Diagnostic ultrasound imaging and fluid flow analysis of the human body.
Device Story
Portable, software-controlled ultrasound system; acquires/displays real-time ultrasound data. Consists of wireless transducers (C35, L75, P25) communicating via direct Wi-Fi to NUSONO App on off-the-shelf iOS/Android mobile devices. App provides interface for mode/setting control, image acquisition, and storage. Features 128-channel beamformer and signal processing. Used by trained healthcare professionals in clinical settings (clinics, hospitals, point-of-care). Output viewed on mobile device screen; assists clinicians in diagnostic imaging and fluid flow analysis; facilitates rapid clinical decision-making at point-of-care.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing only, including measurement accuracy, system sensitivity, thermal/mechanical/electrical safety, biocompatibility, cleaning/disinfection, and software/acoustic output verification.
Technological Characteristics
Wireless handheld ultrasound scanner; 128-channel beamformer; Wi-Fi connectivity to mobile devices. Powered by rechargeable Li-ion battery. Models: C35 (convex), L75 (linear), P25 (phased array). Compliance with IEC 60601-1, IEC 60601-2-37, IEC 62304, ISO 10993, and ISO 14971. Software-based image processing.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis in fetal, abdominal, pediatric, urology, gynecology, lung, small organ (thyroid, prostate, scrotum, breast), musculoskeletal (superficial and conventional), peripheral vessel, carotid, and cardiac (adult and pediatric) applications. Intended for use by trained healthcare professionals in physician offices, clinics, hospitals, and clinical point-of-care settings. Contraindicated in environments with high intensity electromagnetic disturbances.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Leltek Ultrasound Imaging System (Model: LU700 Series) (K222365)
UP600 Diagnostic Doppler Ultrasound System (K121983)
Submission Summary (Full Text)
{0}------------------------------------------------
February 20, 2025
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and name on the right. The Department of Health & Human Services logo is a stylized depiction of a human figure, while the FDA acronym and name are written in blue, with the acronym in a square and the name in a sans-serif font.
Nurodata Inc. Hao-Ching Chiu Project Specialist 5F.-3, No. 6-2, Sec.2, Shengyi Rd. Hsinchu Science Park Zhubei. 302058 Taiwan
Re: K242681
Trade/Device Name: NUSONO Handheld Ultrasound Scanner (NUSONO-C35); NUSONO Handheld Ultrasound Scanner (NUSONO-L75); NUSONO Handheld Ultrasound Scanner (NUSONO-P25) Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: January 23, 2025 Received: January 23, 2025
Dear Hao-Ching Chiu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
### YANNA S. KANG -S
Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K242681
Device Name
NUSONO Handheld Ultrasound Scanner (NUSONO-C35); NUSONO Handheld Ultrasound Scanner (NUSONO-L75); NUSONO Handheld Ultrasound Scanner (NUSONO-P25)
ndications for Use (Describe)
The NUSONO Handheld Ultrasound Scanner is a portable and software-based ultrasound imaging system, indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications:
NUSONO-C35
Fetal, Abdominal, Pediatric, Urology, Gynecology, Lung
NUSONO-L75
Pediatric, Small Organ (Thyroid, Prostate, Scrotum, Breast), Musculoskeletal (Superficial and Conventional), Peripheral Vessel, Others (Carotid), Lung
NUSONO-P25
Fetal, Abdominal, Pediatric, Urology, Gynecology, Cardiac (adult and pediatric), Lung
The system provides diagnostic ultrasound imaging in B mode, M mode, Color Doppler mode, Power Doppler mode and combine mode (B+M. B+CD. B+PD), intended for use in environments where healthcare is provided by trained healthcare professionals. The environments where the system can be used include physician offices, clinics, hospitals, and clinical point of care for diagnosis of patients except environments where intensity of electromagnetic disturbances is high.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "NURODATA" in a stylized font. The first four letters, "NURO," are in blue, while the last four letters, "DATA," are in red. The font is rounded and sans-serif, giving the word a modern and playful appearance.
