: For devices with more than one output mode, the information in Section 1 and Section 3 should be completed for each output mode.
4. Output Channels: For devices with more than one output channel, describe whether the outputs are delivered in a synchronous and/or alternating fashion, and describe the method of achieving channel isolation.
5. The Med Labs EMS-2C predicate device is being compared to the Synchrony Motor Point Probe being used with the Synchrony Stimulator
Page 7 of 11
{11}
Synapse Biomedical Incorporated
Output Specifications
| Output Specifications | Synchrony New Device | Bioness H200 Predicate | Med Labs EMS-2C Predicate^{1} | Difference/Rationale |
| --- | --- | --- | --- | --- |
| Waveform | Balanced Biphasic | Not Publicly Available | Not Publicly Available | Same. No impact on safety and effectiveness. |
| Shape | Rectangular | Not Publicly Available | Not Publicly Available | Same. Synchrony uses a rectangular waveform shape, which is commonly used in surface neuromuscular stimulation. No impact on safety and effectiveness. |
| Maximum Output Voltage | 32V @ 500Ω
96V @ 2KΩ
96V @ 10KΩ
VRMS 4.4V @ 500Ω
VRMS 13.1V @ 2KΩ
VRMS 13.1V @ 10KΩ | Not Publicly Available | Not Publicly Available | Max output voltage at 500 Ohms, 2000 Ohms, and 10,000 Ohms is within the maximum output voltage range for the predicate devices. No impact on safety and effectiveness. |
| Maximum Output Current | 64mA @ 500Ω
48mA @ 2KΩ
9.6mA @ 10KΩ
IRMS 8.7mA @ 500Ω
IRMS 6.5mA @ 2KΩ
IRMS 1.3mA @ 10KΩ | Not Publicly Available | Not Publicly Available | Synchrony applied peak and RMS current are lower at nominal and maximum load compared to predicate
The maximum output current is lower for the EMS-2C device than Synchrony.
The Synchrony and predicate devices both provide effective muscle contractions over their range of current setting within the safe current density limits.
The Synchrony device disperses charge to a lower density and thus utilizes higher currents but lower current density within the safe current density limits. No impact on safety and effectiveness. |
| Pulse Width | 1 - 250 μsec | Not Publicly Available | Not Publicly Available | Synchrony pulse width is programmable in 1 usec increments, rather than pre-set selections on the predicate.
No impact on safety and effectiveness. |
| Frequency | 32-37 Hz | Not Publicly Available | Not Publicly Available | Synchrony frequency range falls within the predicate ranges
Synchrony uses a lower stimulating frequency than predicate. No impact on safety and effectiveness. |
| For interferential modes | N/A | N/A | N/A | N/A |
Page 8 of 11
{12}
Synapse Biomedical Incorporated
| Output Specifications | Synchrony New Device | Bioness H200 Predicate | Med Labs EMS-2C Predicate^{5} | Difference/Rationale |
| --- | --- | --- | --- | --- |
| - Beat Frequency (Hz) | | | | |
| For multiphasic waveforms
- Symmetrical phases
- Phase Duration | -N/A
-N/A | -N/A
-N/A | -N/A
-N/A | N/A |
| Net Charge (μC) | 0 μC @ 500Ω | Not Publicly Available | Not Publicly Available | Charged balanced biphasic waveform for Synchrony and a monophasic waveform for predicate. No impact on safety and effectiveness. |
| Maximum Phase Charge (μC) @ 500Ω | 22-1004 (2x2”):
16μC
22-3001 (2x4”):
16μC
23-3003 (MPP):
11μC | N/A | N/A | Information unavailable for predicate devices. |
| Maximum Current Density (mA/cm²) @ 500Ω | 22-1004 (2x2”):
0.34 mA/cm²
22-3001 (2x4”):
0.17 mA/cm²
23-3003 (MPP):
1.33 mA/cm² | N/A | Not Publicly Available | Synchrony has a lower maximum current density at 500Ω than EMS-2C given the larger probe tip diameter of the MPP. Both current densities comply with FDA expectations that maximum current density must be below 2 mA/cm² for stimulating electrodes under the IPF product code. No impact on safety and effectiveness. |
| Maximum Power Density (W/cm²) @ 500Ω | 22-1004 (2x2”):
0.0015 W/cm²
22-3001 (2x4”):
0.0007 W/cm²
23-3003 (MPP):
0.0040 W/cm² | N/A | Not Publicly Available | Product testing verified safety and effectiveness of power density. |
| Burst Mode
a. Pulses per burst
b. Bursts per second
c. Burst duration (sec) | a. N/A
b. N/A
c. N/A | a. N/A
b. N/A
c. N/A | Not Publicly Available | N/A (No Burst Mode) |
| ON Time (usec) | 0-500 μsec | Not Publicly Available | Not Publicly Available | ON time is based upon the pulse width. Synchrony has a programmable pulse width from 1-250 usec (X2) equals 500 usec. ON time is within the range of the predicate device for safety and effectiveness. No impact on safety and effectiveness. |
Page 9 of 11
{13}
Synapse Biomedical Incorporated
| Output Specifications | Synchrony New Device | Bioness H200 Predicate | Med Labs EMS-2C Predicate^{6} | Difference/Rationale |
| --- | --- | --- | --- | --- |
| OFF Time (usec) | 100 μsec | Not Publicly Available | Not Publicly Available | For a single pulse the OFF time (interpulse delay) is set at 100 usec. The set 100μs interpulse delay allows for propagation of an action potential prior to reversing the depolarizing potential with the recharge phase of the pulse. No impact on safety and effectiveness.^{7} |
The Med Labs EMS-2C predicate device is being compared to the Synchrony Motor Point Probe being used with the Synchrony Stimulator
Page 10 of 11
{14}
Synapse Biomedical Incorporated
Summary of Performance Testing:
| Standard | Edition / Year |
| --- | --- |
| IEC60601-1 Basic Safety & Essential Performance | 2020 |
| IEC60601-1-2 Electromagnetic Disturbances | 2020 |
| IEC60601-1-6 Usability | 2020 |
| IEC60601-1-11 Home Healthcare Environment | 2020 |
| IEC60601-2-10 Nerve and Muscle Stimulators | 2016 |
| IEC62304 Software Life Cycle Processes | 2015 |
| ISO14971 Risk Management | 2019 |
| ISO10993-1 Biological Evaluation | 2018 |
| ISO10993-5 Cytotoxicity | 2009 |
| ISO10993-10 Sensitization | 2021 |
| ISO10993-23 Irritation | 2021 |
| IEC60825-1 Safety of Laser Products | 2014 |
| IEC62366-1 Usability | 2020 |
| IEC 62133-2 Battery testing, Lithium systems | 2021 |
| ASTM D4169 Shipping Containers and System | 2022 |
| ANSI-AAMI ST98 Cleaning Process | 2022 |
| ANSI C63.27 Wireless Coexistence | 2017 |
| 47 CFR Part 15 Subpart B RF Devices | 2023 |
| AAMI TIR57 Cybersecurity Risk Mgt | 2019+(R2023) |
| AAMI TIR97 Cybersecurity Post Market Risk Mgt | 2019+(R2023) |
| ASTM F899-20 Standard Specification for Wrought Stainless Steels for Surgical Instruments | 2023 |
Clinical data: No clinical data were reviewed in this submission in support of the subject device.
Conclusions:
Based upon the comparisons with predicate devices and performance testing conducted, we believe the Synchrony device to be substantially equivalent and safe and effective for its intended use.
Page 12 of 12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.