### 510(k) Summary
### 1. Submitter's Information:
| Submitter: | Hsien-chen Chiou<br>General Manager<br>Nurodata Inc. |
|----------------|--------------------------------------------------------------------------------------------------------------------|
| Address: | 5F.-3, No. 6-2, Sec. 2, Shengyi Rd.,<br>Hsinchu Science Park,<br>Zhubei City, Hsinchu County 302058 Taiwan, R.O.C. |
| Email: | service@nurodata.com |
| Tel: | (+886) 3 6588233 |
| Fax: | (+886) 3 6588232 |
| Website: | www.nurodata.com |
| Date prepared: | 09/06/2024 |
#### 2. Name of Device and Class:
| Device Name: | NUSONO Handheld Ultrasound Scanner (NUSONO-C35) |
|-----------------|-------------------------------------------------|
| | NUSONO Handheld Ultrasound Scanner (NUSONO-L75) |
| | NUSONO Handheld Ultrasound Scanner (NUSONO-P25) |
| Common Name: | Diagnostic Ultrasound System and Accessories |
| Classification: | Class II |
| Panel: | Radiology |
### Classification Name:
| 21 CFR Section | Classification Name | Product Code |
|----------------|------------------------------------------|--------------|
| 892.1550 | Ultrasonic Pulsed Doppler Imaging System | 90 IYN |
| 892.1560 | Ultrasonic Pulsed Echo Imaging System | 90 IYO |
| 892.1570 | Diagnostic Ultrasound Transducer | 90 ITX |
### 3. Predicate Device:
| | Device Name: | 510(k) Number: |
|--|--------------|----------------|
|--|--------------|----------------|
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for NuroData. The word "Nuro" is in blue, and the word "Data" is in red. The font is sans-serif and rounded.
| Primary predicate | Leltek Ultrasound Imaging System (Model:<br>LU700 Series) | K222365 |
|-------------------|-----------------------------------------------------------|---------|
| Reference | Clarius Ultrasound Scanner | K192107 |
| Reference | Leltek Ultrasound Imaging System | K191235 |
| Reference | UP600 Diagnostic Doppler Ultrasound<br>System | K121983 |
#### 4. Description of the Device:
The NUSONO Handheld Ultrasound Scanner is a portable, software-controlled, diagnostic ultrasound system used to acquire and display high-resolution, real-time ultrasound data through an off-the-self (OTS) iOS 14, 15 and Android 12 or higher versions based mobile device. The system consists of a series of wireless transducers employing Wi-Fi-based technology to communicate with the NUSONO App on traditional smartphone/tablet devices via direct Wi-Fi. The NUSONO App provides the interface for mode/setting control and image display, acquisition, and storage functions. The 128-channel beamformer and image signal processing technology maximize the utility of all imaging transducer elements to enhance the diagnostic utility and confidence provided by the system.
Image /page/5/Figure/4 description: The image shows a diagram of a transducer and a screenshot of the NUSONO app. The transducer diagram shows the different components of the device, including the transducer, ultrasound beams and imaging processing, energy management, Wi-Fi connection, and power supply. The NUSONO app screenshot shows an ultrasound image and the app's user interface, which includes controls for depth, contrast, TGC, and filter.
This transportable system is intended for use in environments where healthcare is provided by trained healthcare professionals.
Intended users will be qualified and appropriately trained healthcare professionals with ultrasound.
{6}------------------------------------------------
The NUSONO Handheld Ultrasound Scanner product include:
#### 1. Scanner models:
- NUSONO-C35;
- NUSONO-L75: ●
- NUSONO-P25; ●
#### 2. Components:
- Software: ●
- o NUSONO App for iOS 14, 15;
- NUSONO App for Android 12; o
- Charging dock ●
- USB Type-C charging cable ●
- Adaptor ●
#### 3. Accessories:
- Carrying bag
- Gel dispenser with cap ●
- Probe protecting cap of the corresponding probe ●
#### 5. Intended Use of the Device:
Diagnostic ultrasound imaging and fluid flow analysis of the human body.
#### Indication for Use:
The NUSONO Handheld Ultrasound Scanner is a portable and software-based ultrasound imaging system, indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications:
NUSONO-C35
Fetal, Abdominal, Pediatric, Urology, Gynecology, Lung.
NUSONO-L75
Pediatric, Small Organ (Thyroid, Prostate, Scrotum, Breast), Musculoskeletal (Superficial and Conventional), Peripheral Vessel, Others (Carotid), Lung NUSONO-P25
Fetal, Abdominal, Pediatric, Urology, Gynecology, Cardiac (adult and pediatric), Lung
The system provides diagnostic ultrasound imaging in B mode, M mode, Color Doppler mode, Power Doppler mode and combine mode (B+M, B+CD, B+PD), intended for use in environments where healthcare is provided by trained healthcare professionals. The environments where the system can be used include physician offices, clinics, hospitals, and clinical point of care for diagnosis of patients except environments where intensity of electromagnetic disturbances is high.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "NURODATA" in a stylized font. The first four letters, "NURO", are in blue, while the last four letters, "DATA", are in a shade of red. The letters are rounded and connected, giving the word a modern and fluid appearance.
#### Determination of SE:
The subject device, "NUSONO Handheld Ultrasound Scanner," is a Track 3 system that utilizes the same fundamental scientific technology as the primary predicate device, Leltek Ultrasound Imaging System (K222365), along with the reference devices: Clarius Ultrasound Scanner (K192107), Leftek Ultrasound Imaging System (K191235), and UP600 Diagnostic Doppler Ultrasound System (K121983).
The NUSONO Handheld Ultrasound Scanner shares identical indications for use with the primary predicate device and at least one model of the reference devices. A detailed comparison between the predicate device and the subject device is provided below:
| Items | Application device | Primary Predicate<br>device | Reference device | Reference device | Reference device |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510k number | Current submission | K222365 | K192107 | K191235 | K121983 |
| Device name | NUSONO Handheld<br>Ultrasound Scanner | Leltek Ultrasound<br>Imaging System<br>(Model: LU700<br>Series) | Clarius Ultrasound<br>Scanner | Leltek Ultrasound<br>Imaging System | UP600 Diagnostic<br>Doppler Ultrasound<br>System |
| Product code | IYN, IYO, ITX | IYN, IYO, ITX | IYN, IYO, ITX | IYN, IYO, ITX | IYN, IYO, ITX |
| Intended use | Diagnostic<br>ultrasound imaging<br>and fluid flow<br>analysis of the<br>human body. | Diagnostic<br>ultrasound imaging<br>or fluid flow<br>analysis of the<br>human body. | Diagnostic<br>ultrasound imaging<br>and fluid flow<br>analysis. | Diagnostic<br>ultrasound imaging<br>or fluid flow<br>analysis of the<br>human body. | The device is a<br>general-purpose<br>ultrasonic imaging<br>instrument intended<br>for use by a qualified<br>physician for<br>evaluation of<br>Abdomen, Cardiac,<br>Small Organ (breast,<br>testes, thyroid), heart |
| Items | Application device | Primary Predicate<br>device | Reference device | Reference device | Reference device |
| 510k number | Current submission | K222365 | K192107 | K191235 | K121983 |
| | | | | | soft<br>tissue, Peripheral<br>Vascular,<br>Musculo-skeletal<br>(Conventional and<br>Superficial),<br>Pediatric, Fetal,<br>Cephalic, Ob/Gyn. |
| Indications<br>for use | - Fetal<br>- Abdominal<br>- Pediatric<br>- Small organ<br>-<br>- Musculoskeletal<br>(conventional)<br>- Musculoskeletal<br>(superficial)<br>- Urology<br>- Gynecology<br>- Cardiac adult<br>- Cardiac pediatric | - Ophthalmic<br>- Fetal<br>- Abdominal<br>- Pediatric<br>- Small organ<br>- Neonatal cephalic<br>- Musculoskeletal<br>(conventional)<br>- Musculoskeletal<br>(superficial)<br>- Cardiac adult<br>- Cardiac pediatric | - Fetal<br>- Abdominal<br>- Intra-operative<br>- Pediatric<br>- Adult cephalic<br>- Musculoskeletal<br>(conventional)<br>- Urology<br>- Gynecology<br>- Cardiac adult<br>- Cardiac pediatric | - Abdominal<br>- Small organ<br>- Musculoskeletal<br>(conventional)<br>- Musculoskeletal<br>(superficial) | - Abdominal<br>- Neonatal cephalic<br>- Adult cephalic<br>- Cardiac adult<br>- Cardiac pediatric |
| Items | Application device | Primary Predicate<br>device | Reference device | Reference device | Reference device |
| 510k number | Current submission | K222365 | K192107 | K191235 | K121983 |
| | - Peripheral vessel<br>- Carotid<br>- Lung | - Peripheral vessel<br>- Carotid<br>- Pulmonary<br>- interventional<br>guidance<br>(includes free hand<br>needle/catheter) | - Peripheral vessel | - Peripheral vessel | -<br>-<br>- |
| Modes of<br>operation | - B mode<br>- M mode<br>- Color doppler (CD)<br>- Power doppler<br>(PD) | - B mode<br>- M mode<br>- Color doppler (CD)<br>- Power doppler<br>(PD)<br>- PWD<br>- Combined mode<br>(B+M, B+CD,<br>B+PWD) | - B mode<br>- M mode<br>- Color doppler (CD)<br>- Power doppler<br>(PD)<br>- PWD<br>- Combined mode<br>(B+M, B+CD,<br>B+PD, B+PWD) | - B mode<br>- M mode<br>- Color doppler (CD)<br>- Power doppler<br>(PD)<br>- PWD<br>- Combined mode<br>(B+M, B+CD,<br>B+PWD) | - B mode<br>- M mode<br>- Color doppler (CD)<br>- Power doppler<br>(PD)<br>- PWD<br>- CWD<br>- Combined mode<br>(B+M, B+CD,<br>B+PD, B+PWD,<br>B+CD+PD,<br>B+PD+PWD) |
| Wireless<br>capability | Wi-Fi | Wi-Fi | Wi-Fi<br>Bluetooth | Wi-Fi | N/A |
| Transducer<br>types | - Convex<br>(NUSONO-C35)<br>- Linear | - Linear | - Convex<br>(C3 HD) | - Linear | - |
| Items | Application device | Primary Predicate<br>device | Reference device | Reference device | Reference device |
| 510k number | Current submission | K222365 | K192107 | K191235 | K121983 |
| | (NUSONO-L75)<br>- Phased array<br>(NUSONO-P25) | (LU710L)<br>- Phased array<br>(LU710PA) | - | (LU700L) | - Phased array<br>(P42) |
| Power source | Rechargable battery<br>(Li-ion) | Rechargable battery<br>(Li-ion) | Rechargable battery<br>(Li-ion) | Rechargable battery<br>(Li-ion) | Wall powered |
| Display | iOS or Android<br>mobile device | iOS or Android<br>mobile device and<br>Windows | iOS or Android<br>mobile device | Android mobile<br>device | Color 15-inch LCD<br>display |
| 510(k) Track | Track 3 | Track 3 | Track 3 | Track 3 | Track 3 |
| Standards | - IEC 60601-1 (2020)<br>- IEC 60601-1-2<br>(2020)<br>- IEC 60601-1-6<br>(2020)<br>- IEC 60601-2-37<br>(2015)<br>- IEC 60601-4-2<br>(2016)<br>- UD 2-2004 (R2009)<br>- IEC 62133-2 (2017) | - ANSI/AAMI<br>ES60601-1 (2012)<br>- IEC 60601-1-2<br>(2014)<br>- IEC 60601-1-6<br>(2013)<br>- IEC 60601-2-37<br>(2015)<br>- UD 2-2004 (R2009)<br>- UD 3-2004 (R2009)<br>- IEC 62133 (2012) | - ANSI/AAMI<br>ES60601-1 (2012)<br>- IEC 60601-1-2<br>(2014)<br>- IEC 60601-1-6<br>(2013)<br>- IEC 60601-1-12<br>(2014)<br>- IEC 60601-2-37<br>(2015)<br>- UD 2-2004 (R2009)<br>- IEC 62133 (2012) | - ANSI/AAMI<br>ES60601-1 (2012)<br>- IEC 60601-1-2<br>(2014)<br>- IEC 60601-1-6<br>(2013)<br>- IEC 60601-2-37<br>(2015)<br>- UD 2-2004 (R2009)<br>- UD 3-2004 (R2009)<br>- IEC 62133 (2012) | - EC 60601-1<br>- EN 60601-1<br>- EN 60601-1-2<br>-<br>- EN 60601-2-37<br>- UD 2<br>- UD 3<br>- IEC 62133 (2012) |
| Items | Application device | Primary Predicate<br>device | Reference device | Reference device | Reference device |
| 510k number | Current submission | K222365 | K192107 | K191235 | K121983 |
| | - IEC 62366-1 (2020)<br>- ISO 10993-1 (2018)<br>- ISO 10993-5 (2009)<br>- ISO 10993-10 (2021)<br>- ISO 10993-23 (2021)<br>- IEC 62304 (2015)<br>- ISO 15223-1 (2021)<br>- ISO 14971 (2019)<br>- ISO 13485 (2016) | - IEC 62366 (2014)<br>- ISO 10993-1 (2009)<br>- ISO 10993-5 (2009)<br>- ISO 10993-10 (2010)<br>- IEC 62304 (2006)<br>- ISO 15223-1 (2016)<br>- ISO 14971 (2019)<br>- ISO 13485 (2016) | - ISO 10993-5 (2009)<br>- ISO 10993-10 (2010)<br>- ISO 10993-11 (2017)<br>- IEC 62304 (2006)<br>- ISO 15223-1 (2012)<br>- ISO 14971 (2007) | - IEC 62366 (2014)<br>- ISO 10993-1 (2009)<br>- ISO 10993-5 (2009)<br>- ISO 10993-10 (2010)<br>- IEC 62304 (2006)<br>- ISO 15223-1 (2016)<br>- ISO 14971 (2019)<br>- ISO 13485 (2016) | -<br>- ISO 10993<br>-<br>-<br>-<br>- |
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the word "NURODATA" in a stylized font. The first four letters, "NURO", are in blue, while the last four letters, "DATA", are in red. The letters are rounded and have a modern look. The logo is simple and eye-catching.
#### 6. Non-clinical Performance Test:
Non-clinical performance tests include measurement accuracy, system sensitivity, thermal, mechanical, electrical safety, patient-contact materials, cleaning and disinfection, software and acoustic output. Nonclinical performance tests show compliance to the following standards:
| Standard | Title of Standard |
|------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1 Edition 3.2 2020-08<br>CONSOLIDATED VERSION | Medical electrical equipment - Part 1: General<br>requirements for basic safety and essential<br>performance |
| IEC 60601-1-2 Edition 4.1<br>2020-09 CONSOLIDATED<br>VERSION<br>ANSI AAMI IEC 60601-1-2:2014<br>[Including AMD 1:2021] | Medical electrical equipment – Part 1–2: General<br>requirements for basic safety and essential<br>performance - Collateral Standard: Electromagnetic<br>disturbances - Requirements and tests |
| IEC 60601-1-6 Edition 3.2<br>2020-07 CONSOLIDATED<br>VERSION | Medical electrical equipment - Part 1-6: General<br>requirements for basic safety and essential<br>performance - Collateral standard: Usability |
| IEC 60601-2-37 Edition 2.1 2015 | Medical electrical equipment – Part 2–37: Particular<br>requirements for the basic safety and essential<br>performance of ultrasonic medical diagnostic and<br>monitoring equipment |
| IEC 62133-2 Edition1.0 2017-02 | Secondary cells and batteries containing alkaline or<br>other non-acid electrolytes - Safety requirements for<br>portable sealed secondary cells, and for batteries made<br>from them, for use in portable applications - Part 2:<br>Lithium systems |
| IEC 62304 Edition 1.1 2015-06<br>CONSOLIDATED VERSION | Medical Device Software - Software Life Cycle<br>Processes |
| IEC 62366-1 Edition 1.1 2020-06<br>CONSOLIDATED VERSION | Medical devices - Part 1: Application of usability<br>engineering to medical devices |
| ISO 10993-1 Fifth edition<br>2018-08 | Biological evaluation of medical devices - Part 1:<br>Evaluation and testing within a risk management<br>process |
| ISO 10993-5 Third edition<br>2009-06-01 | Biological evaluation of medical devices - Part 5:<br>Tests for In Vitro Cytotoxicity |
| ISO 10993-10 Fourth edition | Biological evaluation of medical devices - Part 10: |
| 2021-11 | Tests for Skin Sensitization |
| ISO 10993-23 First edition<br>2021-01 | Biological evaluation of medical devices - Part 23:<br>Tests for Irritation |
| ISO 11737-1 Third edition<br>2018-01 [Including AMD1:2021] | Sterilization of health care products - Microbiological<br>methods - Part 1: Determination of a population of<br>microorganisms on products |
| ISO 13485:2016 | Medical Devices - Quality management systems |
| ISO 14971 Third Edition | Medical devices - Applications of risk management to<br>medical devices |
| ISO 15223-1 Fourth edition<br>2021-07 | Medical Devices - Symbols to be used with<br>information to be supplied by the manufacturer - Part<br>1: General requirements |
| ISO 7000 Sixth edition 2019-07 | Graphical Symbols for Use on Equipment |
| IEC 60417:2002 DB | |
| ISO 7010 Third edition 2019-07 | Graphical symbols - Safety colors and safety signs -<br>Registered safety signs |
| 21 CFR 801.109 | Code of Federal Regulations Title 21 - FDA -<br>Prescription devices |
| 2002/96/EC (WEEE) | Waste Electrical and Electronic Equipment Directive |
| IEC TR 60601-4-2 Edition 1.0<br>2016-05 | Medical electrical equipment - Part 4-2: Guidance and<br>interpretation - Electromagnetic immunity:<br>performance of medical electrical equipment and<br>medical electrical systems |
| AAMI TIR69:2017/(R2020) | Technical Information Report Risk management of<br>radio-frequency wireless coexistence for medical<br>devices and systems |
| IEEE ANSI USEMCSC<br>C63.27-2021 | American National Standard for Evaluation of<br>Wireless Coexistence |
| ANSI/AAMI SW96:2023 | Standard For Medical Device Security - Security Risk<br>Management For Device Manufacturers |
| ANSI ISA 62443-4-1-2018 | Security for industrial automation and control systems<br>Part 4-1: Product security development life-cycle<br>requirements. |
| AAMI TIR 57:2016 | Principles for medical device security - Risk<br>management. |
| AAMI TIR 97:2019 | Principles For Medical Device Security - Postmarket<br>Risk Management For Device Manufacturers |
| UL ANSI 2900-2-1 First Edition<br>2017 | Standard for Safety, Software Cybersecurity for<br>Network-Connectable Products, Part 2-1: Particular<br>Requirements for Network Connectable Components<br>of Healthcare and Wellness Systems |
| NIST Cybersecurity Framework<br>(CSF) 2.0 | Ver. 2024/02/26 |
| NIST SP 800-53 Rev. 5, | Security and Privacy Controls for Information |
| IEC 81001-5-1 Edition 1.0<br>2021-12 | Health software and health IT systems safety,<br>effectiveness and security - Part 5-1: Security -<br>Activities in the product life cycle |
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the word "NURODATA" in a stylized font. The first four letters, "NURO", are in blue, while the last four letters, "DATA", are in red. The letters are rounded and have a playful appearance.
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the word "NuroDATA" in a stylized font. The "Nuro" part of the word is in blue, while the "DATA" part is in a shade of red. The letters are rounded and have a playful appearance. The overall design is simple and modern.
#### 7. Clinical Performance Test:
The NUSONO Handheld Ultrasound Scanner did not require clinical testing to establish substantial equivalence.
#### 8. Conclusion:
The NUSONO Handheld Ultrasound Scanner is substantially equivalent to the predicate device. The NUSONO Handheld Ultrasound Scanner functions in a manner similar to and is intended for the same use as the predicate device. Based on the predicate device comparison of indications for use, acoustic output and general safety and effectiveness information, as well as the non-clinical performance test results, it is concluded that this device is as safe and effective as the predicate device for its intended use and performance, and is substantially equivalent to the predicate device.
End of Summary
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.